K Number
K232934
Device Name
Sunsphere
Date Cleared
2024-05-30

(253 days)

Product Code
Regulation Number
870.3300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sunsphere is intended to be used for the embolization of hypervascular tumors, including uterine fibroids, and arteriovenous malformations (AVMs).

Device Description

This proposed device is composed of polyvinyl alcohol embolic microspheres stored in a preservation solution of 0.9% sodium chloride. The proposed devices are compressible hydrogel microspheres with a regular shape, smooth surface, and calibrated size, which are formed as a result of chemical modification on polyvinyl alcohol (PVA) materials. Sunspheres are chemically cross-linked to form the microsphere. Microspheres are round in appearance with uniform size distribution and are available in blue dyed and undyed (colorless) microspheres.

The proposed device available in blue dyed and undyed (colorless) version, which are available in 1 mL and 2 mL volumes, and 6 different calibrated particle sizes from 90 to 1000 um. Different particle sizes are distinguished by the color of the matching vial cap. The proposed device is supplied sterile and packaged in sealed glass vials.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Sunsphere vascular embolization device, and as such, it does not contain the level of detail typically found in a clinical study report for an AI/ML medical device, which would include specific acceptance criteria and performance data for an algorithm.

The provided text focuses on the substantial equivalence determination for a physical medical device (embolization microspheres) by comparing its technological characteristics and non-clinical performance (biocompatibility, physical properties, animal studies) to a predicate device. It does not present information on an AI/ML driven device's performance, nor does it contain a clinical study with human readers (MRMC study) or standalone algorithm performance data.

Therefore, I cannot extract acceptance criteria or study details related to an AI/ML device from this document. The document describes a traditional medical device clearance process for a physical vascular embolization device.

If you are looking for an example of how an AI/ML device's acceptance criteria and study proving its performance are described in an FDA submission, you would need a document from a software as a medical device (SaMD) or an AI-enabled medical device submission, which would typically include sections on algorithm performance, clinical validation, and ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

May 30, 2024

Hangzhou Yangshun Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, Shanghai 200120 CHINA

Re: K232934

Trade/Device Name: Sunsphere Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: II Product Code: KRD, NAJ Received: April 29, 2024

Dear Diana Hong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

The FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is

{1}------------------------------------------------

consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Reginald K. Avery -S

for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K232934

Device Name Sunsphere

Indications for Use (Describe)

Sunsphere is intended to be used for the embolization of hypervascular tumors, including uterine fibroids, and arteriovenous malformations (AVMs).

Type of Use (Select one or both, as applicable)
✓ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Page 1 of 10

510(k) Summary

Sunsphere

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K232934

    1. Date of Preparation: May 30, 2024

Sponsor Identification 2.

Sponsor Name:Hangzhou Yangshun Medical Technology Co. Ltd.
Address:Room 110, Building 2, No.357 Miaohouwang Road, Xixing SubdistrictBinjiang District, Hangzhou, Zhejiang, China
Contact Person:Dong Hui
Position:Registration Director
Phone:+86-13716835300
Email:dongh@sunspheremedical.com

3. Designated Submission Correspondent

Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Phone: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

4. Identification of Proposed Device

Trade Name: Sunsphere Common Name: Polyvinyl Alcohol Embolic Microspheres

Regulatory Information Regulation Name: Vascular Embolization Device Regulation Number: 21 CFR 870.3300

{5}------------------------------------------------

Device Classification: Class II

Product Code: KRD (Device, Vascular, For Promoting Embolization), NAJ (Agents, Embolic, For Treatment Of Uterine Fibroids)

5. Identification of Predicate Device

Device Name: CalliSpheres Embolic Microspheres, 8Spheres Embolic Microspheres 510(k) Number: K173871 Manufacturer: Suzhou Hengrui Callisyn Biomedical Co., Ltd Regulation Number 21 CFR 870.3300 Product Code: KRD, NAJ

The predicate device has not been subject to any design-related recall.

6. Device Description

This proposed device is composed of polyvinyl alcohol embolic microspheres stored in a preservation solution of 0.9% sodium chloride. The proposed devices are compressible hydrogel microspheres with a regular shape, smooth surface, and calibrated size, which are formed as a result of chemical modification on polyvinyl alcohol (PVA) materials. Sunspheres are chemically cross-linked to form the microsphere. Microspheres are round in appearance with uniform size distribution and are available in blue dyed and undyed (colorless) microspheres.

The proposed device available in blue dyed and undyed (colorless) version, which are available in 1 mL and 2 mL volumes, and 6 different calibrated particle sizes from 90 to 1000 um. Different particle sizes are distinguished by the color of the matching vial cap. The proposed device is supplied sterile and packaged in sealed glass vials. The device configurations of Sunspheres are described in Table 1 below.

