K Number
K232645
Manufacturer
Date Cleared
2024-02-16

(170 days)

Product Code
Regulation Number
880.6500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Airgle Room Air Purifier is a free standing, air purifying device utilizing germicidal ultraviolet light (UV-C wavelengths near 254nm) intended for inactivation of indoor airborne aerosols including bacteria and virus in medical facilities and occupied spaces.

The Airgle® Air Purifier has been demonstrated to destroy MS2 bacteriophage bioaerosol entrained on the filter of the subject device in the following exposure/working conditions:

Test Item(s)Test Results (Average Maximum Log Reduction / Entrainment Time at Fan Speed 5)
MS2 BacteriophageModel AG300: >4.0 Logs / 120 minutes
MS2 BacteriophageModel AG600: >4.0 Logs / 120 minutes
MS2 BacteriophageModel AG900: >4.0 Logs / 120 minutes
Device Description

Airgle Room Air Purificr is an air purification device that employs an ultraviolet air purification technology to inactivate viral aerosols. It is a free-standing device, which has a sealed main housing including a Carbon filter, a cHEPA filter and Titanium Pro (UV-C) module. Model AG300 use a 2-in-1 filter which combine the Carbon filter and cHEPA filter.

The operation of the device is as follows:

  1. Room air is drawn into the device via the rear grill.
  2. The process begins when air moves through the Carbon filter, which captures the large particles
  3. Next, the air moves through the cHEPA filter, which captures a particle size greater than 0.3 microns. The filter is an ultra-high performance cHEPA filter, which filters respirable particulate matter with an efficiency reaching 99.3% at least.
  4. Then the air moves through the Titanium Pro® module, where viral organisms are inactivated by UV-C light.
  5. Purified air is then returned to the room via the front grill.
AI/ML Overview

The Airgle Room Air Purifier (Models AG300, AG600, AG900) is a medical ultraviolet air purifier designed to inactivate indoor airborne aerosols, including bacteria and viruses, in medical facilities and occupied spaces. The device's performance was assessed through non-clinical testing to demonstrate its ability to destroy MS2 bacteriophage bioaerosols.

Here's an overview of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Name of Test MethodologyPurposeAcceptance CriteriaReported Device Performance
IEC 60601-1, IEC 60601-1-2Demonstration of basic safety and essential performanceMeets criteria for CISPR Group 1APASS
AAMI TIR69:2017, ANSI C63.27-2017Wireless coexistence testWireless System ability to coexist with interfering networks in the 2.4G band.PASS
Bacteriophage TestDemonstration of log reduction of viral loadAt least 99.99% reduction of MS2 bacteriophage after 2 hours (>4.0 Log Reduction)PASS: Model AG300: >4.0 Logs / 120 minutesModel AG600: >4.0 Logs / 120 minutesModel AG900: >4.0 Logs / 120 minutes
ANSI/ASHRAE Standard 52.2Demonstration of filter efficiencyAt least 99.3% for 0.3-10 umPASS
Software validationDemonstration of software security and good designMeets design criteriaPASS

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical laboratory testing. The sample size for the "Bacteriophage Test" is not explicitly stated as a number of devices or runs within the provided text. However, the results indicate testing was performed for each model (AG300, AG600, AG900).

The data provenance is not explicitly stated in terms of country of origin, but it is implied to be from laboratory testing conducted as part of the device's clearance process. The studies were retrospective in the sense that they were designed to evaluate the finalized product against pre-defined criteria.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes non-clinical performance testing (e.g., viral inactivation, filter efficiency, electrical safety), not a study involving human experts establishing ground truth from medical images or clinical observations. Therefore, this section is not applicable in the context of this device's non-clinical evaluation. The "ground truth" here is established by the well-defined scientific and engineering standards and methodologies for air purification testing.

4. Adjudication Method for the Test Set

Not applicable, as this was non-clinical laboratory testing rather than a study requiring expert adjudication of data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This device is an air purifier, and its efficacy is measured by its ability to inactivate pathogens and filter air, not by its impact on human reader performance in interpreting medical data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The performance data presented is inherently "standalone" in this context. The device (Airgle Room Air Purifier) is the "algorithm" or system being evaluated for its direct efficacy in air purification, without human intervention in its primary function of germicidal UV light and filtration. The reported log reduction of viral load and filter efficiency are direct measurements of the device's standalone performance.

7. The Type of Ground Truth Used

The ground truth for the performance claims (e.g., viral inactivation) was established through laboratory testing using scientifically recognized and standardized methods.

