AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtrue® Multi-Panel Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Phencyclidine, Ecstasy, Methadone, Oxycodone, Propoxyphene , Tramadol and Trivyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

The multi test panels can consist of any analytes listed above in any combination. Only one cut-off concentration will be included per analyte per device.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Antidepressants.

The Chemtrue® Drug Screen Fentanyl / Tramadol Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The Chemtrue® Drug Screen Fentany! / Tramadol Dip Card Test detects and is calibrated against norfentanyl, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentanyl/ Tramadol. The test is for in vitro diagnostic use only.

The Chemtrue® Drug Screen Fentanyl / Tramadol Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. It is an in vitro diagnostic device. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The Chemtrue® Drug Screen Fentany1 / Tramadol Cup Test detects and is calibrated against norfentany1, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to the drug test result, particularly when preliminary positive result is indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentany/ Tramadol. The test is for in vitro diagnostic use only.

The Chemtrue® Multi-Panel Drug Screen Dip Card Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentany, Marijuana, Methadone, Morphine, Opiates, Oxycodone, Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Benzodiazepines, Antidepressants.

The Chemtrue® Multi-Panel Drug Screen Cup Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Oxycodone, Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Tricyclic Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Antidepressants.

Device Description

The Chemtrue® Drug Screen Tests are colloidal gold-based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in Dip Card or Cup formats, as indicated by the test name.

AI/ML Overview

The provided text describes the performance characteristics of the Chemtrue® Drug Screen Fentanyl/Tramadol Cup Test and Dip Card Test. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for qualitative immunoassay diagnostic devices typically involve high levels of agreement with a reference method, especially around the cutoff concentration. While explicit acceptance criteria ("Pass/Fail" thresholds) are not directly stated in the provided text, the reported performance demonstrates very high agreement.

Note: The tables below focus on the Fentanyl (FYL) and Tramadol (TML) tests as these are the new additions being evaluated in this submission. Previous analytes were cleared under older 510(k)s. The percentages under "Reported Device Performance" are calculated from the provided raw counts.

Method Comparison (Accuracy) Studies vs. LC/MS Reference Method

The agreement is calculated as (Number of correct results / Total number of samples in that category) * 100%.

Analyte (Cutoff)Category (LC/MS Conc.)Acceptance Criteria (Implicit: High agreement)Reported Device Performance (Dip Card)Reported Device Performance (Cup Test)
Norfentanyl (5 ng/mL)Positive (Test Positive)100% (35/35)100% (35/35)
Near Cutoff Positive (Cutoff to 150% C/O)100% (6/6)100% (6/6)
Positive (>150% C/O)100% (26/26)100% (26/26)
Negative (Test Negative)93.4% (57/61)95% (58/61)
No drug present100% (22/22)100% (22/22)
<50% of C/O96.7% (29/30)96.7% (29/30)
50% C/O to Cutoff85.7% (6/7)87.5% (7/8)
Tramadol (100 ng/mL)Positive (Test Positive)100% (31/31)100% (31/31)
Near Cutoff Positive (Cutoff to 150% C/O)100% (8/8)100% (8/8)
Positive (>150% C/O)100% (23/23)100% (23/23)
Negative (Test Negative)100% (52/52)100% (52/52)
No drug present100% (20/20)100% (20/20)
<50% of C/O100% (20/20)100% (20/20)
50% C/O to Cutoff100% (12/12)100% (12/12)

OTC Lay-user Accuracy Studies vs. LC/MS Values

Analyte (Cutoff)Category (LC/MS Conc.)Acceptance Criteria (Implicit: High agreement)Reported Device Performance (Dip Card)Reported Device Performance (Cup Test)
Norfentanyl (5 ng/mL)Positive (Test Positive)100% (47/47)100% (45/45)
Near Cutoff Positive (125% C/O)100% (24/24)100% (22/22)
Positive (150% C/O)100% (23/23)100% (23/23)
Negative (Test Negative)100% (139/139)100% (142/142)
No Drug Present100% (93/93)100% (93/93)
Negative (50% C/O)100% (23/23)100% (25/25)
Near Cutoff Negative (75% C/O)100% (23/23)100% (24/24)
Tramadol (100 ng/mL)Positive (Test Positive)100% (46/46)100% (47/47)
Near Cutoff Positive (125% C/O)100% (23/23)100% (24/24)
Positive (150% C/O)100% (23/23)100% (23/23)
Negative (Test Negative)100% (141/141)100% (140/140)
No Drug Present100% (93/93)100% (93/93)
Negative (50% C/O)100% (24/24)100% (24/24)
Near Cutoff Negative (75% C/O)100% (24/24)100% (23/23)

2. Sample Size Used for the Test Set and the Data Provenance

  • Method Comparison (Accuracy) Studies:

    • Sample Size: 176 clinical urine specimens (93 Fentanyl, 83 Tramadol).
    • Data Provenance: Clinical urine specimens, retrospective (since they were confirmed with LC/MS prior to testing with the device). The country of origin is not specified but implied to be within regular clinical testing networks.
  • OTC Lay-user Accuracy Studies:

