(30 days)
No
The document mentions "image processing" but does not contain any terms or descriptions indicative of AI or ML technology.
No
The device is an imaging system (X-ray) used for diagnosis and not for treatment.
Yes
The device is described as a multi-functional general R/F system that performs routine radiography and fluoroscopy exams to retrieve images, which are used for diagnostic purposes. The software system acquires, processes, stores, displays, and exports digital fluoroscopy and radiographic images, which are then interpreted by a healthcare professional to diagnose medical conditions.
No
The device is described as a multi-functional general Radiography/Fluoroscopy (R/F) system with numerous hardware components listed, including a basic unit, workstation, spot film device, X-ray generator, X-ray tube assembly, and flat panel detector. While it includes software (Eleva Workspot), it is an integral part of a larger hardware system.
Based on the provided information, the CombiDiagnost R90 is not an IVD (In Vitro Diagnostic) device.
Here's why:
- Intended Use/Indications for Use: The intended use clearly states it's a multi-functional general R/F system for routine radiography and fluoroscopy exams. These are imaging procedures performed directly on a patient, not tests conducted on biological samples outside the body.
- Device Description: The description details an X-ray system with components like an X-ray tube, generator, detector, and workstation. This is consistent with medical imaging equipment used for visualizing internal structures of the body.
- Input Imaging Modality: The input modality is X-ray, which is a form of radiation used for imaging, not for analyzing biological samples.
- Anatomical Site: While not specified, the description indicates it's suitable for various anatomical sites within the body.
- IVD Definition: In vitro diagnostics are tests performed on samples such as blood, urine, or tissue to detect diseases, conditions, or infections. The CombiDiagnost R90 does not perform such tests.
The device is a medical imaging system used for diagnostic purposes, but it falls under the category of radiological imaging devices, not in vitro diagnostics.
N/A
Intended Use / Indications for Use
The CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.
Product codes
JAA, KPR
Device Description
The CombiDiagnost R90 is a multi-functional general Radiography/Fluoroscopy (R/F) system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.
The CombiDiagnost R90 is a remote-controlled fluoroscopy system in combination with high-end digital radiography. The system is suitable for routine X-ray examinations and special examinations on patients in standing, seated or laying positions. The CombiDiagnost R90 retrieves images by means of a Cesum Iodide flat panel detector.
Philips fluoroscopy systems consist of the following components (standard configuration):
- Basic unit (also called "geometry" or "table unit")
- Workstation Eleva Workspot with integrated generator control, hand switch, keyboard, mouse, touch screen and PC
- Equipped with a dual screen-monitor as standard
- Spot film device (digital camera or flat panel detector)
- X-ray Generator Velara
- X-ray tube assembly mounted in above table mode to be remote controlled
- Receptor: Flat panel detector
Optional components:
- Skyplate wireless portable detectors small and large
- Ceiling Suspension (CSM3)
- Vertical Wall stand (VS2)
- Ceiling Suspension for monitors
- Monitor trolley
- Remote control for RF viewer
- Accessories for "Stitching on the Table"
The CombiDiagnost R90 uses the same workflow from the currently marketed and predicate device, CombiDiagnost R90 (K163210) with only the following modifications:
- additional optional components (like the reference monitor, remote control),
- Eleva Workspot updated to incorporate new imaging features mainly from the previously approved reference device, DigitalDiagnost C90 (K182973) along with functional clusters like Digital Subtraction Imaging and stitching on the table
- undates to improve usability and serviceability.
The Eleva software of the proposed CombiDiagnost R90 is based on a workstation i.e., Eleva Workspot (computer, keyboard, display, and mouse) that is used by an operator to preset examination data and to generate, process and handle digital x-ray images. The Eleva Software system is decomposed into software components. These components are clustered in three component collections like the image handling focused Back-end (BE), the acquisition focused Front-end (FE) and Image Processing (IP). The Eleva software is intended to acquire, process, store, display and export digital fluoroscopy and radiographic images.
