(90 days)
Immunoassay for the in vitro quantitative determination of neutrophil gelatinase-associated lipocalin (NGAL) in human urine.
Determination of NGAL is intended to be used in conjunction with clinical evaluation in pediatric patients (≥ 3 months to < 22 years) without underlying kidney disease admitted to the intensive care unit (ICU) for the management of cardiovascular or respiratory compromise or who have had a solid organ or bone marrow transplant.
ProNephro AKI™ (NGAL) is intended to be used in the first 24 hours of ICU admission. In patients with low SCr levels, indicative of no acute kidney injury (AKI) or AKI Stage 1, the test can and to identify patients at risk to develop moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment. In patients with an elevated SCr level, indicative of Stage 2/3 AKI, the test can be used as an aid to identify patients at risk to have persistent moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment.
The particle-enhanced turbidimetric immunoassay is intended for use on the Roche cobas c 501 clinical chemistry analyzer.
The ProNephro AKI™ (NGAL) Reagent Kit contains Reaction Buffer Reagent (R1) and Immunoparticle Suspension Reagent (R2). The R1 reagent is a ready-to-use tris-buffer solution containing murine protein and preservative. The R2 reagent is a ready-to-use suspension of polystyrene microparticles coated with mouse monoclonal antibodies to NGAL that also contains preservative. The ProNephro AKI Reagent Kit provides enough reagents for 100 tests on the Roche cobas c 501 clinical chemistry analyzer. The finished kit will also include a labeled Roche cassette and two funnels for transferring the reagent into the Roche cassette.
The reagents must be transferred to a Roche cassette prior to the cobas c 501. This is done by pouring the R1 reagent into Position B of the cassette and R2 reagent into Position C. The operator will load the cassette onto the instrument and run the ProNephro AKI™ (NGAL) test per the instructions for use.
The ProNephro AKI® (NGAL) Calibrator Kit consists of five (5) individual ready-to-use calibrator solutions (1 mL each) comprised of different concentrations (50-3000 ng/mL) of recombinant human NGAL in a HEPES buffer and a preservative.
The ProNephro AKI™ (NGAL) Control Kit contains ready-to-use High (500 ng/mL target concentration) and Low (200 ng/mL target concentration) Controls comprised of recombinant human NGAL in HEPES buffer and a preservative. There are three (3) 1 mL bottles of each level included with the kit.
Here's a breakdown of the acceptance criteria and study information for the ProNephro AKI™ (NGAL) device, based on the provided text:
Acceptance Criteria and Device Performance
The direct acceptance criteria for the clinical performance are not explicitly stated in a quantifiable manner (e.g., "Sensitivity must be >= X%"). Instead, the document presents the device's performance metrics from the validation study, implying these met the internal criteria for acceptance. The table below summarizes the reported clinical performance.
Table 1: Acceptance Criteria (Implied) and Reported Device Performance
| Performance Metric | Implied Acceptance Criterion (from Predicate) | Reported Device Performance (ProNephro AKI™ (NGAL)) - Overall Study Population (n=514) | Reported Device Performance - Group 1 (n=422) | Reported Device Performance - Group 2 (n=92) |
|---|---|---|---|---|
| Sensitivity (CI 95%) | 90-93% (83-99%) / 76% (60-91%) | 72.3% (57.4-84.4%) | 61.1% (35.8-82.7%) | 79.3% (60.3-92.0%) |
| Specificity (CI 95%) | 45-49% (39-54%) / 51% (41-60%) | 86.3% (82.8-89.3%) | 88.6% (85.1-91.5%) | 71.4% (58.7-82.1%) |
| Positive Predictive Value (PPV) (CI 95%) | 26-27% (21-33%) / 31% (21-42%) | 34.7% (28.5-41.4%) | 19.3% (10.1-31.9%) | 56.1% (39.8-71.5%) |
| Negative Predictive Value (NPV) (CI 95%) | 96-97% (93-100%) / 88% (79-96%) | 96.9% (95.1-98.0%) | 98.1% (96.1-99.2%) | 88.2% (76.1-95.6%) |
Note: The "Implied Acceptance Criterion" is derived from the performance ranges of the predicate device (NEPHROCHECK® Test System) presented in Table 2 of the document. The device performance for ProNephro AKI™ (NGAL) is directly reported.
Study Information
Here's a breakdown of the requested study information:
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: 514 evaluable subjects for the overall clinical performance validation, comprising 422 subjects in Group 1 and 92 subjects in Group 2.
- Data Provenance: Multi-center prospective clinical study ("GUIDANCE") conducted at 15 sites across the US.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: At least two independent clinical experts.
- Qualifications of Experts: Not explicitly stated in the document. They are referred to as "independent clinical experts."
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: "The clinical data for each subject were captured and provided to at least two independent clinical experts for adjudication." This implies a consensus-based adjudication, but the exact method (e.g., needing agreement from both, or a tie-breaker if they disagree) is not detailed.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a quantitative immunoassay for a biomarker (NGAL), not an imaging AI diagnostic that assists human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the clinical performance results presented (Sensitivity, Specificity, PPV, NPV) represent the standalone performance of the ProNephro AKI™ (NGAL) test. It is intended to be used "in conjunction with clinical evaluation," meaning the result from the test (NGAL concentration) is an input for the clinician's overall assessment, rather than the device itself being a "human-in-the-loop" AI system.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth Type: Expert consensus for "moderate to severe acute kidney injury (Stage 2 to 3 AKI) 48 to 72 hours after the time of assessment," based on clinical data. The text states: "The clinical data for each subject were captured and provided to at least two independent clinical experts for adjudication."
