AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Safety Hypodermic Needles for Single Use are intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

The Sterile Hypodermic Needles for Single Use- Disposable Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.

The Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.

Device Description

The proposed devices are provided in two types of configurations; one type is a Sterile Safety Hypodermic Needles for Single Use, the other is Sterile Hypodermic Needles for Single Use. The Sterile Hypodermic Needles also divided into two types, one type is an ordinary disposable needle contained in a sterility maintenance package, the other is self-sealing hypodermic needle with an end cap, the protective cap and end cap form a sterility maintenance package.

The Sterile Safety Hypodermic Needles for Single Use are intended for single use only, which consists of a hypodermic needle with a safety sheath attached to the needle hub. The proposed device is available in 28-34 gauge and 4-32 mm lengths. The safety sheath will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

The Sterile Hypodermic Needles for Single Use are intended for single use only. The proposed device is available in 18-34 gauge and 2-40 mm lengths.

The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and are supplied in a sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

AI/ML Overview

The provided text describes the 510(k) summary for hypodermic needles and does not contain information about an AI/ML-enabled medical device. Therefore, I cannot address the specific points regarding acceptance criteria and study details for such a device. The document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing for physical, mechanical, chemical, biocompatibility, and sterilization properties of the needles.

  • No AI/ML device: The document pertains to "Sterile Safety Hypodermic Needles for Single Use" and "Sterile Hypodermic Needles for Single Use-Disposable Needles," and "Sterile Hypodermic Needles for Single Use-Self-sealing Hypodermic Needles." These are physical medical devices, not AI/ML-enabled software.
  • No AI/ML performance data: Consequently, there are no acceptance criteria or study details related to the performance of an AI/ML algorithm (e.g., sensitivity, specificity, AUC) or its impact on human reader performance.
  • Non-clinical tests: The document lists numerous non-clinical tests conducted to prove the device meets standards like ISO 7864, ISO 9626, and ISO 80369-7, as well as biocompatibility tests (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogen). It also mentions a "Simulated Clinical Study" and "Safety Feature Test" for the sharps injury prevention feature, though details on the study design or sample size are not provided beyond stating that pre-established criteria were met.

Therefore, I cannot generate the table or information requested as it is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2024

Berpu Medical Technology Co., Ltd. % Diana Hong, General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, Shanghai 200120, China

Re: K232165

Trade/Device Name: Sterile Safety Hypodermic Needles for Single Use; Sterile Hypodermic Needles for Single Use- Disposable Needles: Sterile Hypodermic Needles for Single Use-Self-sealing Hypodermic Needles

Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: January 2, 2024 Received: January 2, 2024

Dear Diana Hong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shruti N. Mistry -S

Shruti Mistry, MS Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K232165

Device Name

Sterile Safety Hypodermic Needles for Single Use Sterile Hypodermic Needles for Single Use- Disposable Needles Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles

Indications for Use (Describe)

The Sterile Safety Hypodermic Needles for Single Use are intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

The Sterile Hypodermic Needles for Single Use- Disposable Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.

The Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K232165 510(k) Summary

    1. Date of Preparation: 2/1/2024
    1. Sponsor Identification

Berpu Medical Technology Co., Ltd.

NO.14 Xingji Road, Yongxing Street, Longwan District, Wenzhou, Zhejiang, China 325000

Establishment Registration Number: 3004496829

Contact Person: Buxin Yu Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: ybx@berpu.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Mr. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{5}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Sterile Safety Hypodermic Needles for Single Use Sterile Hypodermic Needles for Single Use- Disposable Needles Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles

Regulation Name: Hypodermic, Single Lumen Needle Classification: II Regulation Number: 21 CFR 880.5570 Product Code: FMI

Predicate Device: 510(k) Number: K180417 Product Name: Sterile Hypodermic Needles for Single Use Sterile Safety Hypodermic Needles for Single Use

Indications for use:

The Sterile Safety Hypodermic Needles for Single Use are intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

The Sterile Hypodermic Needles for Single Use- Disposable Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.

The Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection.

Device Description

The proposed devices are provided in two types of configurations; one type is a Sterile Safety Hypodermic Needles for Single Use, the other is Sterile Hypodermic Needles for Single Use. The Sterile Hypodermic Needles also divided into two types, one type is an ordinary disposable needle contained in a sterility maintenance package, the other is self-sealing hypodermic needle with an end cap, the protective cap and end cap form a sterility maintenance package.

The Sterile Safety Hypodermic Needles for Single Use are intended for single use only, which consists of a hypodermic needle with a safety sheath attached to the needle hub. The proposed device is available in 28-34 gauge and 4-32 mm lengths. The safety sheath will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

The Sterile Hypodermic Needles for Single Use are intended for single use only. The proposed device is

{6}------------------------------------------------

available in 18-34 gauge and 2-40 mm lengths.

The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 106 and are supplied in a sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

Needle TypeNeedle gaugeNeedle Length
Sterile Safety HypodermicNeedles for Single Use28G, 29G, 30G, 31G, 32G,33G, 34G4mm, 5mm, 6mm, 8mm, 9mm, 10mm,12mm, 13mm, 16mm, 20mm, 22mm,25mm, 30mm, 32mm
Sterile Hypodermic Needlesfor Single Use- DisposableNeedle27G, 30G, 31G, 32G, 33G,34G2mm, 2.5mm, 4mm, 5mm, 6mm, 8mm,9mm, 10mm, 12mm, 13mm, 16mm,20mm, 22mm, 25mm, 30mm, 32mm
Sterile Hypodermic Needlesfor Single Use- Self-sealingHypodermic Needle18G, 19G, 20G, 21G, 22G,23G, 24G, 25G, 26G, 27G,28G, 29G, 30G, 31G, 32G,33G, 34G2mm, 2.5mm, 4mm, 5mm, 6mm, 8mm,9mm, 10mm, 12mm, 13mm, 16mm,20mm, 22mm, 25mm, 30mm, 32mm,35mm, 38mm, 40mm

Table 1 Specifications for Proposed Devices

{7}------------------------------------------------

5. Technological Characteristic Table

ITEMProposed DeviceK232165Predicate DeviceK180417Remark
ProductSterile Safety Hypodermic Needlesfor Single UseSterile Safety Hypodermic Needlesfor Single Use/
Product CodeFMIFMISame
Regulation No.21 CFR 880.557021 CFR 880.5570Same
ClassClass IIClass IISame
Indication for UseThe Sterile Safety Hypodermic Needles for Single Use are intended tobe used with a luer slip or luer locksyringe for aspiration and injection offluids for medical purpose. Afterwithdrawal of the needle from thebody, the attached needle safety shieldcan be manually activated to cover theneedle immediately after use tominimize risk of accidentalneedlestick.The Sterile Safety Hypodermic Needles for Single Use are intended tobe used with a luer slip or luer locksyringe for aspiration and injection offluids for medical purpose. Afterwithdrawal of the needle from thebody, the attached needle safety shieldcan be manually activated to cover theneedle immediately after use tominimize risk of accidentalneedlestick.Same
Configuration andmaterialNeedle hub Polypropylene (PP)Protective cap Polypropylene (PP)Needle Stainless SteelSafety sheath Polypropylene (PP)Needle hub Polypropylene (PP)Protective cap Polypropylene (PP)Needle Stainless SteelSafety sheath Polypropylene (PP)Same
Safety sheathImage: Safety sheathSame
Operation ModeFor manual use onlyFor manual use onlySame
Single UseSingle UseSingle UseSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Needle GaugeAvailable in 28G, 29G, 30G, 31G,32G, 33G, 34GAvailable in 18G, 19G, 20G, 21G,22G, 23G, 24G, 25G, 26G, 27GDifferent
Needle LengthAvailable in 4mm, 5mm, 6mm, 8mm,9mm, 10mm, 12mm, 13mm, 16mm,20mm, 22mm, 25mm, 30mm, 32mmAvailable in 6-50mm
Needle tube colorsComply with ISO 6009Comply with ISO 6009Similar
34G: Orange27G: Medium Grey
33G: Black26G: Brown
32G: Deep green25G: Orange
31G: White24G: Medium Purple
30G: Yellow23G: Dark Blue
29G: Red22G: Black
28G: Blue-green21G: Dark Green
20G: Yellow
19G: Cream
18G: Pink
Needle PerformanceComplied withISO 7864,ISO 9626Complied withISO 7864,ISO 9626Same
Luer ConnectorPerformanceComplied withISO 80369-7Complied withISO 594-2Different
Biocompatibility
CytotoxicityNo cytotoxicityNo cytotoxicitySame
IrritationNo intracutaneous reactivityNo intracutaneous reactivity
SensitizationNo sensitizationNo sensitization
Systemic ToxicityNo systemic toxicityNo systemic toxicity
HemolysisNo HemolysisNo Hemolysis
PyrogenNo PyrogenNo Pyrogen
ParticulateComply with USP<788>no more than 6000 /device≥10µm, andno more than 600 /device≥25µm.Comply with USP<788>
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame
Shelf life5 years5 yearsSame

