(259 days)
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System consists of the CONTOUR® PLUS BLUE meter, the CONTOUR® PLUS blood glucose test strips and the Contour® Diabetes app.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The CONTOUR® PLUS blood glucose test strips are for use with the CONTOUR® PLUS BLUE meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips.
The system is intended for in vitro diagnostic use only.
CONTOUR® PLUS BLUE Blood Glucose Monitoring System is a blood glucose monitoring system with Bluetooth Low Energy technology built in so that the meter can communicate wirelessly to smart phones and tablets. The meter uses the CONTOUR® PLUS blood glucose test strips and CONTOUR® PLUS control solution. The meter can be connected to the Contour® Diabetes app. It utilizes a similar algorithm as the one used in the CONTOUR® NEXT GEN Blood Glucose Monitoring System. It uses two replaceable coin cell batteries. The meter's shape is a traditional oval form factor and it includes an on-screen arrow that points to the color indicating if a glucose result is above, within, or below target range.
This document describes the CONTOUR® PLUS BLUE Blood Glucose Monitoring System, a medical device for quantitative measurement of glucose in fresh capillary whole blood. The submission aims to demonstrate substantial equivalence to a predicate device (CONTOUR® NEXT GEN Blood Glucose Monitoring System (K193407)).
Here's an analysis of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text generally states that the device "met the system specifications" and "performed as intended." However, it does not provide specific numerical acceptance criteria or detailed performance data in a tabular format as requested. Therefore, the table below provides what can be inferred from the text, but lacks the specific quantitative details typically found in such a table.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| Bench Testing | The CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications. |
| Usability Testing | The CONTOUR® PLUS BLUE Blood Glucose Monitoring System was easy to use by typical customers. |
| Clinical Testing | The CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications. |
Important Note: The current document lacks the detailed numerical acceptance criteria (e.g., specific accuracy ranges, precision limits, interference limits) and the corresponding measured performance data. For a complete answer, this information would be required from a more detailed study report.
2. Sample Size and Data Provenance
The document states that "clinical testing showed that the CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications." However, it does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the study).
3. Number of Experts and Qualifications for Ground Truth
The document does not provide any information regarding the number of experts used to establish the ground truth for the test set or their qualifications. This information is typically relevant for studies involving human interpretation (e.g., interpretation of medical images), which is not directly applicable to a blood glucose monitoring system that provides a quantitative reading. The "ground truth" for a device measuring glucose would likely be a highly accurate laboratory reference method.
4. Adjudication Method for the Test Set
The document does not mention any adjudication method. This is usually relevant in studies where multiple human readers or experts are involved in assessing data, and their disagreements need to be resolved. For a blood glucose monitoring system, the "adjudication" would refer to the process by which the reference method's result is established, which is typically a standardized laboratory procedure rather than expert consensus on interpretation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study is not applicable to a blood glucose monitoring system. This type of study evaluates the effect of AI assistance on human reader performance, typically in diagnostic imaging. The CONTOUR® PLUS BLUE system directly measures glucose and provides a reading, without human interpretation in the same way an imaging study would require. Therefore, no such study was performed or needed.
6. Standalone Performance (Algorithm Only)
The document focuses on the overall system performance, which includes the meter, test strips, and app. It states that the system "met the system specifications." While the device uses "a similar algorithm as the one used in the CONTOUR® NEXT GEN Blood Glucose Monitoring System," the document does not provide separate standalone (algorithm-only) performance data. The performance described encompasses the integrated device.
7. Type of Ground Truth Used
While not explicitly stated, for a blood glucose monitoring system, the ground truth is typically established using a highly accurate laboratory reference method for glucose measurement (e.g., a YSI analyzer, or a certified clinical laboratory method). This is the gold standard against which the device's measurements are compared. The document states "clinical testing showed that the CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications," implying comparison to a reference.
8. Sample Size for the Training Set
The document does not provide any information about the sample size for the training set. The device utilizes "a similar algorithm as the one used in the CONTOUR® NEXT GEN Blood Glucose Monitoring System," which suggests that the core algorithm may have been trained previously, and this submission focuses on validating the integrated system.
