(88 days)
The feeding tube is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication, or decompression through an established gastrointestinal stoma tract.
The Kangaroo Skin Level Gastrostomy Device Feed Set Kit with ENFit® Connection is indicated for use in adult and pediatic patients that require enteral feeding, medication administration, or decompression through a Kangaroo Skin Level Device.
The Kangaroo ™ Skin Level Balloon Gastrostomy Kit will be comprised of one Kangaroo™ Skin Level Device and one of each of the feed sets. The feed sets, continuous and bolus, are also available in Feed Set Kits packaged individually and then boxed as a 10-pack.
The Kangaroo™ SLD is constructed primarily of silicone and consists of a hub which is overmolded on to a previously extruded bi-lumen shaft. The tip of the shaft is filled with radiopaque barium RTV. At the distal end, a silicone subsumed tip balloon is reverse bonded to the shaft using a silicone RTV. This subsumed tip balloon accomplishes patient comfort by not exposing the tip of the shaft to the patient as well as being in line with the shaft. In use, the balloon end of the catheter is inserted through a gastrointestinal stoma tract and into the stomach. The balloon is then inflated with sterile or distilled water (per the recommended volume for the specific device size) to secure the device in place. To confirm placement of the device in the stomach, fluoroscopic procedures are typically used since the catheter shaft tip has radiopaque barium stripe present. Accessories for the Kangaroo™ SLD include two feed extension set confiqurations (ENFit bolus and continuous), an inflation syringe, a feeding syringe (ENFit), and gauzes.
The hub has two ports. The top (center) port contains a reflux valve assembly which is keyed to accept the feed set valve connectors. Leaks are prevented by the reflux valve assembly (with cross-slit valve design). The second port is the inflation port and contains the inflation valve.
The device is available in the configurations per determined market needs shown below in Table 5.5-2.
The feed extension sets have a valve connector end that inserts into the Kangaroo™ SLD reflux valve. For these ENFit Feed Sets, the opposite end has an ENFit-compatible connector complying with ISO 80369-3 (for prevention of misconnections with nonenteral feeding devices). The extension set can then connect to any ENFit-compatible enteral giving sets or ENFit-compatible syringes. Finally, both the Kangaroo™ Skin Level Device and the Kangaroo™ Feed Sets are compatible with competitor devices currently on the market.
This document is a 510(k) summary for the Kangaroo Skin Level Balloon Gastrostomy Kit and associated feed sets. It does not describe a study involving an AI/ML device, but rather a traditional medical device (gastrostomy kit). Therefore, many of the requested categories related to AI/ML device studies, such as the use of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable.
However, I can extract information regarding the device's acceptance criteria and how its performance was demonstrated through non-clinical testing.
Here's the summary of the acceptance criteria and study information that is applicable:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a specific table of acceptance criteria with numerical performance results directly in the format requested. Instead, it states that results from all tests "met the acceptance criteria for each test."
However, it lists the types of performance tests conducted based on relevant standards. The "acceptance criteria" can be inferred as the requirements set by these standards:
| Acceptance Criteria Category (Inferred from Standards) | Reported Device Performance |
|---|---|
| Biocompatibility (Cytotoxicity, Sensitization, Irritation) | Met acceptance criteria (deemed applicable from predicate device) |
| Sterilization (EO residuals) | Met acceptance criteria (deemed applicable from predicate device) |
| Resistance to device migration (hub or balloon) | Met acceptance criteria |
| Enteral feeding devices testing (ASTM F2528-06(2014)) | Met acceptance criteria |
| Tensile testing (EN 1615:2000 and EN 1618:1997) | Met acceptance criteria |
| Leak testing (EN1615:2000, EN 1618:1997, ISO 20695:2020) | Met acceptance criteria |
| Valve connector insertion/removal force testing (ISO 20695:2020) | Met acceptance criteria |
| Low profile device-valve connector torque testing | Met acceptance criteria |
| Feed/extension set pinch clamp testing (stop fluid flow) | Met acceptance criteria |
| Resistance to expected chemical exposure | Met acceptance criteria |
| Distribution simulation testing | Met acceptance criteria |
| Design validation testing (simulated use, blenderized diet) | Met acceptance criteria |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes (number of devices) used for each specific test mentioned under "Performance Testing Bench." It only indicates that "all the above tests met the acceptance criteria," implying that sufficient units were tested to demonstrate compliance.
