K Number
K232416
Manufacturer
Date Cleared
2024-05-03

(266 days)

Product Code
Regulation Number
864.5200
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AUTION EYE AI-4510 Urine Particle Analysis System is a fully automated urine particle analyzer for in vitro diagnostic use. AUTION EYE AI-4510 is intended for the quantitative measurement of red blood cells (WBC) and squamous epithelial cells (SQEC), the semi-quantitative measurement of bacteria (BACT) and crystals (CRYS) and the qualitative measurement of white blood cell clumps (WBCC), non-squamous epithelial cells (NSE), hyaline casts (HYAL), non-hyaline casts (NHC), yeast (YST), mucus (MUCS) and sperm (SPRM) in urine samples.

A trained operator can set criteria for flagging speciment analyte image decisions should be reviewed and reclassified as necessary by a trained technologist.

The AUTION EYE AI-4510 analyzer can be used as a standalone unit or combined with an AUTION MAX AX-4060 urine chemistry analyzer.

Device Description

The AI-4510 System (AUTION EYE AI-4510) is a fully automated urine particle analyzer for in vitro diagnostic use that uses flow cell digital imaging technology in a clinical laboratory setting. Based on images captured in the flow method, the instrument automatically classifies the images of various formed elements. The AI-4510 System can quantitatively measure RBC, WBC, and SQEC; semi-quantitatively measure BACT, and CRYS; and qualitatively measure WBCC, NSE, HYAL, NHC, YST, MUCS and SPRM in urine samples. In addition, the AI-4510 System allows trained operators to manually review and reclassify all the element images collected by the system.

AI/ML Overview

This document describes the validation of the AUTION EYE AI-4510 Urine Particle Analysis System. The device is intended for the quantitative, semi-quantitative, and qualitative measurement of various elements in urine samples. The validation primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device (iQ200 Urine Analyzer).

Here's a breakdown of the requested information based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list "acceptance criteria" in a single table for all performance measures. Instead, it states that "all results meeting the predefined acceptance criteria" for precision studies, and that "Quantitative, Semiquantitative and Qualitative parameters met the acceptance criteria" for the method comparison study. The reported performance is presented in various tables throughout the "Summary of Performance Data" section (5.6).

Here's an aggregated table derived from the provided performance data:

Type of MeasurementElementCriteria/MetricReported Performance
Quantitative RepeatabilityRBC%CVLow: 16.2%, MDL: 7.7%, Mid: 3.5%, High: 3.6%
WBC%CVLow: 17.2%, MDL: 12.5%, Mid: 1.8%, High: 1.9%
SQEC%CVLow: 14.7%, MDL: 9.4%, Mid: 4.8%, High: 4.3%
Semi-Quantitative & Qualitative RepeatabilityBACT% Agreement with expected rankLevel 1-4: 100.0%
CRYS% Agreement with expected rankLevel 1-5: 100.0%
NSE% Agreement with expected rankNegative: 100.0%, Low positive: 80.0%, High Positive: 100.0%
HYAL% Agreement with expected rankNegative: 100.0%, Low positive: 100.0%, High Positive: 100.0%
NHC% Agreement with expected rankNegative: 100.0%, Low positive: 100.0%, High Positive: 100.0%
WBCC% Agreement with expected rankNegative: 100.0%, Low positive: 100.0%, High Positive: 100.0%
YST% Agreement with expected rankNegative: 100.0%, Low positive: 70.0%, High Positive: 100.0%
MUCS% Agreement with expected rankNegative: 100.0%, Low positive: 100.0%, High Positive: 100.0%
SPRM% Agreement with expected rankNegative: 100.0%, Low positive: 100.0%, High Positive: 100.0%
Quantitative Within-Laboratory PrecisionRBCWithin Laboratory %CVLow: 25.1%, MDL: 11.9%, High: 6.2%
WBCWithin Laboratory %CVLow: 26.4%, MDL: 12.7%, High: 6.2%
SQECWithin Laboratory %CVLow: 17.0%, MDL: 8.5%, High: 4.4%
Quantitative Reproducibility (All Sites Combined)RBCReproducibility %CVLow: 25.5%, Mid: 10.9%, High: 12.6%
WBCReproducibility %CVLow: 26.5%, Mid: 8.6%, High: 7.7%
SQECReproducibility %CVLow: 15.1%, High: 11.9%
LinearityRBCLinear Range5-1,000 RBC/μL
WBCLinear Range5-1,000 WBC/μL
SQECLinear Range5-180 SQEC/μL
Limit of DetectionRBCLoB, LoD, LoQLoB: 0.0, LoD: 2.3, LoQ: 2.3
WBCLoB, LoD, LoQLoB: 0.0, LoD: 1.5, LoQ: 1.5
SQECLoB, LoD, LoQLoB: 0.2, LoD: 1.6, LoQ: 1.6
CRYSLoB, LoD, LoQLoB: 0.0, LoD: 6.4, LoQ: 6.4
BACTLoB, LoD, LoQLoB: 0.0, LoD: 6.0, LoQ: 6.0
CarryoverAll 12 elementsPresence of carryoverNo carryover effect detected.
InterferenceVarious elements & interferentsConcentration limit with no significant interferenceSee Tables 10, 11, and 12 for specific concentrations.
Sample StabilityAll 12 elementsStability durationRoom temperature (15-30°C) for up to 2 hours; Refrigeration (2-8°C) for up to 6 hours.
Method Comparison (AI-4510 (Manual) vs. iQ200 (Manual) / Manual Microscopy)RBCWeighted Deming R2, Intercept, SlopeR2: 0.918, Intercept: 1.501, Slope: 0.824
WBCWeighted Deming R2, Intercept, SlopeR2: 0.903, Intercept: 0.629, Slope: 0.968
SQECWeighted Deming R2, Intercept, SlopeR2: 0.928, Intercept: -0.451, Slope: 0.933
CRYS (vs. Manual Microscopy)Sensitivity, SpecificitySensitivity: 76.2%, Specificity: 83.7%
BACT (vs. Manual Microscopy)PPA, NPAPPA: 90.5%, NPA: 98.2%
NSE (vs. iQ200)PPA, NPAPPA: 88.7%, NPA: 84.3%
NHC (vs. iQ200)PPA, NPAPPA: 80.2%, NPA: 83.8%
HYAL (vs. iQ200)PPA, NPAPPA: 85.0%, NPA: 89.0%
YST (vs. iQ200)PPA, NPAPPA: 97.1%, NPA: 99.6%
WBCC (vs. iQ200)PPA, NPAPPA: 86.5%, NPA: 89.3%
MUCS (vs. iQ200)PPA, NPAPPA: 81.9%, NPA: 88.0%
SPRM (vs. iQ200)PPA, NPAPPA: 86.2%, NPA: 99.6%

