K Number
K231792
Date Cleared
2023-09-28

(100 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

Device Description

Safety Sterile Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Safety Sterile Syringes have two types: Type F and type D. Type F is a syringe with a fixed needle. Type D is a syringe with a detachable needle. The detachable needle is secured by rotating it through the front needle hub into the back needle hub located on the syringe. Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the spring to minimize the risk of accidental sharps injury. Product is delivered sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in professional healthcare facility/hospital and transport environment.

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA for a medical device called "VeriSafe Safety sterile syringes." This type of document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than detailing a specific clinical study with acceptance criteria in the same way one might for a novel diagnostic or therapeutic AI device.

Therefore, the requested information elements related to AI/algorithm performance, ground truth establishment for AI, sample sizes for test/training sets in an AI context, expert qualifications, and MRMC studies are not directly applicable to this submission. This document describes bench testing verifying physical and material properties of a medical device (a syringe).

However, I can extract the relevant "acceptance criteria" (standards and tests) and how the device "meets" them based on the provided summary.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard/Test)Reported Device Performance
ISO 7886-1:2017Complied with requirements for:
- Extraneous matter- Performance meets required limits
- Limits for acidity or alkalinity- Performance meets required limits
- Limits for extractable metals- Performance meets required limits
- Lubrication- Performance meets required limits
- Tolerance on graduated capacity- Performance meets required limits
- Graduated Scale- Performance meets required limits
- Numbering of Scales- Performance meets required limits
- Overall length of scale to Nominal capacity line- Performance meets required limits
- Position of scale- Performance meets required limits
- Barrel- Performance meets required limits
- Plunger stopper/plunger assembly- Performance meets required limits
- Nozzle- Performance meets required limits
- Dead space- Performance meets required limits
- Freedom from air and liquid leakage past plunger stopper- Performance meets required limits
- Force to operate the piston- Performance meets required limits
- Fit of the plunger stopper/plunger in barrel- Performance meets required limits
ISO 9626:2016Complied with requirements for:
- Stiffness- Performance meets required limits
- Resistance to breakage- Performance meets required limits
- Resistance to corrosion- Performance meets required limits
ISO 7864:2016Complied with requirements for:
- Cleanliness- Performance meets required limits
- Colour coding- Performance meets required limits
- Needle tube-Outside diameter- Performance meets required limits
- Tolerance on length- Performance meets required limits
- Freedom from defects- Performance meets required limits
- Needle points- Performance meets required limits
- Fragmentation test- Performance meets required limits
- Penetration force and drag force- Performance meets required limits
- Bond between hub and needle tube- Performance meets required limits
- Patency of lumen- Performance meets required limits
ISO 23908:2011Complied with requirements for:
- Safety mechanism activation- Performance meets required limits (mechanism activates, needle retracts into barrel)
- Safety overriding and unlocking force after activation- Performance meets required limits
- Simulated clinical Use- Performance meets required limits (simulated use demonstrated effectiveness of safety features)
USP<85> Bacterial Endotoxin Test- Complied with test requirements (Pyrogen: Non pyrogenic)
Biocompatibility (ISO 10993 Series)Complied with ISO 10993 series standards, specifically for:
- ISO 10993-5 (Cytotoxicity)- No cytotoxicity
- ISO 10993-10 (Skin Sensitization)- No evidence of sensitization
- ISO 10993-23 (Irritation)- No evidence of skin irritation
- ISO 10993-11 (Acute Systemic Toxicity)- No systemic toxicity
- ASTM F756-17 (Hemolysis)- No evidence of hemolysis
- ISO 10993-7 (Ethylene oxide sterilization residuals)- Complied with limits for residuals
- USP42-NF37<151> (Pyrogen Test)- Non pyrogenic
- USP<788> (Particulate matter)- Complied with limits for particulate matter
Sterilization (ISO 11135:2014)- EO Sterilization, SAL = 10^-6 (validated)
Package Integrity (ASTM D4169-16, ASTM F88/F88-15, ASTM F1929-15, USP38-NF33_C71)- Packaging deemed acceptable for protection of product and sterility maintenance after environmental conditioning and simulated transportation.- Seal strength, dye penetration, and sterility tests performed and met criteria.
Shelf-Life (ASTM F1980-16)- 3 years (validated using accelerated aging)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: Not specified in terms of specific number of units tested for each standard. The document states "The bench testing performed verifies that the performance of the subject device is substantially equivalent..." implying tests were conducted on representative samples of the device as required by the listed ISO and ASTM standards.
  • Data Provenance: The device manufacturer is Promisemed Hangzhou Meditech Co., Ltd. in Hangzhou, Zhejiang, China. The tests cited are international standards (ISO, ASTM, USP), indicating the testing was conducted to these globally recognized specifications. The document doesn't explicitly state if the testing was performed retrospectively or prospectively, but standard device testing for 510(k) submissions is typically conducted during product development and validation, and is inherently "prospective" in relation to the submission itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is Not Applicable (N/A). The "ground truth" here is defined by performance against established international consensus standards (e.g., ISO, ASTM, USP) for physical, chemical, and biological properties of a medical device (syringe), not by expert clinical judgment or interpretation of data. The standards themselves are developed by committees of experts, but this is not what is typically asked for in terms of "ground truth" for a device's performance in a clinical or AI context.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is Not Applicable (N/A). Performance is measured objectively against the criteria defined by the listed standards, not through adjudication by multiple human reviewers.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is Not Applicable (N/A). This device is a sterile safety syringe and does not involve AI or human "readers" in the context of its function or assessment for this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is Not Applicable (N/A). This device is a sterile safety syringe and does not involve an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is the objective criteria and acceptable limits defined within the cited international consensus standards (ISO 7886-1, ISO 9626, ISO 7864, ISO 23908, ISO 10993 series, ISO 11135, ASTM F1980-16, ASTM D4169-16, ASTM F88/F88-15, ASTM F1929-15, USP<85>, USP42-NF37<151>, USP<788>, USP38-NF33_C71). These standards specify the test methods and the performance thresholds that the device must meet to be considered compliant and safe/effective.

