(228 days)
No
The summary describes standard physiological monitoring and arrhythmia detection based on established standards (ANSI/AAMI EC57). There is no mention of AI, ML, or any learning algorithms. The performance evaluation relies on standard database testing, not training and validation of a learning model.
No
The device is described as a monitoring system that collects and displays physiological vital signs and generates alerts, which indicates it is used for diagnostic and monitoring purposes rather than directly treating a condition.
Yes
The Vios Monitoring System (VMS) is intended for physiological vital signs monitoring and can generate alerts when physiological vital signs fall outside of selected parameters, as well as detecting cardiac arrhythmias. This direct assessment of a patient's physiological state and providing alerts based on deviations classifies it as a diagnostic device. The ECG rhythm analysis also aids in the identification of arrhythmia events and aids in clinical review, which directly contributes to diagnosis.
No
The device description explicitly states that the Vios Monitoring System (VMS) includes hardware components such as a Chest Sensor and Finger Adapter, in addition to the software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for "physiological vital signs monitoring of adult (18+) patients in healthcare facilities." This involves monitoring vital signs like ECG, heart rate, oxygen saturation, blood pressure, and activity. These are measurements taken directly from the patient's body, not from samples of bodily fluids or tissues.
- Device Description: The description details a "wireless mobile medical device platform" with components like a Chest Sensor and Finger Adapter. These are external devices applied to the patient.
- Lack of IVD Characteristics: The text does not mention any of the typical characteristics of an IVD, such as:
- Analyzing samples (blood, urine, tissue, etc.)
- Detecting diseases or conditions based on sample analysis
- Using reagents or other laboratory-based methods
The Vios Monitoring System is a patient monitoring device, which falls under a different regulatory category than IVDs.
N/A
Intended Use / Indications for Use
The Vos Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs montoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when the physiological vital signs fall outside of selected parameters.
VMS can also generate alerts when cardias arrhythmias (Tachycardia, Asystole, Ventricular Fibrillation and Atrial Fibrillation/ Atrial Flutter) are detected.
The ECG rhythm analysis is intended for use by medified professionals in the identification of arrhythmia events and to aid in clinical review of arrhythmias and medical interventions.
The Vos CSM/CS Software is indicated for use by healthcare professionals for the purpose of centralized monitoring of patient data within a healthcare facility. The Vios CSMCS SW receives, stores, and displays patient physiological and waveform data and alams generated by Vios proprietary patient vitals monitoring software.
Product codes
QYW, DQA, DPZ, DRG, DXJ, OUG
Device Description
The Vios Monitoring System (VMS) Model 2050 is a wireless mobile medical device platform that allows caregivers in healthcare settings to monitor patient vitals. The VMS includes a proprietary monitoring software, Chest Sensor, Finger Adapter and Central Server and Central Monitoring Station. The VMS BSM SW Model B2050 is stand-alone software that can receive, analyze, and display physiological vitals data from one or more patient-worn sensors via standard communication protocols (Bluetooth™). It runs on a commercial IT platform and is intended to be used in conjunction with the Vios Chest Sensor and Vios Lead Adapters and can support peripheral, medical grade, Bluetooth™-enabled devices. The VMS Chest Sensor Model CS2050 is a small, patient-worn, non-sterile multiple use,
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Adult (18+) patients
Intended User / Care Setting
medically qualified personnel for physiological vital signs montoring of adult (18+) patients in healthcare facilities.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
The subject device's performance was also evaluated using additional records from LTAF, AAEL, and VFDB database, other than referred in the standard.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The non-clinical tests for evaluation of performance of Vios system with the addition of arrhythmia alarms is based on ANSI/AAMI EC57, showing substantial equivalence to the predicate (K180472). The subject device's performance was also evaluated using additional records from LTAF, AAEL, and VFDB database, other than referred in the standard.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Vios Monitoring System™ Model 2050 (K172586), Vios Central Station Monitor/Central Server Software 2050 (K173107), ViSi Mobile Monitoring System (K180472)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2024
Murata Vios, Inc. John Lansdown General Manager, Quality, Regulatory, & Compliance 700 Commerce Drive. #190 Woodbury, Minnesota 55125
Re: K232354
Trade/Device Name: Vios Monitoring System (Vios Monitoring System(TM) Model 2050; Vios Central Station Monitor/Vios Central Server Software 2050) Regulation Number: 21 CFR 870.2300 Regulation Name: Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: QYW, DQA, DPZ, DRG, DXJ, OUG Dated: February 22, 2024 Received: February 22, 2024
Dear John Lansdown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
1
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely. Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Enclosure
3
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K232354
Device Name
Vios Monitoring System(TM) Model 2050
Indications for Use (Describe)
The Vos Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs montoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when the physiological vital signs fall outside of selected parameters.
VMS can also generate alerts when cardias arrhythmias (Tachycardia, Asystole, Ventricular Fibrillation and Atrial Fibrillation/ Atrial Flutter) are detected.
The ECG rhythm analysis is intended for use by medified professionals in the identification of arrhythmia events and to aid in clinical review of arrhythmias and medical interventions.
The Vos CSM/CS Software is indicated for use by healthcare professionals for the purpose of centralized monitoring of patient data within a healthcare facility. The Vios CSMCS SW receives, stores, and displays patient physiological and waveform data and alams generated by Vios proprietary patient vitals monitoring software.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."
4
510(k) Summary
510(k) Summary
| Submitter: | Murata Vios, Inc.