2. Sample size used for the test set and the data provenance

  • Precision Studies:
    • Repeatability Study: Clinical urine samples were used.
      • Quantitative elements: Not explicitly stated, but "clinical urine samples in the evaluation of repeatability for all twelve (12) elements from low to high concentrations."
      • Semi-quantitative and Qualitative elements: n=10 replicates per test level (e.g., Level 1, 2, 3 etc.).
    • Within-Laboratory Precision Study: ARKRAY control materials prepared using clinical samples.
    • Reproducibility Study: Commercially available control materials and ARKRAY control materials prepared using clinical samples were used.
  • Linearity Testing: Not specified for sample size beyond "one instrument."
  • Limit of Detection: Not specified for sample size.
  • Carryover Testing: High-level and low-level samples, aliquoted into 5 tubes each, measured in sequences (e.g., H1 L1 H2 L2 H3 L3 H4 L4 H5 L5, repeated 5 times).
  • Interference Testing: Not specified for sample size beyond the substances tested.
  • Sample Stability: Positive and negative samples for all 12 elements.
  • Method Comparison:
    • Population for Reference Range results: n=247
    • Quantitative Elements (RBC, WBC, SQEC): n=377 (RBC), n=845 (WBC), n=382 (SQEC) for comparison between AI-4510 (M) and iQ200 (M).
    • Semi-quantitative & Qualitative Elements: n=1474 (CRYS, BACT, WBCC, MUCS, SPRM), n=765 (NSE, NHC, HYAL, YST).
    • Data Provenance: Clinical samples. The method comparison study was conducted at "three (3) CLIA-Moderate complexity laboratories." The document states samples were "collected fresh within two (2) hours or refrigerated up to six (6) hours post collection," implying a prospective collection directly for these studies. The country of origin is not explicitly stated, but the submission is for FDA clearance in the US, and the company and testing sites (CLIA labs) suggest operations relevant to the US market.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for the method comparison study (especially for semi-quantitative and qualitative elements) appears to be established by comparison to the predicate device iQ200 (Manual) and Manual Microscopy.

  • For the method comparison, it refers to "CLIA-trained operators" performing testing and "A trained operator can set criteria for flagging speciment analyte image decisions should be reviewed and reclassified as necessary by a trained technologist." This indicates that trained technologists or CLIA-trained operators (which implies suitable qualifications for laboratory testing) established the "ground truth" or reference values, either by manual microscopy or using the predicate device's manual review function.
  • The document does not specify the exact number of individual experts or their specific qualifications (e.g., specific years of experience, board certification as pathologists or medical technologists). It only refers to "CLIA-trained operators" and "trained technologists."

4. Adjudication method for the test set

  • The document states: "All instrument analyte image decisions should be reviewed and reclassified as necessary by a trained technologist." This implies a form of human override or adjudication post-AI classification.
  • For the "Method Comparison" tables (14, 15, 16), most comparisons are listed as "AI-4510 (Manual) vs. iQ200 (Manual)" or "AI-4510 (Manual) vs. Manual Microscopy." The "(M)" denotes "manually reviewed and reclassified results." This indicates that the results from both the investigational device and the predicate device/manual microscopy were subjected to manual review/adjudication by trained human operators to establish the final classification used for comparison.
  • The specific method of adjudication (e.g., 2+1, 3+1 consensus) among multiple readers for establishing the ground truth is not specified. The comparison is against already "manual" classifications from the predicate or direct manual microscopy, suggesting that the human reading itself serves as the reference, likely by one or more trained technologists.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This document describes performance characteristics of the device itself, often compared to a predicate device or manual microscopy.
  • It does not describe an MRMC comparative effectiveness study where the performance of human readers with AI assistance is directly compared to human readers without AI assistance to quantify improvement or effect size. The AI-4510 System is an automated analyzer with a manual review component, not an AI assistance tool for human interpretation of images outside of the system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The document mentions "The software processes the recorded images, automatically identifying and classifying the formed elements based on the sorting algorithm." (Section 5.5)
  • However, the indications for use explicitly state: "A trained operator can set criteria for flagging specimens. All instrument analyte image decisions should be reviewed and reclassified as necessary by a trained technologist."
  • Furthermore, the "Method Comparison" tables are predominantly listed as "AI-4510 (Manual) vs. iQ200 (Manual)" or "AI-4510 (Manual) vs. Manual Microscopy," where "(M)" denotes "manually reviewed and reclassified results."
  • This strongly suggests that the reported performance data for clinical claims (method comparison) represents the combined human-in-the-loop performance after technologist review and reclassification, particularly for the semi-quantitative and qualitative elements.
  • While the device has an "automatic classification" function (also mentioned in section 5.4 under "Automatic Classification"), the reported clinical performance data does not appear to be purely standalone (algorithm-only) without human intervention.
  • Table 11 (Interference Effect on Auto-classified Results) hints at some testing of the auto-classified performance in specific scenarios (interference), but the bulk of the clinical validation on the main intended use appears to involve human review.

