(448 days)
Reprocessed Cordis Catheters are designed to deliver radiopaque contrast medium to selected sites in the vascular system.
The reprocessed Cordis Catheters are available in a broad variety of French sizes and configurations. These catheters combine an atraumatic tip with a braided body.
The provided text describes the 510(k) summary for Reprocessed Cordis Super Torque and Super Torque Plus Diagnostic Angiographic Catheters. This document is a premarket notification for a medical device and thus does not contain information about studies related to AI or ground truth as it pertains to AI/ML models. Instead, it focuses on demonstrating substantial equivalence to predicate devices through functional and safety testing.
Here's the relevant information based on the provided text, adapted for the requested format where applicable, and noting where information is not present:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Tests Conducted | Reported Device Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity, Hemocompatibility, Sensitization, Irritation, Systemic Toxicity | Not explicitly detailed in terms of specific numeric results. The document states "Bench and laboratory testing was conducted to demonstrate substantially equivalent performance." Implies that the reprocessed device met established biocompatibility standards. |
| Cleaning Validation | (Implicit based on reprocessing nature) | Not explicitly detailed in terms of specific numeric results. The document states "reprocessing of these devices includes removal of visible soil and decontamination." Implies validation demonstrated effective cleaning to specified standards. |
| Sterilization Validation | (Implicit based on reprocessing nature) | Not explicitly detailed in terms of specific numeric results. Implies validation demonstrated effective sterilization to specified standards. |
| Functional Testing | Visual Inspection, Dimensional Verification, Radiopacity Testing, Simulated Use Testing, Leak Testing, Torsion Testing, Tip Buckling Testing, Tensile Testing, Coating Coverage, Particulate Testing | Not explicitly detailed in terms of specific numeric results. The document states "Bench and laboratory testing was conducted to demonstrate substantially equivalent performance." Implies the reprocessed device met established functional performance standards in these areas. |
| Packaging Validation | (Implicit) | Not explicitly detailed in terms of specific numeric results. Implies validation demonstrated effective packaging to maintain sterility and integrity. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each of the listed functional and safety tests. It generally refers to "Bench and laboratory testing." It is a regulatory submission for a reprocessed medical device, not an AI/ML study. Therefore, the concept of "test set" in the context of AI/ML models, data provenance (country of origin), and retrospective/prospective data collection is not applicable here. The testing would have been performed on samples of the reprocessed devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable. The device is a physical medical device (angiographic catheter), and "ground truth" as it relates to expert-labeled data for an AI/ML model is not relevant to this submission. The "ground truth" in this context would be the established performance specifications for diagnostic angiographic catheters, and the testing validates that the reprocessed device meets those specifications.
4. Adjudication Method for the Test Set
This information is not applicable, as it pertains to expert consensus on AI/ML model outputs, which is not relevant for this device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable as this is a physical medical device (angiographic catheter) and does not involve AI or human readers evaluating images or data.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable as the device is a physical medical device and does not involve an algorithm or AI.
7. The Type of Ground Truth Used
The "ground truth" for this device's acceptance criteria would be the established performance specifications and safety standards for diagnostic angiographic catheters as defined by relevant industry standards (e.g., ISO, ASTM) and FDA guidance for these types of devices and for reprocessed devices. This includes physical properties, biocompatibility, sterility, and functional performance. It is based on engineering specifications and scientific testing standards, not expert consensus on medical images or pathology.