    • Sample Size: 140 intended lay-users for each device format (Dip Card and Cup). Each lay-user received two blind-coded samples.
    • Data Provenance: Urine samples spiked with drugs at various concentrations (0, 50%, 75%, 125%, 150% of cutoff) into drug-free urine pools. This is a prospective study design for evaluating lay-user performance with controlled samples. The samples were confirmed with LC/MS/MS prior to the study. The study was conducted at three intended user sites (Shopping Mall, School, and Hotel), suggesting diverse user demographics.
  • Precision/Reproducibility Studies:

    • Sample Size: For each analyte (Norfentanyl and Tramadol), 5 replicates at 6 concentration levels (0%, -50%, -25%, Cutoff, +25%, +50% of the cutoff) were tested with 3 lots of each device format (Cup and Dip Card) by 3 operators over 5 consecutive days.
      • This totals to 5 replicates * 6 concentrations * 3 lots * 3 operators = 270 tests per analyte per device format.
      • Thus, for FYL Dip Card: 270 samples. FYL Cup Test: 270 samples. TML Dip Card: 270 samples. TML Cup Test: 270 samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Method Comparison (Accuracy) Studies: The ground truth was established by LC/MS (Liquid Chromatography/Mass Spectrometry), which is a highly accurate analytical chemistry technique considered the "gold standard" for drug concentration determination. The text states: "Each specimen was confirmed with LC/MS assay with the value assignment." and "The accuracy of the Chemtrue® Test devices were evaluated against ... clinical urine specimens. Each specimen was confirmed with LC/MS assay with the value assignment."

    • This is an instrument-based determination, not typically requiring human expert consensus in the same way as, for example, image interpretation. The "experts" here are the drug analysts who performed the LC/MS analysis and validated the results.
    • The study mentions "One drug analyst with one set of blind code in one device format - Dip Card / Cup were tested by one operator." This refers to the operators performing the rapid immunoassay tests, not the LC/MS ground truth determination.
  • OTC Lay-user Accuracy Studies: The ground truth was established by LC/MS/MS ("Concentrations By LC-MS/MS (ng/mL)") of the spiked urine samples. As above, this is an instrument-based method.

  • Precision/Reproducibility Studies: The ground truth for the spiked urine controls was defined by their known concentrations, which were "LC/MS confirmed concentrations."


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • For the Method Comparison (Accuracy) Studies and OTC Lay-user Accuracy Studies, no human adjudication method (like 2+1 or 3+1) is described for establishing the ground truth. The ground truth (LC/MS or LC/MS/MS results) is considered definitive. The rapid immunoassay device results are compared directly to this definitive ground truth.
    • For the actual reading of the rapid immunoassay devices, 4 operators performed the tests for the method comparison study, but their readings were individual comparisons to the LC/MS reference, not adjudicated against each other.
    • For the OTC study, "intended lay-users" performed the interpretation, and their results were compared to the LC/MS/MS values.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done.
  • This device is a rapid lateral flow immunoassay for drug detection, not an AI-powered diagnostic system requiring human interpretation or assistance from AI, so this type of study is not applicable. The immunoassay itself provides a visual result (line/no line) that lay users interpret.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • This question is not applicable as the device is a rapid lateral flow immunoassay, which is a standalone point-of-care test, not an algorithm. The "performance" is the chemical reaction and visual read-out of the test strip/cup. The "human-in-the-loop" is the user interpreting the visual result. However, the studies evaluating its performance involved human users (operators for method comparison, lay-users for OTC study) interpreting the visual output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The primary ground truth used for all performance studies (accuracy, precision, OTC accuracy) was analytical confirmation by LC/MS or LC/MS/MS. This is a highly precise and quantitative laboratory method considered definitive for identifying and quantifying drug metabolites in urine.

8. The Sample Size for the Training Set

  • No training set is described because this device is a rapid lateral flow immunoassay, not a machine learning or AI-based algorithm that requires a training set. The device operates based on chemical reactions and antigen-antibody binding.

9. How the Ground Truth for the Training Set was Established

  • This question is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Chemtron Biotech, Inc. Jane Zhang Vice President of Quality 9245 Brown Deer Road, Suite B San Diego, California 92121

Re: K232736

Trade/Device Name: Chemtrue® Drug Screen Fentany]/Tramadol Cup Test. Chemtrue® Drug Screen Fentanyl/Tramadol Dip Card Test, Chemtrue® Multi-Panel Drug Screen Cup Test. Chemtrue® Multi-Panel Drug Screen Dip Card Test. Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG, DKZ, DIS, JXM, DIO, DJC, LDJ, LAF, DJR, DNK, LCM, JXN, LFG Dated: November 16, 2023 Received: November 17, 2023

Dear Jane Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Digitally signed by Joseph Joseph A. Kotarek -S Date: 2023.12.20 10:46:07 -05'00' Joseph Kotarek Branch Chief Toxicology Branch Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232736

Device Name

Device Name: Chemtrue® Drug Screen Multi-Panel Dip Card Tests

Indications for Use (Describe)

