K Number
K232827
Date Cleared
2023-09-14

(1 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.

Device Description

The LightForce Orthodontic System (LFO System) is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. The LFO System consists of patient-specific ceramic brackets, patient-specific placement jigs, and TPS that allows the orthodontist to view, measure, and diagnose cases.

The patient-specific brackets are available with slot sizes to orthodontic wires up to 0.022" in height and come in a variety of hook options colors and shades. The patient-specific placement jig is an optional-use lab device to aid orthodontists in bracket placement and bonding and is composed of an FDA cleared, class 1 certification material. The LFO system does not contain commercially available or patient-specific shaped arch wires, ligatures, or adhesives that affix the brackets to the teeth. The LFO system packaging consists of a single product offering box with indirect bonding trays (IDB trays) loaded with brackets as well as a secondary set of individual brackets and, where applicable IDB trays and/or bite turbo indirect bonding travs. The LFO system may also include non-patient-specific brackets for temporary use with predetermined torque, tip, offset and base contour, packaged and labeled separately.

The inclusion of an optional self ligating metal clip is to ligate a wire to the bracket by encapsulating the wire in the slot of the bracket.

AI/ML Overview

This document is a 510(k) summary for the LightForce Orthodontic System (LFO System). The purpose of this summary is to demonstrate substantial equivalence to legally marketed predicate devices. It primarily focuses on the addition of an optional self-ligating nitinol clip to the existing LFO System.

Here's an analysis of the provided information concerning acceptance criteria and the study proving the device meets them, organized as requested:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics for the overall device. Instead, it describes performance testing conducted to assess the differences introduced by the new self-ligating clip.

The key performance characteristics evaluated and their qualitative "performance" against the predicate are:

Acceptance Criteria (What was tested)Reported Device Performance
Tie-wing fracture strength"The tie wing fracture strength of LFO System brackets is equivalent to the predicate device."
Wire friction force"The friction force of the LFO System brackets with the inclusion of self-ligating metal clips is substantially better (lower) than the predicate device." (This implies a lower friction force is a desirable outcome).
Wire escapement testing (clip's ability to retain wire)"The LFO system's metal clip is an effective mechanism, in combination with ceramic brackets, to retain a wire in a slot throughout treatment." (Qualitative assessment of effectiveness).
Biocompatibility Assessment (of nitinol clip)"The results of the evaluation met the requirements of the standards as identified above." (ISO 10993-1:2018 and ANSI/ADA Standard No. 41-2020)

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not specify the sample sizes (e.g., number of brackets, number of tests) used for the non-clinical performance testing (tie-wing fracture, wire friction, wire escapement).
  • Data Provenance: The data is from "Non-clinical Performance Testing" and "Biocompatibility," implying lab-based, retrospective testing performed by the manufacturer to support the 510(k) submission. There is no mention of country of origin of the data, but presumably it was conducted in the US or under standards accepted by the US FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable as the testing described is non-clinical performance testing of physical characteristics and biocompatibility, not a clinical study involving human patients or expert interpretation of medical images/diagnoses. The "ground truth" here is established by the results of the physical and chemical tests based on established engineering and materials science principles and standards.

4. Adjudication Method for the Test Set

This section is not applicable for the same reasons as point 3. Adjudication methods are typically used in clinical studies or studies involving human judgment to establish a consensus "ground truth."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. The document explicitly states: "No clinical performance testing was conducted on LFO System brackets."
  • No AI component that would involve human readers. This device is a physical orthodontic system (brackets and treatment planning software) for correcting malocclusions, not an AI-powered diagnostic imaging tool that would assist human readers in interpreting images.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No standalone algorithm performance study was done in the context of an AI/ML algorithm. The "LFO System" includes "TPS [treatment planning software] that allows the orthodontist to view, measure, and diagnose cases." However, the 510(k) summary focuses on the mechanical and material changes to the brackets (specifically the addition of the self-ligating clip) and does not describe performance testing related to the software's diagnostic or planning capabilities as a standalone algorithm. The software is part of the system that "allows the orthodontist to view, measure, and diagnose cases," implying a human-in-the-loop for diagnosis, not a standalone automated diagnosis.

7. The Type of Ground Truth Used

  • For non-clinical performance testing (tie-wing fracture strength, wire friction, wire escapement): The ground truth is defined by the physical measurements obtained from mechanical testing rigs, compared against predefined engineering specifications or performance of the predicate device.
  • For biocompatibility: The ground truth is established by the results of standardized biological assays and chemical characterization, evaluated against the requirements of international standards (ISO 10993-1:2018) and US FDA guidance.