The proposed CombiDiagnost R90 is same as the predicate device (K203087) with some modifications as described.
The proposed device complies to 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, dated September 1, 2016'. The solid-state imaging components including the detector in the proposed device have the same physical, functional, and operational characteristics as the predicate device (K203087). Also, other image chain components like X-ray tube and generator, which are used for exposure characteristics and clinical performance evaluation remains the same. Hence all the features and characteristics potentially influencing image quality of the proposed are in accordance with FDA guidance.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
Not Found
Indicated Patient Age Range
pediatric work
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
This 510(K) premarket notification includes non-clinical performance testing.
Tests were performed on the proposed ProxiDiagnost N90 according to the following FDA recognized standards and guidance documents: ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, IEC 60601-1-2 Edition 4.0 2014-02, IEC 60601-1-3 Edition 2.1 2013-04, IEC 60601-1-6 Edition 3.1 2013-10, IEC 60601-2-54 Edition 1.1 2015-04, IEC 62304 Edition 1.1 2015-06, ANSI AAMI ISO 14971: 2007/(R)2010, ISO 10993-1, Fifth edition 2018-08, Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices, issued September 1, 2016, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005, Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications, issued November 2017, Guidance for Radio Frequency Wireless Technology in Medical Devices, issued August 14, 2013, Guidance for Management of Cybersecurity in Medical Devices, issued October 2, 2014.
Key results:
- System Verification testing: Tests performed (D000710352, D000906247, D001021873). Result: Pass, System verification test activities substantiate that the system conforms to the system requirements.
- Software verification testing: Tests performed (D000693721, D000457609, D001012699, D000185012, D000560160). Result: Pass, Software verification test activities substantiate that the software conforms to the requirements.
- Risk control: Tests performed (D000234134, D000188652). Result: Pass, System meets the defined risk control measures.
- Cyber Security: Tests performed (D000584899, D000256905, D000585130, D000513810, D001393076). Result: Pass, Verification test activities substantiate that the system meets the defined security risk control measures.
No clinical data was submitted in this 510(k) premarket notification.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1650 Image-intensified fluoroscopic x-ray system.
(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
October 19, 2023
Philips Medical Systems DMC GmbH % Ming Xiao Regulatory Affairs Manager Röntgenstrasse 24 HAMBURG, HAMBURG 22335 GERMANY
Re: K232910
Trade/Device Name: CombiDiagnost R90 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA, KPR Dated: September 19, 2023 Received: September 19, 2023
Dear Ming Xiao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
510(k) Number (if known)
Device Name
CombiDiagnost R90
Indications for Use (Describe)
The CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness is prepared in accordance with 21 CFR 8807.92
§807.92. | |
---|---|
Preparation Date: | September 14th, 2023 |
510(k) Owner: | Philips Medical Systems DMC GmbH |
Röntgenstrasse 24 | |
22335 Hamburg GERMANY | |
Establishment registration number: 3003768251 | |
Primary Contact: | Ming Xiao |
Regulatory Affairs Manager | |
Phone: +49(40) 34971-2306 | |
E-mail: ming.xiao@philips.com | |
Secondary Contact: | Shalin Dave |
Senior Regulatory Affairs Project Manager | |
Phone: +31(6) 42532604 | |
E-mail: shalin.dave@philips.com | |
Proposed Device | |
Device Name | CombiDiagnost R90 |
Legal Manufacturer | Philips Medical Systems DMC GmbH |
Röntgenstrasse 24 | |
22335 Hamburg | |
GERMANY | |
Classification Name: | Image-intensified fluoroscopic x-ray System |
Classification Regulation: | 21 CFR Part 892.1650 |
Classification Panel: | 90 - Radiology |
Device Class: | Class II |
Product Code: | JAA, KPR |
Predicate Device | |
Device Name | CombiDiagnost R90 (K203087, December 03, 2020) |
Legal Manufacturer | Philips Medical Systems DMC GmbH |
Röntgenstrasse 24 | |
22335 Hamburg | |
GERMANY | |
Classification Name: | Image-intensified fluoroscopic x-ray system |
Classification Regulation: | 21 CFR Part 892.1650 |
Classification Panel: | 90 - Radiology |
Device Class: | Class II |
Product Code: | JAA, KPR, MQB |
Reference Device # 1 | |
Device Name | ProxiDiagnost N90 (K212837, cleared September 21, 2021) |
Legal Manufacturer | Philips Medical Systems DMC GmbH |
Röntgenstrasse 24 | |
22335 Hamburg | |
GERMANY | |
Classification Name: | Image-intensified fluoroscopic x-ray System |
Classification Regulation: | 21 CFR Part 892.1650 |
Classification Panel: | 90 - Radiology |
Device Class: | Class II |
4
Device Description:
The CombiDiagnost R90 is a multi-functional general Radiography/Fluoroscopy (R/F) system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.
