(156 days)
The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.
The Endophys BPM-30 Blood Pressure Monitor is an electronic device that provides compatibility between a physiological fiber optic blood pressure sensor (transducer) and conventional invasive arterial blood pressure inputs to a standard physiological patient monitor. The device converts the optical transducer data to electrical signals that are interpreted by a conventional patient monitor and/or are displayed directly on the BPM-30. The BPM-30 accurately emulates a fluidic arterial blood pressure transducer and supplies electrical signals to its output that are indistinguishable from a conventional fluidic blood pressure transducer.
The BPM-30 is implemented as a self-contained unit that has a fiber optic transducer connection as an input source and communicates with a patient monitor as its output. The BPM acts to directly emulate the electrical interface characteristics of conventional fluidic blood pressure transducers (that patient monitors are compatible with) while providing much more precise blood pressure data derived from a fiber optic transducer placed within an artery. Electrically emulating a conventional fluidic transducer uniquely allows a fiber optic pressure sensor to be used with a wide variety of existing physiological patient monitors without modification of those monitors. Systolic, diastolic, and mean blood pressure values are also displayed directly on the BPM-30 every four seconds.
The BPM-30 is powered by a standard AC power adapter. The BPM-30 also has an internal battery supply that will automatically recharge and maintain operation during brief (up to 30 minutes) power interruptions. The BPM-30 is used outside of the sterile environment and has standard alerts and alarms.
The BPM-30 has an operating pressure range of -30 to 300 mmHg with an accuracy of ±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg and within ±3% of reading for pressures ≥ 50 mmHg
The provided text describes a 510(k) premarket notification for the Endophys Blood Pressure Monitor Model BPM-30. It outlines the device's description, intended use, and a comparison to a predicate device (Endophys Blood Pressure Monitor model BPM-20). While it mentions "Performance Data" and lists several types of testing conducted, it does not provide specific acceptance criteria or the reported device performance in terms of accuracy metrics against a ground truth as typically expected for proving the device meets those criteria.
Instead, the performance data section states: "The modified Endophys Blood Pressure Monitor met all specified criteria and did not raise new safety or performance questions. Based on the design verification performance the modified Endophys Blood Pressure Monitor was found to have the safety and effectiveness profile that is similar to the predicate device." This indicates that the device was tested to meet internal design specifications and regulatory standards, and its performance was deemed similar to the previously cleared predicate.
Given the information provided, I cannot construct a table of acceptance criteria and reported device performance with specific numerical accuracy values. The document focuses on regulatory compliance and substantial equivalence to a predicate device, rather than a detailed clinical performance study with defined ground truth and expert evaluation.
However, I can extract the accuracy specification for the device, which would likely form part of its inherent design acceptance criteria.
Here's an attempt to answer the request based only on the provided text, highlighting what is implicitly or explicitly stated and what is missing:
Acceptance Criteria and Device Performance for Endophys Blood Pressure Monitor Model BPM-30
The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (Endophys Blood Pressure Monitor Model BPM-20) rather than presenting a detailed clinical study for novel performance claims. Therefore, specific numerical acceptance criteria and a detailed breakdown of reported device performance in a clinical setting against a robust ground truth (e.g., expert consensus, pathology) are not explicitly provided in the document.
The document does state the performance specifications of the device, which implicitly serve as acceptance criteria for its design and manufacturing.
1. Table of Acceptance Criteria and the Reported Device Performance
Based on the "Summary of Technical Characteristics as Compared to the Predicate Device" and "Device Description" sections, the key performance specification mentioned is accuracy:
| Acceptance Criterion (Specification) | Reported Device Performance (as stated in document) | Notes from text |
|---|---|---|
| Pressure measurement range | -30 to 300 mmHg | Stated as a technical characteristic and part of the device description. |
| Pressure accuracy (≤ 50 mmHg) | ±1 mmHg plus ±1% of reading | Stated as a technical characteristic and part of the device description. The text indicates "The modified Endophys Blood Pressure Monitor met all specified criteria". |
| Pressure accuracy (≥ 50 mmHg) | within ±3% of reading | Stated as a technical characteristic and part of the device description. The text indicates "The modified Endophys Blood Pressure Monitor met all specified criteria". |
It is important to note that the document does not provide raw data, statistical analyses, or specific clinical study results to demonstrate how these accuracy figures were met in a performance study with a test set. It merely states that the "modified Endophys Blood Pressure Monitor met all specified criteria." The "Performance Data" section lists types of testing (Software Verification, Electrical Safety, EMC, Packaging, Functional Testing) but does not detail the results of these tests in terms of the accuracy specifications.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified. The document lists types of performance data (e.g., functional testing, electrical safety) but does not mention the sample size of a patient-based test set or a clinical study.
- Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not specified. There is no mention of the origin or nature of data for any performance testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable/Not specified. The document does not describe a study involving expert readers or the establishment of ground truth in a clinical context (e.g., for interpretation of medical images). The testing appears to be primarily engineering verification and validation against technical specifications rather than a comparative clinical performance study.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable/None specified. Since there's no mention of a clinical test set requiring expert interpretation or adjudication, this information is not relevant to the provided text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device described (Blood Pressure Monitor) is not an AI-assisted diagnostic device that would typically undergo an MRMC study. It is a measurement device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, implicitly. The accuracy specifications (±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg, and within ±3% of reading for pressures ≥ 50 mmHg) are for the device itself, implying its standalone measurement performance. This is typically established through bench testing and calibration, which aligns with "Functional Testing" mentioned in the "Performance Data" section.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For a blood pressure monitor, the ground truth would typically be established by a highly accurate reference blood pressure measurement device (e.g., a calibrated pressure transducer or manometer) or a standardized pressure source in a laboratory setting. The document does not explicitly state the type of ground truth, but it would align with calibrated reference measurements used during functional and design verification testing.
8. The sample size for the training set
Not applicable/Not specified. This device is a traditional medical device, not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This device is not an AI/ML device that requires a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 3, 2023
Endophys Technologies LLC % Mike Winegar Principal Winegar Consulting, Inc. 7829 Ithaca Ln N Maple Grove, Minnesota 55311
Re: K231586
Trade/Device Name: Endophys Blood Pressure Monitor Model BPM-30 Regulation Number: 21 CFR 870.1110 Regulation Name: Blood Pressure Computer Regulatory Class: Class II Product Code: DSK Dated: May 30, 2023 Received: May 31, 2023
Dear Mike Winegar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Form Approved: OMB No. 0910-0120 Food and Drug Administration Expiration Date: 6/30/2023 See PRA Statement below. Indications for Use
510(k) Number (if known)
Device Name
Endophys Blood Pressure Monitor Model BPM-30
Indications for Use (Describe)
Indications for Use
The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Endophys. The logo features the word "Endophys" in bold, with the "Endo" portion in black and the "phys" portion in red. A red line resembling a heartbeat is to the left of the word. Below the word, there is a tagline that reads "Measuring physiology where it happens."
510(k) SUMMARY
| Applicant | Endophys Technologies, LLC1933 E Levee St, Suite BDallas, TX 75207 |
|---|---|
| Contact Person | Mike WinegarPrincipalWinegar Consulting, Inc.763-639-0700mwinegar@comcast.net |
| Summary Date | September 29, 2023 |
| Proprietary Name | Endophys Blood Pressure Monitor model BPM-30 |
| Classification | Class II |
| Classification Name | Computer, Blood-Pressure |
| Regulation Number | 21 CFR 870.1110 |
| Product Code | DSK |
| Predicate Device | The Endophys Blood Pressure Monitor model BPM-30 is substantially equivalent to the currentlymarketed Endophys Blood Pressure Monitor modelBPM-20 (K160945) |
{4}------------------------------------------------
Device Description:
The Endophys BPM-30 Blood Pressure Monitor is an electronic device that provides compatibility between a physiological fiber optic blood pressure sensor (transducer) and conventional invasive arterial blood pressure inputs to a standard physiological patient monitor. The device converts the optical transducer data to electrical signals that are interpreted by a conventional patient monitor and/or are displayed directly on the BPM-30. The BPM-30 accurately emulates a fluidic arterial blood pressure transducer and supplies electrical signals to its output that are indistinguishable from a conventional fluidic blood pressure transducer.
