K Number
K231476
Manufacturer
Date Cleared
2023-12-01

(193 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TRUENESS™ AIR Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™ AIR Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™ AIR Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ AIR Blood Glucose Monitoring System is comprised of the TRUENESS™ AIR blood glucose meter and the TRUENESS™ blood glucose test strip.

The TRUENESS™ Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™ Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™ Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ Blood Glucose Monitoring System is comprised of the TRUENESS™ blood glucose meter and the TRUENESS™ blood glucose test strip.

Device Description

The TRUENESS Blood Glucose Monitoring System and TRUENESS AIR Blood Glucose Monitoring System mainly consist of two parts as below:
(1) TRUENESS Blood Glucose Meter or TRUENESS AIR Blood Glucose Meter (With Bluetooth function)
(2) TRUENESS Blood Glucose Test Strip

A glucose test is based on measurement of electrical current caused by the reaction of glucose with flavin adenine dinucleotide (FAD) glucose dehydrogenase on the electrode of the test strip. The blood or control solution sample is drawn into the tip of the TRUENESS Blood Glucose Test Strip through capillary action. Glucose in the sample reacts with the FAD glucose dehydrogenase and generate electrons. The magnitude of the resultant current is proportional to the concentration of glucose in the blood and is converted to a glucose concentration. The glucose concentration is displayed on the meter display for the user.

AI/ML Overview

The provided text primarily focuses on the FDA's 510(k) clearance for the TRUENESS™ AIR Blood Glucose Monitoring System and the TRUENESS™ Blood Glucose Monitoring System. It outlines the regulatory process, device descriptions, and a high-level summary of testing. However, it does not contain the detailed acceptance criteria or the specific study results in the format requested.

The document indicates that "Design verification and validation testing consisted of performance tests (precision, linearity, interference, flex studies), electrical/mechanical safety tests, as well as disinfection, cleaning, and robustness studies." It also mentions "A user evaluation confirmed the system accuracy, operation according to design, and ease of use to support the intended use as described in the proposed labeling."

Therefore, I cannot populate the table or answer all your questions with the information given. I can, however, extract the general statement about accuracy.

Here's a breakdown of what can and cannot be answered based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
System AccuracyConfirmed by user evaluation
PrecisionTested
LinearityTested
InterferenceTested
Flex studiesTested
Electrical/Mechanical SafetyTested
DisinfectionTested
CleaningTested
RobustnessTested

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for any of the tests.
  • Data Provenance: Not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not specified. The document mentions a "user evaluation" confirmed system accuracy, but does not detail the nature or qualifications of these users as "experts" for establishing ground truth in a clinical sense.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not specified.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable/mentioned. This device is a blood glucose monitoring system, not an AI-assisted diagnostic tool that involves human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • The device performs quantitative measurement of glucose. The "system accuracy" would refer to its standalone performance, confirmed by user evaluation. The specifics of this standalone performance are not detailed in terms of metrics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not specified. For blood glucose monitoring systems, ground truth is typically established by comparing the device's readings against a laboratory reference method (e.g., YSI analyzer), but this is not explicitly stated in the document provided.

8. The sample size for the training set

  • Not applicable/mentioned. Blood glucose monitoring systems typically don't have a "training set" in the machine learning sense. Their performance is validated through defined chemical and user studies against reference methods.

9. How the ground truth for the training set was established

  • Not applicable/mentioned for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2023

Sinocare Inc. Lingjia Zhu Regulatory Affairs Engineer No.265 Guyuan Road, Hi-Tech Zone, Changsha, Hunan Province Changsha, Hunan 410205 China

Re: K231476

Trade/Device Name: TRUENESS™ AIR Blood Glucose Monitoring System; TRUENESS™ Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: May 22, 2023 Received: May 22, 2023

Dear Lingjia Zhu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joshua Balsam -S

Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231476

Device Name

TRUENESS™ AIR Blood Glucose Monitoring System

Indications for Use (Describe)

The TRUENESS™ AIR Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™ AIR Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™ AIR Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ AIR Blood Glucose Monitoring System is comprised of the TRUENESS™ AIR blood glucose meter and the TRUENESS™ blood glucose test strip.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231476

Device Name

TRUENESS™ Blood Glucose Monitoring System

Indications for Use (Describe)

The TRUENESS™ Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™ Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™ Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ Blood Glucose Monitoring System is comprised of the TRUENESS™ blood glucose meter and the TRUENESS™ blood glucose test strip.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Contact details

510(k) Number:K231476
Applicant Name:Sinocare Inc.
Applicant Name:No.265 Guyuan Road, Hi-tech Zone, Changsha,Hunan Province, 410205, China.
Applicant Contact Telephone:+86-18890033693
Applicant Contact:Candy.zhou@sinocare.com
Correspondent Name:Sinocare Inc.
Correspondent Address:No.265 Guyuan Road, Hi-tech Zone, Changsha,Hunan Province, 410205, China.
Correspondent Contact Telephone:+86-15573694818
Correspondent Contact:Lingjia Zhu
Correspondent Contact email:lingjia.zhu@sinocare.com
Applicant Contact email:Candy.zhou@sinocare.com
Prepared on:11/28/2023
    1. Device Name
Trade Name:TRUENESSTM AIR Blood Glucose MonitoringSystem;TRUENESSTM Blood Glucose Monitoring
Common Name:System Blood Glucose Monitoring System
Classification Name:System, Test, Blood Glucose, Over The Counter
Regulation Number:862.1345
Product Code:NBW

