K Number
K232317
Date Cleared
2024-04-26

(267 days)

Regulation Number
864.6550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UC-1800 Automatic Urine Analyzer is automated instrument which is intended for professional, in vitro diagnostic use only.

Depending on the reagent strips being used, the instruments perform semi-quantitative detection of the following analytes in urine: ascorbic acid, microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine hydrometer (optional) can determine the color and turbidity of urine. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 11FA urine reagent strips provide semi-quantitative tests for ascorbic acid, leukocytes, setone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 11FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 12FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

Device Description

UC-1800 Automatic Urine Analyzer is characterized by fully automated and simple operation. All you need to do is to set test strips and samples, press the START key, and the rest of operations are fully automated with UC-1800, which can measure samples continuously. For each measurement, the instrument automatically performs a series of operation: sample transmitting, sample aspirating, sample dropping, rinsing strip feeding and color identifying, etc. The instrument is used in conjunction with a serial of URIT urine test strips for measuring 15 parameters. Measure results are printed through either built-in printer or external printer.

Urine Reagent Strips is used to determine the components to be measured in urine by dry chemistry method together with urine analyzer. Various components to be tested in the urine can result in changes to the colors of corresponding reagent blocks on the Urine Reagent Strips. The depth of reaction color is proportional to the corresponding component to be tested in the urine. Qualitative and semi-quantitative detection can be conducted to the contents of the corresponding detected components. As a reagent for the determination of multiple components in human urine and the most basic test item for clinical urine routine test), it is suitable for the screening test or auxiliary diagnosis for clinical diagnosis, without the specificity for diseases or indications, and urine dry chemistry test is a screening test and cannot be used as a single diagnostic method.

AI/ML Overview

The provided document describes the URIT UC-1800 Automatic Urine Analyzer and its associated reagent strips (URIT 11FA and 12FA Urine Reagent Strips). The information below summarizes the acceptance criteria and the studies performed to demonstrate the device meets these criteria.


1. Table of Acceptance Criteria and Reported Device Performance

The document presents the performance in terms of "Exact agreement" and "± 1 color block" agreement with expected values or between the proposed device and predicate devices. For most analytes, the criteria seem to be high exact agreement and 100% agreement within ±1 color block. Specific thresholds for acceptance were not explicitly stated as global criteria but are implied by the "Qualified" conclusions for individual tests. The comparison to predicates also uses agreement rates.

Below is a summary of the reported device performance from the "Precision / Reproducibility" section (Tables 8, 9, 10, 11) for repeatability and "Comparison Studies" section (Tables a.1, a.2, a.3) for agreement with predicate devices. Given the extensive number of analytes and concentration levels, key representative results are presented.

a. Repeatability (Within-Run Precision)
Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.

Test (Analyte)Expected Value (Example)Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips)Conclusion
Ascorbic acid (11FA)-(0) mg/dL100% (60/60)Qualified
Nitrite (11FA/12FA)- (Negative)100% (60/60)Qualified
Leukocyte (11FA/12FA)-(0) leu/μL100% (60/60)Qualified
Ketone (11FA)-(0) mg/dL100% (60/60)Qualified
Ketone (12FA)+1(15) mg/dL96.7% (58/60)Qualified
Urobilinogen (11FA/12FA)Normal EU/dL100% (60/60)Qualified
Bilirubin (11FA)+2(2.0) mg/dL96.7% (58/60)Qualified
Bilirubin (12FA)+2(2.0) mg/dL91.7% (55/60)Qualified
Glucose (11FA/12FA)All100% (60/60)Qualified
Protein (11FA/12FA)+1(30) mg/dL93.3% (56/60)Qualified
Specific Gravity (11FA)1.01096.7% (58/60)Qualified
Specific Gravity (12FA)All100% (60/60)Qualified
Blood (11FA)+1(25) Cell/μL98.3% (59/60)Qualified
Blood (12FA)All100% (60/60)Qualified
pH (11FA)7.096.7% (58/60)Qualified
pH (12FA)All100% (60/60)Qualified
Microalbumin (12FA)All100% (60/60)Qualified
Creatinine (12FA)All100% (60/60)Qualified
Turbidity (Physical Module)All100% (60/60)Qualified
Color (Physical Module)All100% (60/60)Qualified

b. Reproducibility (Between-Run Precision)
Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.

Test (Analyte)Expected Value (Example)Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips)Conclusion
Ascorbic acid (11FA)+2(50) mg/dL94.2% (113/120)Qualified
Protein (11FA)+2(100) mg/dL99.2% (119/120)Qualified
Blood (11FA)+2(80) Cell/μL95% (114/120)Qualified
pH (11FA)7.099.2% (119/120)Qualified
Nitrite (12FA)All100% (120/120)Qualified
Blood (12FA)+2(80) Cell/μL95.8% (115/120)Qualified
All other analytes (11FA/12FA)Most levels100% (120/120) or very closeQualified

c. Comparison with Predicate Devices

Analyte (Strip)Predicate DeviceAgreement TypeAgreement Rate (%) (Reported value from the table)Conclusion
Ascorbic acid (11FA)Uritest-500B (K082811)Complete agreement97.60% to 100% across all levels (e.g., 99.00% for +1(25) level)Subst. Eq.
Leukocyte (11FA/12FA)Uritest-500B (K082811)Complete agreement95.80% to 100% across all levels (e.g., 95.80% for +2(125))Subst. Eq.
Ketone (11FA/12FA)Uritest-500B (K082811)Complete agreement97.40% to 100% across all levels (e.g., 97.40% for +1(15))Subst. Eq.
Nitrite (11FA/12FA)Uritest-500B (K082811)Overall Agreement (OPA)99.40% (11FA); 98.70% (12FA)Subst. Eq.
Urobilinogen (11FA/12FA)Uritest-500B (K082811)Complete agreement95.50% to 100% across all levels (e.g., 95.50% for +1(2.0) on 12FA)Subst. Eq.
Bilirubin (11FA/12FA)Uritest-500B (K082811)Complete agreement95.20% to 100% across all levels (e.g., 95.20% for +2(2.0) on 12FA)Subst. Eq.
Glucose (11FA/12FA)Uritest-500B (K082811)Complete agreement96.30% to 100% across all levels (e.g., 96.30% for +1(100))Subst. Eq.
Protein (11FA/12FA)Uritest-500B (K082811)Complete agreement95.50% to 100% across all levels (e.g., 95.50% for +1(30) on 12FA)Subst. Eq.
pH (11FA/12FA)Uritest-500B (K082811)Complete agreement97.80% to 100% across all levels (e.g., 97.80% for 6.5)Subst. Eq.
Specific Gravity (11FA/12FA)Uritest-500B (K082811)Complete agreement97.80% to 100% across all levels (e.g., 97.80% for 1.010 on 12FA)Subst. Eq.
Microalbumin (12FA)Mission® U120 Ultra Urine Analyzer (K142391)Complete agreement83.33% for 30mg/L, up to 98.21% for 150mg/LSubst. Eq.
Creatinine (12FA)Mission® U120 Ultra Urine Analyzer (K142391)Complete agreement90.91% for 300mg/dL, up to 97.74% for 50mg/dLSubst. Eq.
Color (Physical module)AUTION MAX AX-4030 Urinalysis System (K093098)Coincidence rate91.29% (Colorless) to 98.48% (Brown)Subst. Eq.
Turbidity (Physical module)AUTION MAX AX-4030 Urinalysis System (K093098)Coincidence rate96.00% (Turbid) to 100% (Micro turbid)Subst. Eq.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Repeatability (within-run) & Reproducibility (between-run):

    • Sample Size:
      • Repeatability: 60 measurements per concentration level per analyte (20 replicates x 3 instruments).
      • Reproducibility: 120 measurements per concentration level per analyte (20 days x 2 runs/day x 1 time/run in 3 sites, with 1 instrument/site, 3 operators).
    • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. It refers to "negative urines and spiked urines of known concentrations." This suggests controlled laboratory-prepared samples rather than direct patient samples.
  • Linearity/Assay Reportable Range:

    • Sample Size: 63 measurements per concentration level per analyte (reference solutions tested 21 times on 3 UC-1800 machines with 3 lots of reagent strips).
    • Data Provenance: Laboratory-prepared reference solutions, not clinical samples.
  • Analytical Sensitivity: No sample sizes mentioned, values are stated directly.

  • Critical Value: No sample sizes mentioned, values are stated directly.

  • Analytical Specificity (Interference, pH, Color, Specific Gravity):

    • Sample Size: For interference studies, samples were tested 5 times on 2 UC-1800 machines with 2 batch numbers of 11FA and 12FA reagent strips. This totals 20 measurements per condition (5 tests x 2 machines x 2 strips).
    • Data Provenance: Laboratory-prepared urine samples (negative samples prepared and spiked with interfering substances or adjusted for pH/color/SG).
  • Comparison Studies with Predicate Devices (Clinical Samples):

    • Uritest-500B (K082811): 1000 clinical urine samples.
    • Mission® U120 Ultra Urine Analyzer (K142391): 979 clinical urine samples.
    • AUTION MAX AX-4030 Urinalysis System (K093098):
      • Color: 1365 clinical urine samples.
      • Turbidity: 1000 clinical urine samples.
    • Data Provenance: The document explicitly states "clinical urine samples were collected." The country of origin is not specified but given the submitter's address (China), it is highly likely that these clinical samples were collected in China and were retrospective as the study compares against existing predicate devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of medical image interpretation or clinical diagnosis. For chemical analyzers, the "ground truth" (or reference standard) is typically established by:

  • Known concentrations: For repeatability, reproducibility, linearity, and analytical sensitivity, samples are prepared with known concentrations of the analytes.
  • Reference methods: For the analytical specificity and comparison studies, the predicate devices themselves or established reference methods (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Lange method for Ketone, etc. as specified in Table 35 "Traceability") serve as the reference for comparison.

Therefore, the concept of "number of experts" and their "qualifications" for ground truth establishment, as typically applied in AI/ML performance evaluation (e.g., for image interpretation), is not directly applicable here. The ground truth is analytical and based on laboratory standards and established measurement techniques.


4. Adjudication Method for the Test Set

Not applicable in the context of this analytical device. As explained above, the "ground truth" refers to known concentrations or results from predicate/reference methods. There isn't a subjective interpretation by multiple experts that would require an adjudication method like 2+1 or 3+1.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an automated urine analyzer, not an AI/ML-driven diagnostic aid for human readers. It performs measurements to provide semi-quantitative results for various urine analytes. Therefore, there is no human-in-the-loop performance or comparison of human reader improvement with or without AI assistance.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are all standalone performance evaluations of the device (UC-1800 Automatic Urine Analyzer with URIT 11FA/12FA Urine Reagent Strips) without human intervention in the measurement process. The device operates automatically to detect and report analyte levels. The comparison studies demonstrate its performance against existing predicate (standalone) analyzers.


7. The Type of Ground Truth Used

The ground truth used in the studies includes:

  • Known Concentrations: For analytical performance studies such as precision (repeatability and reproducibility), linearity, analytical sensitivity, and analytical specificity (interference testing), samples were prepared with known, precisely measured concentrations of the target analytes or interfering substances.
  • Predicate Device Measurements: For method comparison studies, the results obtained from the established predicate devices (Uritest-500B Urine Analyzer, Mission® U120 Ultra Urine Analyzer, and AUTION MAX AX-4030 Urinalysis System) served as the reference standard for comparison. These predicate devices also operate based on defined analytical principles.
  • Reference Methods: Table 35 details the "Reference Method" used for traceability for each analyte (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Glucose Oxidase Method for Glucose, Acidometer measurement for pH, etc.). This indicates fundamental analytical standards are the basis for the stated detection ranges and performance.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning. The UC-1800 Automatic Urine Analyzer uses reflectance photometry and other physical principles (refractometer for specific gravity, light-scattering for turbidity, light-transmission for color) to generate results, not an AI algorithm that learns from a dataset in the conventional sense. The "training" of such a system typically involves calibrating optical sensors and algorithms to known standards, which is part of the engineering design and quality control processes.

If interpretation of "training set" refers to the data used for the initial development and calibration of the device's measurement algorithms:

  • No specific sample size for a "training set" for an AI model is reported because the device does not employ machine learning that requires a distinct "training set" and "test set" in the typical AI/ML development lifecycle.
  • The system is calibrated using URIT urine control materials and calibration test strips (Table 9, "Calibration").

9. How the Ground Truth for the Training Set Was Established

As noted above, the device does not use an AI/ML model with a "training set" in the common understanding. The system's operational parameters and calibration are established using:

  • Reference materials and calibrators: The device is calibrated using "URIT urine control materials and calibration test strips" (Table 9, "Calibration"). These control materials and calibration strips would have their values established using highly accurate reference methods or certified reference materials, ensuring traceability to scientific standards (as indicated in Table 35 "Traceability").
  • Chemical principles: The underlying "ground truth" for the device's internal algorithms (i.e., how they convert optical signals to analyte concentrations) is based on established chemical reactions and physical measurement principles (e.g., reflectance photometry, refractometry, light scattering, dye-binding, enzymatic reactions). The "ground truth" for developing and fine-tuning these algorithms would be derived from rigorous scientific validation against these known chemical and physical properties.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 26, 2024

URIT Medical Electronic Co., Ltd. c/o Dylan Wu, Consultant Shanghai SUNGO Management Consutling Co., Ltd. Room 1401, Dongfang Building, 1500# Century AVE Shanghai 200122, China

Re: K232317

Trade/Device Name: UC-1800 Automatic Urine Analyzer, URIT 11FA Urine Reagent Strips, URIT 12FA Urine Reagent Strips Regulation Number: 21 CFR 864.6550 Regulation Name: Occult Blood Test Regulatory Class: Class II Product Code: JIO, JIL, JFY, CDM, CEN, JIN, JJB, JMT, LJX, JRE, JMA, JIR, KQO Dated: March 29, 2024 Received: March 29, 2024

Dear Dylan Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232317

Device Name

UC-1800 Automatic Urine Analyzer URIT 11FA Urine Reagent Strips URIT 12FA Urine Reagent Strips

Indications for Use (Describe)

The UC-1800 Automatic Urine Analyzer is automated instrument which is intended for professional, in vitro diagnostic use only.

