K Number
K230722
Device Name
Eagle device
Manufacturer
Date Cleared
2023-12-08

(267 days)

Product Code
Regulation Number
886.4390
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Eagle device is a prescription device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam. The Eagle device has the same intended use as the predicate device. The Eagle device has the following Indications for Use (IFU) statement:

The Eagle device is indicated for use in selective laser trabeculoplasty (SLT).

Device Description

The Eagle device is a Q-switched, 532 nm-wavelength, frequency-doubled Nd:YAG laser that is intended for use in performing selective laser trabeculoplasty. The laser spots produced by the Eagle device have a 400 um spot size, a 3 ns pulse duration, and a 50-Hz pulse repetition rate. The sequence of laser spots consists of 120 spots in a predefined circumferential elliptical pattern delivered at a pre-defined pulse energy level. The spots are delivered through the limbus to the trabecular meshwork in a non-contact fashion, without the need for the use of a contact gonioscopy lens. The device automatically locates the treatment location. The treatment location may be adjusted slightly by the operator. Once confirmed by the operator, the device then automatically applies the laser treatment sequence to the limbal region of the eye, while the eye tracker compensates for any eye movement. The default energy setting is 1.8 mJ/pulse.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly list "acceptance criteria" for the device's performance in the typical sense of numerical thresholds for a specific metric that must be met. Instead, it describes a non-inferiority study where the Eagle device (subject device) is compared to a conventional SLT device (predicate). The implicit acceptance criteria appear to be the demonstration of non-inferiority in effectiveness (IOP reduction) and a comparable safety profile to the predicate device.

Implicit Acceptance Criteria and Reported Device Performance

Criteria CategoryImplicit Acceptance CriteriaReported Device Performance (Eagle Device vs. Conventional SLT)
Effectiveness (IOP Reduction)Primary Effectiveness Endpoint: Between-group difference in mean change in unmedicated IOP at 6 months compared to baseline must demonstrate non-inferiority to conventional SLT. (Threshold for clinical non-inferiority is not explicitly stated in this summary, but the study was designed to show non-inferiority.)Mean Reduction in Unmedicated IOP at 6 months (mPP Population): - Eagle: 5.48 ± 0.52 mmHg reduction - Conventional SLT: 6.29 ± 0.53 mmHg reduction Difference (SLT - Eagle): -0.80 mmHg (95% Cl -2.28, 0.68). The study concluded that the Eagle device demonstrated acceptable effectiveness comparable to conventional SLT.
Safety (Adverse Events)Primary Safety Outcome: Rate of ocular adverse events (AEs) in the Eagle treatment group at or prior to 12 months should be comparable to the conventional SLT group.Overall Safety Profile: - Serious Ocular SAEs (0-6 months): 1 in Eagle (0%) vs 1 in conventional SLT (1.1%) (subluxation of pre-existing IOL deemed related to conventional SLT). - Serious Ocular SAEs (6-12 months): 1 in Eagle (1.0%) (acute optic neuropathy, non-glaucomatous) vs 0 in conventional SLT. - Most Common Non-Serious AE (0-6 months): Punctate subconjunctival hemorrhage (20.8% in Eagle vs 1.1% in conventional SLT) - resolved without sequelae. - Cataract progression (0-6 months): 3.1% in Eagle vs 1.1% in conventional SLT. - Overall, comparable safety profile to the conventional SLT device, with some differences in types and frequencies of non-serious AEs.
SLIT Lamp Examination (AC Cells & Flare)Changes in anterior chamber cells and flare from screening should be minimal and comparable between groups.AC Cells (Increase +0.5 or more): - Post Procedure: Eagle 31.3% vs SLT 20.9% - 1 Day: Eagle 18.9% vs SLT 12.1% - Rapid reduction over time, with negligible increases beyond 1 month for both groups. AC Flare (Increase +0.5 or more): - Post Procedure: Eagle 20.8% vs SLT 16.5% - 1 Day: Eagle 11.6% vs SLT 7.7% - Rapid reduction over time, with negligible increases beyond 1 month for both groups.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Randomized Participants (ITT Population): 196 participants (99 to Eagle, 97 to conventional SLT)
    • Participants Underwent Assigned Procedure: 187 participants (96 Eagle, 91 conventional SLT)
    • Modified Per Protocol (mPP) Population: 152 participants (77 Eagle, 75 conventional SLT)
  • Data Provenance: Prospective, multi-center, randomized, controlled trial. The specific countries of origin for the data are not explicitly stated, but it's a multi-center study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This was a clinical trial involving patients, not an AI review of images needing expert consensus for ground truth. The "ground truth" for effectiveness was objective physiological measurements (Intraocular Pressure reduction) and for safety, it was the occurrence of adverse events monitored by clinical investigators. Therefore, standard expert-based ground truth establishment (e.g., from radiologists) is not applicable here. Clinicians at the 14 sites would have assessed patients, but "experts" for ground truth in the AI sense are not mentioned.

