K Number
K231576
Device Name
cNeuro cPET
Manufacturer
Date Cleared
2023-09-14

(106 days)

Product Code
Regulation Number
892.1200
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans.

The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information relative changes in PET-FDG glucose metabolism.

cNeuro cPET additionally allows the user to generate information relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration.

PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy.

cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.

Device Description

cNeuro cPET has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET scans. The software enables the display, co-registration, and fusion of PET images with those from MRI. Additionally, cPET enables automated quantitative and statistical analysis of tracer uptake by registration of a volume-of-interest atlas to the PET images and by comparing voxel and region-based uptake with corresponding uptake in healthy, amyloid-negative subjects. There are two quantification pipelines, one when the patient's MRI is not available (PET-only) and one when the patient's MRI is available (PET-MR). Quantification results are presented using volumes of interest, voxel-based or 3D stereotactic surface projection maps of the brain.

AI/ML Overview

The provided text describes the regulatory filing for cNeuro cPET (K231576), a device intended to aid physicians in the evaluation and quantification of PET brain scans for pathologies, particularly related to FDG glucose metabolism and amyloid load.

Here's an analysis of the acceptance criteria and the study proving the device meets these criteria, based on the provided information:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly derived from the performance evaluation conducted, demonstrating the device's accuracy, robustness, and agreement with established methods and a predicate device. While explicit "acceptance criteria" values are not presented in a table form, the performance summary details the results proving the device meets its intended use and is substantially equivalent to its predicate.

Performance Metric / Acceptance Criteria (Implied)Reported Device Performance (cNeuro cPET)
Agreement with Predicate Device (CortexID Suite)Correlations (R) ranging from 0.98 – 0.99 depending on cohort.
Agreement with FreeSurfer based PET-MR quantificationStrong correlation (R) of 0.98 – 0.99.
Test-Retest Variability (TRT) - Flutemetamol (Pons reference)PET-MR: 1.2% PET-only: 1.7%
Test-Retest Variability (TRT) - Florbetapir (Whole cerebellum reference)PET-MR: 1.3% PET-only: 1.8%
Agreement with SoT (Histopathology) - Florbetaben CategorizationPET-only: 94.4% PET-MR: 96.3%
Agreement with SoT (Majority Visual Read) - Florbetaben CategorizationPET-only: 94.8% PET-MR: 92.8%
Agreement with SoT (Unanimous Visual Read) - Florbetaben CategorizationPET-only: 98.8% PET-MR: 98.2%
Correlation of Centiloids with Centiloid Project valuesR$^2$ ranging from 0.85 - 0.99 depending on tracer and reference region.
Compliance with DICOM standardComplies with NEMA PS 3.1 - 3.20 (2021) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology).

2. Sample Size and Data Provenance

  • Test Set Sample Size: 2,275 subjects.
  • Data Provenance: The text explicitly mentions "a large dataset from the Alzheimer's Disease Neuroimaging Initiative (ADNI)", indicating a combination of retrospective and potentially prospective data, likely from diverse geographical origins given ADNI's multi-national nature, although specific countries are not stated. It also refers to Florbetaben Phase III data. The type of tracers supported (Flutemetamol, Florbetaben, FDG) were used in the testing.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: For the majority visual read ground truth, 5 readers were used.
  • Qualifications: The text does not explicitly state the qualifications of these readers (e.g., "radiologist with 10 years of experience"). However, given the context of PET brain scan interpretation, it can be inferred they are medical professionals specializing in nuclear medicine or radiology.

4. Adjudication Method for the Test Set

  • For the visual read ground truth for Florbetaben categorization: A "majority visual read" by 5 readers was used. This implies a 3 out of 5 agreement for positive/negative classification.
  • Additionally, a subset of scans where there was "unanimous categorization by the readers" (all 5 readers agreed) was analyzed separately.
  • For histopathology-based ground truth, no adjudication method is relevant as it's a direct pathological diagnosis.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • The document does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance is compared.
  • The study focuses on the standalone performance of the cNeuro cPET software and its agreement with established methods (predicate, FreeSurfer) and ground truth (histopathology, expert consensus). While expert visual reads were used to establish ground truth, they were not used in a comparative "human-in-the-loop" study format.

6. Standalone Performance (Algorithm Only)

  • Yes, a standalone performance study was clearly conducted. The reported performance metrics (correlation with predicate, test-retest variability, agreement with histopathology and expert consensus) are all measures of the algorithm's performance independent of human-in-the-loop assistance. The device output includes quantitative metrics (SUVR, Z-scores, Centiloids) and maps for physician interpretation, but the performance testing itself is on the software's ability to generate these outputs accurately.

