K Number
K231764
Manufacturer
Date Cleared
2023-10-23

(129 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.

The clinical applications and exam types include: Fetal (including Obstetrics), Abdominal, Pediatric, Small Organ (also known as Small Parts), Adult Cephalic is also known as Adult trans-cranial), Neonatal Cephalic, Intra-operative, Intra-operative Neuro (also known as Neurosurgery). Trans-raginal. Musculo-skeletal (Conventional and Superficial), Cardiac Adult, and Peripheral Vessel (also known as Peripheral Vascular).

Modes of Operation:

  • · 2D (B-Mode) (including Tissue Harmonic imaging)
  • · M-Mode
  • · Vector Flow Imaging (VFI)
  • Strain Elastography
  • · CW Doppler
  • · Contrast Imaging
  • · PWD Mode
  • · CFM Mode (Includes Color Doppler and Amplitude (Power) Doppler)
Device Description

The Ultrasound System 1300 is a multi-purpose mobile, software-controlled diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms.

The system consists of a mobile console (engine) that provides digital acquisition, processing and display capabilities. The user interface includes a glass touchpad, a 19" Clinical display monitor (CDM). In addition to this a variety of system accessories are available such as baskets, foot switch, printer start-up kit, and extra holders.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Prostate Volume Assist (PVA) AI feature, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated Goal)Reported Device Performance
Segmentation algorithm performance (standard metric of similarity for testing)"The test passed." (No specific metric value provided)
Initial caliper placement for 9018 transducer (compared to manual placement)Expected deviation: 11% +/- 6%
Initial caliper placement for 9048 transducer (compared to manual placement)Expected deviation: 7% +/- 15%
Workflow improvement (overall goal of PVA)Verified by visual evaluations by clinical experts validation testing, comparing initial algorithm placement to manual placement. The purpose was to compare PVA volume calculation with clinical personnel's volume measurement.

2. Sample Size and Data Provenance

  • Test Set (Clinical Validation):
    • Healthy Images: 975
    • Diseased Images: 1461 (includes images from the Holland dataset specifically used for testing)
    • Total Test Images: 2436
  • Data Provenance: Czech Republic, Holland.
  • Retrospective/Prospective: The text does not explicitly state retrospective or prospective, but as it's referred to as "data acquired in Amsterdam, Holland" and "data from curved and linear arrays," it implies retrospective use of existing data for a test set. The use of a "Holland dataset that are only used as Test images" further supports a retrospective approach.

3. Number of Experts and Qualifications for Ground Truth Establishment

  • Number of Experts: Not explicitly stated as a single number, but referred to as "clinical personnel" who had "an average of about 15 years' experience in ultrasound." This suggests multiple experts.
  • Qualifications: Clinical personnel with "an average of about 15 years' experience in ultrasound." (Specific roles like radiologist or sonographer are not mentioned, but the context implies experienced ultrasound practitioners).

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly stated as a formal adjudication process like 2+1 or 3+1. However, the "human caliper setters scored the images individually," and the comparison was made between the "automatic initial placement" and "manual placement by clinical personnel." This indicates a direct comparison to individual expert measurements, rather than a consensus-driven ground truth process for each image.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? A formal MRMC study, as typically understood for comparative effectiveness of AI vs. human readers, was not explicitly detailed. The study focused on comparing the AI's initial caliper placement and volume calculation against manual placement by clinical personnel, framed as a "workflow improvement."
  • Effect Size of Human Reader Improvement: Not applicable, as the study design was not an MRMC to measure human reader improvement with AI assistance. The AI's performance (initial caliper placement deviation) was measured against human manual placement.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? Yes, the "segmentation part of the algorithm is tested using a standard metric of similarity," which represents a standalone performance test. The numerical result for this specific metric is not provided beyond "The test passed."
  • The comparison of "automatic initial placement" to "manual placement by clinical personnel" also evaluates the standalone performance of the AI's caliper placement, even though it's compared against human input.