8. The Sample Size for the Training Set
This information is not applicable. This is not an AI/ML model; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as there is no "training set" for this physical medical device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 1, 2024
Innovative Health, LLC. Amanda Babcock Regulatory Affairs Manager 1435 North Hayden Road Suite 100 Scottsdale, Arizona 85257
Re: K231015
Trade/Device Name: Reprocessed Cordis Super Torque and Super Torque Plus Diagnostic Angiographic Catheters Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: NLI Dated: April 7, 2023 Received: April 10, 2023
Dear Amanda Babcock:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Ariel Ariel G. Ash-G. Ash-shakoor -S Date: 2024.07.01 09:21:23 shakoor -S -04'00'
For
Gregory O'Connell
{2}------------------------------------------------
Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K231015
Device Name
Reprocessed Cordis Super Torque and Super Torque Plus Angiographic Diagnostic Catheters
Indications for Use (Describe)
Reprocessed Cordis Catheters are designed to deliver radiopaque contrast medium to selected sites in the vascular system.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) SUMMARY
Submitter's Name and Address:
Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257
Contact Name and Information:
Meerna Muradvich Regulatory Affairs Engineer Innovative Health, LLC. (480) 692-7176 (office) (888) 965-7706 (fax) mmuradvich@innovative-health.com
Amanda Babcock Regulatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7706 (fax) ababcock@innovative-health.com
Date prepared:
06/26/2024
Device Information:
| Trade/Proprietary Name: | Reprocessed Cordis Super Torque and SuperTorque Plus Diagnostic Angiographic Catheters |
|---|---|
| Common Name: | Angiographic Catheter |
| Classification Name: | Catheter, Angiography, Reprocessed |
| Classification Number: | Class II, 21 CFR 870.1200 |
| Product Code: | NLI |
Predicate Device:
| 510(k) Number | Device | Manufacturer |
|---|---|---|
| K915836 | Cordis 5.0 French Super Torque Catheters | Cordis Corp. |
| K914007 | Cordis 5.2 French Super Torque Plus | Cordis Corp. |
Device Description:
The reprocessed Cordis Catheters are available in a broad variety of French sizes and configurations. These catheters combine an atraumatic tip with a braided body.
The item numbers included in the scope of this submission are as follows:
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
|---|---|---|---|---|---|
| 532-410T | 4 | 65 | 8 | PIG | |
| 532-420T | 4 | 65 | 8 | straight | |
| 532-462 | 4 | 65 | 2 | RDCS | |
| Description | Item Number | French Size | Length | Side holes | Curve |
| ReprocessedCordis SuperTorqueDiagnosticAngiographicCatheter | 532-444 | 4 | 65 | 0 | C3 |
| 532-445 | 4 | 65 | 2 | C3 | |
| 532-473 | 4 | 65 | 0 | SHK 1 | |
| 532-474 | 4 | 65 | 0 | SHK 2 | |