  • MS2 Bacteriophage: A widely accepted surrogate virus for airborne viral studies due to its non-pathogenic nature and environmental robustness.
  • ANSI/ASHRAE Standard 52.2: A standard method for testing general ventilation air-cleaning devices for removal efficiency by particle size.
  • IEC 60601-1, IEC 60601-1-2, AAMI TIR69:2017, ANSI C63.27-2017: Industry standards for electrical safety, electromagnetic compatibility, and wireless coexistence.

8. The Sample Size for the Training Set

Not applicable. This document describes the validation (testing) of a physical device. There is no mention of a "training set" in the context of an AI/ML algorithm being developed. The device's design and engineering would be based on established scientific principles and materials, not a data "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of this device's validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 16, 2024

Airgle Corporation Jet Li Regulation manager 1170 Lincoln Avenue Unit 2 Holbrook, New York 11741

Re: K232645

Trade/Device Name: Airgle Room Air Purifier Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: January 16, 2024 Received: January 16, 2024

Dear Jet Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Christopher K. Dugard -S

Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical

{2}------------------------------------------------

and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232645

Device Name

Airgle Room Air Purifier

Indications for Use (Describe)

The Airgle Room Air Purifier is a free standing, air purifying device utilizing germicidal ultraviolet light (UV-C wavelengths near 254nm) intended for inactivation of indoor airborne aerosols including bacteria and virus in medical facilities and occupied spaces. The Airgle® Air Purifier has been demonstrated to destroy MS2 bacteriophage bioaerosol entrained on the filter of the subject device in the following exposure/working conditions:

Test Item(s)Test Results (Average Maximum Log Reduction / Entrainment Time at Fan Speed 5)
MS2 BacteriophageModel AG300: >4.0 Logs / 120 minutes
MS2 BacteriophageModel AG600: >4.0 Logs / 120 minutes
MS2 BacteriophageModel AG900: >4.0 Logs / 120 minutes

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED

{4}------------------------------------------------

{5}------------------------------------------------

K232645

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a Traditional 510(K) submission, and there were no prior submissions for the subject device.

Date of the summary prepared: February 15, 2024

1. Submitter's Information

Sponsor

  • ♦ Applicant: Airgle Corporation
  • ♦ Address: 1170 Lincoln Avenue Unit 2 Holbrook, NY 11741 USA
  • Manufacturer: Shenzhen Airgle Corporation Limited �
    • ♦ Factory address: Zone A&B, The 11th Floor, KeChuang Building, QuanZhi Science &

Technology Innovation Park, Houting Mao zhoushan Industrial Park, ShaJing Street,

Baoan District, Shenzhen Guangdong, CN 518000

  • ◆ Phone: +86-755-86726696
  • ♦ Email: jerry@airgle.com.cn
  • ◆ Contact Person (including title): Jerry Liu (CTO)

Application Correspondent:

  • ◆ . Airgle Corporation
  • ♦ Address: 1170 Lincoln Avenue Unit 2 Holbrook, NY 11741 USA
  • ♦ Contact Person: Mr. Jet Li
  • ♦ Title: Regulation Manager
  • ♦ Tel: +86-18588874857
  • ♦ Email: jianda-lee@foxmail.com

{6}------------------------------------------------

2. Subject Device Information

  • Type of 510(k) submission: Traditional �
  • � Classification: Medical Ultraviolet Air Purifier
  • � Trade Name: Airgle Room Air Purifier
  • � Model: AG300 /AG600 /AG900
  • � Review Panel: General Hospital
  • � Product Code: FRA
  • � Regulation Number: 21 CFR 880.6500
  • � Regulation Class: 2

3. Predicate / Reference Device Information

Predicate DeviceReference Device
510(k) numberK202339K212824
SponsorMolekule, Inc.Bluezone Products, Inc.
ClassificationMedical Ultraviolet Air PurifierMedical Ultraviolet Air Purifier
Trade NameMolekule Air MiniBluezone Model RX-450 Air Purifier
Review PanelGeneral HospitalGeneral Hospital
Product CodeFRAFRA
Regulation Number21 CFR 880.650021 CFR 880.6500
Regulation Class22

4. Device Description

Airgle Room Air Purificr is an air purification device that employs an ultraviolet air purification technology to inactivate viral aerosols. It is a free-standing device, which has a sealed main housing including a Carbon filter, a cHEPA filter and Titanium Pro (UV-C) module. Model AG300 use a 2-in-1 filter which combine the Carbon filter and cHEPA filter.