7. Type of Ground Truth Used

  • Ground Truth Type: Expert consensus for the training data (implied, see point 9) and individual expert manual measurements for the test data ("The validator used 2 distance measurements (using 4 calipers) on the transverse image and one distance measurement (using 2 calipers) on the sagittal image."). The purpose was to compare the prostate volume calculation of PVA with volume measurement performed by clinical personnel. So the ground truth for volume measurement is based on clinical personnel's manual calculations.

8. Sample Size for the Training Set

  • Healthy Images: 505
  • Diseased Images: 13447
  • Synthesized Data: 4104
  • Total Training Images: 18056

9. How the Ground Truth for the Training Set was Established

  • The document implies that the ground truth for the training set was established through expert annotation/labeling, as it refers to the test volumes being "completely distinct from the ones used during training process" and mentions that the algorithm uses both "segmentation and caliper placement." The process of generating synthesized data also suggests a controlled method for creating labeled examples, but the specifics of human involvement in training data ground truth are not detailed beyond inferring that it must have been expert-derived to train an algorithm providing "initial caliper placement."

{0}------------------------------------------------

October 23, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

BK Medical ApS c/o Bryan Behn Regulatory Affairs Director Mileparken 34 Herlev. 2730 DENMARK

Re: K231764

Trade/Device Name: Ultrasound System 1300 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX, QIH Dated: September 22, 2023 Received: September 25, 2023

Dear Bryan Behn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231764

Device Name Ultrasound System 1300

Indications for Use (Describe)

The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.

The clinical applications and exam types include: Fetal (including Obstetrics), Abdominal, Pediatric, Small Organ (also known as Small Parts), Adult Cephalic is also known as Adult trans-cranial), Neonatal Cephalic, Intra-operative, Intra-operative Neuro (also known as Neurosurgery). Trans-raginal. Musculo-skeletal (Conventional and Superficial), Cardiac Adult, and Peripheral Vessel (also known as Peripheral Vascular).

Modes of Operation:

  • · 2D (B-Mode) (including Tissue Harmonic imaging)
  • · M-Mode
  • · Vector Flow Imaging (VFI)
  • Strain Elastography
  • · CW Doppler
  • · Contrast Imaging
  • · PWD Mode
  • · CFM Mode (Includes Color Doppler and Amplitude (Power) Doppler)

Environment:

The Ultrasound System 1300 is intended for use in the professional healthcare environment (e.g. hospitals, physician offices)

Contraindications:

The Ultrasound System 1300 is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye.

The Cardiac Adult application is not intended for direct use on the heart.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BK Medical, a GE Healthcare company. The logo features the letters "bk" in a bold, green font, stacked above the word "medical" in a smaller, green font. Below "medical" is the text "a GE Healthcare company" in a smaller, lighter font. To the right of the "bk" is a stylized graphic of three curved lines, also in green, that resemble sound waves or a signal.

K231764

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

I. Submitter:BK Medical ApSMileparken 34Herlev 2730Denmark
Manufacturer:BK Medical ApSMileparken 34Herlev 2730Denmark
Primary Contact Person:Inesa CernajuteSenior Regulatory Affairs SpecialistBK MedicalTel: +45 42277733E-mail: inesa.cernajute@ge.com
Date Prepared:October 18, 2023

儿 Device Names / Common Names / Classification Names:

Trade Names:Ultrasound System 1300
Common Name:Ultrasound System
Classification Name:Ultrasonic pulsed doppler imaging system
Product Code:IYN (primary), IYO, ITX, QIH (secondary)
Class:II
Regulation Number:21 CFR §892.1550, §892.1560, §892.1570
Classification Panel:Radiology

{4}------------------------------------------------

III. Identification of Predicate or Legally Marketed Devices:

  • Primary predicate device: Ultrasound System 1300 as cleared under 510(k) premarket notification No K173569.
Trade Name:Ultrasound System 1300
Common Name:Ultrasound System
Classification Name:Ultrasonic pulsed doppler imaging system
Product Code:IYN (primary), IYO, ITX (secondary)
Class:II
Regulation Number:21 CFR §892.1550, §892.1560, §892.1570
Classification Panel:Radiology
  • . Reference predicate device: Ultrasound System 2300 as cleared under 510(k) premarket notification No K223830.
Trade Name:Ultrasound System 2300
Common Name:Ultrasound System
Classification Name:Ultrasonic pulsed doppler imaging system
Product Code:IYN (primary), IYO, ITX (secondary)
Class:II
Regulation Number:21 CFR §892.1550, §892.1560, §892.1570
Classification Panel:Radiology

The Ultrasound system 1300 is marketed under the marketing name bkSpecto

{5}------------------------------------------------

IV. Device Description

The Ultrasound System 1300 is a multi-purpose mobile, software-controlled diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms.

