K Number
K232127
Device Name
21HQ513D, 32HL512D, 31HN713D, 32HQ713D
Date Cleared
2023-08-15

(29 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
21HQ513D, 32HL512D: This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. 31HN713D, 32HO713D: This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Device Description
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.
More Information

Not Found

No
The document describes a medical monitor for displaying images and mentions image processing, but there is no mention of AI, ML, or related concepts like deep learning, training sets, or performance metrics typically associated with AI/ML algorithms. The performance studies focus on display characteristics.

No
The device is a medical monitor intended for displaying radiological images for diagnosis, not for providing therapy or treatment.

Yes

This device is a medical monitor intended for displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The term "diagnosis" directly indicates a diagnostic function.

No

The device description explicitly states "The Medical monitor is intended to provide high resolution color and grayscale medical imaging". This indicates a physical display device, not solely software. The performance studies also refer to testing "display characteristics".

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description clearly states that this device is a Medical Monitor used for displaying radiological images. It is a display device for visual interpretation of medical images, not a test performed on a biological sample.
  • Intended Use: The intended use is for displaying images for review, analysis, and diagnosis by trained medical practitioners. This is consistent with the function of a medical imaging display, not an IVD.
  • Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
    • Analyzing biological samples.
    • Detecting specific analytes (e.g., proteins, DNA, antibodies).
    • Using reagents or assays.
    • Providing quantitative or qualitative results based on biological measurements.

Therefore, this device falls under the category of medical imaging display devices, not in vitro diagnostics.

N/A

Intended Use / Indications for Use

21HQ513D, 32HL512D:
This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

31HN713D, 32HO713D:
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

Product codes (comma separated list FDA assigned to the subject device)

PGY

Device Description

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.

Mentions image processing

Medical image management and processing system

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained medical practitioners

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Test summary:

  1. Electrical Safety and Electromagnetic Compatibility
    The test results demonstrated that the proposed device complies with the following standards:
  • IEC 60601-1:2005/AMD2:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromaqnetic disturbances -Requirements and tests
  1. Software Validation
    The 21HQ513D, 32HL512D, 31HN713D and 32HQ713D contain a MODERATE level of concern software, The software was designed and developed according to a software development process and was verified and validated. There have been firmware updates since FDA 510(K) clearance.
    The LG Calibration Studio Medical, PerfectLum and Calibration Feedback System are moderate level of concern software programs. The software programs were verified and validated according to IEC 62304.
    -The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005
  2. Bench Test – Performance Test Report
    The performance items suggested in the FDA quidance "Display Devices for Diagnostic Radiology" were tested on the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D using Calibration Feedback System.

Measurements and Test Results for 21HQ513D:
a. Spatial resolution: Pass
b. Pixel defects: Pass
c. Artifacts: Pass
d. Temporal response: Pass
e. Luminance: Pass
f. Conformance to a grayscale-to-luminance function: Pass
g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners: N/A
h. Luminance uniformity or Mura test: N/A
i. Stability of luminance and chromaticity response with temperature and time of operation (on-time): N/A
j. Spatial noise: N/A
k. Reflection coefficient: N/A
l. Veiling glare or small-spot contrast: N/A
m. Color tracking: Pass
n. Gray tracking: N/A

Measurements and Test Results for 32HL512D:
a. Spatial resolution: Pass
b. Pixel defects: Pass
c. Artifacts: Pass
d. Temporal response: Pass
e. Luminance: Pass
g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners: N/A
h. Luminance uniformity or Mura test: N/A
i. Stability of luminance and chromaticity response with temperature and time of operation (on-time): N/A
j. Spatial noise: N/A
k. Reflection coefficient: N/A
l. Veiling glare or small-spot contrast: N/A
m. Color tracking: Pass
n. Gray tracking: N/A