4. Adjudication Method for the Test Set

This was a clinical trial with objective measures and adverse event reporting. Therefore, an adjudication method like "2+1" for discrepancies in expert readings (common in AI imaging studies) is not directly applicable. Clinical trial protocols typically have methods for reporting, verifying, and adjudicating adverse events and protocol deviations, but this isn't detailed in the provided summary.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The study directly compared two laser devices (Eagle vs. conventional SLT) in a randomized controlled trial on human patients, measuring clinical outcomes (IOP reduction and adverse events), not human reader performance with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone algorithm-only performance study was not conducted in the traditional sense of evaluating an AI's diagnostic accuracy. The Eagle device is a laser system that automatically locates treatment areas and applies laser sequences, with the operator able to make slight adjustments and confirm. The study evaluates the device's clinical performance when used by an operator, not a purely autonomous AI algorithm against a manual process.

7. The Type of Ground Truth Used

The ground truth used in this clinical study was based on:

  • Effectiveness: Objective physiological measurement of unmedicated Intraocular Pressure (IOP).
  • Safety: Clinical observation and reporting of ocular adverse events (AEs).

8. The Sample Size for the Training Set

The provided document describes a clinical validation study for the Eagle device, which is a medical device (laser system), not an AI algorithm that requires a separate training set. Therefore, there is no "training set" in the context of machine learning. The clinical trial directly evaluated the device's performance in patients.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an AI algorithm, this question is not applicable. The device itself is the subject of the clinical investigation, with its internal automation being part of its design.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 8, 2023

BELKIN Vision Ltd. % Anne-Marie Riplev Clinical & Regulatory Consultant Regulatory Pathways Group, Inc. 440 N. Barranca Ave., #2471 Covina. California 91723

Re: K230722

Trade/Device Name: Eagle Device Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: November 2, 2023 Received: November 2, 2023

Dear Anne-Marie Ripley:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Claudine H. Krawczyk -S

Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230722

Device Name Eagle device

Indications for Use (Describe) The Eagle device is indicated for use in selective laser trabeculoplasty (SLT).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K230722

l. Submitter Information

  • 510(k) Owner: BELKIN Vision Ltd. 13 Gan Rave Blvd. Yavne, Israel, 8122214 +972-8-857-1619
  • Contact Person: Anne-Marie Ripley Regulatory Pathways Group, Inc. 440 N. Barranca Ave., #2471 Covina, CA 91723 aripley@regulatorypathways.com
  • Date Prepared: December 8, 2023

II. Device Name and Classification

  • Device Trade Name: Eagle device
  • Common Name: Ophthalmic laser
  • Classification Name: Ophthalmic laser
  • Regulation Number: 21 CFR 886.4390
  • Device Classification: Class 2
  • Product Code: HQF

III. Predicate Device and Reference Devices

Predicate Device

  • Lumenis Selecta Duet LED (K220877) .

Reference Devices

  • Topcon PSLT for PASCAL Streamline (K171488) ●
  • OD-OS Navilas Laser System 577s (K162191)

{5}------------------------------------------------

Topcon PSLT for PASCAL Streamline (K171488) supports automated delivery of laser spots in a pre-defined pattern for SLT indications and high pulse repetition rate laser spots.

OD-OS Navilas Laser System 577s (K162191), which allows for laser treatment to the retina using an automated delivery system with eye tracking, supports algorithms for eye tracking to ensure that the laser beam position is maintained at the desired treatment location throughout the laser treatment.

