(87 days)
The Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
The Intense pulsed light therapy device is an intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 650-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin. IPL Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores. Based on this, the Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.
The document is a 510(k) Summary for an Intense Pulsed Light (IPL) therapy device (Model: FI-L06) seeking FDA clearance for permanent hair removal. It compares the proposed device to a predicate device (K192519) to establish substantial equivalence.
Here's an analysis based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a numerical or pass/fail format for clinical performance. Instead, it relies on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of technical specifications. The key performance comparison is presented in tables 6-2 and 6-3.
| Feature/Parameter | Proposed Device Performance | Predicate Device (K192519) Performance | Difference Analysis |
|---|---|---|---|
| Product Code | ONF | ONF | SAME |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Class | II | II | SAME |
| Indication for Use | Surgical and aesthetic applications in permanent hair removal. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment. | Surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment. | SAME (for permanent hair removal indication) |
| Light source | Intense pulsed light (Xenon Lamp) | Intense pulsed light | Analysis 1: Both use Xenon lamps; considered SAME type. |
| Wavelength | 650-1200nm | 520-650nm, 800-1200nm, 540-800nm, 640-1200nm | SAME (for hair removal, proposed device's range is within the broader predicate ranges) |
| Deliver system | Sapphire | Sapphire | SAME |
| Energy density | 1-50J/cm² | 1-50 J/cm² | SAME |
| Pulse Width | 0.1-25ms | 1-25ms | SAME (within the common range) |
| Max. Power | 1500W | 3500VA | Analysis 2: Different, but non-clinical tests (AAMI/ANSI/ES 60601-1 and IEC 60601-1-2) showed no safety/effectiveness issues. |
| Spot size | 8×40mm | Small: 40×12 mm, Large: 46×16mm, Ex-Large: 60×20mm | Analysis 3: Different, but only affects treatment area, not therapeutic effect. |
| Wavelength Range (for Hair removal) | 650-1200 | 640-1200 | SAME |
| Energy Range (for Hair removal) | 1-50 | 5-40 | Analysis 4: Slight difference, but within allowable error range of predicate and bench tests support substantial equivalence. |
| Pulse Width (for Hair removal) | 0.1-25 | 1-25 | SAME |
| Spot Size (for Hair removal) | 8×40 | Small: 40×12 mm, Large: 46×16mm, Ex-Large: 60×20mm | Different (same as above for spot size analysis). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. The submission explicitly states: "No clinical study is included in this submission." The evaluation for substantial equivalence relies on non-clinical (bench) testing and comparison to the predicate device's established performance and specifications.
- Data Provenance: Not applicable, as no clinical data was submitted. Non-clinical tests were conducted by the manufacturer, Hebei Zhemai Technology Co., Ltd., in China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Not applicable. No clinical test set requiring expert ground truth was created for this submission. The submission relies solely on non-clinical testing and comparison to an existing predicate device.
4. Adjudication Method for the Test Set
- Not applicable. No clinical test set requiring adjudication was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is an Intense Pulsed Light therapy device for hair removal, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant for this device type.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This device is a physical IPL therapy machine; it does not involve an algorithm running in a standalone mode without human interaction, nor an AI component to evaluate. Its operation is managed by a user.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- Not applicable. No clinical ground truth was established or used as no clinical studies were performed. The "ground truth" for demonstrating substantial equivalence was based on bench testing and comparison against recognized industry standards (IEC 60601-1-2:2020, AAMI/ANSI ES 60601-1: 2012, IEC 60601-2-57: 2011, ISO 10993-5: 2009, ISO 10993-10: 2010, ISO 14971: 2019) and the established specifications and indications for use of the predicate device.
8. The Sample Size for the Training Set
- Not applicable. No AI model or algorithm requiring a training set was part of this submission.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. No training set was used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
December 1, 2023
Hebei Zhemai Technology Co.,Ltd Rav Wang Official Correspondent 1st FL, N of No.1 Warehouse, W of Shuichang Rd, N of Beiyi Rd, Yanjiao Development Zone, Sanhe Langfang, Hebei 065201 China
Re: K232708
Trade/Device Name: Intense pulsed light therapy device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: September 4, 2023 Received: September 5, 2023
Dear Ray Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe
Tanisha L. s 2023.12.01 Hithe -S 15:06:55 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Tab#15 User Manual
{4}------------------------------------------------
Intense pulsed light therapy device Model: FI-L06
User Manual
Caution: Federal law restricts this device to sale by or on the order of a physician.
Hebei Zhemai Technology Co.,Ltd
{5}------------------------------------------------
Catalogue
| Foreword | 3 |
|---|---|
| Special Note | 4 |
| Chapter 1 Symbol Description | 5 |
| Chapter 2 Clinical Application | 7 |
| Chapter 3 Technical Parameters | 10 |
| Chapter 4 Working Principle | 12 |
| Chapter 5 Installation and Commissioning | 14 |
| Chapter 6 Instructions for Use | 18 |
| Chapter 7 Maintenance | 22 |
| Chapter 8 Contact Us | 32 |
{6}------------------------------------------------
Foreword
You are welcome to use the Intense pulsed light therapy device. The Intense pulsed light therapy device produced by our company is a high-tech product independently developed by our company and has independent intellectual property rights.
This manual is applicable to the Intense pulsed light therapy device,the model is FI-L06.
After receiving the device, first check the accessories according to the packaging label. If the accessories delivered with the device are incomplete, please contact us according to the contact information in Chapter 8.
User Notice
If you are using the Intense pulsed light therapy device for the first time, be sure to read this manual carefully before powering up.
It is recommended not to use this device in the presence of electromagnetic interference, vibration and other disturbing environmental conditions. During use, this device will not interfere with other instruments/device.
Avoid using this Intense pulsed light device in the presence of flammable anesthetics or oxidizing gases such as nitrous oxide and oxygen. When there is enough oxygen in the air, some materials, such as cotton and wool, are easily ignited by the high temperatures generated by the normal operation of an Intense pulsed light device. Solvents and flammable solutions used for cleaning or sterilization should evaporate before using the laser device.
