(267 days)
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals.
As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment.
eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
The AgileMD eCARTv5 Clinical Deterioration Suite ("eCART") is a cloud-based software device that is integrated into the electronic health record ("EHR") in order to anticipate clinical deterioration in adult ward patients, which is signified as either of the following two predicted outcomes: (1) death or (2) ICU transfer. The tool synthesizes routine vital signs, laboratory data, and patient demographics into a single value that can be used to flag patients at-risk of the composite outcome of clinical deterioration for additional evaluation and monitoring. eCARTv5 requires the healthcare system within which it will be used, to provide an EHR connection and data interfaces through which the patient data necessary to run the software will be transmitted.
The primary functions of the system are imparted by the Gradient Boosted Machine ("GBM") learning algorithm that takes input directly from the EHR, in real time, to provide an assessment of patients and displays its outputs in an intuitive user interface which drives providers to follow standardized clinical workflows (established by their institutions) for elevated-risk patients.
eCARTv5's end users include med-surg nursing staff, physicians and other providers caring for these patients. The eCARTv5 composite score is determined from the model output (predicted probability of deterioration) scaled from 0-100, based on the specificity (true negative rate). The observed rate of deterioration at each eCART score threshold, displayed as the odds of deterioration in the next 24 hours, is presented to the user along with the scaled score. Default thresholds are set to an eCART of 93 and 97, respectively, for moderate and high risk categorization.
Here's a breakdown of the acceptance criteria and study details for the eCARTv5 Clinical Deterioration Suite, based on the provided FDA 510(k) summary:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for the eCARTv5 device are implicitly defined by the performance metrics reported in the validation studies, specifically Area Under the Receiver Operating Characteristic curve (AUROC), Sensitivity, Specificity, Positive Predictive Value (PPV), and Negative Predictive Value (NPV) for two risk thresholds (Moderate-risk at eCART ≥93 and High-risk at eCART ≥97). The composite outcome of interest is "Deterioration" (death or ICU transfer within 24 hours).
Table: Acceptance Criteria (Implicit) and Reported Device Performance
| Performance Metric | Acceptance Criteria (Implicit Target) | Retrospective Cohort (Deterioration) | Prospective Cohort (Deterioration) |
|---|---|---|---|
| AUROC | (Target > 0.82) | 0.835 (0.834, 0.835) | 0.828 (0.827, 0.829) |
| Moderate-risk threshold (eCART ≥93) | |||
| Sensitivity | (Target ~48-52%) | 51.8% (51.7%, 51.8%) | 48.8% (48.7%, 49.0%) |
| Specificity | (Target ~93-94%) | 93.1% (93.1%, 93.1%) | 93.3% (93.3%, 93.3%) |
| PPV | (Target ~8-9%) | 9.0% (9.0%, 9.1%) | 8.9% (8.8%, 8.9%) |
| NPV | (Target ~99%) | 99.3% (99.3%, 99.3%) | 99.3% (99.3%, 99.3%) |
| High-risk threshold (eCART ≥97) | |||
| Sensitivity | (Target ~33-38 %) | 38.6% (38.5%, 38.7%) | 33.7% (33.6%, 33.9%) |
| Specificity | (Target ~96-97%) | 96.9% (96.9%, 96.9%) | 97.3% (97.3%, 97.3%) |
| PPV | (Target ~14%) | 14.2% (14.1%, 14.2%) | 14.2% (14.1%, 14.3%) |
| NPV | (Target ~99%) | 99.2% (99.2%, 99.2%) | 99.1% (99.1%, 99.1%) |
Note: The "Acceptance Criteria (Implicit Target)" values are inferred based on the consistently reported values that demonstrate performance above random chance and clinical utility for risk stratification. The document does not explicitly state pre-defined quantitative acceptance criteria but rather presents the achieved performance as a demonstration of substantial equivalence.
Study Details
-
Sample sizes used for the test set and data provenance:
- Retrospective Test Set:
- Encounters (N): 1,769,461 unique hospitalizations.