The Chemtrue® Multi-Panel Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Phencyclidine, Ecstasy, Methadone, Oxycodone, Propoxyphene , Tramadol and Trivyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Conc. (ng/mL)
AmphetamineAMPd-Amphetamine300
AmphetamineAMPd-Amphetamine500
AmphetamineAMPd-Amphetamine1,000
BarbituratesBARSecobarbital/Pentobarbital200
BarbituratesBARSecobarbital/Pentobarbital300
BenzodiazepinesBZOOxazepam200
BenzodiazepinesBZOOxazepam300
BuprenorphineBUPBuprenorphine10
CocaineCOCBenzoylecgonine150
CocaineCOCBenzoylecgonine300
EcstasyMDMAd,l-Methylenedioxy methamphetamine500
NorfentanylFYLNorfentanyl5
MethamphetamineMETd-Methamphetamine300
MethamphetamineMETd-Methamphetamine500
MethamphetamineMETd-Methamphetamine1,000
MarijuanaTHC11-nor-Δ9-THC-9-COOH50
MethadoneMTDMethadone300
MorphineMORMorphine300
OpiatesOPIMorphine2,000
OxycodoneOXYOxycodone100
PhencyclidinePCPPhencyclidine25
PropoxyphenePPXPropoxyphene300
Tricyclic AntidepressantsTCANortriptyline1,000
TramadolTMLTramadol100

The multi test panels can consist of any analytes listed above in any combination. Only one cut-off concentration will be included per analyte per device.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Antidepressants.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K232736

Device Name

Device Name: Chemtrue® Drug Screen Multi-Panel Cup Tests

Indications for Use (Describe)

The Chemtrue® Multi-Panel Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Norfentaryl, Marijuana, Methamphetamine, Morphine, Opiates, Phencyclidine, Ecstasy, Methadone, Oxycodone, Propoxyphene , Tramadol and Trivyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Conc. (ng/mL)
AmphetamineAMPd-Amphetamine300
AmphetamineAMPd-Amphetamine500
AmphetamineAMPd-Amphetamine1,000
BarbituratesBARSecobarbital/Pentobarbital200
BarbituratesBARSecobarbital/Pentobarbital300
BenzodiazepinesBZOOxazepam200
BenzodiazepinesBZOOxazepam300
BuprenorphineBUPBuprenorphine10
CocaineCOCBenzoylecgonine150
CocaineCOCBenzoylecgonine300
EcstasyMDMAd,l-Methylenedioxy methamphetamine500
NorfentanylFYLNorfentanyl5
MethamphetamineMETd-Methamphetamine300
MethamphetamineMETd-Methamphetamine500
MethamphetamineMETd-Methamphetamine1,000
MarijuanaTHC11-nor-Δ9-THC-9-COOH50
MethadoneMTDMethadone300
MorphineMORMorphine300
OpiatesOPIMorphine2,000
OxycodoneOXYOxycodone100
PhencyclidinePCPPhencyclidine25
PropoxyphenePPXPropoxyphene300
Tricyclic AntidepressantsTCANortriptyline1,000
TramadolTMLTramadol100

The multi test panels can consist of any analytes listed above in any combination. Only one cut-off concentration will be included per analyte per device.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Antidepressants.

{5}------------------------------------------------

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Indications for Use

510(k) Number (if known) K232736

Device Name

Device Name: Chemtrue® Drug Screen Fentanyl / Tramadol Dip Card Tests

Indications for Use (Describe)

The Chemtrue® Drug Screen Fentany! / Tramadol Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
NorfentanylFYLNorfentanyl5
TramadolTMLTramadol100

The Chemtrue® Drug Screen Fentany! / Tramadol Dip Card Test detects and is calibrated against norfentanyl, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentanyl/ Tramadol. The test is for in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

==============================================================================================================================================================================

{7}------------------------------------------------

Indications for Use

510(k) Number (if known) K232736

Device Name

Device Name: Chemtrue® Drug Screen Fentanyl / Tramadol Cup Tests

Indications for Use (Describe)

The Chemtrue® Drug Screen Fentanyl / Tramadol Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
NorfentanylFYLNorfentanyl5
TramadolTMLTramadol100

The Chemtrue® Drug Screen Fentanyl / Tramadol Cup Test detects and is calibrated against norfentany1, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentany/ Tramadol. The test is for in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{8}------------------------------------------------

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

A. SUBMITTER: Chemtron Biotech, Inc. 9245 Brown Deer Road, Suite B, San Diego, CA

  1. USA.

TEL: 858-450-0044:

FAX: 858-450-0046

Contact Person: Jane Zhang, Vice President of Quality Official FDA Correspondent

9245 Brown Deer Road, Suite B San Diego, CA 92121

Office: (858) 450-0044; FAX: (858) 450-0046

Email: jane(@uschemtronbio.com

Date: September 06, 2023

B. DEVICE

Trade or Proprietary Name: Chemtrue® Drug Screen Fentanyl/Tramadol Cup Test,

Chemtrue® Drug Screen Fentanyl/Tramadol Dip Card Test,

Chemtrue® Multi-Panel Drug Screen Cup Test,

Chemtrue® Multi-Panel Drug Screen Dip Card Test.

Common Name: Single/Multi-Drugs of Abuse Screen Urine Tests

Regulatory Class: Class II

Regulatory Information:

Drug of AbuseProduct CodePanelRegulation Section
Fentanyl (FYL)DJGToxicology 9121 CFR 862.3650, Opiate test system.
Tramadol (TML)DJGToxicology 9121 CFR 862.3650, Opiate test system.