8. The Sample Size for the Training Set

This section is not applicable. The device is an orthodontic appliance system, not an AI/ML device that requires a training set for model development. The "training" in this context refers to manufacturing processes, material selection, and design validation, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.

September 14, 2023

LightForce Orthodontics % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K232827

Trade/Device Name: LightForce Orthodontic System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NJM, PNN Dated: September 13, 2023 Received: September 13, 2023

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Michael E. Adjodha -S

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232827

Device Name LightForce Orthodontic System

Indications for Use (Describe)

The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ട്. 510(k) Summary

510(k) owner's name:LightForce Orthodontics
Address:44 Third Ave.
Burlington, MA 01803
Phone number:800-481-0185
Name of contact person:Jeffrey Roberts
Date the summary was prepared:August 25, 2023
Trade/Proprietary Name of Device:LightForce Orthodontic System
Common or Usual Name:Orthodontic Ceramic Bracket and Accessory
Classification Name:Orthodontic Plastic Bracket
Classification Regulation:21 CFR 872.5470
Product Code:NJM, PNN
Regulatory Class:II

Predicate Devices:

Primary Predicate: LightForce Orthodontic System (K222764) Reference Predicate: Dentsply International IN-OVATION C brackets (K060837)

Device Description:

The LightForce Orthodontic System (LFO System) is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. The LFO System consists of patient-specific ceramic brackets, patient-specific placement jigs, and TPS that allows the orthodontist to view, measure, and diagnose cases.

The patient-specific brackets are available with slot sizes to orthodontic wires up to 0.022" in height and come in a variety of hook options colors and shades. The patient-specific placement jig is an optional-use lab device to aid orthodontists in bracket placement and bonding and is composed of an FDA cleared, class 1 certification material. The LFO system does not contain commercially available or patient-specific shaped arch wires, ligatures, or adhesives that affix the brackets to the teeth. The LFO system packaging consists of a single product offering box with indirect bonding trays (IDB trays) loaded with brackets as well as a secondary set of individual brackets and, where applicable IDB trays and/or bite turbo indirect bonding travs. The LFO system may also include non-patient-specific brackets for temporary use with predetermined torque, tip, offset and base contour, packaged and labeled separately.

The inclusion of an optional self ligating metal clip is to ligate a wire to the bracket by encapsulating the wire in the slot of the bracket.

{4}------------------------------------------------

Intended Use:

The LFO System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.

The indications for use are the same as the primary predicate K222764.

Technological Characteristics:

The LFO System brackets, like its primary predicate LFO system (K222764), are composed of a polycrystalline alumina with a bracket design consisting of tie-wings for ligation, a primary archwire slot and auxiliary slots. There are rounded corners and edges along with a rounded hook on the distal-gingival tie wing to accommodate accessories during orthodontic treatment. These design features allow an archwire to move the bonded brackets along a designated path until the desired tooth position is achieved. The brackets have a mechanical locking base design and are built through additive manufacturing methods. The technological characteristics of the LFO System brackets differ from the predicate LFO system in that the updated LFO system offers the inclusion of an optional self ligating nitinol clip. The self ligating nitinol clip included in the LFO system has the same technological characteristics as the reference predicate Dentsply International IN-OVATION C brackets (K060837).

Non-clinical Performance Testing:

Non-clinical performance testing was performed to ensure that the LFO system brackets were similar to the predicate LFO System. Performance testing was carried out to evaluate the features of the design that were different/changed between the LFO system and its predicate. Performance testing consisted of:

    1. Tie-wing fracture strength is the fracture strength of the tie-wing complex when a load is placed directly under the tie wing. The tie wing fracture strength of LFO System brackets is equivalent to the predicate device.
  • Wire friction force is the force required to drag a ligated stainless-steel wire through the 2. primary slot of the bracket. The friction force of the LFO System brackets with the inclusion of self-ligating metal clips is substantially better (lower) than the predicate device.
    1. Wire escapement testing is an evaluation of the self ligating clip and bracket's capability to successfully hold a wire in place when subjected to normal orthodontic forces. The LFO system's metal clip is an effective mechanism, in combination with ceramic brackets, to retain a wire in a slot throughout treatment.

{5}------------------------------------------------

Clinical Performance Testing:

No clinical performance testing was conducted on LFO System brackets.

Biocompatibility:

A biocompatibility assessment was performed on the proposed modified LFO System to assess the impact of including the nitinol clip to the overall material biocompatibility per ISO 10993-1:2018 and the US FDA's guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." and ANSI/ADA Standard No. 41-2020 Evaluation of Biocompatibility of Medical Devices Used in Dentistry. LFO system brackets are considered mucosal membrane contacting for a duration of greater than 30 days.

