K Number
K233120

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-05-23

(239 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for use in dermatologic and general surgical procedures for electro coagulation.

Device Description

This device is a Fractional RF Surgical Unit with Sterile Micro Needle, which is composed of the main device (generator and user interface display, handpiece, foot switch, neutral electrode, and monopolar microneedle electrodes. It is designed for applying radiofrequency therapy with microneedles, which coagulates skin tissue using 1 MHz or 2 MHz radiofrequency energy delivered from the generator to the microneedles. A Bipolar Handpiece suction filter is also available for use to remove unwanted debris from the skin application site and safeguarding the equipment by preventing foreign substances from entering during use.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (CWM-930S Radio Frequency Therapy System). It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria based on studies involving AI or human reader performance.

Therefore, the document does not contain the information requested regarding acceptance criteria for AI device performance, sample size for test sets (in the context of AI models), data provenance, number and qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone AI performance, type of ground truth used for AI models, training set sample size, or how ground truth for training was established.

The "Performance Testing" section does mention some non-clinical performance tests for the CWM-930S device itself, but these are not related to AI model performance or human reader studies. Specifically, it mentions:

  • Non-Clinical Test Summary [21 CFR 807.92(b)(1)]:

    • Electrical Safety, Electromagnetic Compatibility Testing: Compliance with IEC 60601 series standards.
    • Software Validation: CWM-930S has a MODERATE level of concern software that was designed, developed, verified, and validated according to FDA guidance.
    • Biocompatibility: Tested parts include the Cap, Electrode (needle and coating) made of SUS, and Neutral Electrode composed of Hydrogel, Cloth, Aluminum Film, Transparent Film, and Loctite Dura-Tak 129a. All showed biocompatibility for intact skin contact for limited duration (< 24 hours).
    • Sterilization, Shelf-life Testing: Ethylene Oxide Sterilization Test according to ISO 11135:2014/Amd 1:2018.
    • Performance Test: Validated output characteristics by load resistance, frequency, amplitude, duty cycle, and crest factor.
    • Thermal effect: Ex vivo animal testing using porcine abdominal skin to obtain histological data of values for depth and zone of coagulation and thermal damage immediately post-treatment at low, mid, and high intensity. The study concluded that CWM-930 showed similar performance to predicate devices (VIVACE and SILKRO) regarding thermal effect.
  • Clinical Test Summary [21 CFR 807.92(b)(2)]: States "No clinical studies were considered necessary and performed." This further confirms that no human reader or AI performance studies were conducted as part of this submission.

In summary, the provided document is a regulatory submission for an electrosurgical device, not an AI or imaging diagnostic device. Therefore, the information requested about AI performance acceptance criteria and study details is not present.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 23, 2024

Chungwoo Co., Ltd. Su Jin Lee Regulatory Affair Manager 614. 2. Gasadigital 1-ro Guemcheon-gu, Seoul 13201 Korea, South

Re: K233120

Trade/Device Name: CWM-930S Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 27, 2023 Received: September 27, 2023

Dear Su Jin Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2024.05.23 11:24:41 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233120

Device Name CWM-930S Radio Frequency Therapy System

Indications for Use (Describe)

This device is intended for use in dermatologic and general surgical procedures for electro coagulation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea

TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

K233120

510(k) Summary

[As required by 21 CFR 807.92]

Date Prepared [21 CFR 807.92(a)(a)] 1.

May 22, 2024

Submitter's Information & Contact Person [21 CFR 807.92(a)(1)] 2.

- Name of Manufacturer:Chungwoo Co., Ltd.
- Address:614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea
- Contact Name:Su Jin LEE
- Telephone No.:+82-2-2027-2200
- Fax No .:+82-2-2027-2207
- Email Address:cwra3@mycw.co.kr

Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] 3.

Common name: Radio Frequency Therapy System Trade name: CWM-930S

Classification Description21 CFR SectionProduct Code
Electrosurgical, Cutting and Coagulation and Accessories878.4400GEI

As stated in 21 CFR, parts 878.4400, this generic type of device has been classified as Class II.

{5}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea

TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

The identified predicate devices within this submission are shown as follow:

Predicate device 1

  • 510(k) Number: K210084
  • Applicant: Hironic Co., Ltd
  • electrosurgical, cutting & coagulation & accessories Classification Name: ●
  • Trade Name: SILKRO

Predicate device 2

  • K193070 ● 510(k) Number:
  • Applicant: ShenB Co Ltd.
  • Classification Name: electrosurgical, cutting & coagulation & accessories
  • Trade Name: VIVACE Electrosurgical System

5. Description of the Device [21 CFR 807.92(a)(4)]

This device is a Fractional RF Surgical Unit with Sterile Micro Needle, which is composed of the main device (generator and user interface display, handpiece, foot switch, neutral electrode, and monopolar microneedle electrodes. It is designed for applying radiofrequency therapy with microneedles, which coagulates skin tissue using 1 MHz or 2 MHz radiofrequency energy delivered from the generator to the microneedles. A Bipolar Handpiece suction filter is also available for use to remove unwanted debris from the skin application site and safeguarding the equipment by preventing foreign substances from entering during use.

6. Indications for Use [21 CFR 807.92(a)(5)]

This device is intended for use in dermatologic and general surgical procedures for electro coagulation.

{6}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea

TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]

There are no significant differences between the CWM-930S and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.