- Observations (n): 132,873,833 eCART scores.
- Unique Patients: 934,454
- Data Provenance: Admissions between 2009 and 2023 from three geographically distinct health systems. The specific countries are not mentioned, but "US" is inferred from typical FDA submissions. It is retrospective.
- Prospective Test Set:
- Encounters (N): 205,946 unique hospitalizations.
- Observations (n): 21,516,964 eCART scores.
- Unique Patients: 151,233
- Data Provenance: Non-overlapping admissions between 2023 and 2024 from the same three healthcare systems as the retrospective analysis. It is prospective.
- Retrospective Test Set:
-
Number of experts used to establish the ground truth for the test set and qualifications of those experts:
- The document does not specify the number of experts used or their qualifications for establishing ground truth. The ground truth (death or ICU transfer) appears to be derived directly from Electronic Health Record (EHR) data, which is objective outcome data, rather than requiring expert labeling.
-
Adjudication method for the test set:
- The document does not specify an adjudication method. Given that the ground truth is death or ICU transfer from EHR, a formal adjudication process involving multiple experts for each case may not have been necessary, as these are typically clear clinical outcomes documented in the EHR.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:
- A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned as being performed to compare human readers with and without AI assistance. The performance data presented is for the standalone algorithm.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone study was presented. The performance metrics (AUROC, Sensitivity, Specificity, PPV, NPV) are reported for the eCART algorithm itself, without human intervention in the reported performance. The device is intended as a "clinical decision support device" to "aid clinical teams in identifying which patients are most likely to clinically deteriorate."
-
The type of ground truth used:
- The ground truth used is outcomes data derived from the Electronic Health Record (EHR). Specifically, "Deterioration is defined as death or ward to ICU transfer within 24 hours following a score." Mortality is defined as "death within 24 hours following a score." These are objective clinical events.
-
The sample size for the training set:
- The document states: "eCART's algorithm was trained on ward patients..." but does not explicitly provide the sample size of the training set. It only provides details for the retrospective and prospective validation cohorts which are distinct from the training set.
-
How the ground truth for the training set was established:
- The document does not explicitly detail how the ground truth for the training set was established. However, given the nature of the ground truth for the test set (death or ICU transfer from validated EHR data), it is reasonable to infer that the ground truth for the training set would have been established similarly using objective patient outcomes data from EHRs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 21, 2024
AgileMD, Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street 23rd Floor Philadelphia, Pennsylvania 19103
Re: K233253
Trade/Device Name: eCARTy5 Clinical Deterioration Suite ("eCART") Regulation Number: 21 CFR 870.2210 Regulation Name: Adjunctive Predictive Cardiovascular Indicator Regulatory Class: Class II Product Code: ONL Dated: May 17, 2024 Received: May 17, 2024
Dear Kelliann Payne:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
- d. F
Page 2
LCDR Stephen Browning Assistant Director
{2}------------------------------------------------
Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known) K233253
Device Name
eCART
Indications for Use (Describe)
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals.
As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment.
eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
510(k) SUMMARY
eCARTv5 Clinical Deterioration Suite
Submitter
AgileMD, Inc. 2261 Market Street #4378 San Francisco CA, 94114 Phone: 415-650-0522 Contact Person: Borna Safabakhsh
Date Prepared: June 20, 2024
Name of Device: eCARTv5 Clinical Deterioration Suite ("eCART")
Common or Usual Name: Clinical Monitor
Classification Regulation: 21 CFR 870.2210 (Adjunctive Predictive Cardiovascular Indicator)
Regulatory Class: II
Product Code: QNL
Predicate Device: CLEWICU (K200717)
Reference Device: PeraServer and PeraTrend (K172959)
Device Description
The AgileMD eCARTv5 Clinical Deterioration Suite ("eCART") is a cloud-based software device that is integrated into the electronic health record ("EHR") in order to anticipate clinical deterioration in adult ward patients, which is signified as either of the following two predicted outcomes: (1) death or (2) ICU transfer. The tool synthesizes routine vital signs, laboratory data, and patient demographics into a single value that can be used to flag patients at-risk of the composite outcome of clinical deterioration for additional evaluation and monitoring. eCARTv5 requires the healthcare system within which it will be used, to provide an EHR connection and data interfaces through which the patient data necessary to run the software will be transmitted.