SpecificationDyeColorCalibrated Size (µm)Size Range (µm)Quantity
F2B100909070 - 1201 mL embolic microspheres/7 mL0.9% sodium chloride
F2B10200200150 - 2501 mL embolic microspheres/7 mL0.9% sodium chloride
F2B10400Blue400350 - 4501 mL embolic microspheres/7 mL0.9% sodium chloride
F2B10600600550 - 6501 mL embolic microspheres/7 mL0.9% sodium chloride
F2B10800800700 - 9001 mL embolic microspheres/7 mL0.9% sodium chloride
F2B110001,000900 - 11001 mL embolic microspheres/7 mL0.9% sodium chloride
F2B200909070 - 1202 mL embolic microspheres/7 mL0.9% sodium chloride
F2B20200200150 - 2502 mL embolic microspheres/7 mL0.9% sodium chloride
F2B20400400350 - 4502 mL embolic microspheres/7 mL0.9% sodium chloride
F2B20600600550 - 6502 mL embolic microspheres/7 mL0.9% sodium chloride
F2B20800800700 - 9002 mL embolic microspheres/7 mL0.9% sodium chloride
F2B210001,000900 - 11002 mL embolic microspheres/7 mL0.9% sodium chloride
F2C100909070 - 1201 mL embolic microspheres/7 mL0.9% sodium chloride
F2C10200200150 - 2501 mL embolic microspheres/7 mL0.9% sodium chloride
F2C10400400350 - 4501 mL embolic microspheres/7 mL0.9% sodium chloride
F2C10600600550 - 6501 mL embolic microspheres/7 mL0.9% sodium chloride
F2C10800800700 - 9001 mL embolic microspheres/7 mL0.9% sodium chloride
F2C110001,000900 - 11001 mL embolic microspheres/7 mL0.9% sodium chloride
F2C20090Colorless9070 - 1202 mL embolic microspheres/7 mL0.9% sodium chloride
F2C20200200150 - 2502 mL embolic microspheres/7 mL0.9% sodium chloride
F2C20400400350 - 4502 mL embolic microspheres/7 mL0.9% sodium chloride
F2C20600600550 - 6502 mL embolic microspheres/7 mL0.9% sodium chloride
F2C20800800700 - 9002 mL embolic microspheres/7 mL0.9% sodium chloride
F2C210001,000900 - 11002 mL embolic microspheres/7 mL0.9% sodium chloride

Table 1. Product specification

{6}------------------------------------------------

Page 3 of 10

7. Indications for Use:

Sunsphere is intended to be used for the embolization of hypervascular tumors, including uterine fibroids, and arteriovenous malformations (AVMs).

8. Technological Comparison to Predicate Device

Table 2. General Comparison

{7}------------------------------------------------

ItemProposed DeviceK232934Predicate DeviceK173871Remark
ClassificationIIIISame
Product CodeKRD, NAJKRD, NAJSame
RegulationNumber21 CFR 870.330021 CFR 870.3300Same
Indication for UseThe Sunsphere is indicated foruse in the embolization ofhypervascular tumors, includinguterine fibroids, andarteriovenous malformations(AVMs).CalliSpheres EmbolicMicrospheres and 8SpheresEmbolic Microspheres are intendedto be used for the embolization ofarteriovenous malformations(AVMs) and hypervascular tumors,including uterine fibroids.Same
PolymerizationMethodSuspension polymerizationSuspension polymerizationSame
ResorptionNon-resorbableNon-resorbableSame
AppearanceHydrogel microspheres withregular shape. smooth surfaceand calibrated size,Blue dyed or undyedHydrogel microspheres withregular shape. smooth surface andcalibrated size,CalliSpheres: Blue dyed8Spheres: UndyedSame
MicrospherePolymerCompositionPolyvinyl alcohol (PVA)PVASame
Size Range (µm)Size Range: 70 – 1,100µmLabeled size range:70-120150-250350-450550-650700-900900-1100Size Range: 100 – 1,200µmLabeled size range:100-300300-500500-700700-900900-1200Different.
Size indicated forUAE (µm)550-650700-900900-1100500-700700-900900-1200Different
Storage media0.9% sodium chloride solution0.9% sodium chloride solutionSame
Quantity ofMicrospheres1 mL (in 7 mL 0.9% sodiumchloride)2 mL (in 7 mL 0.9% sodiumchloride)1 mL (in 7 mL 0.9% sodiumchloride)2 mL (in 7 mL 0.9% sodiumchloride)Same