These devices also incorporate the assays previously cleared under 510(k) numbers K153192 and K142396, which consist of any combinations of the following drug tests:

Drug of AbuseProductCodePanelRegulation Section
AmphetamineDKZToxicology 9121CFR 862.3100, Amphetamine Test System
BarbituratesDISToxicology 9121 CFR 862.3150, Barbiturates Test System
BenzodiazepinesJXMToxicology 9121 CFR 862.3170, Benzodiazepines Test System
Buprenorphine (BUP)DJGToxicology 9121CFR 862.3650, Opiate Test System
CocaineDIOToxicology 9121 CFR 862.3250, Cocaine and metabolites Test System
Ecstasy (MDMA)DJCToxicology 9121 CFR 862.3610, Methamphetamine Test System
MarijuanaLDJToxicology 9121 CFR 862.3870, Cannabinoids Test System
MethamphetamineLAFToxicology 9121 CFR 862.3610, Methamphetamine Test System
MethadoneDJRToxicology 9121 CFR 862.3620, Methadone Test System
MorphineDNKToxicology 9121 CFR 862.3640, Morphine Test System
OpiatesDJGToxicology 9121 CFR 862.3650, Opiate Test System
OxycodoneDJGToxicology 9121 CFR 862.3650, Opiate Test System

{9}------------------------------------------------

PhencyclidineLCMToxicology 91Unclassified, Enzyme immunoassay Phencyclidine
PropoxypheneJXNToxicology 9121 CFR 862.3700 Propoxyphene test system
Tricyclic Antidepressants (TCA)LFGToxicology 9121 CFR 862.3910, Tricyclic antidepressant drugs test system.

C. PREDICATE DEVICES

K153192: Chemtrue® Multi-Panel Drug Screen Cup/Dip Card Tests

D. INDICATIONS FOR USE:

The Chemtrue® Drug Screen Fentanyl / Tramadol Dip Card Test is a rapid lateral flow immunoassay for the qualitative detection of Norfentanyl (FYL) 5 and Tramadol (TML) 100 drugs in human urine. It is an in vitro diagnostic device. The test cut-off concentrations and calibrators are listed below:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
NorfentanylFYLNorfentanyl5
TramadolTMLTramadol100

The Chemtrue® Drug Screen Fentanyl / Tramadol Dip Card Test detects and is calibrated against norfentanyl, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to the drug test result, particularly when preliminary positive result is indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentany// Tramadol. The test is for in vitro diagnostic use only.

The Chemtrue® Drug Screen Fentany] / Tramadol Cup Test is a rapid lateral flow immunoassay for the qualitative detection of Norfentany] (FYL) 5 and Tramadol (TML) 100 drugs in human urine. It is an in vitro diagnostic device. The test cut-off concentrations and calibrators are listed below:

AnalyteAbbreviationCalibratorCutoff Concentration
(ng/mL)
NorfentanylFYLNorfentanyl5
TramadolTMLTramadol100

The Chemtrue® Drug Screen Fentanyl / Tramadol Cup Test detects and is calibrated against norfentanyl, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to the drug test result, particularly when preliminary positive result is indicated.

The test is not intended to differentiate between drugs of abuse and prescription use of Fentany/ Tramadol. The test is for in vitro diagnostic use only.

{10}------------------------------------------------

The Chemtrue® Multi-Panel Drug Screen Dip Card Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentany, Marijuana, Methadone, Morphine, Opiates, Oxycodone, Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
AmphetamineAMPd-Amphetamine300
AmphetamineAMPd-Amphetamine500
AmphetamineAMPd-Amphetamine1,000
BarbituratesBARSecobarbital/Pentobarbital200
BarbituratesBARSecobarbital/Pentobarbital300
BenzodiazepinesBZOOxazepam200
BenzodiazepinesBZOOxazepam300
BuprenorphineBUPBuprenorphine10
CocaineCOCBenzoylecgonine150
CocaineCOCBenzoylecgonine300
EcstasyMDMAd,l-Methylenedioxy methamphetamin500
NorfentanylFYLNorfentanyl5
MethamphetamineMETd-Methamphetamine300
MethamphetamineMETd-Methamphetamine500
MethamphetamineMETd-Methamphetamine1,000
MarijuanaTHC11-nor-Δ9-THC-9-COOH50
MethadoneMTDMethadone300
MorphineMORMorphine300
OpiatesOPIMorphine2,000
OxycodoneOXYOxycodone100
PhencyclidinePCPPhencyclidine25
PropoxyphenePPXPropoxyphene300
Tricyclic AntidepressantsTCANortriptyline1,000
TramadolTMLTramadol100

The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Benzodiazepines, Antidepressants.