Table 2 Comparison of Sterile Safety Hypodermic Needles

{8}------------------------------------------------

Different Analysis 1-Needle Size and Length

The needle size and length for the proposed device is different from the predicate devices. This difference is just in dimension. Different size and length devices will be selected by the physician per the patient's condition. Moreover, the needle length of the proposed device with safety needle is included in the range of the needle length of the predicate device. The performance of the needle was tested, and the test results showed that it was in accordance with ISO 7864 and ISO 9626 standards. Therefore, this difference does not raise new questions of safety and effectiveness.

Similar Analysis 2-Needle hub color

{9}------------------------------------------------

The color of needle hub for proposed devices is different from the predicate devices. The color is indicating the nominal outer diameter of the needle. The proposed device and the predicate device have different needle gauges, so the color of the needle hub is different, but the needle hub complies with the ISO 6009 standard. Therefore, this difference does not raise new questions of safety and effectiveness.

Different Analysis 3-Luer Connector Performance

Although the proposed device and the predicate device follow different luer connector standards, this is because ISO 594-2 is replaced by ISO 80369-7 according to current recognitions. The test results of the proposed device show that the connector performance meets the requirements of ISO 80369-7. Therefore, this difference does not raise new questions of safety and effectiveness.

{10}------------------------------------------------

ITEMProposed DeviceK232165Predicate DeviceK180417Remark
ProductSterile Hypodermic Needles for Single Use- Disposable NeedlesSterile Hypodermic Needles for Single Use/
Product CodeFMIFMISame
Regulation No.21 CFR 880.557021 CFR 880.5570Same
ClassClass IIClass IISame
Indication for UseThe Sterile Hypodermic Needles for Single Use-Disposable Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.The Sterile Hypodermic Needles for single Use are intended to be used with a luer slip or luer lock syringe and injection devices for general purpose fluid injection/aspiration.Same
Configuration and materialNeedle Hub Polypropylene (PP)Protective cap Polypropylene (PP)Needle Stainless SteelNeedle hub Polypropylene (PP)Protective cap Polypropylene (PP)Needle Stainless SteelSame
Operation ModeFor manual use onlyFor manual use onlySame
Single UseSingle UseSingle UseSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Needle GaugeAvailable in 27G, 30G, 31G, 32G, 33G, 34GAvailable in 14G, 15G, 16G, 17G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 29G, 30GDifferent
Needle LengthAvailable in 2mm, 2.5mm, 4mm, 5mm, 6mm, 8mm, 9mm, 10mm, 12mm, 13mm, 16mm, 20mm, 22mm, 25mm, 30mm, 32mmAvailable in 6-60mm
Needle hub colorComplied with ISO 600934G: Orange33G: Black32G: Deep green31G: White30G: Yellow27G: Medium GreyComplied with ISO 600930G: Yellow29G: Red27G: Medium Grey26G: Brown25G: Orange24G: Medium Purple23G: Dark Blue22G: Black21G: Dark Green20G: Yellow19G: CreamSimilar
17G: Red Purple
16G: White
15G: Blue Grey
14G: Light Green
Needle PerformanceComplied withISO 7864,ISO 9626Complied withISO 7864,ISO 9626Same
Luer ConnectorPerformanceComplied withISO 80369-7Complied withISO 594-2Different
Biocompatibility
CytotoxicityNo cytotoxicityNo cytotoxicitySame
IrritationNo intracutaneous reactivityNo intracutaneous reactivity
SensitizationNo sensitizationNo sensitization
Systemic ToxicityNo systemic toxicityNo systemic toxicity
HemolysisNo HemolysisNo Hemolysis
PyrogenNo PyrogenNo Pyrogen
ParticulateComply with USP<788>no more than 6000 /device≥10µm, andno more than 600 /device≥25µm.Comply with USP<788>
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame
Shelf life5 years5 yearsSame