The results of the evaluation met the requirements of the standards as identified above.

Substantial Equivalence:

Lightforce Orthodontics claims the LFO System is substantially equivalent to the legally marketed predicate device based on intended use, indication for use, typical clinical use, operational characteristics, and fundamental technological characteristics.

CharacteristicProposedPrimary PredicateReference Predicate
LFO SystemLFO System K222764Dentsply InternationalIN-OVATION Cbrackets K060837
Product Codes /RegulationsNJM (OrthodonticCeramic Bracket,21CFR 872.5470)NJM (OrthodonticCeramic Bracket,21CFR 872.5470)NJM (OrthodonticCeramic Bracket,21CFR 872.5470)
Indications forUseThe LFO System is atreatment planningsoftware andorthodontic appliancesystem used to correctmalocclusions inorthodontic patientsThe LFO System is atreatment planningsoftware andorthodontic appliancesystem used to correctmalocclusions inorthodontic patientsThe In-Ovation C is anorthodontic appliancesystem used to correctmalocclusions inorthodontic patientsusing off-the-shelforthodontic systems.
Table 5-1. Side-by-Side Comparison of the LFO System with the Predicate Device

{6}------------------------------------------------

CharacteristicProposedPrimary PredicateReference Predicate
LFO Systemusing patient-matchedorthodontic appliances.LFO System K222764using patient-matchedorthodontic appliances.Dentsply InternationalIN-OVATION Cbrackets K060837
Sequence ofTreatment Planor Mode of UseLightForceOrthodontics (LFO)receives the patient'sdigital data viacommercially availablecommunications toolsfrom the DDS/DMD.LFO prepares a finaltooth setup and bracketplacement plan andreturns the digital databack to the DDS/DMD.The orthodontist canthen view, measure andmodify the case usingthe TPS. Onceapproved, brackets andjigs are manufacturedand shipped to theDDS/DMD.LightForceOrthodontics (LFO)receives the patient'sdigital data viacommercially availablecommunications toolsfrom the DDS/DMD.LFO prepares a finaltooth setup and bracketplacement plan andreturns the digital databack to the DDS/DMD.The orthodontist canthen view, measure andmodify the case usingthe TPS. Onceapproved, brackets andjigs are manufacturedand shipped to theDDS/DMD.In-Ovation C bracketsare manufactured byDentsply Internationaland shipped to theDDS/DMD or a reseller.
Bracket MaterialCeramicCeramicCeramic
Clip materialNickelTitaniumN/ACobalt chromium,known to containNickel
ClipBiocompatibilitySurfaceTreatmentElectropolished NiTi orGold Rhodium CoatingN/AGold Rhodium Coating
ManufacturingMethod3D Printed3D printedInjection molded
AnalysisMethodsN/AN/AN/A
CharacteristicProposedPrimary PredicateReference Predicate
LFO SystemLFO System K222764Dentsply InternationalIN-OVATION Cbrackets K060837
Non-sterile,single useYesYesYes
Bracket BaseMechanical Lock BaseMechanical Lock BaseMechanical Lock Base
Bracket DesignArch wire slot, tiewings for ligature, andidentification marks forplacement. Hooks forligation, for additionaltooth movement. Aceramic body withrounded corners andedges. Slot to holdorthodontic wires.Self-ligating metal clipto retain orthodonticwires.Arch wire slot, tiewings for ligature, andidentification marks forplacement. Hooks forligation, for additionaltooth movement. Aceramic body withrounded corners andedges. Slot to holdorthodontic wires.Non-self-ligatingArch wire slot, tiewings for ligature, andidentification marks forplacement. Hooks forligation, for additionaltooth movement. Aceramic body withrounded corners andedges. Slot to holdorthodontic wires.Self-ligating metal clipto retain orthodonticwires.

{7}------------------------------------------------

Additional reference predicate device with the same chemical composition as the LFO clip material:

  • CDB Self Ligating Bracket 8F (K160957) by CDB Corporation. CDB Self Ligating ● Bracket 8F is a self ligating ceramic bracket with a Nickel titanium clip.

Conclusion:

Based on the information provided in Table 5-1 and the results from the nonclinical performance testing and the biocompatibility assessment the LFO System brackets with the inclusion of optional metal clips are substantially equivalent to the listed predicate device in terms of indications for use and technological characteristics and effective. Further, any

{8}------------------------------------------------

differences are considered minor and do not raise new issues of safety and effectiveness of the LFO System when compared to the predicate device.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.