(212 days)
Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
The devices subject of this submission, generator AGN-H-1.3, generator AGN-3.3 and AMICA-PROBE 17G & 18G applicators, belong to the HS AMICA devices family and they represent an evolution and implementation of previous models, already authorized by FDA under K182605. The working configuration of the HS AMICA commercial system consists of:
- . A programmable generator for the generation and control of the energy required for the thermoablative treatment;
- . Disposable applied parts (or applicators) for the direct release of energy into the patients or for temperature measurements inside the patient's body.
The result is an integrated system for thermoablation of tissues through controlled emission of nonionizing electromagnetic radiations in the microwave and radiofrequency ranges. The generators and its accessories are able to emit only microwaves (MW, 2450 MHz), or only radiofrequency waves (RF, 450 kHz) or either microwaves or radiofrequency waves (not simultaneously). The new electrosurgical devices introduced by H.S. Hospital Service S.p.A. are the AMICA-GEN models named AGN-H-1.3 and AGN-3.3, along with 17G and 18G AMICA-PROBE disposable applicators.
This document describes the premarket notification (510(k)) for the HS AMICA devices family, specifically the AMICA-GEN AGN-H-1.3 and AGN-3.3 generators, and AMICA-PROBE 17G & 18G applicators. The submission aims to establish substantial equivalence to a previously cleared device (K182605).
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly based on demonstrating substantial equivalence to the predicate devices through adherence to safety and performance standards for electrosurgical cutting and coagulation devices. The reported device performance primarily focuses on the device's functionality, adherence to electrical safety and EMC standards, and a single ex-vivo bench test for the new applicators.
Table of Acceptance Criteria (Implied) and Reported Device Performance:
| Feature/Test | Acceptance Criteria (Implied by Predicate Equivalence and Standards) | Reported Device Performance |
|---|---|---|
| Indications for Use | Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures. | Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures. (Unchanged, met) |
| Generators (AGN-H-1.3, AGN-3.3) | Evolution of predicate models (AGN-H-1.2, AGN-3.2) with increased MW rated power (190W to 250W) in pulsed mode, while maintaining principles of operation, design, architecture, and critical components. | Increased MW rated power from 190W to 250W (CW max) and 140W to 180W (pulsed mode). This change "solely affects the PULSED energy delivery mode and does not alter the total amount of microwave energy that the HS AMICA system may administer to a patient in a single treatment session." Tested for electrical safety and EMC. |
| Applicators (AMICA-PROBE 17G & 18G) | New, smaller diameter needles (17G, 18G) compared to predicate (11G, 14G, 16G); same performance specifications and manufacturing materials as authorized applicators. Sterile and disposable. | New applicators are 17G and 18G. They share the same performance specifications and manufacturing materials with already approved applicators. Sterilization method (Ethylene Oxide) revalidated. Ex-vivo test conducted to validate functional ablation performances, supporting "moving shot" technique. |
| Electrical Safety | Compliance with IEC 60601-2-6 and IEC 60601-2-2. | Complies with IEC 60601-2-6 and IEC 60601-2-2. (Met) |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2. | Complies with IEC 60601-1-2. (Met) |
| Software Verification & Validation | Validation according to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005) and IEC 62304. Moderate Level of Concern (Class B). | Software validated to specified guidance and standards. Considered "Moderate Level of Concern." "Changes introduced to adapt the functioning of the new applicators can be considered marginal and therefore only a few validation tests have been performed." |
| Biocompatibility | Not necessary if using same materials as predicate devices. | Not necessary, as manufactured using same materials as predicate devices. (Met) |
| Performance (Functional Ablation) | Demonstrated ability of new applicators to perform functional ablation, particularly for the thinnest probe (18G). | Ex-vivo test conducted for 18G-gauge AMICA-PROBE to validate functional ablation performances, supporting the "moving shot" technique. Specific quantitative metrics of performance (e.g., ablation zone size, consistency) are not detailed in this summary. |
2. Sample Size Used for the Test Set and Data Provenance:
- The document mentions "ex-vivo test" for the 18G-gauge AMICA-PROBE.
- Sample Size: Not specified for the ex-vivo test.