Depending on the reagent strips being used, the instruments perform semi-quantitative detection of the following analytes in urine: ascorbic acid, microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine hydrometer (optional) can determine the color and turbidity of urine. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 11FA urine reagent strips provide semi-quantitative tests for ascorbic acid, leukocytes, setone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 11FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 12FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K232317

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement by 21 CFR 807.92

Date prepared: 25th, April 2024

1 Submitter's Information

Name: URIT Medical Electronic Co., Ltd. Address: No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China Contact Person: Jun Jiang Title: Manager Tel: 0773-2288586 FAX: 0773-2288560 Email: service@uritest.com 2 Official Contact Person Information

Primary contact: Mr. Dylan Wu Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-18616797137 Email: haiyang.wu@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

3 Subject Device

3.1 Trade Name and Regulatory Information:

No.Trade nameRegulatory Information
1UC-1800Automatic UrineAnalyzerClass I §21 CFR 862.2900Automated Urinalysis System
3URIT 11FA UrineReagent StripsClass II (Blood and Glucose analytesraise system to Class II / 510(k)required)
4URIT 12FA UrineReagent StripsClass II (Blood, Glucose and Creatinineanalytes raise system to Class II /510(k) required)

3.2 Classification Information

No.Regulation21 CFR SectionProductCodeClassificationDescriptionPanel
1862.2900KQOClass IAutomated Urinalysis SystemClinical Chemistry
2862.1225JFYClass IICreatinine test SystemClinical Chemistry
3862.1645JIRClass IProtein or Albumin(Urinary, Non-Quant.)Clinical Chemistry

{4}------------------------------------------------

No.Regulation21 CFR SectionProductCodeClassificationDescriptionPanel
4862.1340JILClass IIGlucose(Urinary, Non-Quantitative)Clinical Chemistry
5864.6550JIOClass IIBlood, Occult, Colorimetric, In UrineHematology
6862.1785CDMClass IUrobilinogen(Urinary Non-Quant.)Clinical Chemistry
7862.1550CENClass IpH(Urinary, Non-Quant.)Clinical Chemistry
8862.1435JINClass IKetones(Urinary, Non-Quant.)Clinical Chemistry
9862.1645JIRClass IProtein or Albumin(Urinary, Non-Quant.)Clinical Chemistry
10862.1115JJBClass IUrinary Bilirubin & Its Conjugates(Urinary, Non-Quant.)Clinical Chemistry
11862.1510JMTClass INitrite (Urinary, Non-Quant.)Clinical Chemistry
12864.7675LJXClass ILeukocyte peroxidase testHematology
13862.2800JREClass ISpecific GravityClinical Chemistry
14862.1095JMAClass IAscorbic Acid Test SystemClinical Chemistry

4 Predicate device

510(k) Number: K082811 Uritest 500B Urine Analyzer Uritest 50 Urine Analyzer Uritest 10G Urine Reagent Strips Uritest 11G Urine Reagent Strips

510(k) Number: K142391 Mission® U120 Ultra Urine Analyzer Mission® Urinalysis Reagent Strips (Microalbumin/Creatinine) Mission® Liquid Urine Controls, Mission® Liquid Diptube Urine Controls

510(k) Number: K093098 AUTION MAX AX-4030 Urinalysis System

5 Device Description

5.1 UC-1800 Automatic Urine Analyzer

UC-1800 Automatic Urine Analyzer is characterized by fully automated and simple operation. All you need to do is to set test strips and samples, press the START key, and the rest of operations are fully automated with UC-1800, which can measure samples continuously. For each measurement, the instrument automatically performs a series of operation: sample transmitting, sample aspirating, sample dropping, rinsing strip feeding and color identifying, etc. The instrument is used in conjunction with a serial of URIT urine test strips for measuring 15 parameters. Measure results are printed through either built-in printer or external printer. The device is described as below.

Test strip measurement principle a

Measurement of test strips is done by the reflectance photometry method.

Placing tube racks loaded with samples on the rack injection mechanism and clicking the START key, the instrument will automatically perform a series of operations, such as transmitting samples, selecting strips, reading barcode, aspirating

{5}------------------------------------------------

samples, dropping samples, measuring samples and printing results, until all tube racks are done. During the measurement, the reacted pads on strip (calibration pad is not involved in reaction, just for reference) will change colors and absorb irradiated monochromatic light as a result of chemical reaction within 60 seconds. Then the optical mechanism will compare the reflective light amount of each reacted pad with the reflective light amount of the concentrations of analyzes will be calculated by CPU and printed together with semi-quantitative symbols.

The measuring system consists of a light source (LED) and a light rom the light source falls on the reacted pads and the calibration pad on the strip. The absorbent and reflective light amounts vary with the color of reagent pads. If the color is darker, more light is absorbed, and less light is reflected, vice versa. i.e., the degree of color development is proportional to the concentration of analyzes in urine. The reflectance is calculated using the following formula:

$$\text{R%} = \frac{T_m \cdot \mathcal{C}_s}{T_s \cdot \mathcal{C}_m} \times 100% \tag{1}$$

R: Reflectance

Tm: Reflective light amount at the reactive pad with the measurement wavelength

Ts: Reflective light amount at the reactive pad with the reference wavelength

Cm: Reflective light amount at the calibration pad with the measurement wavelength

Cs: Reflective light amount at the calibration pad with the reference wavelength

The reflected lights from reagent pads are transmitted in the optical unit and received by the light receptor, where the optical signals are transformed into electrical signals. Then the electrical signals are transformed through I/V converter, processed by CPU and finally printed out.

b Specific gravity measurement principle

Specific gravity measurement method is refractometer, which using the correlation between light refractive index and total solids in the solution to determine.

Refractometer method, available at 15℃~38℃ temperature range of use can be calibrated by the temperature compensation device; available for use the known standard high specific gravity concentration solution and standard low specific gravity deionized water to calibrate; easy to standardization, less quantity of samples, especially suitable for patients with oliguria and pediatric patients. Refractometer method is reference method by Clinical Laboratory Standard Institution, CLSI and Chinese Committee for Clinical Laboratory Standards, CCCLS.

Specific gravity is based on the principle of different concentrations of urine sample which have different refractive indexes to measure, that is uses the same wavelength of monochromatic parallel light comes into the contains urine sample, and then according to position of refracted ray in photoelectric technology detector (displacement sensor) to determine the specific gravity value. Specific gravity measurement principle functional block diagram is shown in Figure 1 below.

Image /page/5/Figure/14 description: The image shows a light ray entering a prism and being refracted. The light ray is labeled "Light ray" and the refracted ray is labeled "Refracted ray". The prism is labeled "Prism" and a detector is labeled "Detector".

Figure 1 Specific gravity measurement principle

Specific gravity results are calculated by the following formulas:

$$SG_X = \left{ SG_H - SG_L \right} \cdot \frac{K_X - K_L}{K_H - K_L} + SG_L \tag{2}$$

can change to

$$\frac{SG_X - SG_L}{SG_H - SG_L} = \frac{K_X - K_L}{K_H - K_L} \tag{3}$$

relationship between them is linear.

SGH: The specific gravity of high concentration solution

{6}------------------------------------------------

SGL: The specific gravity of low concentration solution

SGX: The specific gravity of sample solution

KH: High concentration solution position coefficient

KL: Low concentration solution position coefficient

KX: Sample solution position coefficient

Position coefficient: It is calculated by the detector output data, and has a linear relationship with the refractive index.

Refractive index change depends on the temperature of the sample solution, and the specific gravity value is using the following formula to correct.

$$\text{SG}_t = \text{SG}_X + {T_{SAM} - T_{STD}} \cdot \mathcal{C}_t \tag{4}$$

SGt: The specific gravity of high concentration solution

SGx: The specific gravity of low concentration solution

TSAM: The temperature of sample solution

TSTD: The temperature of low concentration solution

Ct: Temperature coefficient (SG 0.001/3°C) (temperature coefficient)

If the urine sample contains large amounts of glucose or protein, then the specific gravity will be affected, according to WS/T 229-229 "Physical, chemical and microscopic examination of urine" 5.4.1 requirements: 1 g/L protein will increase urine specific gravity 0.0003, 1 g/L glucose will increase urine specific gravity 0.0004. So, the specific gravity results will be corrected through the glucose and protein level which was measured by the test strip.

$$\mathcal{SG} = \mathcal{SG}_t - \mathcal{C}_{GLU} - \mathcal{C}_{PRO} \tag{5}$$

SG: Specific gravity value which after the temperature compensation

SGt: Specific gravity value which gets from formula (4)

CGLU: Glucose correction value

CPRO: Protein correction value

C Turbidity measurement principle

Turbidity module emits light, to make it go through the sample, and then detect how much light is scattered by the particles in the water from the direction at a 90-degree angle to the incident light (The most stable angle of scattered light, is at a right angle to the center line of the incident light, so measuring the scattered light from 90° direction which can minimize the influence of particle size on scattering light intensity). This scattered light measurement method called scattering method. Turbidity measurement principle functional block diagram is shown in Figure 2 below.

Image /page/6/Figure/22 description: The image shows a diagram of a glass making sample pool. A light ray enters the pool from the left and is scattered in different directions. Transmission light exits the pool to the right and is measured by a detector. Scattered light at 90 degrees is also measured by a detector.

Figure 2 Turbidity measurement principle

Turbidity result is calculated by the following formula:

$$T = \left(\mathbb{S}_{\mathcal{S}}/T_{\mathbb{S}} - \mathbb{S}_{W}/T_{W}\right)/K \tag{6}$$

T: Turbidity level

{7}------------------------------------------------

Ss: Sample scattered light level

Ts: Sample transmission light level

Sw: Flushing fluid scattered light level

Tw: Flushing fluid transmission light level

K: Coefficient factor

d Color measurement principle

The color of often seen objects, are actually the objects surfaces absorb a part of chromatic light from the white light(sunlight) that fall on them, and then reflect the other part of chromatic light to human eyes' response. Various frequencies of visible light are mixed together become white, that is to say, it contains all sorts of color of light, such as red(R), yellow(Y), green(G), blue(B), purple(P). According to the German physicist Helinholtz's three primary colors theory, all sorts of color is made of different proportion of three primary colors (red, green, blue).

Primary colors are the "basic color" which cannot be gotten by other colors mixed. But mix the primary colors in different proportion will get other new colors. Three primary colors of light are RGB (Red, Green, and Blue). Equivalent red light +green light=yellow light, green light +blue light = cyan light. Equivalent red light = magenta light, equivalent red +green+ blue=white, and if the intensity of these three lights is zero, it is black (dark).

When the white light through colored solution colors light will be absorbed, so the color of the light through the solution can be expressed as the color of the solution color can be detected by the professional color recognition sensor (filter) which in the back-end of the solution.

Image /page/7/Figure/10 description: The image shows a diagram of a glass making sample pool. A white light ray is shown entering the glass making sample pool, which is represented by a circle. After passing through the sample pool, the light is labeled as transmission light. The transmission light then passes through an RGB light filter before reaching a detector.

RGB light fliter

Figure 3 Principle of color detection

5.2 URIT 11FA/12FA urine reagent strips

Urine Reagent Strips is used to determine the components to be measured in urine by dry chemistry method together with urine analyzer. Various components to be tested in the urine can result in changes to the colors of corresponding reagent blocks on the Urine Reagent Strips. The depth of reaction color is proportional to the corresponding component to be tested in the urine. Qualitative and semi-quantitative detection can be conducted to the contents of the corresponding detected components. As a reagent for the determination of multiple components in human urine and the most basic test item for clinical urine routine test), it is suitable for the screening test or auxiliary diagnosis for clinical diagnosis, without the specificity for diseases or indications, and urine dry chemistry test is a screening test and cannot be used as a single diagnostic method.

a Ascorbic Acid

Based on the principle of Tillman's Reagent, Ascorbic acid can reduce the dye from blue to red. The purpose of the determination of this project is to provide the user with the content of Ascorbic acid in the sample to determine its possible interference.

Nitrite b

In this reaction, nitrate is reduced to nitrite by Gram-negative bacteria in the nitrite will react with arsanilic acid to form a diazonium compound and the diazonium compound with naphthy ethylenediamine dihydrochloride to show a pink color.

{8}------------------------------------------------

Microalbumin C

Based on the dye-binding method, microalbumin can react with the dye to form a pink complex and generate produce a color change, which is particularly sensitive to the reaction of albumin.

Leukocytes d

Based on the principle of esterase method, granulosa cytoplasm contains esterase which can hydrolyze a 3-hydroxyindoxyl ester substrate, release phenol and react with diazo reagent to generate purple-red compounds.

Creatinine e

Based on the principle of displacement reaction, creatinine can displace the dye in the metal chloride-acid dye complex, and the color will change from green to yellow.

f Ketone

Based on the principle of sodium nitroprussiate method, sodium nitroprusside can interact with ketone (acetoacetate) under alkaline conditions to become purple, especially acetoacetate is particularly sensitive to this.

Urobilinogen g

Based on the principle of azo-binding method, urobilinogen is coupled with diazonium salt under strong acid conditions to form carmine pigment.

h Bilirubin

Based on the principle of azo-coupline method,2,4-dichloroaniline diazonium salt can react specifically with bilirubin, and produce different colors depending on the concentration of bilirubin.

i Glucose

Based on the reaction principle of glucose oxidase can specifically oxidize ß-D-glucose to generate glucuronic acid and hydrogen peroxide that will oxidize the indicator under the action of peroxidase and show a purple-red color.