The primary functions of the system are imparted by the Gradient Boosted Machine ("GBM") learning algorithm that takes input directly from the EHR, in real time, to provide an assessment of patients and displays its outputs in an intuitive user interface which drives providers to follow standardized clinical workflows (established by their institutions) for elevated-risk patients.
eCARTv5's end users include med-surg nursing staff, physicians and other providers caring for these patients. The eCARTv5 composite score is determined from the model output (predicted probability of deterioration) scaled from 0-100, based on the specificity (true negative rate). The observed rate of deterioration at each eCART score threshold, displayed as the odds of deterioration in the next 24 hours, is presented to the user along with the scaled score. Default thresholds are set to an eCART of 93 and 97, respectively, for moderate and high risk categorization.
{5}------------------------------------------------
Intended Use / Indications for Use
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals.
As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment.
eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
The differences in indications for use from the predicate device are not critical to the intended use of eCARTv5, nor do they raise different questions of safety or effectiveness when the subject device is used as labeled. The minor differences are supported by adequate performance testing that show the subject device is substantially equivalent for the proposed indications for use. Both devices are intended to provide trained healthcare providers with a patient status score that reflects the underlying patient condition to supplement standard of care and informed decision making.
Summary of Technological Characteristics
At a high level, the subject and predicate devices are based on the following same technological elements:
- · A risk-predictive output generated using a machine-learning algorithm.
- Inputs to the software device include vital signs, assessments, and laboratory data collected . from the hospital EHR system.
- . The output is a color-defined (red-yellow), real-time risk categorization to provide users information regarding potential risk of patient deterioration.
- Embedded workflow features enable users to acknowledge the risk prediction and record next . steps for medical management, based on their independent medical judgment.
The primary technological differences between the devices are that eCART's algorithm was trained on ward patients to predict the probability of ICU transfer or death, whereas the predicate is trained in ICU patients to predict respiratory failure and/or hemodynamic instability. Additionally, the predicate device does not provide a risk score (but rather, a risk category). The minor technological differences between the devices raise no different questions of safety or effectiveness. With respect to the output, the reference device (PeraServer and PeraTrend (K172959) provides a risk score (the Rothman Index), which was validated on the outcome of mortality within 24 hours and included training and testing on data from adult ward patients (similar to eCART). Performance data for eCART, including retrospective and prospective clinical testing as well as human factors validation, demonstrates that the subject device is substantially equivalent to the predicate device.
Performance Data
eCART was assessed in retrospective and prospective validation studies with adult ward patients from three geographically distinct health systems. Test characteristics for both cohorts are shown in
{6}------------------------------------------------
Table A below. The retrospective analysis included admissions between 2009 and 2023 in three health systems. Deterioration is defined as death or ward to ICU transfer within 24 hours following a score. The mortality outcome is defined as death within 24 hours following a score. The prospective validation was performed in the same three healthcare systems where the retrospective analysis was undertaken. This analysis included non-overlapping admissions between 2023 and 2024.
Encounters (N) are defined as unique hospitalizations for a single patients may have more than one encounter in the data set if they were admitted and discharged from a study hospital more than once during the study period. Each encounter includes all eCART scores (observations, n) generated during that hospitalization.