The discarded parts produced by scrapping or replacing the Intense pulsed light therapy device should be sent to the designated institution or manufacturer for professional treatment to avoid polluting the environment.
The power supply device should be grounded reliably.
All contents of this document are protected by copyright laws. Without the express written permission of the company, any organization or individual may not copy or store this manual in a database or retrieval system.
{7}------------------------------------------------
Special Note
Image /page/7/Picture/1 description: The image contains a warning sign for collimated light. The sign features a black triangle with a symbol of a laser beam in the center. Below the triangle, there is a rectangular sign with the text "COLLIMATED LIGHT LEVEL: CLASS 3 B" written in black letters on a white background.
Warning
Intense pulsed light therapy device can emit high-intensity infrared laser beams, and in order to protect the eyes, users and patients are required to wear appropriate safety goggles in accordance with (NOHD) standards.
Please note the following:
Do not point the laser directly at the eyes and skin.
When the device is working, do not cut off the power.
Be sure to familiarize yourself with the structure and specifications of the device before use.
{8}------------------------------------------------
Chapter 1 Symbol Description
Image /page/8/Figure/1 description: The image shows the label of an intense pulsed light therapy device. The label includes the model number FI-L06, power supply AC 110V, 60Hz, laser wavelength 650-1200 nm, and product mass 50kg. It also includes a warning about invisible radiation and the manufacturer's information, Hebei Zhemai Technology Co., Ltd.
| Number | Symbol | Interpretation |
|---|---|---|
| 1 | Image: Manufacturer Symbol | Manufacturer |
| 2 | Image: Manufacturing Date Symbol | Manufacturing Date |
| 3 | Image: Serial Number Symbol | Serial Number |
| 4 | Image: General Warning Sign Symbol | General warning signs |
| 5 | Image: Do not throw away electrical and electronic products symbol | Do not throw away electrical and electronic products at will,and dispose of them separately in the designated garbage area |
| 6 | Image: Type BF applied part symbol | Type BF applied partThe applied components meet the requirements of thestandard IEC 60601-1 to provide protection against electricshock, especially regarding the leakage current not greaterthan the value of human injury |
| 7 | Image: Laser Output Warning Label Symbol | Laser Output Warning Label |
| 8 | Image: No Pushing Symbol | No Pushing |
{9}------------------------------------------------
| 9 | Image: COLLIMATED LIGHT LEVEL: CLASS 3 B | Laser radiation label |
|---|---|---|
| 10 | Image: Follow the user manual | Follow the user manual |
| 11 | Image: Key switch | Key switch |
| 12 | Image: Emergency stop switch | Emergency stop switch |
| 13 | Image: CAUTION Federal law restricts this device to sale by or on the order of a physician | Medical Device Labels for Restricted Sale |
| 14 | Image: FOOT SWICH | Foot pedal |
| 15 | Image: STOP | Emergency Stop Switch Label |
| 16 | Image: Up label | Up label |
| 17 | Image: Fragile label | Fragile label |
| 18 | Image: Keep dry | Keep dry |
| 19 | Image: Don't scroll | Don't scroll |
{10}------------------------------------------------
Chapter 2 Clinical Application
2.1 Indication For Use
The Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
2.2 Clinical features
This product has novel design, stable operation, short irradiation time, lasting effect, convenient operation, good clinical effect and high safety factor.
Convenience: Human-machine interface - touch screen, intuitive and easy to use, so its operation is simple and convenient.
Permanent Hair Removal: It is suitable for hair of all skin tones and has a good hair removal effect.
Safety: stable performance, long service life, and real-time control by intelligent microprocessor.
2.3 Potential risks
The most common complications of Intense pulsed light therapy device are postoperative local erythema and follicular edema, most of which resolve spontaneously within a few hours. complications are localized crusting, purpura, blisters, hyperpigmentation Rare or hypopigmentation, and increased sebum production. Adverse reactions after hair removal are mainly related to energy density and epidermal melanin content, and are also significantly related to seasonal changes, treatment sites, and solar radiation. Adhering to preoperative and postoperative guidance and correct operation can reduce adverse reactions.
2.4 Processing method
(The following processing steps and methods are for reference only, the actual parameters and operating procedures are subject to clinical use)
2.4.1 Preoperative preparation
-
- For patients with dark skin, sun radiation should be avoided as much as possible before surgery. It is best to use sunscreen for 4 to 6 weeks. For those with a tendency to pigmentation, hydroquinone can be added to prevent it.
-
- The treated area must be skin-prepared prior to treatment, so hair, oil and dirt should be thoroughly shaved.
{11}------------------------------------------------
2.4.2 Anesthesia
Whether anesthesia is required during treatment should be determined according to the treatment site and size, whether it is a sensitive area, and the patient's pain. Generally, only topical anesthesia can be used.
2.5 Clinical considerations
-
- Generally closely related to energy density, treatment outcome and inflammatory status after skin damage. High energy density is more effective, but at the same time, the inflammatory response is greater; the deep skin is also damaged. The energy is processed starting at low energy and gradually increasing the energy density until better results are achieved.
-
- The maximum energy allowed is inversely proportional to the skin pigment. Generally, the skin is darker and has a lower energy density to reduce the absorption of more light energy into heat by the epidermis. A good cooling system is very important to the treatment, to protect the skin from burns to the greatest extent, and it is better for dark skin.
-
- Shave excess hair before surgery, as excess hair absorbs energy and heats it on the epidermis, which may cause partial skin burns and damage the probe.
-
- A good cooling system will reduce the temperature of the skin. Cold probes help to increase allowable energy density and partial anesthesia. Cooling is highly recommended throughout the treatment, especially for darker skin. Dark skin absorbs energy more easily than light skin.
2.6 Intense pulse light handle cleaning
When the treatment is complete, remove the handle, clean the top of the handle with a relatively soft cotton cloth and keep it dry.
-
- Cleaning process
After processing is complete, exit the operating program and turn off the power of the device. Soak a soft cotton cloth in 75% alcohol and wipe the top of the handle clean until the alcohol has evaporated again. Then wipe the surface marks with the soft cotton cloth.