The data provenance is from non-clinical bench testing conducted by the manufacturer, Xeridiem Medical Devices, on the proposed devices, or leveraged from previously conducted tests on the identified predicate and reference devices. There is no mention of country of origin for data; it is assumed to be from the manufacturer's internal testing. The data is prospective in the sense that the tests were conducted specifically to support this 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
Not applicable, as this is for a traditional medical device and not an AI/ML diagnostic or interpretive device. The "ground truth" here is the performance of the device against engineering and safety standards, directly measured through physical tests.
4. Adjudication Method for the Test Set
Not applicable. Device performance was assessed against predefined engineering and safety standards, not through human interpretation requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No. This is not an AI/ML device, and therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a traditional medical device.
7. The Type of Ground Truth Used
The "ground truth" for the performance evaluation of this traditional medical device is based on established engineering and safety standards (e.g., ISO 80369-3, EN1615, ISO 20695, ASTM F2528-06, ISO 10993-1). The device's physical and functional characteristics are directly measured and compared against the requirements specified in these standards.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device, so there is no "training set."
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 6, 2023
Xeridiem Medical Devices, A Spectrum Plastics Group Company Mr. Steve Murray Regulatory Affairs Specialist 4700 S. Overland Dr. Tucson, AZ 85714
Re: K232046
Trade/Device Name: Kangaroo Skin Level Balloon Gastrostomy Kit, Kangaroo Bolus Feed Set with ENFIT Connection, Kangaroo Continuous Feed Set with ENFIT Connection Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PIF Dated: July 5, 2023 Received: July 10, 2023
Dear Steve Murray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K232046
Device Name Kangaroo Skin Level Balloon Gastrostomy Kit Kangaroo Bolus Feed Set with ENFIT Connection Kangaroo Continuous Feed Set with ENFIT Connection
Indications for Use (Describe)
The feeding tube is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication, or decompression through an established gastrointestinal stoma tract.
The Kangaroo Skin Level Gastrostomy Device Feed Set Kit with ENFit® Connection is indicated for use in adult and pediatic patients that require enteral feeding, medication administration, or decompression through a Kangaroo Skin Level Device.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5.0 510(k) SUMMARY
5.1 Submitter Information
| Date Prepared: | June 30, 2023 |
|---|---|
| 510(k) Submitter: | Xeridiem Medical Devices, |
| A Spectrum Plastics Group Company | |
| 4700 S. Overland Dr. | |
| Tucson, AZ 85714 | |
| Contact Person: | Steve Murray, Regulatory Affairs Special |
| Contact Person: | Steve Murray, Regulatory Affairs Specialist(520) 882-7794 ext. 178steven.murray@spectrumplastics.com |
|---|---|
| ----------------- | -------------------------------------------------------------------------------------------------------------- |
5.2 Device Information
| Trade Name: | Kangaroo™ Skin Level Balloon Gastrostomy Kit |
|---|---|
| Kangaroo™ Bolus Feed Set with ENFit® Connection | |
| Kangaroo™ Continuous Feed Set with ENFit® Connection |
| Common Name: | Low Profile Gastrostomy Tube |
|---|---|
| Classification Name: | Gastrointestinal tube and accessories(21 CFR 876.5980, Product Code PIF) |
| Classification Panel: | Gastroenterology/Urology |
Predicate Device Information 5.3
| Trade Name: | Entuit® LP Gastrostomy BR Balloon RetentionFeeding Tube Kit with ENFit® Connection |
|---|---|
| Common Name: | Low Profile Gastrostomy Tube |
| Classification Name: | Gastrointestinal tube and accessories(21 CFR 876.5980, Product Code PIF) |
| 510(k) Number: | K180708 |
Reference Device Information 5.4
| Trade Name: | EndoVive™ 3s Low Profile Balloon Kit*EndoVive™ 3s Bolus Extension SetsEndoVive™ 3s Continuous Extension SetsEndoVive™ 3s Medication Extension Set |
|---|---|
| Common Name: | Gastrostomy Tube |
| Classification Name: | Gastrointestinal tube and accessories(21 CFR 876.5980, Product Codes PIF, PIO) |
| 510(k) Number: | K142297 |
{4}------------------------------------------------
- Low profile device from the Low Profile Balloon Kit is the only part of the cleared device beinq used for reference considerations in technological characteristics.