Protein j

Based on the principle of protein error method of dye binding, the protein can combine with the dye to form a complex that produce a color change, especially the response to albumin is more sensitive than that of globulin, hemoglobin, Bence-Jone protein and mucin.

k Blood

Based on the principle of hemoglobin contact activity method, the decomposition of peroxides can be catalyzed through the peroxidase-like action of hemoglobin, so that tetramethylbenzidine is oxidized and colored.

l Specific Gravity

Using the polyelectrolyte method, and based on the principle of ion exchange between electrolytes in urine and polyelectrolytes. In the presence of cations, the polymer hydrogen ions will be released through exchange, and the color of bromothymol blue indicator will change from blue to blue-green and finally to yellow.

pH m

The pH value within the range from 5.0 to 9.0 is measured by pH indicator, and the pH value of fresh urine of normal people is within the range from 5.5 to 7.0.

6 Indications for use/Intended use

The UC-1800 Automatic Urine Analyzer is automated instrument, which is intended for professional, in vitro diagnostic use only.

Depending on the reagent strips being used, the instruments perform semi-quantitative detection of the following analytes in urine: ascorbic acid, microalbumin, letone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine hydrometer (optional) can determine the color and turbidity of urine. Test results may provide information regarding the status of carbohydrate

{9}------------------------------------------------

metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 11FA urine reagent strips provide semi-quantitative tests for ascorbic acid, leukocytes, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 11FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

The URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin, leukocytes, creatinine, ketone, urobilingen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 12FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

Table 1 Technological characteristics of Proposed Device
FeatureUC-1800 Automatic Urine Analyzer
MethodologyReflectance photometer
PrincipleMeasurement of test strips is done by the reflectancephotometry method, using CIS (contact image sensor) imagescanning analysis technology to detect.
ChemistryURIT 11FA Urine Reagent StripsURIT 12FA Urine Reagent Strips
Throughput480test/hour
Memory2 million sample data and 100,000 sample pictures.
PC PortPS/2 interface, serial port, Ethernet interface, USB interface
CapabilitiesBarcode Scanner; Built-in thermal printer or external USBprinter
Available Languages onScreenChinese or English
Environmentrequirement15°C-30°C,RH ≤ 80%
Specimen ID enterManually enter or by bar code reader
Dimensions (L × W × H)653mm×641mm×570mm
Power SourceAC 100V-240V~, 50/60Hz, three-core power supply, goodgrounding.
Weight75kg
CalibrationThe user can use the URIT urine control materials andcalibration test strips to calibrate the instrument

7 Comparison of technological characteristics with the predicate device

Tahlo 1 Tochnological charactorictics of Propo

7.1 Substantial Equivalence

The URIT 11FA/12FA Urine Reagent Strips and the URIT UC-1800 Automatic Urine Analyzer is substantially equivalent to the Uritest 11G Urine Reagent Strips and the Uritest-500B Urine Analyzer (K082811) except Microalbumin and Creatinine. And the URIT 12FA Urine Reagent Strips and the URIT UC-1800 Automatic Urine Analyzer is substantially equivalent to the Mission® Urinalysis Reagent Strips (Microalbumin/Creatinine) and the Mission® U120 Ultra Urine Analyzer(K142391) Microalbumin and Creatinine only

And the URIT UC-1800 Automatic Urine Analyzer is substantially equivalent to AUTION MAX AX-4030 Urinalysis System (K093098) in detecting Specific Gravity, Turbidity and Color detected by Physical Module.

{10}------------------------------------------------

Table 2 Total review of proposed device and predicate devices
Device NamePredicate Device NamePredicate510(K)numberNote
URIT UC-1800Automatic UrineAnalyzerUritest-500B Urine AnalyzersK082811Except Microalbumin/Creatinine
Mission® U120 Ultra Urine AnalyzerK142391Microalbumin/Creatinine only
AUTION MAX AX-4030 UrinalysisSystemK093098Specific Gravity, Turbidity and Colordetected by Physical Module
URIT 11FA UrineReagent StripsUritest 11G Urine Reagent StripsK082811N/A
URIT 12FA UrineReagent StripsUritest 11G Urine Reagent StripsK082811Except Microalbumin/Creatinine
URIT 12FA UrineReagent StripsMission® Urinalysis Reagent Strips(Microalbumin/Creatinine)K142391Microalbumin/Creatinine only

{11}------------------------------------------------

Table 3 Comparison with Predicate Devices (1)
DeviceProposed Device (K232317)Predicate Device 1(K082811)Predicate Device 2(K142391)Predicate Device 3 (K093098)
ManufacturerURIT Medical Electronic Co., Ltd.URIT Medical Electronic Co., Ltd.Acon Laboratories, Inc.ARKRAY, Inc.
Device nameURIT UC-1800 Automatic Urine AnalyzerUritest-500B Urine AnalyzersMission® U120 Ultra Urine AnalyzerAUTION MAX AX-4030 Urinalysis System
Similarities
ClassificationClass II (Blood, Glucose and Creatinine analytes raisesystem to Class II / 510(k) required)Class II (Blood and Glucose analytes raise system toClass II / 510(k) required)Class II (Creatinine analytes raise system toClass II / 510(k) required)Class II (Blood and Glucose analytes raisesystem to Class II / 510(k) required)
Intended UseUC-1800 Automatic Urine Analyzer is automatedinstruments which are intended for professional, in vitrodiagnostic use only.Depending on the reagent strips being used, the instrumentsperform semi-quantitative detection of the followinganalytes in urine: ascorbic acid, microalbumin, leukocytes,creatinine, ketone, urobilinogen, bilirubin, glucose, protein,specific gravity, blood and pH in urine and for qualitativedetermination of nitrite in urine. The urine hydrometer(optional) can determine the specific gravity, color andturbidity of urine. Test results may provide reference forclinical examination and diagnosis.Uritest-500B urine analyzer is semi-automated, benchtop instruments which are intended for prescription, invitro diagnostic use only. The instruments perform semi-quantitative detection of the following analytes in urine:leukocytes, ketone, nitrite, urobilinogen, bilirubin,protein, glucose, specific gravity, blood, pH and ascorbicacid. Test results may provide information regarding thestatus of carbohydrate metabolism, kidney and liverfunction, acid base balance and bacteriuria. Theinstruments use the accompanying check strip for dailycalibration.The Mission U120 Ultra Urine Analyzer is aurinalysis instrument intended for in vitrodiagnostic use. It is intended for professional useonly at point-of-care locations. The Mission U120Ultra Urine Analyzer is intended to read MissionUrinalysis Reagent strips(Microalbumin/Creatinine) for the semiquantitative measurement of Albumin andCreatinine. These measurements are used to assistdiagnosis for kidney function.The AX-4030 is an automated urine analyzer that isdesigned to measure and analyze urine samplesusing measurements that include but are notlimited to; Normal, STAT, Control and Check. Thesemeasurements are used to examine the followinganalytes; glucose (GLU), protein (PRO), bilirubin(BIL), urobilinogen (URO), pH (PH), blood (BLD),ketones (KET), nitrite (NIT) leukocytes (LEU) andspecific gravity (S.G.). In addition, this device isused only with AUTION Sticks 9EB multi-parameter test strips.
Data TypeQualitative, Semi-QuantitativeQualitative, Semi-QuantitativeSemi-QuantitativeSemi-Quantitative
LocationClinical LaboratoryClinical LaboratoryClinical LaboratoryClinical Laboratory
Medical Device /IVDMedical Device /IVDMedical Device /IVDMedical Device /IVDMedical Device /IVD
IndicationsFor prescription use only.For prescription use only.For prescription use only.For prescription use only.
Specimen TypeHuman UrineHuman UrineHuman UrineHuman Urine
AnalyteWith strips: Glucose, Blood, Urobilinogen, pH, Ketones,Protein, Bilirubin, Nitrite, Leukocyte, Specific Gravity,Ascorbic acidGlucose, Blood, Urobilinogen, pH, Ketones, Protein,Bilirubin, Nitrite, Leukocyte, Specific Gravity, AscorbicacidN/AN/A
With strips: Creatinine, MicroalbuminN/ACreatinine, MicroalbuminN/A
With Physical module: Specific Gravity, Turbidity and ColorN/AN/ASpecific Gravity, Turbidity and Color
Methodology/PrincipleTest strips: Reflectance photometer, Measurement of teststrips is done by the reflectance photometry method,using CIS (contact image sensor) image scanning analysistechnology to detect.Specific Gravity: Refractometer methodColor: Light-transmission measurementTurbidity: Light-scattering measurement methodReflectance photometer (Only Test strips)Ingredients that change color in reaction with analytesReflectance photometer (Only Test strips)The Mission® U120 Ultra Urine Analyzer utilizesa CMOS image sensor to measure the intensity oflight. The frequency of the light is determined bythe LED light source.(without Test Strips)Specific gravity: Reflection refractometryColor hue: Light-transmission measurementTurbidity: Light-scattering measurement method
Specimen ID enterManually enter or by bar code readerManually enter or by bar code readerManually enter or by bar code readerBarcode Reader
Environmentrequirement15°C-30°C,RH ≤ 80%15°C-30°C,RH ≤ 80%15-30°C (59-86°F); 20-80% Relative Humidity(noncondensing)Temperature: 10-30°C,Humidity: 20-80% RH(Nocondensation )

9 / 42

{12}------------------------------------------------

Table 4 Comparison with Predicate Devices (2)
DeviceProposed Device (K232317)Predicate Device 1(K082811)Predicate Device 2(K142391)Predicate Device 3(K142392)
ManufacturerURIT Medical Electronic Co., Ltd.URIT Medical Electronic Co., Ltd.Acon Laboratories, Inc.ARKRAY, Inc.
Device nameURIT UC-1800 Automatic Urine AnalyzerUritest-500B Urine AnalyzersMission® U120 Ultra Urine AnalyzerAUTION MAX AX-4030 UrinalysisSystem
Differences
ChemistryURIT 11FA Urine Reagent StripsURIT 12FA Urine Reagent StripsUritest 11G Urine Reagent StripsMission® Urinalysis Reagent Strips(Microalbumin/ Creatinine)None
AvailableLanguages onScreenChinese or EnglishChinese or EnglishEnglish and SpanishEnglish
Dimensions(L × W × H)653mm×641mm×570mm390 mm x 340 mm ×290 mm26.0 (L) x 15.0 (W) x 17.5 (H) cm530 (w) x 530(d) x 530 (h)(mm)
Power SourceAC 100V-240V~, 50/60Hz,three-core power supply, good grounding.AC100-240V, 50/60Hz Input: 61VA100- 240 VAC (adapter), (50-60Hz± 1HZ)6AA batteries with 100 tests/6 new batteries;100-240 VAC, 50/60 Hz
Weight75kg6.5 Kg≤1.66 kg (3.65 lbs.) without batteries orpower supply41Kg (including the Sampler unit 4kg)
PC PortPS/2 interface, serial port,Ethernet interface, USB interfaceRS 232 port.Standard RS232C Port (cable not included),USB Port (cable not included);(Not connected to PC) Bluetooth WirelessRS-232C/ Ethernet
CapabilitiesBarcode ScannerBuilt-in thermal printer or external USB printerInternal or external PrinterInternal thermal printerBarcode reader ConnectorExternal printer (optional)Barcode reader (optional)RJ45 Ethernet; (optional)Built-in printer
CalibrationThe user can use the URIT urine control materials andcalibration test strips to calibrate the instrumentSelf-calibration---white calibration barThe Mission® U120 Ultra Urine Analyzerperforms an automatic calibration each timea test is run.N/A
Throughput480test/hour500 tests/hourSingle Test Mode: 55 tests/hour.Continuous Test Mode: 120 tests/hour.225 samples/hour
Memory2 million sample data and 100,000 sample pictures1,000 resultsLast 2000 test results2500 tests

{13}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China

Characteristic of the URIT 11FA/12FA Urine Reagent Strips are compared with the Uritest 11G Urine Reagent Strips (K082811) for instrument reading in Table 5 to Table 7

