(90 days)
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned dectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
NemoCast is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes STL Models (also called dental casts) for thermoforming aligners, STL files for direct printing aligners and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). The device has no patient contact.
This document (K232549) is a 510(k) premarket notification for the device "NemoCast," an orthodontic software. It establishes substantial equivalence to existing predicate devices, particularly 3Shape Ortho System™ (K152086) and a previous version of NemoCast (K193003).
The key takeaway is that this is primarily a software validation and substantial equivalency claim, rather than a study proving new clinical performance. The manufacturer is demonstrating that their current software performs similarly to a previously cleared version and a predicate device.
Here's an analysis of the provided information regarding acceptance criteria and performance study:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in the typical format of a clinical study summary. Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to predicate devices. The "reported device performance" is the functionality of NemoCast itself, and the "study" is the comparison against the predicate devices.
The "Comparison of Intended Use and Technological Characteristics with the reference Device" table (pages 7-8) serves as the primary evidence of meeting "acceptance criteria" for substantial equivalence. It lists various features and functions, and for each, it aims to show "None" under "Differences" or a difference that does not affect safety and effectiveness.
Here's a condensed version of how that table functions as a de facto "acceptance criteria" and "performance report":
| Feature | Acceptance Criteria (Implied by Predicate) | NemoCast Performance (Reported) |
|---|---|---|
| Product Code | PNN, LLZ | PNN, LLZ |
| Common Name | Orthodontic Software | Orthodontic Software |
| Classification Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket |
| Regulation Number | 21 CFR 872.5470 | 21 CFR 872.5470 |
| Supported anatomic areas | Maxilla and Mandible | Maxilla and Mandible |
| Use by dental professionals in orthodontic treatment planning | Yes (NemoCast K193003: "only before treatment") | Yes (NemoCast K232549: "before, during, after treatment") Difference noted: broadened scope compared to reference device, but aligned with primary predicate. |
| Management of patients and models | Yes | Yes |
| Inspection, measurement and analysis of orthodontic models | Yes | Yes |
| Treatment simulation | Yes | Yes |
| Virtual appliance preparation, handling and export | Yes | Yes |
| Provide digital file and device output | Yes (STL files for dental casts) | Yes (STL files for dental casts, and additionally indirect Bonding Transfer Media) |
| Supported PC formats | Windows | Windows |
| Creating, editing, deleting and copying patient data | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes |
| Surface scan from intra-oral scanner | Yes | Yes |
| Surface scan from STL, PLY, OBJ file | Yes (Predicate: STL only) | Yes (STL, PLY, OBJ) Difference noted: Broader import formats, stated not to affect security/safety. |
| CT image data | DICOM | DICOM |
| 2D overlay | PNG, JPG, BMP | PNG, JPG, BMP |
| Aligning surface scan and CT image | Yes | Yes |
| Aligning cephalometric images | Yes | Yes |
| Alignment of surface scan with 2D overlays | Yes | Yes |
| Ability to check/adjust DICOM visibility | Yes | Yes |
| DICOM scan segmentation | Yes | Yes |
| Occlusal Orientation | Yes (Reference Device) / No (Predicate Device) | Yes Difference noted: Feature present in subject and reference, not in primary predicate. |
| Segmenting teeth roots | Yes (Reference Device) / No (Predicate Device) | Yes Difference noted: Feature present in subject and reference, not in primary predicate. |
| DICOM orientation | Yes | Yes |
| 2D measurement toolbox | Yes | Yes |
| 3D measurement toolbox | Yes | Yes |
| Arch shape analysis | Yes | Yes |
| Wire length analysis | Yes | Yes |
| Tooth width analysis | Yes | Yes |
| Bolton analysis | Yes (Predicate Device) / No (Reference Device) | Yes |
| Space analysis | Yes | Yes |
| Overjet/overbite analysis | Yes (Predicate Device) / No (Reference Device) | Yes |
| Occlusion map | Yes | Yes |
| Treatment analysis and report generation | Yes | Yes |
| 2D & 3D simulation | Yes | Yes |
| Orthodontic appliance search | Yes | Yes |
| Orthodontic appliance virtual preparation | Yes | Yes |
| Orthodontic appliance design | Yes | Yes |
| Orthodontic appliance export | Yes | Yes |
| Virtual articulator | Yes | Yes |
| Intended User | Dental Professionals | Dental Professionals |
| Intended Patient Population | Patients requiring Orthodontic Treatment (Predicate) / Adults requiring Orthodontic Treatment (Reference) | Patients requiring Orthodontic Treatment Difference noted: Aligned with primary predicate. |
2. Sample size used for the test set and the data provenance
The document states: "The performance testing remains unchanged from the company's own reference device submission, NemoCast K193003. The performance testing for the subject device is being leveraged from the company's own reference device including: design verification and validation testing."