- Cleaning process
-
- Warning message
Intense pulsed light processing precision optical components, please avoid falling and make them subject to blunt force impact or not allowed to open, otherwise the intense pulse light module will not emit light normally; This is a large impact of the intense pulse light output. The surface cleanliness of the optical crystal, therefore, before use Inspection and proper cleaning. The handle transfer system should avoid serious damage caused by bending during use.
- Warning message
2.7 Contraindications
- Tanned skin (has been exposed to the sun or used tanning beds in the past three to four
{12}------------------------------------------------
weeks) or has recently used a self-tanning lotion
- Diseases stimulated by light (Lupus, Epilepsy ... ) .
- Use of photosensitizing medications such as: tetracycline or herbal (St John's Wort) etc. .
- Use of anticoagulants or has a history of bleeding coagulopathies. .
- Fragile and dry skin. Use of Accutane (isotrenoin) Roaccutane or Tretinoin Retin A for . the treatment of acne in the previous 6-12 months.
- . Active infections: herpetic lesions, cold sores (place on anti-viral medication pre-treatment)
- Any inflammatory skin condition (e.g. eczema, etc.), in the area to be treated. .
- Pregnancy or trying to conceive. .
- . Tattoos, permanent make-up, suspicious pigmented lesions (can treat around them)
- History of keloidal scarring .
- . For Hair Reduction - Electrolysis, waxing or plucking in last 6 weeks (target has been removed)
- Planning to be in the sun following treatment .
- Diabetes (unless under control) .
- Cancer, in particular, skin cancer .
- Hormonal Disorders (unless under control) .
{13}------------------------------------------------
Chapter 3 Technical Parameters
3.1 Technical parameters
| Model | FI-L06 |
|---|---|
| Light source | Intense pulsed light |
| LCD screen | LCD touch screen 7 inches |
| Wavelength | 650nm-1200nm |
| Spot size | 3.2cm2 (8mm×40mm2) |
| Pulse length | 0.1-25ms |
| Energy density | 1-50J/cm2 |
| Pulse-frequency | 1-10Hz |
| Cooling | Air cooling+water cooling |
| Input voltage | 110V 60Hz |
| size | 89.657.794.3cm (package)794995cm (device) |
| weight | 50kg (device)65kg (package) |
3.2 Conditions of use
3.2.1 Working conditions:
- a) Ambient temperature range: 15℃~30℃;
- b) Relative humidity range: ≤80%;
- c) Atmospheric pressure range: 860hPa~1060hPa;
- d) Power supply: AC 110V, 60Hz.
3.2.2 Transportation and storage environment
- a) Do not place the product in a separate enclosure unless proper ventilation is provided;
- b) No insulation;
- c) Keep away from chemical substances and corrosive substances;
- d) Dustproof;
- e) Special protection controller (LCD screen), handle and crystal, fragile, no pressure.
- 3.2.3 Shelf Life
5 years
Note: The maximum IPL emission number (maximum IPL lens flash) of xenon lamp tube before performance degradation is 100000 times.
3.3 Electromagnetic Interference
{14}------------------------------------------------
The Intense pulsed light therapy device complies with IEC 60601-1-2 "Electromagnetic Compatibility requirements and tests". Part of IEC 60601-1-2 addresses the issue of unintended RF emissions from products. Accidental emissions from products may interfere with the performance of other nearby equipment. Emissions from Intense pulsed light therapy device have been reduced to practical levels without affecting function. If interference from an intense pulsed light therapy device is suspected, ensure that the intense pulsed light therapy device is plugged into an AC power source not shared by the affected device. If the interference persists, move the Intense pulsed light therapy device or affected equipment to another room.
{15}------------------------------------------------
Chapter 4 Working Principle
4.1 Svstem principle
Intense pulsed light is a non-laser light source, which is incoherent light composed of single or multiple pulse sequences, and has the characteristics of wide spectral range and high energy density. The light source of the Intense pulsed light therapy device is a xenon lamp. Its basic working principle is that the trigger applies a high voltage to the xenon gas to trigger the ionization of the xenon gas. Internal xenon avalanche ionization, xenon converts and releases the charged energy in the form of high-intensity light radiation, and this discharge process is a pulse.
The parameters required in the treatment process, such as energy density, etc., should be set by the user according to the actual situation.
4.2 System
The Intense pulsed light therapy device consists of six modules, namely control system, user interface, power supply, laser emission and delivery system, cooling system and safety features. As shown in Figure 4.1.
Image /page/15/Figure/6 description: The image shows a block diagram of a system. The diagram includes the following components: User Interface, Cooling System, Control System, Power Source, Safety Features, and Emission & Delivery system. The Control System is in the center of the diagram, and it is connected to all of the other components via arrows.
Figure 4.1 Block Diagram of Intense pulsed light therapy device System
4.2.1 Control system
The control system consists of a microprocessor, a control circuit and a detection circuit. The control module is highly automated. Users need to turn on the power switch, the key switch, and release the emergency switch. After the system is turned on, the system will automatically perform system self-checking, processing parameter calls and other actions. After system initialization, self-checking, and parameter calling, press the foot switch and the control module will adjust it. The pulse parameters of each release energy, and alarm when the system is abnormal.
4.2.2 User Interface
The display module is mainly composed of a touch screen and a liquid crystal display. The touch screen consists of a touch detection component and a touch screen controller. The touch detection component is installed on the front of the LCD to detect the user's touch position and
{16}------------------------------------------------
send it to the touch screen controller after receiving. The main function of the touch screen is to receive touch information, convert it into the coordinates of the touch point, and then send it to the CPU, where it can receive and execute commands from the CPU. And finally realize the man-machine operation display.
4.2.3 Power
The power module is mainly composed of photon pulse power, DC 12V switching power supply and energy storage capacity. The photon power consists of a charging part, a pre-burning part of the xenon lamp and a discharging part. The DC 12V switching power supply converts input AC to output DC, providing DC 12V power for fans and semiconductor cooling parts.