DeviceProposed DevicePredicate Device 1
ManufacturerURIT Medical Electronic Co., Ltd.URIT Medical Electronic Co., Ltd.
510K numberK232317K082811
Device nameURIT 11FA Urine Reagent StripsURIT 12FA Urine Reagent StripsUritest 11G Urine Reagent Strips
ClassificationClass II (Blood, Glucose and Creatinineanalytes raise system to Class II / 510(k)required)Class II (Blood and Glucose analytes raisesystem to Class II / 510(k) required)
Intended UseThe URIT 11FA urine reagent strips providesemi-quantitative tests for ascorbic acid,leukocytes, ketone, urobilinogen, bilirubin,glucose, protein, specific gravity, blood andpH in urine and for qualitative determinationof nitrite in urine. The URIT 11FA urinereagent strips are for use with the UC-1800Automatic Urine Analyzer and are forprofessional, in vitro diagnostic use only.The URIT 12FA urine reagent strips providesemi-quantitative tests for microalbumin,leukocytes, creatinine, ketone, urobilinogen,bilirubin, glucose, protein, specific gravity,blood and pH in urine and for qualitativedetermination of nitrite in urine. The URIT12FA urine reagent strips are for use with theUC-1800 Automatic Urine Analyzer and arefor professional, in vitro diagnostic use only.Uritest 11G urine reagent strips providesemi-quantitative tests for Ascorbic acid,Nitrite, Leukocyte, Ketone, Urobilinogen,Bilirubin, Glucose, Protein, Blood, SpecificGravity, PH in urine. The Uritest 11 G urinereagent strips are for use with Uritest-500Burine analyzer and are for prescription, invitro diagnostic use only. Test results mayprovide information regarding the status ofcarbohydrate metabolism, kidney and liverfunction, acid-base balance and bacteriuriaThe strips are read instrumentally by theUritest-500B Urine Analyzers
IndicationsProfessional in vitro diagnostic useSame
AnalytesURIT 11FA/12FA Urine Reagent Strips:Nitrite, Leukocyte, Ketone, Urobilinogen,Bilirubin, Glucose, Protein, Blood, SpecificGravity, PHURIT 11FA Urine Reagent Strips also providethe detection of Ascorbic acid,While URIT 12FA Urine Reagent Stripsprovide the detection of microalbumin andcreatinine (Predicated with K142391)Ascorbic acid, Nitrite, Leukocyte, Ketone,Urobilinogen, Bilirubin, Glucose, Protein,Blood, Specific Gravity, PH
SpecimenHuman UrineSame
ArchitectureFirm plastic, dry reagent stripsSame
Table 5 Comparison with Predicate Devices-Reagent strips (1)
DeviceProposed DevicePredicate Device 1
ManufacturerURIT Medical Electronic Co., Ltd.URIT Medical Electronic Co., Ltd.
510K numberK232317K082811
Device nameURIT 11FA Urine Reagent StripsURIT 12FA Urine Reagent StripsUritest 11G Urine Reagent Strips
ClassificationClass II (Blood, Glucose and Creatinine analytes raise system to Class II /510(k) required)Class II (Blood and Glucose analytes raise system toClass II / 510(k) required)
Test PrincipleIngredients that change color in reaction with analytesSame
Blank block(calibrator)The distribution of urine on the test block and the color of the urine itself willgenerally cause errors to the measurement results. The purpose of setting theblank block is to eliminate these factors that may cause error, and the same blankblock is used for each project.Same
Ascorbic acidBased on the principle of Tillman's Reagent, Ascorbic acid can reduce the dye fromblue to red. The purpose of the determination of this project is to provide the userwith the content of Ascorbic acid in the sample to determine its possibleinterference.Same
NitriteIn this reaction, nitrate is reduced to nitrite by Gram-negative bacteria in the urine,the nitrite will react with arsanilic acid to form a diazonium compound, and thediazonium compound will be combined with naphthylethylenediaminedihydrochloride to show a pink color.Same
MicroalbuminBased on the dye-binding method, microalbumin can react with the dye to form apink complex and generate produce a color change, which is particularly sensitiveto the reaction of albumin.None
LeukocyteBased on the principle of esterase method, granulosa cytoplasm contains esterasewhich can hydrolyze a 3-hydroxyindoxyl ester substrate, release phenol and reactwith diazo reagent to generate purple-red compounds.Same
CreatinineBased on the principle of displacement reaction, creatinine can displace the dye inthe metal chloride-acid dye complex, and the color will change from green to yellowNone
KetoneBased on the principle of sodium nitro prussiate method, sodium nitroprusside caninteract with ketone (acetoacetate) under alkaline conditions to become purple,especially acetoacetate is particularly sensitive to this.Same
UrobilinogenBased on the principle of azo-binding method, urobilinogen is coupled withdiazonium salt under strong acid conditions to form carmine pigment.Same
BilirubinBased on the principle of azo-coupling method,2,4-dichloroaniline diazonium saltcan react specifically with bilirubin, and produce different colors depending on theconcentration of bilirubin.Same
GlucoseBased on the reaction principle of glucose oxidase method, glucose oxidase canspecifically oxidize β-D-glucose to generate glucuronic acid and hydrogen peroxidethat will oxidize the indicator under the action of peroxidase and show a purple-red color.Same
ProteinBased on the principle of protein error method of dye binding, the protein cancombine with the dye to form a complex that produce a color change, especially theresponse to albumin is more sensitive than that of globulin, hemoglobin, Bence-Jone protein and mucin.Same
BloodBased on the principle of hemoglobin contact activity method, the decompositionof peroxides can be catalyzed through the peroxidase-like action of hemoglobin, sothat tetramethylbenzidine is oxidized and colored.Same
Specific GravityThis test contains a detergent and Bromthymol blue that indicates the presence ofionic constituents in the urine by changing color from green to yellow.Same
PHThe pH value within the range from 5.0 to 9.0 is measured by pH indicator, and thepH value of fresh urine of normal people is within the range from 5.5 to 7.0.Same
Detection RangeAscorbic Acid (10-100) mg/dLSame
Nitrite (+/-)Same
Leukocytes (15-500)leu/mcLSame
Ketone (5-80) mg/dLSame
Urobilinogen (2-8)EU/dLSame
Bilirubin (0.5-6.0) mg/dLSame
Glucose (50-1000) mg/dLSame
Protein (15-300)mg/dLSame
Blood (10-200) ery/mcLSame
Specific Gravity 1.005-1.030Same
pH 5.0-9.0Same

{14}------------------------------------------------

{15}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China

Characteristic of the URIT 12FA Urine Reagent Strips in detecting Microalbumin and creatinine are compared with the Mission® Urinalysis Reagent Strips (Microalbumin/Creatinine) (K142391) read by Mission®U120 Ultra Urine Analyzer in Table 7.

DeviceProposed DevicePredicate Device 2
ManufacturerURIT Medical Electronic Co., Ltd.Acon Laboratories, Inc.
510K numberK232317K142391
Device nameURIT 12FA Urine Reagent Strips(Only Microalbumin/Creatinine)Mission® Urinalysis Reagent Strips(Microalbumin/Creatinine)
ClassificationClass IIClass II
Similarities
Intended UseThe URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin , leukocytes,creatinine , ketone, urobilinogen, bilirubin,glucose, protein, specific gravity, blood and pH inurine and for qualitative determination of nitritein urine. The URIT 12FA urine reagent strips arefor use with the UC-1800 Automatic UrineAnalyzer and are for professional, in vitrodiagnostic use only.The Mission Urinalysis Reagent strips(Microalbumin/Creatinine) are intended for thesemi quantitative measurement of albumin andcreatinine in urine samples using the MissionU120 Ultra Urine Analyzer. Thesemeasurements are used to assist diagnosis forkidney function. It is intended for professionaluse only at point-of-care locations.The strips are read instrumentally by theMission® U120 Ultra Urine Analyzer
IndicationsProfessional in vitro diagnostic useProfessional use in point-of-care urine testing
AnalytesMicroalbumin and CreatinineMicroalbumin and creatinine
SpecimenHuman UrineHuman Urine
ArchitectureFirm plastic, dry reagent stripsPlastic strips affixed with two separate reagentareas.
Differences
Test PrincipleIngredients that change color in reaction withanalytesIntensity of the light reflected from the reagentareas of a urinalysis reagent strip.
StorageStore at 2°C-30°CStore at 15°C-30°C
MicroalbuminDetectionMethodologyBased on the dye-binding method, microalbumincan react with the dye to form a pink complexand generate produce a color change, which isparticularly sensitive to the reaction of albumin.This test is based on dye binding using a highaffinity sulfone phthalein dye. At a constant pH,the development of any blue color is due to thepresence of albumin. The resulting color rangesfrom pale green to aqua blue
CreatinineDetectionMethodologyBased on the principle of displacement reaction,creatinine can displace the dye in the metalchloride-acid dye complex, and the color willchange from green to yellow.This test is based on the peroxidase-like activityof a copper creatinine complex that catalyzes thereaction of diisopropylbenzene dihydroperoxideand 3,3',5,5'- tetramethylbenzidine. Theresulting color ranges from orange throughgreen to blue
DetectionRangeMicroalbumin (10-150) mg/LDetects albumin between 10-150mg/L
Creatinine(10-300)mg/dLDetects creatinine between 10-300 mg/dL (0.9 -26.5 mmol/L)

Table 7 Comparison with Predicato novicos.Roggont ctrine (2)

Therefore, it can be concluded that the subject device and the proposed device are substantially equivalent.

{16}------------------------------------------------

8 Summary of Non-Clinical Testing

8.1 Analytical performance

Precision /Reproducibility a

The Repeatability (With-in Run) Precision of the URIT UC-1800 Automatic Urine Analyzer was evaluated by using negative urines and spiked urines of known concentrations for each analyte. Each sample was tested in 20 replicates across 3 instruments with the 3lots of URIT 11FA Urine Reagent Strips and 3lots of URIT 12FA Urine Reagent Strips in each instrument for a total of 60 measurements at each concentrations for all tests had 60 of 60 (100%) match at ± one color block. The results are summarized in Table 8 to Table 11.

TestExpected Value
Ascorbic acid (mg/dL)-(0)+-(10)+1(25)+2(50)+3(100)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Nitrite- (Negative)+ (Positive)
Exact agreement100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)
ConclusionQualifiedQualified
Leukocyte(leu/μL)-(0)+-(15)+1(70)+2(125)+3(500)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Ketone (mg/dL)-(0)+-(5)+1(15)+2(40)+3(80)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Urobilinogen (EU/dL)Normal+1(2.0)+2(4.0)+3(8.0)
Exact agreement100% (60/60)100% (60/60)96.7% (58/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified
Bilirubin (mg/dL)-(0)+1(0.5)+2(2.0)+3(6.0)
Exact agreement100% (60/60)100% (60/60)96.7% (58/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified
Glucose(mg/dL)-(0)+-(50)+1(100)+2(250)+3(500)+4(1000)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Protein(mg/dL)-(0)+-(15)+1(30)+2(100)+3(300)
Exact agreement100% (60/60)100% (60/60)93.3% (56/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Specific Gravity1.0051.0101.0151.0201.0251.030
Exact agreement100% (60/60)96.7% (58/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
TestExpected Value
Blood(CELL/µL)-(0)+-(10)+1(25)+2(80)+3(200)
Exact agreement100% (60/60)100% (60/60)98.3% (59/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
pH5.05.56.06.57.07.5
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)96.7% (58/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
pH8.08.59.0
Exact agreement100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualified

Table 8 Summary of Repeatability (With-in Run) Precision of URIT 11FA Urine Reagent Strips

{17}------------------------------------------------

Table 9 Summary of Repeatability (With-in Run) Precision of URIT 12FA Urine Reagent Strips

TestExpected Value
Nitrite- (Negative)+ (Positive)
Exact agreement100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)
ConclusionQualifiedQualified
Leukocyte(leu/µL)-(0)+-(15)+1(70)+2(125)+3(500)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Ketone (mg/dL)-(0)+-(5)+1(15)+2(40)+3(80)
Exact agreement100% (60/60)100% (60/60)96.7% (58/60)96.7% (58/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Urobilinogen (EU/dL)Normal+1(2.0)+2(4.0)+3(8.0)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified
Bilirubin (mg/dL)-(0)+1(0.5)+2(2.0)+3(6.0)
Exact agreement100% (60/60)98.3% (59/60)91.7% (55/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified
Glucose(mg/dL)-(0)+-(50)+1(100)+2(250)+3(500)+4(1000)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Protein(mg/dL)-(0)+-(15)+1(30)+2(100)+3(300)
Exact agreement100% (60/60)100% (60/60)93.3% (56/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Specific Gravity1.0051.0101.0151.0201.0251.030
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)

{18}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China
TestExpected Value
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Blood(CELL/µL)-(0)+-(10)+1(25)+2(80)+3(200)
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
pH5.05.56.06.57.07.5
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
pH8.08.59.0
Exact agreement100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualified
Microalbumin10mg/L30mg/L80mg/L150mg/L
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified
Creatinine10 mg/dL50 mg/dL100 mg/dL200 mg/dL300 mg/dL
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 color block100% (60/60)100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified

URIT MEDICAL ELECTRONIC CO., LTD. District High-Tech Zone Guilin Gr

Table 10 Summary of Repeatability (With-in Run) Precision of Physical Module Item Turbidity

TestExpected Value
TurbidityClearMicro turbidTurbidVery turbid
Exact agreement100% (60/60)100% (60/60)100% (60/60)100% (60/60)
± 1 Block100% (60/60)100% (60/60)100% (60/60)100% (60/60)
ConclusionQualifiedQualifiedQualifiedQualified

Table 11 Summary of Repeatability (With-in Run) Precision of Physical Module Item Color

TestColorExact agreementConclusion
Expected ValueColorless100% (60/60)Qualified
Brown100% (60/60)Qualified
Yellow100% (60/60)Qualified
Red100% (60/60)Qualified
Green100% (60/60)Qualified
Other-Orange100% (60/60)Qualified
Other-Blue100% (60/60)Qualified
Other-Purple100% (60/60)Qualified

The Reproducibility (Between-Run) Precision of the URIT UC-1800 Automatic Urine Analyzer was evaluated by using negative urines and spiked urines of known concentrations for each analyte. Each sample was tested for 20days with 2 runs per day, in 1 time per run in 3 sites, each site with 1 instrument, totally the test performed across 3 instruments with the 3lots of URIT 11FA Urine Reagent Strips and 3lots of URIT 12FA Urine Reagent Strips in each instrument by 3 operators, each operator performing the test for 6-7 days, for a total of 120 measurements at each concentration were tested. All

{19}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China

concentrations for all tests had 120 of 120 (100%) match at ± one color block. The results are summarized in Table 12 to Table 15.