Measurements and Test Results for 31HN713D:
a. Spatial resolution: Pass
b. Pixel defects: Pass
c. Artifacts: Pass
d. Temporal response: Pass
e. Luminance: Pass
f. Conformance to a grayscale-to-luminance function: Pass
g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners: Pass
h. Luminance uniformity or Mura test: Pass
i. Stability of luminance and chromaticity response with temperature and time of operation (on-time): Pass
j. Spatial noise: Pass
k. Reflection coefficient: Pass
l. Veiling glare or small-spot contrast: Pass
m. Color tracking: Pass
n. Gray tracking: Pass

Measurements and Test Results for 32HQ713D:
a. Spatial resolution: Pass
b. Pixel defects: Pass
c. Artifacts: Pass
d. Temporal response: Pass
e. Luminance: Pass
f. Conformance to a grayscale-to-luminance function: Pass
g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners: Pass
h. Luminance uniformity or Mura test: Pass
i. Stability of luminance and chromaticity response with temperature and time of operation (on-time): Pass
j. Spatial noise: Pass
k. Reflection coefficient: Pass
l. Veiling glare or small-spot contrast: Pass
m. Color tracking: Pass
n. Gray tracking: Pass

Key results: All display characteristics of the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D have met the pre-defined criteria. Therefore, the performance of the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D were verified through the performance test.

Clinical Test Summary: No clinical studies were considered necessary and performed.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K223789, K230845

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

LG Electronics Inc. % Daseul An RA Specialist 168. Suchul-daero Gumi-si. Gyeongsangbuk-do 39368 REPUBLIC OF KOREA

August 15, 2023

Re: K232127

Trade/Device Name: 21HQ513D, 32HL512D, 31HN713D, 32HQ713D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: July 10, 2023 Received: July 17, 2023

Dear Daseul An:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K232127

Device Name 21HQ513D, 32HL512D, 31HN713D, 32HQ713D

Indications for Use (Describe) 21HQ513D, 32HL512D

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

31HN713D, 32HO713D

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/0 description: The image shows the logo for LG Electronics. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the text "LG Electronics" in gray. The text is in a simple, sans-serif font.

510(k) Summary

K232127

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

July 11, 2023

2. Submitter's Information [21 CFR 807.92(a)(1)]

Name of Sponsor:LG Electronics Inc.
Address:222, LG-ro, Jinwi-myeon, Pyeongtaek-si,
Gyeonggi-do, 17709, Republic of Korea
Name of Manufacturer:LG Electronics Inc.
Address:168, Suchul-daero, Gumi-si, Gyeongsangbuk-do, 39368
Republic of Korea
Contact Name:Daseul An / Regulatory Affairs Specialist
Telephone No.:+82-10-8914-0116
Email Address:daseul.an@lge.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

  • Trade Name: 21HQ513D, 32HL512D, 31HN713D, 32HQ713D ●
  • Common Name: .

Medical Monitor

  • Classification: ●
Classification NameMedical image management and processing system
Classification Number21 CFR 892.2050
Product CodePGY
Device ClassII
Review PanelRadiology

4

Image /page/4/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" inside, next to the text "LG Electronics" in gray. The "L" in the circle is designed to resemble a smiling face, with the dot above the "L" representing an eye.

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow;

Predicate Device for 21HQ513D

510(k) Number:K223789
Applicant:LG Electronics
Classification Name:Medical image management and processing system
Trade Name:21HQ513D

Predicate Device for 32HL512D, 31HN713D, 32HQ713D

• 510(k) Number:K230845
• Applicant:LG Electronics
• Classification Name:Medical image management and processing system
• Trade Name:32HL512D, 31HN713D, 32HQ713D

5. Description of the Device [21 CFR 807.92(a)(4)]

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.

6. Indications for use [21 CFR 807.92(a)(5)]

- 21HQ513D, 32HL512D

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diaqnosis by trained medical practitioners. The display is not intended for mammography.

- 31HN713D, 32HQ713D

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

5

Image /page/5/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the text "LG Electronics" in gray. The logo is simple and modern, and the colors are eye-catching.