Table 3 Comparison of Sterile Hypodermic Needles for Single Use- Disposable Needles

{11}------------------------------------------------

Different Analysis 4-Needle Size and Length

The needle size and length for the proposed device is different from the predicate devices. This difference is just in dimension. Different size and length devices will be selected by the physician per the patient's condition. Moreover, the needle length of the proposed device with disposable needle is similar to the needle length of the predicate device. The performance of the needle was tested, and the test results showed that it was in accordance with the ISO 7864 and ISO 9626 standards. Therefore, this difference does not raise new questions of safety and effectiveness.

Similar Analysis 5-Needle hub color

The color of needle hub for the proposed devices is different from the predicate devices. The color is indicating the nominal outer diameter of the tube of the needle. The proposed device and the predicate device have different needle gauges, so the color of the needle hub is different, but the needle hub complies with the ISO6009 standard. Therefore, this difference does not raise new questions of safety and effectiveness.

{12}------------------------------------------------

Different-Analysis 6-Luer Connector Performance

Although the proposed device and the predicate device follow different luer connector standards, this is because ISO 594-2 is replaced by ISO 80369-7 according to current recognitions. The test results of the proposed device show that the connector performance meets the requirements of ISO 80369-7. Therefore, this difference does not raise new questions of safety and effectiveness.

{13}------------------------------------------------

ITEMProposed DeviceK232165Predicate DeviceK180417Remark
ProductSterile Hypodermic Needles for Single Use- Self-sealing Hypodermic NeedlesSterile Hypodermic Needles for Single Use/
Product CodeFMIFMISame
Regulation No.21 CFR 880.557021 CFR 880.5570Same
ClassClass IIClass IISame
Indication for UseThe Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles are intended to be used with a luer slip or luer lock syringes and injection devices for general purpose fluid injection/aspiration.The Sterile Hypodermic Needles for single Use are intended to be used with a luer slip or luer lock syringe and injection devices for general purpose fluid injection/aspiration.Same
Configuration and materialNeedle HubPolypropylene (PP)Needle hubPolypropylene (PP)Different
Protective capPolypropylene (PP)Protective capPolypropylene (PP)
NeedleStainless SteelNeedleStainless Steel
End capPolypropylene (PP)/
Operation ModeFor manual use onlyFor manual use onlySame
Single UseSingle UseSingle UseSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Needle GaugeAvailable in 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G, 31G, 32G, 33G, 34GAvailable in 14G, 15G, 16G, 17G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 29G, 30GDifferent
Needle LengthAvailable in 2mm, 2.5mm, 4mm, 5mm, 6mm, 8mm, 9mm, 10mm, 12mm, 13mm, 16mm, 20mm, 22mm, 25mm, 30mm, 32mm, 35mm, 38mm, 40mmAvailable in 6-60mm
Needle hub colorComplied with ISO 600934G: Orange33G: Black32G: Deep green31G: White30G: Yellow29G: Red28G: Blue-green27G: Medium Grey26G: Brown25G: OrangeComplied with ISO 600930G: Yellow29G: Red27G: Medium Grey26G: Brown25G: Orange24G: Medium Purple23G: Dark Blue22G: Black21G: Dark Green20G: YellowSimilar
24G: Medium Purple19G: Cream
23G: Deep Blue18G: Pink
22G: Black17G: Red Purple
21G: Deep Green16G: White
20G: Yellow15G: Blue Grey
19G: Cream14G: Light Green
18G: Pink
Needle PerformanceComplied withISO 7864,ISO 9626Complied withISO 7864,ISO 9626Same
Luer ConnectorPerformanceComplied withISO 80369-7Complied withISO 594-2Different
Biocompatibility
CytotoxicityNo cytotoxicityNo cytotoxicitySame
IrritationNo intracutaneous reactivityNo intracutaneous reactivity
SensitizationNo sensitizationNo sensitization
Systemic ToxicityNo systemic toxicityNo systemic toxicity
HemolysisNo HemolysisNo Hemolysis
PyrogenNo PyrogenNo Pyrogen
ParticulateComply with USP<788>no more than 6000 /device≥10µm, andno more than 600 /device≥25µm.Comply with USP<788>
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame
Shelf life5 years5 yearsSame

