K Number
K231832
Device Name
Access Myoglobin
Date Cleared
2023-09-18

(88 days)

Product Code
Regulation Number
866.5680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Access Myoglobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of myoglobin levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis of heart or renal disease.

Device Description

The Access Myoglobin assay is a sandwich immunoenzymatic assay. The Access Myoglobin assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access Myoglobin assay reagent pack, Access Myoglobin assay calibrators, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

AI/ML Overview

The provided text describes the Beckman Coulter Access Myoglobin assay, which is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of myoglobin levels in human serum and plasma. The K231832 submission seeks to demonstrate substantial equivalence to the predicate device (Access Myoglobin assay, K021229) when run on the Dxl 9000 Access Immunoassay Analyzer.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a separate section with pass/fail thresholds for all performance metrics. However, it presents the results of various performance studies against industry guidelines (CLSI standards), which implicitly serve as the acceptance criteria for demonstrating appropriate performance.

Performance CharacteristicAcceptance Criteria (Implicit from CLSI guidelines and typical assay requirements)Reported Device Performance (Access Myoglobin on Dxl 9000)
Method Comparison (vs. Predicate)Slope near 1.0, Intercept near 0, High correlation (R)Slope: 0.99, Intercept: 0.47, R: 1.00
Imprecision≤ 1.10 ng/mL SD at concentrations ≤ 11.0 ng/mL; ≤ 10.0% CV at concentrations > 11.0 ng/mLRepeatability (Within-Run): 2.0-2.3% CV, 0.18 SD (low)
Between-Run: 2.2-4.0% CV, 0.18-85.9 SD
Between-Day: 0.0-3.0% CV, 0.0-3.0 SD
Within-Laboratory: 3.7-5.4% CV, 0.31-101.9 SD
LinearityDemonstrate linearity across the measuring interval (e.g., polynomial fit not significantly better than linear)Demonstrated linearity across the measuring interval
Limit of Blank (LoB)Establish a LoB that is analytically sound3.0 ng/mL (μg/L)
Limit of Detection (LoD)Establish a LoD that is analytically sound3.0 ng/mL (μg/L)
Limit of Quantitation (LoQ)Establish a LoQ with acceptable imprecision (e.g., ≤ 20% within-lab CV)3.0 ng/mL (μg/L) at ≤ 20% within-lab CV

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison: N = 155 (This refers to 155 patient samples). The data provenance (country of origin, retrospective/prospective) is not specified in the provided text.
  • Imprecision: N = 80 per sample assessed (e.g., Sample 1, Sample 2, etc.), which involved multiple samples tested in duplicate in 2 runs per day for a minimum of 20 days.
  • Linearity: Sample size not explicitly stated, but typically involves a series of diluted/spiked samples.
  • LoB, LoD, LoQ: Not explicitly stated as a single "sample size" but involved multiple reagent lots and 3 instruments over a minimum of 3-5 days.

The data provenance for all studies (country of origin, retrospective or prospective) is not specified in the provided document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The Access Myoglobin assay is an in vitro diagnostic device for quantitative measurement of a biomarker. Its "ground truth" is established through analytical performance studies, not typically by expert review of individual cases as would be done for imaging or clinical decision support AI. Therefore, this section is not applicable in the traditional sense for this type of device. The "truth" is based on the accurate measurement of myoglobin concentration.

4. Adjudication Method for the Test Set

As this is an in vitro diagnostic quantitative assay, there is no adjudication method described or necessary in the context of expert consensus, as might be found for imaging AI. The performance is assessed against reference methods or calibrated controls.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size

No MRMC comparative effectiveness study was done. This type of study is relevant for imaging or clinical decision support AI where human readers interpret results, and the AI might augment their performance. For a quantitative immunoassay, the "reader" is effectively the instrument, and the performance is evaluated on its analytical accuracy and precision.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies presented (Method Comparison, Imprecision, Linearity, LoB/LoD/LoQ) demonstrate the standalone performance of the Access Myoglobin assay on the Dxl 9000 Access Immunoassay Analyzer. These are analytical performance studies of the device itself, without human interpretation as a primary variable.

7. The Type of Ground Truth Used

The ground truth for these analytical performance studies is based on:

  • Reference measurements: For method comparison, the predicate device (Access 2 Immunoassay System) served as the reference standard.
  • Established concentrations/materials: For imprecision, linearity, LoB/LoD/LoQ, the ground truth is based on known concentrations of myoglobin in controls, calibrators, and spiked samples, prepared according to industry standards.
  • Analytical principles: The assay measures myoglobin concentration, and the "ground truth" is the actual quantity of myoglobin present in a sample, determined through scientifically validated methods.

8. The Sample Size for the Training Set

This information is not provided in the document. For an immunoassay like this, there isn't a "training set" in the machine learning sense. The device's operational parameters and assay design are developed through extensive research and development, but this is distinct from an AI training dataset.

