(239 days)
The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
The VenusX Radiotherapy System is a medical linear accelerator that delivers therapeutic radiation to patient in accordance with the physician's prescription. It supports CRT /IMRT Treatment Techniques (Mode). The system consists of a photon therapeutic 6 MV X-ray radiation beam producing component with a photon diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.
The provided text describes a medical device called VenusX, a radiotherapy delivery system, and its submission for FDA clearance. However, it does not include detailed information about specific acceptance criteria related to its performance in terms of accuracy or clinical effectiveness as one would expect for a diagnostic AI device or a device with a direct measurable output related to patient diagnosis/treatment efficacy (e.g., sensitivity, specificity, accuracy).
Instead, the document focuses on demonstrating substantial equivalence to predicate devices (Varian UNIQUE and Varian On Board Imaging) by comparing technological characteristics and ensuring adherence to various safety, electrical, software, and biocompatibility standards.
Therefore, many of the requested points cannot be answered based on the provided text. I will answer the points that can be addressed from the document and explicitly state when information is not available.
1. A table of acceptance criteria and the reported device performance
The document primarily discusses compliance with various standards and safety requirements rather than performance metrics like sensitivity, specificity, or quantifiable clinical outcomes. Therefore, a table listing "acceptance criteria" and "reported device performance" in the typical sense for a clinical study is not possible from this text.
The "acceptance criteria" mentioned are related to:
- Electrical Safety and EMC: Compliance with IEC 60601-1-2:2014+AMD1:2020. Reported performance: "The system complies with the following standards."
- Bench Testing: Conformance to various IEC and ANSI AAMI standards (e.g., ANSI AAMIES/IEC60601-1, IEC 60601-1-3, IEC 60601-2-1, IEC 60601-2-68, IEC 62274, IEC 62366-1, IEC 60601-1-6, IEC 60976, IEC 61217). Reported performance: "Test results met all the pre-determined acceptance criteria."
- Cybersecurity: Compliance with security requirement testing, threat mitigation testing, closed box vulnerability scanning, and penetration testing, with reference to IEC 62443-4-1 Section 9.4. Reported performance: "The results of the cybersecurity testing showed that VenusX meets the cybersecurity requirements under Section524B(b) of FD&C Act."
- Hardware and Software V&V: Conformance to 21 CFR §820, ISO 13485, ISO 14971, and FDA guidance "Content of Premarket Submissions for Device Software Functions." Reported performance: "Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly." Also, "there were no software DRs (Discrepancy Reports) remaining which had a priority of Safety Intolerable or Customer Intolerable." The software is considered "enhanced" level of concern.
- Biocompatibility: Compliance with ISO 10993-1, ISO 10993-5, ISO 10993-10 for the Carbon fiber couch top. Reported performance: "Test results met all the pre-determined acceptance criteria."
The document states that the VenusX device "provides precision radiotherapy" and its "Cone Beam CT Imager device (CBCT) is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks." It does not provide quantitative performance metrics for these functions.
A comparison of technological characteristics between the VenusX and predicate/reference devices is provided (pages 5-8), indicating that the VenusX has similar intended use, indications for use, and core functionalities (e.g., photon energy, maximum treatment field size, patient support, microwave source, interface for external system gating, image registration, data interface, record treatment delivery results, treatment techniques). Some differences are noted, such as FFF mode dose rates, SAD, number of MLC leaves (101 leaves in two orthogonal layers vs. 120), maximum leaf over travel, average leaf transmission (≤ 0.05% for VenusX vs. ≤ 2.0% for predicate), MV Detector size (larger for VenusX: 43cm x 43cm vs. 30cm x 40cm), and MV Pixel (higher resolution for VenusX: 2816x2816 vs. 1024x768). For CBCT, the VenusX also uses the FDK algorithm, with a dose per acquisition ≤ 2cGy (vs. ≤ 1.4cGy for reference device) and a 512x512 reconstruction matrix / 360° acquisition mode.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The testing described is primarily non-clinical (bench testing, electrical safety, cybersecurity, hardware/software V&V, biocompatibility). There is no mention of a clinical "test set" in the context of patient data, nor its provenance or design.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. Since no clinical "test set" with patient data requiring ground truth by experts is described, this point is not applicable.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document. Since no clinical "test set" requiring adjudication is described, this point is not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not provided in the document. The VenusX is a radiotherapy delivery system, not a diagnostic AI system designed to assist human readers in interpreting images or data. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to the described device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document refers to the device's technical performance against engineering and safety standards. It does not describe an "algorithm only" standalone performance in the sense of a diagnostic or analytical algorithm reporting a clinical outcome. The device itself (a linear accelerator) performs the function of radiation delivery, and its imaging components (MV, CBCT) are for positional verification.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical testing, the "ground truth" implicitly refers to the specified technical requirements, standards, and design specifications. For example, in electrical safety, the ground truth is adherence to the IEC 60601-1-2 standard. For hardware/software V&V, it's conformance to requirements specifications and hazard safeguards. There is no mention of ground truth established from clinical data like pathology or outcomes data.