IV. Device Description

The Eagle device is a Q-switched, 532 nm-wavelength, frequency-doubled Nd:YAG laser that is intended for use in performing selective laser trabeculoplasty. The laser spots produced by the Eagle device have a 400 um spot size, a 3 ns pulse duration, and a 50-Hz pulse repetition rate. The sequence of laser spots consists of 120 spots in a predefined circumferential elliptical pattern delivered at a pre-defined pulse energy level. The spots are delivered through the limbus to the trabecular meshwork in a non-contact fashion, without the need for the use of a contact gonioscopy lens. The device automatically locates the treatment location. The treatment location may be adjusted slightly by the operator. Once confirmed by the operator, the device then automatically applies the laser treatment sequence to the limbal region of the eye, while the eye tracker compensates for any eye movement. The default energy setting is 1.8 mJ/pulse.

V. Intended Use and Indications for Use

The Eagle device is a prescription device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam. The Eagle device has the same intended use as the predicate device. The Eagle device has the following Indications for Use (IFU) statement:

The Eagle device is indicated for use in selective laser trabeculoplasty (SLT).

The IFU statement of the Eagle device is not substantially different from that of the predicate device. The non-substantial difference in the IFU statement for the Eagle versus the predicate

{6}------------------------------------------------

device is that the IFU statement for the Eagle device is narrower; it does not include posterior capsulotomy, pupillary membranectomy in aphakic and pseudophakic patients, and iridotomy.

Comparison of Technological Characteristics with the Predicate Device VI.

Although the Eagle device and the predicate do not share identical technological characteristics, these differences do not raise different types of questions of safety and effectiveness.

CharacteristicSubject DeviceBELKIN VisionEagle devicePredicate DeviceLumenis Be Inc.Selecta Duet LEDComparison of Subject Device toPredicate Device
K230722K220877
Device Class22Same
ClassificationProduct CodeHQFHQFSame
Regulation Number886.4390886.4390Same
Intended UseSelective LaserTrabeculoplasty (SLT)Selective LaserTrabeculoplastySame
Indications for useSelective lasertrabeculoplasty (SLT).In SLT mode:Selective lasertrabeculoplasty (SLT)In YAG mode:Photodisruption ofocular tissue using lightenergy emitted by aNd:YAG laser, includingdiscission of theposterior capsule of theeye (posteriorcapsulotomy), anddiscission of pupillarymembranes (pupillarymembranectomy) inaphakic andpseudophakic patients,andiridotomy/iridectomyDifferent, but the differences donot raise different types ofquestions of safety andeffectiveness. Both devices areindicated for SLT.
Laser Parameters
Laser typeQ-switched, frequency-doubled Nd:YAGQ-switched, frequency-doubled Nd:YAGSame
Wavelength532 nm532 nmSame
Laser pulse duration3 ns3 nsSame
CharacteristicSubject DevicePredicate DeviceComparison of Subject Device to Predicate Device
BELKIN VisionEagle deviceK230722Lumenis Be Inc.Selecta Duet LEDK220877
Pulse energy range1.1 – 1.9 mJ0.3 – 2.6 mJDifferent, but the difference does not raise different types of questions of safety and effectiveness. The pulse energy range of Subject Device is within the energy range of the Predicate Device. The energy level of the subject device is fixed for the entire set of 120 spots while the energy level of the predicate device may be changed as laser spots are applied to the trabecular meshwork (TM).The safety and effectiveness of the Eagle device's pulse energy range were supported by non-clinical and clinical performance data.
Pulse repetition rate50 HzUp to 3 HzDifferent, but the difference does not raise different types of questions of safety and effectiveness.The safety and effectiveness of the Eagle device's 50-Hz pulse repetition rate were supported by non-clinical and clinical performance data.
Laser Beam Delivery
Method of laser delivery (pattern)Automated delivery of laser spots in a pre-defined pattern (360° ellipse) through the limbus without the use of a contact gonioscopy lens; the trabecular meshwork (TM) is not directly visualized as laser spots are applied.Manual delivery of laser spots through a contact gonioscopy lens to directly visualize the TM. Spots are applied for usually 180° or 360° of the TM.Same pattern (i.e. 360° ellipse)Method of delivery is different, but the difference does not raise different types of questions of safety and effectiveness.The safety and effectiveness of the Eagle device's automated delivery were supported by non-clinical and clinical performance data.
Target tissueTrabecular meshworkTrabecular meshworkSame
CharacteristicSubject DevicePredicate DeviceComparison of Subject Device toPredicate Device
BELKIN VisionEagle deviceLumenis Be Inc.Selecta Duet LED
K230722K220877
Method ofmaintaining laserbeam targetingAnatomical detectionalgorithm and eyetracking to compensatefor eye movementsduring the procedure.Direct visualization ofaiming beam and targettissue through slit lampmicroscope and handheldgonioscope contact lens.Different, but the difference doesnot raise different types ofquestions of safety andeffectiveness.Reference Device 2 supports thebench testing methods used todemonstrate the safety andeffectiveness of this technologicalfeature.
Spot diameter400 μm400 μmSame
Auxiliary optical characteristics
Aiming laserDiode laserDiode laserSame
Aiming wavelength635 nm635 nmSame
Method ofmaintaining thefocal position oftreatment beamManually set using ajoystick to adjust theaxial position byvisualizing the overlap oftwo ranging (650 nm)diode laser beams.Manually set using ajoystick to adjust the axialposition using aimingbeam and microscopeimage as a guide.Similar
IlluminationLED illumination ring,visible white lightLED illumination source,both visible white lightand filtered lightSimilar
Physical dimensions
System weight30 kg / 66 lbs31 kg / 68 lbsSimilar
System dimensions(H x W X D)52 x 53 x 63 cm20.5 x 21.8 x 25 inches57 x 75 x 44 cm,23 x 30 x 18 inchesSimilar