- Data Provenance: The ex-vivo test was conducted by HS Hospital Service Spa. The country of origin for the test is not explicitly stated, but the company is based in Italy. The study is a bench test, and inherently retrospective in the sense that the results were analyzed after the experiments. It is not a clinical study involving human subjects or patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- This information is not provided. The study appears to be a bench test rather than a clinical study requiring expert consensus on image interpretation or clinical outcomes. The "ground truth" for a bench test would be objective measurements of physical parameters (e.g., ablation volume, temperature distribution), which are typically established by the testing methodology itself, not by expert readers.
4. Adjudication Method for the Test Set:
- Not applicable as this is a bench test, not a study requiring reader adjudication on a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not performed. This submission focuses on demonstrating substantial equivalence for an electrosurgical device family, particularly new models and smaller-gauge applicators, primarily through engineering principles, electrical safety, EMC, software validation, and bench testing. It does not involve AI software or human-in-the-loop performance evaluation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
- No, a standalone algorithm performance study was not performed. This device is a hardware electrosurgical system with integrated software, not a standalone diagnostic AI algorithm.
7. Type of Ground Truth Used:
- For the ex-vivo bench test, the ground truth would be objective physical measurements related to ablation performance (e.g., thermal lesion size, shape, or temperature profiles) achieved in the ex-vivo tissue model. The document states "validate the functional ablation performances," implying that quantitative or qualitative measurements of the ablation were the "ground truth" for this specific test.
8. Sample Size for the Training Set:
- Not applicable. This device is an electrosurgical system, not an AI/ML model that undergoes a "training" phase with a dataset. The software validation is based on standard software engineering principles and testing, not machine learning.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no "training set" for this type of medical device submission.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 9, 2024
HS Hospital Service SPA % Guido Bonapace Consultant Isemed s.r.l Via Palmiro Togliatti 19/X Imola, Bologna 40026 Italy
Re: K232072
Trade/Device Name: HS AMICA devices family Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: December 13, 2023 Received: December 13, 2023
Dear Guido Bonapace:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.09 13:55:28 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K232072
Device Name HS AMICA devices family
Indications for Use (Describe) Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
1. General Information
| Submitter: | HS Hospital Service establishment is at:Via A. Vacchi, 23/25, 04011 Aprilia (LT) – ItalyPhone: +39 06 920 1961Establishment Registration Number: 8010312 |
|---|---|
| Consultant/ Contact: | Guido BonapaceISEMED srlVia P. Togliatti. 19/X,a |
Imola, BO 40026 ITALY Mob. +39 335 537 8686 Tel. +39 0542 683803 Fax +39 0542 698456 Email: regulatory@isemed.eu
Summary Prepared Date:
February 09, 2024
2. Device
Name of Device: HS AMICA devices family (AMICA-GEN AGN-H-1.3 and AGN-3.3, AMICA-PROBE 17G & 18G applicators) Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400) Regulation Class: II Product Code: GEI
3. Predicate Devices
The AMICA-GEN models AGN-H-1.3 and AGN-3.3 and AMICA-PROBE 17G & 18G applicators are claimed substantially equivalent to the following devices legally marketed in the US:
| Applicant | Device name | 510(k) Number |
|---|---|---|
| H.S. Hospital Service S.p.A. | HS AMICA devices family | K182605 |
These predicate devices have not been subject to a design-related recall.
4. Device Description
The devices subject of this submission, generator AGN-H-1.3, generator AGN-3.3 and AMICA-PROBE 17G & 18G applicators, belong to the HS AMICA devices family and they represent an evolution and implementation of previous models, already authorized by FDA under K182605. The working configuration of the HS AMICA commercial system consists of:
- . A programmable generator for the generation and control of the energy required for the thermoablative treatment;
- . Disposable applied parts (or applicators) for the direct release of energy into the patients or for temperature measurements inside the patient's body.
{5}------------------------------------------------
The result is an integrated system for thermoablation of tissues through controlled emission of nonionizing electromagnetic radiations in the microwave and radiofrequency ranges. The generators and its accessories are able to emit only microwaves (MW, 2450 MHz), or only radiofrequency waves (RF, 450 kHz) or either microwaves or radiofrequency waves (not simultaneously). The new electrosurgical devices introduced by H.S. Hospital Service S.p.A. are the AMICA-GEN models named AGN-H-1.3 and AGN-3.3, along with 17G and 18G AMICA-PROBE disposable applicators.