-
The sample size for the training set:
- Training Set Sample Size: The clinical cutoff of 125 ng/mL was established using a multi-center prospective clinical study called EARNEST with 257 pediatric patients (203 evaluable subjects). This study served as the "training" or "cutoff establishment" cohort before validation.
-
How the ground truth for the training set was established:
- The ground truth for the EARNEST study (which established the cutoff) was defined as "moderate to severe acute kidney injury (Stage 2 to 3 AKI) 48 to 72 hours after the time of assessment." The specific method for establishing this ground truth (e.g., based on KDIGO criteria, expert adjudication) is not explicitly detailed for the EARNEST study in the same way it is for the GUIDANCE validation study, but it is implied to be clinical assessment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text.
December 7, 2023
BioPorto Diagnostic Inc. Monika Bak Director Regulatory Affairs Tuborg Havnevej 15, st Hellerup, 2900 Denmark
Re: K232761
Trade/Device Name: ProNephro AKI™ (NGAL) Regulation Number: 21 CFR 862.1220 Regulation Name: Acute Kidney Injury Test System Regulatory Class: Class II Product Code: PIG Dated: September 8, 2023 Received: September 8, 2023
Dear Monika Bak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marianela Perez-torres -S
Marianela Perez-Torres, PhD Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K232761
Device Name ProNephro AKI™ (NGAL)
Indications for Use (Describe)
Immunoassay for the in vitro quantitative determination of neutrophil gelatinase-associated lipocalin (NGAL) in human urine.
Determination of NGAL is intended to be used in conjunction with clinical evaluation in pediatric patients (≥ 3 months to < 22 years) without underlying kidney disease admitted to the intensive care unit (ICU) for the management of cardiovascular or respiratory compromise or who have had a solid organ or bone marrow transplant.
ProNephro AKI™ (NGAL) is intended to be used in the first 24 hours of ICU admission. In patients with low SCr levels, indicative of no acute kidney injury (AKI) or AKI Stage 1, the test can and to identify patients at risk to develop moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment. In patients with an elevated SCr level, indicative of Stage 2/3 AKI, the test can be used as an aid to identify patients at risk to have persistent moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment.
The particle-enhanced turbidimetric immunoassay is intended for use on the Roche cobas c 501 clinical chemistry analyzer.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo features a stylized atom-like graphic to the left of the word "bioporto" in a sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, sans-serif font.
510(k) Summary
ProNephro AKI™ (NGAL)
BioPorto Diagnostics Inc. 117 Fourth Avenue, Suite 202 Needham, MA 02494 USA
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "bioporto DIAGNOSTICS". To the left of the word "bioporto" is a graphic of a central circle with eight smaller circles surrounding it. The word "bioporto" is in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, sans-serif font.
General Information 1.
| Submission Date: | September 08, 2023 |
|---|---|
| Submitter InformationSubmitted By: | BioPorto Diagnostics Inc.117 Fourth Avenue Suite 202Needham MA 02494United States |
| Contact Person: | Monika BakDirector Regulatory Affairsmbk@bioporto.com |
510(k):
Purpose for Submission 2.
To obtain a substantial equivalence determination for the ProNephro AKI™ (NGAL) in a pediatric population (≥ 3 months to < 22 years) admitted to the pediatric intensive care unit (PICU).
3. Measurand
Neutrophil gelatinase-associated lipocalin (NGAL)
4. Type of Test
Quantitative, particle-enhanced turbidimetric immunoassay
5. Applicant
BioPorto Diagnostics Inc.
Proprietary and Established Names 6.
ProNephro AKI™ (NGAL)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the text "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font.
Regulatory Information 7.
Assay
| Trade Name: | ProNephro AKI™ (NGAL) |
|---|---|
| Classification: | Class II, 510(k) required |
| Regulation: | 21 CFR 862.1220 |
| Regulation Name: | Acute Kidney Injury Test System |
| Product Code: | PIG |
| Panel: | Clinical Chemistry |
8. Indications for Use
8.1. Indications for Use
Immunoassay for the in vitro quantitative determination of neutrophil gelatinase-associated lipocalin (NGAL) in human urine.
Determination of NGAL is intended to be used in conjunction with clinical evaluation in pediatric patients (≥ 3 months to < 22 years) without underlying kidney disease admitted to the intensive care unit (ICU) for the management of cardiovascular or respiratory compromise or who have had a solid organ or bone marrow transplant.
ProNephro AKI™ (NGAL) is intended to be used in the first 24 hours of ICU admission. In patients with low SCr levels, indicative of no acute kidney injury (AKI) or AKI Stage 1, the test can be used as an aid to identify patients at risk to develop moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment. In patients with an elevated SCr level, indicative of Stage 2/3 AKI, the test can be used as an aid to identify patients at risk to have persistent moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment.
The particle-enhanced turbidimetric immunoassay is intended for use on the Roche cobas® c 501 clinical chemistry analyzer.
8.2. Special Conditions for Use Statement(s)
For in Vitro Diagnostics Use and Rx only
Warnings and Precautions for the Assay
- . This kit should only be used by qualified laboratory staff.
- Do not shake the reagents.
- . Use only clean containers if transferring reagents.
- . Do not pour reagents back into the original containers once transferred.
- . All specimens used in this test should be considered potentially infectious, and as such be treated with standard precautions.
- . Powder-free gloves should be worn. Avoid direct skin contact. Should reagents come into contact with the skin, eyes or mouth, flush with plenty of water. Seek medical advice if any symptom is observed or when it is considered necessary.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for "bioporto DIAGNOSTICS". The logo features a stylized atom-like graphic to the left of the text. The word "bioporto" is in a bold, sans-serif font, and the word "DIAGNOSTICS" is in a smaller, lighter font below it.
