(186 days)
Permanent sealing of root canal
One-Fil is a convenient premixed ready-to use, injectable calcium silicate composition which requires the presence of water to set and harden. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, and a holder.
The provided text focuses on the 510(k) summary for a dental root canal sealer named "One-Fil." It describes the device, its intended use, and a comparison to a predicate device ("Well-Root™ST"). However, the document does not describe a clinical study that proves the device meets specific acceptance criteria in the way a diagnostic AI/ML device would.
Instead, the performance data presented is based on non-clinical tests (bench testing and biocompatibility) against established standards. Therefore, most of the requested information for acceptance criteria and a study demonstrating performance against those criteria is not applicable in this context.
Here's a breakdown of the relevant information from the document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't define explicit "acceptance criteria" in terms of clinical performance metrics (e.g., sensitivity, specificity, accuracy). Instead, it states that the device was tested in accordance with existing standards.
| Test Category | Standard / Test Performed | Reported Device Performance |
|---|---|---|
| Performance Tests | ISO 6876 (Appearance, Volume, Packaging, Flow Setting time, Film thickness, Solubility, Radiopaque) | Conformed to ISO 6876 |
| Chemical/Physical | pH Test | Test performed (result not specified but deemed acceptable for SE) |
| Biocompatibility | ISO 10993-3, 5, 6, 10, 11 and USP 34, 151> | Tests performed (results not specified but deemed acceptable for SE) |
| Stability | Shelf Life Test | Test performed (result not specified but deemed acceptable for SE) |
Regarding the study proving the device meets acceptance criteria:
The document does not describe a clinical or comparative effectiveness study of "One-Fil." The demonstration of meeting "acceptance criteria" is based on the device conforming to established international standards (ISO 6876, ISO 10993, USP) and demonstrating substantial equivalence to a predicate device through non-clinical testing.
The remaining points of your request are not applicable to this document as it pertains to a dental filling material, not an AI/ML-driven diagnostic device.
2. Sample size used for the test set and the data provenance: Not applicable. This document describes bench testing and biocompatibility assessments, not a clinical study with a "test set" of patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data is not relevant here.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to a clinical test set. For the non-clinical tests, the "ground truth" would be the specifications and requirements defined by the ISO and USP standards.
8. The sample size for the training set: Not applicable. There is no training set mentioned, as this is not an AI/ML device.
9. How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2024
Mediclus Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612
Re: K232182
Trade/Device Name: One-Fil Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: November 28, 2023 Received: November 28, 2023
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name One-Fil
Indications for Use (Describe)
Permanent sealing of root canal
| Type of Use (Select one or both, as applicable) |
|---|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K232182 510(k) Summary
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date:
1. Applicant / Submitter:
MEDICLUS CO., LTD. No. 1210, 134 Gongdan-ro, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Republic of Korea Tel : +82-43-211-2877 Fax : +82-43-211-2866
Submission Correspondent: 2.
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine CA 92620 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com
3. Device:
| Proprietary Name: | One-Fil |
|---|---|
| Common Name: | Dental root canal sealer |
| Classification Name: | Resin, Root Canal Filling |
| Classification: | Class II, 21 CFR 872.3820 |
| Classification Product Code: | KIF |
4. Predicate Device:
Well-Root™ST (K170950) by VERICOM Co., Ltd.
{5}------------------------------------------------
Device Description: 5.
One-Fil is a convenient premixed ready-to use, injectable calcium silicate composition which requires the presence of water to set and harden. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, and a holder.
Indications for Use: 6.
Permanent sealing of root canal
7. Performance Data (Non-Clinical):
The following tests were performed on the subject device and the test results support that the subject device is substantially equivalent to the predicate devices.
- Performance Tests in accordance with ISO 6876 (Appearance, Volume, Packaging, Flow Setting time, Film thickness, Solubility, Radiopaque)
- pH Test
- Biocompatibility Tests in accordance with ISO 10993-3, 5, 6, 10, 11 and USP 34,151> ■
- Shelf Life Test ■
8. Substantial Equivalence
8.1. Comparison Chart
| Subject Device | Primary Predicate Device | |
|---|---|---|
| Trade name | One-Fil | Well-Root™ST |
| Manufacturer | Mediculs Co.,Ltd | VERICOM Co., Ltd. |
| 510K Number | - | K170950 |
| Product Code | KIF | KIF |
| Indications forUse | Permanent sealing of root canal | Permanent sealing of root canal |
| Image | Image: One-Fil product packaging | Image: Well-Root product packaging |
| Raw materials | Insoluble, radiopaque material based on acalcium silicate compound | Insoluble, radiopaque material based on acalcium aluminosiliate compound |
{6}------------------------------------------------
| Principle ofoperation | One-Fil is an insoluble, radiopaquematerial which sets and hardens withmoisture providing from dentin tubulesduring hydration reaction. One-Fil ispackaged in a preloaded syringe and issupplied with disposable tips. | WellRoot™ ST is an insoluble, radiopaquematerial which sets and hardens withmoisture providing from dentin tubulesduring hydration reaction. Well-Root™STis packaged in a preloaded syringe and issupplied with disposable tips. |
|---|---|---|
| PerformanceStandardsConformance | Conformed to ISO 6876 | Conformed to ISO 6876 |
| Biocompatibility | Yes | Yes |
| Use | Prescription / Hospital | Prescription / Hospital |
| Delivery form | Single paste | Single paste |
| Sterility | Non-sterile | Non-sterile |
8.2. Substantial Equivalence Discussion
One-Fil has the same intended use as the predicate devices: Well-Root™ST (K170950) by VERICOM Co., Ltd. The indication of the statement of the subject device is more simplified than the predicate devices but it does not contain new indications.
All of the devices confirm to ISO 6876 and are provided non-sterile. Both One-Fil and Well-Root™ ST are premixed ready-to-use injectable white hydraulic cement paste. The differences between One-Fil and Well-Root™ ST are raw materials. However, performance test according to ISO 6876 and biocompatibility tests according to ISO 10993 demonstrate that the difference does not raise a question in safety and effectiveness. We conclude that One-Fil is substantially equivalent to the predicate device.
9. Conclusion:
Based on the information submitted herein, MEDICLUS CO., LTD. concludes that the One-Fil is substantially equivalent to the predicate device.
§ 872.3820 Root canal filling resin.
(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.