The system consists of a mobile console (engine) that provides digital acquisition, processing and display capabilities. The user interface includes a glass touchpad, a 19" Clinical display monitor (CDM). In addition to this a variety of system accessories are available such as baskets, foot switch, printer start-up kit, and extra holders.

Catalog/Reference(REF)ModelModel Description
13001300-21Image: Ultrasound machineModel with rechargeable battery
13001300-25Model without rechargeable battery

Table 1: Ultrasound System 1300

The various configurations of the Ultrasound System 1300 are intended to be used with various multi-frequency transducers (see Table 2). The indicated uses are different and specific for each transducer listed.

{6}------------------------------------------------

Name Description#Reference
ENDOCAVITY TRANSDUCER E10C49019
BIPLANE ENDOCAVITY TRANSDUCER E11C3b9008
TRIPLANE ENDOCAVITY TRANSDUCER E14C4t9018
ENDFIRE ENDOCAVITY TRANSDUCER E13C29029
BIPLANE ENDOCAVITY TRANSDUCER E14CL4b9048
CURVED ARRAY TRANSDUCER 9C29002
SMALL CURVED ARRAY TRANSDUCER 6C2s9023
CURVED ARRAY TRANSDUCER 6C29040
CURVED ARRAY TRANSDUCER 5C1e9085
PHASED ARRAY TRANSDUCER 5P1e9087
HOCKEY STICK TRANSDUCER X18L5s9009
WIDE LINEAR ARRAY TRANSDUCER 13L4w9011
LINEAR ARRAY TRANSDUCER 8L29032
LINEAR ARRAY TRANSDUCER 14L39051
HIGH-FREQUENCY LINEAR ARRAYTRANSDUCER18L59070
HIGH-FREQUENCY LINEAR ARRAYTRANSDUCER18L5s9081
LINEAR ARRAY TRANSDUCER 14L3e9086
3D ENDOCAVITY TRANSDUCER X14L49038
ANORECTAL TRANSDUCER 20R39052
I-SHAPED INTRAOPERATIVE TRANSDUCER I14C5I9015
T-SHAPED INTRAOPERATIVE TRANSDUCER I14C5T9016
BIPLANE INTRAOPERATIVE TRANSDUCER I12C5b9024
ROBOTIC DROP-IN TRANSDUCER X12C49026
CURVED ARRAY CRANIOTOMY TRANSDUCER N13C59062
BURR HOLE TRANSDUCER N11C5s9063

Table 2: Transducers used with Ultrasound System 1300

V. Indications / Intended Use:

Indications for Use:

The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.

{7}------------------------------------------------

The clinical applications and exam types include: Fetal (including Obstetrics), Abdominal, Pediatric, Small Organ (also known as Small Parts), Adult Cephalic (cephalic is also known as Adult trans-cranial), Neonatal Cephalic, Intra-operative, Intra-operative Neuro (also known as Neurosurgery), Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial), Cardiac Adult, and Peripheral Vessel (also known as Peripheral Vascular).

Modes of Operation:

  • · 2D (B-Mode) (including Tissue Harmonic imaging)
  • M-Mode
  • · Vector Flow Imaging (VFI)
  • Strain Elastography
  • · CW Doppler
  • Contrast Imaging
  • PWD Mode
  • · CFM Mode (Includes Color Doppler and Amplitude (Power) Doppler)

Environment:

The Ultrasound System 1300 is intended for use in the professional healthcare environment (e.g. hospitals, physician offices).