7. The type of ground truth used

The ground truth used for the method comparison study was established through:

  • Comparison to the iQ200 System (Manual): This means the results obtained from the predicate device after its own manual review and reclassification process.
  • Manual Microscopy: This is considered the traditional gold standard for urine particle analysis, established by trained technologists.

Therefore, the ground truth is a combination of expert consensus (implied via "trained technologist" review) and comparison to a legally marketed predicate device (also with human review), with manual microscopy serving as a reference.

8. The sample size for the training set

The document provided does not contain any information about the training set for the AI-4510 System's algorithm. This K submission focuses on device performance studies for validation and comparison to a predicate, not on the developmental aspects of the AI model.

9. How the ground truth for the training set was established

As no information about the training set is provided, how its ground truth was established is also not available in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

May 3, 2024

Arkray Inc. % Daya Ranamukhaarachi VP/Science and Regulatory Affairs ARKRAY, Inc. 5198 West 76th Street Minneapolis, Minnesota 55439

Re: K232416

Trade/Device Name: AUTION EYE AI-4510 Urine Particle Analysis System Regulation Number: 21 CFR 864.5200 Regulation Name: Automated cell counter Regulatory Class: Class II Product Code: LKM Dated: April 5, 2024 Received: April 5, 2024

Dear Daya Ranamukhaarachi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu-S

Min Wu Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232416

Device Name

AUTION EYE AI-4510 Urine Particle Analysis System

Indications for Use (Describe)

AUTION EYE AI-4510 Urine Particle Analysis System is a fully automated urine particle analyzer for in vitro diagnostic use. AUTION EYE AI-4510 is intended for the quantitative measurement of red blood cells (WBC) and squamous epithelial cells (SQEC), the semi-quantitative measurement of bacteria (BACT) and crystals (CRYS) and the qualitative measurement of white blood cell clumps (WBCC), non-squamous epithelial cells (NSE), hyaline casts (HYAL), non-hyaline casts (NHC), yeast (YST), mucus (MUCS) and sperm (SPRM) in urine samples.

A trained operator can set criteria for flagging speciment analyte image decisions should be reviewed and reclassified as necessary by a trained technologist.

The AUTION EYE AI-4510 analyzer can be used as a standalone unit or combined with an AUTION MAX AX-4060 urine chemistry analyzer.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This 510(k) Summary of the AUTION EYE AI-4510 Urine Particle Analysis System is submitted in compliance with 21 CFR807.92 for the purposes of safety and effectiveness.

Date Prepared:April 03, 2024
Establishment:ARKRAY Factory Inc.1480 Koji, Konan-cho,Koka-shiShiga Japan, 520-3306
Submitter:ARKRAY Inc.Yousuien-nai, 59 Gansuin-cho,Kamigyo-kuKyoto Japan, 602-0008
Official US Agent:Daya Ranamukhaarachchi, PhDARKRAY Factory USA Inc.Title: VP/Science and Regulatory AffairsPhone (408) 234-0804Email: ranamukd@arkrayusa.com
Primary Contact:Ryan HansonARKRAY Factory USA Inc.Title: Science Affairs ManagerPhone (952) 646-3121Email: hansonr@arkrayusa.com
Secondary Contact:Marissa BrooksARKRAY Factory USA Inc.Title: Regulatory Affairs Project ManagerPhone (760) 525-8405Email: brooksm@arkrayusa.com

Device Name:

Trade Name: AUTION EYE AI-4510 Urine Particle Analysis System

Common Name: Automated cell counter (Urine particle counter)

{5}------------------------------------------------

Predicate Device:

Trade Name:iQ200 Urine Analyzer (K093861)
Common Name:Automated Urinalysis System

5.1 Regulatory Information:

Regulatory Classification for Automated cell counter (Urine particle counter) and each of the tests are listed in Table 1.

Table 1. Regulatory Classification for the AI-4510 System

TestDescriptionProduct CodeDevice ClassRegulationDevice Panel
Urine ParticleCounterLKMClass II21 CFR864.520081- Hematology
AutomatedUrinalysisSystemKQOClass I21 CFR862.290075- Chemistry

5.2 Intended Use

AUTION EYE AI-4510 Urine Particle Analysis System is a fully automated urine particle analyzer for in vitro diagnostic use. AUTION EYE AI-4510 is intended for the quantitative measurement of red blood cells (RBC), white blood cells (WBC) and squamous epithelial cells (SQEC), the semi-quantitative measurement of bacteria (BACT) and crystals (CRYS) and the qualitative measurement of white blood cell clumps (WBCC), non-squamous epithelial cells (NSE), hyaline casts (HY AL), non-hyaline casts (NHC), yeast (YST), mucus (MUCS) and sperm (SPRM) in urine samples.

A trained operator can set criteria for flagging specimens. All instrument analyte image decisions should be reviewed and reclassified as necessary by a trained technologist.

The AUTION EYE AI-4510 analyzer can be used as a standalone unit or combined with an AUTION MAX AX-4060 urine chemistry analyzer.

5.3 Device Description

The AI-4510 System (AUTION EYE AI-4510) is a fully automated urine particle analyzer for in vitro diagnostic use that uses flow cell digital imaging technology in a clinical laboratory setting. Based on images captured in the flow method, the instrument automatically classifies the images of various formed elements. The AI-4510 System can

{6}------------------------------------------------

quantitatively measure RBC, WBC, and SQEC; semi-quantitatively measure BACT, and CRYS; and qualitatively measure WBCC, NSE, HYAL, NHC, YST, MUCS and SPRM in urine samples. In addition, the AI-4510 System allows trained operators to manually review and reclassify all the element images collected by the system.

5.4 Substantial Equivalence

The following Table 2 shows the comparisons between the predicate and candidate devices and identifies the major technological and performance characteristics.