This implies that the sample size and data provenance for the current 510(k) submission are not new. They are relying on previous testing. The document does not explicitly state the sample size (number of cases/patients) or the data provenance (e.g., country of origin, retrospective/prospective) for the test set. This information would typically be found in the original K193003 submission or internal validation reports, which are not detailed here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. As this is a software substantial equivalence submission leveraging previous testing, details about the ground truth establishment for the test set (number and qualifications of experts) are not specified here.
4. Adjudication method for the test set
This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a MRMC comparative effectiveness study demonstrating improved human reader performance with AI assistance was not done for this submission. The "NemoCast" device is described as a "medical front-end device providing tools for management... analysis, treatment simulation, and virtual appliance design." It is a planning and design software, not an AI-assisted diagnostic tool that would typically involve a human-in-the-loop MRMC study for assessing reader improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document mentions "The software is thoroughly tested in accordance with a documented test plan. This test plan is derived from the specifications and ensures that all controls and features are functioning properly. The software is validated together with end-users." This general statement indicates functional and validation testing that would assess standalone performance (software functioning as intended). However, specific quantifiable metrics of "algorithm-only" performance (e.g., accuracy of automatic measurements) are not reported in this summary. The focus is on the functional equivalence of the tools provided within the software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the type of "ground truth" used for the underlying validation of the software's measurements or simulations. Given the nature of orthodontic planning software, ground truth for features like measurements or segmentations would typically be established through:
- Expert consensus: Manual measurements or segmentations performed by experienced orthodontists on 3D models.
- Physical measurements/phantom data: Validation against known physical dimensions or phantom models.
However, these details are not provided in this 510(k) summary.
8. The sample size for the training set
This information is not provided. As this is a software product, not necessarily one relying on a large deep learning model needing a defined "training set" in the context of AI development for image interpretation, this detail might not be applicable or explicitly stated. If there are features utilizing machine learning (which is not explicitly detailed but possible for functions like segmentation), the training set details are not included. The submission is focused on demonstrating functional equivalence.
9. How the ground truth for the training set was established
This information is not provided and is likely not applicable in the context of this 510(k) submission, as it focuses on demonstrating substantial equivalence of a software tool rather than a novel AI algorithm with a distinct training phase. If machine learning components exist, the process for establishing their "training ground truth" is not disclosed here.
In summary, the provided text details a 510(k) submission for substantial equivalence. It does not contain the detailed clinical study information (like sample sizes, expert qualifications, adjudication methods, or MRMC study results) that would typically accompany a submission for a novel diagnostic AI device where independent performance validation is the primary focus. Instead, it relies on demonstrating that the updated software (NemoCast K232549) is functionally equivalent to its predicate devices (NemoCast K193003, 3Shape Ortho System™ K152086) and that any differences do not impact safety or effectiveness.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
November 21, 2023
Software Nemotec S.L. % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127
Re: K232549
Trade/Device Name: NemoCast Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN, LLZ Dated: August 23, 2023 Received: August 23, 2023
Dear Kevin Walls:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K232549
Device Name
NemoCast
Indications for Use (Describe)
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned dectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
즈 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the letters in a dark blue color. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in a smaller, lighter blue font. To the right of the text is a circular graphic with a dark blue circle in the center and three light blue segments around it.
510(K) Summary - K232549
This summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Applicant Information:
Software Nemotec S.L.