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

The comparison table shows the technical characteristics of the subject device are substantially equivalent to the predicate device. There are no significant differences between the subject devices and the primary predicate devices that would adversely affect the use of the product. The main change is the addition of the available calibration tool which is validated according to IEC 62304.

Proposed DevicePredicate DeviceEquivalence
K NumberTBDK223789-
ManufacturerLG Electronics Inc.LG Electronics Inc.-
Model Name21HQ513D21HQ513D-
Classification
NameMedical image management and
processing systemMedical image management and
processing systemSame
Classification
Number21 CFR 892.205021 CFR 892.2050Same
Indications for UseThis Medical Monitor is indicated
for use in displaying radiological
images for review, analysis, and
diagnosis by trained medical
practitioners. The display is not
intended for mammography.This Medical Monitor is indicated
for use in displaying radiological
images for review, analysis, and
diagnosis by trained medical
practitioners. The display is not
intended for mammography.Same
PowerMAX. 120WMAX. 120WSame
ConsumptionOff Mode ≤ 0.3WOff Mode ≤ 0.3WSame
LCD ScreenTFT LCDTFT LCDSame
Pixel Pitch0.2115 x 0.2115 mm0.2115 x 0.2115 mmSame
Resolution1,536 x 2,048 pixels1,536 x 2,048 pixelsSame
Horizontal
Frequency30 kHz to 130 kHz30 kHz to 130 kHzSame
Vertical Frequency56 Hz to 61 Hz56 Hz to 61 HzSame
Input video signalsDisplayPort x 2
DVI-IN x 1DisplayPort x 2
DVI-IN x 1Same
Calibration ToolPerfectLum / LG Calibration Studio
Medical / Calibration Feedback
SystemPerfectLum / LG Calibration Studio
MedicalModified

Table 1. Comparison of Proposed Device to Primary Predicate Device (21HQ513D)

Table 2. Comparison of Proposed Device to Primary Predicate Device (32HL512D)

Proposed DevicePredicate DeviceEquivalence
K NumberTBDK230845-
ManufacturerLG Electronics Inc.LG Electronics Inc.-
Model Name32HL512D32HL512D-
Classification
NameMedical image management and
processing systemMedical image management and
processing systemSame
Classification
Number21 CFR 892.205021 CFR 892.2050Same

6

Image /page/6/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the text "LG Electronics" in gray. The text is in a sans-serif font and is aligned to the right of the circular logo.

| Indications for Use | This Medical Monitor is indicated
for use in displaying radiological
images for review, analysis, and
diagnosis by trained medical
practitioners. The display is not
intended for mammography. | This Medical Monitor is indicated
for use in displaying radiological
images for review, analysis, and
diagnosis by trained medical
practitioners. The display is not
intended for mammography. | Same |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Power
Consumption | MAX. 65W
Sleep Mode ≤ 0.5W
Off Mode ≤ 0.3W | MAX. 65W
Sleep Mode ≤ 0.5W
Off Mode ≤ 0.3W | Same |
| Screen size | 718.2 x 414.3 mm | 718.2 x 414.3 mm | Same |
| LCD Screen | TFT LCD | TFT LCD | Same |
| Pixel Pitch | 0.18159 x 0.18159 mm | 0.18159 x 0.18159 mm | Same |
| Resolution | 3,840 x 2,160 pixels | 3,840 x 2,160 pixels | Same |
| Horizontal
Frequency | 30 kHz to 135 kHz | 30 kHz to 135 kHz | Same |
| Vertical Frequency | 56 Hz to 61 Hz | 56 Hz to 61 Hz | Same |
| Input video signals | DisplayPort x 2
HDMI x 1 | DisplayPort x 2
HDMI x 1 | Same |
| Calibration Tool | PerfectLum / LG Calibration Studio
Medical / Calibration Feedback
System | PerfectLum / LG Calibration Studio
Medical | Modified |

Table 3. Comparison of Proposed Device to Primary Predicate Device (31HN713D)