4.2.4 Intense pulsed light emission and delivery system
In the standby state, when the switch button on the handle is triggered, the xenon lamp will emit a strong pulse of light through the light spot emission window on the handle.
4.2.5 Cooling system
The cooling system consists of a fan and a water circulation system. After the device is turned on, the cooling system starts to work. The fans air-cool the electrical components of the main unit and the insulated distributor, removing heat generated by the airflow within the machine. The internal circulating water is replaced in time and takes away the heat generated by the processing handle, and is sent to the water temperature cooling system.
4.2.6 Security Features
Intense pulsed light therapy device provides enough safety features to ensure the normal operation of the system, such as key switch, emergency stop switch, internal safety interlock, protective casing and alarm system to immediately shut down the system in an emergency, press the red mushroom Shaped emergency stop switch to terminate IPL energy discharge. The red switch is used to turn off the power in an emergency. Press it and the power can now be turned off. After turning clockwise, the machine will continue to work. Or the machine keeps shutting down.
The Intense pulsed light therapy device contains a "normally closed" safety shutter that prevents IPL emission when in the closed position. The shutter will only open when the system is in ready mode and the foot switch is pressed.
The alarm system consists of buzzer alarm and symbol alarm. Detailed warnings and faults are described below.
Water Flowli Water flow alarm: there is no water or no water circulation inside, the indicator light is displayed, and the alarm sound follows. In response, the machine will immediately stop all work.
Water Temperature alarm: when the temperature is higher than the system setting: 55 ℃, the indicator light shows this, and an alarm sound is issued at the same time. In response, the machine will immediately stop all work.
{17}------------------------------------------------
Chapter 5 Installation and Commissioning
Installation and commissioning of device must be done by qualified professionals and technicians trained by the manufacturer.
5.1 Model FI-L06
5.1.1 Accessories
Image /page/17/Picture/4 description: The image shows a medical device with several labeled components. The device includes a treatment handle and a handle interface, suggesting it's used for some form of therapy or procedure. A screen is present, likely for displaying settings or providing feedback during use. The device also features a key switch and an emergency stop switch, indicating safety and control measures. Finally, it has movable castors, making it easy to transport.
Image /page/17/Picture/5 description: The image shows the back of a medical device with labels pointing to different parts. The labels indicate the location of the foot switch socket, fuse holder, and power cord. The power cord is plugged into the device, and the foot switch socket and fuse holder are located above it. The device is on wheels.
{18}------------------------------------------------
5.1.2 Accessories of the device
Image /page/18/Picture/1 description: The image shows four different medical devices and accessories. The first image shows a white treatment handle with a cable. The second image shows patient goggles and a clear plastic case. The third image shows doctor's goggles and a black case. The fourth image shows an accessory case with an orange bag and a black cable inside.
Image /page/18/Picture/2 description: The image shows four different objects: a foot switch, a power cord, a key, and two fuses. The foot switch is a gray box with a yellow triangle symbol on it. The power cord is black and coiled. The key is silver and has a round head. The fuses are clear glass tubes with metal caps on each end.
5.1.3 Water inlet procedure
Open the rear of the device as shown below
Image /page/18/Picture/5 description: In the image, a person is shown holding a clear plastic tool with a red wire attached to it. The person is using the tool on a white machine with a gray frame. The machine has a series of small holes on its surface. There are also two black cables plugged into the bottom of the machine.
Holding the buckle hand, unscrew the fixing screws one by one
Holding the bottle cap by hand, unscrew the water tank
Image /page/18/Picture/8 description: In the image, a person is shown placing a gray bottle into a gray machine. The bottle has a white cap. The person is using both hands to hold the bottle. The machine has a small opening in the front.
Image /page/18/Picture/9 description: In the image, a person is using a pair of blue scissors to cut a clear tube. The tube is being held by the person's other hand. The cutting is taking place inside of a gray machine. The machine has a white and black component at the top.
Before adding water, please cut the front end of the water pipe by 2 cm
Fill the tank with 2500ml of distilled water
Image /page/18/Picture/12 description: In the image, a person is holding a white plastic bottle and inserting it into a machine. The machine appears to be a medical device or equipment, with a gray frame and a white top. The person's hand is visible, holding the bottle steady as it is being inserted. The bottle is cylindrical in shape and has a white cap. The background is blurred, but it appears to be a medical setting.
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows two different steps for maintaining a machine. The first step shows a person putting a bottle into a compartment. The second step shows a person tightening the screws on the back of the machine. The text in the image provides instructions for the maintenance, including tightening the screws and refilling the water to 2500ml.
- 5.1.4 Connect the pedals as shown below.
Image /page/19/Picture/2 description: In the image, a hand is shown holding a gray cable with a black connector. The connector is plugged into a port on a gray and white machine. The machine has a barcode sticker on the side. The hand appears to be plugging in or unplugging the cable from the machine.
Image /page/19/Picture/3 description: The image shows a close-up of the back of a piece of electronic equipment. A gray cable is plugged into a connector on the back of the equipment. The equipment is light gray and white in color. A silver bolt is visible on the top left of the equipment.
- 5.1.5 After the above operations are completed, please connect offline.
Method: The water contact CPC on the treatment handle plug points to the water CPC socket. The method is the same as connecting the power system. When you hear the sound "pa" twice, it means the connection is successful. As follows.
Image /page/19/Picture/6 description: In the image, a hand is holding a gray and white device that is attached to a white wall. The device has a blue button on top. The device is being held in a way that suggests it is being used or operated. The wall appears to be part of a larger structure, possibly a piece of equipment or a building interior.
Image /page/19/Picture/7 description: The image shows a close-up of a medical device, possibly a handheld ultrasound transducer. The device has a gray body with a blue button on the end. The other end of the device is white and appears to connect to a cable. The device is resting on a white surface.
{20}------------------------------------------------
5.1.6 After inserting the therapy handle, turn on the power, then turn the power emergency switch and the key switch to ensure the machine is powered on.
5.1.7 Change the water every 2 or 3 months, here's how to drain it.
Image /page/20/Picture/2 description: The image shows a medical device with instructions on how to remove the fixing screws and take out the water tank. The instructions indicate that the fixing screws should be removed one by one. The instructions also indicate that the water tank should be drained and replaced with new distilled water. The image is likely a guide for maintenance or repair of the medical device.