The operation of the device is as follows:

    1. Room air is drawn into the device via the rear grill.

{7}------------------------------------------------

  • The process begins when air moves through the Carbon filter, which captures the 2) large particles
    1. Next, the air moves through the cHEPA filter, which captures a particle size greater than 0.3 microns. The filter is an ultra-high performance cHEPA filter, which filters respirable particulate matter with an efficiency reaching 99.3% at least.
  • Then the air moves through the Titanium Pro® module, where viral organisms are 4) inactivated by UV-C light.
  • Purified air is then returned to the room via the front grill. 5)

5. Intended Use / Indications for Use (For models: AG300/AG600/AG900)

The Airgle Room Air Purifier is a free standing, air purifying device utilizing germicidal ultraviolet light (UV-C wavelengths near 254nm) intended for inactivation of indoor airborne aerosols including bacteria and virus in medical facilities and occupied spaces.

The Airgle® Air Purifier has been demonstrated to destroy the following MS2 bacteriophage bioaerosol entrained on the filter of the subject device under the following exposure/working conditions:

Test Item(s)Test Results (Average Maximum Log Reduction /Entrainment Time at Fan Speed 5)
MS2Model AG300: >4.0 Logs / 120 minutes
BacteriophageModel AG600: >4.0 Logs / 120 minutesModel AG900: >4.0 Logs / 120 minutes

6. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, and intended use of Airgle Room Air Purifier are shown below in tabular form.

{8}------------------------------------------------

DeviceProposed DevicePrimary PredicateDevice 1Reference Device 1Comment
K202339K212824
ManufacturerAirgle CorporationMolekule, Inc.Bluezone Products, Inc.None
Product NameAirgle Room AirPurifierAG300/AG600/AG900Molekule Air Mini,Bluezone Model RX-450Air PurifierNone
Device typeMedical Ultraviolet AirPurifierPurifier, Air, Ultraviolet,MedicalMedical Ultraviolet AirPurifierSame
ClassificationClass II, FRA (21CFR880.6500)Class II, FRA (21CFR880.6500)Class II, FRA (21CFR880.6500)Same
Use typeOTCOTCOTCSame
Intended UseDestroy bacteria andviruses by exposure toultraviolet radiation.Destroy bacteria andviruses by exposure toultraviolet radiation.Kill bacteria and virus inthe airSame
Indications for useThe Airgle Room AirPurifier is a freestanding, air purifyingdevice utilizinggermicidal ultravioletlight intended forinactivation of indoorairborne aerosolsincluding bacteria andvirus in medicalfacilities and occupiedspaces.The Airgle® AirPurifier has beendemonstrated todestroy the followingMS2 bacteriophagebioaerosol entrainedon the filter of thesubject device underthe followingexposure/workingconditions:The Molekule Air Mini airpurifier is a deviceintended for medicalpurposes that is usedto capture 95% ofparticulate matter anddestroy bacteria andviruses by exposure toultraviolet radiation.The Molekule Air Mini airpurifier has beendemonstrated to destroythe following MS2bacteriophagebioaerosol entrained onthe filter of the subjectdevice under thefollowingexposure/workingconditions:The Bluezone ModelRX-450 Air Purifier is afree standing, airpurifying device utilizinggermicidal ultravioletlight (UV-C wavelengthsnear 254nm) intendedfor inactivation of indoorairborne aerosolsincluding bacteria, mold,and virus in medicalfacilities and occupiedspaces.The Bluezone Model Rx-450 Air Purifier has beendemonstrated to entrainand destroy the followingexposure/workingconditions:Subject and PredicateDevices tested against thesame organismReference Device testedagainst additionalorganisms

{9}------------------------------------------------

{10}------------------------------------------------

Test Item(s)Test Results (Average Maximum Log Reduction / Entrainment Time at Fan Speed 5)
MS2 BacteriophageModel AG300: >4.0 Logs / 120 minutesModel AG600: >4.0 Logs / 120 minutesModel AG900: >4.0 Logs / 120 minutes
Test ItemTest Result
VirusMS2 bacteriophage$5.09\pm0.42$ / 120 mins
Particulate matter0.3 to 1.0 micron size particles95% or greater according to ASHRAE 52.2
Test itemAverage Net Log Reduction/Time @ High Fan Speed. Room Temperature Test
Bact Bacillus atrophaeus4.95 / 60 minutes
Bact Bacillus atrophaeus4.00 / 48 minutes
Mold Penicillium rocqueforti4.40 / 60 minutes
Mold Penicillium rocqueforti4.00 / 52 minutes
Virus MS-2 bacteriophage5.32 / 60 minutes
Virus MS-2 bacteriophage4.00 / 40 minutes