5.5 Device Description
The proposed devices are listed below in Table 5.5-1 and are described following the table.
| Device PartNumber | Device Brand Name | Hereafter Referred to As |
|---|---|---|
| 50-0508 | Kangaroo™ Skin Level Device | Kangaroo™ SLD |
| 70-0079 | Kangaroo™ Skin Level Balloon Gastrostomy Kit | Kangaroo™ SLD Kit |
| 20-1724 | Kangaroo™ 24" Continuous Feed Extension Set | Continuous ES |
| 20-1725 | Kangaroo™ 12" Bolus Feed Extension Set | Bolus ES |
| 50-0509 | Kangaroo™ Bolus Feed Set with ENFit® Connection (10 pack of Bolus Feed Set) | FSK - Bolus |
| 50-0510 | Kangaroo™ Continuous Feed Set with ENFit® Connection (10 pack of Continuous Feed Set) | FSK - Continuous |
Table 5.5-1 - Proposed Devices
The Kangaroo ™ Skin Level Balloon Gastrostomy Kit will be comprised of one Kangaroo™ Skin Level Device and one of each of the feed sets. The feed sets, continuous and bolus, are also available in Feed Set Kits packaged individually and then boxed as a 10-pack.
The Kangaroo™ SLD is constructed primarily of silicone and consists of a hub which is overmolded on to a previously extruded bi-lumen shaft. The tip of the shaft is filled with radiopaque barium RTV. At the distal end, a silicone subsumed tip balloon is reverse bonded to the shaft using a silicone RTV. This subsumed tip balloon accomplishes patient comfort by not exposing the tip of the shaft to the patient as well as being in line with the shaft. In use, the balloon end of the catheter is inserted through a gastrointestinal stoma tract and into the stomach. The balloon is then inflated with sterile or distilled water (per the recommended volume for the specific device size) to secure the device in place. To confirm placement of the device in the stomach, fluoroscopic procedures are typically used since the catheter shaft tip has radiopaque
{5}------------------------------------------------
barium stripe present. Accessories for the Kangaroo™ SLD include two feed extension set confiqurations (ENFit bolus and continuous), an inflation syringe, a feeding syringe (ENFit), and gauzes.
The hub has two ports. The top (center) port contains a reflux valve assembly which is keyed to accept the feed set valve connectors. Leaks are prevented by the reflux valve assembly (with cross-slit valve design). The second port is the inflation port and contains the inflation valve.
The device is available in the configurations per determined market needs shown below in Table 5.5-2.