Contraindications:

The Ultrasound System 1300 is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye.

The Cardiac Adult application is not intended for direct use on the heart.

{8}------------------------------------------------

Mi Comparison of Technological Characteristics with the Predicate Device

Table 3: Substantial Equivalence Table of the proposed device with its predicate devices

Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
ManufacturerBK Medical ApSBK Medical ApSSame
Common NameUltrasound SystemUltrasound SystemSame
Name(Configurationmodels)bkSpecto (1300-21,1300-25)bkSpecto (1300-21)Added:- bkSpecto 1300-25- amodel without battery.- Prostate Volume Assist(PVA), AI softwarefeature.
Mode of OperationB, M, PW, CFM, P, THI,SE, CWCombination modes:2D+M, 2D+PW,2D+C+PW, 2D+P+PW,2D+2D, 2D+2D (BiplaneImaging), 2D+(2D+C),2D+(2D+P), 2D+THI,2D+SE· 2D (B-Mode)(including TissueHarmonic imaging)• M-Mode· Vector Flow Imaging(VFI)B, M, PW, CFM, P, THI,SE, CWCombination modes:2D+M, 2D+PW,2D+C+PW, 2D+P+PW,2D+2D, 2D+2D (BiplaneImaging), 2D+(2D+C),2D+(2D+P), 2D+THI,2D+SE· 2D (B-Mode)(including TissueHarmonic imaging)• M-Mode• Vector Flow Imaging(VFI)Same - primary predicateEquivalent - ref. predicate
Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
• Strain Elastography• CW Doppler• Contrast Imaging• PWD Mode• CFM Mode(Includes ColorDoppler andAmplitude (Power)Doppler)• Strain Elastography• CW Doppler• Contrast Imaging• PWD Mode• CFM Mode(Includes ColorDoppler andAmplitude (Power)Doppler)
Intended UseIntended Use:The system is adiagnostic ultrasoundimaging system used byqualified and trainedhealthcare professionalsfor ultrasound imaging,human body fluid flowanalysis and puncture andbiopsy guidance.Intended Use:The system is adiagnostic ultrasoundimaging system used byqualified and trainedhealthcare professionalsfor ultrasound imaging,human body fluid flowanalysis and punctureand biopsy guidance.Same
Environment:The Ultrasound System1300 is intended for usein the professionalhealthcare environment(e.g., hospitals, physicianoffices).Contraindications:The Ultrasound SystemEnvironment:The Ultrasound System1300 is intended for usein the professionalhealthcare environment(e.g., hospitals, physicianoffices).Contraindications:The Ultrasound System1300 is not intended for
Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
causing the acoustic beamto pass through the eye.The Cardiac Adultapplication is notintended for direct use onthe heart.causing the acoustic beamto pass through the eye.The Cardiac Adultapplication is notintended for direct use onthe heart.
Indications/ClinicalApplications• Fetal (includingObstetrics)• Abdominal• Pediatric1• Intraoperative1,2• Intraoperative Neuro(also known asNeurosurgery)1• Small Organ (SmallParts)3• Adult Cephalic(cephalic is also knownas Adult trans-cranial)1• Neonatal Cephalic1• Transrectal• Transvaginal• Musculoskeletal(Conventional)• Musculoskeletal(Superficial)• Cardiac Adult• Peripheral Vessel(Peripheral)• Fetal (includingObstetrics)• Abdominal• Pediatric1• Intraoperative1,2• Intraoperative Neuro(also known asNeurosurgery)1• Small Organ (SmallParts)3• Adult Cephalic(cephalic is also knownas Adult trans-cranial)1• Neonatal Cephalic1• Transrectal• Transvaginal• Musculoskeletal(Conventional)• Musculoskeletal(Superficial)• Cardiac Adult• Peripheral Vessel(Peripheral)Identical
Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
ApplicationEnvironmentProfessional healthcarefacility environmentProfessional healthcarefacility environmentSame
UsersQualified ProfessionalusersQualified ProfessionalusersSame