The Chemtrue® Multi-Panel Drug Screen Cup Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Oxycodone,

{11}------------------------------------------------

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
AmphetamineAMPd-Amphetamine300
AmphetamineAMPd-Amphetamine500
AmphetamineAMPd-Amphetamine1,000
BarbituratesBARSecobarbital/Pentobarbital200
BarbituratesBARSecobarbital/Pentobarbital300
BenzodiazepinesBZOOxazepam200
BenzodiazepinesBZOOxazepam300
BuprenorphineBUPBuprenorphine10
CocaineCOCBenzoylecgonine150
CocaineCOCBenzoylecgonine300
EcstasyMDMAd,l-Methylenedioxy methamphetamin500
NorfentanylFYLNorfentanyl5
MethamphetamineMETd-Methamphetamine300
MethamphetamineMETd-Methamphetamine500
MethamphetamineMETd-Methamphetamine1,000
MarijuanaTHC11-nor-Δ9-THC-9-COOH50
MethadoneMTDMethadone300
MorphineMORMorphine300
OpiatesOPIMorphine2,000
OxycodoneOXYOxycodone100
PhencyclidinePCPPhencyclidine25
PropoxyphenePPXPropoxyphene300
Tricyclic AntidepressantsTCANortriptyline1,000
TramadolTMLTramadol100

Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Tricyclic Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Antidepressants.

E. DEVICE DESCRIPTION

The Chemtrue® Drug Screen Tests are colloidal gold-based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in Dip Card or Cup formats, as indicated by the test name.

ட் SUBSTANTIAL EQUIVALENCE INFORMATION:

{12}------------------------------------------------

ItemDevicePredicate - K153192
Indication(s) for UseSameQualitative determination of drugs of abuse in human urine.
Specimen TypeSameHuman urine
Type of TestSameQualitative test of in vitro diagnostic device
Methodology /TechnologicalCharacteristicsSameLateral flow, antibody and antigen competitive binding qualitativeimmunoassay.
ConfigurationsSameCup and Dip Card
StorageSame2 - 30°C

Comparison with the predicate devices is outlined below:

ItemCandidate DeviceDifferences
Predicate - K153192
Calibrator/Cut OffNorfentanyl 5 ng/mLTramadol 100 ng/mLAMP at300/500/1000 ng/mLBAR at200/ 300ng/mLBZO at200/ 300ng/mLBUP at10 ng/mLCOC at150/300ng/mLMDMAat 500ng/mLMET at300/500/1000 ng/mL
MTD at300 ng/mLMOR at300 ng/mLOPI at2000 ng/mLOXY at100 ng/mLPCP at25 ng/mLPPX at300 ng/mLTCA at1000 ng/mL
THC at 50 ng/mL

G. TEST PRINCIPLE

These devices are rapid lateral flow immunoassays in which chemically modified drugs (drugprotein conjugates) compete with drugs that may be present in urine. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band indicates a negative result. If the target drug level exceeds its cutoff concentration, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibodycolloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

Monoclonal anti-drug antibodies that are used on the Fentanyl and Tramadol Test devices are derived from mouse.

{13}------------------------------------------------

H. PERFORMANCE CHARACTERISTICS

Performance data of AMP300/500, BAR200, BZO200, COC150, MET300/500 and PPX were presented in the cleared 510(k) K153192 submission. AMP1000, BAR3000, BZO300, BUP10, COC300, MET1000, MOR, PCP, THC, MDMA, MTD, OPI2000, OXY and TCA analytes were previously cleared under K143599 and K142396. In this submission, Fentanyl (FYL) and Tramadol (TML) study data are summarized and presented below:

    1. Precision/ Reproducibility Studies:
      Precision/reproducibility studies were carried out for Norfentanyl and Tramadol spiked urine controls with LC/MS confirmed concentrations of 0%, -50%, 100%, +25% and +50% of the drug cut-off. All the aliquots were blindly coded by the research coordinator who prepared the samples and didn't perform the testing. These blind coded samples were used for FYLS and TML100 tests in Cup and Dip Card formats in a randomized order. Five (5) replicates at each control level were tested with three (3) lots each device format of FYL and TML Drug Screen Cup/Dip Card Tests by each of the three operator tested the blind-labeled samples according to proposed package inserts, record the test results and sign on the data collection form. The study was performed in five (5) consecutive days. The data is analyzed and summarized in the tables below:
Lot No.0-50%Cut off-25%Cut offCut Off+25%Cut off+50%Cut off
Lot 110-/0+10-/0+10-/0+8+/2-10+/0-10+/0-
Lot 210-/0+10-/0+10-/0+9+/1-10+/0-10+/0-
Lot 310-/0+10-/0+10-/0+8+/2-10+/0-10+/0-
SUM30-/0+30-/0+30-/0+25+/5-30+/0-30+/0-
Table 1b. TML Dip Card Test: Cutoff: 100 ng/mL
Lot No.0-50%Cut off-25%Cut offCut Off+25%Cut off+50%Cut off
Lot 110-/0+10-/0+10-/0+9+/1-10+/0-10+/0-
Lot 210-/0+10-/0+10-/0+7+/3-10+/0-10+/0-
Lot 310-/0+10-/0+10-/0+9+/1-10+/0-10+/0-
SUM30-/0+30-/0+30-/0+25+/5-30+/0-30+/0-
Table 1c. FYL Cup Test: Cutoff: 5 ng/mL
Lot No.0-50%Cut off-25%Cut offCut Off+25%Cut off+50%Cut off
Lot 110-/0+10-/0+10-/0+9+/1-10+/0-10+/0-
Lot 210-/0+10-/0+10-/0+9+/1-10+/0-10+/0-
Lot 310-/0+10-/0+10-/0+8+/2-10+/0-10+/0-
SUM30-/0+30-/0+30-/0+26+/4-30+/0-30+/0-
Table 1d. TML Cup Test: Cutoff: 100 ng/mL
Lot No.0-50%Cut off-25%Cut offCut Off+25%Cut off+50%Cut off