| FR | LENGTH (CM) | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SIZE | 0.8 - | 1.0 1.2 | 1.5 | 1.7 | 2.0 | 2.3 | 2.5 | 2.7 3.0 | 3.5 | 4.0 4.5 | 5.0 | |||
| 12 | イ | > | > | " < > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > | > | " √ √ . | > | > | = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = | > | > | > | ||
| 14 | イ | イ | > | > | > | > | > | V / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / | > = | > | > | イ | ||
| 16 | > | > | > | イ | > | > | > | : | V . | > | > | > | > | > |
| 18 | > | イ | > | 「 ノ ・ | > | > | > | >>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>> | >>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>> | > | イ | " √ . | > | > |
| 20 | > | イ | > | > | > | >> | " √ . | > | " √ . | > = | VI | > | > | |
| 22 | > | イ | > | > | > | > | > | > | V I | > | > | > | > | イ |
| 24 | > | > | > | > | > | ! ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ | > | イ | > | > | V | > | ! ノ ・ |
| Table 5.5-2 – Kangaroo™ SLD Configurations (Length [cm] for Each FR Size) | |
|---|---|
| FR | LENGTH (CM) |
The feed extension sets have a valve connector end that inserts into the Kangaroo™ SLD reflux valve. For these ENFit Feed Sets, the opposite end has an ENFit-compatible connector complying with ISO 80369-3 (for prevention of misconnections with nonenteral feeding devices). The extension set can then connect to any ENFit-compatible enteral giving sets or ENFit-compatible syringes. Finally, both the Kangaroo™ Skin Level Device and the Kangaroo™ Feed Sets are compatible with competitor devices currently on the market.
5.6 Indications for Use
5.6.1 KangarooTM Skin Level Balloon Gastrostomy Kit
The feeding tube is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication administration, or decompression through an established gastrointestinal stoma tract.
The Kanqaroo™ Skin Level Gastrostomy Device Feed Set Kit with ENFit® Connection is indicated for use in adult and pediatric patients that require enteral feeding, medication administration, or decompression through a Kangaroo™ Skin Level Device.
{6}------------------------------------------------
5.6.2 Kangaroo™ Bolus Feed Set with ENFit® Connection and Kangaroo™ Continuous Feed Set with ENFit® Connection (10-Pack Kits)
The Kangaroo™ Skin Level Gastrostomy Device Feed Set Kit with ENFit® Connection is indicated for use in adult and pediatric patients that require enteral feeding, medication administration, or decompression through a Kangaroo™ Skin Level Device.
5.7 Technological Characteristics
The proposed device, as shown in Table 5.7 below with specific similarities and differences, have the same or very similar technological characteristics as the predicate device. The intended uses of the proposed device and predicate device are the same and there are no significant differences noted in the table below that would affect the performance of the proposed device as compared to the predicate device. The reference device is used as the comparison for certain balloon characteristics (see Table 5.7a below the main table).
| PREDICATE DEVICE(K180708) | PROPOSED DEVICE | |
|---|---|---|
| Entuit® LP Gastrostomy BR BalloonRetention Feeding Tube Kit withENFit® Connection | Kangaroo™ Skin Level Device Kitwith Feed Sets | |
| FDA Class | Class II | Same as K180708 |
| FDA Product Code | PIF (for ENFit-compatible) | Same as K180708 |
| FDA Regulation | 876.5980 - Tubes, gastrointestinal(and accessories) | Same as K180708 |
| 510(k) Number | K180708 | TBD |
| Indications for Use | The feeding tube is indicated foruse in percutaneous placement ofan enteral feeding tube in adult andpediatric patients that requireenteral feeding, medicationadministration or decompressionthrough an establishedgastrointestinal stoma tract. | Same as K180708 |
| User Population | Adult and Pediatric | Same as K180708 |
| Operating Principle | Device tip enters through the stomatract and into the stomach, then thedevice is secured in place byinflating the retention balloon. Oncesecured in place, the appropriateextension set is connected to thedevice and it then serves as achannel for administering nutrientsand medication into the stomachand for gastric decompression. | Same as K180708 |
| Table 5.7 - Device Comparison Table (Predicate Device) | |||
|---|---|---|---|
| PREDICATE DEVICE(K180708) | PROPOSED DEVICE | ||
| Entuit® LP Gastrostomy BR BalloonRetention Feeding Tube Kit withENFit® Connection | Kangaroo™ Skin Level Device Kitwith Feed Sets | ||
| Sterile | Yes (ETO) for pouch-sealed lowprofile device | Same as K180708 | |
| Shaft | Silicone 65D | Silicone 65D (same as K180708);adds four additional functionallengths: 1.2cm, 1.7cm, 2.3cm,2.7cm | |