5.1.8 Handle connection
Image /page/20/Picture/4 description: The image shows a person holding a gray and white device. The device has a gray top and a white bottom. The person's hand is visible at the bottom of the image, holding the device. The background is a light-colored surface.
Image /page/20/Picture/5 description: The image shows a close-up of a gray and white device with a blue button on top. The device is attached to a machine, and a gray tube is connected to the bottom of the device. The device appears to be a medical or cosmetic tool, possibly used for treatments or procedures. The background is blurred, but it seems to be a medical or clinical setting.
Insert the handle socket until a "click" sound means a good handle connection.
{21}------------------------------------------------
Chapter 6 Instructions for Use
6.1 General Instructions
The Intense pulsed light therapy device is an intense pulsed light system that provides intense pulsed light in the wavelength range of 650nm-1200nm. Intense pulsed light (IPL) therapy device employ the principle of selective thermal decomposition. That is, thermal damage to the target chromophore is achieved by using light of an appropriate wavelength in a pulse that exceeds the thermal relaxation time of the chromophore, but by limiting the pulse width below the thermal relaxation time of the skin. Thermal damage to the target chromophore is caused by using light of an appropriate wavelength, but preserving normal skin by limiting the pulse width to a pulse width below the skin's thermal relaxation time.
Intense pulsed light differ from lasers in that they provide multiple wavelengths in each light pulse, rather than just one. In general, IPL enhances penetration without using excessive energy levels and is able to target specific chromophores. Based on this, the Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.
Image /page/21/Picture/4 description: The image is a black and white warning sign. The sign is a triangle with a thick black border. Inside the triangle is a symbol of a laser beam, which is represented by a circle with many lines radiating outward. The sign is used to warn people of the potential danger of lasers.
WARNING: HAZARDOUS EXPOSURE IF THE CONTROLS AND REGULATIONS ARE NOT USED OR PERFORMED IN THE SPECIFIC METHODS OF THE LASER SYSTEM BEFORE USE OR OPERATION.
6.2 Start the program
Once filled with water, proceed as follows :
-
- Check that the handle adapter is plugged into the treatment head of the device. (Note: The handle must be connected to the specified position above, otherwise the device will operate incorrectly.)
-
- Check that the foot switch is also connected to the field marked "foot switch".
-
- After the above steps 1 and 2 are completed, insert the three-core power plug into an AC110V socket that can withstand the corresponding current.
-
- Turn on the rocker switch on the back of the chassis, rotate the key clockwise from "off" 90 degrees to "on", then turn the emergency stop switch to the right to enter the release state, at this time the hair removal instrument is turned on, and the touch screen displays the startup image.
-
- After waiting for a few minutes, the Intense pulsed light therapy device will automatically complete the preheating and enter the standby mode screen. (If there is a problem with the
{22}------------------------------------------------
cooling system, the temperature cannot be controlled to increase to the specified temperature, it is regarded as a device failure, the device cannot enter standby mode, and always remains in the initial startup image).
6.3 System Control
- 6.3.1 Initial interface and main interface
Image /page/22/Picture/3 description: The image shows a mobile app interface with a dark blue gradient background. On the left side, there are three icons: a back arrow, a home icon, and a settings gear. In the center of the screen, the logo "HEONE" is displayed in white. Below the logo, there is a graphic of interconnected lines and dots, resembling a circuit board pattern.
Figure 1 Startup interface, enter Figure 2 after 2 seconds
Image /page/22/Figure/5 description: The image shows a screen with the HEONE logo at the top, followed by the text "Matters Needing Attention. Operate safely in the specified environment. Please maintain the equipment regularly." Below this text are five circular buttons, each labeled with a different language: English, Español, Français, русский язык, and بالعربية. The background of the screen is a gradient of dark blue to light blue, with a circuit board pattern at the bottom.
Figure 2 Enter the language options: select English and enter Figure 3
{23}------------------------------------------------
Image /page/23/Figure/0 description: The image shows a screen display with the HEONE logo at the top. Below the logo, the text "Matters Needing Attention" is displayed, followed by instructions to operate safely and maintain the equipment regularly. The screen also features several icons labeled with terms such as "INTERLOCK," "WATER FLOW SENSOR," "REMOTE CONTROL," and "WATER FLOW."
Figure 3 Enter the automatic protection detection of the instrument, enter Figure 4 after 10 seconds
Image /page/23/Figure/2 description: The image shows a graphical user interface (GUI) of a device, possibly a laser or IPL machine, with a dark background and teal-colored interactive elements. The screen displays information such as "Energy" set at "01 J/CM²", "Pattern" set to "HR", and "Refrigeration" at level "3". It also shows "Temperature" at "24" and other parameters like "One operation" and "Cumulative count" with values of "00000000".
Enter the function options, select the treatment item and enter Figure 4
Figure 4 Work interface, parameter adjustment settings
{24}------------------------------------------------
Image /page/24/Figure/0 description: The image shows seven numbered instructions for operating a device. The first two instructions describe how to adjust the energy settings using the up and down arrows. The third instruction explains how to increase or reduce the cooling intensity. The remaining instructions describe the function of the start/standby button, the temporary counter, the water temperature detection, and the total luminous count.
6.3.2 Parameter adjustment settings
The appropriate therapeutic parameters were selected according to the required treatment mode and patient status.
{25}------------------------------------------------
Chapter 7 Maintenance
7.1 System maintenance
System repairs should be performed by a Zheone certified service engineer. For training and information, please contact your local Zheone representative.
7.2 Cleaning and Disinfection
7.2.1 Cleaning external surfaces
The outside of the IPL can be cleaned with a soft cloth moist with mild soapy water solution, do not use harsh detergent. To disinfect the exterior of the laser system, wipe it with a soft cloth stained with 75% medical alcohol. Cleaning frequency should be once a week, in frequent use can be appropriate to increase the cleaning frequency.
