K Number
K233238
Device Name
Air Pressure Therapy System (Model:VU-IPC4M)
Date Cleared
2024-06-14

(260 days)

Product Code
Regulation Number
890.5650
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Air Pressure Therapy System (Model:VU-IPC4M) is intended for home to temporarily relieve minor muscle aches and or pains, and to temporarily increase circulation to the treated areas in people who are in good health.
Device Description
Air Pressure Therapy System (Model:VU-IPC4M) consists of air pressure sensor, air pump, sleeves Etc. working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.
More Information

Not Found

No
The description mentions a "microprocessor" and "software verification and validation testing," which are standard components of many medical devices and do not inherently indicate the presence of AI/ML. There is no mention of AI, ML, deep learning, training data, or performance metrics typically associated with AI/ML algorithms. The device's function appears to be based on a pre-programmed sequence of inflation and deflation cycles.

Yes
The device is described as an "Air Pressure Therapy System" intended to "temporarily relieve minor muscle aches and or pains, and to temporarily increase circulation to the treated areas," which are therapeutic claims.

No

The device is intended to temporarily relieve minor muscle aches and pains and temporarily increase circulation, which are therapeutic rather than diagnostic functions.

No

The device description explicitly lists hardware components such as "air pressure sensor, air pump, sleeves Etc." and describes their physical interaction ("The air pump is connected to the dedicated sleeves via a series of hoses."). The performance studies also include testing related to hardware, such as "seam strength testing" and "failure mode verification and validation testing to ensure mitigation of overpressurization risk if there is a software failure," indicating a physical device with potential hardware failures.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "temporarily relieve minor muscle aches and or pains, and to temporarily increase circulation to the treated areas in people who are in good health." This describes a physical therapy or wellness device, not a device used to examine specimens derived from the human body to provide information for diagnosis, monitoring, or compatibility testing.
  • Device Description: The description details a system that applies air pressure to limbs. This is a mechanical action, not a process that analyzes biological samples.
  • Lack of IVD Characteristics: There is no mention of analyzing blood, urine, tissue, or any other biological specimen. There is no mention of reagents, assays, or any of the typical components or processes associated with in vitro diagnostics.

Therefore, the Air Pressure Therapy System (Model:VU-IPC4M) falls under the category of a physical therapy or medical device used for external application, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

Air Pressure Therapy System (Model:VU-IPCAM) is intended for home to temporarily relieve minor muscle aches and or pains, and to temporarily increase circulation to the treated areas in people who are in good health.

Product codes (comma separated list FDA assigned to the subject device)

IRP

Device Description

Air Pressure Therapy System (Model:VU-IPC4M) consists of air pressure sensor, air pump, sleeves Etc. working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Leg (including of foot, calf, knee, upper leg)
Pants (including of upper leg, glutes, hips, lower back)
Arm (including of entire arm, shoulder)
Half-Leg (including of foot, calf, knee)

Indicated Patient Age Range

Not Found

Intended User / Care Setting

home

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Testing - Clinical: Not Applicable.
Performance Testing - Animal: Not Applicable.

Software Verification and Validation Testing:
The device also passed:

  • pressure accuracy and time accuracy testing
  • seam strength testing to ensure cuffs do not burst if maximum pressure is exceeded
  • failure mode verification and validation testing to ensure mitigation of overpressurization risk if there is a software failure

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K211745, K211283, K210913

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 890.5650 Powered inflatable tube massager.

(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

June 14, 2024

Xiamen Weiyou Intelligent Technology Co., Ltd. % Libray Chang Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609 Room, No. 133 Shengang Avenue, Pudong New District Shanghai, 201306 China

Re: K233238

Trade/Device Name: Air Pressure Therapy System (Model: VU-IPC4M) Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: May 23, 2024 Received: May 23, 2024

Dear Libray Chang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Heather L. Dean -S

Heather Dean, Ph.D Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices

2

OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K233238

Device Name

Air Pressure Therapy System (Model: VU-IPC4M)

Indications for Use (Describe)

Air Pressure Therapy System (Model:VU-IPCAM) is intended for home to temporarily relieve minor muscle aches and or pains, and to temporarily increase circulation to the treated areas in people who are in good health.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

Type of SubmissionTraditional
Date PreparedJune 11, 2024
Submission Sponsor
Manufacturer NameXiamen Weiyou Intelligent Technology Co., Ltd.
AddressUnit 3 No. 6 Xianghong Road, Torch Hi-Tech Zone
Industrial Park, Xiang'an District, Xiamen P.R.China.
Tel086-0592-6251545
Email17916220@qq.com
Contact PersonYong Sun
Device Identification
Trade NameAir Pressure Therapy System (Model:VU-IPC4M)
Regulation Number21 CFR 890.5650
Regulation NamePowered inflatable tube massager
Device ClassificationClass II
Product CodeIRP
PanelPhysical Medicine
Application Correspondent
Company NameShanghai Spica Management Consulting Co., Ltd.
Address609 Room, No.133 Shengang Avenue, Pudong New District, Shanghai, China
Tel86-13020102321
EmailLibray@spicagloble.com
Contact PersonLibray Chang
Indications for Use
Air Pressure Therapy System (Model:VU-IPC4M) is intended for home to temporarily relieve minor
muscle aches and/or pains, and to temporarily increase circulation to the treated areas in people who

Device Description

are in good health.