- . Reagents contain preservatives with 2-methyl-2H isothiazol-3-one (MIT) which may cause an allergic skin reaction. For more information, please refer to the safety data sheet.
- Do not switch caps on reagent containers as it may cause contamination or mix-up. .
- . Place calibrators in the correct order onto the instrument to ensure a correct calibration curve.
- . Do not use reagents after the expiration date on the labels.
- . Reagents with different lot numbers should not be mixed.
- . All solutions and samples must be handled carefully and disposed of in accordance with national and local regulations.
Special Instrument Requirements 8.3.
Roche cobas® c 501 clinical chemistry analyzer
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font. The overall design is clean and modern.
Device Description 9.
Table 1 describes the components required to run a ProNephro AKI™ (NGAL) test.
The ProNephro AKI™ (NGAL) Reagent Kit contains Reaction Buffer Reagent (R1) and Immunoparticle Suspension Reagent (R2). The R1 reagent is a ready-to-use tris-buffer solution containing murine protein and preservative. The R2 reagent is a ready-to-use suspension of polystyrene microparticles coated with mouse monoclonal antibodies to NGAL that also contains preservative. The ProNephro AKI Reagent Kit provides enough reagents for 100 tests on the Roche cobas c 501 clinical chemistry analyzer. The finished kit will also include a labeled Roche cassette and two funnels for transferring the reagent into the Roche cassette.
The reagents must be transferred to a Roche cassette prior to the cobas c 501. This is done by pouring the R1 reagent into Position B of the cassette and R2 reagent into Position C. The operator will load the cassette onto the instrument and run the ProNephro AKI™ (NGAL) test per the instructions for use.
The ProNephro AKI® (NGAL) Calibrator Kit consists of five (5) individual ready-to-use calibrator solutions (1 mL each) comprised of different concentrations (50-3000 ng/mL) of recombinant human NGAL in a HEPES buffer and a preservative.
The ProNephro AKI™ (NGAL) Control Kit contains ready-to-use High (500 ng/mL target concentration) and Low (200 ng/mL target concentration) Controls comprised of recombinant human NGAL in HEPES buffer and a preservative. There are three (3) 1 mL bottles of each level included with the kit.
| Product Trade Name and Model Number | Kit Components |
|---|---|
| ProNephro AKI™ (NGAL) Reagent KitST101UB | Reaction Buffer (R1), 17.9 mLImmunoparticle Suspension (R2), 7.65 mLRoche cobas c 501 cassette and two funnels |
| ProNephro AKI™ (NGAL) Calibrator KitST102UA | Calibrator 1: 1 mLCalibrator 2: 1 mLCalibrator 3: 1 mLCalibrator 4: 1 mLCalibrator 5: 1 mL |
| ProNephro AKI™ (NGAL) Control KitST103UA | Control Low: 3 x1 mLControl High: 3 x1 mL |
Table 1: Test Components
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, sans-serif font. The atom graphic is light blue and gray, while the text is dark gray.
10. Substantial Equivalence Information
10.1. Predicate Device Name(s)
NEPHROCHECK® Test System
10.2. Predicate 510(k) Number(s)
10.3. Comparison With Predicate
Table 2: Predicate Device Comparison
| ltem | ProNephro AKI™ (NGAL) | NEPHROCHECK® Test System(Predicate Device) |
|---|---|---|
| K Number | K232761 | DEN130031 |
| Intended Use /Indications for Use | INDICATIONS FOR USE:Immunoassay for the in vitro quantitativedetermination of neutrophil gelatinase-associated lipocalin (NGAL) in humanurine.Determination of NGAL is intended to beused in conjunction with clinical evaluationin pediatric patients (≥ 3 months to < 22years) without underlying kidney diseaseadmitted to the intensive care unit (ICU) forthe management of cardiovascular orrespiratory compromise or who have had asolid organ or bone marrow transplant.ProNephro AKI™ (NGAL) is intended to beused in the first 24 hours of ICU admission.In patients with low SCr levels, indicative ofno acute kidney injury (AKI) or AKI Stage 1,the test can be used as an aid to identifypatients at risk to develop moderate tosevere AKI (Stage 2/3) 48 to 72 hours afterthe assessment. In patients with an elevatedSCr level, indicative of Stage 2/3 AKI, thetest can be used as an aid to identifypatients at risk to have persistent moderateto severe AKI (Stage 2/3) 48 to 72 hoursafter the assessment.The particle-enhanced turbidimetricimmunoassay is intended for use on theRoche cobas® c 501 clinical chemistryanalyzer. | INTENDED USE:The Astute Medical NEPHROCHECK®Test System is intended to be used inconjunction with clinical evaluation inpatients who currently have or have hadwithin the past 24 hours acutecardiovascular and or respiratorycompromise and are ICU patients as anaid in the risk assessment for moderate orsevere acute kidney injury (AKI) within 12hours of patient assessment. TheNEPHROCHECK® Test System isintended to be used in patients 21 yearsof age or older. |
| Classification | Class II | Same |
| Product Code | PIG | Same |
| Test Type | Quantitative immunoassay | Same |
| Test Technology | Particle Enhanced Turbidimetric assay | Lateral Flow with fluorescent detection |
| Item | ProNephro AKI™ (NGAL) | NEPHROCHECK® Test System(Predicate Device) |
| Instrument | Roche cobas c 501 Clinical ChemistryAnalyzer | ASTUTE140® Meter |
| Measurand | Neutrophil gelatinase-associated lipocalin(NGAL) | Insulin-like growth factor binding protein(IGFBP7) and tissue-inhibitor ofmetalloproteinases 2 (TIMP2) |
| Population | Pediatric (≥ 3 months to < 22 years) | Adult |
| Units of Measure | ng/mL | AKI Risk Score |
