K Number
K232512
Device Name
Masimo W1
Manufacturer
Date Cleared
2023-11-17

(91 days)

Product Code
Regulation Number
870.2340
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Masimo W1™ and the integrated Masimo W1 module are intended for the spot-check determination of Heart Rate using a single-channel electrocardiogram (ECG). The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel ECG for the manual interpretation of heart rate. It is worn on the wrist and also provides other continuous parameters technologies (e.g., pulse oximetry).

The Masimo W1 ™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

Device Description

The Masimo W1 is a watch that incorporates the W1 Module, which is the device that is responsible for the physiological signal detection and algorithm in providing the supported parameters. The module incorporates ECG functionality for Heart Rate Monitoring and Masimo SET Pulse Oximetry technology so that it can provide both ECG (Heart Rate parameter) and Masimo SET Pulse Oximetry parameter output from the W1 Module is displayed on the touchscreen watch interface.

As part of the Masimo W1 watch, the Masimo W1 Module is integrated into the Masimo W1 watch platform, which consists of a typical IT hardware platform to enable other non-medical smart watch features (e.g., step counting, walking, running, fall detection and rise to wake).

AI/ML Overview

The provided text describes the regulatory clearance of the Masimo W1 device. It details various non-clinical and clinical tests performed to demonstrate its performance and substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly lay out a table of "acceptance criteria" in a pass/fail format with specific thresholds before the study results. Instead, it states the device's performance specifications as the target, and then presents the study results demonstrating that the device meets or exceeds these.

Here's a table based on the "Performance Specifications" listed in Table 1 and the "Clinical Performance Testing" results:

Feature/ParameterAcceptance Criteria (Stated Specification)Reported Device Performance (Clinical Study Result)
SpO2 (70-100%)2% ARMS (adults, No Motion/Low Perfusion)1.62% ARMS (after adjusting for repeated measures)
Pulse Rate (25-240 bpm)3 bpm (adults)Not explicitly stated in clinical results, but "supported the claimed pulse rate performance" implies it met the 3bpm spec.
Heart Rate (25-240 bpm)≤ 5 bpm (adults)"The testing supported the claimed heart rate performance and its substantial equivalence."

Note: For Pulse Rate and Heart Rate, the document states that testing supported the claimed performance, rather than providing a specific numerical ARMS or bias/precision value for these parameters from the clinical study itself. However, the stated acceptance criteria (specifications) are given.

2. Sample Sizes Used for the Test Set and Data Provenance

  • SpO2 Clinical Testing:
    • Sample Size: 31 healthy volunteer subjects (13 light, 12 medium, 6 dark pigmented subjects).
    • Data Provenance: Prospective clinical study. Country of origin is not specified, but typically conducted in the country of submission (USA in this case, given FDA submission).
  • Heart Rate (HR) Clinical Testing:
    • Sample Size: 61 subjects.
    • Data Provenance: Prospective clinical study. Country of origin is not specified.
  • ECG Waveform Quality Testing:
    • Sample Size: Not explicitly stated for specific subjects, but relates to "ECG waveforms collected by the Masimo W1."
    • Data Provenance: Not explicitly stated if prospective or retrospective, but implied from clinical validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • ECG Waveform Quality Testing:
    • Number of Experts: Three (3)
    • Qualifications: Board certified cardiologists.

For SpO2 and Heart Rate, the ground truth was established by reference devices (e.g., FDA cleared ECG reference measurement for HR, likely CO-oximetry for SpO2, in accordance with ISO 80601-2-61). The expertise for these would be in the operation and interpretation of these reference standards, not necessarily by individual expert review of each measurement.