Table 12 Summary of Reproducibility (Between-Run) Precision of URIT 11FA Urine Reagent Strips

TestExpected Value
Ascorbic acid (mg/dL)-(0)+-(10)+1(25)+2(50)+3(100)
Exact agreement100% (120/120)100% (120/120)100% (120/120)94.2% (113/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Nitrite- (Negative)+ (Positive)
Exact agreement100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)
ConclusionQualifiedQualified
Leukocyte(leu/μL)-(0)+-(15)+1(70)+2(125)+3(500)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Ketone (mg/dL)-(0)+-(5)+1(15)+2(40)+3(80)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Urobilinogen (EU/dL)Normal+1(2.0)+2(4.0)+3(8.0)
Exact agreement100% (120/120)100% (120/120)98.3% (118/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified
Bilirubin (mg/dL)-(0)+1(0.5)+2(2.0)+3(6.0)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified
Glucose(mg/dL)-(0)+-(50)+1(100)+2(250)+3(500)+4(1000)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Protein(mg/dL)-(0)+-(15)+1(30)+2(100)+3(300)
Exact agreement100% (120/120)100% (120/120)100% (120/120)99.2% (119/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Specific Gravity1.0051.0101.0151.0201.0251.030
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Blood(CELL/μL)-(0)+-(10)+1(25)+2(80)+3(200)
Exact agreement100% (120/120)100% (120/120)100% (120/120)95% (114/120)97.5% (117/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
pH5.05.56.06.57.07.5
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)99.2% (119/120)100% (120/120)
TestExpected Value
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
pH8.08.59.0
Exact agreement100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualified

{20}------------------------------------------------

Table 13 Summary of Reproducibility (Between-Run) Precision of URIT 12FA Urine Reagent Strips

TestExpected Value
Nitrite- (Negative)+ (Positive)
Exact agreement100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)
ConclusionQualifiedQualified
Leukocyte(leu/µL)-(0)+-(15)+1(70)+2(125)+3(500)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Ketone (mg/dL)-(0)+-(5)+1(15)+2(40)+3(80)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Urobilinogen (EU/dL)Normal+1(2.0)+2(4.0)+3(8.0)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified
Bilirubin (mg/dL)-(0)+1(0.5)+2(2.0)+3(6.0)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified
Glucose(mg/dL)-(0)+-(50)+1(100)+2(250)+3(500)+4(1000)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Protein(mg/dL)-(0)+-(15)+1(30)+2(100)+3(300)
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified
Specific Gravity1.0051.0101.0151.0201.0251.030
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
Blood(CELL/µL)-(0)+-(10)+1(25)+2(80)+3(200)
Exact agreement100% (120/120)100% (120/120)100% (120/120)95.8% (115/120)99.2% (119/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified

{21}------------------------------------------------

TestExpected Value
pH5.05.56.06.57.07.5
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualified
pH8.08.59.0
Exact agreement100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualified
Microalbumin10 mg/L30 mg/L80 mg/L150 mg/L
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified
Creatinine10 mg/dL50 mg/dL100 mg/dL200 mg/dL300 mg/dL
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 color block100% (120/120)100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualifiedQualified

Table 14 Summary of Reproducibility (Between- Run) Precision of Physical Module Item Turbidity

TestExpected Value
TurbidityClearMicro turbidTurbidVery turbid
Exact agreement100% (120/120)100% (120/120)100% (120/120)100% (120/120)
± 1 Block100% (120/120)100% (120/120)100% (120/120)100% (120/120)
ConclusionQualifiedQualifiedQualifiedQualified

Table 15 Summary of Reproducibility (Between- Run) Precision of Physical Module Item Color

TestColorExact agreementConclusion
Expected ValueColorless100% (120/120)Qualified
Brown100% (120/120)Qualified
Yellow100% (120/120)Qualified
Red100% (120/120)Qualified
Green100% (120/120)Qualified
Other-Orange100% (120/120)Qualified
Other-Blue100% (120/120)Qualified
Other-Purple100% (120/120)Qualified

b Linearity/assay reportable range

The assay reportable range was evaluated by measuring the samples containing known concentrations of all measurement blocks. Prepare the reference solutions according to the marked values of concentration for each item on the URIT 11FA and 12FA urine reagent strips, and conduct linear range/reportable range experiments on the instrument with the reference solutions. Test each reference solution on three UC-1800 machines of reagent strips. Repeat the test for each batch of test strip for 21 times on each instrument.

The results of assay reportable range of URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips are summarized as Table 16 to Table 18 show below. The Summary of Reportable Results of URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips are summarized as Table 19 show below.

{22}------------------------------------------------

Table 16 Summary of Linearity of URIT 11/12FA Urine Reagent Strips -common analytes
URIT 11/12 FA Urine Reagent Strips (Common Item)
QualitativeSemi-QuantitativeConcentration/LevelAgreement at sameAgreement within + /- one
AnalyteRankRankTestedblockblock
Leukocyte-00 leu/mcL100% (63/63)100% (63/63)
+/-1515 leu/mcL100% (63/63)100% (63/63)
+17070 leu/mcL100% (63/63)100% (63/63)
+2125125 leu/mcL100% (63/63)100% (63/63)
+3500500 leu/mcL100% (63/63)100% (63/63)
Ketones-00 mg/dL100% (63/63)100% (63/63)
+/-55 mg/dL100% (63/63)100% (63/63)
+11515 mg/dL100% (63/63)100% (63/63)
+24040 mg/dL100% (63/63)100% (63/63)
+38080 mg/dL100% (63/63)100% (63/63)
UrobilinogenNormalNormalNormal100% (63/63)100% (63/63)
+12.02.0 mg/dL100% (63/63)100% (63/63)
+24.04.0 mg/dL100% (63/63)100% (63/63)
+38.08.0 mg/dL100% (63/63)100% (63/63)
Bilirubin-00 mg/dL100% (63/63)100% (63/63)
+10.50.5 mg/dL100% (63/63)100% (63/63)
+22.02.0 mg/dL100% (63/63)100% (63/63)
+36.06.0 mg/dL100% (63/63)100% (63/63)
Glucose-00 mg/dL100% (63/63)100% (63/63)
+/-5050 mg/dL100% (63/63)100% (63/63)
+1100100 mg/dL100% (63/63)100% (63/63)
+2250250 mg/dL100% (63/63)100% (63/63)
+3500500 mg/dL100% (63/63)100% (63/63)
+410001000 mg/dL100% (63/63)100% (63/63)
Protein-00 mg/dL100% (63/63)100% (63/63)
+/-1515 mg/dL100% (63/63)100% (63/63)
+13030 mg/dL100% (63/63)100% (63/63)
+2100100 mg/dL100% (63/63)100% (63/63)
+3300300 mg/dL100% (63/63)100% (63/63)
Blood-00 ery/mcL100% (63/63)100% (63/63)
+/-1010 ery/mcL100% (63/63)100% (63/63)
+12525 ery/mcL100% (63/63)100% (63/63)
+28080 ery/mcL100% (63/63)100% (63/63)
+3200200 ery/mcL100% (63/63)100% (63/63)
SpecificGravityN/A1.0051.005100% (63/63)100% (63/63)
1.0101.010100% (63/63)100% (63/63)
1.0151.015100% (63/63)100% (63/63)
1.0201.020100% (63/63)100% (63/63)
1.0251.025100% (63/63)100% (63/63)
1.0301.030100% (63/63)100% (63/63)
pHN/A5.05.0100% (63/63)100% (63/63)
URIT 11/12 FA Urine Reagent Strips (Common Item)
QualitativeRankSemi-QuantitativeRankConcentration/LevelTestedAgreement at sameblockAgreement within +/- oneblock
Analyte6.06.0100% (63/63)100% (63/63)
6.56.5100% (63/63)100% (63/63)
7.07.0100% (63/63)100% (63/63)
7.57.5100% (63/63)100% (63/63)
8.08.0100% (63/63)100% (63/63)
8.58.5100% (63/63)100% (63/63)
9.09.0100% (63/63)100% (63/63)

{23}------------------------------------------------

Table 17 Summary of Linearity of URIT 11FA Urine Reagent Strips

URIT 11 FA Urine Reagent Strips
AnalyteQualitativeRankSemi-QuantitativeRankConcentration/LevelTestedAgreement at sameblockAgreement within +/- oneblock
Ascorbicacid-00mg/dL100% (63/63)100% (63/63)
+11010mg/dL100% (63/63)100% (63/63)
+22525mg/dL100% (63/63)100% (63/63)
+35050mg/dL100% (63/63)100% (63/63)
+4100100mg/dL100% (63/63)100% (63/63)

Table 18 Summary of Report range of URIT 12FA Urine Reagent Strips

URIT 12FA Urine Reagent Strips (Microalbumin & Creatinine)
AnalyteSemi-Quantitative RankConcentration/Level TestedAgreement at same blockAgreement within +/- one block
Microalbumin1010mg/L100% (63/63)100% (63/63)
3030mg/L100% (63/63)100% (63/63)
8080mg/L100% (63/63)100% (63/63)
150150mg/L100% (63/63)100% (63/63)
Creatinine1010mg/dL100% (63/63)100% (63/63)
5050mg/dL100% (63/63)100% (63/63)
100100mg/dL100% (63/63)100% (63/63)
200200mg/dL100% (63/63)100% (63/63)
300300mg/dL100% (63/63)100% (63/63)

Table 19 Summary of Reportable Results of URIT 11/12FA Urine Reagent Strips

ItemURIT Urine Reagent Strips
11FA12FA
Ascorbic acid0, 10, 25, 50, 100 mg/dL-, +/-, +1, +2, +3N/A
MicroalbuminN/A10,30,80,150 mg/L
Leukocyte0, 15, 70, 125, 500 leu/mcL-, +/-, +1, +2, +3
CreatinineN/A10,50,100,200,300 mg/dL
Ketone0,5,15,40,80 mg/dL-, +/-, +1, +2, +3
UrobilinogenNormal,2.0,4.0,8.0 EU/dLNormal, +1, +2, +3

{24}------------------------------------------------

ItemURIT Urine Reagent Strips
11FA12FA
Bilirubin0,0.5,2.0,6.0mg/dL-, +1, +2, +3
Glucose0,50,100,250,500,1000 mg/dL-, +/-, +1, +2, +3, +4
Protein0,15,30,100,300 mg/dL-, +/-, +1, +2, +3
Blood0, 10,25,80,200 ery/mcL-, +/-, +1, +2, +3
Specific Gravity1.005, 1.010, 1.015, 1.020, 1.025, 1.030
pH5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0

c

The results of Analytical Sensitivity of URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips are summarized as Table 20.

Table 20 Summary of Analytical Sensitivity of Urine Reagent Strips
ItemURIT Urine Reagent Strips
11FA12FA
Ascorbic acid8~10mg/dLN/A
Nitrite0.1-0.2mg/dL
MicroalbuminN/A20-30mg/L
Leukocyte10~15leu/mcL
CreatinineN/A25~50mg/dL
Ketone4~5mg/dL
Urobilinogen1~2EU/dL
Bilirubin0.4~0.5mg/dL
Glucose40~50mg/dL
Protein10~15mg/dL
Blood5~10ery/mcL

Analytical sensitivity is not applicable for both strips in detection of pH and Specific Gravity

The results of Critical value of URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips are summarized as Table 21 show.

Table 21 Summary of Critical value of Urine Reagent Strips
AnalyteConcentrationCritical valueAgreement rate%
11FA12FA
Ascorbic acid25 (1+)12 mg/dL76
Ascorbic acid50 (2+)32 mg/dL89N/A
Ascorbic acid100 (3+)72 mg/dL66
Microalbumin85.2mg/dLN/A88
Microalbumin1511.3mg/dL79
Leukocyte70(1+)38 leu/mcL9274
Leukocyte125(2+)88 leu/mcL8564
Leukocyte500 (3+)337 leu/mcL7769
Creatinine10074mg/dL77
Creatinine200142mg/dLN/A75
Creatinine300239mg/dL79

{25}------------------------------------------------

AnalyteConcentrationCritical valueAgreement rate%
Ketone15 (1+)8 mg/dL8274
Ketone40 (2+)25 mg/dL8276
Ketone80(3+)63 mg/dL7271
Urobilinogen4 (2+)1.8 EU/dL8177
Urobilinogen8 (3+)4.5 EU/dL7174
Bilirubin2(2+)1.3 mg/dL8878
Bilirubin6(3+)3 mg/dL7973
Glucose100 (1+)75 mg/dL8673
Glucose250 (2+)165 mg/dL7774
Glucose500 (3+)360 mg/dL6676
Glucose1000 (4+)735 mg/dL7174
Protein30 (1+)21 mg/dL7783
Protein100 (2+)65 mg/dL7570
Protein300 (3+)180 mg/dL7970
Blood25 (1+)16 ery/mcL7977
Blood80 (2+)65 ery/mcL7779
Blood200 (3+)130 ery/mcL7971
pH5.55.28373
pH65.88679
pH6.56.38277
pH7.06.78178
pH7.57.37682
pH8.07.88276
pH8.58.27294
pH9.08.86877

{26}------------------------------------------------

d Analytical specificity

d.1 Exogenous and Endogenous Interference

Select urine sample and mix it well to prepare negative samples for each dry chemistry item. Prepare the first positive-grade sample (excluding pH and specific gravity) by adding the substance to be tested, Dissolve the interfering substance with an appropriate solvent according to the chemical properties of the interfering substance. The concentration of the stock solution shall be appropriate to reduce dilution of the basic sample matrix. The dilution ratio of the sample matrix shall not exceed 5%, Absorb an appropriate volume of interfering substance stock solution according to a certain proportion and add it into the volumetric flask, and then fix to the scale with the basic sample to prepare a sample containing a single interfering substance. The concentration of interfering substances in different samples is shown in Table 22.