| Assay MeasuringRange | 50 ng/mL - 3000 ng/mL | Score 0.04 - 10.0 |
| Specimen Type | Urine | Same |
| Controls | 2 levels of controls: low and high | Same |
| Calibrators | N=5 levels across the measuring range ofthe assay | Same |
| Special Controls | Complies | Same |
| Repeatability (within-run) andIntermediatePrecision (within-laboratory) | Repeatability (within-run):• 3.8 to 5% below the cutoff (125 ng/mL)• 0.9 to 2.3% from 125 to 500 ng/mL• 1.2 to 1.5% above 500 ng/mLIntermediate Precision (within-laboratory):• 4.9 to 6.6% below the cutoff (125ng/mL)• 2.2 to 3.4% from 125 to 500 ng/mL• 2.2 to 2.7% above 500 ng/mL | Repeatability (within-run):• 9.5 to 9.7% below the cutoff score (0.3)• 7.4 to 7.8% above the cutoff score (0.3)Intermediate precision (within-laboratory):• 10.2 to 11% below the cutoff score (0.3)• 8.0 to 10.2% above the cutoff score(0.3) |
| Reproducibility(multi-site) | Total Imprecision:• 6.2 to 6.6% below the cutoff (125ng/mL)• 3.6 to 4.8% from 125 to 500 ng/mL• 3.6 to 4.3% above 500 ng/mL | Total Imprecision:• 10.4 to 18.3% below and at the cutoffscore (0.3)• 9.1 to 11.7% above the cutoff score(0.3) |
| Analytical Sensitivity | LoB = 9 ng/mLLoD = 14 ng/mLLoQ = 18 ng/mL | LoB = 0.0002LoD = 0.002LoQ = 0.002 |
| Linearity | Linearity confirmed between 49 and 3105ng/mL | The reportable range of theNEPHROCHECK AKI Risk Score is 0.4 to10.0. The markers that go into the scorewere assessed and found to be linear.However, the AKI Risk Score itself is notexpected to be linear. |
| High Dose HookEffect | No Hook Effect at NGAL concentrationsup to 40,000 ng/mL | No Hook Effect at AKI Risk Score valuesup to 250. |
| Item | ProNephro AKI™ (NGAL) | NEPHROCHECK® Test System(Predicate Device) |
| Clinical Performance | Complete Study Population (n=514): | Sensitivity (CI 95%): |
| Sensitivity (CI 95%): | Study A, tested in 3 laboratories: | |
| 72.3% (57.4-84.4%) | 90-93% (83-99%) | |
| Specificity (CI 95%): | Study B, tested in 1 laboratory: | |
| 86.3% (82.8-89.3%) | 76% (60-91%) | |
| PPV (CI 95%): | ||
| 34.7% (28.5-41.4%) | Specificity (CI 95%): | |
| NPV (CI 95%): | Study A, tested in 3 laboratories: | |
| 96.9% (95.1-98-0%) | 45-49% (39-54%) | |
| Group 1 - patients with Day 0 SCr | Study B, tested in 1 laboratory: | |
| indicative of no AKI or AKI Stage 1 | 51% (41-60%) | |
| (n=422): | ||
| Sensitivity (CI 95%): | PPV (CI 95%): | |
| 61.1% (35.8-82.7%) | Study A, tested in 3 laboratories: | |
| Specificity (CI 95%): | 26-27% (21-33%) | |
| 88.6% (85.1-91.5%) | Study B, tested in 1 laboratory: | |
| PPV (CI 95%): | 31% (21-42%) | |
| 19.3% (10.1-31.9%) | ||
| NPV (CI 95%): | NPV (CI 95%): | |
| 98.1% (96.1-99.2%) | Study A, tested in 3 laboratories: | |
| Group 2 - patients with Day 0 SCr | 96-97% (93-100%) | |
| indicative of AKI Stage 2/3 (n=92) | Study B, tested in 1 laboratory: | |
| Sensitivity (CI 95%): | 88% (79-96%) | |
| 79.3% (60.3-92.0%) | ||
| Specificity (CI 95%): | ||
| 71.4% (58.7-82.1%) | ||
| PPV (CI 95%): | ||
| 56.1% (39.8-71.5%) | ||
| NPV (CI 95%): | ||
| 88.2% (76.1-95.6%) |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized molecular structure graphic on the left, followed by the text "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized, light blue, circular graphic to the left of the company name. The company name, "bioporto" is in a dark gray sans-serif font, with the word "DIAGNOSTICS" in smaller font below.
11. Standards/Guidance Documents Referenced
- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved . Guideline - Third Edition
- . CLSI EP06 2nd, Edition, Evaluation of the Linearity of Quantitative Measurement Procedures
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement . Procedures; Approved Guideline - Second Edition
- CLSI EP07 3rd, Edition, Interference Testing in Clinical Chemistry. .
- . CLSI EP28-A3C Defining Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
- . ISO 14155 Third edition 2020-07, Clinical investigation of medical devices for human subjects -Good clinical practice
- ISO 14971 Third Edition 2019-12, Medical devices Application of risk management to medical ● devices
- . ISO 15223-1 Fourth edition 202-07, Medical devices- Symbols to be used with information to be supplied by the manufacturer - Part 1: General Requirements
- . ISO 17511 Second edition 2020-04, In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators trueness control materials and human samples
- ISO 2859-1 Second edition 1999-11-15, Sampling procedures for inspection by attributes - Part 1: Sampling schemes indexed by acceptance quality limit (AQL) for lot-by-lot inspection [Including: Technical Corrigendum 1 (2001) Amendment 1 (2011)]
- IEC 62366-1 Edition 1.0. 2015-02, Medical devices - Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
12. Test Principle
ProNephro AKI (NGAL) is a particle-enhanced turbidimetric immunoassay for the quantitative determination of NGAL in pediatric urine. A urine sample is mixed with reaction buffer. After a short incubation, the reaction is started by the addition of an immunoparticle suspension (polystyrene microparticles coated with mouse monoclonal antibodies to NGAL). NGAL in the sample causes the immunoparticles to aggregate. The degree of aggregation is proportional to the loss of intensity of transmitted light, due to the scattering effect of particles suspended in solution. The NGAL concentration in the sample is determined by interpolation on an established calibration curve.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like symbol on the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, sans-serif font.