Av. Juan Caramuel, №1
Leganés, Madrid, Spain, 28919
Submission Correspondent:
Kevin Walls Principal Consultant, Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127
Device Information:
| Trade Name: | NemoCast |
|---|---|
| Common Name: | Orthodontic Software |
| Classification Name: | Orthodontic Plastic Bracket |
| Classification Regulation: | 21CFR 872.5470 |
| Device Class: | II |
| Product Code: | PNN, LLZ |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the "NEMO" in a darker blue and the "TEC" in a lighter blue. Below the name is the text "A BIOTECH DENTAL COMPANY" in a smaller, lighter font. To the right of the name is a circular graphic divided into three sections of varying shades of blue, with a dark blue circle in the center.
Reference Device:
NemoCast, Software Nemotec, K193003
Primary Predicate:
3Shape Ortho System™, 3Shape A/S, K152086
Date Prepared:
Aug 08, 2023
Predicate Device:
The NemoCast Software has the same intended uses and technical characteristics as primary predicate 3Shape Ortho System™ (K152086) from 3Shape A/S and reference predicate NemoCast from Software Nemotec (K193003) as listed in the table below:
Based on the information and supporting documentation provided, the NemoCast Software and the primary predicate (3Shape Ortho System™) have the same intended use.
Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering management of patients and models, inspection, 2D and 3D measurement and orthodontic analysis of models, 2D & 3D treatment simulation, as well as virtual appliance preparation, handling and export, and they are both providing device output. Additionally, the indirect bonding functionality of both systems is intended for use with commercially available brackets and wires. Therefore, the NemoCast Software and the primary predicate (3Shape) are found to be similar in their intended use, supported anatomic areas and the majority of the available features and functionalities.
Device Description:
NemoCast is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes STL Models (also called dental casts) for thermoforming aligners, STL files for direct printing aligners and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). The device has no patient contact.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the first three letters in a darker blue and the last four letters in a lighter blue. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in a smaller, sans-serif font. To the right of the word "NEMOTEC" is a circular graphic divided into three sections, with a smaller circle in the center.
Indications for Use:
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in large, bold, dark blue letters. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in smaller, light blue letters. To the right of the text is a circular graphic with three sections in different shades of blue and a dark blue circle in the center.
Comparison of Intended Use and Technological Characteristics with the reference Device:
| Substantial Equivalence Table | ||||
|---|---|---|---|---|
| Feature name | SUBJECTDEVICENemoCast | REFERENCEDEVICENemoCast(K193003) | PRIMARYPREDICATE3Shape OrthoSystem ™,K152086 | Differences |
| Product Code | PNN, LLZ | PNN, LLZ | PNN, LLZ | None |
| Common Name | OrthodonticSoftware | OrthodonticSoftware | OrthodonticSoftware | None |
| Classification Name | OrthodonticPlasticBracket | OrthodonticPlastic Bracket | OrthodonticPlastic Bracket | None |
| Regulation Number | 21 CFR872.5470 | 21 CFR872.5470 | 21 CFR872.5470 | None |
| Supported anatomic areas | Maxilla andMandible | Maxilla andMandible | Maxilla andMandible | None |
| Intended Use | ||||
| Use by dental professionals inorthodontic treatmentplanning (before, during, aftertreatment). | Yes | Yes | Yes | Reference deviceNemoCast wasused by dentalprofessionalsonly beforetreatment. |
| Management of patients andmodels. | Yes | Yes | Yes | None |
| Inspection, measurement andanalysis of orthodonticmodels. | Yes | Yes | Yes | None |
| Treatment simulation. | Yes | Yes | Yes | None |
| Virtual appliance preparation(including dental casts),handling and export. | Yes | Yes | Yes | None |
| Provide digital file and deviceoutput. | Yes | Yes | Yes | All 3 devicesoutput STL filesfor fabrication ofdental casts.The subjectDevice andpredicate Device |
| both additionallyoutput indirectBonding TransferMedia. | ||||
| Supported PC formats | Windows | Windows | Windows | None |
| Managing patient and case base data: | ||||
| Creating, editing, deletingand copying patient data | Yes | Yes | Yes | None |
