K Number
K232550
Date Cleared
2023-10-20

(58 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Nitrile Examination Gloves, Dual Color Black-Blue are Class I Patient Examination Gloves. The glove is made dual color by double dipping. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19. The gloves are single use, disposable, and provided non-sterile.

AI/ML Overview

This document describes the non-clinical performance data for the "Nitrile Examination Gloves, Dual Color Black-Blue" (K232550) and compares it to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

MethodologyTest PerformedAcceptance CriteriaReported Device Performance
ASTM D6319-19Physical Dimensions LengthMinimum 220mm for size XS-S; Minimum 230mm for size M-XXLPass
ASTM D6319-19Physical Dimensions Palm WidthXS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
ASTM D6319-19Physical Dimensions ThicknessFinger: 0.05mm (min); Palm: 0.05mm (min)Pass
ASTM D6319-19 & ASTM D412-16(2021)Physical PropertiesTensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) MinPass
ASTM D6319-19 & ASTM D5151-19Water leak testG-I, AQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19 & ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
ISO 10993-10 & 23:2021Irritation and Skin SensitizationSkin sensitization and Skin irritation requirements (e.g., non-sensitizing, non-irritating)Is non-sensitization and Non-irritation
ISO 10993-5:2009CytotoxicityCytotoxicity reactivity (no adverse reaction)showed potential toxicity to L929 cells.
ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each specific test in the "Summary of Non-Clinical Performance Data" section. It refers to various ASTM and ISO standards, which typically specify sample sizes for their respective test methodologies. The data provenance is not specified, but it can be inferred that these are laboratory tests performed on samples of the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. These are laboratory-based performance tests for a medical device (examination gloves), not studies requiring expert interpretation of medical images or patient data. The "ground truth" is established by the standardized test methodologies themselves.

4. Adjudication Method for the Test Set

Not applicable. This is not a study requiring human adjudication. The results are based on objective measurements and observations according to standardized test protocols.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical cases. This document pertains to the performance of examination gloves.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This document is for a physical medical device (examination gloves), not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance criteria is established by the specifications defined within the referenced ASTM and ISO standards (e.g., minimum length, maximum powder residue, specified AQL for water leak test, biological response classifications).

8. The Sample Size for the Training Set

Not applicable. This document pertains to the performance of a physical medical device (examination gloves) through non-clinical testing, not a machine learning or AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable (as explained in point 8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 20, 2023

Shanxi Hongjin Plastic Technology Co., Ltd. Zhigang Wu General Manager Coal Bed Gas Industrial Zone, Qu'e Town, Daning County Linfen, Shanxi 042300 China

Re: K232550

Trade/Device Name: Nitrile Examination Gloves, Dual Color Black-Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 20, 2023 Received: August 23, 2023

Dear Zhigang Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K232550

Device Name

Nitrile Examination Gloves, Dual Color Black-Blue

Indications for Use (Describe)

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

The assigned 510(K) numbers: K232550 Date Prepared: October 13, 2023

1. Owner's Identification:

Mr. Wu Zhigang Shanxi Hongjin Plastic Technology Co., Ltd. Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province Email: fdareg@hongray.com.cn

Contact: Mr. Wu Zhigang/General Manager Address: 3973 Schaefer Avenue, Chino, CA 91710, USA Tel: 909-590-1611 Email: janicema@hongrayusa.com

2. Name of the Device:

Trade / Product Name: Nitrile Examination Gloves, Dual Color Black-Blue Common Name: Exam Gloves Classification Name: Non-powdered Patient Examination Glove Classification Regulation: 21 CFR 880.6250 Product Code: LZA Classification Panel: General Hospital Device Class: Class I

3. Predicate Device Information:

Careplus (M) SDN BHD. Powder Free Nitrile Examination Glove Black and Dual Color White-Blue (K201857)

4. Device Description:

Nitrile Examination Gloves, Dual Color Black-Blue are Class I Patient Examination Gloves. The glove is made dual color by double dipping. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL.

Gloves meet the specification of ASTM D6319-19. The gloves are single use, disposable, and provided non-sterile.