Interfering substanceConcentrationInterfering substanceConcentrationInterfering substanceConcentration
Exogenous interfering substancesEndogenous interfering substances
Acetaminophen300 mg/dLQuaternary Ammonium200mg/dLHydroxybutyrate dehydrogenase450 mg/dL
Amoxicillin1333 mg/dLLithium acetoacetate250 mg/dLAmmonium chloride2500 mg/dL
Ascorbic acid400 mg/dLPeroxide10%Bilirubin170 mg/dL
Biotin100 mg/dLPeroxidase20mg/dLCalcium chloride300 mg/dL
Cefoxitin1200 mg/dLSodium thiosulfate20mg/dLCreatinine1500 mg/dL
Furosemide200 mg/dLPotassium chloride1500 mg/dLGlucose10000mg/dL
Gabapentin1200 mg/dLSodium Bicarbonate1500 mg/dLHGB83 mg/dL
Gentamicin sulphate40 mg/dLCitric acid150 mg/dLProtein500 mg/dL
Ibuprofen250 mg/dLCreatine10 mg/dLNitrite11 mg/dL
Levodopa125 mg/dLFructose100mg/dLUrea20000 mg/dL
Lisinopril27 mg/dLGalactose80 mg/dLUrea acid155mg/dL
Metformin850 mg/dLOxalic acid70mg/dLUrobilinogen300mg/dL
Methyldopa200 mg/dLVitamin B10mg/dLHuman immunoglobulin IgG500mg/dL
Methylamine + methylene blue400 + 66.5 mg/dLSodium acetate2.25mg/dLLeukocyte2500leu/uL
Acetylcysteine20 mg/dLSodium chloride5500mg/dLCysteine20mg/dL
Ofloxacin90 mg/dLSodium nitrate10mg/dLBlood0.05%
Phenazopyridine30 mg/dLSodium phosphate500mg/dLGlycine450 mg/dL
Salicylic acid600 mg/dLTheophylline100 mg/dLLactose10mg/dL
Tetracycline50 mg/dLLeukocyte2500leu/uL
Table 22 Interfering substance and concentration
--------------------------------------------------------------

Add the solvent of the same volume as the experimental sample (the solvent used to prepare the stock solution) into the second volumetric flask and fix to the scale with the basic sample (experimental sample and control sample) shall be tested for five times on two UC-1800 machines, two batch numbers of 11FA and 12FA reagent strips. Results are summarized in the tables below.

1. The substances show no interference for all analytes

The following substances, i.e. Cefaxitin, Phenazopyridine, Salicylic acid, Tetracycline, Hydroxybutyrate delydrogenase, Protein, Lactose, Leukocyte, Potassium chloride, Citric acid, Creatine, Oxalic acid, Vitamin B, Sodium acetate, Sodium chloride, Sodium nitrate, Theophylline show no interference for all analytes at tested are summarized in Table 23.

Interfering substanceMaximum concentrationwithout interference
Cefoxitin1200 mg/dL
Ofloxacin90 mg/dL

Table 23 The substances that do not cause interference

{27}------------------------------------------------

Interfering substanceMaximum concentrationwithout interference
Phenazopyridine30 mg/dL
Salicylic acid600 mg/dL
Tetracycline50 mg/dL
Hydroxybutyrate dehydrogenase450 mg/dL
Protein500 mg/dL
Lactose10mg/dL
Leukocyte2500leu/uL
Potassium chloride1500 mg/dL
Citric acid150 mg/dL
Creatine10mg/dL
Fructose100mg/dL
Galactose80 mg/dL
Oxalic acid70mg/dL
Vitamin B10mg/dL
Sodium acetate2.25mg/dL
Sodium chloride5500mg/dL
Sodium nitrate10mg/dL
Theophylline100 mg/dL

{28}------------------------------------------------

2.The substances that cause interference

For those substances that on initial screening with certain analytes, dose response testing was conducted to establish the concentration limitible) which no significant interference is expected. The results are given in the tables below

Table 24 The substances that cause interference

AnalyteInterfering substancesMaximum concentrationwithout interference (mg/dL)Interferenceconcentration (mg/dL)Specific interference effectStrip
LeukocyteUrobilinogen2550From - to +- (False Positive)
Amoxicillin7001050From +- to 1 (elevated positive result)
Ibuprofen125187.5From +- to 2 (elevated positive result)
Methylamine + methylene blue200+35300+52.5From - to 1 (False Positive)URIT 11/12 FAUrine ReagentStrips
Bilirubin6080From +- to - (False Negative)
Glucose15001666.7From +- to - (False Negative)
Sodium Bicarbonate375750From +- to - (False Negative)
Glycine225337.5From +- to - (False Negative)
Sodium phosphate250375From +- to - (False Negative)
KetonesMethyldopa50100From +- to - (False Negative)
Methylamine + methylene blue100+17.5150From - to +- (False Positive)
Acetylcysteine3.3200+35From +- to 1 (elevated positive result)URIT 11/12 FAUrine ReagentStrips
Ammonium chloride12505From +- to - (False Negative)
Bilirubin605From - to +- (False Positive)
Creatinine11251875From +- to - (False Negative)
Gabapentin1580From +- to - (False Negative)
UrobilinogenMethylamine + methylene blue66.7+11.71500From - to +- (False Positive)
Bilirubin4022.5From 1 to Normal (False Negative)
Nitrite0.8200+35From Normal to 1 (False Positive)URIT 11/12 FAUrine ReagentStrips
BilirubinUrobilinogen12.5100+17.5From 1 to 2 (elevated positive result)
Ascorbic acid15060From 1+ to 2+ (elevated positive result)
Methylamine + methylene blue200+351.7From 1+ to Normal (False Negative)
Nitrite525From - to 1 (False Positive)URIT 11/12 FAUrine ReagentStrips
GlucoseLithium acetoacetate8050From - to 2 (False Positive)
Peroxide5%25From 1 to 2 (elevated positive result)
Ascorbic acid50200From 1 to - (False Negative)
Levodopa10.8300+52.5From 1 to - (False Negative)
Methylamine + methylene blue100+17.510From 1 to - (False Negative)
Bilirubin40125From +- to - (False Negative)
ProteinUrea100507.50%From - to +- (False Positive)URIT 11/12 FAUrine ReagentStrips
Quaternary Ammonium507.50%From +- to 1 (elevated positive result)
Sodium Bicarbonate750100From +- to - (False Negative)
Amoxicillin70021.7From +- to - (False Negative)
Gabapentin7.5200+35From +- to - (False Negative)
Ibuprofen62.560From +- to - (False Negative)
Methylamine + methylene blue66.7+11.715025From +- to - (False Negative)
Ammonium chloride625100From - to +- (False Positive)
Bilirubin40150From - to 1 (False Positive)
Calcium chloride150100From +- to 1 (elevated positive result)
Creatinine375150From +- to 2 (elevated positive result)
HGB12501125From +- to 1 (elevated positive result)
Urea100501050From +- to - (False Negative)
15From +- to 1+ (elevated positive result)
125From +- to - (False Negative)
200+35From - to +- (False Positive)
100+17.5From +- to 1+ (elevated positive result)URIT 11/12 FAUrine ReagentStrips
1250From +- to - (False Negative)
60From - to +- (False Positive)
225From +- to - (False Negative)
1500From - to +-/1+ (False Positive and elevatedpositive result)
750From +- to 1+ (elevated positive result)
3750From - to +-/1+ (False Positive and elevatedpositive result)
2500From +- to 1+ (elevated positive result)
15025From +- to 1+ (elevated positive result)
HGB125037502500From - to +/-/1+ (False Positive and elevated positive result)From +- to 1+ ( elevated positive result)
Urea1005015025From +- to 1+ ( elevated positive result)

{29}------------------------------------------------

AnalyteInterfering substancesMaximum concentrationwithout interference (mg/dL)Interferenceconcentration (mg/dL)Specific interference effectStrip
BloodPeroxidase510From - to +- (False Positive)
Sodium Bicarbonate375750From +- to 1 (elevated positive result)
Glycine112.5225From +- to - (False Negative)
Ascorbic acid50100From +- to 1 ( elevated positive result)
Biotin25003750From +- to - (False Negative)
Furosemide50100From - to +- (False Positive)URIT 11/12 FAUrine ReagentStrips
Ibuprofen187.5250From +- to - (False Negative)
Levodopa10.821.7From +- to - (False Negative)
Methyldopa16.733.3From - to +- (False Positive)
Methyldopa16.7150From +- to 1+ ( elevated positive result)
Methylamine + methylene blue66.7+11.7100+17.5From - to +- (False Positive)From +- to 1+ ( elevated positive result)
Specific GravitySalicylic acid300450600From 1.010 to 1.020 (Result rise)From 1.010 to 1.030 (Result rise)URIT 11/12 FAUrine ReagentStrips
Uric acid116.25155From 1.015 to 1.025 (Result rise)
Sodium bicarbonate11251500From 1.020 to 1.005 (Result reduction)
Urobilinogen12.525From - to + (False Positive)
NitriteSodium Bicarbonate7501125From + to - (False Negative)
Ascorbic acid50100From + to - (False Negative)URIT 11/12 FAUrine ReagentStrips
Methylamine + methylene blue200+35300+52.5From - to + (False Positive)
Bilirubin6080From + to - (False Negative)
Creatinine7501125From + to - (False Negative)
Ascorbic acidSodium thiosulfate1015From - to +- (False Positive)From +- to 1 ( elevated positive result)
Cysteine1015From - to +- (False Positive)From +- to 1 ( elevated positive result)URIT 11 FAUrine ReagentStrips
Sodium phosphate250375From +- to 1 ( elevated positive result)
Levodopa5.410.8From +- to 1+ (elevated positive result)
Acetylcysteine1015From +- to 1+ (elevated positive result)
Ammonium chloride18752500From +- to 1+ (elevated positive result)
Quaternary Ammonium50100From 10mg/L to 30mg/L (False Positive)From 30mg/L to 80mg/L ( elevated positive result)
MicroalbuminBlood0.0375%0.05%From 10mg/L to 30mg/L (False Positive)From 30mg/L to 80mg/L ( elevated positive result)
Human immunoglobulin lgG41.6783.33From 10mg/L to 30mg/L (False Positive)From 10mg/L to 80mg/L (False Positive)From 30mg/L to 80mg/L ( elevated positive result)From 30mg/Lto 150mg/L( elevated positive result)
Ascorbic acid200300From 30 mg/L to 10 mg/L (False Negative)URIT 12 FAUrine ReagentStrips
Gabapentin7.530From 10 mg/L to 30 mg/L (False Positive)From 30 mg/L to 80 mg/L ( elevated positive result)
Methylamine + methylene blue66.7+11.7100 + 17.5From 10mg/L to 30mg/L (False Positive)From 30mg/L to 80mg/L ( elevated positive result)
Ammonium chloride6251250From30mg/L to 10mg/L (False Negative)From10mg/L to 30mg/L (False Positive)
Creatinine6001500From30mg/L to 80mg/L ( elevated positive result)
HGB83750830From10mg/L to 30mg/L (False Positive)From30mg/L to 80mg/L ( elevated positive result)
Acetaminophen225208300From 50 mg/dL to 100 mg/dL ( elevated positive result
Biotin8301250From 50 mg/dL to 100 mg/dL(elevated positive result)
CreatinineFurosemide100150From 50 mg/dL to 100 mg/dL (elevated positive result)URIT 12 FA
Gabapentin1522.5From 50 mg/dL to 100 mg/dL (elevated positive result)Urine Reagent
Gentamicin sulphate2030From 50 mg/dL to 10 mg/dL (False Negative)Strips
Acetylcysteine3.3200From 10 mg/dL to 50 mg/dL (False Positive)
5From 50 mg/dL to 100 mg/dL (elevated positive result)
Ammonium chloride100104.2From 50 mg/dL to 10 mg/dL (False Negative)

{30}------------------------------------------------

d.2 Effect of urine pH

Collect mixed negative urine samples and divise their pH values with 1M hydrochloric acid aqueous solution and 1M sodium hydroxide aquous solution to prepare samples with pH value of 4.5, 5.5, 6.5, 7.5, and 8.5, respectively. Each sample shall be tested for 5 times on 2 UC-1800 machines, two batch numbers of 11FA and 12FA test strips, respectively For all test strip and specific gravity, For leukorytes, unine pH values lover than 5.5 may yield a false negative result. For specific result. For spec No significant interference was observed for the urine pH ranges from 5.5 to 7.5, Urine pH values lower than 5.5 may yield a false increased of Specific Gravity, and urine pH values higher than 7.5 may yield decreased of specific gravity. The results were summarized as Table 25 shows.

Table 25 Summary of effect of urine pH
URIT 11/12 FA Urine Reagent Strips
ItemNo interference pHInterference condition
LeukocytespH>5.5pH=4.5, From +- to - (False Negative)
Specific Gravity5.5-7.5pH=4.5, From 1.015 to 1.025 (False Positive)
pH=4.5, From 1.020 to 1.030 (False Positive)
pH≥8.5, From 1.015 to 1.005 (False Negative)
pH≥8.5, From 1.020 to 1.010 (False Negative)

d.3 Effect of urine color

Collect mixed negative une samples and divide them into 3 parts tellow, ligin, lemon yellow, Naphthol green, Bright blue and Crystal violet to prepare the samples into the red, orage, brown, yellow purple samples ball be tested for 5 times on two UC-1800 machines of 11FA and 1274 test stips, respectively. All the results dentified that all color showed no interference on each and and a colle 26 to Table 31 shows.