4. Adjudication Method for the Test Set

  • ECG Waveform Quality Testing: The text states, "the ECG waveforms collected by the Masimo W1 were compared to Lead I of a gold standard reference of a 12-Lead ECG by three board certified cardiologists." It doesn't specify an explicit adjudication method (e.g., 2+1, 3+1). It implies independent review by each, leading to an overall assessment of acceptability. Given "compared to... by three board certified cardiologists," it suggests independent assessments that would then be aggregated or reconciled for the overall conclusion of "acceptability."
  • SpO2 and Heart Rate: No human adjudication method described. Ground truth was established using objective reference standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study involving human readers assisting with AI vs. without AI assistance was not described. The device's performance was evaluated for its ability to accurately measure physiological parameters (HR, SpO2, PR) against reference standards, and for the interpretability of its ECG waveforms by cardiologists, not for its role in assisting human interpretation or decision-making in a comparative setting.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, for the primary physiological parameters (SpO2, Pulse Rate, Heart Rate), the device's performance relies on its integrated module and algorithms. The studies evaluate the accuracy of these measurements compared to reference standards, which is a standalone performance assessment.

  • For ECG HR, the W1 module "detects and calculates the heart rate" and the single-channel ECG for the manual interpretation of heart rate. The clinical study for HR compared W1 measurements to an FDA cleared ECG reference. This suggests a standalone algorithmic performance for heart rate detection from the ECG signal.
  • For SpO2 and PR, the W1 module "continuously detects and processes the optical signals" and "utilizes multiple wavelengths of light and advanced signal processing techniques" to determine these. This is a standalone algorithmic performance.

While the ECG waveform itself is for "manual interpretation of heart rate," the accuracy of the Heart Rate determination by the device itself is a standalone function.

7. The Type of Ground Truth Used

  • SpO2: Clinical study results for SpO2 were conducted "in accordance with the ISO 80601-2-61," which typically means comparison to arterial blood gas measurements (CO-oximetry) as the reference standard. This is an objective, gold-standard reference method.
  • Heart Rate (HR): Ground truth established by an "FDA cleared ECG reference measurement." This is an objective, reference standard device.
  • ECG Waveform Quality: Ground truth established by comparison to "Lead I of a gold standard reference of a 12-Lead ECG by three board certified cardiologists." This involves expert consensus/interpretation against a clinical gold standard.

8. The Sample Size for the Training Set

The document does not provide information on the sample size used for the training set of the device's algorithms. The provided details focus on the data used for performance validation (i.e., test set). Medical device submissions often do not disclose the specifics of internal development and training datasets.

9. How the Ground Truth for the Training Set was Established

As the training set information is not provided, the method for establishing its ground truth is also not detailed in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

November 17, 2023

Masimo Corporation Sindura Penubarthi Associate Director, Regulatory Affairs 52 Discovery Irvine, California 92618

Re: K232512

Trade/Device Name: Masimo W1 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQA, DXH Dated: October 23, 2023 Received: October 23, 2023

Dear Sindura Penubarthi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K232512

Device Name

Masimo W1

Indications for Use (Describe)

Masimo W1™ and the integrated Masimo W1 module are intended for the spot-check determination of Heart Rate using a single-channel electrocardiogram (ECG). The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel ECG for the manual interpretation of heart rate. It is worn on the wrist and also provides other continuous parameters technologies (e.q., pulse oximetry).

The Masimo W1 ™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a stylized red checkmark-like symbol on the left, followed by the company name "Masimo" in a simple, sans-serif black font. The checkmark symbol is rounded and appears to be slightly tilted to the right.

Submitter and Address of ManufacturingFacility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7592
Date:November 17, 2023
Contact:Sindura PenubarthiAssociate Director, Regulatory AffairsMasimo CorporationPhone: (949) 396-4041
Trade Name:Masimo W1
Common Name:Electrocardiograph
Classification Regulation/ Product Code:21 CFR 870.2340, Class II/DPS
Additional Product Codes:DQADXH
Establishment Registration Number:3011353843
Reason for Premarket Notification:New Device
Predicate Device:K201456 – Withings Scan Monitor
Reference Device:K221260 - CSF-3

1.0 Device Description

The Masimo W1 is a watch that incorporates the W1 Module, which is the device that is responsible for the physiological signal detection and algorithm in providing the supported parameters. The module incorporates ECG functionality for Heart Rate Monitoring and Masimo SET Pulse Oximetry technology so that it can provide both ECG (Heart Rate parameter) and Masimo SET Pulse Oximetry parameter output from the W1 Module is displayed on the touchscreen watch interface.