Patient PopulationAdult, PediatricAdult, PediatricSame
Transducer typesSurfaceEndocavityIntraoperativeSurfaceEndocavityIntraoperativeIdentical
System Transducers9002, 9008, 9009, 9011,9015, 9016, 9018, 9019,9023, 9024, 9026, 9029,9032, 9038, 9040, 9048,9051, 9052, 9062, 9063,9070, 9081, 9085, 9086,90879002, 9008, 9009, 9011,9015, 9016, 9018, 9019,9022, 9023, 9024, 9026,9029, 9032, 9038, 9040,9048, 9051, 9052, 9062,9063, 9067, 9070, 9081,9085, 9086, 9087Since 2017 the followingtransducers have beenterminated:9022 and 9067
BiocompatibilityThe Ultrasound Systemdoes not come in contactwith the patient.The Ultrasound Systemdoes not come in contactwith the patient.Same
HardwareClinical display monitor(CDM):• 19" Optical bondedglass front.• Can be tilted andmoved sideways.Cart:• adjustable height andwith 4 lockable wheelsKeyboard:Glass touch UIClinical display monitor(CDM):• 19" Optical bondedglass front.• Can be tilted andmoved sideways.Cart:• adjustable height andwith 4 lockable wheelsKeyboard:Glass touch UISame - primary predicateEquivalent - ref. predicate
Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
Scan engine:• 3 Transducer ports• 64 TX/RX channelsScan engine:• 3 Transducer ports64 TX/RX channels
OS SoftwareWindows 10Windows 8.1 IndustrialUpdated to Windows 8 OS to Windows 10.
Options- 3D Freehand- 3D Professional- DICOM Encrypted- Vector Flow Imaging (VFI)- Varian Interface- Strain Elastography- Needle Enhancement (X-shine)- Wi-Fi- bkViewer (SW running on a mac/windows pc) – not a medical device- Prostate Volume Assist (PVA)- 3D Freehand- 3D Professional- DICOM Encrypted- Vector Flow Imaging (VFI)- Varian Interface- Strain Elastography- Needle Enhancement (X-shine)- Wi-Fi- Proposed a new Prostate Volume Assist (PVA), AI software feature that has been recently cleared under K223830 for the Reference Predicate device Ultrasound System 2300. PVA provides a workflow improvement to an existing prostate volume measurement and calculation tool.It will require the same biplane/triplane transrectal transducers (9008, 9018, 9048) that are being used for existing prostate volume calculations and measurements.
Image featuresSpeckle reduction, compound imaging, tissue harmonic imaging (thi), trapezoid scanning (virtual convex) strain, elastography (se)Speckle reduction, compound imaging, tissue harmonic imaging (thi), trapezoid scanning (virtual convex) strain, elastography (se)Same
Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
UI Design-19-inch Clinical Monitorand touch input devicefor user interaction.-Touchpad track pad forcursor control (GlassTouch UI)-Full configurableinteraction controls(size/position)-19-inch Clinical Monitorand touch input devicefor user interaction.-Touchpad track pad forcursor control (GlassTouch UI)-Full configurableinteraction controls(size/position)Same – primary predicateEquivalent - ref. predicate
StandardsELECTROMAGNETICCOMPATIBILITYComplies withrequirements for Class Adevices of IEC 60601-1-2SAFETYANSI/ AAMI/ES 60601-1,IEC 60601-2-37SOFTWARE/FIRMWAREIEC62304CLEANINGVALIDATIONAAMI TIR12, AAMITIR30DICOMNEMA PS3.1-3.20Digital Imaging andCommunications inMedicine (DICOM)ELECTROMAGNETICCOMPATIBILITYComplies withrequirements for Class Adevices of IEC 60601-1-2SAFETYANSI/ AAMI/ES 60601-1,IEC 60601-2-37SOFTWARE/FIRMWAREIEC62304CLEANINGVALIDATIONAAMI TIR12, AAMITIR30DICOMNEMA PS3.1-3.20Digital Imaging andSame - primary predicateEquivalent - ref. predicate
Ultrasound System 1300Ultrasound System 1300
CharacteristicProposed device(K231764)Primary Predicate(K173569)Comment onComparison
Communications inMedicine (DICOM)