8. The sample size for the training set
This information is not provided in the document. The document describes a medical linear accelerator system, not an AI model that requires a "training set" of data in the typical sense. The software and hardware development involves verification and validation, but not a data-driven training process in the way a machine learning model would.
9. How the ground truth for the training set was established
This information is not provided in the document. As noted in point 8, there is no mention of a "training set" for an AI model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2024
LinaTech LLC Jonathan Yao President 1294 Kifer Road #705 Sunnyvale, California 94086
Re: K232489
Trade/Device Name: VenusX Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: March 12, 2024 Received: March 12, 2024
Dear Jonathan Yao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K232489
Device Name VenusX
Indications for Use (Describe)
The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services | |
| Food and Drug Administration | |
| Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff | |
| PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." |
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
510(k) SUMMARY - K232489
As required by 21 CFR 807.92
SUBMITTER l.
| Submitter's Name: | LinaTech1294 Kifer Road #705Sunnyvale, CA 94086, USA |
|---|---|
| Primary ContactPerson: | Jonathan YaoPhone: (408) 733 2051Fax: (408) 733 2045E-mail: Jonathanyao@LinaTech.com |
| Date Prepared: | 11 April 2024 |
II. DEVICE
lll.
| Proprietary Name: | VenusX |
|---|---|
| Common/Usual Name: | Medical Linear Accelerator |
| Classification Name: | Medical charged-particle radiation therapy system |
| Regulation: | 21 CFR §892.5050 |
| Regulatory Class: | Class II |
| Product Code: | IYE |
| PREDICATE DEVICE | |
| Predicate Devices: | Varian UNIQUE (K101751) |
| Reference Device: | Varian On Board Imaging (OBI) (K042720) |
IV. DEVICE DESCRIPTION
Device Description: The VenusX Radiotherapy System is a medical linear accelerator that delivers therapeutic radiation to patient in accordance with the physician's prescription. It supports CRT /IMRT Treatment Techniques (Mode). The system consists of a photon therapeutic 6 MV X-ray radiation beam producing component with a photon diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located
outside the treatment room.
{4}------------------------------------------------
V. INTENDED USE AND INDICATIONS FOR USE
| Intended Use: | The intended use is the same with precision radiotherapy as the predicate device. VenusX does not support stereotactic radiosurgery.The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
|---|---|
| Indications for Use: | The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
VI. TECHNOLOGICAL CHARACTERISTICS
| SummaryOf TechnologicalCharacteristics: | Both the VenusX and the predicate device contain similartechnological characteristics and functional scientific technologyto deliver radiation therapy by authorized medical practitioners.A subset of technological characteristics and features of thecurrent device is different from the predicate. Thebiocompatibility of patient-contacting components remains thesame as the predicate device. The results of the verification,validation and safety standards testing demonstrate that thereare no changes to the safety profile of the device. |
|---|---|
| ------------------------------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
The feature comparison chart below notes the similarities and differences between predicate, reference, and subject device.