COMPONENT/CHARACTERISTICPROPOSEDPREDICATE
510(K) NumberK232416K093861
DeviceAUTION EYE AI-4510 UrineParticle Analysis SystemiQ200 System with LaminaCradle
Device TypeAutomated Cell CounterAutomated Cell Counter
FDA Product CodeLKM, KQOLKM, KQO
FDA Authorization UseLaboratory UseLaboratory Use
Intended Use andIndication for UseAUTION EYE AI-4510 UrineParticle Analysis System is afully automated urine particleanalyzer for in vitro diagnosticuse. AUTION EYE AI-4510 isintended for the quantitativemeasurement of red blood cells(RBC), white blood cells(WBC) and squamousepithelial cells (SQEC), thesemi-quantitative measurementof bacteria (BACT) and crystals(CRYS) and the qualitativemeasurement of white bloodcell clumps (WBCC), non-squamous epithelial cells(NSE), hyaline casts (HYAL),non-hyaline casts (NHC), yeastThe iQ®200 System is an in-vitro diagnostic device used toautomate the completeurinalysis profile, includingurine test strip chemistry paneland microscopic sedimentanalysis.
(YST), mucus (MUCS) andsperm (SPRM) in urinesamples.A trained operator can setcriteria for flagging specimens.All instrument analyte imagedecisions should be reviewedand reclassified as necessary bya trained technologist.The AUTION EYE AI-4510analyzer can be used as astandalone unit or combinedwith an AUTION MAX AX-4060 urine chemistry analyzer.Optionally, the iQ200 analyzercan be used as a stand-alone unit,or the results from the iQ200analyzer can be combined withother urine chemistry resultsreceived from an LIS. Itproduces quantitative orqualitative counts of all formedsediment elements present inurine, including cells, casts,crystals, and organisms. Acompetent human operator canset criteria for auto-reporting andflagging specimens for review.All instrument analyte imagedecisions may be reviewed andoverridden by a trainedtechnologist.
Measurement PrincipleSame as predicateFlow digital imaging
ParametersRed blood cells (RBC), Whiteblood cells (WBC), White bloodcell clumps (WBCC), Squamousepithelial cells (SQEC), Non-squamous epithelial cells (NSE),Hyaline casts (HYAL), non-hyaline casts (NHC ), Bacteria(BACT), Crystals (CRYS), Yeast(YST),Mucus (MUCS) andSperm (SPRM)Red blood cells (RBC), Whiteblood cells (WBC), White bloodcell clumps (WBCC), Squamousepithelial cells (SQEC), Non-squamous epithelial cells (NSE),Hyaline casts (HYAL),Unclassified casts (NHC orUNCC),Bacteria (BACT), Crystals(CRYS), Yeast (YST), Mucous(MUCS),Sperm (SPRM),Dysmorphic RBC andUnclassified Crystals
Automatic ClassificationRed blood cells (RBC), whiteblood cells (WBC), white bloodcell clumps (WBCC),squamous epithelial cells(SQEC), non-squamous epithelialcells (NSE), hyaline casts(HYAL), non-hyaline casts(NHC), bacteria (BACT), crystal(CRYS), yeast (YST), mucus(MUCS) and sperm (SPRM)Red Blood Cells (RBC), WhiteBlood Cells (WBC), WhiteBlood Cell Clumps (WBCC),Squamous Epithelial Cells(SQEC), Non
(While a dysmorphic red bloodcell is counted as a red blood cell,the flag can be automaticallyassigned for the dysmorphic redblood cell.)Squamous Epithelial Cells(NSE), Bacteria (BACT),Crystals (UNCX), Hyaline Cast(HYAL), Unclassified Cast(UNCC), Yeast (BYST/HYST),Sperm(SPRM), Mucus (MUCS)
Manual ClassificationRed blood cells: 2 itemsDysmorphic red blood cells(DRBC) and red blood cellclumps (RBCC)Red blood cells: 2 itemsDysmorphic red blood cells(DRBC) and red blood cellclumps (RBCC)
Crystal: 10 items Calcium oxalatecrystal (CAOX), calciumphosphate crystal (CAPH),magnesium ammoniumphosphate crystal (TPO4),leucine crystal (LEUC), uric acidcrystal (URIC), calciumcarbonate crystal (CACB),cystine crystal (CYST), tyrosinecrystal (TYRO),amorphous salts (phosphoricacid, uric acid) (AMOR) andunclassified crystal (UNCX)Crystal: 10 items Calciumoxalate crystal (CAOX),calcium phosphate crystal(CAPH), magnesiumammonium phosphate crystal(TPO4), leucine crystal (LEUC),uric acid crystal (URIC),calcium carbonate crystal(CACB), cystine crystal(CYST), tyrosine crystal(TYRO),amorphous salts (phosphoricacid, uric acid) (AMOR) andunclassified crystal (UNCX)
Casts: 6 itemsGranular casts (GRAN), waxycasts (WAXY), RBC casts(RBCT), WBC casts (WBCT),epithelial casts (EPIC) and fattycasts (FATC)Casts: 6 itemsGranular casts (GRAN), waxycasts (WAXY), RBC casts(RBCT), WBC casts (WBCT),epithelial casts (EPIC) and fattycasts (FATC)
Yeast: 1 itemBudding yeast (BYST)Yeast: 1 itemBudding yeast (BYST)
Epithelial cells: 2 items Renaltubular epithelial cells (REEP),transitional epithelial cells(TREP)Epithelial cells: 2 items Renaltubular epithelial cells (REEP),transitional epithelial cells(TREP)
Fat: 2 itemsFat: 2 items
Fat (FAT) and oval fat body(OVFB)Fat (FAT) and oval fat body(OVFB)
Other categories: 5 itemsTrichomonas (TRCH), inclusionOther categories: 5 items
cells (INCC), parasite (PARA),artifact (ART) and unclassified(UNCL)Trichomonas (TRCH), inclusioncells (INCC), parasite (PARA),artifact (ART) and unclassified(UNCL)
Specimen Type /Measurement ObjectUrineUrine
ReagentsAUTION EYE Sheath SolutionConcentrated Washing Solution 3AUTION EYE CalibratoriQ Lamina Iris Diluent IrisSystem CleanseriQ Focus Solution
Quality Control SolutionAUTION EYE Control Solution(It consists of High and Lowlevels.)iQ Positive control,iQ Negative control
Sample ConsumptionVolumeMaximum 1 mLApproximately 1.3 mL
Required Volume2 mL or moreMinimum volume 3 mL
Measurement TimeApproximately 45 seconds/test(Excluding cases in which largenumbers of formed elements arepresent or in the case of controlmeasurement)101 seconds/test
Warm up timeMaximum 10 minutes (Durationafter the power is turned ON. Thewarm- up time is extended if areadjustment process is addedduring startup.)The iQ200 System should beallowed to warm up for 1 or 2hours if it was turned off formore than 6 hours.
Memory CapacityMeasurement results (Normal):10,000 testsOnboard storage of up to 10,000patient results
Number of CapturedImagesThree digital cameras capture1,500 frames per sampleA digital camera captures 500frames per sample
Capturing ImagesTakes images with 3 differentfocal points using 3 digitalcamerasPresents a specimen to amicroscope coupled to a 1.3-megapixel CCD (chargecoupling device) digital camera
Power100 - 240 V AC, 50/60 Hz(Including fluctuations of±10%)Microscopy module - 90-240VAC 50-60 Hz200 watts max
IntendedEnvironmentIndoor use onlyIndoor use only
MeasurementEnvironmentTemperature: 59 to 86 °F,Humidity: 20 to 80% (Nocondensation)Temperature: 15-30°C (59-86°F)Humidity: 20% - 80% non-condensing
MeasurementsourceStrobe lampStrobe lamp
Dimensions530 (W) × 600 (D) × 650 (H) mm(including the sampler)530 (W) × 200 (D) × 135 (H) mm(sampler only)Microscopy module - 56 H x 59W x 64.8 D cm (22 H x 23 W x25.5 D in.)
WeightAnalyzer (with the sampler):Approx. 57 kg Sampler: Approx.4 kgMicroscopy module -45.4 kg (100 lbs.)