URIT 11/12FA Urine Reagent Strips
AnalyteConcentrationRedOrangeBrown
HGBConcentrationResultSunset yellowConcentrationResultLigninConcentrationResult
Leukocyte- (0 leu/mcL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+/- (15 leu/mcL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Ketones- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+/- (5 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
UrobilinogenNormal (Normal)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
1 (2.0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Bilirubin- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
1 (0.5 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Glucose- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+/- (50 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Protein- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+/- (15 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Blood- (0 ery/mcL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+/- (10 ery/mcL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Specific Gravity1.015 (1.015)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
1.020 (1.020)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
pH6.5 (6.5)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
7.5 (7.5)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Nitrite- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
+ (0.5 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference

Table 26 Summary of effect of urine color - Red, Orange, Brown (11/12FA)

Table 27 Summary of effect of urine color – Yellow, Green, Blue, Purple (11/12FA)

URIT 11/12FA Urine Reagent Strips
YellowGreenBluePurple
AnalyteConcentrationLemon yellowConcentrationResultNaphthol greenConcentrationResultBright blueConcentrationResultCrystal violetConcentrationResult
- (0 leu/mcL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Leukocyte+/- (15 leu/mcL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Ketones+/- (5 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Normal (Normal)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Urobilinogen1 (2.0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Bilirubin1 (0.5 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Glucose+/- (50 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Protein+/- (15 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 ery/mcL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Blood+/- (10 ery/mcL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
1.015 (1.015)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Specific Gravity1.02 (1.02)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
pH6.5 (6.5)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
7.5 (7.5)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Nitrite+ (0.5 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference

{31}------------------------------------------------

RedOrangeBrown
AnalyteConcentrationHGBConcentrationResultSunset yellowConcentrationResultLigninConcentrationResult
Ascorbic acid- (0 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Ascorbic acid+/- (10 mg/dL)0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference

Table 29 Summary of effect of urine color – Yellow, Green, Blue, Purple (11FA)

AnalyteConcentrationYellowGreenBluePurple
Lemon yellowConcentrationResultNaphthol greenConcentrationResultBright blueConcentrationResultCrystal violetConcentrationResult
Ascorbic acid- (0 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
+/- (10 mg/dL)0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference

Table 30 Summary of effect of urine color - Red, Orange, Brown (12FA)

AnalyteConcentrationRedOrangeBrown
HGB ConcentrationResultSunset yellow ConcentrationResultLignin ConcentrationResult
Microalbumin0 mg/L0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Microalbumin30 mg/L0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Creatinine0 mg/dL0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference
Creatinine50 mg/dL0,750,300 mg/LNo interference0,312.5,150 mg/LNo interference0;1;0.4 mg/LNo interference

Table 31 Summary of effect of urine color – Yellow, Green, Blue, Purple (12FA)

ConcentrationYellowGreenBluePurple
AnalyteLemon yellowResultNaphthol greenResultBright blueResultCrystal violetResult
ConcentrationConcentrationConcentrationConcentration
Microalbumin0 mg/L0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
30 mg/L0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
Creatinine0 mg/dL0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference
50 mg/dL0;20;3 mg/LNo interference0;40;20 mg/LNo interference0;10;1 mg/LNo interference0;10;3 mg/LNo interference

d.1 Effect of urine Specific Gravity

Collect mixed negative urine samples and divide to adjust their specific gravity to prepare the samples with specific gravity of 1.005, 1.100, 1.015, 1.130, 1.135, 1.130, 1.045, and 1.150 respective); Each sample shall be tested for 5 times on two UC-800 machines, two batch numbers of 11FA and 12H test strips.

Table 32 Summary of Effect of urine Specific Gravity

URIT 11/12 FA Urine Reagent Strips
ItemNo interference SGInterference situation
LeukocytesSG<1.035SG≥1.040, From +- to- (false negative)

{32}------------------------------------------------

e Assay Cut-off

The results of assay Cut-off of URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips are summarized as Table 33.

Table 33 Summary of Analytical Sensitivity of Urine Reagent Strips
URIT 11FA Urine Reagent StripsURIT 12FA Urine Reagent Strips
AnalyteConcentrationCut-offAgreement%Cut-offAgreement%
Glucosemg/dL100 (1+)75867573
250 (2+)1657716574
500 (3+)3606636076
1000 (4+)7357173574
Bloodery/mcL25 (1+)16791677
80 (2+)65776579
200 (3+)1307913071
UrobilinogenEU/dL4 (2+)1.8811.877
8 (3+)4.5714.574
Ketonesmg/dL15 (2+)882874
40(3+)25822576
80(4+)63726371
Proteinmg/dL30 (1+)21772183
100 (2+)65756570
300 (3+)1807918070
Bilirubinmg/dL2(2+)1.3881.378
6(3+)379373
Leukocyteleu/mcL70(1+)38923874
125(2+)88858864
500 (3+)3377733769
pH5.55.2835.273
6.05.8865.879
6.56.3826.377
7.06.7816.778
7.57.3767.382
8.07.8827.876
8.58.2728.294
9.08.8688.877
Ascorbic acidmg/dL25 (1+)1276N/AN/A
50 (2+)3289N/AN/A
100 (3+)7266N/AN/A
Creatinine(mg/dL)100N/AN/A6.577
200N/AN/A12.575
300N/AN/A2179
Microalbuminmg/L80N/AN/A5288
150N/AN/A11379

f Carryover

Evaluate the carryover by alternately testing the high-concentration samples (UQ14 NO. II/III) and low-concentration samples (UQ14 NO. I), namely, conduct tests in the sequence of high, low, high, low, high, low, high and low. For all

{33}------------------------------------------------

analytes, all negative/normal samples are read to be negative/normal. UC-1800 automatic urine analyzer with 11FA, 12FA test strip test carryover verification meets the requirements (All low-value samples are negative). The results of carryover presented in Table 34 below.

Test strip typeURIT 11FAUrine Reagent StripsURIT 12FAUrine Reagent Strips
NitriteQualifiedQualified
LeukocytesQualifiedQualified
KetoneQualifiedQualified
UrobilinogenQualifiedQualified
BilirubinQualifiedQualified
GlucoseQualifiedQualified
ProteinQualifiedQualified
BloodQualifiedQualified
Ascorbic acidQualifiedN/A
MicroalbuminN/AQualified
CreatinineN/AQualified

Table 34 Summary of Carryover of Urine Reagent Strips

g Reference Range

The Reference Range Result were confirmed as Table 35 shows below and listed in the IFU of Urine Reagent Strips.

ItemReference value (reference interval)ItemReference value (reference interval)
LeukocytesNegativeSpecific Gravity1.003~1.035
KetoneNegativepH4.5~8.0
NitriteNegativeBloodNegative
Urobilinogen(0.2~1.0) EU/dLGlucoseNegative
BilirubinNegativeCreatinine(10~300)mg/dL
ProteinNegativeMicroalbumin<20 mg/L
Ascorbic acid2-10 mg/dL//

Table 35 Reference Range Result of Urine Reagent Strips

The Reference Ranges are included in the labeling and are taken from literature references:

  1. Shchersten B, Fritz H. Subnormal Levels of Glucose in Urine. JAMA 201:129-132, 1967.

  2. McGarry JD, Lilly: Lecture, 1978: New Perspectives in the Regulation of Ketogenesis. Diabetes 28: 517-523 May, 1978.

  3. Williamson DH. Physiological Ketoses, or Why Ketone Bodies? Postgrad. Med. J. (June Suppl.): 372-375, 1971.

  4. Paterson P, et al. Maternal and Fetal Ketone Concentrations in Plasma and Urine. Lancet: 862-865; April 22, 1967.

  5. Fraser |, et al. Studies with a Simplified Nitroprusside Test for Ketone Bodies in Urine, Serum, Plasma and Milk. Chin. Chem. Acta II: 372-378, 1965.

  6. Henry JB, et al. Clinical Diagnosis and Management by Laboratory Methods, 20th Ed. Philadelphia. Saunders. 371-372, 375, 379, 382, 385, 2001.

  7. Tietz NW. Clinical Guide to Laboratory Tests. W.B. Saunders Company. 1976.

  8. Position Statement: Diabetic Nephropathy. Diabetes Care 20: S24-S27; 1997.

  9. Burtis, C.A. and Ashwood, E.R.: Tietz Textbook of Clinical Chemistry, 3rd ed. Philadelphia: Saunders; 1999; pp. 483-484.

  10. Mangili, R. et al.: Prevalence of Hypertension and Microalbuminuria in Adult Type 1 (Insulin Dependent) Diabetic Patients without Penal Failure in Italy - Validation of Screening techniques to detect microalbuminuria, Acta Diabetol. 29: 156-166; 1992.

  11. American Diabetes Association, Clinical Practice Recommendations, Diabetes Care, Vol. 31, Suppl. 1, January 2008.

{34}------------------------------------------------

h Traceability

Each assay of the UC-1800 Automatic urinalysis test System is traceable to the following standards.
NoAnalyteReference MethodStandardization
1Ascorbic Acid2,6-Dichlorophenolindophenol TitrationMethodAscorbic Acid
2NitriteDiazotization colorimetric methodSodium nitrite
3LeukocytesWBC counting methodLeukocyte esterase
4KetoneLange methodLithium acetoacetate
5UrobilinogenEhrlich acetaldehyde test tube reactionmethod2,5-dimethylpyrrole
6BilirubinBiochemical testing/Dichloroaniline DiazoSalt MethodBilirubin standardBilirubin substitute TB
7Glucoseglucose oxidase methodGlucose purity standard substance,glucosum anhydricum
8ProteinHuman serum albumin standard substance,Bovine serum albuminHuman serum albumin standardsubstance, Bovine serum albumin
9BloodErythrometryBovine hemoglobin albumin
10SpecificGravitySpecific gravity measurementSpecific gravity standard buffer
11pHAcidometer measurementpH buffer solution (pH 4.0, 6.86, 9.18)
12MicroalbuminBiochemical testBiochemical immunoturbidimetryHuman serum albumin
13CreatinineBiochemical testBiochemical immunoturbidimetryCreatinine standard
14TurbidityDirect turbidimeter measurement methodTurbidity standard buffer
15ColorDirect method of measurement of Sysmex UC3500Pigments

{35}------------------------------------------------

8.2 Comparison Studies

Method comparison with predicate device

a.1 Comparison with Automatic Urine Analyzer URIT-500B([K082811](#K082811))

A total of 1000 clinical urine samples were collected for the experiment of comparison with Automatic Urine Analyzer

URIT-500B(K082811), and the comparison results presented in the tables below.

Ascorbic acid (N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer(11FA test strips)
-(0)+-(10)+1(25)+2(50)+3(100)
-(0)7160000
Predicate device(K082811)Uritest-500B urine analyzer(11G test strip)+-(10)095000
+1(25)009810
+2(50)001401
+3(100)000048
Total71695994149
Complete agreement rate100%100%99.00%97.60%98.00%
General agreement rate100%100%100%100%100%
Leukocyte (N=1000) (Leu/mcL)URIT UC-1800 Automatic Urine Analyzer
(11FA test strips)
-(0)+-(15)+1(70)+2(125)+3(500)
Predicate device (K082811)-(0)5324000
Predicate device (K082811)+-(15)3205100
Uritest-500B urine analyzer+1(70)0010720
(11G test strip)+2(125)001690
+3(500)000175
Total5352091097275
Complete agreement rate99.40%98.10%98.20%95.80%100%
General agreement rate100%100%100%100%100%
Ketone(N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer
(11FA test strips)
-(0)+-(5)+1(15)+2(40)+3(80)
-(0)8280000
Predicate device (K082811)+-(5)145100
Uritest-500B urine analyzer+1(15)003710
(11G test strip)+2(40)000371
+3(80)000049
Total82945383850
Complete agreement rate99.90%100%97.40%97.40%98.00%
General agreement rate100%100%100%100%100%

{36}------------------------------------------------

URIT UC-1800 Automatic Urine Analyzer(11FA test strips)
Nitrite (N=1000)+-
Predicate device (K082811)+2163
Uritest-500B urine analyzer(11G test strip)-3778
Total219781
Complete agreement rate98.60%99.60%
95% confidence interval
Overall agreement rate OPA99.40%99.72%
Positive percentage agreement rate PPA98.63%99.53%
Negative percentage agreement rate NPA99.62%99.87%
URIT UC-1800 Automatic Urine Analyzer
Urobilinogen(N=1000) (EU/dL)(11FA test strips)
Normal+1(2.0)+2(4.0)+3(8.0)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)Normal900100
+1(2.0)14210
+2(4.0)00290
+3(8.0)00026
Total901433026
Complete agreement rate99.90%97.70%96.70%100%
General agreement rate100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
Bilirubin(N=1000) (mg/dL)(11FA test strips)
-(0)+1(0.5)+2(2.0)+3(6.0)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)-(0)931000
+1(0.5)12500
+2(2.0)00200
+3(6.0)00023
Total932252023
Complete agreement rate99.90%100%100%100%
General agreement rate100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
Blood (N=1000) (ery/mcL)(11FA test strips)
-(0)+-(10)+1(25)+2(80)+3(200)
5790000
Predicate device (K082811)0160100
Uritest-500B urine analyzer0212000
(11G test strip)000550
000083
Total5791621215583
Complete agreement rate100%98.80%99.20%100%100%
General agreement rate100%100%100%100%100%

{37}------------------------------------------------

Glucose (N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer
(11FA test strips)
-(0)+-(50)+1(100)+2(250)+3(500)+3(1000)
-(0)81000000
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)+-(50)1421000
+1(100)0026000
+2(250)0001900
+3(500)0000301
+3(1000)0000070
Total8114227193071
Complete agreement rate99.90%100%96.30%100%100%98.60%
General agreement rate100%100%100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
Protein (N=1000) (mg/dL)(11FA test strips)
-(0)+-(15)+1(30)+2(100)+3(300)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)-(0)6981000
+-(15)095100
+1(30)006400
+2(100)001601
+3(300)000059
Total69896666060
Complete agreement rate100%99.00%97.00%100%98.30%
General agreement rate100%100%100%100%100%
PH(N=1000)URIT UC-1800 Automatic Urine Analyzer(11FA test strips)
5.05.56.06.57.07.58.08.59.0
5.016120000000
5.502720000000
6.001212000000
Predicate device (K082811)6.500113410000
Uritest-500B urine analyzer(11G test strip)7.000031110000
7.500000048000
8.000000002700
8.500000000200
9.00000000007
Total1612752131371124827207
Complete agreement rate100%98.90%99.50%97.80%99.10%100%100%100%100%
General agreement rate100%100%100%100%100%100%100%100%100%

{38}------------------------------------------------

SG(N=1000)URIT UC-1800 Automatic Urine Analyzer
(11FA test strips)
1.0051.0101.0151.0201.0251.030
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)1.0053300000
1.01001343000
1.01502212000
1.02000027300
1.02500012020
1.03000002138
Total33136215274204138
Complete agreement rate100%98.50%98.60%99.60%99.00%100%
General agreement rate100%100%100%100%100%100%
Leukocyte (N=1000) (Leu/mcL)URIT UC-1800 Automatic Urine Analyzer(12FA test strips)
-(0)+-(15)+1(70)+2(125)+3(500)
05324000
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)+-(15)3205100
+1(70)0010630
+2(125)001690
+3(500)000076
Total5352091087276
Complete agreement rate99.40%98.10%98.10%95.80%100%
General agreement rate100%100%100%100%100%
Ketone(N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer(12FA test strips)
-(0)+-(5)+1(15)+2(40)+3(80)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)-(0)8280000
+-(5)145100
+1(15)003710
+2(40)000371
+3(80)000049
Total82945383850
Complete agreement rate99.90%100%97.40%97.40%98.00%
General agreement rate100%100%100%100%100%
Nitrite (N=1000)URIT UC-1800 Automatic Urine Analyzer(12FA test strips)
+-
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)+2163
-3778
Total219781
Complete agreement rate98.60%99.60%
95% confidence interval

{39}------------------------------------------------

URIT MEDICAL ELECTRONIC CO., LTD.

