(60 days)
This device is used with an ultrasound endoscope for fine needle biopsy (FNB), of submucosal and extramural lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the GI tract.
The EchoTip® AcuCore™ Ultrasound Biopsy Needle (ECHO-BX-3-22) is an endoscopic ultrasound needle consisting of a needle assembly and syringe. The needle assembly is comprised of a handle, sheath, stylet and needle cannula. This device is available in one 22Ga needle size.
Here's a breakdown of the acceptance criteria and the study information for the EchoTip® AcuCore™ Ultrasound Biopsy Needle, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Test | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Mechanical Performance | Needle Crumpling Compression Test | Withstand handle angulation | All test articles met the acceptance criteria. |
Suction Test | Ability to perform suction at endoscopic ultrasound needle device based on the suction pressure created by the syringe. | All test articles met the acceptance criteria. | |
Needle Puncture Test | Comparative performance and function of Endoscopic ultrasound needles in terms of puncture force. (Exact quantitative criteria not specified, but implied to be comparable to predicate or sufficient for intended use). | All test articles met the acceptance criteria. | |
Needle Extension Length Test | Meet the needle extension length requirement. (Exact quantitative criteria not specified). | All test articles met the acceptance criteria. | |
Stylet Pull Test | Evaluate the removal force for the device's stylet. (Exact quantitative criteria not specified but implied to be within acceptable range). | All test articles met the acceptance criteria. | |
Functional/Simulated Use | Simulated Use Testing | Qualitatively evaluate the performance of the device using a tortuous path that simulates the intended use conditions. (Specific criteria not detailed, but implied to be successful tissue acquisition/maneuverability). | All acceptance criteria as defined in the study protocol were met. |
Material/Biocompatibility | Biocompatibility Assessment and Biological Testing | Biocompatibility of the subject device when used as intended. (Specific criteria not detailed, but implied to be non-toxic, non-irritating, etc.). | Supports the biological safety of the subject device. |
Shelf Life | Simulated Use Testing (on aged devices) | Support a 1-year shelf life. (Exact quantitative criteria not specified, but implied to maintain performance after aging). | Supports a 1-year shelf life for the subject device. |
Missing Information: It's important to note that the document extensively references "acceptance criteria" but rarely provides specific quantitative values for these criteria. It generally states that "All test articles met the acceptance criteria" or "meets the relevant acceptance criteria."
2. Sample Size Used for the Test Set and Data Provenance:
- Test Set Sample Size: The document does not specify the exact sample sizes for each non-clinical test. It refers to "All test articles" meeting the acceptance criteria, suggesting a finite number of devices were tested for each performance characteristic.
- Data Provenance: The studies were retrospective in the sense that they were conducted by the manufacturer (Cook Ireland Ltd.) for regulatory submission to demonstrate substantial equivalence. The data is from internal testing (Cook), rather than from external clinical trials on human subjects. The country of origin for the manufacturing and potentially the testing is Ireland, as indicated by "Cook Ireland Ltd."
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
- Not Applicable. The document describes non-clinical performance testing (e.g., puncture force, suction) and simulated use testing in a laboratory setting. There is no mention of a "test set" requiring expert ground truth establishment in the context of diagnostic accuracy, as this is a medical device for biopsy, not an AI or diagnostic imaging system. Therefore, no experts were used for this purpose in this context.
4. Adjudication Method for the Test Set:
- Not Applicable. As the performance testing is non-clinical and objective (e.g., measuring force, length), an adjudication method (like 2+1 or 3+1 used for expert discrepancies) is not relevant. The results are based on direct measurements and pre-defined acceptance thresholds.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No. An MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging or AI systems where human readers interpret medical images/data, and the AI's impact on their performance (with vs. without AI assistance) is evaluated. The EchoTip® AcuCore™ is a physical biopsy needle, and its performance is assessed through mechanical, functional, and simulated use testing, not through human reader interpretation of data.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Not Applicable. This question typically pertains to AI algorithms. The EchoTip® AcuCore™ is a physical medical device (biopsy needle), not an algorithm, so the concept of "standalone performance" in this context is not relevant. The device's performance is inherently tied to its use by a human clinician.
7. The Type of Ground Truth Used:
- The "ground truth" for the non-clinical tests was established by engineering specifications, material science standards, and established physical performance benchmarks. For example, the puncture test would have a "ground truth" derived from target tissue models (e.g., worst-case material representative of liver and pancreatic tissue) and measured force, while the extension length test would have a "ground truth" based on the design specifications and measurements. Biocompatibility relies on established regulatory standards and biological testing outcomes.
- For the "Indications for Use" change, the "ground truth" for the justification essentially relies on alignment with market standards (other cleared devices like Boston Scientific's Acquire) and the fact that procedural steps/risks were not identified as new by the internal quality system.
8. The Sample Size for the Training Set:
- Not Applicable. This question is relevant for machine learning algorithms. The EchoTip® AcuCore™ is a physical medical device and does not involve AI or machine learning, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. As there is no training set for this device, the question of how its ground truth was established is not relevant.
§ 876.1075 Gastroenterology-urology biopsy instrument.
(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.