As part of the Masimo W1 watch, the Masimo W1 Module is integrated into the Masimo W1 watch platform, which consists of a typical IT hardware platform to enable other non-medical smart watch features (e.g., step counting, walking, running, fall detection and rise to wake).

See Table 1 below for the Masimo W1 Specifications.

Table 1 - Specifications
FeatureSpecifications
Continuous Display of Parameter DataYes
User InterfaceTouchscreen

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a stylized red checkmark-like symbol on the left, followed by the company name "Masimo" in a bold, sans-serif font. The checkmark is a vibrant red color, while the company name is in black.

Table 1 - Specifications
FeatureSpecifications
Performance Specifications
SpO2, No Motion/ Low Perfusion (70-100%)2% adults
Pulse Rate, No Motion (25-240 bpm)3 bpm adults
Heart Rate (25-240 bpm)≤ 5 bpm adults
Electrical Specifications
BatteryInternal Rechargeable Li-Ion
Mechanical Specifications
Size40 mm (1.57")
Display TypeTouchscreen
Weight54 g (including watchband)
Environmental Specifications
Operating Temperature0 to 35 °C (32 to 95°F)
Operating Humidity10% to 95% RH (non-condensing)
Storage/Transport Temperature-20°C to 60°C (-4°F to 140°F)
Storage/Transport Humidity10% to 95% RH (non-condensing)
Classification per IEC 60601-1
Electrical SafetyIEC 60601-1
EMCIEC 60601-1-2
Electrical Isolation TypeInternally Powered
Applied Part TypeCF Applied Part
Ingress ProtectionIP24
Mode of OperationContinuous

2.0 Intended Use/ Indications for Use

Masimo W1™ and the integrated Masimo W1 module are intended for the spot-check determination of Heart Rate using a single-channel electrocardiogram (ECG). The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel ECG for the manual interpretation of heart rate. It is worn on the wrist and also provides other continuous parameters technologies (e.g., pulse oximetry).

The Masimo W1™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

3.0 Technological Characteristics

Principle of Operation

Electrocardiogram (ECG)

The ECG feature on Masimo W1 relies on the principle that the electrical signals can be detected as different parts of the heart contract and relax during a cardiac cycle allowing the detection of heart activity and for the estimation of heart rate (HR). The change in polarization of the heart muscles creates electrical signals that are propagated so that they can be detected at the skin surface of the wrist.

Pulse Oximetry-Based Parameters

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red checkmark symbol to the left of the word "Masimo" in a bold, sans-serif font. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

The Masimo SET pulse oximeter technology relies on the Beer-Lambert law and the following principles of pulse oximetry:

  • . Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
  • . The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.

Mechanism of Action for Achieving the Intended Effect

The Masimo W1 watch achieves its intended effect through the integrated Masimo W1 Module that provides both the electrical and optical sensing technology and parameter algorithms. The W1 Module is integrated into the W1 watch so that the sensing components contact the skin on the wrist. The top of the watch is provided with an electrical sensing pad that is contacted by the two fingers on the opposing hand. Contacting the electrical sensing pad allows for the detection of the ECG signal when the ECG feature is activated on the watch touchscreen. Once the ECG feature is activated, it detects and calculates the heart rate.

The Masimo SET pulse oximetry-based parameters are supported by the W1 Module's optical sensing components located on the bottom of the module that contacts the skin on the wrist. The W 1 Module continuously detects and processes the optical signals that change with the transmission of LED light into the wrist tissue. The W1 Module utilizes multiple wavelengths of light and advanced signal processing techniques to isolate the arterial signal from other static factors (e.g., skin pigment) to establish the ratio used in the estimation of the SpO2. The pulse rate is determined by the periodic changes in the photoplethysmograph (PPG). The parameters are then continuously updated and displayed on the watch so that it can be viewed, recorded, and/or transferred. The use of the Masimo W1 watch can be discontinued by taking off the watch or deactivating the continuous parameters from the watch's touchscreen.

