K Number
K232574
Manufacturer
Date Cleared
2023-10-11

(47 days)

Product Code
Regulation Number
870.1330
Panel
CV
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.

Device Description

The VASSALLO® GT Crossing14 Peripheral Guide Wire in this submission is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) in available lengths of 190cm and 300cm. The outer coil wire diameter has been modified to 0.072 mm. The devices have a solid core with a hvdrophilic coated coil-type distal end. The coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy. The core shaft surface is coated with Polytetrafluoroethylene (PTFE). About 2 cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the quide wire with a length of less than 300cm. A Torque device is included in the same package.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and device performance:

It's important to note that the provided document is a 510(k) summary for a medical device (a guide wire), not a publication about an AI/ML medical device. Therefore, many of the requested elements for AI/ML device studies (e.g., ground truth methods, expert qualifications, effect size for human readers with AI assistance) are not applicable to this document. This summary focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing and biocompatibility assessments, not on the performance of a diagnostic or assistive AI system.

Here's the information that can be extracted from the document, with explanations for what cannot be:


1. A table of acceptance criteria and the reported device performance

The document states that "The in vitro bench tests demonstrated that the VASSALLO GT Crossing14 met all acceptance criteria and performed similarly to the predicate devices." However, the specific quantitative acceptance criteria for each test and the corresponding precise device performance values are not explicitly detailed in this 510(k) summary. It only lists the tests performed.

Acceptance Criteria (General)Reported Device Performance (General)
Meets all specified acceptance criteriaMet all acceptance criteria for listed non-clinical tests.
Performs similarly to the predicate devicesPerformed similarly to the predicate devices.
Functions as intendedFunctions as intended.
Safety and effectiveness profile similar to predicateHas a safety and effectiveness profile similar to the predicate devices.
BiocompatibleBiocompatibility of the subject device is supported by leveraging testing performed on the predicate device.
Sterilizable with existing validated cycleThe minor differences do not represent a greater sterilization challenge, and the device was adopted into a product family with a validated sterilization cycle.

Non-clinical tests performed (details on specific criteria and results are not provided):

  • Dimensional Verification
  • Visual Inspection
  • Tensile Strength / Tip Pull
  • Torque Strength
  • Torqueability
  • Coating Adhesion/Integrity
  • Catheter Compatibility / Lubricity
  • Corrosion Resistance
  • Kink Resistance
  • Tip Flexibility
  • Radiopacity

Biocompatibility tests performed (details on specific criteria and results are not provided; leveraged from predicate):

  • Cytotoxicity
  • Sensitization
  • Intracutaneous Irritation
  • Acute Systemic Toxicity
  • Material Mediated Pyrogenicity
  • Hemolysis
  • Partial Thromboplastin Time
  • In Vivo Thromboresistance
  • Complement Activation (SC5b-9)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document refers to "non-clinical laboratory testing" and "in vitro bench tests." It does not specify sample sizes for these tests (e.g., how many guidewires were tested for tensile strength). Since these are bench tests, "data provenance" in terms of patient data (country of origin, retrospective/prospective) is not applicable. The tests were performed to support a device manufactured by FILMECC CO., LTD. in Japan.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is not an AI/ML medical device or a clinical study requiring expert ground truth for interpretation. The "ground truth" for these non-clinical tests would be the physical and chemical properties measured against established engineering and material standards.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is not an AI/ML medical device or a clinical study involving subjective interpretation requiring adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document describes a traditional medical device (guide wire) and does not involve AI assistance or human readers.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This document describes a traditional medical device (guide wire) and does not involve an algorithm.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For non-clinical bench testing, the "ground truth" is defined by established engineering standards, material specifications, and physical/chemical measurements. For biocompatibility, it's defined by biological responses according to ISO standards for biocompatibility testing. It is not expert consensus, pathology, or outcomes data.


8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.


9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device that requires a training set.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.