Table 2. Comparison Between AUTION EYE AI-4510 and iQ200 SYSTEM

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

5.5 Test Principle and Technology

The principle of operation of the AI-4510 System is based on image capture and the digital processing of images for particle analysis and recognition from a locked sorting algorithm for clustering and classification.

The instrument uses flow digital imaging technology with a flow cell and Complementary Metal- Oxide- Semiconductor (CMOS) image sensors. As the urine passes through the flow cell, it is illuminated by a strobe lamp. Color images are recorded from three different focal points by the CMOS image sensors. The software processes the recorded images, automatically identifying and classifying the formed elements based on the sorting algorithm. The images and the classification of the formed elements are also displayed on the device screen for the operator. The results of the measurements of formed elements for each sample can be centrally controlled by the computer. Images that could not be classified automatically can be classified manually.

5.6 Summary of Performance Data

Clinical and bench testing were conducted to verify the performance characteristics of this device. This testing showed acceptable device performance that is substantially equivalent to the performance of the predicate device.

Precision

Precision for the AI-4510 System was evaluated to estimate repeatability, within-laboratory precision, and reproducibility for quantitative analytes: Red Blood Cell (RBC), White Blood Cell (WBC), and Squamous Epithelial Cells (SQEC), and repeatability for all semi-quantitative and qualitative elements: Bacteria (BACT), Crystals (CRYS), White blood cell clumps (WBCC), Nonsquamous epithelial cells (NSE), Hyaline casts (HYAL), non-hyaline casts (NHC), Yeast (YST), Mucus (MUCS) and Sperm (SPRM).

Precision studies were conducted in accordance with the general guidelines established in CLSI EP05-A3; (Third Edition)- Evaluation of Precision Devices; Approved Guideline

Precision studies were conducted to evaluate measurement imprecision using both controlled materials and clinical samples, with all results meeting the predefined acceptance criteria.

There were three major studies conducted to demonstrate the precision of the AI-4510 System:

{12}------------------------------------------------

Repeatability Study- (Quantitative, Semiquantitative, and Qualitative Elements) Results were collected with one (1) instrument, at one (1) site, with one (1) operator and 10 replicates, using clinical urine samples in the evaluation of repeatability for all twelve (12) elements from low to high concentrations.

Within-Laboratory Precision Study- (Quantitative Elements) Results collected over twenty (20) days with two (2) runs per day and two (2) replicates per run using ARKAY control materials prepared using clinical samples.

Reproducibility Study-(Quantitative Elements) Results collected from three (3) sites, with one (1) analyzer per site, and testing conducted over five (5) days with two (2) runs per day and three (3) replicates per run using commercially available control materials and ARKRAY control materials prepared using clinical samples.

Results from the analysis conducted for each study is presented in Tables 3-6 below.