| Creating, editing, deletingand copying case data | Yes | Yes | Yes | None |
| Collection of study material | ||||
| Surface scan fromintra-oral scanner | Yes | Yes | Yes | None |
| Surface scan from STL, PLY,OBJ file | Yes | No | Yes, only STL | Althoughpredicate deviceonly import STLformat, the OBJand PLY formatsare otherequivalentformats which donot affect thesecurity andsafety of thedevice. |
| CT image data | DICOM | DICOM | DICOM | None |
| 2D overlay | PNG, JPG,BMP | PNG, JPG,BMP | PNG, JPG,BMP | None |
| Alignment of study material | ||||
| Aligning surface scan andCT image | Yes | Yes | Yes | None |
| Aligning cephalometricimages | Yes | Yes | Yes | None |
| Alignment of surface scanwith 2D overlays | Yes | Yes | Yes | None |
| Ability to check/adjustDICOM visibility | Yes | Yes | Yes | None |
| DICOM scan segmentation | Yes | Yes | Yes | None |
| Occlusal Orientation | Yes | Yes | No | Although thepredicate devicedoes not performocclusal |
| orientation, the | ||||
| reference device | ||||
| contains this | ||||
| Segmenting teeth roots | Yes | Yes | No | option.Although the |
| predicate device | ||||
| does not | ||||
| segment teeth | ||||
| roots, the | ||||
| reference device | ||||
| contains this | ||||
| option. | ||||
| DICOM orientation | Yes | Yes | Yes | None |
| Measuring study material | ||||
| 2D measurement toolbox | Yes | Yes | Yes | None |
| 3D measurement toolbox | Yes | Yes | Yes | None |
| Analyzing study material | ||||
| Arch shape | Yes | Yes | Yes | None |
| Wire length | Yes | Yes | Yes | None |
| Tooth width | Yes | Yes | Yes | None |
| Bolton | Yes | No | Yes | None |
| Space analysis | Yes | Yes | Yes | None |
| Overjet/overbite | Yes | No | Yes | None |
| Occlusion map | Yes | Yes | Yes | None |
| Treatment analysis andreport generation | Yes | Yes | Yes | None |
| Treatment simulation | ||||
| 2D & 3D simulation | Yes | Yes | Yes | None |
| Virtual appliance design | ||||
| Orthodontic appliancesearch | Yes | Yes | Yes | None |
| Orthodontic appliance | Yes | Yes | Yes | None |
| virtual preparation | ||||
| Orthodontic appliancedesign | Yes | Yes | Yes | None |
| Orthodontic applianceexport | Yes | Yes | Yes | None |
| Virtual articulator | Yes | Yes | Yes | None |
| Indications for Use | NemoCast is | NemoCast is | 3Shape Ortho | The indications |
| intended for | intended for | System™ is | for use for the 3 | |
| use as a | supporting the | intended for | devices are the | |
| medical | diagnostic and | use as a | same, with the | |
| front-end | treatment | medical | exception of the | |
| device | plan processes | front-end | virtual design | |
| providing | for | device | options. | |
| tools for | orthodontic | providing | (Compared to the | |
| management | procedures | tools for | reference device, | |
| of | related to | tools for | the Subject | |
| orthodontic | minor anterior | management | Device | |
| models, | tooth | of orthodontic | additionally is | |
| systematic | movements; it | models, | used for design | |
| inspection, | provides tools | systematic | of Indirect | |
| detailed | for | inspection, | Bonding Transfer | |
| analysis, | management | detailed | Media. | |
| treatment | of orthodontic | analysis, | ||
| simulation | models, | treatment | The Subject | |
| and virtual | systemic | simulation and | Device virtual | |
| appliance | inspection, | virtual | design options | |
| design | detailed | appliance | are the same as | |
| options | analysis, | design options | the Virtual design | |
| (Export of | treatment | (Custom Metal | options for the | |
| Models, | simulation | Bands, Export | predicate | |
| Indirect | and virtual | of Models, | Device.) | |
| Bonding | appliance | Indirect | ||
| Transfer | design options | Bonding | ||
| Media) based | (export of | Transfer | ||
| on 3D models | models) based | Media) based | ||
| of the | on 3D models | on 3D models | ||
| patient's | of the | of | ||
| dentition | patient's | the patient's | ||
| before the | dentition | dentition | ||
| start of an | before the | before the start | ||
| orthodontic | start of an | of an | ||
| treatment. It | orthodontic | orthodontic | ||
| can also be | treatment. It | treatment. It | ||
| applied | can also be | can also be | ||
| during the | applied to | applied during | ||
| treatment to | evaluate if the | the treatment | ||
| inspect and | outcome is | to inspect and | ||
| analyze the | consistent | analyze the | ||
| progress of | with | progress of the | ||
| the | the | treatment. It | ||