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China
Overall agreement rate OPA99.40%98.70%99.72%
Positive percentage agreement rate PPA98.63%96.05%99.53%
Negative percentage agreement rate NPA99.62%98.88%99.87%
Urobilinogen(N=1000) (EU/dL)URIT UC-1800 Automatic Urine Analyzer
(12FA test strips)
Normal+1(2.0)+2(4.0)+3(8.0)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)Normal900100
+1(2.0)14210
+2(4.0)01280
+3(8.0)00026
Total901442926
Complete agreement rate99.90%95.50%96.60%100%
General agreement rate100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
Bilirubin(N=1000) (mg/dL)(12FA test strips)
-(0)+1(0.5)+2(2.0)+3(6.0)
Predicate device (K082811)Uritest-500B urine analyzer-(0)931000
+1(0.5)12410
+2(2.0)00200
+3(6.0)00023
(11G test strip)Total932242123
Complete agreement rate99.90%100%95.20%100%
General agreement rate100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
(12FA test strips)
-(0)+-(10)+1(25)+2(80)+3(200)
-(0)5790000
Predicate device (K082811)+-(10)0160100
Uritest-500B urine analyzer(11G test strip)+1(25)0212000
+2(80)000550
+3(200)000083
Total5791621215583
Complete agreement rate100%98.80%99.20%100%100%
General agreement rate100%100%100%100%100%
Glucose (N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer(12FA test strips)
-(0)+-(50)+1(100)+2(250)+3(500)+3(1000)
Predicate device (K082811)-(0)81000000
+-(50)1421000
Uritest-500B urine analyzer(11G test strip)+1(100)0026000
+2(250)0001900
+3(500)0000301

{40}------------------------------------------------

URIT MEDICAL ELECTRONIC CO., LTD. No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China +3(1000) 0 0 0 0 1 69 27 Total 811 42 19 31 70 99.90% 100% 96.30% 96.80% 100% 98.60% Complete agreement rate 100% General agreement rate 100% 100% 100% 100% 100%

Protein (N=1000) (mg/dL)URIT UC-1800 Automatic Urine Analyzer (12FA test strips)-(0)+-(15)+1(30)+2(100)+3(300)
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)-(0)6981000
+-(15)094200
+1(30)006400
+2(100)001601
+3(300)000059
Total69895676060
Complete agreement rate100%98.90%95.50%100%98.30%
General agreement rate100%100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
PH(N=1000)(12FA test strips)
5.05.56.06.57.07.58.08.59.0
5.016120000000
5.502720000000
6.002211000000
Predicate device (K082811)6.500113410000
Uritest-500B urine analyzer(11G test strip)7.000031110000
7.50000048000
8.00000002700
8.50000000200
9.0000000007
Total1612762121371124827207
Complete agreement rate100%98.60%99.50%97.80%99.10%100%100%100%100%
General agreement rate100%100%100%100%100%100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer (12FA test strips)
SG(N=1000)1.0051.0101.0151.0201.0251.030
Predicate device (K082811)Uritest-500B urine analyzer(11G test strip)1.0053300000
1.01001343000
1.01503211000
1.02000027300
1.02500012020
1.03000002138
Total33137214274204138
Complete agreement rate100%97.80%98.60%99.60%99.00%100%
General agreement rate100%100%100%100%100%100%

a.2 Comparison with Mission® U120 Ultra Urine Analyzer (K142391)

A total of 979 clinical urine samples were collected for the experiment of comparison with Mission® U120 Ultra Urine

{41}------------------------------------------------

Analyzer (K142391), and the Comparison results presented in the tables below.

URIT UC-1800 Automatic Urine Analyzer
Microalbumin(N=979)(12FA test strips)
10mg/L30mg/L80mg/L150mg/L
Predicate device (K142391)527800
Mission® U120 Ultra Urine Analyzer147590
Mission® Urinalysis Reagent Strips(Microalbumin/Creatinine)071683
150mg/L003165
Total54190180168
Complete agreement rate97.41%83.33%93.33%98.21%
General agreement rate100%100%100%100%
URIT UC-1800 Automatic Urine Analyzer
Creatinine(N=979)(12FA test strips)
10mg/dL50mg/dL100mg/dL200mg/dL300mg/dL
10mg/dL523000
Predicate device (K142391)Mission® U120 Ultra Urine AnalyzerMission® Urinalysis Reagent Strips(Microalbumin/Creatinine)50mg/dL3303500
100mg/dL0441870
200mg/dL0071512
300mg/dL000420
Total5531043016222
Complete agreement rate94.55%97.74%97.21%93.21%90.91%
General agreement rate100%100%100%100%100%

a.3 Comparison with AUTION MAX AX-4030 Urinalysis System (K093098)

a.3.1 Color

Collect 1365 clinical urine sample results shall cover all grades of color of predicate device AX-4030. The test shall be completed within 2h, and the interval of test between different instruments shall be controlled within 2h. According to the requirements of product instructions, conduct comparison test on the ARKRAY AUTION MAX AX-030 and UC 1800 automatic urine analyzer (physical mode), once for each sample (retest if the results are abnormal due to instrument failure), and record the color results. After completing the test, review and compare the results; before test, quality control calibration shall be carried out on the corresponding instruments to ensure accuracy of the instruments. The comparison results of Color are presented in the Table 36 below.

AX-4030
ColorColorlessYellowRedBrownGreenOrangeBlueViolet
UC1800Colorless28310000000
Yellow27717100400
Red026310002
Brown001650000
Green000037010
Other00002723938
Total310729656639764040
Coincidence rate91.29%98.35%96.92%98.48%94.87%94.74%97.50%95.00%
ConclusionQualifiedQualifiedQualifiedQualifiedQualifiedQualifiedQualifiedQualified

Table 36 The comparison results of two methods-Color

{42}------------------------------------------------

a.3.2 Turbidity

Collect 1000 clinical urine sample results shall cover all grades of turbidity of predicate device AX-4030. The test shall be completed within 2h, and the time interval of test between different instruments shall be controlled within 2h. According to the requirements of product instructions, conduct comparison test on the ARKRAY AUTION MAX AX-4030 and UC 1800 automatic urine analyzer (physical mode), once for each sample (retest ifthe results are abnormal due to instrument failure), and record the Turbidity results. After completing the test, review and compare the results; before test, quality control calibration shall be carried out on the corresponding instruments to ensure accuracy of the instruments. The comparison results of Turbidity are presented in the Table 37 below.

AX-4030
Turbidity-+1+2
UC1800Clear89800
Micro turbid10420
Turbid0251
Very turbid0024
Total9086725
Coincidence rate98.90%100%96.00%
SampleMethod 1Method 2
11.21.1
21.51.4
31.81.7
42.12.0
52.42.3

b Matrix comparison

Not applicable. This device is for testing with human urine only.

9 Clinical Test Conclusion

No clinical study is included in this submission.

9.1 Clinical studies

Clinical Sensitivity a

Not applicable

b Clinical specificity

Not applicable

c Other clinical supportive data (when a. and b. are not applicable)

Not applicable

10 Stability Test Conclusion

10.1URIT 11FA Urine Reagent Strips

After a study on the stability of Urine Reagent Strips in the shelf life, the results show: various performance indexes of the Urine Reagent Strips within 25 months after being stored at 30°C (extreme storage condition) all meet the relevant performance index requirements in the technical requirements for the requirement that the validity period of the Urine Reagent Strips is expected to be 24 months. Therefore, in consideration of design margin, we determine that the shelf life of Urine Reagent Strips is 24 months.

After a study on Open bottle stability, the results performance indexes of the Urine Reagent Strips stored at 30°C(extreme storage condition), humidity $\leq$ 80% for 14 weeks after opening all meet the performance index requirements in the technical requirements for the product, but various performance indexes at the 15th week do not meet these requirements. To be prudent, we determine that the validity period of Urine Reagent Strips after opening is 3 months (12 weeks) after opening at 30°C(extreme storage condition), humidity $\leq$ 80%. The Urine Reagent Strips within the shelf life and the validity period after opening has reliable properties and reliable test results as reference for clinical diagnosis.

After a study on stability during simulant transport, after the Urine Reagent Strips are stored respectively at the high

{43}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China

temperature of 50°C, at the low temperature of -20°C, and at the high temperature of 40°C and a high humidity of no less than 80% for 30 days, they are stored at 2℃ to 30℃ in a dry and dark place until the validity period expires, and compared with Urine Reagent Strips always stored on the conditions specified in the performance indexes remain basically unchanged and meet the appearance and performance in the technical requirements for the product, which shows that the performance of the product is accurate and reliable during transport (within 30 days) even under extremely harsh transport conditions.

In addition, various performance indexes of the Reagent Strips within 25 hours after being put in the reagent strip bin of Automatic Urine Analyzer on experimental conditions all meet the relevant performance index requirements in the technical requirements for the product. To be prudent, we determine that the On-board stability time of Urine Reagent Strips is 24 hours.

10.2 URIT 12FA Urine Reagent Strips

After a study on the stability of Urine Reagent Strips in the shelf life, the results show: various performance indexes of the Urine Reagent Strips within 25 months after being stored at 30°C (extreme storage condition) all meet the relevant performance index requirements in the technical requirements for the requirement that the validity period of the Urine Reagent Strips is expected to be 24 months. Therefore, in consideration of design margin, we determine that the shelf life of Urine Reagent Strips is 24 months.

After a study on Open bottle stability, the results performance indexes of the Urine Reagent Strips stored at 30°C (extreme storage condition),humidity ≤ 80% for 14 weeks after opening all meet the performance index requirements in the technical requirements for the product, but various performance indexes at the 15th week do not meet these requirements. To be prudent, we determine that the validity period of Urine Reagent Strips after opening is 3 months (12 weeks) after opening at 30°C(extreme storage condition),humidity ≤ 80%. The Urine Reagent Strips within the shelf life and the validity period after opening

has reliable properties and can bring accurate and reliable test results as reference for clinical diagnosis.

After a study on stability during simulant transport, after the Urine Reagent Strips are stored respectively at the high temperature of 50°C, at the low temperature of -20°C, and at the high temperature of 40°C and a high humidity of no less than 80% for 30 days, they are stored at 2°C to 30°C in a dry and dark place until the validity period expires, and compared with Urine Reagent Strips always stored on the conditions specified in the performance indexes remain basically unchanged and meet the appearance and performance in the technical requirements for the product, which shows that the performance of the product is accurate and reliable during transport (within 30 days) even under extremely harsh transport conditions.

In addition, various performance indexes of the Urine Reagent Strips within 25 hours after being put in the reagent strip bin of Automatic Urine Analyzer on experimental conditions all meet the relevant performance index requirements in the technical requirements for the product. We determine that the On-board stability time of Urine Reagent Strips is 24 hours.

11 Conclusion

The analytical performance studies demonstrated substantial equivalency between the proposed device and predicate devices. Which demonstrated that the URIT 11FA Urine Reagent Strips and URIT 12FA Urine Reagent Strips read by URIT UC-1800 Automatic Urine Analyzer are safe, effective and such are substantially equivalent to the Uritest 11G Urine Reagent Strips read by the Urine Analyzer (K082811) currently sold on the U.S. market for professional use in detecting Ascorbic acid, Nitrite, Leukocyte, Creatinine, Ketone, Bilirubin, Glucose, Protein, Blood, Specific Gravity and pH in/of human urine.

And the URIT 12FA Urine Reagent Strips read by URIT UC-1800 Automatic Urine Analyzer is safe, effective and such is substantially equivalent to the Mission® Urinalysis Reagent Strips (Microalbumin/Creatinine) read by the Mission® U120 Ultra Urine Analyzer(K142391) currently sold on the U.S. market for professional use at point-of-care locations in detecting Microalbumin and Creatinine in human urine.

{44}------------------------------------------------

No. D-07 Information Industry District, High-Tech Zone, Guilin, Guangxi 541004, P. R. China

And the URIT UC-1800 Automatic Urine Analyzer is safe, effective and such is substantially equivalent to the AUTION MAX AX-4030 Urinalysis System ([K093098](#K093098)) currently sold on the U.S. market for professional use in detecting specific gravity, turbidity and color of human urine.

§ 864.6550 Occult blood test.

(a)
Identification. An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)(b)
Classification. Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.