Proposed DevicePredicate DeviceEquivalence
K NumberTBDK230845-
ManufacturerLG Electronics Inc.LG Electronics Inc.-
Model Name31HN713D31HN713D-
Classification
NameMedical image management and
processing systemMedical image management and
processing systemSame
Classification
Number21 CFR 892.205021 CFR 892.2050Same
Indications for UseThis Medical Monitor is indicated
for use in displaying radiological
images (including full-field digital
mammography and digital breast
tomosynthesis) for review, analysis,
and diagnosis by trained medical
practitioners.This Medical Monitor is indicated
for use in displaying radiological
images (including full-field digital
mammography and digital breast
tomosynthesis) for review, analysis,
and diagnosis by trained medical
practitioners.Same
Power
ConsumptionMAX. 150W
Sleep Mode ≤ 0.5W
Off Mode ≤ 0.3WMAX. 150W
Sleep Mode ≤ 0.5W
Off Mode ≤ 0.3WSame
Screen size676.9 x 459.7 mm676.9 x 459.7 mmSame
LCD ScreenTFT LCDTFT LCDSame
Pixel Pitch0.1554 x 0.1554 mm0.1554 x 0.1554 mmSame
Resolution4,200 x 2,800 pixels4,200 x 2,800 pixelsSame

7

Image /page/7/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with the letters "L" and "G" inside, followed by the words "LG Electronics" in gray. The logo is simple and modern, and it is easily recognizable.

| Horizontal

Frequency30 kHz to 175 kHz30 kHz to 175 kHzSame
Vertical Frequency56 Hz to 61 Hz56 Hz to 61 HzSame
Input video signalsDisplayPort x 2
HDMI x 1DisplayPort x 2
HDMI x 1Same
Calibration ToolPerfectLum / LG Calibration Studio
Medical / Calibration Feedback
SystemPerfectLum / LG Calibration Studio
MedicalModified

Table 4. Comparison of Proposed Device to Primary Predicate Device (32HQ713D)

Proposed DevicePredicate DeviceEquivalence
K NumberTBDK230845-
ManufacturerLG Electronics Inc.LG Electronics Inc.-
Model Name32HQ713D32HQ713D-
Classification
NameMedical image management and
processing systemMedical image management and
processing systemSame
Classification
Number21 CFR 892.205021 CFR 892.2050Same
Indications for UseThis Medical Monitor is indicated
for use in displaying radiological
images (including full-field digital
mammography and digital breast
tomosynthesis) for review, analysis,
and diagnosis by trained medical
practitioners.This Medical Monitor is indicated
for use in displaying radiological
images (including full-field digital
mammography and digital breast
tomosynthesis) for review, analysis,
and diagnosis by trained medical
practitioners.Same
Power
ConsumptionMAX. 120W
Off Mode ≤ 0.3WMAX. 120W
Off Mode ≤ 0.3WSame
Screen size730.7 x 425.2 mm730.7 x 425.2 mmSame
LCD ScreenTFT LCDTFT LCDSame
Pixel Pitch0.18159 x 0.18159 mm0.18159 x 0.18159 mmSame
Resolution3,840 x 2,160 pixels3,840 x 2,160 pixelsSame
Horizontal
Frequency30 kHz to 135kHz30 kHz to 135kHzSame
Vertical Frequency56 Hz to 61 Hz56 Hz to 61 HzSame
Input video signalsDVI IN x 1,
DP IN x 1
DP OUT x 1DVI IN x 1,
DP IN x 1
DP OUT x 1Same
Calibration ToolPerfectLum / LG Calibration Studio
Medical / Calibration Feedback
SystemPerfectLum / LG Calibration Studio
MedicalModified

8

Image /page/8/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the words "LG Electronics" in gray. The "L" and "G" are designed to resemble a winking face, with the "L" forming the nose and the "G" forming the face.