Table 4 Comparison of Sterile Hypodermic Needles for Single Use- Self-sealing Hypodermic Needles

{14}------------------------------------------------

Different Analysis 7- Configuration

The configuration of the proposed device is similar as the configurations of the predicate device. The proposed device has an end cap, and the predicate devices do not have an end cap. The end cap is the protection, and it's taken off when it's used. The presence of an end cap does not affect the intended use of the product, nor does it affect its use. Therefore, the differences in configuration does not raise new questions of safety and effectiveness.

Different Analysis 8-Needle Size and Length

The needle size and length for the proposed device is different from the predicate devices. This difference is just in dimension. Different size and length devices will be selected by the physician per the patient's condition. Moreover, the needle length of the proposed device with self-sealing needle is similar to the needle length of the predicate device. The performance of the needle was tested, and the test results showed

{15}------------------------------------------------

that it was in accordance with the ISO 7864 and ISO 9626 standards. Therefore, this difference does not raise new questions of safety and effectiveness.

Similar Analysis 9-Needle hub color

The color of needle hub for the proposed devices is different from the predicate devices. The color is indicating the nominal outer diameter of the tube of the needle. The proposed device and the predicate device have different needle gauges, so the color of the needle hub is different, but the needle hub complies with the ISO6009 standard. Therefore, this difference does not raise new questions of safety and effectiveness.

Different Analysis 10-Luer Connector Performance

Although the proposed device and the predicate device follow different luer connector standards, this is because ISO 594-2 is replaced by ISO 80369-7 according to current recognitions. The test results of the proposed device show that the connector performance meets the requirements of ISO 80369-7. Therefore, this difference does not raise new questions of safety and effectiveness.

  • Non-Clinical Test Conclusion 6.
    Non-clinical tests were conducted to verify that the proposed device met all design specifications and is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • △ ISO 10993-5: 2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity

  • ♪ ISO 10993-10: 2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization.

  • レ ISO 10993-11: 2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity

  • ▶ ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

  • ► ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection

  • ► ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)

  • ♪ ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dve Penetration

  • レ ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods

  • レ ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices

  • ► ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications

  • ♪ ISO 23908:2011 Sharps injury protection - Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling

{16}------------------------------------------------

  • レ ISO 10993-7:2008 Biological evaluation of medical devices-Part 7: Test of Ethylene Oxide Residues.
  • レ USP<85> Bacterial Endotoxins Test
  • レ USP<151> Pyrogen Test
  • レ USP <788> Particulate Testing
  • レ ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and System

Physical, Mechanical, Chemical testing listed in below were performed on the proposed device. The test results show that the device conforms with the requirements of related standards.

