(190 days)
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
The 4x reprocessed Monarch Bronchoscope. MBR-000211-B. (hereafter referred to as "Proposed Device") is similar to the 2x reprocessed Monarch Bronchoscope, MBR-000211-B, (hereafter referred to as "Predicate Device").
The Original Monarch Bronchoscope. MBR-000211-A. (hereafter referred to as "Co-Predicate Device") cleared under K193534 is identical to the 2x reprocessed Monarch Bronchoscope ("Predicate Device"), cleared under K203614.
Both device designs are controlled by Auris Health, Inc., throughout their total product lifecycle.
The Monarch Bronchoscope is a component of, and must be used with, the Monarch Platform, cleared under K193534. The Monarch Bronchoscope is connected to the robotic arms of the Monarch Platform to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Proposed Device is a reprocessed single-use device that is either discarded or returned to Auris for reprocessing after clinical use. The original singleuse Monarch Bronchoscope was cleared for one clinical use under K193534. The reprocessed single-use (rSUD) Monarch Bronchoscope was cleared for two (2) clinical uses under K203614. This premarket notification intends to gain market clearance for a reprocessed single-use (rSUD) Monarch Bronchoscope that can be reprocessed up to three (3) times, for a total of four (4) clinical uses.
Identical to the Predicate Device, the Reprocessed Monarch Bronchoscope, consisting of the Inner Scope ("scope") and the Outer Sheath ("sheath"), has 4-way articulation controlled by continuous, direct, visual control of the physician using the Monarch Platform. The Proposed Device contains a working channel through which biopsy devices, or other working channel instruments, may be introduced. The distal tip of the Proposed Device has a camera control unit (CCU) that collects live images that are then transmitted to the physician's display interface of the Monarch Platform. The camera transmits vision data to the Monarch Tower through the camera cable. The single-use, manually controlled, working channel instruments compatible with the Proposed Device are identical to the working channel instruments compatible with the Predicate Device.
The provided text describes the regulatory clearance of a reprocessed medical device, the Monarch Bronchoscope, and its performance testing to demonstrate substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria for a diagnostic algorithm, nor does it detail a study that proves a device meets such criteria in terms of diagnostic performance (e.g., sensitivity, specificity).
The document is a 510(k) premarket notification letter from the FDA, confirming clearance for the "4x reprocessed Monarch Bronchoscope" (meaning it can be reprocessed up to three times for a total of four clinical uses). The performance testing described focuses on the physical and functional integrity of the reprocessed device, not on diagnostic accuracy of an algorithm or an AI component.
Therefore, many of the requested items (e.g., acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert qualifications, MRMC studies, standalone algorithm performance, type of ground truth) are not applicable or cannot be extracted from this document.
Here's an attempt to answer the questions based only on the provided text, highlighting where information is absent or not relevant to diagnostic algorithm validation:
1. A table of acceptance criteria and the reported device performance
The document provides a list of performance tests conducted for the reprocessed bronchoscope, with "Pass" as the result, indicating that the device met internal design specifications and applicable standards for safe and effective reprocessing. These are functional and safety-related criteria for the physical device, not diagnostic performance metrics.
| Test Name | Acceptance Criteria (Implicit: Met relevant standards/specifications) | Reported Device Performance |
|---|---|---|
| Cleaning Validation | Device can be returned to a like-new condition | Pass |
| Functional Performance and System Compatibility | Performance not negatively impacted by reprocessing | Pass |
| Electromagnetic Compatibility and Electrical Safety | Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-18 | Pass |
| Biocompatibility | Evaluates end-of-life biocompatibility per ISO 10993-1:2018 | Pass |
| Sterilization | Assessed per ISO 11135:2018; residuals per ISO 10993-7:2008 | Pass |
| Sterility Assurance Level (from Sterilization test) | Minimum of 10^-6 | Achieved (Implied by Pass) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The performance tests refer to "worst case device conditioning" but do not specify the number of devices or clinical cases used in these tests. The data provenance is not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable to the type of device and testing described. The document pertains to the reprocessing of a physical bronchoscope, not a diagnostic algorithm requiring expert ground truth for image interpretation or diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable to the type of device and testing described. Adjudication methods are typically relevant for human-in-the-loop or diagnostic studies, which are not detailed here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable to the type of device and testing described. The document does not mention an AI component or any MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable to the type of device and testing described. The Monarch Bronchoscope is a physical device; there is no mention of a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable in the context of diagnostic performance. For the safety and functional tests performed, the "ground truth" would be the established engineering and biological standards and specifications the device aims to meet (e.g., sterility, electrical safety, cleaning effectiveness).