13. Performance Data – Analytical Studies
13.1. Precision
13.1.1. Repeatability and Intermediate Precision
Precision measurements were conducted to evaluate the repeatability (within-run precision) and intermediate precision (within-laboratory precision) according to CLSI guideline EP05-03. The protocol consisted of testing seven native pediatric urine sample pools and two controls with three lots of reagent for 20 days with two runs per day and three replicates per sample (n=360/sample) on one cobas c 501 analyzer.
| Repeatability(Within-Run) | Between-Run | Between-Day | Within Laboratory(total) | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Mean(ng/mL) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) | |
| Control Low | 203 | 3.5 | 1.7 | 4.1 | 2.0 | 3.4 | 1.7 | 6.4 | 3.1 |
| Control High | 498 | 4.6 | 0.9 | 7.4 | 1.5 | 7.8 | 1.6 | 11.7 | 2.3 |
| Ped. Urine 1 | 71 | 3.6 | 5.0 | 2.2 | 3.1 | 0.0 | 0.0 | 4.7 | 6.6 |
| Ped. Urine 2 | 100 | 3.5 | 3.5 | 1.7 | 1.7 | 1.2 | 1.2 | 4.5 | 4.5 |
| Ped. Urine 3 | 102 | 3.8 | 3.8 | 2.3 | 2.2 | 1.0 | 1.0 | 5.0 | 4.9 |
| Ped. Urine 4 | 142 | 3.3 | 2.3 | 2.6 | 1.9 | 1.6 | 1.1 | 4.9 | 3.4 |
| Ped. Urine 5 | 165 | 3.7 | 2.2 | 2.4 | 1.5 | 2.5 | 1.5 | 5.2 | 3.1 |
| Ped. Urine 6 | 304 | 4.5 | 1.5 | 2.9 | 1.0 | 3.6 | 1.2 | 7.3 | 2.4 |
| Ped. Urine 7 | 512 | 6.0 | 1.2 | 4.0 | 0.8 | 6.5 | 1.3 | 11.2 | 2.2 |
| Ped. Urine 8 | 1049 | 12.4 | 1.2 | 8.5 | 0.8 | 12.8 | 1.2 | 22.7 | 2.2 |
| Ped. Urine 9 | 2776 | 42.4 | 1.5 | 27.9 | 1.0 | 34.8 | 1.3 | 75.2 | 2.7 |
Table 3: Repeatability and Intermediated Precision Results
{13}------------------------------------------------
Table 4: Lot-to-Lot Imprecision
| Mean (ng/mL) | Within-Lot* | Between-Lot | |||
|---|---|---|---|---|---|
| Std Dev (ng/mL) | CV (%) | Std Dev (ng/mL) | CV (%) | ||
| Control Low | 203 | 6.4 | 3.1 | 0.0 | 0.0 |
| Control High | 498 | 11.7 | 2.3 | 0.0 | 0.0 |
| Ped. Urine 1 | 71 | 4.2 | 5.9 | 2.1 | 3.0 |
| Ped. Urine 2 | 100 | 4.1 | 4.1 | 2.0 | 2.0 |
| Ped. Urine 3 | 102 | 4.6 | 4.5 | 2.1 | 2.0 |
| Ped. Urine 4 | 142 | 4.5 | 3.2 | 1.9 | 1.3 |
| Ped. Urine 5 | 165 | 5.0 | 3.1 | 1.1 | 0.7 |
| Ped. Urine 6 | 304 | 6.4 | 2.1 | 3.5 | 1.1 |
| Ped. Urine 7 | 512 | 9.7 | 1.9 | 5.5 | 1.1 |
| Ped. Urine 8 | 1049 | 19.7 | 1.9 | 11.3 | 1.1 |
| Ped. Urine 9 | 2776 | 61.5 | 2.2 | 43.3 | 1.6 |
*The "Within-Lot" precision is a summation of the variances for the lower components: repeatability + between run + between day
13.1.2. Reproducibility
Precision measurements were conducted to evaluate the reproducibility according to CLS! guideline EP05-03. The protocol consisted of testing seven native pediatric urine samples and sample pools and two controls with one reagent lot for five days with two runs per day and three replicates per sample at three sites (one internal, two external) on one cobas c 501 analyzer per site and two operators per site. 30 data points were generated per sample and per site.
Table 5: Reproducibility Summary, per Site and Overall
| Site 1 | Site 2 | Site 3 | Overall | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample | Mean(ng/mL) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) | SD(ng/mL) | CV(%) |
| Control Low | 213 | 8.9 | 4.3 | 4.0 | 1.8 | 8.2 | 3.8 | 9.3 | 4.4 |
| Control High | 532 | 22.0 | 4.2 | 17.1 | 3.2 | 16.1 | 3.0 | 19.5 | 3.7 |
| Ped. Urine 1 | 67 | 4.3 | 6.4 | 3.8 | 5.6 | 4.1 | 6.3 | 4.4 | 6.6 |
| Ped. Urine 2 | 89 | 5.9 | 6.6 | 6.1 | 6.6 | 4.5 | 5.1 | 5.6 | 6.2 |
| Ped. Urine 3 | 133 | 7.8 | 6.0 | 5.8 | 4.3 | 5.5 | 4.1 | 6.5 | 4.8 |
| Ped. Urine 4 | 160 | 7.0 | 4.5 | 7.9 | 4.9 | 5.5 | 3.4 | 7.5 | 4.7 |
| Ped. Urine 5 | 493 | 16.6 | 3.4 | 13.0 | 2.6 | 22.0 | 4.5 | 17.6 | 3.6 |
| Ped. Urine 6 | 1213 | 46.3 | 3.8 | 36.9 | 3.0 | 43.0 | 3.6 | 43.4 | 3.6 |
| Ped. Urine 7 | 2629 | 111.3 | 4.3 | 88.2 | 3.3 | 115.6 | 4.5 | 113.2 | 4.3 |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font.