{7}------------------------------------------------

{8}------------------------------------------------

Summary of Non-Clinical Testing VII.

Based on the risk assessment and design control requirements, the following verification and validation testing was performed:

  • Device performance testing ●
  • Validation of laser fluence delivery to the target tissue
  • . Environmental and transportation testing were conducted per elements of:
    • o ASTM D4332-14 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing

{9}------------------------------------------------

  • ASTM D6653-13 Standard Test Methods for Determining the Effects of High o Altitude on Packaging Systems by Vacuum Method
  • o ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems (Assurance Level II)
  • ASTM D5276-19 Standard Test Method for Drop Test of Loaded Containers by O Free Fall
  • ASTM D4728-17 Standard Test Method for Random Vibration Testing of о Shipping Containers
  • Laser and optical radiation hazard evaluation and safety testing was conducted per: ●
    • IEC 60601-2-22: 2007 (3rd Ed.) Medical electrical equipment Part 2: Particular o requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
    • IEC 60825-1:2014 Safety of laser products Part 1: Equipment classification and o requirements
  • Product reliability testing was conducted per Telcordia SR-332.
  • Biocompatibility testing: The biocompatibility profile of the tissue contacting components of the device were assessed for cytotoxicity, sensitization (Guinea pig maximization) and intracutaneous reactivity testing as recommended by FDA's 2020 Biocompatibility Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'" and relevant parts of International Standard Organization (ISO) 10993 Biological evaluation of medical devices standard series.
  • . Software verification and validation testing: BELKIN developed and verified the software in accordance with a major level of concern described in the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and also per the IEC 62304:2006 and A1:2015 Medical Device Software - Software Life Cycle Processes standard.
  • . Electromagnetic compatibility (EMC) was conducted per IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests.
  • Electrical safety testing was conducted per IEC 60601-1:2005 + CORR1:2006 + . CORR2:2007 + AMD1:2012 Medical electrical equipment – Part 1: General requirements for safety 1: collateral standard: Safety Requirements for Medical Electrical Systems.
  • Human factors testing was conducted per
    • FDA guidance, "Applying Human Factors and Usability Engineering to Medical o Devices"
    • ANSI/AAMI IEC 62366-1 Medical devices Part 1: Application of usability o engineering to medical devices, and
    • IEC 60601-1-6 Medical electrical equipment Part 1-6: General requirements for o basic safety and essential performance – Collateral standard: Usability
  • Animal testing was conducted per 21 CFR Part 58 (Good Laboratory Practices) in rabbit to evaluate the acute and subacute safety of the device compared to the standard

{10}------------------------------------------------

selective laser trabeculoplasty (SLT). Macroscopic treatment-related changes were comparable in eyes undergoing SLT with the Eagle device and the SLT comparator device. Mild changes were limited in acute phase and resolved by day 29. The study provides evidence to support the substantial equivalence of the Eagle device to the predicate device in post-treatment tissue reactions.

Results of the non-clinical testing support a substantial equivalence determination. The Eagle device is substantially equivalent to its predicate device for the indications for use.