Air Pressure Therapy System (Model:VU-IPC4M) consists of air pressure sensor, air pump, sleeves

5

Etc. working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.

Predicate and Reference Device Information

Predicate (Primary):THERAGUN, Inc.
Trade/Device NameRecoveryAir PRO
510(K) numberK211745
Predicate (Secondary):Wonjin Mulsan Co., Ltd.
Trade/Device NameCompressible Limb and Circulation Therapy System
510(K) numberK211283
Predicate (Tertiary):SLK Medical GmbH
Trade/Device NameV12 PRO
510(K) numberK210913

Performance Testing - Clinical

Not Applicable.

Performance Testing - Animal

Not Applicable.

6

Table 6A: Summary of Comparison

| | Subject Device | Primary Predicate | Secondary Predicate | Tertiary Predicate | Differences
Discussion |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Air Pressure Therapy
System
(Model:VU-IPC4M) | RecoveryAir PRO (Model
RecoveryAir PRO) | Compressible Limb and
Circulation Therapy System | V12 PRO | N/A |
| 510(k) number | K233238 | K211745 | K211283 | K210913 | N/A |
| Manufacturer | Xiamen
Weiyou
Intelligent Technology
Co., Ltd. | Theragun Inc | Wonjin Mulsan Co., Ltd. | SLK Medical
GmbH | N/A |
| Product
regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
| Classification
name | Massager, Powered
Inflatable Tube | Massager, Powered Inflatable
Tube | Massager, Powered Inflatable
Tube | Massager, Powered
Inflatable Tube | Same |
| Regulation class | 2 | 2 | 2 | 2 | Same |
| Product code | IRP | IRP | IRP | IRP | Same |
| Indications for
use | Air Pressure Therapy
System
(Model: VU-IPC4M) is
intended for home to
temporarily relieve
minor muscle aches
and/or pains, and to
temporarily increase | The RecoveryAir PRO is an
air compression therapy
device intended to provide
graduated pressure to
compression garments.
The RecoveryAir PRO is
indicated for the temporary
relief of minor muscle aches | POWER-Q2300 is intended
for the temporary relief of
minor muscle aches and
pains, and for temporary
increase in blood circulation
to the treated areas in people
who are in good health.
POWER-Q2300 simulates | V12 PRO is a
Compression
Therapy System is
intended for the
temporary relief of
minor muscle aches
and pains and for
the temporary | Same |
| Rx or OTC | | | | | |
| | circulation to the
treated areas in people
who are in good health. | and pains, and for temporary
increase in blood circulation
to the treated areas in people
who are in good health.
The RecoveryAir PRO
simulates kneading and
stroking of tissues by using
an inflatable garment. | kneading and stroking of
tissues by using an inflatable
garment (cuff). | increase in
circulation to the
treated areas in
people who are in
good health.
The V12 PRO
Therapy System
simulates kneading
and stroking of
tissues by using an
inflatable garment. | Same |
| Rx or OTC | OTC | OTC | OTC | OTC | Same |
| Power Source | 11.1 V / 1600mAh
Rechargeable Li-ion
battery (100-240V AC
input) | 100-240V AC, 50/60 Hz,
12V or internal battery | 110-120VAC, 50/60Hz | 110-230V 50/60 Hz | Similar |
| Dimensions
(WHL) | 25.426.036.92CM | 8.6in (L) *6.7in (W) *5.1in
(H) | 290 x 260 x 172 mm | 6.69"x 7.87"x7.87" | Similar |
| Weight | 0.58 kg | 4.202 pounds | 3.1 kg | 4.4 pounds | Similar |
| Housing
Materials | Molded ABS enclosure | Molded ABS enclosure | Molded ABS enclosure | Molded ABS
enclosure | Same |
| Number of
Chambers | 4-chamber | 4-chamber | 4 Chambers | 12 Chambers | Similar |
| Sleeve Materials | Nylon cloth +PUether
amine | Polyether Nylon Fabric | / | Nylon with a
Polyurethane
laminate | Different |
| Modes | Mode A: In this mode | Sequential, ISO or Rehab | Mode A: | Inflates chambers | Different. |
| ((Inflation
sequences)) | | (wave), Flow cycles | | | |
| | ①chamber is inflated
and the pressure is
maintained after filling;
②chambers are
inflated, chambers are
inflated and ①
chamber is deflated
while maintaining
pressure after filling;
Similarly, the pressure
is maintained after ③
chambers are filled, ④
chambers are inflated,
and ② chambers are
deflated; Then the cycle
repeats. | Sequential mode that applies
a directional massage,
starting at the base of the
treated area, and progresses
upwards towards the torso
and then releases