7. Type of Ground Truth Used

The study utilized multiple types of ground truth:

  • Expert Consensus: "Majority visual read" by 5 readers for Florbetaben scan categorization.
  • Pathology: "Histopathology" for Florbetaben scan categorization.
  • Reference Methods/Data:
    • Comparison with the predicate device (CortexID Suite).
    • Comparison with a FreeSurfer based PET-MR quantification method used in ADNI.
    • Validation against Centiloid Project values (a standardized method for quantitative amyloid plaque estimation).

8. Sample Size for the Training Set

  • The document does not specify the sample size used for the training set. The 2,275 subjects are explicitly stated as the dataset for "testing of cNeuro cPET". Without further information, it's unclear if parts of this dataset were used for initial training or if a separate, larger training set was used. It states that the "ADNI" dataset was used for "assessment of robustness," which might imply testing, but ADNI datasets are also often used for model development.

9. How Ground Truth for Training Set Was Established

  • Since the training set size is not provided, the method for establishing its ground truth is also not explicitly described. However, based on the performance evaluation, it is highly probable that similar methods (expert consensus, established reference methods, or potentially a form of "weak" or "noisy" labeling for initial training) would have been employed. The document mentions the device performs "automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects," which implies the use of a "normal database" that would have required some form of ground truth or normative data establishment during development.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

September 14, 2023

Combinostics Oy % Erin Gontang Senior Consultant Ram+ 2251 San Diego Avenue, B-257 San Diego, California 92110

Re: K231576

Trade/Device Name: cNeuro cPET Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, LLZ Dated: May 31, 2023 Received: May 31, 2023

Dear Erin Gontang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Signature

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K231576

Device Name

cNeuro cPET

Indications for Use (Describe)

cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans.

The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information relative changes in PET-FDG glucose metabolism.

cNeuro cPET additionally allows the user to generate information relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration.

PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy.

cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of blue and red squares on the left, followed by the word "Combinostics" in a dark gray, sans-serif font. The brain design is abstract, with the squares arranged to suggest the shape of a brain.

510(k) Summary (K231576)

DATE PREPARED

August 17, 2023

MANUFACTURER AND 510(k) OWNER

Combinostics OY Hatanpään valtatie 24 FI 33100 Tampere, Finland Telephone: Official Contact: Lennart Thurfjell, CEO

REPRESENTATIVE/CONSULTANT

Erin A. Gontang, Ph.D. Allison C. Komiyama, Ph.D., RAC RQM+ +1 (412) 816-8253 Telephone: Email: egontang@rqmplus.com akomiyama@rqmplus.com

DEVICE INFORMATION

Proprietary Name/Trade Name:cNeuro cPET
Common Name:System, Image Processing, RadiologicalSystem, Tomography, Computed, Emission
Regulation Number:892.2050892.1200
Class:Class II
Product Code:LLZKPS
Premarket Review:Radiology
Review Panel:Office of Radiological Health (OHT8)Div. of Imaging Devices and Electronic Products (DHT8B)

PREDICATE DEVICE IDENTIFICATION

The cNeuro cPET is substantially equivalent to the following predicate:

510(k) NumberPredicate Device Name / ManufacturerPrimary Predicate
K141074CortexID Suite / GE Medical Systems, LLC

The predicate device have not been subject to a design related recall.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of blue and red squares on the left, followed by the company name "Combinostics" in a dark gray, sans-serif font. The logo is simple and modern, with a focus on the company name.

DEVICE DESCRIPTION

cNeuro cPET has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET scans. The software enables the display, co-registration, and fusion of PET images with those from MRI. Additionally, cPET enables automated quantitative and statistical analysis of tracer uptake by registration of a volume-of-interest atlas to the PET images and by comparing voxel and region-based uptake with corresponding uptake in healthy, amyloid-negative subjects. There are two quantification pipelines, one when the patient's MRI is not available (PET-only) and one when the patient's MRI is available (PET-MR). Quantification results are presented using volumes of interest, voxel-based or 3D stereotactic surface projection maps of the brain.

INDICATIONS FOR USE

cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans.

The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information regarding relative changes in PET-FDG glucose metabolism.

cNeuro cPET additionally allows the user to generate information regarding relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration.

PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy.

cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Combinostics believes that the cNeuro cPET is substantially equivalent to the predicate device based on the information summarized here:

The subject device and predicate device are software for the automated quantification, visualization, and reporting of PET FDG and amyloid brain scans. Both devices compute standard uptake value ratios (SUVR) where the value in a target region or voxel is divided with the value in a reference region. In addition, both devices provide comparison to tracer specific

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped icon made up of small squares in red and blue, followed by the word "Combinostics" in a dark gray, sans-serif font. The logo is simple and modern, and the colors are eye-catching.

reference data with results presented as region and voxel-based Z-scores and with results summarized in a report.