| treatment. It | planned/desir | can be used at | ||
| can be used | ed treatment | the end of the | ||
| at the end of | objectives. | treatment to | ||
| the treatment | NemoCast | evaluate if the | ||
| to evaluate if | requires the | outcome is | ||
| the outcome | user to have | consistent with | ||
| is consistentwith theplanned/desired treatmentobjectives. Itcan also beappliedduring thetreatment toinspect andanalyze theprogress ofthetreatment.The use oftheNemoCastrequires theuser to havethe necessarytraining anddomainknowledge inthe practiceoforthodontics,as well as tohave receiveda dedicatedtraining in theuse of thesoftware. | the necessarytraining anddomainknowledge inthe practice oforthodontics,as wellas to havereceived adedicatedtraining in theuse of thesoftware. | theplanned/desired treatmentobjectives.The use of theOrthoSystemTMrequires theuser to havethe necessarytraining anddomainknowledge inthe practice oforthodontics,as well as tohave received adedicatedtraining in theuse of thesoftware. | ||
| Intended User | DentalProfessionals | DentalProfessionals | DentalProfessionals | None |
| Intended PatientPopulation | PatientsrequiringOrthodonticTreatment | AdultsrequiringOrthodonticTreatment | PatientsrequiringOrthodonticTreatment | The intendedpatientpopulation forthe subjectdevice is thesame as thepredicate device.The referencedevice intendedpatientpopulation was |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the first four letters in a darker blue and the last three in a lighter blue. Below the company name, the words "A BIOTECH DENTAL COMPANY" are written in a smaller font. To the right of the company name is a circular graphic with a dark blue dot in the center and three sections in different shades of blue surrounding the dot.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the letters in a dark blue color. Below the company name is the text "A BIOTECH DENTAL COMPANY" in a smaller, lighter font. To the right of the company name is a circular graphic with a dark blue dot in the center.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "NEMOTEC" in blue font, with the words "A BIOTECH DENTAL COMPANY" in smaller font underneath. To the right of the word "NEMOTEC" is a circular logo with a blue dot in the center. The logo is made up of three different shades of blue.
Image /page/10/Picture/1 description: The image shows a circular diagram divided into three sections. The sections are colored in varying shades of blue, with the darkest shade at the top right, a medium shade at the bottom, and the lightest shade at the top left. There is a solid dark blue circle in the center of the diagram.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the letters in a dark blue color. Below the company name, the words "A BIOTECH DENTAL COMPANY" are written in a smaller, lighter blue font. To the right of the company name is a circular graphic with a dark blue circle in the center.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the first two letters in a darker blue and the rest in a lighter blue. Below the company name, the words "A BIOTECH DENTAL COMPANY" are written in a smaller, sans-serif font. To the right of the company name is a circular graphic with three sections in different shades of blue and a white circle in the center with a dark blue dot.
TECHNOLOGICAL CHARACTERISTICS
NemoCast; Software Nemotec, is software for simulating/evaluating orthodontic treatment programmed in C++ language and running on the Windows operating system.
TESTING & VALIDATION
The software is thoroughly tested in accordance with a documented test plan. This test plan is derived from the specifications and ensures that all controls and features are functioning properly. The software is validated together with end-users.
The performance testing remains unchanged from the company's own reference device submission, NemoCast K193003. The performance testing for the subject device is being leveraged from the company's own reference device including: design verification and validation testing.
Differences:
The Indications for Use statement between the subject and predicate devices are equivalent; minor differences in wording do not alter the intended use of the subject device. In addition, there are minimum differences in the software functions or technical requirements; none of these differences affect the safety and effectiveness of the subject device relative to the predicate, both can be considered equivalent.
§ 872.5470 Orthodontic plastic bracket.
(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.