Warn Do not attempt to access any internal components. Electrical shock and/or accidental laser exposure may result. Do not spray or pour 75% medical alcohol directly on the system console.
7.2.2 Cleaning the system touch screen display clean the system touch screen display with a damp soft cloth
Warn
Do not spray or pour 75% medical alcohol directly onto the system console or control screen, or you may damage the console, touch screen, and system electronics.
7.2.3 Cleaning the handle
7.2.3.1 Cleaning the protective windows
The protective window is an optical device. Essentially, with use, it breaks down, and once broken, the protective window must be replaced. However, inspection and cleaning after each procedure will extend its life.
Warn
Hair from the patient may collect on the windows. During treatment, windows may need to be washed regularly. This window must be cleaned using the techniques described below.
{26}------------------------------------------------
Cautious
If the window is not properly maintained, debris can build up on the surface of the window, causing permanent damage to the window and ultimately to the phone optics.
Warn
Before checking the window, the system must be shut down. Always wear the correct glasses when the system is on. Serious eye injury can occur in the event of an accidental laser emission.
Notice
Stick and cotton applicator are needed, but do not use products that contain adhesives.
- λ Make sure the system is turned off.
- 2 If necessary, place a small amount of 75% medical alcohol on a cotton swab; Shake excess alcohol off the cotton swab before cleaning the protective window.
- λ Gently wipe the surface of the window.
- ^ Check the window. If necessary, clean the window again with a new cotton-tipped applicator.
Notice
Windows must be kept clean before treating patients. Chips, pits, cracks, or burns in Windows are not acceptable. Windows with these abnormalities must be replaced. If the exception on the window cannot be resolved, you must replace the window.
7.2.3.2 Tips for cleaning your phone
Sterilize patients who use 75% medical alcohol. Windows should be cleaned as described in "Cleaning Protective Windows".
Warn
Before disinfecting the handle tips, make sure the system is turned off or in standby mode. Always wear appropriate protective glasses when the system is turned on. If accidental laser emission occurs, it can cause serious damage to the eyes.
7.2.4 Handling of handle components
For infection control reasons, disinfecting the handle and then disposed of according to local agency policies and procedures.
{27}------------------------------------------------
7.3 Calibration procedure
If self-calibration fails, Zheone authorized personnel should perform the following calibration. Questions regarding this procedure should be directed to your local Zheone representative.
Disclaimer Warning
Calibration is a service procedure and can only be performed by a Zheone-certified service engineer or a customer who has attended and passed a Zheone-certified service training course. Adjustments by anyone other than a trained Zheone service engineer or certified customer will void any existing manufacturer's warranty on the instrument.
A service manual for this system can be purchased from the Zheone Microservice Department, however, possession of service instructions or service tools does not authorize the repair or modification of the Zheone Microsystem by non-certified personnel.
7.3.1 Equipment
-
- Laser safety glasses for everyone in the room (produces the appropriate optimum density at the wavelength).
- ^ Lambda laser energy meter, the meter used must have received a NIST traceable calibration or a calibration to an applicable standard (international) within the past 12 months (in the U.S.).
7.3.2 Calibration Instructions
- 2 Places the laser energy meter in a convenient location so that the sensor head can be easily accessed with a mobile phone. Set the meter display unit to power mode.
-
- Remove the handle tip and place the handle laser hole in front of the energy meter sensor head.
- ^ Provides power for the system.
- 2 Select the low flux range, small spot size, and set the repetition rate to 1 Hz.
- λ Puts the system in ready mode.
- λ Press the foot switch and fire a shot in the energy meter.
- λ Record the corresponding reading on the energy meter.
- 2 Repeat steps 4 to 7 for the handle selectable intermediate and higher flux settings, and set the repetition rate to 4 Hz.
-
- Zheone authorized personnel will compare the recorded data to a standard energy meter and then take further steps to complete the calibration.
Notice
{28}------------------------------------------------
The distance from the energy meter sensor to the laser aperture should not exceed 10mm and should remain constant. The window surface must be parallel to the meter sensor surface.
Cautious
Work by any unauthorized person will void all warranties.
7.4 Check or replace the protection tube of the power supply
Turn off all power switches, unplug the power cord, open the fuse protector with a small screwdriver, and turn it counterclockwise. Then remove the fuse protector, replace the fuse protector only with the standard model specified by the device supplier. (Model of fuse protector: 10A/250VAC) Put in the fuse protector and jacket, and tighten the jacket clockwise. Plug the back of the device into the power cord, turn on the power switch, and test if the machine works properly.
7.5 Maintenance of cooling system
Please check regularly whether the cooling fan is working properly. If the cooling fan or cooling module does not work, the device cannot exchange heat with the outside, and the heat will accumulate inside the machine, which will cause the laser to not work properly or even damage the machine.
When the maintenance personnel test the machine, they should check the water storage capacity, water quality and cooling capacity of the water tank. If there is any abnormality, please add or change the water in time.