AGN-H-1.3 and AGN-3.3, are identical to the approved AMICA-GEN models named AGN-H-1.2 and AGN-3.2, respectively, as for principles of operation, design, architecture, and critical components; however, the new models feature a higher MW rated power, increased from 190W to 250W. This power increase solely affects the PULSED energy delivery mode and does not alter the total amount of microwave energy that the HS AMICA system may administer to a patient in a single treatment session. To support the increased MW power rating, new generator models use AC-DC power supply units with 600W DC output, whereas previously approved models used 500W AC-DC power supply units.
Both new AMICA-GEN models and previously approved models feature a monochromatic touchscreen LCD display (serving as human-machine interface), a rotary knob for parameters selection, a set of connection ports, buttons and indicators, and a built-in peristaltic pump for inner convective cooling of the applied parts used in conjunction with the generator: none of these features is changed in new models with respect to previously approved models.
The new variants of AMICA-PROBE disposable applicators are characterized by a smaller diameter of the needle and shaft length options limited to 70mm, 100mm or 150mm; these applicators better serve the purpose of further minimizing treatment invasiveness and further enhancing safety in clinical applications that prioritize the tailoring of the treatment area to the target tissues geometry rather than the absolute maximization volume. The new models of AMICA-PROBE share the same performance specifications and the same manufacturing materials with already approved applicators (K182605). The difference with previous models resides in the probe gauge (i.e., the outer diameter of the introducing cannula), which is 17G and 18G in the new models vs 11G, 14G or 16G in the previously approved devices.
As predicate devices, the new applicators are supplied in sterile conditions and cannot be resterilized; the sterile kits composed by the microwave applicators and their disposable accessories (a scalpel for skin pre-incision, and a self-adhesive drape for better delimitation of the operating area) are the same as for predicate devices. The sterilization method used for the sterile kits is Ethylene Oxide (unchanged with respect to what was described and validated in K182605).
5. Indications for Use
Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
6. Substantial equivalence comparison
In the table below a detailed comparison between subject devices (AMICA-GEN AGN-H-1.3, AGN-3.3 and AMICA PROBE applicators) and predicate devices HS AMICA devices family (K182605) is provided.
{6}------------------------------------------------
| Characteristic | Proposed Device | Main Predicate |
|---|---|---|
| AMICA-GEN AGN-H-1.3AMICA-GEN AGN-3.3 | HS AMICA devices family(AGN-H-1.2 and AGN-3.2) | |
| Manufacturer | H.S. Hospital Service S.p.A. | K182605H.S. Hospital Service S.p.A. |
| Classification Name | Electrosurgical cutting andcoagulation device and accessories. | Electrosurgical cutting and coagulation device and accessories. |
| Regulation number | 878.4400 | 878.4400 |
| Regulatory class | II | II |
| ClassificationProduct Code | GEI | GEI |
| Indications for use | Coagulation (thermoablation) of softtissue. Not for use in cardiacprocedures. | Coagulation (thermoablation) of soft tissue. Not for use incardiac procedures. |
| Prescription or OTC | Prescription use. | Prescription use. |
| Mechanism of action | Thermoablation of soft tissuesthrough controlled emission of MWenergy or RF energy(nonsimultaneous), respectively atthe tip of a microwave coaxialantenna lodged into a needle or atthe tip of RF electrodes. | Thermoablation of soft tissues through controlled emission ofMW energy or RF energy (nonsimultaneous), respectively at thetip of a microwave coaxial antenna lodged into a needle or at thetip of RF electrodes. |
| Classification | Risk class (93/42/CE as modified by2007/47/EEC): IIbElectrical class (EN60601-1): I BFEMC Class (EN60601-1-2): Class A,group 2Software Class (EN62304): B(Moderate Level of Concern) | Risk class (93/42/CE as modified by 2007/47/EEC): IIbElectrical class (EN60601-1): I BFEMC Class (EN60601-1-2): Class A, group 2Software Class (EN62304): B (Moderate Level of Concern) |
| Technical features | ||
| Design structure | ■ Generator with RF and MWmodule, integrated peristalticpump■ Footswitch | ■ Generator with RF and MW module, integrated peristalticpump■ Footswitch |