| DETERIORATION | MORTALITY | ||||
|---|---|---|---|---|---|
| Retrospective | Prospective | Retrospective | Prospective | ||
| N=1,769,461 | N=205,946 | N=1,769,461 | N=205,946 | ||
| AUROC | 0.835 (0.834, 0.835) | 0.828 (0.827, 0.829) | 0.923 (0.923, 0.924) | 0.913 (0.911, 0.914) | |
| n=132,873,833 | n=21,516,964 | n=132,873,833 | n=21,516,964 | ||
| Outcome Prevalence | 1.3% | 1.3% | 0.2% | 0.2% | |
| 1,744,044/132,873,833 | 284,678/21,516,964 | 249,636/132,873,833 | 39,609/21,516,964 | ||
| Moderate-riskthreshold | PositivityRate | 7.5% | 7.3% | 7.5% | 7.3% |
| 9,982,577/132,873,833 | 1,563,648/21,516,964 | 9,982,577/132,873,833 | 1,563,648/21,516,964 | ||
| Sensitivity | 51.8% (51.7%, 51.8%) | 48.8% (48.7%, 49.0%) | 75.8% (75.6%, 75.9%) | 71.7% (71.3%, 72.1%) | |
| 902,951/1,744,044 | 139,043/284,678 | 189,171/249,636 | 28,399/39,609 | ||
| Specificity | 93.1% (93.1%, 93.1%) | 93.3% (93.3%, 93.3%) | 92.6% (92.6%, 92.6%) | 92.9% (92.8%, 92.9%) | |
| 122,050,163/131,129,789 | 19,807,681/21,232,286 | 122,830,791/132,624,197 | 19,942,106/21,477,355 | ||
| eCART≥93 | PPV | 9.0% (9.0%, 9.1%) | 8.9% (8.8%, 8.9%) | 1.9% (1.9%, 1.9%) | 1.8% (1.8%, 1.8%) |
| 902,951/9,982,577 | 139,043/1,563,648 | 189,171/9,982,577 | 28,399/1,563,648 | ||
| NPV | 99.3% (99.3%, 99.3%) | 99.3% (99.3%, 99.3%) | 100.0% (100.0%, 100.0%) | 99.9% (99.9%, 99.9%) | |
| 122,050,163/122,891,256 | 19,807,681/19,953,316 | 122,830,791/122,891,256 | 19,942,106/19,953,316 | ||
| Risk Diff | 8.4% (8.3%, 8.4%) | 8.2% (8.1%, 8.2%) | 1.8% (1.8%, 1.9%) | 1.8% (1.7%, 1.8%) | |
| High-riskthresholdeCART≥97 | PositivityRate | 3.6% | 3.1% | 3.6% | 3.1% |
| 4,753,332/132,873,833 | 675,279/21,516,964 | 4,753,332/132,873,833 | 675,279/21,516,964 | ||
| Sensitivity | 38.6% (38.5%, 38.7%) | 33.7% (33.6%, 33.9%) | 64.7% (64.5%, 64.9%) | 58.1% (57.6%, 58.6%) | |
| 673,539/1,744,044 | 96,051/284,678 | 161,531/249,636 | 23,025/39,609 | ||
| Specificity | 96.9% (96.9%, 96.9%) | 97.3% (97.3%, 97.3%) | 96.5% (96.5%, 96.5%) | 97.0% (97.0%, 97.0%) | |
| 127,049,996/131,129,789 | 20,653,058/21,232,286 | 128,032,396/132,624,197 | 20,825,101/21,477,355 | ||
| PPV | 14.2% (14.1%, 14.2%) | 14.2% (14.1%, 14.3%) | 3.4% (3.4%, 3.4%) | 3.4% (3.4%, 3.5%) | |
| 673,539/4,753,332 | 96,051/675,279 | 161,531/4,753,332 | 23,025/675,279 | ||
| NPV | 99.2% (99.2%, 99.2%) | 99.1% (99.1%, 99.1%) | 99.2% (99.2%, 99.2%) | 99.9% (99.9%, 99.9%) | |
| 127,049,996/128,120,501 | 20,653,058/20,841,685 | 127,049,996/128,120,501 | 20,825,101/20,841,685 | ||
| Risk Diff | 13.3% (13.3%, 13.4%) | 13.3% (13.2%, 13.4%) | 3.3% (3.3%, 3.3%) | 3.3% (3.3%, 3.4%) |
Table A. eCART Prediction of Deterioration and Mortality in the Full External Retrospective and Prospective Cohorts
Note: The encounter and observation data above represents 934,454 and 151,233 unique patients in the retrospective and prospective cohorts, respectively. 95% confidence intervals were calculated using the Clopper-Pearson method.