| 532-447 | 4 | 65 | 0 | JC 1 | |
| 532-448 | 4 | 65 | 0 | JC 2 | |
| 532-449 | 4 | 65 | 0 | JC 3 | |
| SR3740 | 4 | 80 | 0 | JL 1.5 | |
| SRD6054 | 4 | 80 | 0 | JL 2 | |
| SR4533 | 4 | 80 | 0 | JL 2.5 | |
| SRD6920 | 4 | 80 | 4 | PIG 145° | |
| 532-432 | 4 | 80 | 2 | multipurpose small | |
| 532-478 | 4 | 80 | 2 | RDCS | |
| 532-411T | 4 | 90 | 8 | PIG | |
| 532-421T | 4 | 90 | 8 | straight | |
| SRD5816 | 4 | 100 | 0 | IM-MOD-5 | |
| 532-412T | 4 | 100 | 8 | PIG | |
| 532-422T | 4 | 100 | 8 | straight | |
| 532-430 | 4 | 100 | 0 | multipurpose small | |
| 532-482 | 4 | 100 | 2 | multipurpose small | |
| 532-414 | 4 | 100 | 0 | SIM 1 | |
| 532-415 | 4 | 100 | 0 | SIM 2 | |
| 532-461 | 4 | 100 | 0 | H1 | |
| 532-470 | 4 | 100 | 0 | MAN | |
| 532-405 | 4 | 100 | 0 | BERN | |
| 532-436 | 4 | 100 | 0 | JB1 | |
| 532-437 | 4 | 100 | 0 | JB2 | |
| 532-438 | 4 | 100 | 0 | JB3 | |
| 532-497 | 4 | 100 | 0 | VERT | |
| 532-413T | 4 | 110 | 8 | PIG | |
| 532-524V8 | 5 | 65 | 8 | PIG | |
| 532-564V8 | 5 | 65 | 8 | straight | |
| 532-576 | 5 | 65 | 0 | multipurpose (subintimalrecanalization) | |
| 532-510 | 5 | 65 | 2 | RDCS | |
| 532-514 | 5 | 65 | 0 | C3 | |
| 532-517 | 5 | 65 | 2 | C3 | |
| 532-519 | 5 | 65 | 0 | SHK 1 | |
| 532-522 | 5 | 65 | 0 | JC 2 | |
| 532-523 | 5 | 65 | 0 | JC 3 | |
| 532-579 | 5 | 80 | 0 | MPA | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| 532-511 | 5 | 80 | 2 | RDCS | |
| 532-509 | 5 | 80 | 0 | RDCA | |
| 532-567 | 5 | 80 | 0 | SHK 2 | |
| 532-539F8 | 5 | 90 | 8 | PIG | |
| 532-555H8 | 5 | 100 | 8 | PIG | |
| 532-569H8 | 5 | 100 | 8 | straight | |
| 532-578 | 5 | 100 | 0 | MPA | |
| 532-501 | 5 | 100 | 0 | SIM 1 | |
| 532-546 | 5 | 100 | 2 | SIM 1 | |
| 532-502 | 5 | 100 | 0 | SIM 2 | |
| 532-547 | 5 | 100 | 2 | SIM 2 | |
| 532-503 | 5 | 100 | 0 | SIM 3 | |
| 532-504 | 5 | 100 | 0 | H1 | |
| 532-520 | 5 | 100 | 0 | HN4 | |
| 532-571 | 5 | 100 | 0 | MAN | |
| Reprocessed | 532-541H0 | 5 | 100 | 0 | JB1 |
| Cordis Super | 532-543H0 | 5 | 100 | 0 | JB2 |
| Torque | 532-549H0 | 5 | 100 | 0 | VERT |
| Diagnostic | 455-610T | 6 | 65 | 12 | PIG |
| AngiographicCatheter | 455-672 | 6 | 65 | 0 | C3 |
| 455-689 | 6 | 80 | 2 | RDCA | |
| 455-698 | 6 | 80 | 4 | NIH | |
| 532-506 | 6 | 80 | 2 | multipurpose small | |
| 455-623 | 6 | 80 | 0 | MPA | |
| 455-636 | 6 | 80 | 2 | MPA | |
| 455-686 | 6 | 80 | 0 | RDCA | |
| 455-611T | 6 | 90 | 12 | PIG | |
| 532-618 | 6 | 100 | 0 | JL 3.5 | |
| 532-620 | 6 | 100 | 0 | JL 4 | |
| 532-627 | 6 | 100 | 0 | JL 4.5 | |
| 532-622 | 6 | 100 | 0 | JL 5 | |
| 532-624 | 6 | 100 | 0 | JL 6 | |
| 532-645 | 6 | 100 | 0 | AL 1 | |
| 532-646 | 6 | 100 | 0 | AL 2 | |
| 532-647 | 6 | 100 | 0 | AL 3 | |
| 532-619 | 6 | 100 | 0 | JR 3.5 | |
| 532-621 | 6 | 100 | 0 | JR 4 | |
| 532-623 | 6 | 100 | 0 | JR 5 | |
| 532-625 | 6 | 100 | 0 | JR 6 | |
| 532-648 | 6 | 100 | 0 | AR MOD | |