3. Legally Marketed Predicate Devices

Contour® next GEN Blood Glucose Monitoring System (K193407)

Product Code:

Page (1) of Page (4)

NBW

{5}------------------------------------------------

4. Device Description Summary

A Device description

The TRUENESS Blood Glucose Monitoring System and TRUENESS AIR Blood Glucose Monitoring System mainly consist of two parts as below:

  • (1) TRUENESS Blood Glucose Meter or TRUENESS AIR Blood Glucose Meter (With Bluetooth function)
  • (2) TRUENESS Blood Glucose Test Strip

B Principle of Operation

A glucose test is based on measurement of electrical current caused by the reaction of glucose with flavin adenine dinucleotide (FAD) glucose dehydrogenase on the electrode of the test strip. The blood or control solution sample is drawn into the tip of the TRUENESS Blood Glucose Test Strip through capillary action. Glucose in the sample reacts with the FAD glucose dehydrogenase and generate electrons. The magnitude of the resultant current is proportional to the concentration of glucose in the blood and is converted to a glucose concentration. The glucose concentration is displayed on the meter display for the user.

5. Intended Use/Indications For Use

The TRUENESS™ Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™ Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™ Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ Blood Glucose Monitoring System is comprised of the TRUENESS™ blood glucose meter and the TRUENESS™ blood glucose test strip.

The TRUENESS™AIR Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of their diabetes control program. The TRUENESS™AIR Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is for in vitro diagnostic use only. The TRUENESS™AIR Blood Glucose Monitoring System is not intended for the diagnosis of, or screening for diabetes. It is not intended for use on neonates.

The TRUENESS™ AIR Blood Glucose Monitoring System is comprised of the TRUENESS™ AIR blood glucose meter and the TRUENESS™ blood glucose test strip.

{6}------------------------------------------------

6. Indications for Use Comparison

The indications for use of the TRUENESS Blood Glucose Monitoring System/TRUENESS AIR Blood Glucose Monitoring System and predicate device are the same.

7. Technological Comparison

Predicate deviceSubject Devices
FeaturesContour® next GEN(K193407)TRUENESSTRUENESSAIR(with Bluetooth)
Specimen TypeCapillary whole bloodCapillary whole bloodCapillary whole blood
Where Device UsedHomeHomeHome
Reagent FormElectrochemical TestStripElectrochemical Test StripElectrochemical Test Strip
MethodologyElectrochemicalBiosensorElectrochemical BiosensorElectrochemical Biosensor
Storage Environment41°F -86°F41°F -86°F41°F -86°F
Test Time5 seconds5 seconds5 seconds
Units of Measuremg/dLmg/dLmg/dL
Reagent EnzymeFAD- GlucoseDehydrogenaseFAD- Glucose DehydrogenaseFAD- Glucose Dehydrogenase
Averages Calculation7/14/30/90-day averages7/14/30/90-day averages7/14/30/90-day averages
Test strip calibrationNo codingNo codingNo coding
BLE PairingYesYes (TRUENESS AIR only)Yes (TRUENESS AIR only)
WirelessCommunicationBluetooth Low EnergyBluetooth Low Energy(TRUENESS AIR only)Bluetooth Low Energy(TRUENESS AIR only)
OperatingEnvironment41°F-113°F (5°C-45°C)50°F-104°F (10°C-40°C)50°F-104°F (10°C-40°C)
Hematocrit Range20%-55%15%-60%15%-60%
Test ResultsPlasma/serum glucosecalibratedPlasma glucose calibratedPlasma glucose calibrated
Measurement Range20 – 600 mg/dL40 - 600 mg/dL40 - 600 mg/dL

{7}------------------------------------------------

Power SourceTwo 3-volt CR2032 orDL2032 coin cellbatteries2 AAA alkaline batteries
Sample Size0.6μL1 µL
AltitudeUp to 20,674 ftUp to 10,100 ft

8. Non-Clinical and/or Clinical Tests Summary & Conclusions

Design verification and validation testing consisted of performance tests (precision, linearity, interference, flex studies), electrical/mechanical safety tests, as well as disinfection, cleaning, and robustness studies. Software validation studies were performed for moderate level of concern device per FDA Guidance Content of Premarket Submissions for Software Contained in Medical Devices. Results from these studies show that the subject device met design specifications and requirements.

A user evaluation confirmed the system accuracy, operation according to design, and ease of use to support the intended use as described in the proposed labeling.

The conclusions from the nonclinical and clinical tests demonstrated that the device is as safe, as effective, and performs as well as legally marketed device identified above.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.