The CombiDiagnost R90 is a remote-controlled fluoroscopy system in combination with high-end digital radiography. The system is suitable for routine X-ray examinations and special examinations on patients in standing, seated or laying positions. The CombiDiagnost R90 retrieves images by means of a Cesum Iodide flat panel detector.
Philips fluoroscopy systems consist of the following components (standard configuration):
- Basic unit (also called "geometry" or "table unit") ●
- . Workstation Eleva Workspot with integrated generator control, hand switch, keyboard, mouse, touch screen and PC
- Equipped with a dual screen-monitor as standard
- Spot film device (digital camera or flat panel detector)
- . X-ray Generator Velara
- . X-ray tube assembly mounted in above table mode to be remote controlled
- Receptor: Flat panel detector Optional components:
- . Skyplate wireless portable detectors small and large
- Ceiling Suspension (CSM3) ●
- . Vertical Wall stand (VS2)
- . Ceiling Suspension for monitors
- Monitor trolley
- . Remote control for RF viewer
- Accessories for "Stitching on the Table" ●
The CombiDiagnost R90 uses the same workflow from the currently marketed and predicate device, CombiDiagnost R90 (K163210) with only the following modifications:
- additional optional components (like the reference monitor, remote control), -
- -Eleva Workspot updated to incorporate new imaging features mainly from the previously approved reference device, DigitalDiagnost C90 (K182973) along with functional clusters like Digital Subtraction Imaging and stitching on the table
- undates to improve usability and serviceability. -
The Eleva software of the proposed CombiDiagnost R90 is based on a workstation i.e., Eleva Workspot (computer, keyboard, display, and mouse) that is used by an operator to preset examination data and to generate, process and handle digital x-ray images. The Eleva Software system is decomposed into software components. These components are clustered in three component collections like the image handling focused Back-end (BE), the acquisition focused Front-end (FE) and Image Processing (IP). The Eleva software is intended to acquire, process, store, display and export digital fluoroscopy and radiographic images.
The proposed CombiDiagnost R90 is same as the predicate device (K203087) with some modifications as described.
The proposed device complies to 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, dated September 1, 2016'. The solid-state imaging components including the detector in the proposed device have the same physical, functional, and operational characteristics as the predicate device (K203087). Also, other image chain components like X-ray tube and generator, which are used for exposure characteristics and clinical performance evaluation remains the same. Hence all the features and characteristics potentially influencing image quality of the proposed are in accordance with FDA guidance
5
document. Additionally, image quality testing has been performed on the proposed device for the changes that are affecting the image quality.
Indications for Use:
CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.