| 532-642 | 6 | 100 | 2 | MPA 2 | |
| Description | ltemNumber | FrenchSize | Length | Sideholes | Curve |
| 532-649 | ర్ | 100 | 2 | MPB 2 | |
| 532-640 | 6 | 100 | 0 | SON 1 | |
| 532-664 | 6 | 100 | 4 | SON 2 | |
| 532-672 | 6 | 100 | 0 | LCB | |
| 532-670 | ರಿ | 100 | 0 | RCB | |
| 532-660 | 6 | 100 | 0 | IM | |
| Reprocessed | 455-637 | 6 | 100 | 2 | MPA |
| Cordis Super | 455-660 | 6 | 100 | 0 | SIM 1 |
| Torque | 455-660D | 6 | 100 | 2 | SIM 1 |
| Diagnostic | 455-661 | 6 | 100 | 0 | SIM 2 |
| Angiographic | 455-662 | 6 | 100 | 0 | ടിM 3 |
| Catheter | 455-663 | 6 | 100 | 0 | SIM 4 |
| 455-665 | 6 | 100 | 0 | H1 | |
| 455-666 | 6 | 100 | 0 | H1 | |
| 532-650S | 6 | 110 | 6 | PIG | |
| 532-652S | 6 | 110 | 6 | PIG 145° | |
| 532-654S | ರಿ | 110 | 6 | PIG 145° MOD | |
| 455-613E | රි | 110 | 8 | PIG | |
| Reprocessed | 533-562 | 5.2 | 65 | 2 | MPA 2 |
| Cordis Super | 533-579 | 5.2 | 80 | 0 | MPA |
| Torque PlusDiagnostic | SR3216 | 5.2 | 100 | 0 | JL 3 |
| Angiographic | 533-551 | 5.2 | 100 | 0 | JL 3.5 |
| Catheter | 233-223 | 5.2 | 100 | 0 | JL 4 |
| 533-527 | 5.2 | 100 | 0 | JL 4.5 | |
| 233-229 | 5.2 | 100 | 0 | ി 5 | |
| 533-561 | 5.2 | 100 | 0 | JL 6 | |
| 533-542 | 5.2 | 100 | 0 | AL 2 | |
| 533-544 | 5.2 | 100 | O | AL 3 | |
| SR4615 | 5.2 | 100 | O | XB 3.5 ST | |
| 233-250 | 5.2 | 100 | O | JR 3.5 | |
| 533-552 | 5.2 | 100 | O | JR 4 | |
| 533-528 | 5.2 | 100 | 0 | JR 4 MOD | |
| 533-563 | 5.2 | 100 | 0 | JR 4 ST | |
| 533-558 | 5.2 | 100 | 0 | JR 5 | |
| 533-560 | 5.2 | 100 | O | JR 6 | |
| SR1924 | 5.2 | 100 | O | MPA | |
| 533-582 | 5.2 | 100 | 2 | MPA 2 | |
| 533-584 | 5.2 | 100 | O | CAS 1 | |
| 533-585 | 5.2 | 100 | 0 | CAS 2 | |
| 59950072 | 5.2 | 100 | 2 | RBL-JK | |
| 599500T1 | 5.2 | 100 | 1 | RBL-TG | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| 533-572 | 5.2 | 100 | 0 | LCB | |
| 533-570 | 5.2 | 100 | 0 | RCB | |
| 533-580 | 5.2 | 100 | 0 | IM | |
| SR4658 | 5.2 | 100 | 0 | IM-MOD | |
| SR4268 | 5.2 | 100 | 0 | IM-MOD-2 | |
| SR4685 | 5.2 | 100 | 0 | IM-MOD-3 | |
| 533-533 | 5.2 | 110 | 6 | PIG | |
| 533-534A | 5.2 | 110 | 6 | PIG 145° | |
| ReprocessedCordis SuperTorque PlusDiagnostic | 533-537S | 5.2 | 110 | 6 | PIG 145° MOD |
| 533-533A | 5.2 | 110 | 6 | PIG 155° | |
| 533-633 | 6 | 80 | 0 | MPA 1 | |
| 533-629 | 6 | 80 | 2 | MPA 2 | |
| 533-682 | 6 | 80 | 4 | CAS 2 | |
| 533-635 | 6 | 80 | 0 | NIH | |
| 533-618 | 6 | 100 | 0 | JL 3.5 | |
| 533-620 | 6 | 100 | 0 | JL 4 | |
| 533-627 | 6 | 100 | 0 | JL 4.5 | |
| 533-622 | 6 | 100 | 0 | JL 5 | |
| 533-624 | 6 | 100 | 0 | JL 6 | |
| 533-645 | 6 | 100 | 0 | AL 1 | |
| 533-646 | 6 | 100 | 0 | AL 2 | |
| 533-647 | 6 | 100 | 0 | AL 3 | |
| 533-619 | 6 | 100 | 0 | JR 3.5 | |
| 533-621 | 6 | 100 | 2 | JR 3.5 | |
| 533-687 | 6 | 100 | 0 | JR 4 CLASSIC | |
| 533-689 | 6 | 100 | 0 | JR 4 RECESSEDBRAIDING | |
| 533-628 | 6 | 100 | 0 | JR 4 MOD | |
