(85 days)
Sterile and Non-Sterile Ultrasound Gels is intended to be used on external, intact skin as a coupling contact medium for ultrasonic procedures. The sterile Ultrasound Gel is also used for intra operative, biopsy applications. The Ultrasound Gel can be used during the procedures that involved adults and pediatrics in professional healthcare facility by qualified and trained healthcare professional.
The proposed device, Sterile and Non-Sterile Ultrasound Gels, are non- irritating, nonsensitizing acoustic couplants intended for use as a scanning gel in medical diagnostic ultrasound procedures. The gel is an accessory used to couple sound waves between the patient and the medical imaging electronic transducers during diagnostic ultrasound procedures. Ultrasound gel is a type of conductive medium that enables a tight bond between the skin and the probe or transducer, allowing the waves to transmit directly to the tissues beneath and to the parts that need to be imaged. As such, the gel is formulated to improve the acoustic transmission of sound waves to create the ultrasound image.
The provided text describes the 510(k) premarket notification for "Sterile and Non-Sterile Ultrasound Gels" by Jianerkang Medical Co., Ltd. This document is a regulatory submission for a medical device accessory, specifically an ultrasound coupling gel, not an AI software or a device that has complex diagnostic or interventional functions.
Therefore, many of the requested elements for AI/imaging device performance studies (like MRMC studies, effect size of human reader improvement with AI, ground truth establishment for training/test sets, number of experts for adjudication, etc.) are not applicable to this type of device and submission.
The acceptance criteria for this device are based on demonstrating substantial equivalence to a legally marketed predicate device (HR Pharmaceuticals, Inc., EcoVue® Sterile and Non-Sterile Ultrasound Gels, K181363) through bench testing, biocompatibility testing, sterilization validation, and packaging/shelf-life validation. The "study that proves the device meets the acceptance criteria" refers to these non-clinical performance tests.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are primarily based on demonstrating performance similar to or within acceptable ranges compared to the predicate device, and compliance with recognized standards.
| Acceptance Criteria (Requirement Specification) | Reported Device Performance (Test Results) |
|---|---|
| Physical & Chemical Properties (Measured at 35°C): | |
| Sound Velocity (Acoustic Velocity): 1520 ~ 1620 m/s | 1569 m/s ~ 1570 m/s |
| Acoustic Impedance: $1.5 \times 10^6$ ~ $1.7 \times 10^6$ Pa·s/m | $1.593 \times 10^6$ Pa·s/m ~ $1.597 \times 10^6$ Pa·s/m |
| Sound Attenuation: ≤ 0.1 dB/(cm·MHz) | 0.042 dB/(cm·MHz) ~ 0.049 dB/(cm·MHz) |
| Physical Properties (Measured at 25°C): | |
| Viscosity: ≥ 4000 cP (1000 Cp = 1 Pa.s) | 8844 cP ~ 13720 cP |
| Density: 0.85 - 1.15 g/cm³ | 1.021 - 1.022 |
| pH: 5.5 ~ 8.0 | 6.68 ~ 6.73 |
| Compositional/Visual Inspection: | |
| Appearance: Colorless or light transparent gel, no or only a few bubbles, non-soluble foreign bodies | Pass |
| Microbiological Purity (Non-Sterile Gel): | |
| TAMC: ≤ 100 CFU/g | < 10 CFU/g |
| TYMC: ≤ 100 CFU/g | < 10 CFU/g |
| Sterility (Sterile Gel): | |
| Product should be sterile after being sterilized by confirmed gamma sterilization. | Pass (Confirmed by validation per ISO 11137-1, -2, -3) |
| Biocompatibility: | |
| Meets ISO 10993-1:2018 requirements for irritation, sensitization, and cytotoxicity. | Acceptable results (Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Pyrogen tests per ISO 10993 series) |
| Packaging and Shelf Life: | |
| Packaging integrity to maintain sterility and product quality for 3 years. | Passed packaging validation and accelerated aging testing confirming 3-year shelf life. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify exact sample sizes for each bench test (e.g., how many batches were tested for viscosity or pH) but rather reports ranges or single values, implying sufficient testing was performed to establish these properties. For biocompatibility, the tests (e.g., cytotoxicity, irritation) are typically performed on a statistically relevant number of samples (e.g., extracts from the product) as per ISO standards, which is implied by the "acceptable results" statement. For sterilization validation, specific sample sizes are required by ISO 11137, though not explicitly stated in the summary.
- Data Provenance: The data is derived from non-clinical bench testing conducted by Jianerkang Medical Co., Ltd. No specific country of origin for the data is mentioned for testing labs, but the manufacturer is based in China. The data is prospective in the sense that it was generated specifically for this 510(k) submission to demonstrate the device's properties.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable. For a device like ultrasound gel, "ground truth" as it relates to clinical outcomes or image interpretation (which would typically involve experts like radiologists) is not relevant. The "ground truth" in this context is the physical, chemical, and biological properties of the gel itself, established by standardized laboratory testing methods and measurements. These tests are performed by trained laboratory personnel, not clinical experts for diagnostic interpretation.