4.0 Summary of Technological Characteristics of Subject Device Compared to Predicate Device

Similarities and Differences between Predicate and Subject Device

The subject device, Masimo W1, and the predicate device, Scan Monitor (K201456), have the following key similarities:

  • · Both devices have the same intended use;
  • · Both devices rely on the same principles of operation;
  • · Both devices are indicated for the same population of prescription and OTC users;

The subject device, Masimo W1, and the predicate device, Scan Monitor (K201456), have the following key differences:

  • · The subject device does not support ECG rhythm classifications (e.g., AFib);
  • · The subject device continuously updates pulse oximetry parameter data.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

The subject device and predicate devices were found to have intended use without any technological differences that raise different questions of safety and effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo features a red circle with a white line running through it on the left, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo, the text "MASIMO CORPORATION" is displayed above the address "52 Discovery, Irvine, CA 92618".

Feature510(k) NumberMasimo W1Subject DeviceWithings Scan MonitorPrimary Predicate(K201456)Comparison toPredicate Device
General Information
ClassificationRegulation/ ProductCode21 CFR 870.2340, Class II/DPS21 CFR 870.2340, Class II/DPSSame
Product Code(s)DQADXHDQADXHSame
Indications for UseMasimo W1™ and the integrated Masimo W1module are intended for the spot-checkdetermination of Heart Rate using a single-channel electrocardiogram (ECG). The MasimoW1 and the integrated Masimo W1 modulerecords, stores, transfers, and displays thesingle-channel ECG for the manualinterpretation of heart rate. It is worn on thewrist and also provides other-continuousparameters technologies (e.g., pulse oximetry).The Masimo W1TM and the integrated MasimoW1 module are also intended for the spot-checking of functional oxygen saturation ofarterial hemoglobin (SpO2) and pulse rate(PR). The Masimo W1 and the integratedMasimo W1 Module are indicated for adults inhospitals, clinics, long-term care facilities, andhomes.The Scan Monitor is intended to record,store and transfer single-channelelectrocardiogram (ECG) rhythms. TheScan Monitor also displays ECG rhythmsand detects the presence of atrialfibrillation (when the monitor isprescribed or used under the care of aphysician). The Scan Monitor is intendedfor use by healthcare professionals,patients with known or suspected heartconditions and health-consciousindividuals.The Scan Monitor is also indicated foruse in measuring and displayingfunctional oxygen saturation of arterialhemoglobin (SpO2). The Scan Monitor isintended for spot-checking of adultpatients in hospitals, clinics, long-termcare facilities and homes.Similar. The subjectdevice has similarindications as thepredicate for ECG HRand Pulse Oximetry
Technological Characteristics
Feature510(k) NumberMasimo W1Subject DeviceWithings Scan MonitorPrimary Predicate(K201456)Comparison toPredicate Device
Principles of OperationThe W1 Module software analyzes thepatterns in the ECG waveform to determinemanual interpretation of Heart Rate.The Masimo SET pulse oximeter technologyrelies on the Beer-Lambert law and thefollowing principles of pulse oximetry:Oxyhemoglobin (oxygenated blood) anddeoxyhemoglobin (non-oxygenatedblood) differ in their absorption of redand infrared light (spectrophotometry).The amount of arterial blood in tissuechanges with your pulse(photoplethysmography). Therefore, theamount of light absorbed by the varyingquantities of arterial blood changes aswell.The iECG software analyzes the patternsin the ECG waveform to determinedifferent types of known heart activity(e.g., Heart Rate, Atrial Fibrillation).Scan Monitor pulse oximetry technologyrelies on the differential absorption byblood of red (660nm) and infrared light(940nm) which relies on the Beer-Lambert law for the principles of pulseoximetry.Similar. The subjectdevice utilizes the sameprinciples related to ECGand Pulse Oximetry.
Supported MeasuredParametersHR, SpO2, PRHR, SpO2Different. The subjectdevice providesadditional pulseoximetry-based features(i.e., PR).Testing is provided tosupport the PRperformance.
Feature510(k) NumberMasimo W1Subject DeviceWithings Scan MonitorPrimary Predicate(K201456)Comparison toPredicate Device
Supported CalculatedfeaturesPiHRV, AFib ClassificationDifferent. The subjectdevice provides Pi.Testing is provided tosupport the Pi feature.
User InterfaceTouchscreenTouchscreenSame
Performance Specifications (Arms)
SpO2 (70-100%)2%, adults (No Motion/ Low Perfusion)3%, adults (No Motion)Different. Subject deviceincludes performancetesting to supportimproved specification.
Pulse Rate (25-240 bpm)3 bpmNot supportedDifferent. Subject deviceincludes performancetesting to supportimproved specification.
Heart Rate (25-240 bpm)5 bpmNot KnownTesting is provided tosupport the specification.
Electrical Specifications
BatteryInternal RechargeableInternal RechargeableSame
Mechanical Specifications
Watch Face Size40 mm (1.57")38 mm or 42 mmSimilar. Testing isprovided to support thesubstantial equivalence.
Weight54 g (including watchband)58 gms or 83gmsSimilar. The weightdifference is minor and isconsistent with othermarketed watches.
Environmental Specifications
Feature510(k) NumberMasimo W1Subject DeviceWithings Scan MonitorPrimary Predicate(K201456)Comparison toPredicate Device
Operating Temperature0 to 35 °C (32 to 95°F)-10 to 45 °C (14 to 113°F)Similar. Testing isprovided to support thespecification.
Operating Humidity10% to 95%, non-condensingNot KnownTesting is provided tosupport the specification.
Storage/TransportTemperature-20 to 60°C (-4 to 140°F)-20 to 85°C (4 to 185°F)Similar. Testing isprovided to support thespecification.
Storage/TransportHumidity10% to 95%, non-condensingNot KnownTesting is provided tosupport the specification.
Classification per IEC 60601-1
Electrical SafetyIEC 60601-1IEC 60601-1Same
EMCIEC 60601-1-2IEC 60601-1-2Same
Ingress ProtectionIP24Not KnownTesting is provided tosupport the specification.
Mode of OperationContinuousContinuousSame