| 533-626 | 6 | 100 | 0 | JR 4 ST | |
| 533-623 | 6 | 100 | 0 | JR 5 | |
| 533-625 | 6 | 100 | 0 | JR 6 | |
| 533-648 | 6 | 100 | 0 | AR MOD | |
| 533-641 | 6 | 100 | 0 | AR 1 MOD | |
| 533-643 | 6 | 100 | 0 | AR 2 MOD | |
| 533-640 | 6 | 100 | 0 | MPA 1 COURNAD | |
| 533-642 | 6 | 100 | 2 | MPA 2 | |
| 533-649 | 6 | 100 | 2 | MPB 2 | |
| 533-634 | 6 | 100 | 6 | MPB 3 | |
| 533-630 | 6 | 100 | 4 | SON 1 | |
| 533-631 | 6 | 100 | 0 | SON 2 | |
| 533-632 | 6 | 100 | 4 | SON 3 | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| ReprocessedCordis SuperTorque PlusDiagnosticAngiographicCatheter | 533-684 | 6 | 100 | 0 | CAS 1 |
| 533-685 | 6 | 100 | 0 | CAS 2 | |
| 533-686 | 6 | 100 | 0 | CAS 3 | |
| 599600R401 | 6 | 100 | 1 | RBL 4.0 | |
| 599600R451 | 6 | 100 | 1 | RBL 4.5 | |
| 599600R501 | 6 | 100 | 1 | RBL 5.0 | |
| 599600J2 | 6 | 100 | 2 | RBL-JK | |
| 599600T1 | 6 | 100 | 1 | RBL-TG | |
| 533-637 | 6 | 100 | 0 | EGB 1 | |
| 533-638 | 6 | 100 | 0 | EGB 2 | |
| 533-672 | 6 | 100 | 0 | LCB | |
| 533-670 | 6 | 100 | 0 | RCB | |
| 533-660 | 6 | 100 | 0 | IM | |
| 533-636 | 6 | 100 | 0 | NIH | |
| 533-650F | 6 | 110 | 4 | PIG | |
| 533-650S | 6 | 110 | 6 | PIG | |
| 533-650E | 6 | 110 | 8 | PIG | |
| 533-652S | 6 | 110 | 6 | PIG 145° | |
| 533-653S | 6 | 110 | 6 | PIG 145° MOD | |
| 533-654S | 6 | 110 | 6 | PIG 155° | |
| 533-655S | 6 | 110 | 6 | PIG 155° MOD |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Table 1: Device Scope
Indications for Use:
Reprocessed Cordis Catheters are designed to deliver radiopaque contrast medium to selected sites in the vascular system.
Technological Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Cordis Super Torque and Super Torque Plus diagnostic angiographic catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate substantially equivalent performance of the Reprocessed Cordis Super Torque and Super Torque Plus Catheters. This included the following:
{10}------------------------------------------------
- Biocompatibility ●
- Cytotoxicity .
- Hemocompatibility
- Sensitization
- . Irritation
- . Systemic Toxicitv
- Cleaning Validation ●
- Sterilization Validation
- Functional testing ●
- Visual Inspection .
- . Dimensional Verification
- l Radiopacity Testing
- l Simulated Use Testing
- . Leak Testing
- I Torsion Testing
- Tip Buckling Testing I
- 트 Tensile Testing
- . Coating Coverage
- . Particulate Testing
- . Packaging Validation
The Reprocessed Cordis Super Torque and Super Torque Plus Catheters are reprocessed no more than one (1) time. Each device is marked, serialized and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
Conclusion:
Innovative Health concludes that the Reprocessed Cordis Super Torque and Super Torque Plus Diagnostic Angiographic catheters do not raise new questions of safety and effectiveness and supports substantial equivalence to the predicate devices described herein.
