K Number
K231838
Manufacturer
Date Cleared
2023-08-15

(54 days)

Product Code
Regulation Number
886.4300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3-piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device (Model 63002) is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

Device Description

The RxSight Insertion Device (Model 63002) is a sterile, single-use device to be used to fold and insert intraocular lenses (IOL) into the eye through a small incision during cataract surgery. The RxSight Insertion Device (Model 63002) consists of:

  • A single-use, disposable, sterile IOL inserter with a non-pyrogenic cartridge
  • A single-use, disposable, sterile haptic puller.
AI/ML Overview

The provided text describes the 510(k) summary for the RxSight® Insertion Device (Model 63002). This document focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical study with specific acceptance criteria and detailed performance data often seen for novel AI/ML devices. As such, information regarding AI-specific acceptance criteria, test set details (sample size, provenance, expert adjudication, MRMC studies), ground truth establishment for training and test sets, and separate standalone algorithm performance is not present in this submission.

The device in question, an intraocular lens insertion device, does not appear to be an AI/ML product. The performance testing is described as non-clinical and pertains to the physical manipulation and integrity of the intraocular lenses after insertion, following ISO standards.

Here's a breakdown of the available information based on your request, with N/A for criteria not addressed in the provided document, as it is not an AI/ML device submission:

Acceptance Criteria and Device Performance

The submission focuses on demonstrating substantial equivalence through a comparison of technological characteristics and non-clinical performance testing. The "acceptance criteria" are implied by adherence to ISO standards for IOLs, indicating that the device should not adversely affect the IOL's mechanical dimensions, sagitta, optical properties, or homogeneity.

Acceptance Criteria (Implicit from ISO Standards)Reported Device Performance (Summary)
Maintenance of IOL mechanical dimensions and sagitta post-injection (per ISO 11979-3:2012)"mechanical dimensions and sagitta were verified following lens delivery and compared to the measurements performed prior to the lens delivery per ISO 11979-3:2012." Conclusion: Testing established safety and effectiveness.
Maintenance of IOL optical properties post-injection (per ISO 11979-2:2014)"optical properties... were tested pre and post lens injection per ISO 11979-2:2014." Conclusion: Testing established safety and effectiveness.
Maintenance of IOL overall surface and bulk homogeneity post-injection (per ISO 11979-2:2014)"overall surface and bulk homogeneity were tested pre and post lens injection per ISO 11979-2:2014." Conclusion: Testing established safety and effectiveness.
Device ensures safe and effective delivery of specified IOLs (e.g., RxSight Light Adjustable Lens, Light Adjustable Lens+)."Non-clinical data establish that the device is safe and effective in the delivery of the RxSight Light Adjustable Lens and Light Adjustable Lens+ and is shown to be substantially equivalent to the predicate devices."

Study Details (as applicable for a non-AI/ML device submission)

  1. A table of acceptance criteria and the reported device performance: See table above.

  2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for specific tests, but general non-clinical performance testing was conducted.
    • Data Provenance: Not specified, but generally, non-clinical lab testing data.
    • Retrospective or Prospective: N/A (for non-clinical lab testing).
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A. This is a non-clinical device; "ground truth" as you'd define it for AI/ML diagnostics (e.g., disease presence as determined by experts) does not apply. Performance is measured against physical and optical standards.

  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A. Not relevant for this type of non-clinical device testing.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A. This is not an AI/ML device.

  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A. This is not an AI/ML device.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" in this context refers to the defined specifications and standards for IOLs (e.g., mechanical dimensions, optical properties, homogeneity) as outlined in ISO 11979-2:2014 and ISO 11979-3:2012.

  8. The sample size for the training set: N/A. This is not an AI/ML device, so there is no "training set."

  9. How the ground truth for the training set was established: N/A. Not an AI/ML device.

{0}------------------------------------------------

August 15, 2023

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration".

RxSight, Inc. Maureen OConnell Senior Vice President, Clinical and Regulatory Affairs 100 Columbia Aliso Viejo, California 92656

Re: K231838

Trade/Device Name: RxSight® Insertion Device (63002) Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I, reserved Product Code: MSS Dated: June 22, 2023 Received: June 22, 2023

Dear Ms. Maureen OConnell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Y. I

for Bennett Walker, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231838

Device Name

RxSight® Insertion Device (63002)

Indications for Use (Describe)

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3-piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device (Model 63002) is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY RxSight® Insertion Device (Model 63002)

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(a).

APPLICANT:RxSight, Inc.100 ColumbiaAliso Viejo, CA 92656
CONTACT PERSON:Maureen O'ConnellSr. Vice President Clinical and Regulatory Affairsmoconnell@rxsight.comTel: (978) 207-1245
DATE SUMMARYPREPARED:August 15, 2023
TRADE NAME:RxSight Insertion Device (Model 63002)
COMMON NAME:IOL Injector
CLASSIFICATION NAME:Intraocular Lens (IOL) Guide
DEVICE CLASSIFICATION:Class I; 21 CFR 886.4300
PRODUCT CODE:MSS
PREDICATE DEVICE(S):Primary: RxSight Insertion Device, K231466Secondary: pioli IOL Delivery System, K172228

