K Number
K233325
Date Cleared
2024-02-21

(145 days)

Product Code
Regulation Number
886.5916
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzman's Nodular Degeneration), dystrophies (e.g. Cogan's dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring from infection or trauma.

The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management of ocular surface disease including dry eye (e.g. ocular manifestations of Graft-versus-Host disease, Sjogren's syndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical and thermal burns, radiation, filamentary keratitis), epidermal ocular disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the Visionary Optics Scleral Contact Lens may incidentally provide correction of refractive error in persons with myopia, astigmatism or presbyopia.

Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Device Description

The Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Visionary Optics Scleral Contact Lens is lathe cut from one of the following hydrophobic, FDA Group #3 fluoro-silicone acrylate materials: roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, or fluoroxyfocon A.

AI/ML Overview

The provided FDA 510(k) summary (K233325) for the Visionary Optics Scleral Contact Lens describes non-clinical and clinical studies conducted to support the substantial equivalence claim. However, it does not provide specific acceptance criteria or detailed study results in the format requested. The document primarily focuses on demonstrating equivalence to predicate devices rather than providing a standalone performance validation for a novel device with explicit performance metrics.

Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes for test sets, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies is not available within the provided text.

Here's a breakdown of what can be extracted and what is not available:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Not explicitly stated as numerical targets)Reported Device Performance
Non-clinical:
Ability to manufacture lenses from fluoroxyfocon A blanks to prescribed parameters (manufacturing verification)All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.
Bioburden (CFU per lens)Less than 100 CFU per lens.
Clinical:
Safety and effectiveness in managing specified ocular conditions (implied: no serious adverse reactions, stable or improved ocular condition and vision)No serious or significant adverse reactions reported. For all patients, the ocular condition(s) and vision remained stable or improved.

Explanation: The document does not present specific numerical acceptance criteria (e.g., "X% accuracy," "Y precision"). Instead, for non-clinical studies, it states that manufacturing tolerances were met and bioburden was "within the established acceptance criteria of less than 100 CFU per lens." For clinical studies, the "acceptance criteria" appear to be implied by the outcome: no serious adverse events and stable/improved patient conditions.


2. Sample size used for the test set and the data provenance

  • Test set sample size: 33 patients (63 total eyes).
  • Data provenance: Not explicitly stated, but the study involved "Nine (9) independent practitioners," suggesting a multi-center, prospective clinical evaluation. The document does not specify the country of origin of the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of experts: Nine (9) independent practitioners.
  • Qualifications of experts: Not explicitly stated other than "independent practitioners." Their specific specializations (e.g., ophthalmologists, optometrists, or subspecialties) or years of experience are not provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication method: Not described. The text states that "The practitioners reviewed all patients therapeutically managed... and reported the outcome of each patient." There is no mention of an adjudication process among these practitioners for establishing a unified ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC comparative effectiveness study: No, this document pertains to a medical device (scleral contact lens), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to AI assistance for human readers is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone performance: No, this device is a physical contact lens, not an algorithm. The clinical performance inherently involves human fitting and patient use.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of ground truth: The "ground truth" for the clinical study appears to be the observed clinical outcome reported by the independent practitioners, specifically: "no serious or significant adverse reactions reported" and "for all patients, the ocular condition(s) and vision remained stable or improved." This is a form of outcomes data based on clinical observation and assessment by practitioners.

8. The sample size for the training set

  • Training set sample size: Not applicable and not mentioned. This is a medical device, not a machine learning model. The "training" described in the context of this submission refers to the historical data and previous 510(k) clearances that establish the safety and effectiveness of the materials used in the lens.

9. How the ground truth for the training set was established

  • Ground truth establishment for training set: Not applicable and not mentioned. As this is not an AI/ML device, the concept of a "training set ground truth" as you've framed it is not relevant. The safety and effectiveness of the materials (roflufocon D, roflufocon E, etc.) used in the device were established through "reference to previous 510(k) clearances" and "previous 510(k) clearances," implying that historical data and prior regulatory assessments for these materials served as the basis for their use in the current device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 21, 2024

Visionary Optics LLC % Bret Andre Principal Consultant Andre Vision and Device Research 6119 Canter Lane West Linn. OR 97068

Re: K233325

Trade/Device Name: Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: September 25, 2023 Received: January 23, 2024

Dear Bret Andre:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233325

Device Name

Visionary Optics Scleral Contact Lens (roflufocon E. hexafocon A. hexafocon B. tisilfocon A. fluoroxyfocon A)

Indications for Use (Describe)

The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzman's Nodular Degeneration), dystrophies (e.g. Cogan's dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring from infection or trauma.