| Balloon | Silicone 24D | Silicone 24D (with reverse bondingfor subsumed device tip)* | |
| Balloon Inflation Rating | 12-16 FR tubing size: 5-7 mL18-24 FR tubing size: 5-10 mL | 12 FR tubing size: 4 mL*14 FR tubing size: 4.5 mL*16 FR tubing size: 5 mL*18 FR tubing size: 6 mL*20 FR tubing size: 7 mL*22 FR tubing size: 8 mL*24 FR tubing size: 10 mL* | |
| Reflux Valve | Silicone cross-slit valve | Silicone cross-slit valve (withsilicone back-up seal) | |
| Radiopaque Marker | Barium sulfate | Same as K180708 | |
| Balloon Inflation Valve | Thermoplastic housing with siliconevalve | Same as K180708 | |
| Inflation Valve Port | Port on Silicone 60D hub whichhouses inflation valve | Same as K180708 | |
| Connectors (FeedExtension Sets) | Valve Connector - Polycarbonate,straight (bolus) or right angle(continuous) | Same as K180708 | |
| Pinch Clamps (FeedExtension Sets) | Feeding Connector – Nylon (ENFit)Both Bolus and Continuous ES -print "Not for IV Use" symbol and"For Enteral Use Only" warning onpinch clamp;Bolus ES - 6.8 mm ODpolypropylene pinch clampContinuous ES – 3.0 mm ODpolypropylene pinch clamp | Same as K180708 | |
| Low Profile KitContents | Inflation syringe, feeding syringe,gauze, Bolus Extension Set,Continuous Extension Set | Same as K180708 | |
| Replacement FeedExtension Kit Contents | 1 each Bolus Extension Set andContinuous Extension Set | 10 each Bolus Feed Extension Setsor 10 each Continuous FeedExtension Sets | |
| Packaging System/Materials | Tyvek Pouch (only with LPK),Polybag, IFU, Corrugate Case &Shipper | Same as K180708 | |
| Competitor DeviceCompatibility | No | Yes - for select competitor devices(competitor ES with Kangaroo SLD;competitor LP device withKangaroo Feed Sets) |
{7}------------------------------------------------
{8}------------------------------------------------
*The Reference Device comparison in Table 5.7a supports these balloon characteristics.
| Table 5.7a - Device Comparison Table (Reference Device) | ||
|---|---|---|
| REFERENCE DEVICE(K142297) | PROPOSED DEVICE | |
| EndoViveTM 3s Low ProfileBalloon Kit | KangarooTM Skin Level DeviceKit with Feed Sets | |
| Indications for Use(applicable to low profiledevice) | The EndoVive 3S Low ProfileBalloon is indicated for use inadult and pediatric patients whorequire long term feeding, areunable to tolerate oral feeding,are at low risk for aspiration orrequire gastric decompressionand/or medication deliverydirectly into the stomach. | The feeding tube is indicated foruse in percutaneous placementof an enteral feeding tube inadult and pediatric patients thatrequire enteral feeding,medication administration ordecompression through anestablished gastrointestinalstoma tract. |
| Balloon | Silicone 24D (with reversebonding for subsumed devicetip) | Same as K142297 |
| Balloon Inflation Rating | 12 FR tubing size: 4 mL14 FR tubing size: 4.5 mL16 FR tubing size: 5 mL18 FR tubing size: 6 mL20 FR tubing size: 7 mL22 FR tubing size: 8 mL24 FR tubing size: 10 mL | Same as K142297 |
As can be seen from the above tables, the changes are limited in scope and many characteristics are unchanged from the predicate device. The limited differences are as follows:
- . Addition of four additional functional lengths on the Skin Level Device (low profile gastrostomy tube) only offers additional options but does not impact functionality of the device;
- . Balloon is reverse bonded to create a subsumed device tip for enhanced patient comfort while retaining equivalent performance in retention;
- . The balloon inflation volume rating difference is due to the reverse bonded balloon process; the balloon is reversed and trimmed, resulting in less surface area of the balloon; the balloon is then positioned over the shaft, crating a subsumed tip when inflated; this process results in enhanced patient comfort but a decrease in inflation volume for most FR sizes; however, performance testing indicates retention requirements are met;
- . The updated reflux valve design (cross-slit valve with silicone back-up seal) performs the same function as the predicate device's reflux valve and is shown to be equivalent or better based on performance testing;
- The change from providing replacement feed extension sets in a pack of 1 each ● bolus/continuous to 10-pack of either bolus or continuous simply provides
{9}------------------------------------------------
convenient options for replacing the type set a specific user needs and has no impact on safety or effectiveness;
- Compatibility of the Kangaroo™ devices with select competitor devices offers . more flexibility to the end user and has been verified through performance testing.
