(241 days)
The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in healthcare facilities and home monitoring) and inform on bed exit.
The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care.
The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
Neteera 130H-Plus device is a modification of the predicate device Neteera 130H (K212143). It is a contact-free vital-signs monitor based on a high frequency (122.25-123 GHz) micro-radar on-chip and algorithm, capable of detecting Respiration Rate (RR) and Heart Rate (HR) during rest or subject's mild body movement, with the additional capability of identifying and notifying on bed exit.
Neteera's micro- radar-based solution enables remote measurement, in real-time, in a non-invasive and contact-free manner. The system works by measuring only the ballistocardiograph micro-movements of of the skin (BCG) through nonmetallic materials such as furniture and clothing at a high resolution, and it has several different mounting options and measuring ranges.
This FDA 510(k) summary provides information on the Neteera 130H-Plus Vital Sign Monitoring Sensor, which is a modification of a previously cleared device. The summary primarily focuses on demonstrating substantial equivalence to the predicate device rather than detailing comprehensive acceptance criteria and a standalone study for the current device. However, it does provide some clinical validation details.
Here's an attempt to extract and interpret the information based on your requested format:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics for HR, RR, or bed exit detection. It states that "all outcomes met the same pre-specified performance criteria" from the predicate device's clearance, but these criteria are not detailed in this submission. To fully answer this, one would need to refer to the K212143 submission for the predicate device.
However, based on general expectations for vital sign monitors, the reported measurement ranges can be inferred as indirectly related to performance.
| Parameter | Acceptance Criteria (Inferred/Referenced) | Reported Device Performance (Implied) |
|---|---|---|
| Heart Rate (HR) | Within range of 40-160 Beats Per Minute with performance comparable to FDA-cleared ECG reference. | Met pre-specified performance criteria. |
| Respiration Rate (RR) | Within range of 5-40 Breaths Per Minute with performance comparable to FDA-cleared Capnograph reference. | Met pre-specified performance criteria. |
| Bed Exit Detection | New feature; performance criteria not specified in this summary. | Functioned as intended (implied by clearance). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size (Clinical Validation): 46 subjects.
- Data Provenance: The study was a "GCP-compliant confirmatory clinical validation study under Independent Ethical Committee approval." This implies a prospective study. The country of origin is not explicitly stated, but the sponsor, Neteera Technologies Ltd., is based in Jerusalem, Israel, which might suggest the study was conducted there.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The ground truth was established by "FDA-cleared devices ECG and Capnograph," which are considered reference standards, not human experts for interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the ground truth was established by FDA-cleared reference devices (ECG and Capnograph), not through human expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This device is a standalone vital signs monitor, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Standalone Performance: Yes, the clinical validation study was for the standalone performance of the Neteera 130H-Plus device, comparing its measurements directly against FDA-cleared reference devices (ECG and Capnograph). The device is described as "contact-free vital-signs monitor based on a high frequency (122.25-123 GHz) micro-radar on-chip and algorithm."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: Measurements from FDA-cleared reference devices: ECG (for Heart Rate) and Capnograph (for Respiration Rate).
8. The sample size for the training set
This information is not provided in the 510(k) summary. The summary focuses on the clinical validation of the modified device, not its initial algorithm development or training data.
9. How the ground truth for the training set was established
This information is not provided in the 510(k) summary.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
February 9, 2024
Neteera Technologies Ltd. Rakefet Shohat VP QA/RA/CA Prof Rakah 3 St. Jerusalem, Jerusalem 9139002 Israel
Re: K231733
Trade/Device Name: Neteera 130H-Plus Vital Sign Monitoring Sensor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, BZQ Dated: January 11, 2024 Received: January 11, 2024
Dear Rakefet Shohat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hetal B. Odobasic -S for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
510(k) Number (if known)
Device Name
Neteera 130H-Plus Indications for Use (Describe)
The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in healthcare facilities and home monitoring) and inform on bed exit.
The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care.
The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
Type of Use (Select one or both, as applicable)
□ Over-The-Counter Use (21 CFR 801 Subpart C) 2 Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{3}------------------------------------------------
510(k) Summary
Device Name: Neteera 130H-Plus Vital Signs Monitoring Sensor
1. 510K Submission Sponsor
Neteera Technologies Ltd. Prof. Rakah 3 St. PO Box 39088 Jerusalem 9139002, ISRAEL Company Phone No.: +972-2-5808733
Contacts:
- a. Janice M. Hogan, Consultant, Hogan Lovells US LLP Email: mjanice.hogan@hoganlovells.com
- b. Rakefet Shohat, VP QA/RA/CA Email: rakefet.shohat(@neteera.com
2. Date Prepared
9-Jan 2024
3. Device Identification
Trade/Proprietary Name: Neteera 130H-Plus Common/Usual Name: Vital Signs Monitoring Sensor Classification Name: Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) & Monitor, Breathing Frequency Regulation Number: 21 CFR 870.2300 (Product Code DRT) & 21 CFR 868.2375 (Product Code BZQ) Product Code: DRT & BZQ Device Class: 2 Classification Panel: Cardiovascular
-
- Legally Marketed Predicate Device(s): Neteera's Vital Sign Monitoring Sensor Model: 130H (K212143)
-
- Indication for Use Statement: The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in facilities and home monitoring) and inform healthcare on bed exit.