ElementTest LevelMean Value (/μL)Repeatability
SD%CV
RBCLow7.51.216.2
MDL42.13.27.7
Mid468.816.63.5
High858.730.63.6
WBCLow7.81.317.2
MDL38.54.812.5
Mid425.27.81.8
High918.717.51.9
SQECLow6.10.914.7
MDL39.63.79.4
Mid91.64.44.8
High161.56.94.3

Table 3. Repeatability Results - Quantitative Elements

Table 4. Repeatability Results - Semi-Quantitative and Qualitative Elements

ElementnTest LevelAvg. Conc.(/μL)% Agreement with expectedrank
BACT10Level 10.0100.0%
10Level 255.7100.0%
10Level 378.9100.0%
10Level 4166.5100.0%
CRYS10Level 10.0100.0%
10Level 241.7100.0%
10Level 3118.1100.0%
10Level 4200.7100.0%
10Level 5330.1100.0%

{13}------------------------------------------------

ElementnTest LevelAvg. Conc.(/μL)% Agreement with expectedrank
NSE10Negative0.0100.0%
NSE10Low positive4.880.0%
NSE10High Positive15.5100.0%
HYAL10Negative0.0100.0%
HYAL10Low positive1.2100.0%
HYAL10High Positive8.2100.0%
NHC10Negative0.0100.0%
NHC10Low positive1.2100.0%
NHC10High Positive12.9100.0%
WBCC10Negative0.0100.0%
WBCC10Low positive3.3100.0%
WBCC10High Positive15.5100.0%
YST10Negative0.0100.0%
YST10Low positive11.470.0%
YST10High Positive49.3100.0%
MUCS10Negative0.1100.0%
MUCS10Low positive32.8100.0%
MUCS10High Positive106.2100.0%
SPRM10Negative0.0100.0%
SPRM10Low positive7.4100.0%
SPRM10High Positive32.2100.0%

Table 5. Within-Laboratory Precision Results

Mean ValueRepeatabilityBetween RunBetween DayWithin Laboratory
Level(/μL)SD%CVSD%CVSD%CVSD%CV
RBCLow8.12.024.70.33.70.22.32.025.1
RBCMDL43.14.29.82.96.70.00.05.111.9
RBCHigh515.722.84.418.43.613.22.632.26.2
WBCLow7.91.721.91.214.80.00.02.126.4
WBCMDL42.93.47.94.09.31.53.55.512.7
WBCHigh470.113.42.823.85.110.62.329.36.2
SQECLow8.61.416.10.00.00.55.71.517.0
SQECMDL39.53.17.90.71.71.12.73.48.5
SQECHigh158.06.34.02.01.32.21.47.04.4

{14}------------------------------------------------

LevelMeanValue(cells/µL)RepeatabilityBetween RunBetween DayBetween SiteReproducibility
SD%CVSD%CVSD%CVSD%CVSD%CV
RBC Low8.42.023.40.00.00.78.00.56.22.125.5
RBC Mid86.26.67.61.51.70.91.16.57.59.410.9
RBC High675.753.98.00.00.013.42.064.39.585.012.6
WBC Low5.71.424.50.58.20.34.80.23.21.526.5
WBC Mid86.86.47.42.83.21.72.01.92.27.48.6
WBC High645.020.93.220.43.213.82.137.65.849.67.7
SQEC Low14.92.114.40.00.00.32.00.64.22.315.1
SQEC High87.26.57.42.83.20.00.07.68.710.311.9

Table 6. All Sites Combined - Reproducibility Study Results

Linearity

Linearity testing was performed for the AI-4510 System using one instrument to establish the linear interval for quantitative elements: RBC, WBC, and SQEC. The study was conducted following CLSI EP06-2" Edition and met the predefined acceptance criteria, establishing the linearity interval of the AI-4510 System quantitative elements as shown in Table 7 below.

Table 7. AI-4510 System Linearity Interval
Quantitative ElementLinear Range
RBC5-1,000 RBC/μL
WBC5-1,000 WBC/μL
SQEC5-180 SQEC/μL

7 AI 4510 System Linearity Interval

Limit of Detection

Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were conducted for the AI-4510 System quantitative (RBC, WBC, SQEC) and semiquantitative (CRYS, BACT) elements following CLSI-EP17-A2 Evaluation of Detection Capability For Clinical Laboratory Measurement Procedures; Approved Guideline -Second Edition. Based on the results of the study, as shown below in Table 8, all elements met the predefined acceptance criteria, thereby establishing the LoB, LoD, and LoQ for each element.

Table 8. Results of Detection Limits
Element(cells/µL)LoBLoDLoQ
RBC0.02.32.3
WBC0.01.51.5
SQEC0.21.61.6
Table 8. Results of Detection Limits
--------------------------------------------

{15}------------------------------------------------

Element(cells/µL)LoBLoDLoQ
CRYS0.06.46.4
BACT0.06.06.0

Carryover

Carryover testing was performed for the AI-4510 System to confirm that there are no carryover contaminations for all twelve (12) elements. This testing was conducted using pre-prepared concentrations of high-level and low-level samples which were aliquoted into 5 tubes each and measured alternatively as shown in Table 9 for each tested element. Based on the predefined acceptance criteria, no carryover effect was detected.

Measurement sequenceSample placement in rack
Set 1H1 L1 H2 L2 H3 L3 H4 L4 H5 L5
Set 2H1 L1 H2 L2 H3 L3 H4 L4 H5 L5
Set 3H1 L1 H2 L2 H3 L3 H4 L4 H5 L5
Set 4H1 L1 H2 L2 H3 L3 H4 L4 H5 L5
Set 5H1 L1 H2 L2 H3 L3 H4 L4 H5 L5

H: High level sample, L: Low level sample

Interference

Interference testing was performed on the AI-4510 System to evaluate the influence of interfering substances on the measurement of twelve (12) particle elements using the AI-4510 System. A dose dependency evaluation was performed for those substances which were confirmed to have interference effects on the measurements to determine the highest concentration at which no interference effect is observed.

The results of this interference study are summarized in Tables 10-12 below.