9. How Ground Truth for Training Set Was Established
The document does not provide any information on how the ground truth for the training set was established. Given it states the algorithm is "similar" to a previously approved device, it's possible the training data and ground truth protocols for the original algorithm are foundational, but the specific details are not present in this submission summary. As with the test set, it would expectedly involve comparison to a laboratory reference method.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
February 23, 2024
Ascensia Diabetes Care Shahrir Alam Regulatory Affairs Product Manager 100 Summit Lake Drive Valhalla, New York 10595
Re: K231679
Trade/Device Name: CONTOUR® PLUS BLUE Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: January 19, 2024 Received: January 22, 2024
Dear Shahrir Alam:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Leslie Landree -S
for Joshua Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Submission Number (if known)
K231679 Device Name
CONTOUR® PLUS BLUE Blood Glucose Monitoring System
Indications for Use (Describe)
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System consists of the CONTOUR® PLUS BLUE meter, the CONTOUR® PLUS blood glucose test strips and the Contour® Diabetes app.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The CONTOUR® PLUS blood glucose test strips are for use with the CONTOUR® PLUS BLUE meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips.
The system is intended for in vitro diagnostic use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Ascensia Diabetes Care. The logo features a stylized "A" shape in purple next to a blue teardrop shape. Below the shapes, the word "ASCENSIA" is written in a dark purple sans-serif font, and below that, the words "Diabetes Care" are written in a smaller font.
510(k) Summary: K231679
Date prepared: February 22, 2024
According to the requirements of 21 CFR 807.92, the following information is being submitted in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.
| 1) Submitter | Shahrir AlamRegulatory Affairs Product ManagerAscensia Diabetes Care100 Summit Lake DriveValhalla, NY 10595Phone: 914-266-8655Email address: shahrir.alam@ascensia.com |
|---|---|
| 2) Device name: | Trade name: CONTOUR® PLUS BLUE BloodGlucose Monitoring SystemCommon name: Glucose test systemClassification name: System, Test, Blood Glucose,Over The CounterRegulation Number: 21 CFR 862.1345Product Code: NBW |
| 3) Predicate device: | CONTOUR® NEXT GEN Blood Glucose MonitoringSystem (K193407) |
| 4) Device description: | CONTOUR® PLUS BLUE Blood Glucose MonitoringSystem is a blood glucose monitoring system withBluetooth Low Energy technology built in so that themeter can communicate wirelessly to smart phonesand tablets. The meter uses the CONTOUR® PLUSblood glucose test strips and CONTOUR® PLUScontrol solution. The meter can be connected to theContour® Diabetes app. It utilizes a similar algorithmas the one used in the CONTOUR® NEXT GENBlood Glucose Monitoring System. It uses tworeplaceable coin cell batteries. The meter's shape is atraditional oval form factor and it includes an on-screen arrow that points to the color indicating if aglucose result is above, within, or below target range. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ascensia Diabetes Care. The logo features a stylized "A" shape, with the left side in purple and the right side represented by a blue teardrop shape. Below the shape, the word "ASCENSIA" is written in a bold, sans-serif font, and below that, the words "Diabetes Care" are written in a smaller, lighter font.
- Intended Use: The CONTOUR® PLUS BLUE Blood Glucose Monitoring System consists of the CONTOUR® PLUS BLUE meter, the CONTOUR® PLUS blood glucose test strips and the Contour® Diabetes app.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® PLUS BLUE Blood Glucose Monitoring System is intended for selftesting outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR® PLUS BLUE Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The CONTOUR® PLUS blood glucose test strips are for use with the CONTOUR® PLUS BLUE meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips.
The system is intended for in vitro diagnostic use onlv.
Indications for Use The subject CONTOUR® PLUS BLUE Blood Glucose Comparison: Monitoring System has the same intended use/indications for use as the predicate CONTOUR® NEXT GEN Blood Glucose Monitoring System.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ascensia Diabetes Care. The logo features a stylized purple triangle next to a blue teardrop shape. Below these shapes, the word "ASCENSIA" is written in a bold, sans-serif font, with the words "Diabetes Care" written in a smaller font underneath.
- Technological Comparison
The subject CONTOUR® PLUS BLUE Blood Glucose Monitoring System is substantially equivalent to its predicate CONTOUR® NEXT GEN Blood Glucose Monitoring System (K193407). The subject device has been designed with the same intended use, measurement principle, operating principle, electrochemical reaction as its predicate. In comparison to its predicate, CONTOUR® PLUS BLUE is also similar in terms of performance characteristics.
Non-Clinical and Clinical Tests Summary & Conclusions
Bench testing showed that the CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications. Usability testing was conducted to ensure that the CONTOUR® PLUS BLUE Blood Glucose Monitoring System was easy to use by typical customers.
Clinical testing showed that the CONTOUR® PLUS BLUE Blood Glucose Monitoring System performed as intended and met the system specifications.
Based on the outcome of the nonclinical and clinical testing conducted, the CONTOUR® PLUS BLUE Blood Glucose Monitoring System is substantially equivalent to the predicate CONTOUR® NEXT GEN Blood Glucose Monitoring System (K193407).
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.