13.2. Analytical Sensitivity
The Limit of Blank, Limit of Detection and Limit of Quantitation were determined in accordance with CLSI EP17-A2 requirements.
The Limit of Blank, Limit of Detection and Limit of Quantitation were below the lower limit of the measuring range.
The Limit of Blank is determined as the 95th percentile value from n ≥ 60 measurements of analyte-free samples over several independent series. The Limit of Blank corresponds to the concentration below which analyte-free samples are found with a probability of 95% (empirical value determined on the cobas c 501 analyzer: 9 nq/mL).
The Limit of Detection is determined based on the Limit of Blank and the standard deviation of low concentrated samples.
The Limit of Detection corresponds to the lowest analyte concentration which can be detected (value above the Limit of Blank with a probability of 95%) (empirical value determined on the cobas c 501 analyzer: 14 ng/mL).
The Limit of Quantitation is the lowest analyte concentration that can be accurately quantified with a CV ≤ 20%. It has been determined using samples with low NGAL concentrations (empirical value determined on the cobas c 501 analyzer: 18 ng/mL).
13.3. Linearity
The linearity study was conducted to demonstrate that measurements across the claimed measuring range are linear. The study was performed in accordance with CLS! guideline EP06 2™ edition. Three linearity series were tested, each consisting of one pediatric urine samples pool with a high NGAL concentration (above the upper limit of the measuring range) diluted with one pediatric urine sample pool with a low NGAL concentration (below the lower limit of the measuring range) to create 14 samples per series, with concentrations covering and extending the measuring range. Each series was tested in one run with one reagent lot on one cobas c 501 analyzer.
The data support the claim that ProNephro AK/™ (NGAL) measurements are linear between 50 and 3000 ng/mL.
High-Dose Hook Effect 13.4.
High-dose hook effect was evaluated on a single Roche cobas c 501 analyzer. The protocol consisted of testing two native pediatric samples with concentrations of approximately 40,000 ng/mL, using one reagent lot and four replicates per sample.
No sample showed a result that fell within the measuring range. No high-dose hook effect was observed up to 40,000 ng/mL NGAL.
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the logo for "bioporto DIAGNOSTICS". The logo consists of a stylized image of a molecule or network to the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font. The overall design is clean and modern.
13.5. Interference
The effect of the following endogenous substances, contrast agents, pH, and common and special pharmaceutical compounds on the assay's performance were tested. The protocol consisted of testing two pediatric samples pools with NGAL concentrations of approximately 150 ng/mL, and approximately 1,500 ng/mL, respectively, on one cobas c 501 analyzer with one reagent lot and six replicates for the lower sample, and four replicates for the higher sample. Interferences were tested up to the listed concentrations and showed non-significant interference.
Criterion: Recovery within ± 10% of initial value
Table 6: Endogenous Substances
| Compound | Concentration (mg/L) |
|---|---|
| Albumin | 60,000 |
| Ammonium Chloride | 2,000 |
| Bilirubin, conjugated | 800 |
| Calcium Chloride | 2,500 |
| Citric Acid | 2,500 |
| Creatinine | 10,000 |
| Glucose | 14,000 |
| Hemoglobin | 2,500 |
| Immunoglobulin G | 1,100 |
| Magnesium | 2,000 |
| Oxalic Acid | 300 |
| Sodium Phosphate | 4,000 |
| Urea | 38,000 |
| Uric Acid | 1,400 |
| Urobilinogen | 150 |
Table 7: Contrast Agents
| Compound | Concentration (g/L) |
|---|---|
| Visipaque / Iodixanol | 20 |
| Omnipaque / Iohexol | 60 |
| Optiray / loversol | 100 |
Sample pH: pH-values between 3.3 and 10.3 were tested and showed non-significant interference.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for bioporto Diagnostics. The logo consists of a stylized atom-like symbol on the left, followed by the word "bioporto" in a bold, sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter font. The overall design is clean and modern.
Table 8: Common Drugs
| Compound | Concentration (mg/L) |
|---|---|
| Acetaminophen | 3,000 |
| Ascorbic Acid | 4,000 |
| N-Acetylcysteine | 150 |
| Ibuprofen | 4,000 |
| Levodopa | 1,000 |
| Methyldopa | 2,000 |
| Phenazopyridine | 300 |
| Salicyluric Acid | 6,000 |
Table 9: Special Drugs and Immunosuppressive Agents
| Compound | Concentration (mg/L) |
|---|---|
| Cefazolin | 20,000 |
| Cefepime | 65,000 |
| Ceftriaxone | 90,000 |
| Cyclosporine | 200 |
| Diazepam | 500 |
| Diphenhydramine | 15,000 |
| Everolimus | 5 |
| Famotidine | 250 |
| Fentanyl | 100 |
| Furosemide | 3,500 |
| Hydromorphone | 300 |
| Ketamine | 2,500 |
| Ketorolac | 900 |
| Levetiracetam | 25,000 |
| Lorazepam | 300 |
| Midazolam | 3,000 |
| Morphine | 4,000 |
| Mycophenolate Mofetil | 1,000 |
| Mycophenolic Acid Glucuronide (MPAG) | 20,000 |
| Ondansetron | 500 |
| Oxycodone | 1,500 |
| Piperacillin / Tazobactam | 10,000 / 1,250 |
| Prednisone | 3 |
| Prednisolone | 3 |
| Tacrolimus | 2,5 |
| Vancomycin | 20,000 |
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a dark gray sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, lighter gray font.