{11}------------------------------------------------

{12}------------------------------------------------

DeviceProposed DevicePrimary PredicateDevice 1K202339Reference Device 1K212824Comment
Mechanism of ActionUV-C inactivation ofmicroorganismsUV light of sufficientenergy (UV-A) activatesphotocatalyst thatdestroysmicroorganismsentrained on the filterthrough a hotochemicalreaction.UV-C inactivation ofmicroorganismsSubject, Predicate andReference Devices allreduce viable and no-viableaerosols by entrainmentonto filter media andbiocidal action of UV Light
InstallationFree standingFree standingFree standingSame
Elements of designFan circulates airthrough a shieldedchamber where UVlight irradiatesmicrobial aerosolsFan circulates airthrough a shieldedchamber where UV lightirradiates microbialaerosolsFan circulates airthrough a shieldedchamber where UV lightirradiates microbialaerosolsSame
FilterYesYesYesSame
Filtration typeParticle filterParticle filterParticle filterSame
DeviceProposed DevicePrimary PredicateDevice 1K202339Reference Device 1K212824Comment
Internal FanYesYesYesSame
Fan speed5 speeds5 Speeds2 SpeedsNo new safety or efficacyquestions raised bydifferences. Anyperformance differencesmitiagted by test results
Germicidal UVYesYesYesSame
UV Optic typeQuartz tube with lowpressure mercuryvaporUV Light Source:LEDQuartz tube with lowpressure mercury vaporNo new safety or efficacyquestions raised bydifferences. Anyperformance differencesmitiagted by test results.
UV wavelength254nm320-400 nm254nmNo new safety or efficacyquestions raised bydifferences. Any
DeviceProposed DevicePrimary PredicateDevice 1K202339Reference Device 1K212824Comment
performance differencesmitigated by test results.
Ozone EmissionDoes not exceed aconcentration of 0.050ppm according to UL867Does not exceed aconcentration of 0.050ppm according to UL867Does not exceed aconcentration of 0.050ppm according to UL867Same
Removal of bacterialsporesNot claimedNot claimed99.99%No new safety or efficacyquestions raised bydifferences. Performancedifferences mitigated bylabeling claims.
Removal of moldsporesNot claimedNot claimed99.99%No new safety or efficacyquestions raised bydifferences. Performancedifferences mitigated bylabeling claims
DeviceProposed DevicePrimary PredicateDevice 1K202339Reference Device 1K212824Comment
Removal ofAerosolized MS2Bacteriophage4.0 net LOG reductionin a 30 m3 space at120 minutes4 log reduction(99.99%)4.0 net LOG reduction ina 16m3 space at 35minutes.No new safety or efficacyquestions raised bydifferences. Performancedifferences mitigated bylabeling claims
UV radiance andleakage in occupiedspacesNoneNoneEffective irradiance <0.1uW/cm2Same
Chemical AdditivesNoNoNoSame
Electrical safetyIEC 60601-1AAMI TIR69:2017ANSI C63.27-2017UL 507UL 507No new safety or efficacyquestions raised bydifferences in Standardsapplied.
EMCIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2Same

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

7. Summary for non-clinical test

The non-clinical testing was provided to demonstrate the subject device met the acceptance criteria of the test methodology or standards listed below. Airgle Room Air Purifier complies with voluntary standards for electrical safety and electromagnetic compatibility.

Name of testmethodologyPurposeAcceptance CriteriaResults
IEC 60601-1IEC 60601-1-2Demonstration of basic safety and essential performanceMeets criteria for CISPR Group 1APASS
AAMI TIR69:2017ANSI C63.27-2017wireless coexistence testWireless System ability to coexist with interfering networks in the 2.4G band.PASS
Bacteriophage TestDemonstration of log reduction of viral loadAt least 99.99% reduction of MS2 bacteriophage after 2 hoursPASS
ANSI/ASHRAEStandard 52.2Demonstration of filter efficiencyAt least 99.3% for 0.3-10 umPASS
Software validationDemonstration of software security and good designMeets design criteriaPASS

8. Summary for clinical test

Not applicable

9. Conclusion

The Conclusion drawn from the non-clinical testing demonstrates that the subject device, Airgle Room Air Purifier, is as safe, as effective, and performs as well for the documented intended use as the legally marketed predicate device cleared under K202339 and the reference device cleared under K212824.

§ 880.6500 Medical ultraviolet air purifier.

(a)
Identification. A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.(b)
Classification. Class II (performance standards).