K Number
K231210
Device Name
DGA Abutment
Date Cleared
2024-03-05

(313 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DGA Abutment is a patient-specific prosthetic component directly connected to the endosseous dental implant (details as below) and is intended for use as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Device Description

DGA Abutment is endosseous implant abutment which is placed into a corresponding dental implant to provide support for a prosthetic restoration. DGA Abutment is made of titanium grade Ti-6Al-4V ELI (meets ASTM Standard F136). The DGA Abutment is mounted on the implant with a screw. Each abutment is supplied with two identical screws which are used for:
a. For fixing into the endosseous implant.
b. For dental laboratory use during fabrication of restoration.
The screw is also made of titanium grade Ti-6AI-4V ELI (meets ASTM Standard F136).

AI/ML Overview

The provided text is a 510(k) summary for the DGA Abutment, a dental implant component. It addresses substantial equivalence rather than presenting an AI/ML device study with acceptance criteria and performance data. Therefore, the document does not contain the information requested regarding acceptance criteria, study details, sample sizes, expert involvement, or MRMC studies for an AI/ML device.

The document discusses the mechanical properties and biocompatibility of the DGA Abutment, comparing it to predicate devices. It states that fatigue testing was conducted according to ISO 14801:2016 and FDA guidance, and biocompatibility testing was done as per ISO 10993 series. However, it does not provide specific acceptance criteria or reported performance data in a table format for these tests, nor does it refer to an AI/ML algorithm.

Therefore, I cannot directly answer your prompt using the provided text. The prompt's questions pertain to acceptance criteria and study methodologies for an AI/ML device, which are not found in this 510(k) submission for a physical dental abutment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 5, 2024

Modern Dental Laboratory (DG) Co., Ltd. Robin Liu Compliance Manager Room 102 & 1102 & Floor 4-10, Block 1 Modern Dental Industrial Park NO. 7 Nantou, Songshan Lake Hi-Tech Industrial Zone Dongguan, Guangdong 523000 CHINA

Re: K231210

Trade/Device Name: DGA Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 5, 2024 Received: February 5, 2024

Dear Robin Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K231210

Device Name

DGA Abutment

Indications for Use (Describe)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5. 510(k) Summary

5.1 Submitter Information

Company Name:Modern Dental Laboratory (DG) Co.,Ltd.
Company Address:Room 102&1102&Floor 4-10,Block 1 Modern DentalIndustrial Park, No.7 Nantou, Songshan Lake Hi-TechIndustrial Zone 523000 Dongguan,Guangdong People'sRepublic of China
Company Phone:0769-22899211
Contact Person:Robin Liu
Email:Robin.liu@moderndentallab.com
Date PreparedMarch 5,2024

5.2 Device Identification

Device Model Name:DGA Abutment
Classification Name:Endosseous Dental Implant Abutment
Regulation Number:872.3630
Product Code:NHA
ClassII
PanelDental

5.3 Predicated Devices

Primary Predicate Device:K213961
Reference Device:K151295, K142813

5.4 Device Description

DGA Abutment is endosseous implant abutment which is placed into a corresponding dental implant to provide support for a prosthetic restoration. DGA Abutment is made of titanium grade Ti-6Al-4V ELI (meets ASTM Standard F136). The DGA Abutment is mounted on the implant with a screw. Each abutment is supplied with two identical screws which are used for:

  • a. For fixing into the endosseous implant.
  • b. For dental laboratory use during fabrication of restoration.

The screw is also made of titanium grade Ti-6AI-4V ELI (meets ASTM Standard F136).

BrandSystemPlatformImplant body diameter510(k) number
Tissue Level(ST)RN4.1mm,4.8mmK122855
StraumannWN4.8mmK122855

DGA Abutment is compatible with the following implant

{5}------------------------------------------------

Abutment postheight (lengthabove collar/gingival height)AngulationAbutmentdiameterWall thicknessGingival height
4-9 mm(±0.2mm)0°(±0.1°)4-7 mm(±0.03mm)≥0.4mm(±0.05mm)0-7mm(±0.2mm)

The range of design parameters of DGA Abutment please refer to the following sheet.

5.5 Indication for Use

DGA Abutment is a patient-specific prosthetic component directly connected to the endosseous dental implant (details as below) and is intended for use as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

BrandSystemPlatformImplant body diameter510(k) number
StraumannTissue Level (ST)RN4.1mm,4.8mmK122855
WN4.8mmK122855

5.6 Technological Characteristics

  • . Performance Specification
    The DGA Abutment is designed to serve as a connection component between a dental implant and a dental prosthesis and to remain for indefinitely in the oral cavity. As such it must fulfil the following -

  • a. Made with a biocompatible material suitable for permanent mucosal contact.