Cefepime was shown to interfere at concentrations above 4.062.5 mg/L. The maximum daily dose of Cefepime in pediatrics is 150 mg/kg and the calculated maximum concentration in the urine could be up to 6 g/L, depending on the age, weight, and urine output of the patient. The labeling will include a statement that patients being treated with Cefepime may have lower than expected NGAL results as interference was shown above a Cefepime concentration of 4 g/L.
Analytical Specificity / Cross-Reactivity 13.6.
Cross-reactivity studies were performed to determine if other biomarkers that have been reported to be elevated for patients with AKI interfere with ProNephro AKI (NGAL). The testing was performed in alignment with CLSI EP07-A3 requirements. The protocol consisted of testing two pediatric sample pools with NGAL concentrations of approximately 150 ng/mL, and approximately 1,500 ng/mL on one cobas c 501 analyzer with one reagent lot and six replicates for the lower sample, and four replicates for the higher sample. Cross-reactivity was tested up to the listed concentrations and no impact on results was observed.
Criterion: Recovery within ± 10% of initial value
Table 10: Potential Cross Reactants
| Compound | Concentration (mg/L) |
|---|---|
| Alpha 1 Microglobulin (α1Μ) | 1,000 |
| Beta 2 Microglobulin (ß2M) | 15 |
| Cystatin C | 8 |
| Interleukin 18 | 0.001 |
| Kidney Injury Molecule 1 (KIM-1) | 0.015 |
| Liver Type Fatty Acid Binding Protein (L-FABP) | 0.6 |
| N-acetyl H3-D-glucosaminidase (NAG) | 15 U/L |
| Pi-Glutathione s-Transferase (p-GST) | 0.5 |
| Insulin-like Growth Factor binding protein (IGFBP-7) | 0.03 |
| Tissue inhibitor of Metalloproteinase 2 (TIMP-2) | 0.01 |
13.7. Specimen Stability
Analyte stability in pediatric urine samples was studied. Samples remain stable:
- . 1 day at 15-30°C
- . 3 days at 2-8°C
- . 12 months at or below -70°C, up to 2 freeze/thaw cycles
13.8. Reagent Stability and Shelf-Life
Reagent stability and shelf-life studies were performed. The reagent stability studies support the claims as reported in the IFU. The results for the shelf-life studies support a claim of 18 months.
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for bioporto Diagnostics. To the left of the company name is a graphic of a central dot with eight smaller dots surrounding it. The company name is in a sans-serif font, with the word "bioporto" in a larger font than the word "Diagnostics" which is directly below it.
14. Clinical Studies
14.1. Reference Range
In a multi-center, cross sectional, prospectively enrolled reference range study, the ProNephro AKI™ (NGAL) test was used on urine samples from 348 apparently healthy pediatric subjects between ≥ 90 days to < 22 years of age.
The majority of samples in this reference range study vielded results below the lower limit of the measuring range of 50 ng/mL. Even though the true concentration of these samples could be far below 50 ng/mL, the results were reported as "50 ng/mL" for the statistical evaluations.
The following results were obtained for the central 95% reference intervals:
| Table 11: Summary of Results from the Reference Range Study | |||
|---|---|---|---|
| Age groupMALE subjects | N | Median | Central 95%(2.5th – 97.5th) | N / %NGAL results≥125 ng/mL |
|---|---|---|---|---|
| ≥ 90 days to < 22 years (all) | 189 | ≤50 | ≤50 - ≤50 | 0 / 0 |
| ≥ 90 days to < 2 years | 42 | ≤50 | ≤50 - 59 | 0 / 0 |
| ≥ 2 years to < 12 years | 75 | ≤50 | ≤50 - ≤50 | 0 / 0 |
| ≥ 12 years to < 22 years | 72 | ≤50 | ≤50 - ≤50 | 0 / 0 |
| Age groupFEMALE subjects | N | Median | Central 95%(2.5th - 97.5th) | N / %NGAL results≥125 ng/mL |
|---|---|---|---|---|
| ≥ 90 days to < 22 years (all) | 159 | ≤50 | ≤50 - 96 | 2 / 1.26* |
| ≥ 90 days to < 2 years | 31 | ≤50 | ≤50 - 107 | 0 / 0 |
| ≥ 2 years to < 12 years | 64 | ≤50 | ≤50 - 67 | 1 / 1.56* |
| ≥ 12 years to < 22 years | 64 | ≤50 | ≤50 - 101 | 1 / 1.56* |
*Elevated NGAL values might be attributed to a borderline UTI result which did not justify the exclusion of the subject from the reference range study, as per predefined exclusion criteria in the study protocol
Clinical Performance 14.2.
A multi-center prospective clinical study called EARNEST was conducted at 6 clinical sites to establish the NGAL cutoff to identify patients at risk of moderate to severe acute kidney injury (Stage 2 to 3 AKI) 48 to 72 hours after the time of assessment. 257 pediatric patients (≥ 90 days to < 22 years of age) were enrolled in the study with 203 evaluable subjects. The cut-off was established at 125 ng/mL with the sample collection timepoint within the first 24 hours of ICU admission.