4. Adjudication Method for the Test Set
- Not Applicable. As the "test set" consists of laboratory measurements and not clinical image interpretations, there is no need for an adjudication method. Test results are objective quantitative measurements or qualitative observations (e.g., "Pass" for appearance).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. An MRMC study is relevant for evaluating the impact of an AI algorithm on human reader performance, typically in diagnostic imaging. This device (ultrasound gel) is an accessory and does not involve image interpretation or AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This device is an ultrasound gel, not an algorithm.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance is established through objective, standardized measurement of its physical, chemical, and biological properties as defined by relevant ISO and ASTM standards (e.g., sound velocity, pH, sterility, biocompatibility). It is not based on expert consensus, pathology, or outcomes data, as these are clinical validation methods for diagnostic/therapeutic devices, which an ultrasound gel is not.
8. The Sample Size for the Training Set
- Not Applicable. There is no "training set" in the context of an ultrasound gel. This is not an AI/machine learning device. The product formulation development (analogous to "training") would involve R&D and quality control processes to achieve the desired properties, but this is not a data-driven training process in the AI sense.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no "training set" as understood in AI/machine learning, this question doesn't apply. The "ground truth" for developing the gel's formulation is the target specification for its physical, chemical, and biological properties, informed by regulatory requirements for safety and effectiveness as an ultrasound coupling medium.
{0}------------------------------------------------
December 15, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The seal is a stylized depiction of an eagle and a caduceus, while the FDA acronym and name are written in blue, with the word "Administration" appearing below the acronym and name.
Jianerkang Medical Co., Ltd % Gu Mike RA Manager Suzhou Device Innovation Medical Consulting Co., Ltd Room 1001, Building 19, No. 3188 Renmin Road Suzhou, Jiangsu 215000 CHINA
Re: K232957
Trade/Device Name: Sterile and Non-Sterile Ultrasound Gels Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: MUI Dated: September 21, 2023 Received: September 21, 2023
Dear Gu Mike:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device. or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}------------------------------------------------
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K232957
Device Name Sterile and Non-Sterile Ultrasound Gels
Indications for Use (Describe)
Sterile and Non-Sterile Ultrasound Gels is intended to be used on external, intact skin as a coupling contact medium for ultrasonic procedures. The sterile Ultrasound Gel is also used for intra operative, biopsy applications. The Ultrasound Gel can be used during the procedures that involved adults and pediatrics in professional healthcare facility by qualified and trained healthcare professional.
| Type of Use (Select one or both, as applicable) |
|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) SUMMARY For K232957 Sterile and Non-Sterile Ultrasound Gels
I. SUBMITTER:
Jianerkang Medical Co., Ltd. No.1 Jianerkang Road, Zhixi Town Industrial Zone, Jintan District, Changzhou City, Jiangsu Province, China
Telephone Number: 86-519-82446609 Fax Number: 86-519-82442535
Contact Person: Hongfang Tang Date Prepared: 09/30/2023
II. DEVICE: Name of Device: Sterile and Non-Sterile Ultrasound Gels Common or Usual Name: Media, Coupling, Ultrasound Classification Name: Diagnostic ultrasonic transducer (21 CFR 892.1570) Regulatory Class: II Product Code: MUI
III. PREDICATE DEVICES:
HR Pharmaceuticals, Inc., EcoVue® Sterile and Non-Sterile Ultrasound Gels, K181363
No reference devices were used in this submission.
IV. DEVICE DESCRIPTION:
The proposed device, Sterile and Non-Sterile Ultrasound Gels, are non- irritating, nonsensitizing acoustic couplants intended for use as a scanning gel in medical diagnostic ultrasound procedures. The gel is an accessory used to couple sound waves between the patient and the medical imaging electronic transducers during diagnostic ultrasound procedures.
Ultrasound gel is a type of conductive medium that enables a tight bond between the skin and the probe or transducer, allowing the waves to transmit directly to the tissues beneath and to the parts that need to be imaged. As such, the gel is formulated to improve the acoustic transmission of sound waves to create the ultrasound image.