9. How the Ground Truth for the Training Set Was Established

As there is no explicit "training set" in the AI sense for this immunoassay submission, this question is not directly applicable. The "ground truth" during the development and optimization of such an assay would involve internal analytical studies using characterized materials and reference methods to ensure the assay accurately measures myoglobin concentrations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 18, 2023

Beckman Coulter, Inc. Kate Oelberg Senior Staff Quality and Regulatory Affairs 1000 Lake Hazeltine Drive Chaska, Minnesota 55318

Re: K231832

Trade/Device Name: Access Myoglobin Regulation Number: 21 CFR 866.5680 Regulation Name: Myoglobin Immunological Test System Regulatory Class: Class II Product Code: DDR Dated: June 21, 2023 Received: June 22, 2023

Dear Kate Oelberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231832

Device Name Access Myoglobin

Indications for Use (Describe)

The Access Myoglobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of myoglobin levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis of heart or renal disease.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number K231832

Submitter Name and Address:

Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318

Primary Contact:

Kate Oelberg, Senior Staff Quality and Regulatory Affairs Email: kmoelberg@beckman.com Phone: (612) 431-7315

Alternate Contact:

Kuljeet Kaur, Requlatory Affairs Manager Email: kkaur@beckman.com Office Phone: (952) 465-1914

Trade Name: Access Myoglobin Common Name: Myoqlobin Classification Regulation: 21 CFR 866.5680 Classification Product Code: DDR

Predicate Device:

Access Myoglobin 510(k) Number K021229

Device Description

The Access Myoglobin assay is a sandwich immunoenzymatic assay. The Access Myoglobin assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access Myoqlobin assay reagent pack, Access Myoqlobin assay calibrators, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

Intended Use

The Access Myodobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of myoglobin levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis of heart or renal disease.

{4}------------------------------------------------

Comparison of Technological Characteristics to the Predicate (Assay)

SystemAttribute/CharacteristicPredicate Access Myoglobinassay (K021229) run on theAccess 2 ImmunoassaySystemAccess Myoglobin assay runon the Dxl 9000 AccessImmunoassay AnalyzerInstrument
Intended Use/Indications for UseThe Access Myoglobin assayis a paramagnetic particle,chemiluminescentimmunoassay for thequantitativedetermination of myoglobinlevels in human serum andplasma using the AccessImmunoassay Systems.Same
Assay PrinciplesThe Access Myoglobin assayis a two-site immunoenzymatic("sandwich") assay. A sampleis added to a reaction vesselwith mouse monoclonal anti-myoglobin-alkalinephosphatase conjugate,mouse monoclonal anti-myoglobin-biotinconjugate, and paramagneticparticles coated with goat anti-biotin.Same
Solid SupportParamagnetic ParticlesSame
Detection SystemUtilizes dioxetane-basedchemiluminescent substrate;Measures light production froma chemiluminescent reactionSame
CalibratorsLiquid calibrators preparedfrom buffered bovine proteinmatrix and human skeletalMyoglobin at various levelsSame
Sample TypeSerum/Plasma (heparin orEDTA)Same
Measuring Range1 - 4000 ng/mL3.0 - 4000 ng/mL
Expected ResultsSeparate ranges for Heparinplasma/serum vs EDTAplasmaSame ranges
InstrumentAccess Immunoassay systemDxl 9000 AccessImmunoassay Analyzer
SubstrateAccess SubstrateLumi-Phos PRO substrate

{5}------------------------------------------------

Standard/Guidance Document Referenced (if applicable):

CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition

CLSI EP06-2nd Edition : Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline

CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline - Second Edition

CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples-Third Edition

CLSI EP28-A3c Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory - Third Edition

CLSI EP35 Assessment of Equivalence of Suitability of Specimen Types for Medical Laboratory Measurement Procedures - First Edition

Summary of Studies:

Method Comparison:_A study based on CLSI EP09c, 3rd Edition using Weighted Deming regression and Pearson's correlation compared the Access 2 Immunoassay System and the Dxl 9000 Access Immunoassay Analyzer.

NConcentration Range*(ng/mL)SlopeSlope95% CIInterceptIntercept95% CIR
1558.2 – 39000.990.98 – 1.000.47-0.10 – 1.01.00

*Range is Access 2 values

Imprecison: The assay was designed to have within-laboratory imprecision as listed below:

≤ 1.10 ng/mL (µg/L) SD at concentrations ≤ 11.0 ng/mL (µg/L)

≤ 10.0% CV at concentrations > 11.0 ng/mL (uq/L)

A study based on CLSI EP05-A3 performed on the Dxl 9000 Access Immunoassay Analyzer tested multiple samples in duplicate in 2 runs per day for a minimum of 20 days.

ng/mL (µg/L)Repeatability(Within-Run)Between-RunBetween-DayWithin-Laboratory
SampleNMeanSD%CVSD%CVSD%CVSD%CV
Sample 1807.90.182.30.182.20.172.20.313.9
Sample 2801012.22.14.04.03.03.05.55.4
Sample 3804658.01.715.13.20.00.017.03.7
Sample 480176335.02.057.53.30.10.067.33.8
Sample 580271954.82.085.93.20.00.0101.93.7

{6}------------------------------------------------

Linearity: A study based on CLSI EP06-Ed2 performed on the Dxl 9000 Access Immunoassay Analyzer determined the assay demonstrated linearity across the measuring interval.

Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ): Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) studies were conducted on the Dxl 9000 Access Immunoassay Analyzer following CLSI guideline EP17-A2 The LoB study included multiple reagent lots and 3 instruments over a minimum of 3 days. The LoD and LoQ studies included multiple reagent lots and 3 instruments over a minimum of 5 days.

ng/mL (μg/L)
Limit of Blank (LoB)3.0
Limit of Detection (LoD)3.0
Limit of Quantitation (LoQ)≤ 20% within-lab CV3.0

Substantial Equivalence Comparison Conclusion

Beckman Coulter's Access Myoglobin assay on the Dxl 9000 Access Immunoassay Analyzer is substantially equivalent to Myoglobin assay on the Access Immunoassay System (K021229) as demonstrated through the information and data provided in this submission. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.

§ 866.5680 Myoglobin immunological test system.

(a)
Identification. A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.(b)
Classification. Class II (performance standards).