The clinical performance of ProNephro AKI™ (NGAL) was validated in a multi-center prospective clinical study called GUIDANCE based on the established cutoff (125 ng/mL) identifying patients at risk of moderate to severe acute kidney injury (Stage 2 to 3 AKI) 48 to 72 hours after the time of assessment. The clinical performance evaluation was performed with 660 pediatric patients (≥ 90 days to < 22 years of
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for "bioporto DIAGNOSTICS". To the left of the text is a graphic of a central dot with eight lines radiating outward, each ending in a dot. The text "bioporto" is in a bold, sans-serif font, and the word "DIAGNOSTICS" is in a smaller, lighter font below and slightly to the right of "bioporto".
age) were enrolled at 15 sites across the US, with 514 evaluable subjects. The clinical data for each subject were captured and provided to at least two independent clinical experts for adjudication. The clinical performance of ProNephro AKI (NGAL) was evaluated against the adjudicated results.
Sensitivity and specificity and their corresponding 95% confidence interval (CI) were calculated to be 72.3% (57.4-84.4%) and 86.3% (82.8-89.3%) respectively. The negative predictive value (NPV) was 96.9% (95.1-98.0% 9) and the positive predictive value (PPV) was 34.7% (28.5-41.5%).
The indication for use does not exclude patients with diagnosed AKI per KDIGO guidelines or patients with elevated SCr values prior to or within the first 12 hrs of ICU admission. These patients were included in the GUIDANCE study. For these patients, the ProNephro AKI (NGAL) test is used to assess the risk of persistent AKI 48 to 72 hrs after the time of the NGAL assessment.
The subgroups of patients within the intended use population can be summarized into patients that present:
-
- Without a SCr value indicative of Stage 2-3 AKI at ICU admission,
- a. AND developing Stage 2-3 AKI 48 to 72 hours after the time of assessment,
- b. OR not developing Stage 2-3 AKI 48 to 72 hours after the time of assessment.
-
- With an elevated SCr value indicative of Stage 2-3 AKI at ICU admission,
- a. AND have persisting Stage 2-3 AKI 48 to 72 hours after the time of assessment,
- b. OR do not have persisting Stage 2-3 AKI 48 to 72 hours after the time of assessment.
The following clinical performance characteristics for sensitivity, negative predictive value, and positive predictive value were calculated for the two subgroups of patients within the intended use population when assessed separately.
| AKI+ | AKI- | Total | |
|---|---|---|---|
| NGAL+ | 11 | 46 | 57 |
| NGAL- | 7 | 358 | 365 |
| Total | 18 | 404 | 422 |
| Prevalence | 18/422 (4.3%) | ||
| Sensitivity(95 CI) | 61.1%(35.8 - 82.7%) | ||
| Specificity(95 CI) | 88.6%(85.1 – 91.5%) | ||
| Positive Predictive Value(95 CI) | 19.3%(10.1 - 31.9%) | ||
| Negative Predictive Value(95 CI) | 98.1%(96.1 - 99.2%) | ||
| Risk of AKI for NGALnegative Results | 1.9% | ||
| Risk of AKI for NGALpositive Results | 19.3% |
Table 11: Group 1 - GUIDANCE Study Results in Patients with Day 0 SCr indicative of no AKI or AKI stage 1 (n=422)
{20}------------------------------------------------
Image /page/20/Picture/1 description: The image shows the logo for Bioporto Diagnostics. The logo consists of a stylized atom-like graphic on the left, followed by the word "bioporto" in a sans-serif font. Below "bioporto" is the word "DIAGNOSTICS" in a smaller, sans-serif font. The logo is simple and modern, and the colors are muted.
| AKI+ | AKI- | Total | |
|---|---|---|---|
| NGAL+ | 23 | 18 | 41 |
| NGAL- | 6 | 45 | 51 |
| Total | 29 | 63 | 92 |
| Prevalence | 29/92 (31.5%) | ||
| Sensitivity | 79.3% | ||
| (95 CI) | (60.3 - 92.0%) | ||
| Specificity | 71.4% | ||
| (95 CI) | (58.7 - 82.1%) | ||
| Positive Predictive Value | 56.1% | ||
| (95 CI) | (39.8 - 71.5%) | ||
| Negative Predictive Value | 88.2% | ||
| (95 CI) | (76.1 - 95.6%) | ||
| Risk of AKI for NGALnegative Results | 11.8% | ||
| Risk of AKI for NGALpositive Results | 56.1% |
Table 12: Group 2 - GUIDANCE Study Results in Patients with Day 0 SCr indicative of AKI stage 2/3 (n=92)
15. Proposed Labeling
The labeling satisfies the requirements of 21 CFR Parts 801 and 809 as well as the special controls for an Acute Kidney Injury Test System.
16. Conclusion
The complete results of the clinical and analytical studies submitted in this 510(k) premarket notification demonstrate that the ProNephro AKI™ (NGAL) test is substantially equivalent to the predicate device.
§ 862.1220 Acute kidney injury test system.
(a)
Identification. An acute kidney injury test system is a device that is intended to measure one or more analytes in human samples as an aid in the assessment of a patient's risk for developing acute kidney injury. Test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must detail an appropriate end user device training program that will be offered while marketing the device as part of your efforts to mitigate the risk of incorrect interpretation of test results.
(2) As part of the risk management activities performed as part of your 21 CFR 820.10(c) design and development activities, you must document the appropriate end user device training program provided in your premarket notification submission to satisfy the special control in paragraph (b)(1) of this section that will be offered while marketing the device as part of your efforts to mitigate the risk of incorrect interpretation of test results.
(3) Robust clinical data demonstrating the positive predictive value, negative predictive value, sensitivity and specificity of the test in the intended use population must be submitted as part of the premarket notification submission.