{11}------------------------------------------------
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
|---|---|---|---|---|---|
| 532-410T | 4 | 65 | 8 | PIG | |
| 532-420T | 4 | 65 | 8 | straight | |
| 532-462 | 4 | 65 | 2 | RDCS | |
| 532-444 | 4 | 65 | 0 | C3 | |
| 532-445 | 4 | 65 | 2 | C3 | |
| 532-473 | 4 | 65 | 0 | SHK 1 | |
| 532-474 | 4 | 65 | 0 | SHK 2 | |
| 532-447 | 4 | 65 | 0 | JC 1 | |
| 532-448 | 4 | 65 | 0 | JC 2 | |
| 532-449 | 4 | 65 | 0 | JC 3 | |
| SR3740 | 4 | 80 | 0 | JL 1.5 | |
| SRD6054 | 4 | 80 | 0 | JL 2 | |
| SR4533 | 4 | 80 | 0 | JL 2.5 | |
| SRD6920 | 4 | 80 | 4 | PIG 145° | |
| 532-432 | 4 | 80 | 2 | multipurpose small | |
| 532-478 | 4 | 80 | 2 | RDCS | |
| 532-411T | 4 | 90 | 8 | PIG | |
| 532-421T | 4 | 90 | 8 | straight | |
| ReprocessedCordis SuperTorqueDiagnosticAngiographicCatheter | SRD5816 | 4 | 100 | 0 | IM-MOD-5 |
| 532-412T | 4 | 100 | 8 | PIG | |
| 532-422T | 4 | 100 | 8 | straight | |
| 532-430 | 4 | 100 | 0 | multipurpose small | |
| 532-482 | 4 | 100 | 2 | multipurpose small | |
| 532-414 | 4 | 100 | 0 | SIM 1 | |
| 532-415 | 4 | 100 | 0 | SIM 2 | |
| 532-461 | 4 | 100 | 0 | H1 | |
| 532-470 | 4 | 100 | 0 | MAN | |
| 532-405 | 4 | 100 | 0 | BERN | |
| 532-436 | 4 | 100 | 0 | JB1 | |
| 532-437 | 4 | 100 | 0 | JB2 | |
| 532-438 | 4 | 100 | 0 | JB3 | |
| 532-497 | 4 | 100 | 0 | VERT | |
| 532-413T | 4 | 110 | 8 | PIG | |
| 532-524V8 | 5 | 65 | 8 | PIG | |
| 532-564V8 | 5 | 65 | 8 | straight | |
| 532-576 | 5 | 65 | 0 | multipurpose (subintimalrecanalization) | |
| 532-510 | 5 | 65 | 2 | RDCS | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| 532-514 | 5 | 65 | 0 | C3 | |
| 532-517 | 5 | 65 | 2 | C3 | |
| 532-519 | 5 | 65 | 0 | SHK 1 | |
| 532-522 | 5 | 65 | 0 | JC 2 | |
| 532-523 | 5 | 65 | 0 | JC 3 | |
| 532-579 | 5 | 80 | 0 | MPA | |
| 532-511 | 5 | 80 | 2 | RDCS | |
| 532-509 | 5 | 80 | 0 | RDCA | |
| 532-567 | 5 | 80 | 0 | SHK 2 | |
| 532-539F8 | 5 | 90 | 8 | PIG | |
| 532-555H8 | 5 | 100 | 8 | PIG | |
| 532-569H8 | 5 | 100 | 8 | straight | |
| 532-578 | 5 | 100 | 0 | MPA | |
| 532-501 | 5 | 100 | 0 | SIM 1 | |
| 532-546 | 5 | 100 | 2 | SIM 1 | |
| 532-502 | 5 | 100 | 0 | SIM 2 | |
| 532-547 | 5 | 100 | 2 | SIM 2 | |
| 532-503 | 5 | 100 | 0 | SIM 3 | |
| 532-504 | 5 | 100 | 0 | H1 | |
| 532-520 | 5 | 100 | 0 | HN4 | |
| ReprocessedCordis SuperTorqueDiagnosticAngiographicCatheter | 532-571 | 5 | 100 | 0 | MAN |
| 532-541H0 | 5 | 100 | 0 | JB1 | |
| 532-543H0 | 5 | 100 | 0 | JB2 | |
| 532-549H0 | 5 | 100 | 0 | VERT | |
| 455-610T | 6 | 65 | 12 | PIG | |
| 455-672 | 6 | 65 | 0 | C3 | |
| 455-689 | 6 | 80 | 2 | RDCA | |