VIII. Clinical Performance Testing

A prospective, multi-center, randomized, controlled trial was conducted to evaluate the safety and effectiveness of laser trabeculoplasty using the Eagle device to compared to conventional selective laser trabeculoplasty (SLT). 276 participants age 40 years or older with primary openangle glaucoma, pseudoexfoliation glaucoma, pigmentary glaucoma, or ocular hypertension on three or fewer IOP-lowering medications at baseline were enrolled across 14 sites. Prior incisional or laser-based glaucoma procedures, severe glaucoma in either eye, dense pigmentation or hemorrhage in the perilimbal conjunctival area or anterior sclera were exclusionary. Eligible participants underwent baseline IOP-lowering medication washout and those with post-washout IOP between 22 – 35 mmHg were randomized 1:1 to 360° treatment with either the Eagle device or with a conventional SLT device. Participants were followed for 12 months post-treatment. The primary effectiveness endpoint was the between-group difference in the mean change in unmedicated IOP at 6 months compared to baseline. The primary safety outcome was the rate of ocular adverse events (AEs) in each treatment group at or prior to 12 months.

A total of 196 participants were enrolled and randomized, 99 to Eagle and 97 to conventional SLT. Of these, 187 participants, 96 Eagle and 91 conventional SLT, underwent the assigned laser procedure. The mean age of participants was 65.6 ± 9.4 years (range 40 – 88 years) and 39.0% were women. 98% of participants were white, 1% were Black, 0.5% were Asian and 0.5% were mixed race. 87% of participants had glaucoma. 73% had primary open angle glaucoma, 11% had pseudoexfoliative glaucoma and 3% had pigmentary glaucoma. The remaining 13% of participants had ocular hypertension. Mean screening IOP was 21.5 ± 5.4 mmHg for the Eagle group and 20.5 ± 4.5 mmHg for the conventional SLT group. At the time of enrollment, 32.2% of participants were not taking any hypotensive medications. The average number of ocular hypotensive medications at screening was 1.2 ± 1.0 for the Eagle group compared to 1.1 ± 1.0 for the SLT group. Post-washout baseline IOP was 26.5 ± 3.6 mmHg in the Eagle group and 25.8 ± 3.6 mmHg in the SLT group.

Effectiveness

The Eagle procedure provided a mean (±SE per the least-square estimation) reduction of unmedicated IOP at 6 months of 5.5 ± 0.5 mmHg (95% Cl -6.5 to -4.5), compared with 6.3 ± 0.5 mmHg (95% Cl -7.3 to -5.2) in the conventional SLT group (Table 1). The difference in mean

{11}------------------------------------------------

reduction in IOP between the two groups (SLT-Eagle) was -0.80 mmHg (95% Cl -2.28 to 0.68 mmHg). Table 2 presents the primary effectiveness endpoint analysis for the Eagle group, stratified by the energy level used.

MethodEagleNMean ± SE3(95% CI)SLTNMean ± SE3(95% CI)ANCOVA p-value2p-valueH0: Δ ≤ -1.95 mmHgHa: Δ > -1.95 mmHgΔ = Difference in Mean(SLT - Eagle)
ITT Population4 (N= 196)
Primary Analysis99-4.63 ± 0.49(-5.60, -3.67)97-5.70 ± 0.50(-6.69, -4.71)0.209
Difference in mean = SLT – Eagle(95% CI)-1.07 (-2.45, 0.32)
mPP Population5 (N= 152)
Primary Analysis77-5.48 ± 0.52(-6.52, -4.45)75-6.29 ± 0.53(-7.34, -5.23)0.127
Difference in mean = SLT - Eagle(95% CI)-0.80 (-2.28, 0.68)

Table 1: Primary Effectiveness Endpoint 6 Month Unmedicated IOP Change from Baseline Primary Analysis1 per Least Squares Estimation of ANCOVA Model2

Baseline value was used to impute 6-month for participants with SSI or unsafe to washout at 6 months. The SSIs were defined as 1 surgical procedures that might affect the level of IOP (such as iridotomy, trabeculectomy, glaucoma shunt implantation, argon laser trabeculoplasty, selective laser trabeculoplasty, cataract surgery that might affect IOP).

2 ANCOVA Model: Yij = μ + ti + ß*(Xij – average of Xij) + rj + eij, Yij is the change-from-baseline IOP (the 6-month IOP value was subtracted from baseline IOP for each eye) for treatment i and study eve i, Xij is the corresponding baseline IOP measurement, μ is the overall mean, ti is the treatment indicator, ß is the effect of baseline IOP, rj is an indicator of beta-blocker use in the fellow eye, and eij is the error term.