Any differences noted above do not raise new or different questions of safety or effectiveness as compared to the predicate device. The proposed device is deemed substantially equivalent.
5.8 Performance Data
5.8.1 Non-Clinical Tests
Based on detailed analysis of the proposed devices as compared to the predicate device (documented design equivalency analysis) and considering previous testing on materials in both the predicate and reference devices, it was determined what new Verification and Validation Testing was required for the proposed devices and what testing can be legitimately leveraged from the predicate and reference devices. Based on the combination of new and leveraged testing, it was found the proposed devices are in compliance with design and performance requirements according to ISO 80369-3:2016/AMD1:2019 Small-bore connectors for liquids and gases in healthcare applications – Part 3: Connectors for enteral applications, EN1615:2000 Enteral feeding catheters and enteral giving sets for single use and their connectors - Design and testing (with additional support from ISO 20695:2020 Enteral feeding systems - design and testing), and biocompatibility requirements per ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
The following testing was conducted on proposed devices or leveraged from the predicate device or reference device (leveraging based on design equivalency) to support substantial equivalence of the proposed devices:
Biocompatibility
-
Per prior assessment consistent with ISO 10993-1:2018, confirming cytotoxicity, ● sensitization and irritation tests completed on predicate device and deemed applicable to proposed device per design equivalency assessment
Sterilization -
Based on sterilization adoption assessment and design equivalency assessment, ● EO residuals testing on predicate device deemed applicable to proposed device
-
Performance Testing Bench (combined new and leveraged)
- Resistance to device migration (hub or balloon) ●
- Enteral feeding devices (with a retention balloon) testing per ASTM F2528-. 06(2014)
{10}------------------------------------------------
- Tensile testing per EN 1615:2000 and EN 1618:1997 ●
- Leak testing per EN1615:2000, EN 1618:1997 and ISO 20695:2020 ●
- . Valve connector insertion/removal force testing per ISO 20695:2020
- Low profile device-valve connector torque testing
- Feed/extension set pinch clamp testing (able to stop fluid flow) ●
- Testing for resistance to expected chemical exposure .
- Distribution simulation testing
- Design validation testing per simulated use conditions, including blenderized diet ●
Results from all the above tests met the acceptance criteria for each test.
5.8.2 Animal Tests
Animal tests were not required to demonstrate the performance of the proposed devices. Product functionality has been adequately assessed by the non-clinical tests discussed in 5.8.1.
5.8.3 Clinical Tests
Clinical tests were not required to demonstrate the performance of the proposed devices. Product functionality has been adequately assessed by the non-clinical tests discussed in 5.8.1.
5.9 Conclusion
The conclusions drawn from the device comparison for technological characteristics and results of the non-clinical testing demonstrate that the proposed devices in the Kangaroo™ device family are as safe, effective, and perform as well or better than the legally marketed predicate device identified in 5.3, and/or the legally marketed reference device identified in 5.4.
(End of Section)
§ 876.5980 Gastrointestinal tube and accessories.
(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.