K Number
K231614
Date Cleared
2023-12-13

(194 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MiroTract Wound Matrix is intended for the management of wounds including:

· Partial and full thickness wounds

  • Pressure ulcers
  • Venous ulcers
  • · Chronic vascular ulcers
  • · Diabetic ulcers
  • · Tunneled, undermined wounds
  • · Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
  • · Draining wounds

· Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence)

Device Description

The Reprise Biomedical MiroTract device consists of the MiroTract Wound Matrix and disposable delivery system. The MiroTract Wound Matrix is a sterile, single use, noncrosslinked acellular wound dressing that is derived from porcine liver tissue. The porcine liver is perfusion decellularized resulting in a collagen matrix that is dried, cut to size, and radially compressed onto the guidewire of the MiroTract delivery system. The delivery system includes a guidewire and tamp tube to manually push the MiroTract Wound Matrix off the guidewire into a wound. The MiroTract Wound Matrix porous scaffold provides a protective environment for wound healing. The MiroTract Wound Matrix is provided in two diameters (3 mm and 5 mm) and two lengths (5 cm and 9 cm). The MiroTract device is packaged dry in a plastic tray placed inside a foil pouch. The device is terminally sterilized in its packaging by e-beam irradiation. An optional Introducer Set accessory consisting of a tearaway sheath and dilator is provided in a separately packaged sterile pouch that is placed in a shelf box along with a MiroTract packaged device.

AI/ML Overview

The provided text does not contain information about the acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of clinical performance or diagnostic accuracy.

The document describes a 510(k) submission for a medical device called "MiroTract Wound Matrix". A 510(k) submission is primarily focused on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving the device meets new acceptance criteria through a clinical study.

The "Performance Data" section (VIII) lists several types of testing performed:

  1. Biocompatibility Testing: This ensures the device is safe for biological contact.
  2. Bench Testing: This includes:
    • Package integrity
    • Package and product stability
    • Collagen denaturation (a material property)
    • Dimensional testing (ensuring it's within specified size tolerances)
    • MR testing (for MRI compatibility)
    • Deliverability assessment (how well the device can be delivered)

These tests are primarily focused on the safety, material properties, and physical performance characteristics of the device, and ensuring it performs as intended from an engineering and safety perspective. They are not designed to determine clinical efficacy or diagnostic accuracy, which would typically involve human subject studies with specific clinical acceptance criteria.

Therefore, I cannot provide the requested information for the following reasons:

  • No acceptance criteria for clinical performance: The document does not specify clinical acceptance criteria (e.g., wound healing rates, infection reduction percentages, diagnostic sensitivity/specificity).
  • No study proving clinical acceptance criteria met: The listed "performance data" is bench and biocompatibility testing, not a clinical study designed to measure effectiveness in human patients against predefined clinical endpoints.
  • No information on sample size for a test set (clinical): Since no clinical test set for efficacy is described, there's no sample size, data provenance, or details on expert adjudication for ground truth.
  • No MRMC study: A multi-reader multi-case study is not applicable as this is a wound matrix, not an imaging diagnostic device.
  • No standalone algorithm performance: This is a physical wound matrix, not a software algorithm.
  • No ground truth type for clinical efficacy: As above, no clinical efficacy study is detailed.
  • No sample size for training set (clinical/AI): Given it's not an AI/diagnostic device, this is not applicable.
  • No ground truth establishment for training set: As above, not applicable.

In summary, the provided text describes the technical and material safety testing of a wound matrix device to demonstrate substantial equivalence, not its clinical efficacy against specific performance criteria derived from a human study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 13, 2023

Reprise Biomedical, Inc. % Kathy Herzog Sr Regulatory, Quality, and Compliance Consultant DuVal & Associates 1820 Medical Arts Building Suite 1820 Minneapolis, Minnesota 55402

Re: K231614

Trade/Device Name: MiroTract Wound Matrix Regulatory Class: Unclassified Product Code: KGN Dated: November 9, 2023 Received: November 13, 2023

Dear Kathy Herzog:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231614

Device Name MiroTract Wound Matrix

Indications for Use (Describe)

The MiroTract Wound Matrix is intended for the management of wounds including:

· Partial and full thickness wounds

  • Pressure ulcers
  • Venous ulcers
  • · Chronic vascular ulcers
  • · Diabetic ulcers
  • · Tunneled, undermined wounds
  • · Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
  • · Draining wounds

· Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence)

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K231614 510(k) Summary

This 510(k) summary is prepared in accordance with the requirements of 21 CFR §807.92:

SUBMITTER I.