8. The sample size for the training set

This section is Not Applicable (N/A). There is no AI model involved, therefore no training set.

9. How the ground truth for the training set was established

This section is Not Applicable (N/A). There is no AI model involved, therefore no training set or ground truth establishment for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 28, 2023

Promisemed Hangzhou Meditech Co., Ltd. % Wei Hsu Regulatory Manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong China

Re: K231792

Trade/Device Name: VeriSafe Safety sterile syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG Dated: August 29, 2023 Received: August 30, 2023

Dear Wei Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Juliane C. Lessard -S

Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231792

Device Name VeriSafe Safety Sterile Syringes

Indications for Use (Describe)

It is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K231792-510(K)Summary

Date Prepared 1

September 18, 2023

2 Submitter's Information

Name of Sponsor:

Promisemed Hangzhou Meditech Co., Ltd.

Address:

No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou

City, 311121 Zhejiang, China

Contact Name:

Zearou Yang

Telephone No.:

+86 571 88772985

Fax No.:

+86 571 88772985

Email Address:

zearou.yang@promisemed.ca

ന Trade Name, Common Name, Classification

Trade/Product Name: VeriSafe Safety sterile syringes Common Name: Piston Syringe Classification name: Syringe, Piston; Syringe, Antistick Regulation Number: 21 CFR 880.5860 Device Class: Class II Product Code: FMF, MEG

  • ব ldentification of Predicate Device K213560 Verifine Safety Syringe with Fixed Needle

{4}------------------------------------------------

Description of the Device റ

Safety Sterile Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Safety Sterile Syringes have two types: Type F and type D. Type F is a syringe with a fixed needle. Type D is a syringe with a detachable needle. The detachable needle is secured by rotating it through the front needle hub into the back needle hub located on the syringe.

Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the spring to minimize the risk of accidental sharps injury.

Product is delivered sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in professional healthcare facility/hospital and transport environment.

Coding rule:
MWS-032525
TypeCapacityGaugeNominal length of needle tubeWall
Type D1mL,2mL,18G, 20G, 22G,23G, 25G,27G,28G, 29G,30G, 31G, 32G6mm,8mm,12mm,25mm,38mmRegular wall (RW)
Type F3mL,5mL,10mLThin wall (TW)

6 Indication

lt is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury.

{5}------------------------------------------------

7 Similarities and Differences of the Proposed Devices to the Predicate Devices

The VeriSafe Safety sterile syringe is substantially equivalent to the predicate device, Verifine Safety Syringe with Fixed Needle (K213560) in that these devices have same intended use and technological characteristics. Using plunger to fill syringe as well as discharge the fluid is the technological principle of both the subject device and predicate device. The needle of both devices is made of stainless steel (X5CrNi18-10). The performance of both devices fulfills the requirement of ISO 7886-1, ISO 9626, ISO 7864 and ISO 23908. Both the subject and predicate devices are disposable, EO sterilized, single patient use devices. Both the subject and predicate device are used in Professional healthcare facility and transport environment.

