K Number
K231960
Date Cleared
2023-10-03

(92 days)

Product Code
Regulation Number
876.1310
Panel
GU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NaviCam Xpress Stomach Capsule is intended for visualization of the stomach of patients >= 6 years old with BMI =

Device Description

NaviCam® Xpress Stomach Capsule Endoscope System enables the operator to control the capsule endoscope inside the stomach. The capsule endoscope can be moved in any direction during examination. The direction and angulation of the capsule camera head can be adjusted in real-time, hence realizing complete control of capsule endoscope movement and attitude during examination in the X left and right, Y forward and backward, Z up and down directions as well as rotation clockwise and counterclockwise.

AI/ML Overview

This document (K231960) describes a 510(k) premarket notification for the NaviCam Xpress Stomach Capsule Endoscope System. The filing is an expansion of the Indications for Use for an already cleared device, primarily expanding age and BMI limitations. As such, the performance study focuses on demonstrating that the device's performance is not adversely affected by these expanded patient characteristics.

Here's an analysis of the acceptance criteria and study information provided:

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in terms of metrics like sensitivity, specificity, or accuracy for a specific diagnostic task from an AI component. This is because this 510(k) notification is for an expanded indication for use of an existing device, not for the clearance of a new AI-powered diagnostic algorithm.

The "performance data" section focuses on demonstrating that the existing device's performance is not affected by the expanded age and BMI patient populations. Therefore, the "acceptance criteria" are implicitly that the device continues to perform as intended and safely within the expanded patient population.

Acceptance Criteria (Implied)Reported Device Performance
Device performance is not adversely affected by increased BMI up to 65."The company provided technical data demonstrating capsule and system performance would not be affected by increasing the BMI of the patient up to the limitations stated in the report."
Device performance is not adversely affected by lower age patients (down to 6 years old)."The clinical data demonstrate using the system on lower age patients... doesn't affect performance of the device."
Device remains safe in expanded patient populations."This real-world evidence demonstrates the safety..." and "does not pose any new risks to the patient as demonstrated using the device in clinical settings."

2. Sample Size and Data Provenance

  • Test Set Sample Size: The document mentions "clinical data" and "real-world evidence" but does not provide a specific sample size for a test set. This suggests that a formal, prospective, rigorously controlled "test set" in the context of an AI study was not conducted. Instead, they relied on existing clinical literature and internal reports.
  • Data Provenance: The document does not specify the country of origin. The data is described as "published in the scientific literature along with unpublished clinical reports" and "real-world evidence." This implies a mix of retrospective and possibly ongoing observational data.

3. Number of Experts and Qualifications

The document does not mention the number or qualifications of experts used to establish ground truth. This is expected given that the filing is not for an AI diagnostic algorithm requiring this type of ground truth establishment for its performance evaluation. The "ground truth" seems to be the continued functionality and safety of the device itself in the expanded population.

4. Adjudication Method

No adjudication method is mentioned, as it's not applicable to this type of device modification submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was done, as this submission is not about the effectiveness of a diagnostic algorithm or how it improves human reader performance.

6. Standalone (Algorithm Only) Performance Study

No standalone algorithm performance study was done. This device is a capsule endoscope system, and while it has software, the submission is not for a new AI-driven diagnostic or analytical algorithm.

7. Type of Ground Truth Used

The "ground truth" implicitly used for this submission is clinical experience and technical (bench) data demonstrating the continued functional performance and safety of the existing device when used in the expanded age and BMI populations. It is not about diagnostic accuracy against a specific disease state.

8. Sample Size for Training Set

No training set sample size is mentioned. This is not a submission for a new AI/ML algorithm that would undergo a training phase.

9. How Ground Truth for Training Set was Established

Not applicable, as no training set for an AI/ML algorithm is discussed.

§ 876.1310 Magnetically maneuvered capsule endoscopy system.

(a)
Identification. A magnetically maneuvered capsule endoscopy system consists of an ingestible capsule and magnetic controller and is used for visualization of the stomach and duodenum. The ingestible capsule contains a camera that wirelessly captures images of the mucosa. The magnetic controller is used outside of the patient and is magnetically coupled with the capsule to control its location and viewing direction.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing with the device under anticipated conditions of use must evaluate visualization of the intended region and document the adverse event profile.
(2) Non-clinical testing data must demonstrate the optical, mechanical, and functional integrity of the device under physically stressed conditions. The following performance characteristics must be tested, and detailed protocols must be provided for each test:
(i) A bite test must be performed to ensure that the capsule can withstand extreme cases of biting;
(ii) A pH resistance test must be performed to evaluate integrity of the capsule when exposed to a physiological relevant range of pH values;
(iii) A battery life test must be performed to demonstrate that the capsule's operating time is not constrained by the battery capacity;
(iv) A shelf life test must be performed to demonstrate that the device performs as intended at the proposed shelf life date;
(v) Optical testing must be performed to evaluate fundamental image quality characteristics such as resolution, field of view, depth of field, geometric distortion, signal to noise ratio, dynamic range, and image intensity uniformity;
(vi) A color performance test must be performed to compare the color differences between the input scene and output image;
(vii) A photobiological safety analysis must be performed based on maximum (worst-case) light exposure to internal gastrointestinal mucosa, and covering ultraviolet, visible, and near-infrared ranges, as appropriate. A mitigation analysis must be provided;
(viii) Performance testing must demonstrate that the viewing software clearly presents the current frame rate, which is either adjustable manually by the user or automatically by the device. Testing must demonstrate that the viewing software alerts the user when the video quality is reduced from nominal due to imaging data communication or computation problems;
(ix) A data transmission test must be performed to verify the robustness of the data transmission between the capsule and the receiver. This test must include controlled signal attenuation for simulating a non-ideal environment; and
(x) Magnetic field strength testing characterization must be performed to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices, or objects.
(3) Software validation, verification, and hazard analysis must be provided.
(4) Electrical safety, thermal safety, mechanical safety, and electromagnetic compatibility testing must be performed.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Performance data must validate the reprocessing instructions for the reusable components of the device.
(7) Performance data must demonstrate the sterility of any device components labeled sterile.
(8) Human factors testing must demonstrate that the intended users can safely and correctly use the device, based solely on reading the instructions for use.
(9) Clinician labeling must include:
(i) Specific instructions and the clinical and technical expertise needed for the safe use of the device;
(ii) A detailed summary of the clinical testing pertinent to use of the device, including information on effectiveness and device- and procedure-related complications;
(iii) The patient preparation procedure;
(iv) A detailed summary of the device technical parameters;
(v) Magnetic field safe zones;
(vi) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices, or objects near the external magnet;
(vii) Reprocessing instructions for reusable components;
(viii) Shelf life for single use components; and
(ix) Use life for reusable components.
(10) Patient labeling must include:
(i) An explanation of the device and the mechanism of operation;
(ii) The patient preparation procedure;
(iii) A brief summary of the clinical study; and
(iv) A summary of the device- and procedure-related complications pertinent to use of the device.