| Interface | Monochromatic display touchscreen | Monochromatic display touchscreen |
| Input AC power supply | 100-240 Vac / 50-60Hz 600VA | 100-240 Vac / 50-60Hz max 500VA |
| Fuses | 2 x F6.3AL 250V 5x20mm | 2 x T5A, 5x20, 250V |
| Protection, alarms andsafety features | Immediate suspension and disablingof microwave and / or radiofrequency supply, appropriateacoustic and light signals and displayof an appropriate error message onthe LCD in the event of:• Disconnection of the applicator;• Attempt to use incompatible orunrecognized applied parts;• Hardware anomalies at the level ofthe microwave or radio frequencymodule;Impedance detected betweenradiofrequency electrode anddispersion plates less than 25 Ohmor greater than 1000 OhmAnomalies detected by theelectrical continuity monitoringcircuit of the neutral electrodes.Excessive reflections on themicrowave coaxial line | Immediate suspension and disabling of microwave and / orradio frequency supply, appropriate acoustic and light signalsand display of an appropriate error message on the LCD in theevent of:• Disconnection of the applicator;• Attempt to use incompatible or unrecognized applied parts;• Hardware anomalies at the level of the microwave or radiofrequency module;• Impedance detected between radiofrequency electrode anddispersion plates less than 25 Ohm or greater than 1000 Ohm• Anomalies detected by the electrical continuity monitoringcircuit of the neutral electrodes.• Excessive reflections on the microwave coaxial lineHardware / software protections added to the aforementionedsoftware protections:• Availability of a button (Output Enable) with direct action, thatis not mediated by the software, for the emergency stop of theenergy supply.The Brown Out Detection circuit is activated on the CPUs thatgovern the control electronics: in the event of disturbances onthe power supply such as to bring the DC voltage applied tothe CPUs below a fixed threshold (2.7V), the reset is forced ofthe machine after a few ms, deactivating the energy supply |
| Hardware / software protections added to the aforementioned software protections:• Availability of a button (Output Enable) with direct action, that is not mediated by the software, for the emergency stop of the energy supply.• The Brown Out Detection circuit is activated on the CPUs that govern the control electronics: in the event of disturbances on the power supply such as to bring the DC voltage applied to the CPUs below a fixed threshold (2.7V), the reset is forced of the machine after a few ms, deactivating the energy supply• Filtering grid applied to the bottom surface of the cabinet, to prevent the entry of dust and other foreign bodies. | • Filtering grid applied to the bottom surface of the cabinet, to prevent the entry of dust and other foreign bodies. | |
| Mechanical features –Dimensions | Width 47 cm (18,5 in) x Depth 35 cm (13,8 in) x Height 13 cm (5,1 in);Bending tips applied to the cabinet bottom lid and lateral strap handle for easy transportation and field placing. | Width 47 cm (18,5 in) x Depth 35 cm (13,8 in) x Height 13 cm (5,1 in)Bending tips applied to the cabinet bottom lid and lateral strap handle for easy transportation and field placing. |
| Mechanical features –Weight | • AGN-H-1.3: 13 Kg (28.5 lbs)• AGN-3.3: 10.5Kg (23 lbs) | • AGN-H-1.2: 13 Kg (28.5 lbs)• AGN-3.2: 10.5Kg (23 lbs) |
| Measurements | • Measurements of forward and reflected microwave power and dissipated power on radio frequencies.Measurement range: 1-200W.Accuracy equal to ± 20% for powers> 10W and even to ± 2W for powers less than 10W• Radio frequency load impedance measurements.Measurement range: 10- 1500 Ohm.Accuracy of ± 20%• Temperature measurements:Measurement range: 10-150° C.Accuracy equal to ± 3 °C. | • Measurements of forward and reflected microwave power and dissipated power on radio frequencies.Measurement range: 1-200W.Accuracy equal to ± 20% for powers> 10W and even to ± 2W for powers less than 10W• Radio frequency load impedance measurements.• Measurement range: 10- 1500 Ohm. Accuracy of ± 20%• Temperature measurements:Measurement range: 10-150° C. Accuracy equal to ± 3 °C. |
| Radiofrequencyapplicatorconnection | Yes for AGN-H-1.3, not for AGN-3.3 | Yes for AGN-H-1.2, not for AGN-3.2 |
| Cooling systemmanagement | Automatic or manual | Automatic or manual |
| Treatment time range(min) | Up to 15 | Up to 15 |