8. Non-Clinical Test summary

    1. Electrical Safety and Electromagnetic Compatibility
      The test results demonstrated that the proposed device complies with the following standards:
  • IEC 60601-1:2005/AMD2:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromaqnetic disturbances -Requirements and tests

    1. Software Validation

The 21HQ513D, 32HL512D, 31HN713D and 32HQ713D contain a MODERATE level of concern software, The software was designed and developed according to a software development process and was verified and validated. There have been firmware updates since FDA 510(K) clearance.

The LG Calibration Studio Medical, PerfectLum and Calibration Feedback System are moderate level of concern software programs. The software programs were verified and validated according to IEC 62304.

  • -The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005

3) Bench Test – Performance Test Report

The performance items suggested in the FDA quidance "Display Devices for Diagnostic Radiology" were tested on the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D using Calibration Feedback System.

MeasurementsTest Result
a.Spatial resolutionPass
b.Pixel defectsPass
c.ArtifactsPass
d.Temporal responsePass
e.LuminancePass
f.Conformance to a grayscale-to-luminance functionPass
g.Luminance at 30° and 45° in diagonal, horizontal, and
vertical directions at center and four cornersN/A
h.Luminance uniformity or Mura testN/A
i.Stability of luminance and chromaticity response with
temperature and time of operation (on-time)N/A
j.Spatial noiseN/A
k.Reflection coefficientN/A
l.Veiling glare or small-spot contrastN/A
m.Color trackingPass
n.Gray trackingN/A
  • -21HQ513D

9

Image /page/9/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the words "LG Electronics" in gray. The text is in a sans-serif font and is slightly blurred.

32HL512D -

MeasurementsTest Result
a.Spatial resolutionPass
b.Pixel defectsPass
c.ArtifactsPass
d.Temporal responsePass
e.LuminancePass
g.Luminance at 30° and 45° in diagonal, horizontal, andN/A
vertical directions at center and four cornersN/A
h.Luminance uniformity or Mura testN/A
i.Stability of luminance and chromaticity response withN/A
temperature and time of operation (on-time)N/A
j.Spatial noiseN/A
k.Reflection coefficientN/A
l.Veiling glare or small-spot contrastN/A
m.Color trackingPass
n.Gray trackingN/A

31HN713D -

MeasurementsTest Result
a.Spatial resolutionPass
b.Pixel defectsPass
c.ArtifactsPass
d.Temporal responsePass
e.LuminancePass
f.Conformance to a grayscale-to-luminance functionPass
g.Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four cornersPass
h.Luminance uniformity or Mura testPass
i.Stability of luminance and chromaticity response with temperature and time of operation (on-time)Pass
j.Spatial noisePass
k.Reflection coefficientPass
l.Veiling glare or small-spot contrastPass
m.Color trackingPass
n.Gray trackingPass

32HQ713D -

MeasurementsTest Result
a. Spatial resolutionPass
b. Pixel defectsPass
c. ArtifactsPass
d. Temporal responsePass
e. LuminancePass
f.Conformance to a grayscale-to-luminance functionPass
g. Luminance at 30° and 45° in diagonal, horizontal, andPass

10

Image /page/10/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized white "L" and "." inside, next to the gray text "LG Electronics". The logo is simple and modern, and it is easily recognizable.

MeasurementsTest Result
vertical directions at center and four corners
h.Luminance uniformity or Mura testPass
i.Stability of luminance and chromaticity response with
temperature and time of operation (on-time)Pass
j.Spatial noisePass
k.Reflection coefficientPass
l.Veiling glare or small-spot contrastPass
m.Color trackingPass
n.Gray trackingPass

All display characteristics of the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D have met the pre-defined criteria. Therefore, the performance of the 21HQ513D, 32HL512D, 31HN713D and 32HQ713D were verified through the performance test.

Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on September 28, 2022

Clinical Test Summary:

No clinical studies were considered necessary and performed.

9. Conclusion [21 CFR 807.92(b)(3)]

The Medical Monitor 21HQ513D, 32HL512D, 31HN713D and 32HQ713D are found to be substantially equivalent in safety and effectiveness to the predicate devices based on the information provided in this premarket notification.