Table 1a. FYL Dip Card Test: Cutoff: 5 ng/mL

{14}------------------------------------------------

Lot 110-/ 0+10-/ 0+10-/ 0+9+/ 1.10+/ 0-10+/ 0-
Lot 210-/ 0+10-/ 0+10-/ 0+7+/ 3-10+/ 0-10+/ 0-
Lot 310-/ 0+10-/ 0+10-/0+8+/ 2.10+/ 0-10+/ 0-
SUM30-/ 0+30-/ 0+30-/ 0+24+/ 6-30+/ 0-30+/ 0-
    1. Specificity Study: These studies were conducted by adding various drugs, drug metabolites, and other structurally-similar compounds likely to be present in the actual urine specimen.
      The following structurally-related compounds were tested for cross-reactivity and found to be positive if the levels were greater than the following listed concentrations:

Norfentanyl 5 related compounds:

SubstancesConcentration ng/mL% Cross Reactivity
Norfentanyl5100
Fentanyl10100
4-Fluoro-isobutyryl Fentanyl>20,000<1
9-HydroxyRisperidone10,000<1
Acetyl Fentanyl2002.5
Acetyl Norfentanyl2002.5
(±)-β-Hydroxythiofentanyl Hcl2025
Acryl Fentanyl3016.7
Alfentanil1,000<1
Butyryl Fentanyl1533.3
Carfentanil Oxalate>10,000<1
Cis-d, I 3-Methylfentanyl707.1
Despropionylfentanyl (4-ANPP)>20,000<1
Furanyl Fentanyl806.25
Isobutyryl Fentanyl5,000<1
Labetalol Hydrochloride>100,000<1
MT-457,500<1
Norcarfentail Oxalate>20,000<1
Ocfentanil1,000<1
Para-fluoro butyrl Fentanyl (P-FBF)2025
para-Fluorofentanyl1050
Remifentanil>20,000<1
Risperidone10,000<1
Sufentanil3,000<1
Thienyl Fentnayl4012.5
Trans-d, I 3-Methylfentanyl3016.7
Trazodone>100,000<1
U-47700>100,000<1
Valeryl Fentanyl>100,000<1

Note: The cross-reactivity of o-1-Hydroxyfentanyl was not tested in this product and may be a source of false positive results.

Tramadol 100 related compounds:

{15}------------------------------------------------

SubstancesConcentration ng/mL% Cross Reactivity
Tramadol100100
n-Desmethyl Tramadol40025
o-Desmethyl Tramadol1,00010
Venlafaxine>100,000<1
o-Desmethyl Venlafaxine>10,000<1

3. Interference:

Following potential interferents were tested with one lot each of the test device format. It was confirmed not to cross-react, when tested at concentrations of 100 µg/mL at ±25% of the drug cutoff concentrations:

Endogenous Compounds:

AlbuminCreatininer-GlobulinOctopamine
AcetoneDopamineHemoglobinRiboflavin
Ascorbic AcidEthanolHuman serum AlbuminSodium Chloride
AtropineGalactoseB-Hydroxybutyric Acid(F)Uric Acid
BilirubinGlucoseOxalic AcidUrea
Cholesterol

Structurally unrelated compounds:

Amlodipine besylateI-ErythromycinNorethindrone
7-AminonitrazepamEstradiolNoscapine
AmoxicillinEstroneOctopamine
AmpicillinFenfluraminePapaverine
ApomorphineFenofibratePenicillin-G
AspirinFluphenazine(F)Pentazocine
AspartameFotemustinePerphenazine
BaclofenFurosemidePhenelzine
Benzocaine6GemfibrozilPhenylethylamine
BenzylpiperiazineGuaiacolglyceryl etherPhentermine
Benzoic AcidGentisic acidPrednisone
4-Bromo-2,5,DimethoxyphenethylamineHexobarbitalPromazine
CarisoprodolHydralazinePromethazine
ClomipramineHydrocortisonePropoxyphene
Cetirizine3-HydroxytyraminePropranolol
Chloramphenicolβ-Hydroxybutyric AcidPyridoxine
ChlordiazepoxideIbuprofenPyrilamine
Chlorpheniramined,l-IsoproterenolPyrogallol
ChlorpromazineKetamineNorpropoxyphene
ClofibrateLamotrigineQuinidine
ClonidineLisinoprilQuinine
CortisoneLoratidineQuinolinic Acid
I-CotinineMaprotilineRanitidine
Creatine HydrateMeprobamateSalicylic Acid
CyclobenzaprineMetoprololSodium Azide
Cyclodextrin-rMethapyrileneSulfamethazine
CyproheptadineMethylphenidateSulindac
DemoxepamNalidixic AcidTetracycline

{16}------------------------------------------------

DeoxycorticosteroneNaloxoneTetrahydrozoline
DextromethorphanNaltrexoneThiamine
DiclofenacNaproxenThioridazine
DiflunisalNiacinamideTrifluoromethylphenyl-piperazine
Dimethyl-aminoantipyrineN-desmethylapentadolTrifluoperazine
DiphenhydramineNicotinic AcidTryptamine
DiphenylhydantoinNifedipineTyramine
Zolpidem