  • Provide an aesthetic profile from the implant up to the dental prosthesis. b.

  • Provide sufficient strength to withstand masticatory forces. C.

  • . Mechanical Properties

The fatigue testing of the DGA Abutment was conducted in accordance with ISO 14801: 2016 Dentistry -- Implants -- Dynamic loading test for endosseous dental implants and FDA guidance " Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Implant Abutments", the finished assembled implant,abutment and screw systems were used to tested.

. Material Properties

The DGA Abutment is made of titanium grade Ti-6AI-4V which meets ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI Alloy for Surgical Implant Applications.To further confirm finished DGA Abutment's safety, some biocompatibility testing were conducted as per ISO 10993 series of standards, such as cytotoxicity, sensitization and irritation, these tests suggest that DGA Abutment has no cytotoxicity, no sensitization and no irritation and is safety for patient.

{6}------------------------------------------------

. Manufacturing Process

The DGA Abutment should be machined with computer numerical control equipment and tungsten carbide cutting tools ,and water soluble coolants should be employed during the machining process. Finished DGA Abutments should be thoroughly rinsed with clean water to remove the coolants.

DGA Abutment is delivered non-sterile for single use and must be cleaned and sterilized prio to use.The sterilization was carried out according to ISO 17665-1:2006. The recommended sterilization method is provided as below:

MethodMoist heat sterilization
CyclePre-vacuum
Temperature132°C (270° F)
Exposure time4 min
Drying Time20 min

● Mode of Action

The mechanism of action is exactly same with the primary predicate device and reference device, and supports a determination of substantial equivalence.The infrastructure of the DGA Abutment is connected with the implant using screw-retained, and the superstructure of the DGA Abutment is connected with the dental prosthesis by cemented.The DGA Abutment is surgically placed in the maxillary and mandibular arches for supporting the dental restorations to restore patient's chewing function.

. MR Compatibility

Non-clinical worst-case MRI review was performed to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental abutments, and fixation screws) and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque.

Reverse Engineering Analysis ●

Dimensional analysis and reverse engineering of the implant-to-abutment connection platform were performed, including an assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible OEM implant body, OEM abutment, and OEM fixation screw. The testing demonstrated implant to abutment compatibility and has established substantial equivalency of the proposed device with predicate

{7}------------------------------------------------

devices.Static and fatigue testing according to ISO 14801 were performed on worst case scenario. The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate.

5.7 Risk Management

The risks to health have been identified, identified risks generally associated with the use of the DGA Abutment. The identifying approach of risk is specified as per the standard ISO 14971: 2019 Medical devices - Application of risk management to medical deices and ISO TR 24971: 2020 Medical devices - Guidance on the application of ISO 14971.