{8}------------------------------------------------

PackagingDevice are sealed in borosilicate glass vialBoth CalliSpheres and 8Spheres are sealed in borosilicate glass vialSame
Compressibility30% by diameter30% by diameterSame
Delivery MethodDelivered to target position by Micro-Catheter under DSA.Delivered to target position by Micro-catheter under DSA.Same
Compatible Delivery Catheter1.7-4.0Fr1.7-4.0FrSame
Radiopacity MethodMixed with non-ionic contrast agent prior to injectionMixed with non-ionic contrast agent prior to injectionSame
Method of SupplySterile and single useSterile and single useSame
Shelf Life2 years2 yearsSame
SterilizationMoist heat and non-pyrogenicMoist heat and non-pyrogenicSame

Both devices have the same intended use, and use the same embolization particle composition, vial sizes, and sterilization method. The minimum particle size for the proposed device (70 um) is smaller than the minimum particle size for the predicate device (100 µm). The size ranges indicated for UAE for the proposed device are different from the predicate device size range for UAE. The proposed device range (550-1100 um) is within the range listed for the predicate device (500-1200 um). The differences in the particle sizes does not raise different questions of safety and effectiveness.

9. Non-Clinical Test Conclusion

The device is subject to Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices issued on December 29, 2004. The safety and effectiveness of the Sunsphere has been evaluated by non-clinical performance testing including:

Performance testing

The following specifications and performance testing were evaluated on the subject device:

  • . Appearance
  • Microsphere Compressibility .
  • Microsphere Quantity in each Vial .
  • Microsphere Size ●
  • . Time in Suspension
  • Catheter Deliverability ●
  • Water Content per USP 39 NF 34 (2016) <731> Loss on drying
  • pH per USP 39 NF 34 (2016) <791> ●
  • Impurities and residual solvents per USP 39 NF 34 <1086>

Above, all tests were passed and demonstrated the proposed devices can meet the product requirements.

{9}------------------------------------------------

Biocompatibility testing

The biocompatibility evaluation for the proposed device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", issued September 8, 2023. The following tests were conducted for Sunsphere, for a permanent implant device in contact with blood:

Biological EndpointTest MethodResults
MTT Cytotoxicity TestISO 10993-5:2009Non-cytotoxic
ISO Guinea Pig MaximizationSensitization TestISO 10993-10:2010Non-sensitizer
ISO Intracutaneous Study inRabbitsISO 10993-10:2010Non-irritant
ISO Acute Systemic ToxicityStudy in MiceISO 10993-11:2017No mortality or evidence of systemictoxicity
ASTM Hemolysis StudyISO 10993-4:2017ASTM F756:2017Non-hemolytic
ASTM Partial ThromboplastinTimeASTM F2382:2018Non-activator
SC5b-9 Complement ActivationAssayISO 10993-4:2017Both the test article and the sponsorprovided control were considered to bepotential activators of the complementsystem.The test article was comparedstatistically to the control and the testarticle was lower than the control andwas statistically different.
USP Rabbit Pyrogen Study,Material MediatedISO 10993-11:2017Non-pyrogenic
Genotoxicity MouseLymphoma AssayISO 10993-3:2014Non-mutagenic
Biological EndpointTest MethodResults
Bacterial Reverse MutationStudyISO 10993-3:2014Non-mutagenic
ISO Systemic Toxicity Study inRats FollowingSubcutaneous Implantation,13 WeeksISO 10993-6:2016ISO 10993-11:2017There was no evidence of systemictoxicity from the test article followingsubcutaneous implantation in the rat.Microscopically, the test article wasclassified as causing minimal to noreaction.
ISO Muscle Implantation Studyin Rabbits, 1 WeekISO 10993-6:2016The macroscopic reaction was notsignificant as compared to the negativecontrol article.Microscopically, the test article caused aminimal to no reaction as compared tothe negative control article.
ISO Muscle Implantation Studyin Rabbits, 4 WeeksISO 10993-6:2016The macroscopic reaction was notsignificant as compared to the negativecontrol article.Microscopically, the test article caused aminimal to no reaction as compared tothe negative control article.
ISO Muscle Implantation Studyin Rabbits, 13 WeeksISO 10993-6:2016The macroscopic reaction was notsignificant as compared to the negativecontrol article.Microscopically, the test article caused aminimal reaction as compared to thenegative control article.
ISO Muscle Implantation Studyin Rabbits, 26 weeksISO 10993-6:2016The macroscopic reaction was notsignificant as compared to the negativecontrol article.Microscopically, the test article causedno reaction as compared to the negativecontrol article.
Biological EndpointTest MethodResults
Chronic Systemic Toxicity andCarcinogenicity EvaluationISO 10993-18:2023A chemical characterization study andtoxicological risk assessment were usedto support chronic systemic toxicity andcarcinogenicity endpoints. The study wasbased on chemical characterizationresults and supported by information onthe Sunsphere materials of construction,gathered toxicological data on thesematerials. A toxicological riskassessment was completed based on theresults from the chemicalcharacterization.