The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management of ocular surface disease including dry eye (e.g. ocular manifestations of Graft-versus-Host disease, Sjogren's syndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical and thermal burns, radiation, filamentary keratitis), epidermal ocular disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the Visionary Optics Scleral Contact Lens may incidentally provide correction of refractive error in persons with myopia, astigmatism or presbyopia.

Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the

for this collection of information is estimated to average 79 hours per time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510 (k) SUMMARY

This special 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:K233325
I. SUBMITTER
Date Prepared:September 25, 2023
Name:AddressVisionary Optics LLC1325 Progress DriveFront Royal, VA 22630
Contact Person:Donald R. SandersManager and CEO
Phone number:(630) 530-9700
Consultant/Correspondent:Andre Vision and Device ResearchBret Andre6119 Canter LaneWest Linn, OR 97068
Phone number(503) 372-5226
II. DEVICE
Trade Name:Visionary Optics Scleral Contact Lens(roflufocon D, roflufocon E, hexafocon A,hexafocon B, tisilfocon A, fluoroxyfocon A)
CommonName:Daily wear rigid gas permeable contact lens
ClassificationName:886.5916)Rigid gas permeable contact lens. (21 CFR)
RegulatoryClass:Class II
Product Code:HQD

{5}------------------------------------------------

III. PREDICATE DEVICE

The Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) is substantially equivalent to the following predicate devices:

  • · Visionary Optics Scleral Contact Lens (roflufocon E, hexafocon A, hexafocon A, hexafocon B, tisilfocon A) Manufactured by Visionary Optics LLC 510(k) number; K223394
  • · Acuity 200 (Fluoroxyfocon A) Rigid Gas Permeable Contact Lens Manufactured by Acuity Polymers, Inc. 510(k) number: K203571

IV. DEVICE DESCRIPTION

The Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Visionary Optics Scleral Contact Lens is lathe cut from one of the following hydrophobic, FDA Group #3 fluoro-silicone acrylate materials: roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, or fluoroxyfocon A.

ROFLUFOCON DROFLUFOCON EHEXAFOCONAHEXAFOCONBTISILFOCONAFLUOROXYFOCONA
Refractive Index1.43331.43321.4151.42401.43781.430
Light Transmission(clear)>97%>97%->95%--
Light Transmission(tinted)>90%>90%>92%>83%>91%>87%
Water Content<1%<1%<1%<1%<1%<1%
OxygenPermeability(Dk)ISO/FATT Method$100 \times 10^{-11}$ (cm²/sec)(ml O₂/ml x mm Hg @35°C)$125 \times 10^{-11}$ (cm²/sec)(ml O₂/ml x mm Hg @35°C)$140 \times 10^{-11}$(cm²/sec) (ml O₂/ml xmm Hg @ 35°C)$141 \times 10^{-11}$(cm²/sec) (ml O₂/ml xmm Hg @ 35°C)$180 \times 10^{-11}$(cm²/sec) (ml O₂/mlx mm Hg @ 35°C)$200 \times 10^{-11}$ (cm²/sec)(ml O₂/ml x mm Hg @ 35°C)
Contain one ormore of thefollowing coloradditivesconforming to:21 CFR Part 73 &74, Subpart DD & C Green No.6, FD & C RedNo. 17,CI Solvent Yellow18D & C Green No.6, FD & C RedNo. 17,CI SolventYellow 18D&C GreenNo. 6; D&CYellow No. 18;D&C VioletNo. 2;D&C GreenNo. 6; C.I.Solvent YellowNo. 18; D&CViolet No. 2;D&C Red No.17D&C GreenNo. 6, C.I.SolventYellow No.18, D&CViolet No. 2and D&C RedNo. 17D&C Green No. 6;D&C Violet No. 2;Solvent Yellow 18:D&C Red No. 17
UV Light BlockingYesYesYesYesYesYes

The physical properties of the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) are as follows:

{6}------------------------------------------------

The available parameters for the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) are as follows:

ParameterRangeTolerance
Base Curve5.5mm to 25.00mm$\pm$ 0.2mm
Center Thickness0.10mm to 3.00mm$\pm$ 0.1mm
Diameter12.00mm to 26.00mm$\pm$ 0.20mm
Spherical Power-35.00 D to +35.00 D(in .12D steps)$\pm$ 0.12 (0 to = 5D)<br $\pm$ 0.18 (5 to = 10.0D)<br $\pm$ 0.25 (10 to = 15D)<br $\pm$ 0.37 (15 to = 20D)<br $\pm$ 0.50 (over 20D)
Cylindrical Power+10.00 D to -10.00 D(in .12 D steps)$\pm$ 0.25 (0 to = 2D)<br $\pm$ 0.37 (2 to = 4D)<br $\pm$ 0.50 (over 4D)
Cylindrical Axis1° to 180° (in 1º steps)$\pm$ 5°
Bifocal Add+.12 D to +6.00 D(in .12 D steps)$\pm$ 0.25D

The Visionary Optics Scleral Contact Lens may be shipped "dry" or "wet". The primary container for shipping the Visionary Optics Scleral Contact Lens is the PolyVial Contact Lens Case. When shipped "wet", the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) are packaged and shipped in Boston Simplus Multiaction Solution. When shipped "wet", the Visionary Optics Scleral Contact Lens (tisilfocon A) is packaged and shipped in Menicon Unique pH Solution.

V. INDICATIONS FOR USE

The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophies (e.g. Cogan's dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring from infection or trauma.

The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management of ocular surface disease including dry eye (e.g. ocular manifestations of Graftversus-Host disease, Sjogren's syndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical and thermal burns, radiation, filamentary keratitis), epidermal ocular disorders, disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic. paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the Visionarv Optics Scleral Contact Lens may incidentally provide correction of refractive error in persons with myopia, hyperopia, astigmatism or presbyopia.

{7}------------------------------------------------

Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care svstem.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) is substantially equivalent to the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon B, tisilfocon A) (predicate device - K223394) in the following areas:

  • Components/Materials/Formulation (roflufocon D, roflufocon E, hexafocon A, hexafocon B, and tisilfocon A contact lens materials)
  • Manufacturing facility, procedures and controls ●
  • Product code (HQD) ●
  • . Classification (Class II) - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
  • FDA material group - group # 3 fluoro silicone acrylate
  • . Lathe cut manufacturing process
  • Scleral (large diameter) design ●
  • Actions and intended use ●
  • . Therapeutic (Irregular Cornea, Ocular Surface Disease) indications for use

The Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) is substantially equivalent to the Acuity 200 (Fluoroxyfocon A) Rigid Gas Permeable Contact Lens (predicate device - K203571) in the following areas:

  • Components/Materials/Formulation (fluoroxyfocon A contact lens material) ●
  • Product code (HQD) ●
  • Classification (Class II) Lenses, Rigid Gas Permeable, Daily Wear (21 CFR . 886.5916)
  • FDA material group - group # 3 fluoro silicone acrylate
  • Lathe cut manufacturing process
  • Scleral (large diameter) design ●
  • Actions and intended use .
  • Therapeutic (Irregular Cornea) indications for use ●

{8}------------------------------------------------

The following table depicts the classification and technical characteristics of the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) in comparison with the predicate devices.

Visionary Optics ScleralContact LensVisionary Optics ScleralContact LensAcuity 200 (Fluoroxyfocon A)Rigid Gas Permeable ContactLens
Subject DevicePredicate DevicePredicate Device
510(k) NumberK223394K203571
Intended UseDaily WearDaily WearDaily Wear
Device andClassificationClass IILenses, Rigid GasPermeable, Daily WearHQDClass IILenses, Rigid Gas Permeable,Daily WearHQDClass IILenses, Rigid Gas Permeable,Daily WearHQD
Product CodeHQDHQDHQD
Production MethodLathe-cutLathe-cutLathe-cut
Material (USAN)Roflufocon DRoflufocon EHexafocon AHexafocon BTisilfocon AFluoroxyfocon ARoflufocon DRoflufocon EHexafocon AHexafocon BTisilfocon AFluoroxyfocon A
FDA Group #Group # 3 Fluoro SiliconeAcrylateGroup # 3 Fluoro SiliconeAcrylateGroup # 3 Fluoro SiliconeAcrylate
Water Content<1%<1%<1%
UV Absorber/BlockeravailableYESYESYES
Indications for Use
VisionaryOptics ScleralContact Lens(SubjectDevice)The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management ofmultiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus,keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophies (e.g. Cogan'dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant,LASIK, radial keratotomy), and corneal scarring from infection or trauma.
The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management ofocular surface disease including dry eye (e.g. ocular manifestations of Graft-versus-Host disease, Sjogren'ssyndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical andthermal burns, radiation, filamentary keratitis), epidermal ocular disorders, disorders of the skin (e.g. atopy,ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia),and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tearreservoir and protection against an adverse environment. When prescribed for therapeutic use for distortedcornea or ocular surface disease, the Visionary Optics Scleral Contact Lens may incidentally providecorrection of refractive error in persons with myopia, hyperopia, astigmatism or presbyopia.Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning,disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lensmay be cleaned and disinfected using a chemical (not heat) lens care system.