The subject device and predicate device achieve their intended use based on a similar principle, which is the quantification system relies on fitting a volume-of-interest atlas to the data, and through statistical comparison to reference data. The main difference being that the subject device performs the analysis in PET native space whereas the predicate device performs the analysis in a standardized template space.

Subject DevicePredicate Device
Combinostics OYGE Healthcare
cNeuro cPETCortexID Suite
K231576K141074
Indications for UsecNeuro cPET aids physicians in theevaluation of patient pathologies viaassessment and quantification of PETbrain scans.CortexID software has been developedto aid physicians in the evaluation ofpatient pathologies via assessment andquantification of PET brain scans.
The software aids in the assessment ofhuman brain PET scans enablingautomated analysis throughquantification of tracer uptake andcomparison with the correspondingtracer uptake in normal subjects. Theresulting quantification is presentedusing volumes of interest and voxel-based maps of the brain. cNeuro cPETallows the user to generate informationregarding relative changes in PET-FDGglucose metabolism.The software aids in the assessment ofhuman brain PET scans enablingautomated analysis throughquantification of tracer uptake andcomparison with the correspondingtracer uptake in normal subjects. Theresulting quantification is presentedusing volumes of interest, voxel-basedor 3D stereotactic surface projectionmaps of the brain. The package allowsthe user to generate informationregarding relative changes in the PET-FDG glucose metabolism.
cNeuro cPET additionally allows the userto generate information regardingrelative changes in PET brain amyloidload between a subject's images and anormal database, which may be theresult of brain neurodegeneration.PET co-registration and fusion displaycapabilities with MRI allow PET findingsto be related to brain anatomy.cNeuro cPET aids physicians in the imageinterpretation of PET studies conductedon patients being evaluated forCortexID Suite additionally allows theuser to generate information in PETbrain amyloid load between a subject'simages and a normal database, whichmay be the result of brainneurodegeneration.PET co-registration and fusion displaycapabilities with CT and MR allow PETfindings to be related to brain anatomyand offers visualization of structuralabnormalities, which may result frombrain injury, trauma, disorder, disease or
cognitive impairment, or other causes ofcognitive decline.dysfunction, such as subduralhematoma, tumor, stroke, orcerebrovascular disease, etc.
CortexID Suite may aid physicians in theimage interpretation process of PETstudies conducted on patients beingevaluated for cognitive impairment, orother causes of cognitive decline, and isan adjunct to other diagnosticevaluations.
Product Code(s)LLZ - System, Image Processing,RadiologicalKPS - System, Tomography, Computed,EmissionKPS - System, Tomography, Computed,Emission
Regulation Number/Description21 CFR 892.2050 / Medical imagemanagement and processing system21 CFR 892.1200 / Emission computedtomography system21 CFR 892.1200 / Emission computedtomography system
Import of PET andMRI ImagesUpload DICOM files from folder orthrough connection to PACS. PET imagesare mandatory, but MRI images areoptional.Upload DICOM files from folder orthrough connection to PACS. PET imagesare mandatory, but MRI images areoptional.
Supported tracersSupports FDG, Flutemetamol,Florbetaben and Florbetapir.Supports FDG, Flutemetamol,Florbetaben and Florbetapir, and theresearch tracer C11 PIB.
Volumes of interest(VOIs)A VOI-template is defined outlining VOIscorresponding to the different brainlobes, anterior and posterior cingulate,as well as reference VOIs correspondingto cerebellar gray matter, wholecerebellum and pons. There is also acortical composite VOI defined bymerging all the cortical VOIs.A VOI-template is defined outlining VOIscorresponding to the different brainlobes, anterior and posterior cingulate,as well as reference VOIs correspondingto cerebellar gray matter, wholecerebellum and pons. There is also acortical composite VOI defined bymerging all the cortical VOIs.
Method forquantificationFully automated registration to establishthe transformation between the PETimage and a standard template space. AVOI-template is then transformed to PETnative space and is used to quantifytracer uptake.Fully automated registration to establishthe transformation between the PETimage and a standard template space.The PET image is then transformedstandard space and a VOI-template isused to quantify tracer uptake.
Comparison to tracerspecific referencedataProvides reference data for allsupported tracers. The reference data isused to compute regional and voxel-based Z-scores.Provides reference data for FDG andFlutemetamol. The reference data isused to compute regional and voxel-based Z-scores.
Output fromquantificationResults are presented as SUVR images,Z-score images, regional SUVR and Z-scores and 3D SSP Z-score images. Foramyloid tracers results includeResults are presented as SUVR images,Z-score images, regional SUVR and Z-scores and 3D SSP SUVR and Z-scoreimages
Centiloids for a cortical composite region.
Co-registration with MRIIf the patient's MRI is available, this is co-registered with the PET for display purposes, and it is also used during image quantification.If the patient's MRI is available, this is co-registered with the PET for display purposes.
ReportProvides a PDF report that can be sent to PACS. The report summarizes region-based SUVR and Z-scores and shows 3D stereotactic surface projections of voxel-based Z-scores as well as Centiloids for a cortical composite region for amyloid tracers.Provides a PDF report that can be sent to PACS. The report summarizes region-based SUVR and Z-scores and shows 3D stereotactic surface projections of voxel-based Z-scores.