{5}------------------------------------------------
The proposed device is available in the below sizes:
- 2.20Z (non sterile) ● Tubes
- 20ml (sterile) Pouches
- Bottles 8.5OZ, 5000ml (non sterile) ●
V. INDICATIONS FOR USE:
Sterile and Non-Sterile Ultrasound Gels is intended to be used on external, intact skin as a coupling contact medium for ultrasonic procedures. The sterile Ultrasound Gel is also used for intra operative, biopsy applications. The Ultrasound Gel can be used during the procedures that involved adults and pediatrics in professional healthcare facility by qualified and trained healthcare professional.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES:
The subject and predicate devices have the same technological principle whereby, the gel is an accessory to the medical imaging equipment acting as a conductive medium that promotes a tight bond between the skin and the transducer, allowing the waves to transmit directly to the body tissues beneath and to the parts that need to be imaged. As such, ultrasound gels are formulated to act as a coupling agent and reduce static.
At a high level, the proposed Ultrasound Gels and the predicate devices are based on the following same elements:
- Same Intended Use ●
- Same Technological Characteristics ●
- Similar Physical Characteristics ●
- Similar Performance Specifications
- Packaged in similar quantities ●
- Provided in both sterile and non-sterile product offerings ●
- Biocompatible
- Water-based gels ●
Additionally, side-by-side testing confirmed that the subject devices are substantially equivalent to the predicate devices. A more detailed comparison of the predicate and subject devices is presented in the table below.
{6}------------------------------------------------
| Proposed DeviceEcoVue® Sterile and Non-Sterile UltrasoundGels | Predicate DeviceSheathes Sterile Ultrasound Gel (K181363) | ||
|---|---|---|---|
| Manufacturer | Jianerkang Medical Co., Ltd | HR Pharmaceuticals, Inc. | |
| Device Name | Media, Coupling, Ultrasound | Media, Coupling, Ultrasound | |
| Description | Diagnostic ultrasonic transducer(accessory) | Diagnostic ultrasonic transducer(accessory) | |
| Medical Specialty | Radiology | Radiology | |
| Product Code | MUI | MUI | |
| Reg Number | 892.1570 | 892.1570 | |
| Class | 2 | 2 | |
| Indications for Use | Sterile and Non-Sterile Ultrasound Gels is intendedto be used on external, intact skin as a couplingcontact medium for ultrasonic procedures. Thesterile Ultrasound Gel is also used for intraoperative, biopsy applications. The Ultrasound Gelcan be used during the procedures that involvedadults and pediatrics in professional healthcarefacility by qualified and trained healthcareprofessional. | EcoVue® Sterile and Non-Sterile Ultrasound Gelsare ultrasound couplants intended to be used on intactskin during non-invasive medical ultrasoundprocedures to couple sound waves between a patientand the medical imaging electronics. The gels areintended for use in all diagnostic ultrasoundprocedures which require coupling gel or fluid. | |
| Patient-contacting material | Water-based gel (sterile and non-sterile) | Water-based gel (sterile) | |
| Sizes | Tubes 2.2OZ (non sterile)Pouchs 20ml (sterile)Bottle 8.5OZ, 5000ml (non sterile ) | 20g individual packets (both sterile and non-sterile)250g pouch (non-sterile) | |
| PhysicalandChemicalProperties | Appearance | Colorless or light transparent gel, no or only a fewbubbles, non-soluble foreign bodies | Clear to Hazy Color; free from foreignmatter |
| SoundVelocity(AcousticVelocity) | 1520~1620m/s | 1398-1750 m/s | |
| Proposed DeviceEcoVue® Sterile and Non-Sterile UltrasoundGels | Predicate DeviceSheathes Sterile Ultrasound Gel (K181363) | ||
| AcousticImpedance | $1.5×10^6~1.7×10^6$ | 1.40-1.80 Mrayl | |
| SoundAttenuation | ≤0.1dB/(cm.MHz) | 0.32-0.95 dB/cm at 5 MHz0.65-1.10 dB/cm at 7.5 MHz0.85-1.55 dB/cm at 10 MHz | |
| Viscosity | ≥4000 cP (1000 Cp=1Pa.s) | >35,000 cP | |
| Density | 0.85-1.15 g/cm³ | 0.85-1.15 g/cm³ | |
| pH | 5.5-8 | 5.5 - 7.8 | |
| Biocompatibility | Meets ISO 10993-1:2018 requirements for irritation,sensitization, and cytotoxicity | Meets ISO 10993-1:2009 requirements for irritation,sensitization, and cytotoxicity | |
| Utility | Single-Use (20ml pouch, sterile packet)Multi-Use (2.2OZ tube, non-sterile packet; 8.5OZ,5000ml bottles, non-sterile packet ) | Single-Use (20g sterile and non-sterile packets)Multi-Use (250g pouch) | |
| Sterility (Method) | Sterile (Gamma);Non-Sterile | Sterile (Gamma);Non-Sterile | |
| Environment of Use | Healthcare Facility | Healthcare Facility | |
| Target Population | Adult and Pediatric | Adult and Pediatric | |
| Labeling | Prescription Use | Prescription Use | |
| Shelf Life | 3 year | 1 year |
{7}------------------------------------------------
- Values obtained from performance testing conducted per HR Pharmaceuticals' protocols.