5. Indications for Use:

The Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

6. Comparison of Subject Device and Predicate Device:

Subject DeviceK232550Predicate DeviceK201857ComparisonResult
Trade NameNitrile Examination Gloves, Dual Color Black-BluePowder Free Nitrile Examination Glove Black and Dual Color White-BlueSimilar
Product CodeLZALZASame
Regulation Number21 CFR 880.625021 CFR 880.6250Same

{5}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

ClassIISame
Indications for UseNitrile Examination Gloves, DualColor Black-Blue is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patient andexaminer.Powder Free Nitrile Examination GloveBlack and Dual Color White-Blue is adisposable device intended for medicalpurpose that is worn on the examiner'shand or finger to prevent contaminationbetween examiner and patient.Same
MaterialNitrileNitrileSame
Powder or Powder FreePowder FreePowder FreeSame
ColorBlack-BlueWhite-BlueSimilar
Single useSingle useSingle useSame

Technological Characteristic Comparison Table:

TechnologicalCharacteristicsSubject DeviceK232550Predicate DeviceK201857Comparison
LengthMinimum 220mm for size XS-SMinimum 230mm for size M-XXLMinimum 220mm for size XS-SMinimum 230mm for size M-XXXLDifferent*
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
XXL130±10130±10Same
XXXLNo126-132Different*
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, BeforeAging14MPa, min14MPa, minSame
UltimateElongation, BeforeAging500%, min500%, minSame
Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation,After Accelerated Aging400%, min400%, minSame
Watertight (1000ml)21 CFR 800.20ASTM D5151G-I, AQL 2.521 CFR 800.20ASTM D5151G-I, AQL 2.5Same

{6}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

Powder-Content≤ 2 mg per glove≤ 2 mg per gloveSame
10993-23:2021 SkinIrritation StudyUnder the conditions of thestudy, not an irritantUnder the conditions of the study,the subject device is non-irritatingSame
10993-10:2021Maximization SensitizationStudyUnder the conditions of the study,not a sensitizerUnder the conditions of the study,the subject device is nonsensitizationSame
10993-5:2009 CytotoxicityTestUnder the conditions of this study,the test article extract showedpotential toxicity to L929 cells.Cytotoxic (grade >2) at undiluted,1:2, 1:4, 1:8 and Noncyctotoxic,grade <2 at dilutions;1:16,1:32 and1:64.Similar
ISO 10993-11:2017 AcuteSystemic toxicity studyUnder the conditions of this study,there was no evidence of systemictoxicity.Under the conditions of this study,the test material did not revealsystemic toxicitySame

*The difference is for dimension and does not raise different questions of safectiveness and does not affect the substantial equivalence in effectiveness and safety.

7. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

MethodologyTest PerformedAcceptance CriteriaResults
ASTM D6319-19Physical Dimensions LengthMinimum 220mm for size XS-SMinimum 230mm for size M-XXLPass
ASTM D6319-19Physical Dimensions Palm WidthXS: 70±10mmS: 80±10mmM: 95±10mmL:110±10mmXL: 120±10mmXXL: 130±10mmPass
ASTM D6319-19Physical Dimensions ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Pass
ASTM D6319-19ASTM D412-16(2021)Physical PropertiesTensile Strength (Min14 MPa) and Elongation (Before Aging 500% and after aging 400%) MinPass
ASTM D6319-19ASTM D5151-19Water leak testG-I, AQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
ISO 10993-10 &23:2021Irritation and Skin SensitizationSkin sensitization and Skin irritationIs non-sensitization and Non-irritation
ISO 10993-5:2009CytotoxicityCytotoxicity reactivityshowed potential toxicity to L929 cells.
ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. ●

{7}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ●
  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves .
  • ASTM D412-16 (2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension
  • ISO 10993-10:2021 Biological Evaluation of Medical Devices Part 10: Tests For Skin Sensitization. ●
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: Tests For In Vitro Cytotoxicity ●
  • ISO 10993-23:2021 Biological Evaluation of Medical Devices Part 10: Tests For Skin Irritation. .
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity ●

8. Clinical Performance Data

N/A

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.