Fundamental Scientific Technology:
The proposed CombiDiagnost R90 is substantially equivalent to the manufacturer's legally marketed devices:
- . Predicate Device: CombiDiagnost R90, (K203087, SE date December 31, 2020), Philips Medical Systems DMC
A detailed comparison of the proposed and predicate device (K203087) is provided in Table 1.
| | Predicate Device, CombiDiagnost
R90 (K203087) | Proposed Device, CombiDiagnost
R90 |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Legal
Manufacturer | Philips Medical Systems DMC GmbH | Identical |
| Classification | Class II per 21 CFR Part 892.1650, Primary Product
code: JAA
Secondary Product code: KPR, MQB | Identical |
| Regulation Name | Stationary x-ray system | Identical |
| Class | II | Identical |
| Review Panel | Radiology | Identical |
| Indications for Use | CombiDiagnost R90 is a multi-functional general R/F
system. It is suitable for all routine radiography and
fluoroscopy exams, including specialist areas like
angiography or pediatric work, excluding
mammography. | Identical |
| Energy Source | X-ray | Identical |
| Principle of
Operation | CombiDiagnost R90 is intended for the medical
application procedures radiography and fluoroscopy.
Depending on the specific indication, X-ray procedures
vary in patient positioning and the modification of
radiographic parameters. The system is suitable for
routine X-ray examinations and special examinations
on patients in standing, seated or laying positions,
depending on the specific indication. | Identical |
| Table Features | | |
| Working height
(table top center to
floorplate) | 65 cm – 133 cm | Identical |
| Table tilt
movement | -90° to +90° | Identical |
| Table top
suspension | Two sides suspensions | Identical |
| | Predicate Device, CombiDiagnost
R90 (K203087) | Proposed Device, CombiDiagnost
R90 |
| Table top material | Plastic laminate or carbon fiber | Identical |
| Table top
movement | Lateral: 32 cm (12.6") (± 16 cm (6.3"))
Longitudinal: Only detector movements to
improve patient comfort | Identical |
| Table top
absorption | Plastic, with Carbon fiber: 0.6mm Al @ 100kV,
HVL = 3.6mm Al | Identical |
| Maximum patient
weight | 284 kg (626 lbs) | Identical |
| Lateral scan
distance | 32 cm ± 16 cm | Identical |
| Lateral scan speed | 5 cm/s, soft start and stop Auto centering | Identical |
| Longitudinal scan
distance | 148 cm (58.3") longitudinal, motorized | Identical |
| Longitudinal scan
speed / Detector
movement | 3 cm - 20 cm / sec | Identical |
| Table column
angulation | -40° to +40° | Identical |
| Source image
distance | 113cm - 183cm | Identical |
| Other Components | | |
| Generator | Philips Velara GCF/RF, 65 kW, optional
80 kW
SPDU initial version (4512-202-03291) | Philips Velara GCF/RF, 65 kW
optional 80 kW
SPDU upgrade version (4512-
202-03293)
This functionality has been
verified to have no impact on the
product safety and effectiveness
of the proposed device. It is
substantial equivalent to reference
device ProxiDiagnost N90
(Κ212837). |
| Tube | Philips SRO 33100 ROT 380 or SRM 0608 ROT GS
505 | Identical |
| Fixed Detector
(Fluoroscopy) | Pixium FE 4343 F | Identical |
| Fixed Detector
(Optional for
Radiography) | Pixium 4343RCE | Pixium 4343RCE2
Substantial equivalent, This
change has no impact on
product Intended Use, safety,
functionality, performance
and effectiveness. |
6
7
| Modulation
Transfer Function
(MTF) (according to
IEC 62220-1-3