8. The sample size for the training set
This information is not provided and is not applicable as there is no mention of an algorithm or machine learning model being developed or trained.
9. How the ground truth for the training set was established
This information is not applicable as there is no mention of algorithmic training or ground truth for such a purpose.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
November 28, 2023
Auris Health Inc., a Johnson and Johnson Family Company Patrick Garvey Director, Regulatory Affairs 150 Shoreline Drive Redwood City, California 94065
Re: K231473
Trade/Device Name: Monarch Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: QNW Dated: October 30, 2023 Received: October 31, 2023
Dear Patrick Garvey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K231473
Device Name Monarch Bronchoscope
Indications for Use (Describe)
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
General Information
| 510(k) Submitter | Auris Health, Inc.150 Shoreline DriveRedwood City, CA 94065USA |
|---|---|
| Original DeviceManufacturer | Auris Health, Inc.150 Shoreline DriveRedwood City, CA 94065USA |
| FDA Registration Number | 3014447948 |
| Reprocessed DeviceManufacturer | Sterilmed, Inc., a Johnson and Johnson FamilyCompany5010 Cheshire Pkwy N, Suite 2, Plymouth, MN55446 |
| FDA Registration Number | 3011276773 |
| Primary Correspondent | James Patrick Garvey IIDirector, Regulatory AffairsAuris Health, Inc. |
| Contact Information | Email: pgarvey@its.jnj.comPhone: (650) 590-6907 |
| Date Prepared | 28 November 2023 |
Device Identification
Proposed Reprocessed Device:
| Proprietary Name | Monarch Bronchoscope |
|---|---|
| Common Name | Reprocessed Bronchoscope |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 21 CFR 874.4680 |
| Product Code | QNW |
| Regulatory Class | II |
| Model Number | MBR-000211-B |
Primary Predicate Device:
| Proprietary Name | Monarch Bronchoscope |
|---|---|
| Common Name | Reprocessed Bronchoscope |
| Premarket Notification | K203614 |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 21 CFR 874.4680 |
| Product Code | QNW |
| Regulatory Class | II |
| Model Number | MBR-000211-B |
{4}------------------------------------------------
| Proprietary Name | Monarch Platform |
|---|---|
| Common Name | Bronchoscope (Flexible or Rigid) |
| Premarket Notification | K193534 |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 21 CFR 874.4680 |
| Product Code | EOQ |
| Regulatory Class | II |
| Model Number | MBR-000211-A |
Secondary Predicate OEM Device:
Device Description
The 4x reprocessed Monarch Bronchoscope. MBR-000211-B. (hereafter referred to as "Proposed Device") is similar to the 2x reprocessed Monarch Bronchoscope, MBR-000211-B, (hereafter referred to as "Predicate Device").
The Original Monarch Bronchoscope. MBR-000211-A. (hereafter referred to as "Co-Predicate Device") cleared under K193534 is identical to the 2x reprocessed Monarch Bronchoscope ("Predicate Device"), cleared under K203614.
Both device designs are controlled by Auris Health, Inc., throughout their total product lifecycle.
The Monarch Bronchoscope is a component of, and must be used with, the Monarch Platform, cleared under K193534. The Monarch Bronchoscope is connected to the robotic arms of the Monarch Platform to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Proposed Device is a reprocessed single-use device that is either discarded or returned to Auris for reprocessing after clinical use. The original singleuse Monarch Bronchoscope was cleared for one clinical use under K193534. The reprocessed single-use (rSUD) Monarch Bronchoscope was cleared for two (2) clinical uses under K203614. This premarket notification intends to gain market clearance for a reprocessed single-use (rSUD) Monarch Bronchoscope that can be reprocessed up to three (3) times, for a total of four (4) clinical uses.