Subgroup analyses for select comorbidities are presented in Tables B-C below for deterioration and mortality, respectively, and show comparable performance across conditions. Sequencing data for the SARS-CoV-2 variants included in the COVID subgroup analyses was not available, but the sample likely included the common variants circulating in the US between 2020 and the early part of 2024 (see https://cov-lineages.org/lineage list.html). The comparative effectiveness of the model between existing SARS-CoV-2 strains or to future strains is unknown.
{7}------------------------------------------------
| Comorbidity | |||||
|---|---|---|---|---|---|
| DETERIORATION | Congestive HeartFailureN = 306,140 | COVID-19N = 49,834 | Chronic PulmonaryDiseaseN = 443,263 | SepsisN = 639,802 | |
| AUROC | 0.810 (0.809, 0.810) | 0.858 (0.857, 0.859) | 0.824 (0.823, 0.824) | 0.836 (0.836, 0.836) | |
| n = 31,445,409 | n = 5,961,084 | n = 38,638,180 | n = 72,055,421 | ||
| Outcome Prevalence | 2.2% | 2.3% | 1.7% | 1.8% | |
| 695,781/31,445,409 | 136,077/5,961,084 | 646,095/38,638,180 | 1,295,361/72,055,421 | ||
| Sensitivity | 53.9% (53.8%, 54.1%) | 69.6% (69.4%, 69.9%) | 53.9% (53.8%, 54.0%) | 56.4% (56.3%, 56.5%) | |
| 375,328/695,781 | 94,726/136,077 | 348,271/646,095 | 730,846/1,295,361 | ||
| Moderate- | Specificity | 89.8% (89.7%, 89.8%) | 84.6% (84.6%, 84.7%) | 91.1% (91.0%, 91.1%) | 90.8% (90.8%, 90.8%) |
| risk | 27,600,037/30,749,628 | 4,929,294/5,825,007 | 34,593,348/37,992,085 | 64,243,126/70,760,060 | |
| threshold | PPV | 10.6% (10.6%, 10.7%) | 9.6% (9.5%, 9.6%) | 9.3% (9.3%, 9.3%) | 10.1% (10.1%, 10.1%) |
| (≥93) | 375,328/3,524,919 | 94,726/990,439 | 348,271/3,747,008 | 730,846/7,247,780 | |
| NPV | 98.9% (98.8%, 98.9%) | 99.2% (99.2%, 99.2%) | 99.1% (99.1%, 99.1%) | 99.1% (99.1%, 99.1%) | |
| 27,600,037/27,920,490 | 4,929,294/4,970,645 | 34,593,348/34,891,172 | 64,243,126/64,807,641 | ||
| Sensitivity | 40.2% (40.1%, 40.3%) | 55.9% (55.6%, 56.2%) | 40.1% (40.0%, 40.3%) | 42.6% (42.6%, 42.7%) | |
| 279,748/695,781 | 76,086/136,077 | 259,354/646,095 | 552,417/1,295,361 | ||
| Specificity | 95.2% (95.1%, 95.2%) | 90.7% (90.6%, 90.7%) | 95.8% (95.8%, 95.9%) | 95.6% (95.6%, 95.6%) | |
| High-risk | 29,259,307/30,749,628 | 5,280,458/5,825,007 | 36,413,561/37,992,085 | 67,648,075/70,760,060 | |
| threshold(≥97) | PPV | 15.8% (15.8%, 15.9%) | 12.3% (12.2%, 12.3%) | 14.1% (14.1%, 14.2%) | 15.1% (15.1%, 15.1%) |
| 279,748/1,770,069 | 76,086/620,635 | 259,354/1,837,878 | 552,417/3,664,402 | ||
| NPV | 98.6% (98.6%, 98.6%) | 98.9% (98.9%, 98.9%) | 98.9% (98.9%, 99.0%) | 98.9% (98.9%, 98.9%) | |
| 29,259,307/29,675,340 | 5,280,458/5,340,449 | 36,413,561/36,800,302 | 67,648,075/68,391,019 |
Table B. eCART Prediction of Clinical Deterioration in Select Comorbidities in the Full External Retrospective Cohort
Note: The encounter and observation data above represents 136,197 unique patients with congestive heart failure, 44,231 with chronic obstructive pulmonary disease, and 385,997 with sepsis were calculated using the Cloper-Pearson method.