ElementInterferentConcentrationLimit with NoSignificantInterferenceConcentration of Interferent and Observed Effect
RBCTrichomonas50 /μLAt 120 /µL falsely increased for negative samples
Turbidity1+ **At 2+ falsely decreased for positive samples
Bacteria4000 /μLAt 6000 /µL falsely decreased for positive samples
WBCYeast60 /μLAt 80 /µL falsely increased for negative samples andlow positive samples
Trichomonas- *At 10 /µL falsely increased for negative samples
Red Blood Cells1500 /μLAt 5000 /µL falsely increased for positive samples

Table 10. Interference Effect on Manually Reviewed and Reclassified Results

{16}------------------------------------------------

Turbidity1+ **At 2+ falsely decreased for low positive samples
WBCCBacteria7500 /μLAt 15000 /μL falsely decreased for low positive samples
Trichomonas50 /μLAt 120 /μL falsely increased for negative samples
SQECRed Blood Cells1200 /μLAt 1500 /μL falsely decreased for positive samples
Turbidity1+ **At 2+ falsely decreased for positive samples
Bacteria7500 /μLAt 15000 /μL falsely decreased for positive samples
BACTAmorphousUrates120 /μLAt 160 /μL falsely increased for low positive samplesAt 200 /μL falsely increased for negative samples andhigh positive samples
Turbidity1+ **At 2+ falsely increased for low positive samples
Yeast60 /μLAt 80 /μL falsely increased for low positive samples
MUCSRed Blood Cells1500 /μLAt 5000 /μL falsely increased for high positive samples
Bacteria6000 /μLAt 7500 /μL falsely decreased for high positive samples
SPRMBacteria6000 /μLAt 7500 /μL falsely decreased for high positive samples
  • Concentration limit with no significant interference was not evaluated

** The qualitative rank as measured by an AUTION MAX AX-4060 Urine Chemistry Analyze

Table 11. Interference Effect on Auto-classified Results
----------------------------------------------------------------------
Interference effect on Auto-classified results
ElementInterferentConcentration limit with no significant interferenceConcentration of Interferent and observed effect
RBCYeast40 /uLAt 60/uL falsely increased for negative samplesAt 80/uL falsely increased for low positive samples
Crystal-*At 330/uL falsely increased for negative samples
Trichomonas50 /uLAt 120/uL falsely increased for negative samples
Turbidity1+ **At 2+ falsely decreased for positive samples
Bacteria4000/uLAt 6000/uL falsely decreased for positive samples
WBCRed blood cells400 /uLAt 800/uL falsely increased for negative samplesAt 1200/uL falsely increased for low positive samplesAt 1500/uL falsely increased for high positive samples
Yeast40 /uLAt 60/uL falsely increased for negative samplesAt 80/uL falsely increased for low positive samples
Trichomonas-*At 10/uL falsely increased for negative samples andpositive samples
Turbidity1+ **At 2+ falsely decreased for low positive samples
Bacteria7500 /uLAt 15000 /uL falsely decreased for low positivesamples
WBCCRed blood cells1200 /uLAt 5000 /uL falsely increased for negative samples

{17}------------------------------------------------

Yeast60 /uLAt 80 /uL falsely increased for negative samples
Trichomonas20 /uLAt 50 /uL falsely increased for negative samples
Red blood cells1200 /uLAt 1500/uL falsely decreased for positive samples
SQECTurbidity1+ **At 2+ falsely decreased for positive samples
Bacteria7500 /uLAt 15000 /uL falsely decreased for positive samples
NHCYeast60 /uLAt 80 /uL falsely increased for negative samples andlow positive samples
BACTAmorphousurates120 /uLAt 200 /uL falsely increased for negative samples andhigh positive samples
Yeast60 /uLAt 80 /uL falsely increased for low positive samples
Turbidity1+ **At 2+ falsely decreased for low positive samples
YSTRed blood cells1200 /uLAt 5000/uL falsely increased for negative samplesand low positive samples
MUCSRed blood cells1500 /uLAt 5000/uL falsely decreased for high positivesamples
Bacteria6000 /uLAt 7500/uL falsely decreased for high positivesamples
SPRMBacteria6000 /uLAt 7500/uL falsely decreased for high positivesamples
  • Concentration limit with no significant interference not evaluated.

** The qualitative rank as measured by an AUTION MAX AX-4060 urine chemistry analyzer

ElementInterfering SubstancesHighest Concentration Tested with No Interference
RBCViscosity1.24 mPa · s
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Yeast100 /μL
WBCViscosity1.24 mPa · s
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Turbidity2+
WBCCViscosity1.24 mPa · s
Red Blood Cells5000 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Bacteria15000 /μL
Yeast100 /μL
SQECViscosity1.24 mPa · s
Trichomonas120 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Yeast100 /μL
Turbidity2+
Viscosity1.24 mPa · s
NSETrichomonas120 /μL
Red Blood Cells5000 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Bacteria15000 /μL
Yeast100 /μL
Turbidity2+
HYALViscosity1.24 mPa · s
Trichomonas120 /μL
Red Blood Cells5000 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Bacteria15000 /μL
Yeast100 /μL
Turbidity2+
NHCViscosity1.24 mPa · s
Trichomonas120 /μL
Red Blood Cells5000 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Bacteria15000 /μL
Yeast100 /μL
BACTViscosity1.24 mPa · s
Trichomonas120 /μL
Red Blood Cells5000 /μL
MUCS200 /μL
Calcium Oxalate Crystal330 /μL
Turbidity2+
CRYSViscosity1.24 mPa · s
Trichomonas120 /μL
Red Blood Cells5000 /μL
MUCS200 /μL
Amorphous Urates320 /μL
Bacteria15000 /μL
Yeast100 /μL
Turbidity2+
YSTViscosity1.24 mPa · s
Trichomonas120 /μL
Red Blood Cells5000 /μL
MUCSMUCS200 /μL
Amorphous Urates320 /μL
Calcium Oxalate Crystal330 /μL
Bacteria15000 /μL
Turbidity2+
MUCSViscosity1.24 mPa・s
Trichomonas
Amorphous Urates
Calcium Oxalate Crystal
Yeast
Turbidity2+
Viscosity1.24 mPa・s
Trichomonas120 /μL
SPRMRed Blood Cells5000 /μL
MUCS
Amorphous Urates
Calcium Oxalate Crystal
Yeast100 /μL

Table 12. Elements for which No Interference was Observed

{18}------------------------------------------------

{19}------------------------------------------------

Sample Stability

Sample stability was evaluated on the AI-4510 System to confirm the sample stability of twelve (12) elements when measured with the AI-4510 System. Testing was performed on both positive and negative samples for all twelve (12) elements at room temperature (15-30°C) and under refrigeration (2-8°C). In conclusion, all samples tested for each element met the predefined acceptance criteria, therefore establishing the sample stability at room temperature (15-30°C) for up to 2 hours and under refrigeration (2-8°C) for up to 6 hours for all twelve (12) elements measured by the AI-4510 System.