{10}------------------------------------------------

{11}------------------------------------------------

Page 8 of 10

Chemical Characterization

Chemical characterization testing was conducted on Sunsphere in accordance with ISO 10993-18:2022. The testing consisted of exhaustive extraction of the microspheres in purified water, ethanol, and hexane. Extracts and normal saline were analyzed per Direct Injection Gas Chromatography Mass Spectrometry (DI-GCMS), Headspace Gas Chromatography Mass Spectrometry (HS-GC-MS), Liquid Chromatography High-Resolution Mass Spectrometry (LC-HRMS) and Inductively Coupled Plasma Mass Spectrometry (ICP-MS) screening analysis.

Toxicological Risk Assessment

The toxicological risk assessment was performed to address the chronic systemic toxicity and carcinogenicity end point. For each detected compound, a toxicological risk assessment was conducted in accordance with ISO 10993-17: 2002 and ISO/TS 21726: 2019. The toxicological risk assessment was conducted on the detected compounds from the chemical characterization study. The amount of each detected compound is below the threshold of toxicological concern (TTC) value. The appropriate uncertainty factors are used to derive the tolerable intake (TI) limit as recommended in ISO 10993-17:2002. The results demonstrated that the margins of safety (MoS) for potential chemical exposures represented a low risk of chronic toxicity and carcinogenicity. The Sunsphere is not expected to raise an unacceptable toxicity risk to the exposed patients when used as intended.

Sterility, Shipping, and Shelf -life

The proposed device is sterilized by moist heat steam which under validated parameters (121°C, 30 minutes) after sealing the vials to achieve a sterility assurance level (SAL) of 106 in accordance with USP <71> and

{12}------------------------------------------------

ISO 17665:2015. Bacterial Endotoxin Levels were below the level of 0.5 EU/ml in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 2 year shelf-life.

  • Package integrity test after environmental conditioning, simulated transportation testing in accordance . to ASTM D4169-22 on final, packaged, and sterile device.
  • . Sterile Barrier Packaging performed on the proposed device:
    • Dye penetration ASTM F1929-15 O
  • Shelf-life of 2-years is validated using FDA recognized standard ASTM F1980 -21 Standard Guide for ● Accelerated Aging of Sterile Barrier Systems for Medical Devices.

Animal Performance Testing Conclusion

This study simulated the clinical use of Sunsphere and performed partial embolization of the right renal artery of healthy domestic pigs by interventional surgical procedures. The test article was Sunsphere, with particle size of 90 um; the control article was the cleared predicate device 8Spheres (polyviny) alcohol embolic microspheres - colorless), with particle size of 100 ~ 300 µm. The efficacy was evaluated by investigating the delivery performance, embolization effect immediate and within the study cycle. The test article in the study cycle was evaluated by investigating the postoperative animal recovery, clinical pathology, local embolization, gross anatomical observation, and systemic histopathological results at postoperative timepoints of 3, 8 and 28 days after embolization.

No clinically relevant intraoperative, postoperative, hematologic, coagulation, or major organ abnormalities other than the right kidney occurred. Among the serum biochemical parameters, the alanine aminotransferase (ALT) and aspartate aminotransferase (AST) in the test group and the control group were slightly increased in the short term after surgery, but there was no significant difference between the two groups and no physiological abnormality was reached. Histopathology results showed only a small microsphere shift and secondary tissue response in the upper pole of the right kidney on Day 8, but there was no significant difference between the test group and control group, and its severity and extent of involvement were very mild. Overall, the lower pole of the right kidney in the test group and control group presented similar gross lesions and histopathological features. Vascular changes and necrosis were observed to be equivalent in the test group on Day 8 and Day 29, consistent with the equivalent long term embolization performance results of test article. In conclusion, compared with the control article demonstrated efficacy and safety equivalent to the predicate in this animal study.

10. Conclusion

As described above, the subject devices have intended use as the predicate device. The differences in technological characteristics between the subject devices and the predicate do not raise different questions of safety and effectiveness. To evaluate the impact of the technological differences, performance testing was conducted as described above. The testing demonstrate that the subject devices, Sunsphere, is

{13}------------------------------------------------

Page 10 of 10

as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 870.3300 Vascular embolization device.

(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).