Identical to the Predicate Device, the Reprocessed Monarch Bronchoscope, consisting of the Inner Scope ("scope") and the Outer Sheath ("sheath"), has 4-way articulation controlled by continuous, direct, visual control of the physician using the Monarch Platform. The Proposed Device contains a working channel through which biopsy devices, or other working channel instruments, may be introduced. The distal tip of the Proposed Device has a camera control unit (CCU) that collects live images that are then transmitted to the physician's display interface of the Monarch Platform. The camera transmits vision data to the Monarch Tower through the camera cable. The single-use, manually controlled, working channel instruments compatible with the Proposed Device are identical to the working channel instruments compatible with the Predicate Device.
{5}------------------------------------------------
Intended Use/Indications for Use
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Summary of Key Technological Characteristics
The Proposed Device is a single-use device similar to the Predicate Device in design. Both devices have the same clinical applications, patient population, performance specifications, and principles of operation. All key technological characteristics including articulation, vision, and compatibility with working channel instruments are identical. Key technological characteristics are listed in the table below.
| Key Attributes | Proposed Device (MBR-000211-B) |
|---|---|
| Product Code | QNW |
| Regulation Number | 21 CFR 874.4680 |
| Classification | II |
| Intended for Single Use | Yes |
| Field of View (FOV) in air | 90 degrees |
| Direction of view | 0 degrees |
| FOV depth | 3-30mm |
| Imaging type | CMOS Imager |
| Illumination type | LED |
| Active angulation degrees up/down or4 directions | 180/180/180/180 |
| Pixel resolution | 200 x 200 |
| Camera lens | Aluminosilicate glass |
| Light source | LED (covered in cyanoacrylate adhesive) |
| Working Channel InstrumentsCompatibility | Auris working channel instruments and third partyinstruments that meet working channel length anddiameter requirements (e.g., REBUS probe) |
This premarket notification is submitted to demonstrate the ability to reprocess the Monarch Bronchoscope three (3) times without impacting safety or effectiveness as compared to the predicate device. Reprocessing of the device includes removal of adherent soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging, labeling, and sterilization operations.
{6}------------------------------------------------
Performance Testing
The Reprocessed Monarch Bronchoscope was tested for performance in accordance with internal design specification and with the applicable performance standards to demonstrate safety and effectiveness to support substantial equivalence. This includes the following tests:
| Test Name | Description | Results |
|---|---|---|
| Cleaning Validation | Ensures the device can be returned to a like new condition.Worst case device conditioning was taken intoconsideration while conducting this testing. | Pass |
| FunctionalPerformance andSystemCompatibility | Functional performance and system compatibility wereperformed to verify the performance of the Proposed Devicewas not negatively impacted by reprocessing. Worst casedevice conditioning was taken into consideration whileconducting this testing. | Pass |
| ElectromagneticCompatibility andElectrical Safety | The Proposed Device has been fully evaluated for electricalsafety and EMC compliance to the following standards:AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-2-18. | Pass |
| Biocompatibility | Evaluates the end of life biocompatibility of the limitedcontact device in accordance with ISO 10993-1:2018. Worstcase device conditioning was taken into consideration whileconducting this testing. | Pass |
| Sterilization | Sterilization was assessed in accordance with ISO 11135:2018. Sterilization residuals were assessed in accordancewith ISO 10993-7:2008. The Proposed Device uses avalidated ethylene oxide sterilization process using a half-cycle overkill approach to achieve a minimum sterilityassurance level of 10-6. | Pass |
The performance testing demonstrates that the 4x reprocessed Monarch Bronchoscope is as safe and effective as the legally marketed 2x reprocessed Monarch Bronchoscope and operates as originally intended.
Conclusion
The indications for use/intended use, clinical applications, patient population, performance specifications, technological characteristics, biological safety, and principles of operation of the Proposed Device are identical to the Predicate Device.
Test results demonstrated substantial equivalence to the Predicate Device with respect to safety and effectiveness. We therefore conclude that the 4x reprocessed Monarch Bronchoscope is as safe, and effective as, and substantially equivalent to the Predicate Device, 2x reprocessed Monarch Bronchoscope.
{7}------------------------------------------------
The following device is included in this submission:
| Description | OEM Model Number | Reprocessed ModelNumber |
|---|---|---|
| Monarch Bronchoscope | MBR-000211-A | MBR-000211-B |
§ 874.4680 Bronchoscope (flexible or rigid) and accessories.
(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.