{5}------------------------------------------------
The feature comparison with predicate device:
| Attribute/ Feature | Predicate DeviceVarian UNIQUEK101751 | Subject DeviceLinaTech VenusX |
|---|---|---|
| Intended Use | The UNIQUE is intended toprovide stereotacticradiosurgery and precisionradiotherapy for lesions,tumors and conditionsanywhere in the body whenradiation treatment isindicated. | The VenusX radiotherapydelivery system is intendedto provide precisionradiotherapy for lesions,tumors, and conditionsanywhere in the body whereradiation treatment isindicated. |
| Indications for Use | The UNIQUE is indicated forstereotactic radiosurgery andprecision radiotherapy forlesions, tumors andconditions anywhere in thebody when radiationtreatment is indicated. | The VenusX radiotherapydelivery system is intendedto provide precisionradiotherapy for lesions,tumors, and conditionsanywhere in the body whereradiation treatment isindicated. |
| Photon Energy (MV)available | 6 | 6x (FFF) |
| Dose Rates | 100 - 600 cGy/minfor FF mode | 600、800、1000 cGy/minfor FFF mode |
| SAD (cm) | 100 | 90 |
| Maximum standardtreatment field size | 40cmx40cm | 40cmx40cm |
| Number of MLC leaves | 120 | 101 leaves in each of twoorthogonal layers |
| Maximum leaf overtravel distance (cm) | 14 | Upper layer 20Lower layer 14 |
| Average leaftransmission (%) | ≤ 2.0 | ≤ 0.05 |
| Leaf end physicalposition accuracy (cm) | ≤ ± 0.1 | ≤ ± 0.1 |
| Attribute/ Feature | Predicate DeviceVarian UNIQUEK101751 | Subject DeviceLinaTech VenusX |
| Maximum leaf speed(cm/sec) is greater than2.5 | Yes | Yes |
| ElectronTreatmentMode | No | No |
| Non-arc photontreatment modeprogrammable MUmaximum dose is largerthan 1999 MU | Yes | Yes |
| Patient support surface:couch pedestal withcarbon fiber couch topfor improved MVimaging | Yes | Yes |
| Microwave Source isMagnetron | Yes | Yes |
| Interface for ExternalSystem Gating | Yes | Yes |
| MV Detector size | 30cm x 40cm | 43cm x 43cm |
| MV Pixel | 1024 x 768 | 2816×2816 |
| Radiation dose for MVImage is less than 1MU | Yes | Yes |
| Image registration withplanning image | Yes | Yes |
| Basis of imageregistration are forminimum of | soft tissue, bony anatomy,fiducial markers, digitalrepresentation of treatmentapertures. | soft tissue, bony anatomy,fiducial markers, digitalrepresentation of treatmentapertures. |
| Data Interface is usingDICOM RT/3.0 data andwith image | Yes | Yes |
| Attribute/ Feature | Predicate DeviceVarian UNIQUEK101751 | Subject DeviceLinaTech VenusX |
| Record treatmentdelivery results | Yes | Yes |
| Treatment Techniques(Mode) | CRT, IMRT | CRT, IMRT |
| Compatible TPS and OISfor the VenusX andcompare with predicate. | UNIQUE is compatible with acommercially availableindependent third-partyRadiation TreatmentPlanning System (TPS), Suchas Eclips TPS, and also tocompatible with the ARIARadiation TherapyManagement System (FDANo.: K173838) (OIS). | VenusX is compatible with acommercially availableindependent third-partycompatible with TiGRT TPS(FDA No.: K090893)( TPS ) and also ARIA(FDA No.: K173838) (OIS)system. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The feature comparison with reference device:
| Attribute / Feature | Reference DeviceVarian OBI K042720 | Subject DeviceLinaTech VenusX |
|---|---|---|
| Intended Use | The Varian On-BoardImager device (OBI) is usedfor verification of correctpatient position in relationto isocenter andverification of thetreatment fields in relationto anatomical and/orfiducial landmarks. | The VenusX's Cone Beam CTImager device (CBCT) isused for verification ofcorrect patient position inrelation to isocenter andverification of thetreatment fields in relationto anatomical and/orfiducial landmarks. |
| kV imaging | Yes | Yes |
| Cone-beam CT imageReconstruction | Yes | Yes |
| General type ofReconstructionmethod/algorithm | FDK (Feldkamp-Davis-Kress) algorithm | FDK (Feldkamp-Davis-Kress)algorithm |
| Dose per CBCTacquisition | ≤ 1.4cGy | ≤ 2cGy |
| CBCT acquisition mode(pixels / degrees) | 512×512reconstructionmatrix / 200°(head) or360° (body) | 512×512reconstructionmatrix / 360° |