ItemStandard
CleanlinessClause 4.3 of ISO 7864:2016
Limits for acidity or alkalinityClause 4.4 of ISO 7864:2016
Limits for extractable metalsClause 4.5 of ISO 7864:2016
Size designationClause 4.6 of ISO 7864:2016
Colour codingClause 4.7 of ISO 7864:2016
Needle hubClause 4.8 of ISO 7864:2016
Needle CapClause 4.9 of ISO 7864:2016
Needle tubeClause 4.10 of ISO 7864:2016
Needle pointClause 4.11 of ISO 7864:2016
Bond between hub and needle tubeClause 4.12 of ISO 7864:2016
Patency of lumenClause 4.13 of ISO 7864:2016
ItemStandard
Surface finish and appearanceClause 5.2 of ISO 9626:2016
CleanlinessClause 5.3 of ISO 9626:2016
Limits for acidity and alkalinityClause 5.4 of ISO 9626:2016
Size designationClause 5.5 of ISO 9626:2016
DimensionsClause 5.6 of ISO 9626:2016
StiffnessClause 5.8 of ISO 9626:2016
Resistance to breakageClause 5.9 of ISO 9626:2016
Resistance to corrosionClause 5.10 of ISO 9626:2016
ItemStandard
Fluid leakageClause 6.1 of ISO 80369-7:2021
Sub-atmospheric pressure air leakageClause 6.2 of ISO 80369-7:2021
Stress crackingClause 6.3 of ISO 80369-7:2021

Resistance to overriding

Particulate testing

Resistance to separation form axial load

Resistance to separation form unscrewing

USP <788>

Clause 6.4 of ISO 80369-7:2021

Clause 6.5 of ISO 80369-7:2021

Clause 6.6 of ISO 80369-7:2021

Sterile barrier packaging testing was performed on the proposed device, which included visual inspection

{17}------------------------------------------------

(ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.

Sterilization and shelf-life testing listed below were performed on the proposed device. EO/ECH residuals did not exceed the limit of ISO 10993-7. The Endotoxin limit did not exceed 20EU/device. Shelf life test result showed that the device can maintain its performance during the claimed shelf life.

EO residueISO 10993-7:2008
ECH residueISO 10993-7:2008
Bacteria Endotoxin LimitUSP <85>
Shelf-Life EvaluationPhysical, Mechanical, Chemical, Package Testswere performed on aging samples to verify theclaimed shelf life of the device

Biocompatibility testing

The proposed device has an indirect contact with the blood path, and is a limited contact duration (<24 hours) device. The proposed device was evaluated for the following tests. The results for the biocompatibility testing show that there are no negative impacts from the materials that are used in the proposed device.

  • A Cytotoxicity,
  • A Sensitization,
  • A Irritation,
  • Acute Systemic Toxicity,

  • Hemolysis.

  • Pyrogen

Simulated Clinical Study

A simulated clinical study was performed on the proposed device according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature", issued on August 9, 2005 and ISO 23908:2011 to evaluate the safety mechanism of the proposed device. The results demonstrate that the proposed device meets the pre-established criteria.

Safety Feature Test

The safety feature test was performed on both the proposed device and the predicate device to determine its safety feature. The results demonstrate that both the proposed device and predicate device meet the acceptance criteria.

7. Clinical Test Conclusion

No clinical study is included in this submission.

    1. Conclusions
      The differences between the predicate device and the subject device do not raise new or different question

{18}------------------------------------------------

of safety or effectiveness. The Sterile Safe Safety Hypodermic Needles for Single Use, The Sterile Hypodermic Needles for Single Use-Disposable Needles and The Sterile Hypodermic Needles for Single Use-Self-sealing Hypodermic Needles are substantially equivalent to K180417 Sterile Hypodermic Needles for Single Use and Sterile Safety Hypodermic Needles for Single Use with respect to Indications for Use, target populations, treatment method, and technological characteristics.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).