| 455-698 | 6 | 80 | 4 | NIH | |
| 532-506 | 6 | 80 | 2 | multipurpose small | |
| 455-623 | 6 | 80 | 0 | MPA | |
| 455-636 | 6 | 80 | 2 | MPA | |
| 455-686 | 6 | 80 | 0 | RDCA | |
| 455-611T | 6 | 90 | 12 | PIG | |
| 532-618 | 6 | 100 | 0 | JL 3.5 | |
| 532-620 | 6 | 100 | 0 | JL 4 | |
| 532-627 | 6 | 100 | 0 | JL 4.5 | |
| 532-622 | 6 | 100 | 0 | JL 5 | |
| 532-624 | 6 | 100 | 0 | JL 6 | |
| 532-645 | 6 | 100 | 0 | AL 1 | |
| 532-646 | 6 | 100 | 0 | AL 2 | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| ReprocessedCordis SuperTorqueDiagnosticAngiographicCatheter | 532-647 | 6 | 100 | 0 | AL 3 |
| 532-619 | 6 | 100 | 0 | JR 3.5 | |
| 532-621 | 6 | 100 | 0 | JR 4 | |
| 532-623 | 6 | 100 | 0 | JR 5 | |
| 532-625 | 6 | 100 | 0 | JR 6 | |
| 532-648 | 6 | 100 | 0 | AR MOD | |
| 532-642 | 6 | 100 | 2 | MPA 2 | |
| 532-649 | 6 | 100 | 2 | MPB 2 | |
| 532-640 | 6 | 100 | 0 | SON 1 | |
| 532-664 | 6 | 100 | 4 | SON 2 | |
| 532-672 | 6 | 100 | 0 | LCB | |
| 532-670 | 6 | 100 | 0 | RCB | |
| 532-660 | 6 | 100 | 0 | IM | |
| 455-637 | 6 | 100 | 2 | MPA | |
| 455-660 | 6 | 100 | 0 | SIM 1 | |
| 455-660D | 6 | 100 | 2 | SIM 1 | |
| 455-661 | 6 | 100 | 0 | SIM 2 | |
| 455-662 | 6 | 100 | 0 | SIM 3 | |
| 455-663 | 6 | 100 | 0 | SIM 4 | |
| 455-665 | 6 | 100 | 0 | H1 | |
| 455-666 | 6 | 100 | 0 | H1 | |
| 532-650S | 6 | 110 | 6 | PIG | |
| 532-652S | 6 | 110 | 6 | PIG 145° | |
| 532-654S | 6 | 110 | 6 | PIG 145° MOD | |
| 455-613E | 6 | 110 | 8 | PIG | |
| ReprocessedCordis SuperTorque PlusDiagnosticAngiographicCatheter | 533-562 | 5.2 | 65 | 2 | MPA 2 |
| 533-579 | 5.2 | 80 | 0 | MPA | |
| SR3216 | 5.2 | 100 | 0 | JL 3 | |
| 533-551 | 5.2 | 100 | 0 | JL 3.5 | |
| 533-553 | 5.2 | 100 | 0 | JL 4 | |
| 533-527 | 5.2 | 100 | 0 | JL 4.5 | |
| 533-559 | 5.2 | 100 | 0 | JL 5 | |
| 533-561 | 5.2 | 100 | 0 | JL 6 | |
| 533-542 | 5.2 | 100 | 0 | AL 2 | |
| 533-544 | 5.2 | 100 | 0 | AL 3 | |
| SR4615 | 5.2 | 100 | 0 | XB 3.5 ST | |
| 533-550 | 5.2 | 100 | 0 | JR 3.5 | |
| 533-552 | 5.2 | 100 | 0 | JR 4 | |
| 533-528 | 5.2 | 100 | 0 | JR 4 MOD | |
| 533-563 | 5.2 | 100 | 0 | JR 4 ST | |
| Description | Item Number | French Size | Length | Side holes | Curve |
| 533-558 | 5.2 | 100 | 0 | JR 5 | |
| 533-560 | 5.2 | 100 | 0 | JR 6 | |
| SR1924 | 5.2 | 100 | 0 | MPA | |
| 533-582 | 5.2 | 100 | 2 | MPA 2 | |
| 533-584 | 5.2 | 100 | 0 | CAS 1 | |
| 533-585 | 5.2 | 100 | 0 | CAS 2 | |
| 599500J2 | 5.2 | 100 | 2 | RBL-JK | |
| 599500T1 | 5.2 | 100 | 1 | RBL-TG | |
| 533-572 | 5.2 | 100 | 0 | LCB | |
| 533-570 | 5.2 | 100 | 0 | RCB | |
| 533-580 | 5.2 | 100 | 0 | IM | |
| SR4658 | 5.2 | 100 | 0 | IM-MOD | |