ISO mode that applies a
directional massage to a
smaller, user selected area.
The first chamber inflates,
and after a few seconds, the
second chamber starts to
inflate until both chambers
reach the set pressure. Then
both chambers deflate,
and after a pause the process
starts again

Wave mode: The first
chamber inflates, and after a
few seconds, the 1st and 2nd
chamber starts to inflate until
both chambers
reach the set pressure. Then | Image: Mode B | from bottom up but
maintains pressure
in lower chambers
as works its way to
top. Then all
chambers release
pressure at same
time once all
chambers have
sequentially
inflated.
Sequential | Although the
subject device
provides 7 kinds
of work mode,
the Mode A and
Mode B are the
same with
Sequential, ISO
or Rehab (wave)
of Primary
predicate device
(K211745), while
the other work
modes of subject
device just have
difference about
inflatable order of
the different
chambers. The
treatment
pressure range are
the same under
different work
modes, so the
difference of
pressure range
would not raise |
| | Mode B: In this mode,
①chamber is inflated
and the pressure is
maintained after filling;
②chamber inflation;
the same as ③chamber
and ④chamber; Then
the cycle repeats. | | Image: Mode C | | |
| | Mode C: In this mode, | | Image: Mode D | | |

7

8

9

the four chambers arethe deflation on 1st chamber ,adversely impact
inflated at the samethe 2nd and 3rd chamberson safety and
time, the pressure isstarts to inflate until the botheffectiveness.
maintained after filling,chambers reach the set
and then the fourpressure; then deflate the 2nd
chambers are deflatedchambers inflate 3rd and 4th
at the same time;Afterchambers until the both
deflation, the fourchambers reach the set
pressure
chambers are inflated at
the same time;Then theFlow cycles
cycle repeats.Progress 1: first inflate
Chamber 1 to target pressure,
then hold & release. then go
to Progress 2.
Mode D: In this mode,Progress 2: first inflated
only a single chamberChamber 1 to target pressure,
is inflated at a time.theninflate Chamber 2 and
Starting from thehold Chamber 1. when
chamber① andChamber 2 reach the target
working up to thepressure, hold chamber 1 & 2
chamber④. Then thefor specified time, then
cycle repeats.release totally, then go to
Progress 3
Mode E: in this mode.Progress 3: first inflated
①③are inflated,thenChamber 1 to target pressure,
defalted, chamber ②then inflate Chamber 2 and
④are inflated,then
deflated;Then the cycle
repeats.
Mode F: in this mode.hold Chamber 1. when
①② are inflated, thenChamber 2 reach the target
defalted, chamber ③pressure, hold chamber 1 &
④ are inflated, then2, & start to inflate
deflated; Then the cycleChamber 3, when Chamber 3
repeats.reach to the target pressure,
then hold chamber 1, 2, 3 for
specified time then release
ModeG: In this mode,totally. Then go to Progress 4
chamber 1 inflates toProgress 4: first inflated
set pressure, it pulsesChamber 1 to target pressure,
and holds pressure;then inflate Chamber 2 and
Chamber 2 inflates tohold Chamber 1. when
set pressure, pulses andChamber 2 reach the target
holds pressure,pressure, hold chamber 1 &
chamber 1 is holding2, & start to inflate
pressure at the sameChamber 3, when Chamber 3
time; Chamber 3reach to the target pressure,
inflates to set pressure,then hold chamber 1, 2, 3,&
pulses and holdsstart to inflate Chamber 4,
pressure, chamber 1 &when Chamber 4 reach to the
2 are holding pressuretarget pressure, then hold
at the same time; Whilechamber 1, 2,
chamber 1 is3 &4 for specified time then
deflating, chamber 4release totally. then go back
inflates to set pressure,to Progress 1 again.
pulses and holds

10

11

Image /page/11/Figure/3 description: The image shows different modes of operation for a compression device. The modes include sequential cycle mode, sequential rehab (wave), wave cycle mode, and ISO cycle mode. The image also shows different programs (A, B, D, E, and F) with different patterns of chamber inflation and deflation.