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

VII. Performance Data

Summary of non-clinical /Performance - Bench Testing

The proposed Ultrasound System 1300 and applied transducers, have been tested and conform to the following standards:

  • . ANSI/AAMI/ES 60601-: 2005/ (R) 2012 and A1:2012, C1:2009/ (R) 2012 and A2:2010/ (R) 2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2: Ed. 4.0, 2014 Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-2-37 Medical Electrical Equipment Part 2-37: Ed. 2.1, 2017 . Requirements for the Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
  • IEC 62359: Ed. 2.1, 2017 Ultrasonics Field Characterization Test Methods for ● the Determination of Thermal and Mechanical Indices Related to Medical Diagnostic Ultrasonic Fields
  • AAMI/ANSI/ISO 10993-1: 2018 Biological Evaluation of Medical Devices Part ● 1: Evaluation and Testing within a Risk management Process
  • AAMI TIR-12:2010 and AAMI TIR-30:2011 ●
  • IEC 62304: 2006/A1:2016 - Medical Device Software Life-Cycle Processes (Software / Informatics)
  • . NEMA PS3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM)
  • ISO14971: 2019 Application of risk management to medical devices .

The following quality assurance measures are applied to the development of the system:

  • · Risk Analysis
  • Requirements Reviews
  • · Design Reviews
  • · Testing on unit level (Module verification)
  • · Integration testing (System verification)
  • · Performance testing (Verification)
  • · Safety testing (Verification)

{16}------------------------------------------------

AI Feature Summary of Testing - Prostate Volume Assist (PVA)

Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.

Patient TypeTrainingImagesDeepLearningValidationimagesTest images(Clinicalvalidation)Comment
Healthy505190975
Diseased1344711891461Includes images fromHolland dataset that areonly used as Testimages.
Synthesized data4104480
  • · The table summarizes number of datasets used for each different purpose.
    • Gender: Male.

· Age: patients scheduled for prostates biopsies (54 - 78 yrs) and healthy patients 30 -60 yrs old; the specific age of each individual patient is not collected.

· Ethnicity/Country: Czech Republic, Holland. Note, European population is representative of the US population for the PVA - workflow improvement case.

· Mix of data from curved and linear arrays.

· For the testing process, the segmentation part of the algorithm is tested using a standard metric of similarity. The test passed. The final algorithm including both segmentation and caliper placement is verified by visual evaluations by clinical experts validation testing comparing the initial caliper placement by the algorithm compared to manual placement by clinical personnel. The purpose is to compare the prostate volume calculation of PVA with volume measurement performed by clinical personnel. The clinical personnel had an average of about 15 years' experience in ultrasound. We used a freeware tool developed by the Oxford Imaging Group for manual caliper placement on given images. For testing, we use two data sets that have not been part of the training of the algorithm. Representing the 9018, we clinical end-user test in the Czech Republic and representing the 9048 using data acquired by in Amsterdam, Holland. For the latter, we generated images that correspond the scanner preset. Prostate L, as the image depth is 6.5 cm, the recommended minimum depth for PVA. The human caliper setters scored the images individually. They received the following instructions; The calipers were set

{17}------------------------------------------------

in pairs corresponding to a distance measurement, so calipers 1 and 2 form one distance measurement and calipers 3 and 4 form another distance measurement.

The validator used 2 distance measurements (using 4 calipers) on the transverse image and one distance measurement (using 2 calipers) on the sagittal image. The automatic initial placement was tested and determined for the 9018 to deviate 11% +/-6% compared to manual placement and for the 9048, the initial placement of the caliper is expected to deviate 7% +/- 15% compared to manual placement.

· The volumes used for test/validation purpose are completely distinct from the ones used during training process and there is no overlap between the two.

Animal Testing

Not applicable - animal testing was not required to support substantial equivalence to the predicate device.

Clinical Studies

Not applicable - clinical studies were not required to support substantial equivalence to the predicate device.

VIII. Conclusion

BK Medical ApS considers the proposed device to be as safe, as effective and performance is substantially equivalent to the predicate device(s).

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.