The following compounds showed no interference at the concentrations below:

CompoundConc. (ng/mL)CompoundConc. (ng/mL)
Acetylsalicylic Acid500,000Metformin25,000
7-Aminoflunitrazepam25,000Norpseudoephedrine25,000
Bupropion25,000Oxazepam Glucuronide25,000
Caffeine500,000Sufentanil Citrate25,000
Carbamazepine25,00011-nor-9 carboxy THC50,000
Lorazepam Glucuronide50,000L-thyroxine25,000
LSD50,000Zolpidem Tartrate50,000

In addition, further testing with the following opioids compounds /drugs at a concentration of 100 ug/mL in ±25% of the cutoff drug urine controls confirm that no interference or crossreactivity was observed with the Chemtrue® Drug Screen FYL/TML Tests.

6-AcetylmorphineHydromorphoneNorcodeine
AmphetamineLevorphanolNorketamine
BuprenorphineMethadoneNormorphine
CodeineMorphineNoroxycodone
DihydrocodeineMorphine-3-glucuronideOxycodone
EDDPNaloxoneOxymorphone
FluoxetinenorbuprenorphinePentazocine (Talwin)
HeroinMorphineTramadol
HydrocodoneMorphine-3-glucuronide

The following opioids compounds /drugs do not interfere or cross-react at a concentration of 10 ug/mL in ±25% of the cutoff drug urine controls:

MeperidineNormeperidineRisperidone
----------------------------------------

Usability Interference Study: In addition to the cross-reactivity and interference studies presented above in this submission, the drug tests were tested with each of the drug analytes at 150% and 50% of the drug cut-off urine samples. The results confirmed that the no interference or cross-reactivity among these drug tests.

Table 3-3. Usability Interference Study Result Summary:

Targeted Drug Test/Cutoff(ng/mL)Chemtrue® Drug Screen FYL/TML Test Results with AMP/BAR/BZO/BUP/COC/MDMA/MET/MOR300/MTD/OPI2000 (In separate device)/ OXY/PCP/PPX/TCA/THC Test Strips50% of the Targeted Drug Cutoff(PAC: All the results must be -)150% of the Targeted Drug Cutoff(PAC: Only the targeted drug tests shouldbe + ; The other test strips must be -)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{17}------------------------------------------------

# of "+" Results# of "-" ResultsNumber of "-"# of "+" Results# of "-"
from the targetedfrom the targetedresults from thefrom the targetedResults
drug test stripsdrug test stripsother test stripsdrug test stripsfrom the other
test strips
FYL 5Calibrator Norfentanyl 501218012180
TML 100 calibrator Tramadol01218012180
AMP 300/500/1000 d-Amphetamine03627036270
BAR 300 Calibrator Pentobarbital01218012180
Calibrator Secobarbital01218012180
BAR200 3 clinical samples test dataN/A0012180
BZO300: Calibrator Oxazepam01218012180
BZO200: 2 clinical samples test dataN/A008120
TCA 1000 Calibrator Nortripytline01218012180
4 clinical samples test dataN/A0016240
COC150/300 Calibrator Benzoylecgonine02436012180
BUP 10 Calibrator Buprenophine01218012180
MDMA500 Calibrator d,1-Methylenedioxymethamphetamine01218012180
MTD 300 Calibrator Methadone01218012180
MOR300/OPI2000 Calibrator Morphine01218012180
PPX 300 Calibrator Propoxyphene01218012180
MET300/500/1000 d-Methamphetamine03627036270
OXY100 Calibrator Oxycodone01218012180
PCP 25 Calibrator Phencyclidine01218012180
THC50 Calibrator 11-nor-A°-THC9-COOH01218012180
SUBTOTAL02463,4202883,780
TOTAL3,6724,068

CONCLUSION: The usability interference study results demonstrate that no interference or cross reactivity among the 17 drug tests of the Chemtrue® Drug Screen Tests, when testing 50% of the drug cut off. No false positive result was presented. When testing with 150% of the drug cut off urine samples, only the targeted drug tests presented positive results, including the class-drug clinical samples, such as BAR, BZO and TCA that contain multiple drugs in the drug class. No false negative or false positive results were observed.

    1. Effect of Urine pH and Specific Gravity Studies: The testing results demonstrate that the urine pH ranges from 2.0 to 9.0 at ±25% of the drug cut-off concentrations do not affect the test performance. The specific gravity (SG) ranges of 1.010, 1.025 and 1.030 at ±25% of the drug cut-off concentrations do not affect the test results.
    1. Stability Study: To establish and support the shelf life and expiration date, stability studies were conducted under accelerated temperature (at 60°C, 50°C and 40°C), and real time (2°C to 30°C) with three (3) lots of each device format. The stability study results support two (2) years shelf life of the products at (2℃ to 30℃). The real time stability study is still on going.

{18}------------------------------------------------

To validate robustness of the Chemtrue® Drug Screen FYL/TML Test devices, studies were conducted after the devices (Cup and Dip Card Tests) were exposed to various lighting, humidity and temperature conditions. The results support the robustness of the Chemtrue® Devices.