12

Flow cycles
Image: Flow cycle mode
Anatomical siteLeg (including of foot,
calf, knee, upper leg)
Pants (including of
upper leg, glutes, hips,
lower back)
Arm (including of
entire arm, shoulder)
Half-Leg (including of
foot, calf, knee)RecoveryAir Compression
Boots: 95*28.7 cmLeg (including of foot, calf,
knee, upper leg)
Hip (including of upper leg,
glutes, hips, lower back)
Arm (including of entire arm,
shoulder)
Half-Leg (including of foot,
calf, knee)Leg (consisting of
foot, calf, knee,
upper leg) Pant
(consists of foot,
calf, knee, upper
leg, glutes, hips,
lower back)
Arm (consisting of
entire arm,
shoulder, upper
chest and back)
Jacket (consisting of
entire arms,
shoulder, upper
chest and
back)Similar

13

510(k) Summary K233238

Image /page/13/Figure/1 description: The image shows a comparison of different types of compression cuffs and pants. On the left, there are two sets of leg cuffs and a pair of pants, labeled with "Appearance and size of Cuffs". In the middle, there is a leg cuff with dimensions "Medium: 310x940mm" and "Large: 390x990", along with measurements of 310 and 940. On the right, there is a leg cuff with dimensions "32" x 30"", a pant, and the word "Similar".

14

Image /page/14/Figure/1 description: The image shows a collection of medical compression garments. There are pants, arm sleeves, a waist wrap, hip wraps, and a jacket. Some of the items have dimensions listed, such as the arm sleeves which are available in medium (255x870mm) and large (330x895mm).

15

| Image: black sleeve | | Half-Leg
Image: Half-Leg device with dimensions 300 and 700
300x700mm | | | |
|-----------------------|------------|-----------------------------------------------------------------------------|-------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Pressure range | 30~180mmHg | 20-100mmHg | 20-200 mmHg | 15-80 mmHg | Similar to
K211283,
Although the
pressure value is
different from
that of the
Primary Predicate
and Tertiary
Predicate, it is
included in the
pressure value
range of the
Secondary |

16

| | | | | | | | Predicate, so the
pressure value of
the Subject
Device can be
proven to not
affect its safety
and effectiveness |
|----------------|---------------------|------------|---------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------------|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Treatment Time | Default
Settings | 30min, APP | 10min-90min, step of 5min | User can select operation
time from 15 and 30 minutes. | User determines
therapy time. Choose
from 15 to 60
minutessession
time. | | Different,
Although the
treatment time
range of subject
device is set by
APP, but the
default value is
30min which is
included in the
range of the
comparison
equipment. In the
process of use,
the user can start
or stop the button
on the hand
controller at any
time, so the
difference of
Treatment time |

17

| | | | | | would not raise
adversely impact
on safety and
effectiveness. |
|--------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Hold time within
cycle | 30-180 sec | 2-10 sec. | 1 min 20 sec | 1 min 15 sec | Similar |
| Pause interval -
between cycles | 15 sec | 10 - 70 sec | Adjustable at interval of
0/5/10/30 sec at modes A and
B
Fixed at 30 sec at modes C
and D | / | Similar |
| Mobile
application | Bluetooth
communication | Bluetooth communication | / | / | Same |
| Patient contact | Non-conductive
appliances | Non-conductive appliances | / | / | Same |
| Software/Firmw
are/Microproces
sor Control | Microprocessor | Microprocessor | / | Firmware/Micropro
cessor | Similar |
| Technology | Compressor and valve
system which
sequentially inflates
inflatable chambers | Compressor and valve system
which sequentially inflates
cells of appliance | Compressor and valve
system which sequentially
inflates cells of appliance | Compressor
and
valve system which
sequentially inflates
cells sequentially | Same |
| Electrical safety
EMC | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11 | IEC 60601-1:2005;
IEC 60601-1-2:2014;
IEC60601-1-11:2015 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11 | Same |
| Biocompatibility | ISO 10993-5
ISO 10993-10 | ISO 10993-5
ISO 10993-10 | ISO 10993-5
ISO 10993-10 | ISO 10993 | Same |

18

Performance Characteristic

The device meets all the applicable technical requirements of :

IEC 60601-1:2005 - Medical electrical equipments - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-11: 2015 - Medical electrical equipments for Basic Safety and Essential Performance - Collateral Standard: Requirements for medical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-2: 2014 - Medical electrical equipments for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility

IEC 62133-2: 2017 - Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

ISO 10993-10: 2010 - Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization

ISO 10993-5:2009 - Biological Evaluation of Medical Device - Part 5: Tests for in vitro Cytotoxicity

IEEE ANSI USEMCSC C63.27-2021-American National Standard for Evaluation of Wireless Coexistence

Software Verification and Validation Testing

The device also passed:

  • pressure accuracy and time accuracy testing i.
  • ii. seam strength testing to ensure cuffs do not burst if maximum pressure is exceeded
  • iii. failure mode verification and validation testing to ensure mitigation of overpressurization risk if there is a software failure

Conclusion

Based on the indications for use, technological characterstics, and non-clinical performance data, "Air Pressure Therapy " is as safe, as effective, and performs as well as the legally marketed predices. Therefore, the substantially equivalent to the predicate device.