standard) | lp/mm | % | lp/mm | % |
---|---|---|---|---|
1 | 66 | 1 | 63 | |
2 | 35 | 2 | 35 | |
3 | 19 | 3 | 19 | |
3.4 | 15 | 3.4 | 15 has | |
Similar, the MTF remained | ||||
the same, with on | ||||
decrease, thus, ov |
| | Predicate Device, CombiDiagnost
R90 (K203087) | Proposed Device, CombiDiagnost
R90 |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | there is no impact to safety and
effectiveness. |
| Detective Quantum
Efficiency (DQE)
(according to IEC
62220-1-3
standard) | DQE at 2 µGy
lp/mm % 0.05
1 51
2 41
3 27
3.4 18 | DQE at 2 µGy
lp/mm % 0.05
1 69
51
2 41
3 27
3.4 18
Similar, the DQE has remained
essentially the same, with one
slight increase, thus, overall, there
is no impact to safety and
effectiveness. |
| System Control | Remote, optional nearby with nearby control trolley | Identical |
| Collimator
(Fluoroscopy) | Motorized automatic collimation | Identical |
| Grid | Parkable | Identical |
| Image Acquisition | Eleva WorkSpot | Identical |
| Operating System | Microsoft Windows 10 | Identical |
| Image Processing | UNIQUE 2 | Identical |
| Image processing | | |
| Image processing functionality | | |
| DSA (Digital
Subtraction
Angiography)
(Optional) | Digital Subtraction Angiography functionality is
available. It is used for angiography and provides
interactive viewing operations on a vascular run, so
that the vascular anatomy becomes visible. | Identical |
| Predefined
annotations
integrated into the
dynamic viewer | Predefined annotations are available which enables
the user to assign annotation to the image with the
help of a set of predefined annotations. This speeds
the user workflow. | Identical |
| Image processing functionality | | |
| Bone Suppression
(Optional) | Bone Suppression post-processing application is
available. It is intended to generate secondary digital
radiographic image of the chest by suppressing bones
from the original image. | Identical |
8
| Intuitive User
Interface for
Processing
Parameters | An intuitive way of modifying the available
parameters is introduced (e.g., contrast, brightness,
noise limit, detail enhancement etc.) on the Eleva user
interface. | Identical |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Deviation and
Target Exposure
Indices | Deviation Index (DI) function is available to quantify
the deviation of the Exposure Index (EI s) from the
Target Exposure Index (EI T). This feature is used by
the clinical user only to identify whether a certain
image has been correctly exposed. | Identical |
| SkyFlow
(Optional) | Skyflow is extended to imaging of other anatomies
(e.g., Leg, hand etc.) including the chest AP/PA
imaging, with Skyplate detector. | Identical |
| | Predicate Device, CombiDiagnost
R90 (K203087) | Proposed Device, CombiDiagnost
R90 |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Access to and
Export of Original
Image Data | Access to and Export of Original Image Data feature
enables the access to an export of original image data
(clean raw images without any modification) on the
system. | Identical |
| One Button Stores
All (OBSA):
Content and
performance
Improvement | System can additionally use OBSA for sending logfile
(Alert files) to central database (RSN, e.g., RADAR
or M2M server) frequently by single button click. | Identical |
| View Selection for
Changed X-Ray
Generation Data
Sets | Users can change a data set only for current
examination or the examinations that are using this
data set.