1.1 DEVICE DESCRIPTION

The RxSight Insertion Device (Model 63002) is a sterile, single-use device to be used to fold and insert intraocular lenses (IOL) into the eye through a small incision during cataract surgery. The RxSight Insertion Device (Model 63002) consists of:

  • A single-use, disposable, sterile IOL inserter with a non-pyrogenic cartridge ●
  • A single-use, disposable, sterile haptic puller. ●

1.2 INDICATIONS FOR USE

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight

{4}------------------------------------------------

Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

1.3 TECHNOLOGICAL CHARACTERISTICS COMPARISON

RxSight believes that the RxSight Insertion Device (Model 63002) described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to legally marketed predicate devices that are Class I medical devices. The following table compares the RxSight Insertion Device (Model 63002) with the primary predicate device which is the RxSight Insertion Device (Model 63000/63001) cleared most recently in K231466 and the secondary predicate device, the AST Products pioli IOL Delivery System cleared in K172228.

Proposed DevicePrimary PredicateDeviceSecondary PredicateDevice
RxSight Insertion Device(Model 63002)RxSight Insertion Device(Model 63000/63001)AST Productspioli IOL Delivery System
510(k) Number-K231466K172228
Product CodeMSSMSSMSS
Intended UseIntraocular Lens DeliveryIntraocular Lens DeliveryIntraocular Lens Delivery
Indications for UseThe RxSight® InsertionDevice (Model 63002) isindicated for the folding andinsertion of a 3-piecesilicone, intraocular lens intothe human eye through asurgical incision. TheRxSight Insertion Device(Model 63002) is intendedfor the insertion of theRxSight Light AdjustableLens®, RxSight LightAdjustable Lens+, and IOLmodels validated for use withthis device in IOL approvedlabeling.The RxSight® InsertionDevice is indicated for thefolding and insertion of a 3-piece silicone, intraocularlens into the human eyethrough a surgical incision.The RxSight Insertion Deviceis intended for the insertionof the RxSight LightAdjustable Lens®, theRxSight Light AdjustableLens+, the Bausch & LombLI61A0 IOL and IOL modelsvalidated for use with thisdevice in IOL approvedlabeling.The pioli IOL Delivery systemis a single-use, sterile deviceintended to insert a single-piece foldable intraocular lens(IOL) into the human eyethrough a surgical procedure.The system provides a tubularpathway for lens implantationthrough an incision. The pioliIOL Delivery System is onlyfor the insertion of the LenstecSoftec I IOL and IOL modelsvalidated for use with thisdevice as indicated in the IOLapproved labeling.
Operating PrincipleAn IOL is placed in a loadingcartridge. Screw plungeradvances the IOL through thecartridge which folds the IOLand advances it into the eye.An IOL is placed in a loadingcartridge. Cartridge snappedinto the handpiece. Screwplunger advances the IOLthrough the cartridge whichfolds the IOL and advances itinto the eye.Cartridge is back loaded intothe injection system with aplunger that advances the IOLthrough the cartridge and intothe eye
Pre-loaded IOLNoNoNo
Material (Injector)PolycarbonateTitaniumPolypropylene

COMPARISON OF THE RXSIGHT INSERTION DEVICE (MODEL 63002) TO THE PREDICATE DEVICES

{5}------------------------------------------------

Proposed DevicePrimary PredicateDeviceSecondary PredicateDevice
RxSight Insertion Device(Model 63002)RxSight Insertion Device(Model 63000/63001)AST Productspioli IOL Delivery System
Polybutylene terephthalatepolymer
Material (cartridge)PolypropylenePolypropylenePolypropylene
Cartridge CoatingLubriMATRIX™LubriMATRIX™LubriMATRIX™
How Supplied(Reusable/SingleUse)Single Use, supplied sterileHandpiece - ReusableCartridge - Single Use,supplied sterileSingle Use, supplied sterile
Method ofSterilizationEthylene Oxide to SAL 10-6Handpiece - AutoclaveCartridge - Ethylene Oxide toSAL 10-6Ethylene Oxide to SAL 10-6

1.4 SUMMARY OF PERFORMANCE TEST RESULTS

The descriptive characteristics are well-defined and adequate to ensure equivalence of the RxSight Insertion Device with the predicate devices.

Non-clinical performance testing included simulated surgical manipulation and recovery of properties including both pre- and post-injection evaluation of the IOLs in accordance with ISO 11979-3:2012. Specifically, mechanical dimensions and sagitta were verified following lens delivery and compared to the measurements performed prior to the lens delivery per ISO 11979-3:2012. Additionally, optical properties and overall surface and bulk homogeneity were tested pre and post lens injection per ISO 11979-2:2014.

Conclusion

The RxSight Insertion Device (Model 63002) has the same intended use as the legally marketed predicate devices identified in this 510(k) premarket notification and other IOL injectors regulated under 21 CFR 886.4300. This device meets all product design requirements and applicable standards and embodies technological characteristics similar to the predicate devices. Non-clinical data establish that the device is safe and effective in the delivery of the RxSight Light Adjustable Lens and Light Adjustable Lens+ and is shown to be substantially equivalent to the predicate devices.

§ 886.4300 Intraocular lens guide.

(a)
Identification. An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.(b)
Classification. Class I (general controls). Except when used as folders or injectors for soft or foldable intraocular lenses, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9.