700 Commerce Drive, Suite 190
Woodbury, MN 55125 |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John Lansdown
General Manager, Quality, Regulatory, & Compliance
john.lansdown@murata.com
Office: 651-888-8125 |
| Date Prepared: | 8/4/2023 |
| Trade Names: | Vios Monitoring System™ Model 2050
Vios Central Station Monitor/Central Server Software 2050 |
| Common Name: | Vital Signs monitor |
| Regulation: | 21 CFR § 870.1025 Arrhythmia detector and alarm (including ST-segment measurement
and alarm).
21 CFR § 870.2300 Cardiac Monitor (including cardiotachometer and rate alarm)
21 CFR § 870.2910 Radiofrequency physiological signal transmitter and receiver
21 CFR § 870.2700 Oximeter
21 CFR § 870.2710 Ear oximeter
21 CFR § 870.2450 Medical cathode-ray tube display
21 CFR § 880.6310 Medical device data system |
| Classification: | Class II |
| Review Panels: | Cardiovascular, Anesthesiology |
| Product Codes: | QYW, DQA, DPZ, DRG, DXJ, OUG |
| Predicate Devices: | Vios Monitoring System™ Model 2050 (K172586)
Vios Central Station Monitor/Central Server Software 2050 (K173107)
ViSi Mobile Monitoring System (K180472) |
| Device Description: | The Vios Monitoring System (VMS) Model 2050 is a wireless mobile medical device
platform that allows caregivers in healthcare settings to monitor patient vitals.
The VMS includes a proprietary monitoring software, Chest Sensor, Finger Adapter and
Central Server and Central Monitoring Station.
The VMS BSM SW Model B2050 is stand-alone software that can receive, analyze, and
display physiological vitals data from one or more patient-worn sensors via standard
communication protocols (Bluetooth™). It runs on a commercial IT platform and is
intended to be used in conjunction with the Vios Chest Sensor and Vios Lead Adapters
and can support peripheral, medical grade, Bluetooth™-enabled devices.
The VMS Chest Sensor Model CS2050 is a small, patient-worn, non-sterile multiple use, |
| Indications for Use: | The Vios Monitoring System (VMS) is intended for use by medically qualified personnel
for physiological vital signs monitoring of adult (18+) patients in healthcare facilities. It is
indicated for use in monitoring of 7-lead ECG, heart rate, respiratory rate, pulse rate,
functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure, and
patient posture and activity. VMS allows for the input of body temperature, and can
display data from peripheral devices. VMS can generate alerts when the physiological
vital signs fall outside of selected parameters.
VMS can also generate alerts when cardiac arrhythmias (Tachycardia, Bradycardia,
Asystole, Ventricular Tachycardia/ Ventricular Fibrillation and Atrial Fibrillation/ Atrial
Flutter) are detected.
The ECG rhythm analysis is intended for use by medically qualified professionals in the
identification of arrhythmia events and to aid in clinical review of arrhythmias and
medical interventions.
The Vios CSM/CS Software is indicated for use by healthcare professionals for the
purpose of centralized monitoring of patient data within a healthcare facility. The Vios
CSM/CS SW receives, stores, manages, and displays patient physiological and waveform
data and alarms generated by Vios proprietary patient vitals monitoring software. |
| Summary of
Substantial
Equivalence | The Vios Monitoring System has the following similarities and differences with the
indicated predicate devices:
Intended use -The proposed device, Vios Monitoring System monitors the same vital
signs as the predicate, previously cleared Vios Monitoring System (K172586) except for
the addition of arrhythmia detection feature, which is included in the predicate, Visi
Monitoring System (K180472). Considered together, the intended use of the proposed
device is the same as the Vios Monitoring System the addition of arrhythmia detection
like the Visi System. The Vios Monitoring System, is similar to the predicate devices, is
intended to be used by licensed practitioners in a healthcare environment.
Device Performance – The modified Vios Monitoring System with the addition of
arrhythmia detection has been verified to meet the performance requirements as outlined
in the consensus standard ANSI/AAMI EC57:2012 - Testing and reporting performance
results of cardiac rhythm and ST-segment measurement algorithms. The subject device's
performance was also evaluated using additional records from LTAF, AAEL and VFDB
database. The predicate device (K180472) has also been tested for compliance to ANSI/
AAMI EC57 for the arrhythmia features.
Technology -The Vios Monitoring System is unchanged from its predicate submission
(K172586) except for the addition of arrhythmia detection measurement. The VMS
arrhythmia detection measurement feature are based on similar characteristic features to
|
| Summary of Non-
Clinical, Clinical, and
Conformance Testing | The non-clinical tests for evaluation of performance of Vios system with the addition of
arrhythmia alarms is based on ANSI/AAMI EC57, showing substantial equivalence to the
predicate (K180472). The subject device's performance was also evaluated using
additional records from LTAF, AAEL, and VFDB database, other than referred in the
standard. |
| | Other system performance remains unchanged from the predicate (K172586) except the
following standard: |
| | • AAMI/ANSI EC57 |
| | The safety and effectiveness of the VMS Model 2050 based on the following standards
remains unchanged: |
| | • Electrical safety, EMC, and vitals sign monitoring standards (IEC 60601-1, IEC
60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, EC53)
• Biocompatibility standards (ISO 10993)
• Transportation Simulation testing (ASTM D4169-16)
• Software development life cycle (EN 62304)
• Usability and human factors standards (EN 62366)
• Risk Management (ISO 14971)
• Pulse oximetry clinical testing (ISO 80601-2-61)
• Respiratory Rate clinical testing
• Standard for evaluation of Wireless Coexistence (ANSI C63.27) |
| Conclusions: | Based on the results from the non-clinical tests performed, Murata Vios considers the
Vios Monitoring System to be as safe, as effective and substantially equivalent to the
legally marketed predicate devices. |
5
6