{9}------------------------------------------------
VII. SUMMARY OF PERFORMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination.
Electrical Safety and Electromagnetic Compatibility (EMC)
Electromagnetic compatibility (EMC) testing was conducted on the subject device. The system complies with the following standards:
- IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Bench Test
The VenusX was thoroughly tested on bench to evaluate and verify that it meets the required performance specifications. The bench testing plan was developed based on the risk assessment of the device, and the recommendations outlined in the applicable FDA guidance documents, ISO and IEC standards. VenusX conforms to the following standards:
- ANSI AAMIES/IEC60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical eguipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-1:2020 Medical electrical equipment - Part 2-1: Requirements for the basic safety and essential performance of electron accelerators in the range 1 MeV to 50 MeV
- IEC 60601-2-68:2014 Medical electrical equipment - Part 2-68: Requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment
- IEC 62274:2005 Medical electrical equipment— Safety of radiotherapy record and verify systems
- IEC 62366-1:2015+AMD1:2020 Medical devices - Part 1: Application of usability engineering to medical devices
- IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60976:2007 Medical electrical equipment - Medical electron accelerators - Functional performance characteristics
- IEC 61217:2011 Radiotherapy equipment - Coordinates, movements, and scales
Test results met all the pre-determined acceptance criteria.
{10}------------------------------------------------
Cybersecurity
In addition to the bench testing, the VenusX was also thoroughly tested to verify that its cybersecurity objectives are fully met, and security risk control measures are effective. Specifically, the following cybersecurity testing was conducted:
— Security requirement testing that demonstrates requirement was implemented correctly.
— Threat mitigation testing demonstrating effective risk control measures according to the threat modeling.
— A closed box testing of known vulnerability scanning by qualified external testers, with reference to Section 9.4 of "IEC 62443-4-1 Security for industrial automation and control systems Part 4-1: Secure product development lifecycle requirements ".
- Penetration testing by qualified external testers.
The results of the cybersecurity testing showed that VenusX meets the cybersecurity requirements under Section524B(b) of FD&C Act.
Hardware and Software Verification and Validation Testing
Hardware and software verification and validation process were conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards.
Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions," issued June 2023. The software for this device is considered as a "enhanced" level of concern.
Biocompatibility Testing
Biocompatibility testing was conducted on the Carbon fiber couch top in accordance with ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The Biocompatibility testing conducted for the VenusX included the following:
- ISO 10993-5 Cytotoxicity Test
- ISO 10993-10 Skin Sensitization Test
- ISO 10993-10 Skin Irritation Test
Test results met all the pre-determined acceptance criteria.
CONCLUSION OF NON-CLINICAL TESTING VIII.
The results of testing showed that VenusX conforms to the defined user needs and intended uses and that there were no software DRs (Discrepancy Reports) remaining which had a priority of Safety Intolerable or Customer Intolerable.
{11}------------------------------------------------
IX. CONCLUSION FOR SUBSTANTIAL EQUIVALENCE TO THE PREDICATE AND REFERENCE DEVICE
The VenusX is substantially equivalent to the UNIQUE (K101751) (predicate device). The intended use is the same with precision radiotherapy as the predicate device. VenusX does not support stereotactic radiosurgery. A subset of technological characteristics and features of the current device is different to the predicate. However, the Verification and Validation demonstrate that the device is as safe and effective as the predicate and reference device. Based on the comparison and analysis above, the VenusX (subject device) is substantially equivalent to the UNIQUE (K101751) (predicate device).
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.