{29}------------------------------------------------
7.6 EMC
Table 1: Guidance and manufacturer's declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS
| Guidance and manufacturer's declaration - electromagneticemissions | ||
|---|---|---|
| The Intense pulsed light therapy device is intended for use in the electromagnetic environmentspecified below. The customer or user of the equipment should ensure that it is used in such an | ||
| environment. | ||
| Emissions Test | Compliance | Electromagnetic Environment -Guidance |
| RF emissionsCISPR 11 | Group 1 | The Intense pulsed light therapy deviceuses RF energy only for its internalfunction. Therefore its RF emissions arevery low and are not likely to cause anyinterferenceinnearbyelectronicequipment |
| RF emissionsCISPR 11 | Class A | The Intense pulsed light therapy device issuitable for use in all establishments, other |
| Harmonic emissionsIEC 61000-3-2 | Not applicable | than domestic establishments and thosedirectlyconnectedtothepubliclow-voltage power supply network thatsupplies buildings used for domesticpurposes |
| Voltage fluctuations /flicker emissionsIEC 61000-3-3 | Not applicable |
{30}------------------------------------------------
Table 2: Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
| ImmunityTest | IEC 60601 TestLevel | ComplianceLevel | ElectromagneticEnvironment - Guidance |
|---|---|---|---|
| Electrostaticdischarge (ESD)IEC 61000-4-2 | $\pm$ 8 kV contact$\pm$ 2 kV, $\pm$ 4 kV, $\pm$ 8kV, $\pm$ 15 kV air | $\pm$ 8 kV contact$\pm$ 2 kV, $\pm$ 4 kV, $\pm$ 8kV, $\pm$ 15 kV air | Floor should be wood,concrete or ceramic tile. Iffloors are covered withsynthetic material, the relativehumidity should be at least30% |
| Electrostatictransient / burstIEC 61000-4-4 | $\pm$ 2 kV forpower supply lines100 kHz repetitionfrequency$\pm$ 1 kV forinput / outputlines | $\pm$ 2 kV forpower supplylines100 kHz repetitionfrequency | Mains power quality should bethat of a typical commercial orhospital environment |
| SurgeIEC 61000-4-5 | $\pm$ 0.5kV, $\pm$ 1kVdifferential modeline-line$\pm$ 0.5kV, $\pm$ 1kV,$\pm$ 2kV differentialmode line-ground | $\pm$ 0.5kV, $\pm$ 1kVdifferential modeline-line$\pm$ 0.5kV, $\pm$ 1kV,$\pm$ 2kV differentialmode line-ground | Mains power quality should bethat of a typical commercial orhospital environment |
| Voltage dips,shortinterruptionsand voltagevariations onpower supplyinput linesIEC 61000-4-11 | 0 % UT(100 % dip in UT)for 0.5 cycle at 0°,45°, 90°,135°,180°, 225°, 270°,and 315°0 % UT(100 % dip in UT)for 1 cycle at 0°70 % UT(30 % dip in UT)for 25/30 cyclesat0°0 % UT(100 % dip in UT)for 250/300 cycle | 0 % UT(100 % dip in UT)for 0.5 cycle at 0°,45°, 90°,135°,180°, 225°, 270°,and 315°0 % UT(100 % dip in UT)for 1 cycle at 0°70 % UT(30 % dip in UT)for 25/30 cyclesat0°0 % UT(100 % dip in UT)for 250/300 cycle | Mains power quality should bethat of a typical commercial orhospital environment. If theuser of the equipment requirescontinued operation duringpower mains interruptions, it isrecommended that theequipment be powered froman uninterruptible powersupply or a battery |
| Powerfrequency (50 /60 Hz)magnetic fieldIEC 61000-4-8 | 30 A/m, 50/60Hz | 30 A/m, 50/60Hz | Power frequency magneticfields should be at levelscharacteristic of a typicallocation in a typicalcommercial or hospitalenvironment |
| Note: UT is the a.c. mains voltage prior to application of the test level. |
{31}------------------------------------------------
| Guidance and manufacturer's declaration - electromagneticimmunity | |||
|---|---|---|---|
| The Intense pulsed light therapy device is intended for use in the electromagnetic environment specified below. The customer or the user of the Intense pulsed light therapy device should assure that it is used in such an environment. | |||
| Immunity Test | IEC 60601Test Level | ComplianceLevel | ElectromagneticGuidance |
| Conducted RFIEC 61000-4-6 | 3Vrms150kHz to80MHz6 Vrms150 kHz to80 MHzoutside ISMbandsa | 3Vrms150kHz to80MHz6 Vrms150 kHz to80 MHzoutside ISMbandsa | Portable and mobile RF communication equipment should be used no closer to any part of the Intense pulsed light therapy device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended separation distance$d = \left[\frac{3.5}{V_1}\right] \sqrt{P}$ $d = \left[\frac{3.5}{E_1}\right] \sqrt{P}$ 80MHz to 800MHz $d = \left[\frac{7}{E_1}\right] \sqrt{P}$ 800MHz to 2.7GHz |
| Radiated RFIEC 61000-4-3 | 3V/m80MHz to2.7GHz | 3V/m | Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b |
| Interference may occur in the vicinity of equipment marked with the following symbol: |
Table 3: Guidance and MANUFACTURER'S declaration – electromagnetic IMMUNITY
{32}------------------------------------------------
a The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 - MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
C Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the IPL is used exceeds the applicable RF compliance level above, the IPL should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Intense pulsed light therapy device.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
{33}------------------------------------------------
Table 4: Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM
| Recommended | separation | distances between | portable and |
|---|---|---|---|
| mobile RF communications equipment and the Intense pulsed light therapy | |||
| The Intense pulsed light therapy device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the IPL can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Intense pulsed light therapy device as recommended below, according to the maximum output power of the communications equipment. | |||
| Ratedmaximumoutput oftransmitterW | 150 kHz ~ 80 MHz | 80 MHz ~ 800 MHz | 800 MHz ~ 2.5 GHz |
| $d = \left[ \frac{3.5}{V_1} \right] \sqrt{P}$ | $d = \left[ \frac{3.5}{E_1} \right] \sqrt{P}$ | $d = \left[ \frac{7}{E_1} \right] \sqrt{P}$ | |
| 0.01 | 0.12 | 0.12 | 0.23 |
| 0.1 | 0.38 | 0.38 | 0.73 |
| 1 | 1.2 | 1.2 | 2.3 |
| 10 | 3.8 | 3.8 | 7.3 |
| 100 | 12 | 12 | 23 |
| For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. | |||
| Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applied. |
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
{34}------------------------------------------------
Table 5: Recommended Specifications for Enclosure Port Immunity to RF Wireless Communications Equipment
| Recommended separation distances between RF wireless communications equipment | |||||
|---|---|---|---|---|---|
| The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment. | |||||
| FrequencyMHz | MaximumPowerW | Distance | IEC60601Test Level | ComplianceLevel | ElectromagneticEnvironment -Guidance |
| 385 | 1.8 | 0.3 | 27 | 27 | RF wireless communications |
| 450 | 2 | 0.3 | 28 | 28 | communications |
| 710 | equipment should be | ||||
| 745 | 0.2 | 0.3 | 9 | 9 | used no closer to any |
| 780 | part of the device, | ||||
| 810 | including cables, than | ||||
| 870 | 2 | 0.3 | 28 | 28 | the recommended |
| 930 | separation distance | ||||
| 1720 | calculated from the | ||||
| 1845 | 2 | 0.3 | 28 | 28 | equation applicable to |
| 1970 | the frequency of the | ||||
| 2450 | 2 | 0.3 | 28 | 28 | |
| 5240 | |||||
| 5500 | |||||
| 5785 | 0.2 | 0.3 | 9 | 9 | transmitter.Recommended separation distance$E = 6 / d * sqrt{P}$Where P is the maximumoutput power rating ofthe ransmitter inwatts(W) according to thetransmittermanufacturer and d isthe recommendedseparation distance inmeters (m).Fieldstrengths from fixed RFtransmitter,as determinedby anelectromagnetic sitesurvey, should be lessthan the compliancelevel in each frequencyrange. Interference may equipment marked withthe following symbol:((•)) |
| Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. |
{35}------------------------------------------------
Chapter 8 Contact Us
Hebei Zhemai Technology Co., Ltd.