Neteera Technologies Ltd
Page 1 of 5
{4}------------------------------------------------
The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care.
The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
-
- Device Description: Neteera 130H-Plus device is a modification of the predicate device Neteera 130H (K212143). It is a contact-free vital-signs monitor based on a high frequency (122.25-123 GHz) micro-radar on-chip and algorithm, capable of detecting Respiration Rate (RR) and Heart Rate (HR) during rest or subject's mild body movement, with the additional capability of identifying and notifying on bed exit.
Neteera's micro- radar-based solution enables remote measurement, in real-time, in a non-invasive and contact-free manner. The system works by measuring only the ballistocardiograph micro-movements of of the skin (BCG) through nonmetallic materials such as furniture and clothing at a high resolution, and it has several different mounting options and measuring ranges.
- Device Description: Neteera 130H-Plus device is a modification of the predicate device Neteera 130H (K212143). It is a contact-free vital-signs monitor based on a high frequency (122.25-123 GHz) micro-radar on-chip and algorithm, capable of detecting Respiration Rate (RR) and Heart Rate (HR) during rest or subject's mild body movement, with the additional capability of identifying and notifying on bed exit.
7. Performance Testing (Non-Clinical):
As part of demonstrating Substantial Equivalence for safety (electrical, EMC, etc.) and effectiveness of the Neteera 130H-Plus to the predicate device, Neteera Technologies Ltd. repeated several non-clinical performance (SW) and electrical safety tests that were performed for the predicate device. Testing results confirmed that the design output meets the design inputs and specifications of the subject device.
Testing and/or compliance with the following standards has been achieved for the Neteera 130H-Plus Device:
- IEC 60601-1-2: 2014 + A1: 2020 Edition 4.1Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [tested as Class B, Group 2].
- EN 60601-1-2:2015 + A1: 2021 Edition 4.1 - Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-6:2010/AMD1:2013 for use in conjunction with IEC 62366:2007, ● AMD1:2014 and IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 Medical electrical equipment - Part 1-6: General requirements for safety -Collateral standard: Usability.
- EN 60601-1-6:2010/AMD1:2015 for use in conjunction with EN 60601-1:2006 + A11:2011 + A1:2013.
- EN 60601-1-11:2015+A1:2021 Medical electrical equipment. ●
- IEC 60601-1-11:2015+A1:2020 Medical electrical equipment Part 1-11 for use ● in conjunction with IEC 60601-1:2005, AMD1:2012: General requirements for basic safety and essential performance - Collateral Standard: Requirements for
Neteera Technologies Ltd
Page 2 of 5
{5}------------------------------------------------
medical electrical equipment and medical electrical systems used in the home healthcare environment.
- EN 60601-1:2006 +A11:2011 + A1: 2013 Medical electrical equipment General requirements for basic safety and essential performance.
- IEC 60601-1:2005 AMD1: 2012 Medical electrical equipment General ● requirements for basic safety and essential performance.
- IEC 62366-1 Edition 1.0 2015-02 Medical devices - Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
- IEC 62304:2006/A1:2015 Medical device software - Software life cycle processes.
- AIM 7351731 Rev. 3.00 (2021-06-04) Medical Electrical Equipment and ● System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
- IEEE/ANSI C63.27-2017– American National Standard for Evaluation of ● Wireless Coexistence.
- AAMI TIR69 2017 - Risk management of radio-frequency wireless coexistence for medical devices and systems.
- EN 305 550-1 V1.2.1 sections 7.1 thru 7.5, sec 8.1. Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Radio equipment to be used in the 40 GHz to 246 GHz frequency range; Part 1: Technical characteristics and test methods.
- FCC 47 CFR part 15 (Operation in the bands 116-123 GHz) section 15.258(b), transmitter power test; section 15.215(c), occupied bandwidth; section 15.258(c) (2), Out of band radiated emissions below 40 GHz; section 15.258(c) (3), Out of band radiated emissions below 40 GHz up to 370 GHz; section 15.258(d), Frequency stability test; section 15.203, Antenna requirement.
- EN/IEC 60601-1-8 General requirements for safety - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
In addition to the testing above, software validation was also conducted. This validation demonstrated that the Neteera 130H-Plus performed and functioned as intended and according to the design specifications and as such is equivalent to the predicate.