Reprise Biomedical, Inc. 17400 Medina Road, Suite 100 Plymouth, MN 55447 763-284-6795

Contact Person: Carrie Powers Date Prepared: December 12, 2023

II. DEVICE

Trade/Proprietary Names:MiroTract Wound Matrix
Common Name:Animal-derived, extracellular matrix wound care product
Regulation Number:Unclassified
Regulation Name:NA
Device Class:Unclassified
Product Code:KGN
Panel:General & Plastic Surgery

III. PREDICATE DEVICE

Miro3D Wound Matrix (K223257). This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Reprise Biomedical MiroTract device consists of the MiroTract Wound Matrix and disposable delivery system. The MiroTract Wound Matrix is a sterile, single use, noncrosslinked acellular wound dressing that is derived from porcine liver tissue. The porcine liver is perfusion decellularized resulting in a collagen matrix that is dried, cut to size, and radially compressed onto the guidewire of the MiroTract delivery system. The delivery system includes a guidewire and tamp tube to manually push the MiroTract Wound Matrix off the guidewire into a wound. The MiroTract Wound Matrix porous scaffold provides a protective environment for wound healing. The MiroTract Wound

{5}------------------------------------------------

Matrix is provided in two diameters (3 mm and 5 mm) and two lengths (5 cm and 9 cm). The MiroTract device is packaged dry in a plastic tray placed inside a foil pouch. The device is terminally sterilized in its packaging by e-beam irradiation. An optional Introducer Set accessory consisting of a tearaway sheath and dilator is provided in a separately packaged sterile pouch that is placed in a shelf box along with a MiroTract packaged device.

V. INDICATIONS FOR USE

The MiroTract Wound Matrix is indicated for the following:

The MiroTract Wound Matrix is intended for the management of wounds including:

  • . Partial and full thickness wounds
  • Pressure ulcers ●
  • Venous ulcers ●
  • Chronic vascular ulcers ●
  • Diabetic ulcers ●
  • Tunneled, undermined wounds ●
  • Trauma wounds (abrasions, lacerations, partial thickness burns, and skin tears) ●
  • Draining wounds ●
  • Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence)

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The subject MiroTract Wound Matrix has the same intended use and similar technological characteristics as the predicate Miro3D Wound Matrix. Both the MiroTract and Miro3D wound matrices are made from porcine liver that is decellularized using the same perfusion process, materials, and concentrations. For Miro3D, the prepared liver tissue is dried and cut into four defined sizes in cube shape. For MiroTract, the prepared liver tissue is dried and cut into a defined size, threaded and radially compressed onto the guidewire of the MiroTract delivery system to create four MiroTract Wound Matrix sizes (3 and 5 mm Ø each offered in a 5 cm or 9 cm length). The change in wound matrix configuration (dry, uncompressed for Miro3D and dry, compressed onto a guidewire for MiroTract) does not change the principle of operation of the collagen wound matrices for wound management. Both the subject and predicate wound matrices are hydrated with the same hydrating solutions, either before or after wound placement. The MiroTract device includes a delivery system that consists of a nitinol guidewire with stainless steel tip and Pebax tamp tube. An optional Introducer Set accessory that consists of a tearaway sheath and dilator is also provided. MiroTract uses the same tray packaging material and outer foil pouch as Miro3D and is sterilized by the same sterilization method, e-beam

{6}------------------------------------------------

irradiation. The MiroTract packaging includes a packaging stopper disc to prevent the tamp tube from sliding off the guidewire when the device is removed from the tray.