Tel: 0086-10-18678802036
After-sales service department
Email: feijiakeji@zoho.com.cn
Image /page/35/Picture/5 description: The image shows the logo for Internet Explorer. The logo is a blue lowercase "e" with a gold ring encircling it. The "e" is stylized with a curved shape, and the gold ring appears to orbit around it.
Home: http://www.ftelaser.cn/
A
Address: 1st Floor, North of No.1 Warehouse, West of Shuichang Road, North of Beiyi
Road, Yanjiao Development Zone, Sanhe, Langfang, Hebei, China, 065201
{36}------------------------------------------------
510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K232708
-
- Date of Preparation: 11/09/2023
-
- Sponsor
Hebei Zhemai Technology Co.,Ltd
1st Floor, North of No.1 Warehouse, West of Shuichang Road, North of Beiyi Road, Yanjiao Development Zone, Sanhe, Langfang, Hebei, China, 065201 Contact Person: Yuting Qu Position: Sales Manager Tel: +86-18678802036 Email: feijiakeji@zoho.com.cn
-
- Submission Correspondent
Contact Person: Ray Wang Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
- Submission Correspondent
-
- Proposed Device Identification
Trade Name: Intense pulsed light therapy device Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Model(s): FI-L06 Regulatory Information: Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Classification: II Product Code: ONF Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Indication For Use Statement:
- Proposed Device Identification
{37}------------------------------------------------
The Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
న్. Device Description
The Intense pulsed light therapy device is an intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 650-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.
IPL Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores.
Based on this, the Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.
6. Predicate Device Identification
Predicate Device:
510(k) Number: K192519 Product Name: Multi-modality workstation Model: AAD Dual Light Manufacturer: Beijing Superlaser Technology Co., Ltd.
7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- A AAMI/ANSI ES 60601-1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- A IEC 60601-2-57: 2011 Medical Electrical Equipment - Part 2-57: Particular requirements for
{38}------------------------------------------------
the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
-
ISO 10993-5: 2009 Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity
-
ISO 10993-10: 2010 Biological evaluation of medical devices - Part 10: irritation and skin sensitization
-
ISO 14971: 2019 Medical devices - Application of risk management to medical devices
-
- Clinical Test Conclusion
No clinical study is included in this submission.
9. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device (K192519) | Remark |
|---|---|---|---|
| Product Code | ONF | ONF | SAME |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Class | II | II | SAME |
| Indication for Use | The Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen. | The Multi-modality workstation (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen. | SAME |
Table 6-1 General Comparison
Table 6-2 Performance Comparison
| ITEM | Proposed Device | Predicate Device (K192519) | Remark |
|---|---|---|---|
| Light source | Intense pulsed light(Xenon Lamp) | Intense pulsed light | Different 1 |
| Wavelength | 650-1200nm | 520-650nm | SAME |
| 800-1200nm | |||
| 540-800nm | |||
| 640-1200nm | |||
| Deliver system | Sapphire | Sapphire | SAME |
| Energy density | 1-50J/cm² | 1-50 J/cm² | SAME |
{39}------------------------------------------------
| Pulse Width | 0.1-25ms | 1-25ms | SAME |
|---|---|---|---|
| Max. Power | 1500W | 3500VA | Different 2 |
| Spot size | 8×40mm | Small: 40×12 mmLarge: 46×16mmEx-Large: 60×20mm | Different 3 |
Table 6-3 Hair removal Comparison
| ITEM | Proposed Device | Predicate Device (K192519) | Remark |
|---|---|---|---|
| Wavelength Range (nm) | 650-1200 | 640-1200 | SAME |
| Energy Range (J/cm²) | 1-50 | 5-40 | Different 4 |
| Pulse Width (ms) | 0.1-25 | 1-25 | SAME |
| Spot Size (mm) | 8×40 | Small: 40×12 mmLarge: 46×16mmEx-Large: 60×20mm | Different 3 |
Analysis 1:
Xenon lamp is a light source output of Intense pulsed light, and the system controls the energy emitted by the xenon lamp. Intense pulsed light is emitted by the xenon lamp. As far as we know, the predicate device also uses xenon lamps to generate Intense pulsed light, so in terms of light source type, we are the same as the predicate device, and there is no difference between the two.
Analysis 2:
The proposed device is different in Max. Power from the predicate device. By complying with AAMI/ANSI/ES 60601-1 and IEC 60601-1-2, the results shown that the Max. Power of the proposed device will not raise any issues in safety and effectiveness. Therefore, this difference will not affect safety and effectiveness of the proposed device.
Analysis 3:
The proposed device is different in Spot Size from the predicate device, Spot size only affects the area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness.
Analysis 4:
The proposed device only has slight difference in Energy Range with the predicate device. The Energy Range of the proposed device is within the allowable error range of the predicate device, which can justify that the difference in the parameter of Energy Range will not raise new safety issues of the proposed device. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device. So the slight difference is considered to have no effect on effectiveness and safety.
{40}------------------------------------------------
10. Substantially Equivalent (SE) Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device (K192519).
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.