Subject DevicePredicate Device (K213560)
Trade NameVeriSafe Safety Sterile SyringesVerifine Safety Syringewith Fixed Needle
ManufacturerPromisemed HangzhouMeditech Co., LtdPromisemed HangzhouMeditech Co., Ltd
Device ClassClass IIClass IISame
Product CodeFMF, MEGFMF, MEGSame
Regulation number880.5860880.5860Same
Regulation NamePiston syringePiston syringeSame
Intended Use/Indications for UseIt is intended to inject fluidsinto the body or withdrawblood for medical purposes.After injection, the needletube will be retracted into thebarrel when the safetyfeatures are manuallyactivated to minimize the riskof accidental sharps injury.It is intended to injectfluids into or withdrawfluids from the body formedical purposes.Same
OperatingPrinciplePlunger is used to fill syringeas well as discharge the fluid.After use, the needle tube willbe retracted into the barrel bypressing the push-button todepress the plunger all of theway within the barrel.Plunger is used to fillsyringe as well as dischargethe fluid.The protective shield willpermanently be locked inplace by pulling forwardtill click, providingprotection against needlesticks.Different-See Comment #1
Safety mechanismThe safety mechanism isactivated by pressing the push-button down to depress theplunger all of the way withinThe safety mechanism isactivated by pushing thesliding sleeves. WhenDifferentSee Comment#2
the barrel. When activated,needle is retracted into thebarrel.activate, needle is coveredby sliding sleeves.
Volume1mL,2mL,3mL,5mL,10mL0.5ml,1.0mlDifferent
See Comment #3
Tip typeFixed Needle and detachableneedleFixed NeedleDifferent
See Comment #4
Needle andsyringe tipconfigurationType FImage: Type F syringeType DImage: Type D syringeImage: SyringeDifferent
See Comment #5
Gauge18G, 20G, 22G, 23G, 25G,27G,28G, 29G, 30G, 31G, 32G21G, 22G, 23G, 24G, 25G,26G, 27G, 28GDifferent
See Comment #6
Needle Length6mm,8mm,12mm,25mm,38mm12mm, 16mm, 25 mm,32mm, 38mmDifferent
See Comment #7
GradationslegibilityLegibleLegibleSame
LubricantSilicone oilSilicone oilSame
BarreltransparencyTransparentTransparentSame
Reuse durabilitySingle UseSingle UseSame
BiocompatibilityComplied with ISO10993 seriesstandards, and the followingtests are performedComplied with ISO10993series standards, and thefollowing tests areperformedSame
- Cytotoxicity: No cytotoxicity
- Skin Irritation: No evidenceof skin irritation- Cytotoxicity: Nocytotoxicity
- Skin Irritation: Noevidence of skin irritation
- Skin Sensitization: Noevidence of sensitization- Skin Sensitization: Noevidence of sensitization
-Acute Systemic Toxicity: Nosystemic toxicity-Acute Systemic Toxicity:No systemic toxicity
-Hemolysis: No evidence ofhemolysis-Hemolysis: No evidence ofhemolysis
-Pyrogen: Non pyrogenic-Pyrogen: Non pyrogenic
-USP <788> Particulate matter-USP <788> Particulatematter
Configuration andMaterialsNeedle, spring: Stainless Steel(X5CrNi18-10)Needle: Stainless Steel(X5CrNi18-10)Similar
Barrel/push-button:PolypropyleneBarrel/push-button:PolypropyleneSee Comment# 8
Plunger: PolypropylenePlunger: Polypropylene
Needle cap: PolypropyleneNeedle cap: Polyethylene
Piston: Nitrile butadienerubber/ThermoplasticelastomerPiston: Polyisoprene rubber
Sliding Sleeve, Cover:Polypropylene
Retraction cylinder sealingring: Silicone rubberPlunger stopper, Tight ring:Acrylonitrile ButadieneStyrene
Needle hub: Polypropylene
Needle hub sealing ring:Thermoplastic elastomer
Retraction cylinder:Acrylonitrile Butadiene Styrene
PerformanceComplied with ISO 7886-1, ISO9626, ISO 7864, ISO 23908Complied with ISO 7886-1,ISO 9626, ISO 7864, ISO23908Same
Sterilizationmethod and SALEO SterilizationSAL = 10 $-6$EO SterilizationSAL = 10 $-6$Same
Environment ofuseProfessional healthcare facilityand transport environment.Professional healthcarefacility and transportenvironment.Same
shelf life3 years5 yearsDifferent $9$See Comment# 9

A detailed comparison to the predicate is provided in Table 1.