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in black, sans-serif font. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, gray, sans-serif font, followed by the address "82 Discovery, Irvine, CA 92618" also in a smaller, gray, sans-serif font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized red symbol to the left of the word. To the right of the logo is the text "MASIMO CORPORATION", followed by the address "52 Discovery, Irvine, CA 92618" in a smaller font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the company name is the text "MASIMO CORPORATION, 52 Discovery, Irvine, CA 92618" in a smaller, lighter font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Masimo Corp. The logo consists of a red symbol on the left, followed by the word "Masimo" in a bold, sans-serif font. To the right of the word "Masimo" are the words "MASIMO CORP" in a smaller font, followed by the address "52 Discovery, Irvine, CA 92618".

5.0 Performance Data

The following non-clinical testing was provided to support the non-parameter specifications and the substantial equivalence of the subject device.

  • Biocompatibility in accordance with ISO 10993-1 ●
  • EMC testing per IEC 60601-1-2
  • . Electrical safety testing per IEC 60601-1
  • Environmental and Mechanical testing
  • Cleaning Validation
  • Software verification and validation testing per FDA Software Guidance ●
  • Human Factors Usability testing per FDA Human Factors and Usability Guidance

The following are the list of standards that were used as part of the evaluation:

  • IEC 60601-1:2005/2012
  • IEC 60601-1-2:2014
  • IEC 60601-1-6:2013 ●
  • IEC 60601-1-11:2015
  • IEC 60601-2-27:2011 ●
  • IEC 60601-2-47:2012
  • ISO 80601-2-61:2017 ●
  • ISO 10993-1:2018
  • IEC 62304:2015
  • IEC 62366-1:2015

Performance Bench Testing

Performance bench testing for the Masimo W1 is included in this submission to support both ECG HR and Pulse Oximetry based parameter performance.