| SR4268 | 5.2 | 100 | 0 | IM-MOD-2 | |
| SR4685 | 5.2 | 100 | 0 | IM-MOD-3 | |
| 533-533 | 5.2 | 110 | 6 | PIG | |
| 533-534A | 5.2 | 110 | 6 | PIG 145° | |
| ReprocessedCordis SuperTorque PlusDiagnosticAngiographicCatheter | 533-537S | 5.2 | 110 | 6 | PIG 145° MOD |
| 533-533A | 5.2 | 110 | 6 | PIG 155° | |
| 533-633 | 6 | 80 | 0 | MPA 1 | |
| 533-629 | 6 | 80 | 2 | MPA 2 | |
| 533-682 | 6 | 80 | 4 | CAS 2 | |
| 533-635 | 6 | 80 | 0 | NIH | |
| 533-618 | 6 | 100 | 0 | JL 3.5 | |
| 533-620 | 6 | 100 | 0 | JL 4 | |
| 533-627 | 6 | 100 | 0 | JL 4.5 | |
| 533-622 | 6 | 100 | 0 | JL 5 | |
| 533-624 | 6 | 100 | 0 | JL 6 | |
| 533-645 | 6 | 100 | 0 | AL 1 | |
| 533-646 | 6 | 100 | 0 | AL 2 | |
| 533-647 | 6 | 100 | 0 | AL 3 | |
| 533-619 | 6 | 100 | 0 | JR 3.5 | |
| 533-621 | 6 | 100 | 2 | JR 3.5 | |
| 533-687 | 6 | 100 | 0 | JR 4 CLASSICJR 4 RECESSED | |
| 533-689 | 6 | 100 | 0 | BRAIDING | |
| 533-628 | 6 | 100 | 0 | JR 4 MOD | |
| 533-626 | 6 | 100 | 0 | JR 4 ST | |
| 533-623 | 6 | 100 | 0 | JR 5 | |
| 533-625 | 6 | 100 | 0 | JR 6 | |
| 533-648 | 6 | 100 | 0 | AR MOD | |
| Description | ItemNumber | FrenchSize | Length | Sideholes | Curve |
| ReprocessedCordis SuperTorque PlusDiagnosticAngiographicCatheter | 533-641 | 6 | 100 | 0 | AR 1 MOD |
| 533-643 | 6 | 100 | 0 | AR 2 MOD | |
| 533-640 | 6 | 100 | 0 | MPA 1 COURNAD | |
| 533-642 | 6 | 100 | 2 | MPA 2 | |
| 533-649 | 6 | 100 | 2 | MPB 2 | |
| 533-634 | 6 | 100 | 6 | MPB 3 | |
| 533-630 | 6 | 100 | 4 | SON 1 | |
| 533-631 | 6 | 100 | 0 | SON 2 | |
| 533-632 | 6 | 100 | 4 | SON 3 | |
| 533-684 | 6 | 100 | 0 | CAS 1 | |
| 533-685 | 6 | 100 | 0 | CAS 2 | |
| 533-686 | 6 | 100 | 0 | CAS 3 | |
| 599600R401 | 6 | 100 | 1 | RBL 4.0 | |
| 599600R451 | 6 | 100 | 1 | RBL 4.5 | |
| 599600R501 | 6 | 100 | 1 | RBL 5.0 | |
| 599600J2 | 6 | 100 | 2 | RBL-JK | |
| 599600T1 | 6 | 100 | 1 | RBL-TG | |
| 533-637 | 6 | 100 | 0 | EGB 1 | |
| 533-638 | 6 | 100 | 0 | EGB 2 | |
| 533-672 | 6 | 100 | 0 | LCB | |
| 533-670 | 6 | 100 | 0 | RCB | |
| 533-660 | 6 | 100 | 0 | IM | |
| 533-636 | 6 | 100 | 0 | NIH | |
| 533-650F | 6 | 110 | 4 | PIG | |
| 533-650S | 6 | 110 | 6 | PIG | |
| 533-650E | 6 | 110 | 8 | PIG | |
| 533-652S | 6 | 110 | 6 | PIG 145° | |
| 533-653S | 6 | 110 | 6 | PIG 145° MOD | |
| 533-654S | 6 | 110 | 6 | PIG 155° | |
| 533-655S | 6 | 110 | 6 | PIG 155° MOD |
The item numbers included in the scope of this submission are as follows:
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
Table 1: Device Scope
§ 870.1200 Diagnostic intravascular catheter.
(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).