| Microwave and radiofrequency emission features | ||
| Microwaveoutput power | 250 W CW max | 190 W CW max |
| MicrowaveOutput frequency | 2450 MHz | 2450 MHz |
| Radiofrequency OutputPower | 200W @ 50 Ohm | 200W @ 50 Ohm |
| Radiofrequency Outputfrequency | 450 kHz | 450 kHz |
| Max nominal voltage | 100 Vrms | 100 Vrms |
| Max current | 2 Arms | 2 Arms |
| Microwave powerdelivery control | Continuous or Pulsed | Continuous or Pulsed |
| Maximum Microwavepower in continuousmode | 80W | 80W |
| Maximum Microwavepower in pulsed mode | 180W (duty cycle 40% on - 60% off =72W) | 140W (duty cycle 40% on - 60% off = 56W) |
| Radiofrequency powerdelivery control | Manual or automatic or withtemperature control | Manual or automatic or withtemperature control |
| Radiofrequency poweroutput mode | Monopolar | Monopolar |
| Operative impedancerange (Ω) | From 25 to 1000 | From 25 to 1000 |
| Microwave applicators features | ||
| Microwaveapplicatorneedlediameter(Gauge) | 17-18 | 11-14-16 |
| Microwaveapplicatorlengths (mm) | 70-100-150 | 150-200-270 |
| Material in contact withthe tissue | Polytetrafluoroethylene (PTFE) | Polytetrafluoroethylene (PTFE) |
| Conditions ofuse ofapplicators | Sterile - disposable | Sterile - disposable |
{7}------------------------------------------------
{8}------------------------------------------------
None of the performance or technological differences between the new HS AMICA family devices and the identified predicate devices raises new questions of safety and effectiveness.
7. Performance Data
Subject devices have been subjected to validation testing for electrical safety, electromagnetic compatibility and software verification and validation. These tests verified and validated the proper operation and the performance of the system. The new applied parts of the system are represented by new AMICA PROBE applicators characterized by a smaller diameter compared to already authorized probes. The materials used to manufacture them are the same of the already authorized devices (K182605). These applied parts are provided in sterile condition, which was revalidated. This present in 8. Conclusion.
Biocompatibility
Biocompatibility testing was not necessary, as the proposed devices are manufactured using the same materials of predicate devices.
Electrical safety and Electromagnetic compatibility
The HS AMICA family generators, AMICA-GEN AGN-3.3 and AGN-H-1.3, have been tested for electrical safety according to IEC 60601-2-6 and IEC 60601-2-2 and for electromagnetic compatibility (EMC) according to IEC 60601-1-2. They comply with all of the medical electrical safety and electromagnetic compatibility requirements.
Software verification and validation
The software of the HS AMICA devices family has been validated according to Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices requirements FDA— May 2005 and IEC 62304, and it demonstrated to be safe and effective for its intended use. The software is a Moderate Level of Concern as per FDA guidance. All required items related to software by FDA guidance and modified with respect to the original Submission have been included in this submission.
The changes introduced to adapt the functioning of the new applicators can be considered marginal and therefore only a few validation tests have been performed and only the
{9}------------------------------------------------
documents involved in the changes have been updated with respect to those of the previous Submission (K182605).
It is believed that the documentation developed in recent years up to now can be considered consistent also in the light of the new Guideline "Content of Premarket Submission for Device Software Function" issued on June 14, 2023.
Performance bench testing
Hs Hospital Service Spa conducted an ex-vivo test in order to validate the functional ablation performances of the new 18G-gauge AMICA-PROBE applied part. This new probe represents the thinnest of the applicators present in this family. The measurements performed support the use of the "moving shot" technique, which best suits the thin probes operation.
8. Conclusions
The technological comparison and performance test data provided in this submission support the conclusion that the new HS AMICA devices, generator AGN-H-1.3, generator AGN-3.3 and AMICA-PROBE applicators 17G and 18G, are substantially equivalent to their predicate devices for requested indications for use.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.