{9}------------------------------------------------

VisionaryOptics ScleralContact Lens(PredicateDevice:K223394)The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management ofmultiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus,keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophies (e.g. Cogan'sdystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-surgery (e.g. corneal transplant,LASIK, radial keratotomy), and corneal scarring from infection or trauma.The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management ofocular surface disease including dry eye (e.g. ocular manifestations of Graft-versus-Host disease, Sjogren'ssyndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical andthermal burns, radiation, filamentary keratitis), epidermal ocular disorders, disorders of the skin (e.g. atopy,ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia),and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tearreservoir and protection against an adverse environment. When prescribed for therapeutic use for distortedcornea or ocular surface disease, the Visionary Optics Scleral Contact Lens may incidentally providecorrection of refractive error in persons with myopia, hyperopia, astigmatism or presbyopia.Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning,disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lensmay be cleaned and disinfected using a chemical (not heat) lens care system.
Acuity 200(FluoroxyfoconA) Rigid GasPermeableContact Lens(PredicateDevice:K203571)The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens is indicated for daily wear for thecorrection of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakicpersons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyesthat require a rigid contact lens for the management of irregular corneal conditions such as keratoconus,pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens may be cleaned and disinfectedusing a chemical (not heat) lens care system.

VII. PERFORMANCE DATA

~ Non-Clinical Studies ~

Non-clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A has been addressed by reference to previous 510(k) clearances.

Additionally, the following testing was performed on finished Visionary Optics Scleral Contact Lenses (fluoroxyfocon A):

Bench Testing-manufacturing verification testing was conducted to demonstrate the ability of Visionary Optics LLC to manufacture lenses, on a repeatable basis, from fluoroxyfocon A supplied lens blanks to a variety of prescribed parameters. All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.

Bioburden Testing-bioburden testing conducted on fluoroxyfocon A rigid gas permeable lenses manufactured at Visionary Optics LLC demonstrated that the colony forming units (CFU) per lens was within the established acceptance criteria of less than 100 CFU per lens.

{10}------------------------------------------------

~ Clinical Studies ~

Clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D, roflufocon E, hexafocon A, hexafocon B and tisilfocon A with the labeled indications for use has been addressed through previous 510(k) clearances.

Nine (9) independent practitioners evaluated 33 patients (63 total eyes) presenting with various conditions that were managed using scleral contact lenses manufactured by Visionary Optics from Acuity 200 (fluoroxyfocon A) material. The practitioners reviewed all patients therapeutically managed for ocular surface disease with Visionary Optics Scleral Contact Lens (fluoroxyfocon A) for at least 3 months, and reported the outcome of each patient-including any ocular adverse reactions or worsening of the patients' conditions-over the treatment period. All patients were fit with successful outcomes over a treatment follow-up period of at least 3 months (Min: 90 days; Max: 259 days; Median: 113 days). The total number of days of treatment follow-up for all patients combined was There were no serious or significant adverse reactions reported. For all 4.306 days. patients, the ocular condition(s) and vision remained stable or improved during management with the Visionary Optics Scleral Contact Lens (fluoroxyfocon A).

~ Conclusions Drawn from Testing ~

Results from testing presented in this premarket notification for the Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) demonstrate no relevant differences from the predicate devices and supports the substantial equivalence claim.

VIII. CONCLUSIONS

Substantial Equivalence

Information presented in this premarket notification establishes that Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A) for daily wear are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indications.

Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.