Table C. eCART Prediction of Mortality in Select Comorbidities in the Full External Retrospective Cohort
| MORTALITY | Comorbidity | |||
|---|---|---|---|---|
| Congestive HeartFailure | COVID-19N = 49,834 | Chronic PulmonaryDiseaseN = 443,263 | SepsisN = 639,802 | |
| N = 306,140 | ||||
| AUROC | 0.893 (0.892, 0.894) | 0.924 (0.923, 0.926) | 0.912 (0.911, 0.913) | 0.914 (0.914, 0.915) |
| n = 31,445,409 | n = 5,961,084 | n = 38,638,180 | n = 72,055,421 | |
| Outcome Prevalence | 0.3%103,194/31,445,09 | 0.5%30,714/5,961,084 | 0.2%85,582/38,638,180 | 0.2%160,348/72,055,421 |
| Moderate-riskthreshold(≥93) | Sensitivity73.6% (73.3%, 73.9%)75,961/103,194 | Sensitivity86.7% (86.3%, 87.1%)26,639/30,714 | Sensitivity76.1% (75.8%, 76.4%)65,111/85,582 | Sensitivity77.% (76.8%, 77.2%)123,467/160,348 |
| Specificity89.0% (89.0%, 89.0%)27,893,257/31,342,215 | Specificity83.7% (83.7%, 83.8%)4,966,570/5,930,370 | Specificity90.4% (90.4%, 90.5%)34,870,701/38,552,598 | Specificity90.1% (90.1%, 90.1%)64,770,760/71,895,073 | |
| PPV2.2% (2.1%, 2.2%)75,961/3,524,919 | PPV2.7% (2.7%, 2.7%)26,639/990,439 | PPV1.7% (1.7%, 1.8%)65,111/3,747,008 | PPV1.7% (1.7%, 1.7%)123,467/7,247,780 | |
| NPV99.9% (99.9%, 99.9%)27,893,257/27,920,490 | NPV99.9% (99.9%, 99.9%)4,966,570/4,970,645 | NPV99.9% (99.9%, 99.9%)34,870,701/34,891,172 | NPV99.9% (99.9%, 99.9%)64,770,760/64,807,641 | |
| High-riskthreshold(≥97) | Sensitivity61.7% (61.4%, 62.0%)63,697/103,194 | Sensitivity79.2% (78.7%, 79.6%)24,315/30,714 | Sensitivity64.9% (64.5%, 65.2%)55,507/85,582 | Sensitivity66.1% (65.8%, 66.3%)105,915/160,348 |
| Specificity94.6% (94.5%, 94.6%)29,635,843/31,342,215 | Specificity89.9% (89.9%, 90.0%)5,334,050/5,930,370 | Specificity95.4% (95.4%, 95.4%)36,770,227/38,552,598 | Specificity95.1% (95.%, 95.1%)68,336,586/71,895,073 | |
| PPV3.6% (3.6%, 3.6%)63,697/1,770,069 | PPV3.9% (3.9%, 4.0%)24,315/620,635 | PPV3.0% (3.0%, 3.0%)55,507/1,837,878 | PPV2.9% (2.9%, 2.9%)105,915/3,664,402 | |
| NPV99.9% (99.9%, 99.9%)29,635,843/29,675,340 | NPV99.9% (99.9%, 99.9%)5,334,050/5,340,449 | NPV99.9% (99.9%, 99.9%)36,770,227/36,800,302 | NPV99.9% (99.9%, 99.9%)68,336,586/68,391,019 |
Note: The encounter and observation data above represents 136,197 unique patients with congestive heart failure, 44,8231 with chronic obstructive pulmonary disease, and 385,997 with sepss. 95% confidence intervals were calculated using the Cloper-Pearson method.