Neteera Technologies Ltd
Page 3 of 5
{6}------------------------------------------------
8. Clinical Testing
As part of the validation of the effectiveness of the device, Neteera Technologies Ltd. completed a GCP-compliant confirmatory clinical validation study under Independent Ethical Committee approval to demonstrate that minor changes to the device did not affect clinical performance. The study compared the subject device to FDA-cleared devices in the same way and using the same FDA-cleared comparators as done for the clearance of the predicate device. Specifically, the comparators consisted of FDAcleared devices ECG and Capnograph, EPM-10M S/N: AD1-03 for measuring Spot HR, Spot RR, Continuous HR and Continuous RR. Testing was performed on 46 subjects and all outcomes met the same pre-specified performance criteria.
9. Statement of Substantial Equivalence
The Neteera 130H-Plus device has the same intended use and same indications for use and technological characteristics as the predicate device, Neteera 130H, except it has the additional ability to detect and notify on Bed Exit of the monitored person, and minor hardware modifications were introduced.
The changes introduced for the 130H-Plus device and warranted this submission are hardware modification that required performance validation using human subjects and notification on Bed Exit which entails a change in the Intended Use (labeling).
The minor difference of having a feature that identifies and notifies on Bed Exit of the monitored subjects, and the hardware modifications do not raise different questions regarding the safety and effectiveness of the device. A Substantial Equivalence Table is presented below.
Therefore, based on the information provided in this 510(k) premarket notification, Neteera Technologies Ltd. concludes that the Neteera 130H-Plus device is substantially equivalent to the Neteera 130H predicate device.
| Subject Device | Predicate Device | |
|---|---|---|
| 510(K) Number | K231733 | K212143 |
| Device Name | Vital Sign Monitoring SensorNeteera 130H-Plus | Vital Sign Monitoring SensorNeteera 130H |
| Common name | Heart Rate and Respiratory RateMonitor | Heart Rate and Respiratory RateMonitor |
| Manufacturer | Neteera Technologies Ltd. | Neteera Technologies Ltd. |
| Intended Use | The Neteera System, Neteera130H-Plus device, is intendedfor spot and continuousmeasurement of heart rate andrespiration rate in adult patients(in healthcare facilities andhome monitoring) and informon bed exit.The indications provided are tobe used by health careprofessionals and are intended | The Neteera System, Neteera130/131H device, is intended forspot and continuousmeasurement of heart rate andrespiration rate in adult patients(in healthcare facilities andhome monitoring).The indications provided are tobe used by health careprofessionals and are intended to |
| to be reviewed by clinicians toinform patient care.The Neteera device is notintended to be used as an alarmsystem for potentially acutelife-threatening situations inwhich medical intervention isnecessary (e.g., ICU). | be reviewed by clinicians toinform patient care.The Neteera device is notintended to be used as an alarmsystem for potentially acute life-threatening situations in whichmedical intervention is necessary(e.g., ICU). | |
| Components | Sensor; USB Cable, PowerSupply | Sensor; USB Cable, PowerSupply |
| Technology | Contactless, Radar-basedmeasurement of micro-motions | Contactless, Radar-basedmeasurement of micro-motions |
| MedicalParametersMonitored andDisplayer | Heart Rate; Respiratory Rate,bed exit. | Heart Rate; Respiratory Rate |
| Patient Type | Adult | Adult |
| Use Environment | Healthcare facilities and homemonitoring.The indications provided are tobe used by health careprofessionals and are intended tobe reviewed by clinicians toinform patient care.Not intended for potentiallyacute life-threatening situations. | Healthcare facilities and homemonitoring.The indications provided are tobe used by health careprofessionals and are intended tobe reviewed by clinicians toinform patient care.Not intended for potentially acutelife-threatening situations. |
| HR MeasurementRange | 40-160 Beats Per Minute | 40-160 Beats per Minute |
| RR MeasurementRange | 5-40 Breaths Per Minute | 5-40 Breaths per Minute |
| EMCCompliance | IEC 60601-1-2 | IEC 60601-1-2 |
| ProductElectrical Safetyand coexistence | IEC 60601-1;IEC 60601-1-6;IEC 60601-1-11AIM 7351731 Rev. 3.00IEEE/ANSI C63.27-2017AAMI TIR69 2017EN 305 550-1 V1.2.1FCC 47 CFR part 15EN/IEC 60601-1-8 | IEC 60601-1;IEC 60601-1-6;IEC 60601-1-11AIM 7351731 Rev. 3.00IEEE/ANSI C63.27-2017AAMI TIR69 2017EN 305 550-1 V1.2.1FCC 47 CFR part 15EN/IEC 60601-1-8 |
| Data Display | Cloud Dashboard | Cloud Dashboard |
Table 5A - Substantial Equivalence Chart for the Neteera 130H-Plus
Neteera Technologies Ltd
Page 4 of 5
{7}------------------------------------------------
Neteera Technologies Ltd
Page 5 of 5
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).