In addition to the current examinations, the user can
select to apply the changes to subset of all
examinations too that are using this data set. | Identical |
| Avoid Ghosting in
Verification
Images of Portable
Detectors | The feature to avoid ghosting in verification images of
portable detectors is available. It enables the system to
display verification images on wireless (SkyPlate)
detector without ghosting artifacts | Identical |
| Software | | |
| Operating System | Windows 10 | Identical |
| Image Chain
(fluoroscopy) | Philips Dynamic Eleva Image Chain | Identical |
| Image acquisition | Eleva Workspot (Increment 42) | Identical |
| Ceiling Suspension (optional) for Radiography examinations only | | |
| Type | Four-part aluminum telescopic column with spring
counter balanced holder for X-ray tube assembly;
adaptable to individual room heights | Identical |
| Movement | 3440 mm to 6140 mm | Identical |
| Transverse travel | 1500 mm to 3220 mm | Identical |
| Vertical travel | 1650mm to 1705mm | Identical |
| | Predicate Device, CombiDiagnost
R90 (K203087) | Proposed Device, CombiDiagnost
R90 |
| | Live Camera (optional) on tube head for patient positioning support With 2 Lasers (inside the collimator) Detector Calibration with a Filter integrated into the Collimator | |
| Fixed vertical stand
(Optional) | Hardware-Counterbalanced rugged column for motorized and manual vertical movement of the detector, Vertical travel - 30 cm to 180 cm Installation-Floor and wall attachment or floor only | Identical |
| Wireless detector
(Optional) in
vertical stand and
table bucky
(Radiography) | SkyPlate Large (pixium 3543EZ) SkyPlate Small (pixium 2430EZ) | Identical |
| SkyPlate Detector
Sharing | Yes | Identical |
| Tube | High power X-ray Tube,
Philips SRO 33100 ROT 380 or SRM 0608 ROT GS 505 | Identical |
| Automatic Image
Stitching
(Optional) | Yes | Identical |
| X-ray tube
assembly rotation | around vertical axis: ±180°, lock-in position every 45° around horizontal axis: ±115°, lock-in position at 0°, ±90° | Identical |
| Tube Head control | Tube head operation user interface: The User interface on Eleva tube head is touch control. Display: 12" color graphics display with touch control functionality for tube head operation. Eleva screen display in the examination room enables the user to use all the control room parameters from examination room as well. Control Handle: Control handle with flat capacitive sensor for releasing brakes for the ceiling suspension movement. | Identical |
| Collimator | Ralco P 225 ACS DHHS Motorized automatic collimation Manual overrule possible With light field indicator | Identical |
9
10
The outcome of this technological characteristics comparison and risk assessment demonstrate that the minor differences in the technological characteristics do not affect the safety or effectiveness of the proposed CombiDiagnost R90, when compared to the legally marketed predicate device (K203087). This thus demonstrates the substantial equivalence of the proposed device with the predicate device (K203087).
Summary of Non-Clinical and Clinical Performance Data:
This 510(K) premarket notification includes non-clinical performance testing.
Tests were performed on the proposed ProxiDiagnost N90 according to the following FDA recognized standards and guidance documents: The verification and validation methods and acceptance criteria) used to evaluate the proposed CombiDiagnost R90 are the same as those used for the predicate device (K203087) and follow FDA recognized consensus standards and guidance documents applicable to this device type:
- . ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (Recognition #19-4)
- IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests (Recognition #19-8)
- . IEC 60601-1-3 Edition 2.1 2013-04, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment (Recognition # 12-269)
- . IEC 60601-1-6 Edition 3.1 2013-10, Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability (Recognition # 5-89)
- . IEC 60601-2-54 Edition 1.1 2015-04, Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy (Recognition # 12-296)
- . IEC 62304 Edition 1.1 2015-06, Medical device software - Software life cycle processes (Recognition # 13-79)
- . ANSI AAMI ISO 14971: 2007/(R)2010, Medical devices-Application of risk management to medical devices (Recognition # 5-40)
- ISO 10993-1, Fifth edition 2018-08, Biological evaluation of medical devices Part 1: Evaluation and . testing within a risk management process (Recognition # 2-258)
- . Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices, issued September 1, 2016
- . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005
- . Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications, issued November 2017
- . Guidance for Radio Frequency Wireless Technology in Medical Devices, issued August 14, 2013
- . Guidance for Management of Cybersecurity in Medical Devices, issued October 2, 2014
The proposed device complies with the Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices. The solid-state imaging components including the detector in the proposed device have the same physical, functional, and operational characteristics as the predicate device (K203087). Also, other image chain components like X-ray tube and generator, which are used for exposure characteristics and clinical performance evaluation remain the same. Hence, all the features and characteristics potentially influencing image quality of the proposed are in accordance with FDA guidance document 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, dated September 1, 2016'. Additionally,
11
Image quality testing has been performed on the proposed device for the changes that are affecting the image quality.