3 Standard Error (i.e., standard deviation of mean)

4 The ITT Population is defined as all randomized participants.

5 The Modified Per Protocol (mPP) population included participants who met the Per Protocol (PP) definition. In addition, participants who met any of the following criteria during the first 6 months were included in the mPP population as failures: Had a secondary surgical intervention (SSI) in the study eye that could affect IOP; had an ocular SAE in the study eye; or participant for whom 6-month washout was considered unsafe. For these participants, 6-month IOP was imputed with their baseline IOP measurement for the primary endpoint analysis. Participants who were missing their reasons were not included in the mPP population.

Per protocol (PP) was defined as all enrolled and randomized participants who were treated and for whom data concerning the primary effectiveness endpoint measure was available and who had no major protocol deviations (e.g. inclusion and/or exclusion criteria violations, treatment not according to randomization, IOP measured in an unmasked fashion). Participants with major protocol deviations were excluded from the PP population.

{12}------------------------------------------------

Table 2: Primary Effectiveness Endpoint, Stratified by Energy Level (Eagle Group Only) 6 Month Unmedicated IOP Change from Baseline Primary Analysis1 per Least Squares Estimation of ANCOVA Model4

Primary Effectiveness Endpoint1.1 mJ/shot1.2 mJ/shot1.3 mJ/shot1.4 mJ/shot1.5 mJ/shot1.6 mJ/shot1.7 mJ/shot1.8 mJ/shot1.9 mJ/shot
ITT Population2
N42147939264
Change in unmedicated IOP from baseline-5.29 ± 2.51-7.66 ± 3.45-8.63 ± 4.95-6.10 ± 2.44-1.88 ± 1.85-7.23 ± 1.64-4.93 ± 0.78-3.40 ± 0.95-4.87 ± 2.48
to 6 months (mmHg) Mean (SE)3(-10.29, -0.29)(-14.52, -0.79)(-18.47, 1.20)(-10.96, -1.24)(-5.57, 1.80)(-10.50, -3.96)(-6.49, -3.38)(-5.30, -1.51)(-9.81, 0.07)
mPP Population5
N42123933203
Change in unmedicated IOP from baseline-4.83 ± 2.53-7.36 ± 3.46-8.52 ± 4.95-7.83 ± 3.45-5.32 ± 2.83-7.28 ± 1.64-5.50 ± 0.85-4.30 ± 1.10-7.30 ± 2.87
to 6 months (mmHg) Mean (SE)3(-9.88, 0.23)(-14.27, -0.45)(-18.40, 1.37)(-14.72, -0.95)(-10.96, 0.33)(-10.56, -4.00)(-7.19, -3.80)(-6.50, -2.11)(-13.03, -1.58)

I - Baseline value was used to imports with SS or unsfe to washout at 6 months. The SS were edition as surgical rocess that might affect the level of CP (such as inidorin, in trabeculectory, glaucoma shunt implantation, argon laser trabeculoplasty, cataract surgery), or other surgery that might affect (O.

2 The ITT Population is defined as all randomized participants. The ITT analysis because they did not undergo the assered asser are asser as a care as a care a coredure.

3 Standard Error (i.e., standard deviation of mean)

4 ANCOVA Model: Yij = i + i + f + f + if + i + ij, tij it the change-from-baseline (0 V the E-month (0 P rach qe) for earth eye for energy level and study eye i, liji sthe corresponding baseline (OP neasurenent, ui s the energy level indicator, Sis the effect of backer use in the fellow eye, and eij is the erot term.

The Modiled Percocol (nP) population in the Per Protocol (P) definition. In addition, participants who net any of the fille for the first Gronths wee included in the nP population as fallures: Halo secondary ($$) in the study eye that could affect OP; had an coular SAE in the study ever on participant for whon F-nonth was considered wasi these participants, Frinth OP was inqueted to the primary endopint analyss. Participants who were missing their F-month data for other reasons were not included in the mP population.

Per protocol (PP) was defined and andomized participats who whom data concerning the primary effectiveness endopint measure was wailoble and who had no maip rotocol devations (e, inclusion and or exclusions, treatment on teaching to randomization, (0º measured in an unmasked fashion). Participants with napo probod devaitions were ecclud from the PP population.