{8}------------------------------------------------

5.8 Substantial Equivalence Comparison

ElementsSubject DevicePrimary Predicate DeviceReference DeviceReference DeviceExplanation
NameDGA AbutmentTruAbutment DSBiodenta CustomisedAbutment -TitaniumDuplex Abutment
510(k)NumberNew deviceK213961K151295K142813
ManufacturerModern DentalLaboratory (DG) Co.,Ltd.TruAbutment Inc.Biodenta Swiss AGBiogenesis Co., Ltd.
RegulationNumber21 CFR 872.363021 CFR 872.363021 CFR 872.363021 CFR 872.3630Same
Classification NameEndosseous DentalImplant AbutmentEndosseous Dental ImplantAbutmentEndosseous DentalImplant AbutmentEndosseous Dental ImplantAbutmentSame
ProductCodeNHANHANHANHASame
DeviceClassClass IIClass IIClass IIClass IISame
Indicationfor UseDGA Abutment is apatient-specificprosthetic componentdirectly connected tothe endosseous dentalimplant (details asbelow) and is intendedfor use as a support forTruAbutment DS is apatient-specific CAD/CAMabutment, directly connected toendosseous dental implants andis intended for use as an aid inprosthetic rehabilitation. It iscompatible with the followingThe BiodentaCustomized Abutment-Titanium is intended foruse with dentalimplants as a supportfor single or multipletooth prostheses in theThe Biogenesis ImplantSystem - Kisses is indicatedfor use in partially or fullyedentulous mandibles andmaxillae, in support of singleor multiple -unit restorationsincluding; cemented retained,Same. Slightlydifference indescription, howeverthe content issubstantialequivalence.
single or multiple toothprostheses in themaxilla or mandible of apartially or fullyedentulous patient.Brand: StraumannSystem: Tissue Level(ST)Platform & Implant bodydiameter: RN (4.1mm,4.8mm), WN (4.8mm)510(k) number:K122855systems:• Straumann Tissue Level Implant(K122855, K202942): 4.1(RN),4.8(RN), 6.5(WN) mmmaxilla or mandible of apartially or fullyedentulous patient.screw retained,oroverdenture restorations, andterminal or intermediateabutment support for fixedbridgework. The BiogenesisImplant System - Kisses is forsingle and two stage surgicalprocedures.It is for delayedloadingThe material of subjectdevice is same as thatof predicate deviceand reference deviceK151295.
Abutmentand ScrewmaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELITi Gr.4
AbutmentfixationScrew-retainedScrew-retainedScrew-retainedScrew-retainedSame
AbutmentAngulation0°(±0.1°)Max 25°Max 30°straightthe subject's allowedangulation is smallerthan that of predicate/reference( K151295)device, and is same asthat of referencedevice K142813.
AbutmentDiameter4-7 mm(±0.03mm)5.2-8.0 mm4.5-15 mm4.0-6.5mmBelong to patient -specific slightly
difference in
parameter in diameter,however it doesn'taffect the substantialequivalence.
AbutmentPost Height(length above collar/gingival height)4-9 mm(±0.2mm)4.0-7.0 mm5.0 - 12.3 mm4.0-7.0mmBelong to patient-specific,slightlydifference inparameter in height,however it doesn'taffect the substantialequivalence.
WallthicknessMin 0.4mm(±0.05mm)Min 0.4mmUnknownunknownSame as predicatedevice
Belong to patient-specific,slightly
Gingivalheight0-7mm(±0.2mm)UnknownUnknown1.0-7.0mmdifference inparameter in height,however it doesn'taffect the substantialequivalence.
Installationprocedure1. The final DGAAbutment is attached tothe implant using thecorresponding dentalscrew.2. The dental screw isa. After osseointegration hasbeen successfully achieved forthe fixture, design an appropriateCAD/CAM custom abutmentsuitable to the oral environment.b. Reproduce the configurationAn abutment screwprovided by Biodenta isused to secure theabutment to theimplant, whereas theprosthesis is customIt's not available from publicdatabase.Subject device issubstantialequivalence to primarypredicate andreference device(K151295) . Slightly
tightened in theof the oral cavity of the patientmade anddifference in
recommended torquethrough impression and send thecement-retained on thedescription, however
3. In order to obtain theimpression to a dental laboratoryabutment. Thethe content is
recommended torque, ato fabricate the final prosthesis.abutment screw shouldsubstantial
dental torque wrenchOnce the final prosthesisbe tightened with aequivalence.
with a suitable screwfabrication is completed, deliver itmanual torque wrench
driver must be used into the patient to improve theiraccording to the
accordance with themasticating and aestheticimplant manufacturer's
relevant manufacturer'sfunctions.specifications.
instructionsWhen affixing an abutment
4. The screw channelonto a fixture, the recommended
must always be sealedtorque value is equivalent to the
after the abutment ismanufacturers' value for the
attached to the implant.respective implant system. The
torque value should be
determined based on the clinical
assessment of the bone quality,
fixture dimensions, and
prosthesis type, etc.
CAD/CAMProcessingMilled at ModernDental Laboratoryunder QSR control.All digitally designed abutmentsand/or coping for use with theTruAbutment DS abutments areintended to be sent to aTruAbutment-validated millingcenter for manufacture.Milled in Biodentamilling center underQSR control.It's not available from publicdatabase.Subject device issubstantialequivalence to primarypredicate andreference device(K151295)
CompatibleImplantStraumann TissueStraumann Tissue Level ImplantDentsply, AstraTechOsseoSpeedBiogenesis Implant System -KissesSubject device andprimary predicate
SystemLevel Implant(K122855, K202942): 4.1(RN),Biomet 3i, Certaindevice both are
(K122855): 4.1(RN),4.8(RN), 6.5(WN) mmInternalcompatible with
4.8(RN) ,4.8(WN) mmZimmer, Screw VentStraumann Tissue
Nobel Biocare,NobelReplaceNobel Biocare,NobelActiveStraumann, Bone LevelBioHorizons,InternalOsstem, GS/TSBiodenta, Bone Leveland TaperedLevel Implant
SterilizationNon-sterileNon-sterileNon-SterileSame
Non-Sterile

Page 5 of 10

{9}------------------------------------------------

Page 6 of 10

{10}------------------------------------------------

Page 7 of 10

{11}------------------------------------------------

Page 8 of 10

{12}------------------------------------------------

{13}------------------------------------------------

5.9 Substantial Equivalence Conclusion

The DGA Abutment is substantially equivalent to the primary predicate device and the reference devices in terms of indication for use, technical characteristics and function. Although there are slightly differences in size parameter of devices, these differences are very minor not affecting substantial equivalence.

Therefore, the subject device, the predicate device and the reference devices are substantial equivalence.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)