The BPM-30 is implemented as a self-contained unit that has a fiber optic transducer connection as an input source and communicates with a patient monitor as its output. The BPM acts to directly emulate the electrical interface characteristics of conventional fluidic blood pressure transducers (that patient monitors are compatible with) while providing much more precise blood pressure data derived from a fiber optic transducer placed within an artery. Electrically emulating a conventional fluidic transducer uniquely allows a fiber optic pressure sensor to be used with a wide variety of existing physiological patient monitors without modification of those monitors. Systolic, diastolic, and mean blood pressure values are also displayed directly on the BPM-30 every four seconds.
The BPM-30 is powered by a standard AC power adapter. The BPM-30 also has an internal battery supply that will automatically recharge and maintain operation during brief (up to 30 minutes) power interruptions. The BPM-30 is used outside of the sterile environment and has standard alerts and alarms.
The BPM-30 has an operating pressure range of -30 to 300 mmHg with an accuracy of ±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg and within ±3% of reading for pressures ≥ 50 mmHg
Intended Use:
The Endophys BPM-30 Blood Pressure Monitor is meant to be used in minimally invasive vascular procedures and critical patient care situations where the accuracy and timeliness of arterial blood pressure measurements are important. It supports fiber optic transducers that are incorporated into the Endophys PressureSense™ hydrophilic coated Pressure Sensing Sheath (PSS3).
When used with the Endophys Pressure Sensing Sheath, the BPM-30 is designed to provide continuous and accurate systolic, diastolic and mean blood pressure data with no other catheter, pressure transducer or fluid line required.
{5}------------------------------------------------
Indications for Use:
The Endophys Blood Pressure Monitor Model BPM-30 is intended to continuously provide systolic, diastolic and mean blood pressure based on the output of the Endophys PSS3 Pressure Sensing Sheath in patients undergoing therapeutic and/or diagnostic procedures involving percutaneous vascular access. The Endophys Blood Pressure Monitor Model BPM-30 monitor is additionally intended for use in transport situations within hospital environments.
Summary of Technical Characteristics as Compared to the Predicate Device:
Blood Pressure Monitors, by nature, are instruments intended to provide clinicians with the physiological blood pressure of patients reported in milligrams of mercury (mmHg). Specialized computer based instruments have been developed which allow active monitoring of blood pressure during interventional procedures. Blood pressure monitoring using a fiber optic communication is the technological principle for both the subject and predicate devices. The technology requires use of a compatible intravascular pressure sensing sheath to provide this data to the clinician.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Calculates and displays hemodynamic (invasive blood pressure) data when attached to a compatible intravascular pressure sensing device (Endophys PSS3 Pressure Sensing Sheath)
- Optical signal (fiber optic) measurement of blood pressure ●
- Pressure measurement range -30 to 300 mmHg ●
- Pressure accuracy: System accuracy ±1 mmHg plus ±1% of reading for pressures ≤ 50 mmHg ● and within ±3% of reading for pressures ≥ 50 mmHg
The following technological differences exist between the subject and predicate devices:
- Modification of device design to be compatible with FISO Technologies Inc. Fabrey-Perot 0 fiber optic pressure sensors and circuitry incorporated into the Endophys Pressure Sensing Sheath (PSS3)
- Combine the functionality of the daughterboard onto the main circuit board, which were . separate in the predicate device
- Addition of an internal battery supply that will automatically recharge and maintain 0 operation during brief (up to 30 minutes) patient transport or power interruptions
{6}------------------------------------------------
Performance Data
The following performance data were provided in support of the substantial equivalence:
- o Software Verification
- Electrical Safety and Essential Performance Testing (IEC60601-1, IEC60601-1-8, o IEC60601-2-34)
- EMC Testing (IEC60601-1-2)
- Packaging Design Verification ●
- Package, Device, and Labeling Inspection ●
- Functional Testing ●
The modified Endophys Blood Pressure Monitor met all specified criteria and did not raise new safety or performance questions. Based on the design verification performance the modified Endophys Blood Pressure Monitor was found to have the safety and effectiveness profile that is similar to the predicate device.
Conclusion:
Endophys believes the proposed BPM-30 Blood Pressure Monitor is substantially equivalent to the legally marketed predicate device. The indications for use, methods of operation, design and materials used are either identical or substantially equivalent to the existing legally marketed predicate product.
§ 870.1110 Blood pressure computer.
(a)
Identification. A blood pressure computer is a device that accepts the electrical signal from a blood pressure transducer amplifier and indicates the systolic, diastolic, or mean pressure based on the input signal.(b)
Classification. Class II (performance standards).