{8}------------------------------------------------
VII. PERFORMANCE DATA
There are no FDA device-specific guidance documents, special controls document, and/or requirements in a device-specific regulation for acoustic coupling gels. Additionally, no performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for acoustic coupling gels as they are accessories to diagnostic ultrasonic medical devices.
Summary of non-clinical and performance testing Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 10993-1, ISO 11137-1, ISO 11607-1. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.
| Test Item | Description | Test Results |
|---|---|---|
| Sound Velocity(AcousticVelocity) | Measured at 35°C ,The Sound Velocity(Acoustic Velocity) shall be 1520 ~ 1620m/s | 1569m/s~1570m/sPass |
| AcousticImpedance | Measured at 35°C ,Acoustic Impedanceshall be $1.5\times10^6$ ~ $1.7\times10^6$ Pa·s/m | $1.593\times10^6$ Pa·s/m~ $1.597\times10^6$ Pa·s/mPass |
| SoundAttenuation | Measured at 35°C ,Sound Attenuation shallbe ≤0.1dB/(cm·MHz) | 0.042dB/(cm·MHz)-0.049 dB/(cm·MHz)Pass |
| Viscosity | Measured at 25°C, ≥4000cP(1000Cp=1Pa.s) | 8844 cP ~13720 cPPass |
| Density | The density of the product shall be0.85-1.15 | 1.021-1.022Pass |
| pH | The pH value of the product should be5.5~8.0 | 6.68~6.73Pass |
| Sterility | The product should be sterile after beingsterilized by confirmed gamma sterilization. | Pass |
| MicrobiologicalGrowth | TAMC≤100CFU/gTYMC≤100CFU/g | TAMC<10CFU/gTYMC<10CFU/gPass |
| Appearance | Colorless or light transparent gel, no or onlya few bubbles, non-soluble foreign bodies | Pass |
Bench Testing
Biocompatibility Testing
The biocompatibility evaluation for the proposed Ultrasound Gel was conducted in
{9}------------------------------------------------
accordance with the FDA Final Guidance "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" September 04, 2020, and International Standard ISO 10993-1:2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatibility testing included the following and yielded acceptable results:
- . Cytotoxicity - ISO 10993-5:2009 (R) 2014 Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity
- Irritation ISO 10993-23:2021-Biological evaluation of medical devices -● Part 23: Tests for irritation
- . Sensitization - ISO 10993-10:2021—Biological Evaluation of Medical Devices-Part 5: Tests for Irritation and Skin Sensitization
- Acute Systemic Toxicity -ISO 10993-11 : 2017 Biological evaluation of ● medical devices - Part 11: Tests for systemic toxicity
- Pyrogen -ISO 10993-11 : 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
Ultrasound Gels fall into the category of limited (<24 hour) contact surface devices and are intended for direct contact with intact skin.
Sterilization
Sterile Ultrasound Gel was validated for sterilization using gamma irradiation to achieve a Sterility Assurance Level of 106 in accordance with the following recognized standards:
- . ANSI/ AAMI/ISO 11137-1:2006/(R) 2010. Sterilization of health care products -Radiation-Part1: Requirements for development validation, and routine control of a sterilization process for medical devices.
- ANSI/AAMI/ISO 11137-2:2013. Sterilization of health care products-● Radiation-Part 2: Establishing the sterilization dose.
- ANSI/AAMI/ISO 11137-3:2006 (R) 2010. Sterilization of healthcare products -● Radiation-Part 3: Guidance on dosimetric aspects.
- ANSI/AAMI/ISO 11737-1: 2006/(R) 2011. Sterilization of health care products, . Microbiological Methods-Part 1: Determination of a population of microorganisms on product.
Packaging Validation and Shelf Life
Sterile and Non-Sterile Ultrasound Gels underwent packaging validation testing and accelerated aging testing per the below standards in order to support three years shelf life.
- AAMI/ ISO 11607-1: 2006 (R) 2010, Packaging for terminally sterilized medical ● devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems.
{10}------------------------------------------------
- ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials.
- ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of ● Seals for Flexible Packaging by Visual Inspection.
- . ASTM F2096-11, Standard Test Method for Detecting Gross Leaks in Porous Medical Packaging by Internal Pressurization (Bubble Leak).
- ASTM F1980-22, Standard Guide for Accelerated Aging of Sterile Barrier ● Systems for Medical Devices.
VIII. CONCLUSION
All necessary testing has been performed on the Jianerkang Sterile and Non-Sterile Ultrasound Gels to support substantial equivalence. Additionally, the performance testing concluded that the subject devices are substantially equivalent to the legally marketed predicate devices.
§ 892.1570 Diagnostic ultrasonic transducer.
(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.