    1. Method Comparison (Accuracy) Studies:
      Chemtrue® Drug Screen FYL/TML Tests were compared to the LC/MS Reference Method. The accuracy of the Chemtrue® Test devices were evaluated against one hundred and seventy-six (176) clinical urine specimens. Each specimen was confirmed with LC/MS assay with the value assignment. Ninety-three (93) Fentanyl clinical urine specimens and eighty-three (83) Tramadol urine specimens in blind-code label were tested in this correlation study against the LC/MS reference method. Four (4) operators performed the tests. One drug analyst with one set of blind code in one device format - Dip Card / Cup were tested by one operator. Each blind-labeled sample was randomly distributed to each operator for each device format by the Clinical Research Cooperator. The results are summarized in the tables below:
ChemtrueⓇDrug ScreenDip CardConcentration By LC/MS (ng/mL)%Agreement
(-) Negative
No drugpresent<50% ofthe C/O50% of theC/O to cutoff(+) Near cutoff positive(Cutoff to 150% ofthe C/O)(+) LC/MS Positive(>150% of thecutoff)
FYL 5(+)013626100%
(-)222960093.4%
TML100(+)000823100%
(-)20201200100%

Table 6a. Method comparison study summary: Chemtrue® Drug Screen Dip Card Test results vs LC/MS

Table 6b. Method comparison study summary - Chemtrue® Drug Screen Cup Test results vs LC/MS

Chemtrue®Drug ScreenCupConcentration By LC/MS (ng/mL)% Agreement
No drug present(-)< 50% of the C/ONegative50% of the C/O to cutoff(+)Near cutoff positive(Cutoff to 150% of the C/O)Positive(>150% of the C/O)
FYL5 (+)012626100%
FYL5 (-)222970095%
TML100 (+)000823100%
TML100 (-)20201200100%

DISCORDANT RESULTS:

Cutoff Value (ng/mL)Analyte assay(POS/NEG)Drug AnalyteLC/MS Value(ng/mL)
Norfentanyl 5 Cup Test+Norfentanyl3.38
Norfentanyl 5 Cup Test+Norfentanyl2.03*
Norfentanyl 5 Cup Test+Norfentanyl4.94
Norfentanyl 5 Dip Card Test+Norfentanyl3.38
Norfentanyl 5 Dip Card Test+Norfentanyl2.03*

{19}------------------------------------------------

+Norfentanyl3.97
+Norfentanyl4.94

Five (5) discordant results were at or near the norfentanyl drug cutoff level of 5 ng/mL. One sample (it was tested with one Cup and one Dip Card device to present two test data) with a norfentanyl concentration of 2.03 ng/mL and Fentanyl concentration of 15.43 ng/ml returned a posstive result. Further analysis indicated that the cross-reactive level of Fentanyl is 10 ng/ml. So the positive result could have been caused by the high concentration of Fentanyl in the urine sample.

    1. OTC Lay-user Accuracy Studies:
      One hundred and forty (140) intended lay-users participated in the evaluation for each of the device format (Dip Card and Cup) for OTC accuracy and usability study from three (3) intended user sites (Shopping Mall, School and Hotel) with LC/MS confirmed urine sample concentrations are consisted of no drug present (0), 50%, 75%, 125% and 150% of the cutoff by spiking drugs into drug-free urine pool. Each sample was aliquot into an individual blind-labeled container. Each lay-user was provided with a package insert in English only and two (2) random blind-coded samples with one each of the device format. The results are summarized in tables below:
Chemtrue®Drug Screen DipCard TestNo DrugPresentConcentrations By LC-MS/MS (ng/mL)%Agreement
(-)(+)
Negative(50% of the C/O)Near cutoffNegative(75% of the C/O)Near cutoffPositive(125% of the C/O)Positive(150% of the C/O)
FYL 5 (+)0002423100%
FYL 5 (-)93232300100%
TML 100 (+)0002323100%
TML 100 (-)93242400100%

Table 7a. OTC Accuracy study summary between Chemtrue® Dip Card Tests and the LC/MS values

Table 7b. OTC Accuracy study summary between Chemtrue® Cup Tests and the LC/MS values

Chemtrue®Drug Screen CupTestConcentrations By LC/MS (ng/mL)%Agreement
(-)(+)
No DrugPresentNegative(50% of the C/O)Near cutoffNegative(75% of the C/O)Near cutoffPositive(125% of the C/O)Positive(150% of the C/O)
FYL 5(+)0002223100%
(-)93252400100%
TML 100(+)0002423100%
(-)93242300100%

The results demonstrate that the agreement between the Chemtrue® Drug Screen FYL/TML test devices and LC/MS assayed values is 100%.

These lay-users were also given surveys on the ease of understanding the package insert instructions. The results demonstrate that over 92% of the lay users can easily follow the instructions to perform the test and interpret the results. A Flesch-Kincaid reading analysis supports a 7th grade reading level.

{20}------------------------------------------------

I. CONCLUSION:

Based on the test principle and performance characteristics of the device, it is concluded that the Chemtrue® Drug Screen FYL/TML Test devices are substantially equivalent to the predicate.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).