A comparison of the subject device and predicate device is provided below.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of small squares in blue and red, followed by the word "Combinostics" in a dark gray sans-serif font. The logo is simple and modern, with a focus on the company's name.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped icon made up of small squares in red and blue, followed by the word "Combinostics" in a dark gray sans-serif font. The logo is simple and modern, and the brain icon suggests that the company is involved in some kind of cognitive or analytical work.

SUMMARY OF PERFORMANCE TESTING

Summary of Non-Clinical Testing

The cNeuro cPET software complies with NEMA PS 3.1 - 3.20 (2021) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard.

cNeuro cPET employs the same fundamental scientific technology as its predicate device, CortexID Suite. cNeuro cPET uses the equivalent DICOM image data input requirements. It has equivalent display, formatting, archiving and visualization technologies compared to the predicate device. cNeuro cPET utilizes essentially the same methodology to quantify and assess uptake of FDG and amyloid tracers. The information is presented using volumes of interest or voxel-based maps of the brain. cNeuro cPET utilizes existing PET co-registration and fusion technologies and provides for PET & MR co-registration and fusion capabilities. Thorough testing of these capabilities has not raised any safety or effectiveness issues.

The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • Requirements Reviews
  • Design Reviews
  • Integration testing (System verification)
  • Performance testing (Bench testing, verification)
  • o Safety testing (Verification)

The substantial equivalence determination is based on the software documentation for a MODERATE level of concern device.

Combinostics performed testing of cNeuro cPET using PET images from 2,275 subjects. Data were from all tracers supported by cNeuro cPET, i.e., Flutemetamol, Florbetaben and FDG. The performance evaluation included a comparison with the predicate device,

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a stylized brain made of blue and red squares on the left, followed by the word "Combinostics" in a dark gray sans-serif font. The brain graphic is composed of multiple squares of varying sizes, arranged to resemble the shape of a brain.

assessment of robustness using a large dataset from the Alzheimer's Disease Neuroimaging Initiative (ADNI), evaluation of test-retest variability, and comparison with Standard of Truth (SoT) based on majority/consensus visual read and histopathology using Florbetaben Phase III data. Centiloids for amyloid tracers were validated according the methods described by W. E. Klunk et al., The Centiloid Project: standardizing quantitative amyloid plaque estimation by PET. Alzheimers. Dement. 11, 1-4 (2015).

Results showed agreement with the predicate device with correlations (R) ranging from 0.98 – 0.99 depending on cohort. Furthermore, a similar strong correlation (R=0.98 – R=0.99) was obtained for the comparison with a FreeSurfer based PET-MR quantification method used in ADNI. Evaluation of the consistency of cNeuro cPET quantification showed low test-retest variability (TRT) in the quantification of repeat scans. For Flutemetamol, the best results were obtained using Pons as the reference region (TRT = 1.2% for PET-MR, TRT = 1.7% for PET-only), and for Florbetapir, the lowest variability was obtained by using the whole cerebellum as the reference region (TRT = 1.3% for PET-MR, TRT = 1.8% for PET-only).

Categorization of Florbetaben scans into positive and negative based on cNeuro cPET quantification showed agreement with SoT based on histopathology (94.4% for PET-only and 96.3% for PET-MR). The agreement in the comparison of cNeuro cPET categorization with SoT based on majority visual read (5 readers) gave similar agreement (94.8% for PET-only and 92.8% for PET-MR). When excluding all scans where readers had different opinions, i.e., there was unanimous categorization by the readers of the scans into negative or positive, the agreement with cNeuro cPET categorization was 98.8% for PET-only and 98.2% for PET-MR.

The correlation of cNeuro cPET Centiloids for Flutemetamol, Florbetaben and Florbetapir with the corresponding values produced by the Centiloid Project ranged from R2 = 0.85 - 0.99 depending on tracer and reference region.

The results show that cNeuro cPET is accurate and robust, as well as effective and is suitable for its intended use. cNeuro cPET is substantially equivalent to the predicate device.

CONCLUSION

Based on performance testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed cNeuro cPET are assessed to be substantially equivalent to the predicate device.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.