{20}------------------------------------------------

Method Comparison

A method comparison study was conducted at three (3) CLIA-Moderate complexity laboratories to evaluate the agreement of urine sample test results between the candidate device, the AI-4510 System, and the FDA-cleared predicate device, the iQ200 System (K093861) and manual microscopy. Testing was done by CLIA-trained operators using de-identified urine samples collected fresh within two (2) hours or refrigerated up to six (6) hours post collection. The study was conducted to evaluate the following 12 analytes: Quantitative analysis of red blood cells (RBC), white blood cells (WBC), and squamous epithelial cells (SQEC); Semiquantitative analysis of bacteria (BACT), and crystals (CRYS); Qualitative analysis of hyaline casts (HYAL), non-hyaline casts (NHC), non-squamous epithelial cells (NSE), yeast (YST), white blood cell clumps (WBCC), mucus (MUCS) and sperm (SPRM). Tables 13-16 summarize the reference range analysis as well as the overall clinical performance data for all sites combined. In summary, Quantitative, Semiquantitative and Qualitative parameters met the acceptance criteria.

Element95% Reference Limit90% CI
RBC9.027.34 - 10.70
WBC4.813.80 - 5.50
SQEC6.464.50 - 9.30
NSE3.212.50 - 4.54
NHC0.110.00 - 0.30
HYAL0.510.26 - 0.82
BACT119.666.28 - 180.66
CRYS0.030.00 - 0.00
YST00.00 - 0.00
WBCC0.050.00 - 0.30
MUCS26.620.50 - 35.39
SPRM00.00 - 0.00

Table 13. Reference Range results (n=247)

{21}------------------------------------------------

Weighted Deming Regression
ElementComparisonNR2InterceptSlope
RBCAI(M) vs. IQ(M)3770.9181.501 (0.432, 2.569)0.824 (0.763, 0.886)
WBCAI(M) vs. IQ(M)8450.9030.629 (-0.300. 1.558)0.968 (0.930, 1.006)
SQECAI(M) vs. IQ(M)3820.928-0.451 (-1.269, 0.367)0.933 (0.883, 0.982)

Table 14. Method Comparison Results for Quantitative Elements

  • (M) denotes manually reviewed and reclassified results.
Passing-Bablok Regression
ElementComparisonNR2InterceptSlope
RBCAI(M) vs. IQ(M)3770.9180.332 (-0.500, 1.490)0.824 (0.774, 0.886)
WBCAI(M) vs. IQ(M)8450.9031.992 (1.140, 3.067)0.891 (0.860, 0.925)
SQECAI(M) vs. IQ(M)3820.9280.641 (-0.066, 1.493)0.873 (0.826, 0.923)
  • (M) denotes manually reviewed and reclassified results.

Table 15. Method Comparison results for Semi-quantitative elements

511
-------------
AI-4510 (Manual) vs. Manual Microscopy
N1474
Sensitivity76.2% (72.9%, 79.3%)
Specificity83.7% (81.0%, 86.1%)

CRYS

AI-4510 (Manual) vs. Manual Microscopy
N1474
PPA90.5% (85.1%, 94.1%)
NPA98.2% (97.3%, 98.8%)

Table 16. Method Comparison results for Qualitative elements NSE

AI-4510 (Manual) vs. iQ200 (Manual)
N765
PPA88.7% (77.4%, 94.7%)
NPA84.3% (81.4%, 86.8%)

NHC

AI-4510 (Manual) vs. iQ200 (Manual)
N765
PPA80.2% (71.6%, 86.7%)
NPA83.8% (80.8%, 86.4%)

{22}------------------------------------------------

HYAL

AI-4510 (Manual) vs. iQ200 (Manual)
N765
PPA85.0% (78.4%, 89.9%)
NPA89.0% (86.3%, 91.2%)

YST

AI-4510 (Manual) vs. iQ200 (Manual)
N765
PPA97.1% (85.1%, 99.5%)
NPA99.6% (98.8%, 99.9%)

WBCC

AI-4510 (Manual) vs. iQ200 (Manual)
N1474
PPA86.5% (80.5%, 90.8%)
NPA89.3% (87.5%, 90.8%)

MUCS

AI-4510 (Manual) vs. iQ200 (Manual)
N1474
PPA81.9% (76.3%, 86.5%)
NPA88.0% (86.1%, 89.7%)

SPRM

AI-4510 (Manual) vs. iQ200 (Manual)
N1474
PPA86.2% (75.1%, 92.8%)
NPA99.6% (99.2%, 99.8%)

5.7 Proposed Labeling

Labeling adequately communicates device intended use, safety precautions and directions for use. It satisfies 21 CFR Part 809.10 for in vitro diagnostic devices.

5.8 Conclusion

ARKRAY has demonstrated the AI-4510 System is substantially equivalent to the predicate device, the iQ200 System, based upon design, test results, and indications for use. Any noted differences do not raise new issues of safety and effectiveness.

§ 864.5200 Automated cell counter.

(a)
Identification. An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.(b)
Classification. Class II (performance standards).