{8}------------------------------------------------
Subgroup analyses for race are presented in Table D, below, for deterioration. All five groups met the performance thresholds for AUROC and sensitivity.
| Table D. eCART Retrospective Validation Test Characteristics for the Outcomes of Clinical | ||||||
|---|---|---|---|---|---|---|
| Deterioration by Race | ||||||
| DETERIORATION | Race | American Indian orAlaska NativeN = 6,468 | Asian/Mideast IndianN = 26,681 | Black/African-AmericanN = 252,982 | NativeHawaiian/OtherPacific IslanderN = 2,496 | White/CaucasianN = 1,384,075 |
| AUROC | 0.814 (0.808, 0.820)486,615 | 0.847 (0.844, 0.850)1,694,681 | 0.831 (0.830, 0.832)20,057,696 | 0.862 (0.854, 0.870)177,513 | 0.834 (0.833, 0.834)104,522,118 | |
| Outcome Prevalence | 1.2%6,026/486,615 | 1.2%20,736/1,694,681 | 1.2%234,979/20,057,696 | 1.3%2,395/177,513 | 1.3%1,403,867/104,522,118 | |
| Moderate-riskthreshold(≥93) | PosRate | 6.7%32,455/486,615 | 7.2%121,532/1,694,681 | 6.5%1,301,626/20,057,696 | 7.2%12,738/177,513 | 7.7%8,085,868/104,522,118 |
| Sensitivity | 46.6% (45.3%, 47.9%)2,809/6,026 | 52.9% (52.2%, 53.6%)10,968/20,736 | 48.1% (47.9%, 48.3%)112,939/234,979 | 56.5% (54.5%, 58.5%)1,353/2,395 | 52.3% (52.2%, 52.3%)733,566/1,403,867 | |
| Specificity | 93.8% (93.8%, 93.9%)450,943/480,589 | 93.4% (93.4%, 93.4%)1,563,381/1,673,945 | 94.0% (94.0%, 94.0%)18,634,030/19,822,717 | 93.5% (93.4%, 93.6%)163,733/175,118 | 92.9% (92.9%, 92.9%)95,765,949/103,118,251 | |
| PPV | 8.7% (8.4%, 9.0%)2,809/32,455 | 9.0% (8.9%, 9.2%)10,968/121,532 | 8.7% (8.6%, 8.7%)112,939/1,301,626 | 10.6% (10.1%, 11.2%)1,353/12,738 | 9.1% (9.1%, 9.1%)733,566/8,085,868 | |
| NPV | 99.3% (99.3%, 99.3%)450,943/454,160 | 99.4% (99.4%, 99.4%)1,563,381/1,573,149 | 99.3% (99.3%, 99.4%)18,634,030/18,756,070 | 99.4% (99.3%, 99.4%)163,733/164,775 | 99.3% (99.3%, 99.3%)95,765,949/96,436,250 | |
| Risk Diff | 7.9% (7.6%, 8.3%) | 8.4% (8.2%, 8.6%) | 8.0% (8.0%, 8.1%) | 10.0% (9.5%, 10.5%) | 8.4% (8.4%, 8.4%) | |
| High-riskthreshold(≥97) | PosRate | 3.1%14,933/486,615 | 3.6%60,178/1,694,681 | 3.1%611,210/20,057,696 | 3.5%6,174/177,513 | 3.7%3,831,413/104,522,118 |
| Sensitivity | 35.2% (34.0%, 36.4%)2,122/6,026 | 40.9% (40.3%, 41.6%)8,491/20,736 | 34.9% (34.7%, 35.1%)81,982/234,979 | 40.0% (38.0%, 42.0%)957/2,395 | 39.1% (39.0%, 39.1%)548,404/1,403,867 | |
| Specificity | 97.3% (97.3%, 97.4%)467,778/480,589 | 96.9% (96.9%, 96.9%)1,622,258/1,673,945 | 97.3% (97.3%, 97.3%)19,293,489/19,822,717 | 97.0% (96.9%, 97.1%)169,901/175,118 | 96.8% (96.8%, 96.8%)99,835,242/103,118,251 | |
| PPV | 14.2% (13.7%, 14.8%)2,122/14,933 | 14.1% (13.8%, 14.4%)8,491/60,178 | 13.4% (13.3%, 13.5%)81,982/611,210 | 15.5% (14.6%, 16.4%)957/6,174 | 14.3% (14.3%, 14.3%)548,404/3,831,413 | |
| NPV | 99.2% (99.1%, 99.2%)467,778/471,682 | 99.3% (99.2%, 99.3%)1,622,258/1,634,503 | 99.2% (99.2%, 99.2%)19,293,489/19,446,486 | 99.2% (99.1%, 99.2%)169,901/171,339 | 99.2% (99.1%, 99.2%)99,835,242/100,690,705 | |
| Risk Diff | 13.4% (12.8%, 13.9%) | 13.4% (13.1%, 13.6%) | 12.6% (12.5%, 12.7%) | 14.7% (13.8%, 15.6%) | 13.5% (13.4%, 13.5%) |
Note: The encounter and observation data above represents 3,282 unique American Indian or Alaska Native patients, 17,657 Asian/Mideast Asian patients, 116,642 Black/African-American patients, 1,427 Native Hawaiian/Other Pacific Islander patients, 95% confidence intervals were calculated using the Clopper-Pearson method.