Biocompatibility Testing

Biocompatibility testing in accordance with ISO 10993-1 is included as part of this submission to support the acceptability of the biocompatibility risks.

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

EMC testing was conducted in accordance with IEC 60601-1-2: 4.1 Edition and the electrical safety in accordance with the IEC 60601-1 standard. Environmental, cleaning, and chemical resistance testing was also provided to support the substantial equivalence of the Masimo W1.

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Masimo Corporation. The logo features a red circle with a white checkmark inside, followed by the word "Masimo" in black, sans-serif font. To the right of the logo, the text "MASIMO CORPORATION" is displayed, along with the address "62 Discovery, Irvine, CA 92618".

recommended by FDA Guidance for the Content of Premarket Submissions for Software Device Software Functions, June 2023. The software was found to fit in the category of products that would require Basic Documentation Level because the failure or latent flaw of the device software function would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device or others in the environment of use prior to the implementation of the risk controls. The software does not provide any data that is considered life supporting.

Human Factors Usability Testing

To support the usability of the Masimo W1, human factors and usability risks were evaluated to be acceptably mitigated in accordance with FDA Guidance, Applying Human Factors and Usability Engineering to Optimize Medical Device Design, dated February 3, 2016. The testing was found to support acceptability of the human factors and usability risks.

Clinical Performance Testing

To support the performance of the Masimo W1, clinical data is provided to support of the performance of the Masimo W1. To address potential concerns related to skin pigment discrepancies in pulse oximetry, the SpO2 clinical testing included 31 subjects with varying skin pigmentations measured using a colorbased scale, Massey-Martin. The results separated by skin pigment supported the performance of the Masimo W1 across different skin tones.

The SpO2 performance testing was conducted in accordance with the ISO 80601-2-61. The prospective clinical study included 31 healthy volunteer subjects, including 13 light, 12 medium and 6 dark pigmented subjects, which exceeded the ISO 80601-2-61 minimum sample size requirements and the minimum number of dark pigmented subjects in accordance with the FDA Guidance for Pulse Oximeters. Subjects were classified as dark and light based upon their Massey-Martin scores, 1-3 classified as "Light", 4-6 classified as "Medium" and 7-10 as "Dark" or subjects with more skin pigment.

The overall performance was 1.62% Arms after adjusting for repeated measures. The results support the specification of 2% Arms across the range of 70%-100% SaO2. The breakdown of the performance for dark, medium, and light subjects is provided below.

PigmentationBiasPrecisionARMSAdjustedPrecisionAdjustedARMSAdjustedLOANsubjNpairs
Dark(Massey 7-10)0.371.641.681.781.82[-3.14, 3.88]6329
Medium(Massey 4-6)-0.261.581.601.651.67[-3.49, 2.97]12724
Light(Massey 1-3)0.561.411.511.481.58[-2.34, 3.46]13831

To clinically validate the performance of the heart rate (HR) feature on the Masimo W1 watch, a prospective clinical study was conducted on 61 subjects where the spot-check HR measurements obtained

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

from the W1 module were compared to an FDA cleared ECG reference measurement. The testing supported the claimed heart rate performance and its substantial equivalence.

To clinically validate the ECG waveform quality of the Masimo W1, the ECG waveforms collected by the Masimo W1 were compared to Lead I of a gold standard reference of a 12-Lead ECG by three board certified cardiologists. The testing supported the acceptability of the detected ECG waveforms.

6.0 Conclusion

The data supported the substantial equivalence of the Masimo W1.

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).