FeatureMiroTract Wound Matrix(Subject Device)Miro3DWound Matrix(Predicate Device, K223257)
ClassificationUnclassified (pre-amendment)Unclassified (pre-amendment)
Product CodeKGNKGN
ClassUnclassifiedUnclassified
Intended UseWound managementWound management
Indications ForUseThe MiroTract Wound Matrixis intended for the managementof wounds including:· Partial and full thicknesswounds• Pressure ulcers• Venous ulcers• Chronic vascular ulcers• Diabetic ulcers• Tunneled, underminedwounds• Trauma wounds (abrasions,lacerations, partial thicknessburns, and skin tears)• Draining wounds· Surgical wounds (donorsites/grafts, post-Mohs'surgery, post-laser surgery,podiatric, wound dehiscence)The Miro3D Wound Matrix isintended for the management ofwounds including:• Partial and full thicknesswounds• Pressure ulcers• Venous ulcers• Chronic vascular ulcers• Diabetic ulcers• Tunneled, underminedwounds· Trauma wounds (abrasions,lacerations, partial thicknessburns, and skin tears)• Draining wounds• Surgical wounds (donorsites/grafts, post-Mohs'surgery, post-laser surgery,podiatric, wound dehiscence)
UserPhysician or other cliniciantrained in wound carePhysician or other cliniciantrained in wound care
Intended UseEnvironmentSurgical suite, hospital,ambulatory surgery center orout-patient clinicSurgical suite, hospital,ambulatory surgery center orout-patient clinic
DescriptionResorbable, animal-sourced,non-crosslinked, acellularcollagen tissue matrixResorbable, animal-sourced,non-crosslinked, acellularcollagen tissue matrix
Principle ofOperationProvide a protectiveenvironment for woundhealingProvide a protectiveenvironment for woundhealing
FeatureMiroTract Wound Matrix(Subject Device)Miro3DWound Matrix(Predicate Device, K223257)
Wound MatrixConfigurationand SizesCompressed, cylindrical,provided in 4 sizes: Diameter(mm) x length (cm) (ModelNumber)• 3 x 5 (5000)• 3 x 9 (5010)• 5 x 5 (5020)• 5 x 9 (5030)Cube shape provided in foursizes (W x L x H) (cm) (ModelNumber)• 2 x 2 x 2 (3000)• 3 x 3 x 2 (3005)• 5 x 5 x 2 (3010)• 10 x 5 x 2 (3015)
Wound MatrixPreparationUse dry; may rehydrate afterplacement with either sterilesaline or lactated Ringer'ssolutionRehydrate a minimum of fiveminutes in either sterile salineor lactated Ringer's solutionprior to application
Delivery MethodManually placed using:• Delivery system (nitinolguidewire with stainlesssteel tip and Pebax tamptube) with or without anoptional Introducer Set, or• General surgicalinstrumentsManually placed usinggeneral surgical instruments
Packaging• Device package:Packaged dry in a PETGtray with solid snap-onPETG lid and siliconefoam stopper disc• Sterile barrier:Aluminum laminate foilpouch• Shelf box: Cardboard• Device package:Packaged dry in a PETGtray with notched PETGsnap-on lid with Tyveklid seal• Sterile barrier: Aluminumlaminate foil pouch• Shelf box: Cardboard
StorageConditionsNo special storage conditionsrequiredNo special storage conditionsrequired
Shelf Life2 months (as of submissiondate)25 months
SterilizationMethodElectron beam irradiationElectron beam irradiation
SterilizationAssurance Level(SAL)10-610-6
FeatureMiroTract Wound Matrix(Subject Device)Miro3DWound Matrix(Predicate Device, K223257)
MRCompatibilityMiroTract Wound Matrix:MR SafeDelivery System: MR UnsafeMR Safe
BiocompatibilityBiocompatibility testing orjustification for all applicablebiological endpoints per ISO10993-1:2018 were completedBiocompatibility testing orjustification for all applicablebiological endpoints per ISO10993-1:2018 were completed

Table 1: Subject vs. Predicate Wound Matrix Comparison

{7}------------------------------------------------

{8}------------------------------------------------

VII. PERFORMANCE DATA

The following performance testing were provided in support of the substantial equivalence determination:

MiroTract Wound Matrix

1. Biocompatibility Testing

  • Cytotoxicity, ISO 10993-5:2009 a.
  • b. Sensitization, ISO 10993-10:2021
  • Intracutaneous Reactivity, ISO 10993-10:2021 C.
  • Acute Systemic Toxicity, ISO 10993-11: 2017 d.
  • Material Mediated Pyrogenicity, ISO 10993-11:2017 e.
  • Systemic Toxicity, ISO 10993-11:2017 f.
  • Implantation Testing, ISO 10993-6:2016 ಇತ
  • Genotoxicity (Ames Bacterial Reverse Mutation and Mouse Lymphoma), ISO h. 10993-3:2014

2. Bench Testing

  • Package integrity testing per ASTM F88-09, ISO 11607-1:2019, ASTM F2825a. 18, ASTM D4169-16, and ASTM F2096-11
  • b. Package and product stability testing
  • Collagen denaturation ASTM D3418 C.
  • d. Dimensional: MiroTract Wound Matrix is within tolerance of labeled dimensions for Length and Diameter after sterilization, environmental conditioning, and real time aging
  • MR testing per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19, ASTM e. F2119-07
  • Deliverability assessment of MiroTract Wound Matrix used with the delivery f. system components.

VIII. CONCLUSIONS

The subject MiroTract Wound Matrix has the same Intended Use as the predicate Miro3D Wound Matrix for wound management. The differences in technological

{9}------------------------------------------------

differences have been demonstrated to not raise different questions of safety and effectiveness. Performance testing provides evidence that the MiroTract Wound Matrix performs as intended and is as safe and effective as the predicate device. Therefore, the subject MiroTract Wound Matrix device is substantially equivalent to the predicate Miro3D Wound Matrix device (K223257).

N/A