Conclusion
eCARTv5 has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate device, CLEWICU. The minor differences in indications for use do not alter the intended fundamental clinical purpose of the device, nor affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between eCARTv5 and its predicate device raise no different questions of safety or effectiveness and are further supported by performance data. Thus, the device is substantially equivalent.
§ 870.2210 Adjunctive predictive cardiovascular indicator.
(a)
Identification. The adjunctive predictive cardiovascular indicator is a prescription device that uses software algorithms to analyze cardiovascular vital signs and predict future cardiovascular status or events. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.(b)
Classification. Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must be provided, including:
(i) A full characterization of the software technical parameters, including algorithms;
(ii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications;
(iii) A description of sensor data quality control measures;
(iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy;
(v) A description of the expected time to patient status or clinical event for all expected outputs, accounting for differences in patient condition and environment; and
(vi) The sensitivity, specificity, positive predictive value, and negative predictive value in both percentage and number form.
(2) A scientific justification for the validity of the predictive cardiovascular indicator algorithm(s) must be provided. This justification must include verification of the algorithm calculations and validation using an independent data set.
(3) A human factors and usability engineering assessment must be provided that evaluates the risk of misinterpretation of device output.
(4) A clinical data assessment must be provided. This assessment must fulfill the following:
(i) The assessment must include a summary of the clinical data used, including source, patient demographics, and any techniques used for annotating and separating the data.
(ii) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(iii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
(iv) The assessment must evaluate how the device output correlates with the predicted event or status.
(5) Labeling must include:
(i) A description of what the device measures and outputs to the user;
(ii) Warnings identifying sensor acquisition factors that may impact measurement results;
(iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information;
(iv) A specific time or a range of times before the predicted patient status or clinical event occurs, accounting for differences in patient condition and environment;
(v) Key assumptions made during calculation of the output;
(vi) The type(s) of sensor data used, including specification of compatible sensors for data acquisition;
(vii) The expected performance of the device for all intended use populations and environments; and
(viii) Relevant characteristics of the patients studied in the clinical validation (including age, gender, race or ethnicity, and patient condition) and a summary of validation results.