Proposed DevicePredicate Device 1Predicate Device 2
K Number-K210084K193070
ManufacturerChungwoo Co., Ltd.Hironic Co., LtdShenB Co Ltd.
ModelCWM-930SSILKROVIVACEElectrosurgical Device
Intended UseThis device is intended for use indermatologic and general surgicalprocedures for electro coagulation.This device is intended for use indermatologic and general surgicalprocedures for electro coagulation.The VIVACEElectrosurgical System is intended foruse indermatologic and general surgicalprocedures for electrocoagulation andhemostasis and the percutaneoustreatment of facial wrinkles for use withFitzpatrick Skin Type I to Skin Type Vwhen using the 1MHz setting. The2MHz setting has not been evaluated foruse in the percutaneous treatment offacial wrinkles and is intended for use indermatologic and general surgicalprocedures for electrocoagulation andhemostasis.
Mode ofoperationBipolar Handpiece + Microelectrodes, MonopolarBipolar Handpiece + Microelectrodes, MonopolarBipolar Handpiece + Microelectrodes
OperatingFrequency1MHz, 2MHz2MHz1 MHz, 2 MHz
Rated InputAC 100- 240V,50/60HzAC 100- 240V,50/60HzAC 120V 50/60 Hz
Output1Mhz: 35W±20% (500Ω)25W ± 20%, 500Ω1MHz: 36W±10%(load resistance 500Ω)
Proposed DevicePredicate Device 1Predicate Device 2
2Mhz : 25W±20% (500Ω)2MHz : 23.3 W(load resistance 500Ω)
RF Intensity1 ~ 10Level1 ~ 10 Level1-10 Level
RF Duration10ms~900ms50 ms ~ 950 ms100ms-800ms (100ms increments)
Treatment Time5 ~ 15 min5 ~ 15 min-
Needle insertdepthBipolar :0.5mm3.5mm(14pin, 25pin, 49pin) /0.5mm7.0mm(36pin) ±15% / Adjustmentunit 0.1mmmonopolar : 0.5mm~5.0mm (0.1mmstep)Bipolar : 0.5 ~ 3.5 mm(0.1 step)monopolar : 1.2,1.5, 1.8 mm0.5~ 3.5mm (0.1mm step)
Number of pinsBipolar : RF Microneedle 14pin,25pin,36pin, 49pin monopolar : 1pinBipolar : RF Microneedle 25pin, 49pinmonopolar : 1pin36 needle electrode
Needleapplicationrange1pin(1.0 x 1.0mm), 25Pin, 49Pin(7.8 x7.8mm), 14pin(7.8 x 2.0mm),36pin(11x11mm)25pin(8x8mm), 49mm(12x12mm)-
SterilizationEO GasEO GasSterility Assurance Level (SAL) of 10-6
BiocompatibilitytestedYesYesYes

{7}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

{8}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

Non-Clinical Test Summary [21 CFR 807.92(b)(1)]

  1. Electrical Safety, Electromagnetic Compatibility Testing

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

Standard(Edition)Standard Title
IEC 60601-1:2005/AMD2:2020Medical electrical equipment - Part 1: General requirements for basicsafety and essential performance
IEC60601-1-6:2010Medical electrical equipment - Part 1-6: General requirements for basicsafety and essential performance - Collateral standard: Usability
IEC 60601-2-2:2017Medical electrical equipment - Part 2-2: Particular requirements for thebasic safety and essential performance of high frequency surgicalequipment and high frequency surgical accessories
IEC 60601-1-2:2014 +A12020Medical electrical equipment - Part 1-2: General requirements for basicsafety and essential performance - Collateral Standard: Electromagneticdisturbances - Requirements and tests

2) Software Validation

The CWM-930S contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on June 14, 2023.

3) Biocompatibility

PartMaterialPatientContactDuration of Contactby ISO 10993-1Bio-compatibility
Cap,Electrode(needleand coating)SUSIntact SkinLimited(< 24 hours)Yes
Neutral Electrode- Hydrogel- Cloth- Aluminum Film- Transparent Film- Loctite Dura-Tak129aIntact SkinLimited(< 24 hours)Yes

{9}------------------------------------------------

614, 2, Gasandigital1-ro, Geumcheon-gu, Seoul, Republic of Korea TEL: +82 2-2027-2200 FAX: +82 2-2027-2207

  • · Sterilization, Shelf-life Testing
    Ethylene Oxide Sterilization Test: ISO 11135:2014/Amd 1:2018, Sterilization of health care products -- Ethylene oxide -- Requirements for development, validation and routine control of a sterilization process for medical devices

    1. Performance Testing

• Performance test

The purpose of testing is to validate the performance of this device and ensure that the performance effectiveness of this device is substantially equivalent to predicate device.

  • Output characteristics by load resistance
  • Frequency
  • Amplitude
  • Duty cycle
  • Crest factor

· Thermal effect

Ex vivo animal testing using porcine abdominal skin was also conducted to obtain histological data of values for depth and zone of coagulation and thermal damage immediately post treatment was performed at the intensity(power) low, mid, high. Based on this animal test, it was confirmed through mechanical and histological evaluation that micro needle can affect tissue by thermal effect of each output condition.

The results of the tests show that CWM-930 and predicate devices show similar performance in regards to the Thermal effect. Therefore, this study concludes that the CWM-930 is equivalent to the predicate device VIVACE and SILKRO.

Clinical Test Summary [21 CFR 807.92(b)(2)]

No clinical studies were considered necessary and performed.

Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Chungwoo Co., Ltd. concludes that the CWM-930S is substantially equivalent to and is as safe and effective as the predicate devices as described herein.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.