{6}------------------------------------------------

{7}------------------------------------------------

Discussions of differences in technological characteristics

Comment #1 and #2 : The subject device has different safety mechanisms to protect against sharps injury. The sharps protection feature of both devices are addressed through testing per ISO 23908. The difference in sharps protection feature does not affect the effectiveness and safety of the device.

{8}------------------------------------------------

Comment #3 : The subject device has a different capacity range. The specification of capacity of both devices fulfils the requirement of ISO 7886-1. The difference in capacity does not affect the effectiveness and safety of the device.

Comment # 4 : The subject device has an additional type of detachable needle tip. The performance of detachable needle type fulfils the requirement of ISO 9626:2016, ISO 7864:2016 and ISO 7886-1. The difference in tip type does not affect the effectiveness and safety of the device.

Comment # 5: Needle and syringe tip configuration of the subject device is different from that of predicate device. The performance of the subject device fulfill the requirements of ISO 9626:2016, ISO 7886-1:2016, ISO 7864:2016, ISO 23908:2011. The difference in needle and syringe tip configuration does not affect the effectiveness and safety of the device.

Comment #6 and #7 : The gauge and needle length of the subject device is different from the predicate device. The specification of gauge and needle length of both devices fulfil the requirement of ISO 9626:2016 and ISO 7864:2016. The difference in gauge and needle length does not affect the effectiveness and safety of the device.

Comment #8: The syringe bodies of subject device and predicate device have material differences. The biocompatibility tests conducted demonstrated the subject device. The difference in materials does not affect the effectiveness and safety of the device.

Comment # 9 : The subject device has shorter shelf life. The performance of aged subject device has been validated in accordance with ASTM F1980-16, ISO 9626:2016, ISO 7886-1:2016, ISO 7864:2016, ISO 23908:2011. The difference in shelf life does not affect the effectiveness and safety of the device.

The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question

{9}------------------------------------------------

is raised regarding effectiveness and safety. The subject device is substantially equivalent to the identified predicate device.

8 Performance Testing Summary

The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:

  • ISO 7886-1:2017, Sterile hypodermic syringes for single use Part 1: Syringes for manual use
    • Extraneous matter, Limits for acidity or alkalinity, Limits for extractable metals, Lubrication, Tolerance on graduated capacity, Graduated Scale, Numbering of Scales, Overall length of scale to Nominal capacity line, Position of scale, Barrel, Plunger stopper/ plunger assembly, Nozzle, Dead space, Freedom from air and liquid leakage past plunger stopper, Force to operate the piston, Fit of the plunger stopper/plunger in barrel.
  • · ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
  • Stiffness, Resistance to breakage, Resistance to corrosion
  • · ISO 7864:2016, Sterile hypodermic needles for single use Requirements and test methods
  • . Cleanliness, Colour coding, Needle tube-Outside diameter, Tolerance on length,Freedom from defects, Needle points, Fragmentation test, Penetration force and drag force, Bond between hub and needle tube, Patency od lumen
  • ISO 23908:2011, Sharps injury protection Requirements and test methods Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
    • Safety mechanism activation. Safety overriding and unlocking force after activation, Simulated clinical Use
  • USP<85> Bacterial Endotoxin Test

{10}------------------------------------------------

Biocompatibility

In accordance with ISO 10993-1 the device is classified as Externalcommunicating device, Blood path, indirect, with limited contactduration (≤ 24h). The biosafety tests that required consideration were validated:

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5:Tests for in Vitro Cytotoxicity
  • ISO 10993-10:2021 Biological Evaluation of Medical Devices -- Part 10: Tests for Skin Sensitization
  • ISO 10993-23:2021 Biological Evaluation of Medical Devices -Part 23: Test for irritation
  • ISO 10993-11:2017- Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Acute systemic toxicity
  • ASTM F756-17- Standard practice for assesment of hemolytic properties of materials
  • ISO 10993-7:2008- Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
  • USP42-NF37<151> Pyrogen Test
  • USP<788> Particulate matter in injections

Sterility, Shipping, and Shelf-Life

  • ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices
  • Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169-16, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
  • · Sterile Barrier Packaging Testing performed on the proposed device:
    • Seal strength ASTM F88/F88-15
    • Dye penetration ASTM F1929-15
    • Sterility test USP38-NF33_C71
  • Shelf life of 3 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
  • の Conclusion

Based on the information provided within this 510(k) submission, the differences between the predicate and the subject device do not raise any new or different questions of safety

{11}------------------------------------------------

or effectiveness. The proposed subject device is substantially equivalent to the predicate device and is as safe and as effective as the legally marketed predicate device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).