The proposed device complies with the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005. The changes to the proposed device the intended use or the fundamental scientific technology when compared to the predicate device (K203087). The software for the proposed device has the same 'Level of concern (Moderate)' as that of the predicate device (K203087). The software verification testing has been conducted as per the level of concern.
The proposed device complies with the Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications, issued November 2017. The changes made to the proposed device do not affect the pediatric application. The indication for use of the proposed device is the same as the predicate device (K203087).
The proposed device complies with Guidance for Radio Frequency Wireless Technology in Medical Devices, issued August 14, 2013. All the radiofrequency components of the predicate device (K203087) and proposed device are the same except for the replacement of Bluetooth remote control in the predicate device by infrared remote control in the proposed device. The verification testing has been conducted for the safety and efficacy of the remote control.
The proposed device complies with the Guidance for Management of Cybersecurity in Medical Devices, issued October 2, 2014. A set of cybersecurity controls to assure proposed device's cybersecurity and maintain medical device functionality and safety are in place. Cybersecurity plan and risk document are prepared to assess the proposed device for the following: o Identification of assets, threats, and vulnerabilities:
Assessment of the impact of threats and vulnerabilities on device functionality and end o users/patients;
- O Assessment of the likelihood of a threat and of a vulnerability being exploited;
- Determination of risk levels and suitable mitigation strategies; 0
- Assessment of residual risk. O
12
Refer Table 2 for the non-clinical testing that were performed on the proposed device with respect to the changes. Test results demonstrate that the proposed CombiDiagnost R90 meets acceptance criteria and is adequate for its intended use. Risk assessment activities show that the risks are sufficiently mitigated.
Tests | Document number | Test results |
---|---|---|
System | ||
Verification testing | • D000710352: System Verification Report | |
CombiDiagnost R90 Rel. 1.1.1 & 1.1.2 | ||
• D000906247, System Verification Report | ||
CombiDiagnost R90_R1.1.4 | ||
• D001021873: SVR_CombiDiagnost R90_R1.1.5 | Pass | |
System verification test | ||
activities substantiate that the | ||
system conforms to the | ||
system requirements | ||
Software | ||
verification testing | • D000693721: Verification report SubSystem Eleva | |
Inc.42.1 | ||
• D000457609: Subsystem Verification Report R1.1.4 | ||
• D001012699: Subsystem Verification Report 42.0 | ||
Combi R1.1.5 | ||
• D000185012: Configuration Management Plan Eleva | ||
• D000560160: TEST Report IEC 62304 Medical device | ||
software – Software life-cycle processes including | ||
software maintenance plan | Pass | |
Software verification test | ||
activities substantiate that the | ||
software conforms to the | ||
requirements | ||
Risk control | • D000234134: Risk Matrix Status Report for | |
CombiDiagnost R90 | ||
• D000188652: RMSR for subsystem Eleva | Pass | |
System meets the defined risk | ||
control measures | ||
Cyber Security | • D000584899: Cyber Security Plan CombiDiagnost R90 | |
• D000256905: Cyber Security Risk Management Matrix | ||
CombiDiagnsot R90 | ||
• D000585130: Cyber Security Risk Statement | ||
CombiDiangost R90 | ||
• D000513810: Manufacturer Disclosure Statement for | ||
Medical Device Security for CombiDiagnost R90 | ||
• D001393076: SBOM | Pass | |
Verification test activities | ||
substantiate that the system | ||
meets the defined security | ||
risk control measures |
Table 2 Testing Summary for the changes on the proposed device | |||
---|---|---|---|
There is no clinical data submitted in this 510(k) premarket notification.
Substantial Equivalence Conclusion:
The proposed device, CombiDiagnost R90 is substantially equivalent to the currently marketed and predicate device, (K203087) in terms of the of design features, technological characteristics, indications for use, and safety and effectiveness.
Additionally, substantial equivalence was demonstrated with non-clinical performance tests and testing as per FDA-recognized consensus standards. The test results demonstrates that the device conforms to its specifications and is safe and effective for its intended use.