{13}------------------------------------------------

Safety

No ocular serious adverse events (SAEs) were reported in the first six months. One ocular SAE of subluxation of a pre-existing intraocular lens (IOL) was reported in one participant in the conventional SLT group on post-procedure day 1, which was corrected with surgical repositioning. One ocular SAE was reported in an Eagle group participant who experienced a decline in visual field due to non-glaucomatous acute optic neuropathy at the 12-month followup visit.

The most commonly reported non-serious AE was punctate subconjunctival hemorrhage (21% in the Eagle group vs 1% in the conventional SLT group). These events resolved without clinical sequelae. Elevated IOP was reported in two participants in the Eagle group and two participants in the conventional SLT group. The proportion of participants with a worsening of visual field mean deviation by ≥2.5 dB was 6.6% and 9.9% in the Eagle group vs. 12.6% and 15.7% in the conventional SLT group at 6 and 12 months, respectively. In the first six months, progression of cataracts was reported in three Eagle participants and one conventional SLT participant. Between 6 and 12 months, three Eagle and four conventional SLT participants had progression of cataracts and one conventional SLT participant developed a new cataract.

Secondary surgical interventions (SSIs) were reported in four Eagle participants (trabeculectomy [N=1], cataract surgery [N=2], and laser retinopexy [N=1] to repair a retinal tear) and three conventional SLT participants (cataract surgeries [N=2], IOL repositioning [N=1]). Table 3 and Table 4 present study eye ocular AEs within the first 6 months and between 6-12 months, respectively. Ocular adverse events in the study eye are shown stratified by energy level in Tables 7 and 8.

Anterior chamber (AC) cells and flare findings at each study visit for change from screening in are shown in Tables 5 and 6. Mild corneal haze was reported in one Eagle participant at Month 12. No moderate or severe corneal haze was reported in either group. At the 12-month assessment of 5-minute delayed corneal staining, 17% of the Eagle participants and 20% of the conventional SLT participants had staining in one quadrant. 5% of the Eagle participants and 4% of the conventional SLT participants had staining in two quadrants. No participants in either group had staining in three or more quadrants. Conjunctival staining scores at Month 12 were 2.11±3.01 in the Eagle group and 1.83±2.22 in the conventional SLT group.

{14}------------------------------------------------

Ocular EventsEagleSLT
N = 96N = 91
# of Eventsn (%)# of Eventsn (%)
Serious Adverse Eventsl------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
Subluxation of intra-ocular lens------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
Non-Serious Adverse Events4434 (35.4%)2620 (22.0%)
Subconjunctival hemorrhage, punctate2020 (20.8%)11 (1.1%)
Foreign body sensation33 (3.1%)33 (3.3%)
Cataract progressionર્ રે3 (3.1%)11 (1.1%)
Superficial punctate keratitis22 (2.1%)44 (4.4%)
Elevated IOP (IOP increase from baseline22 (2.1%)22 (2.2%)
> 10 mmHg)
Conjunctivitis22 (2.1%)----
Mild or moderate anterior chamber11 (1.0%)33 (3.3%)
inflammation
Transient blurred vision1 (1.0%)22 (2.2%)
Corneal erosion11 (1.0%)11 (1.1%)
Eye discharge11 (1.0%)11 (1.1%)
Eye discomfort11 (1.0%)11 (1.1%)
Eye pain11 (1.0%)11 (1.1%)
Hordeolum externum11 (1.0%)11 (1.1%)
Blepharitis11 (1.0%)--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Eye itchiness11 (1.0%)---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Periodical hyperlacrimation1 (1.0%)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Retinal tear11 (1.0%)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Subconjunctival hemorrhage, moderate11 (1.0%)---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Cystoid macular edema------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
Elevated IOP (IOP increase from baseline---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
< 10 mmHg)<
Eye burning sensation-----11 (1.1%)
Floater--11 (1.1%)
Periorbital pain--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
Any adverse events4434 (35.4%)2720 (22.0%)
Any adverse events (excludingSubconiunctival hemorrhage, punctate)2419 (19.8%)2619 (20.9%)

Table 3: Study Eye Adverse Events from 0-6 Months (Safety Population)

The counts (n) are the number of participants reported with the corresponding events. % = n / N x 100%. Multiple events could be reported for the same participant.

1 Miettinen-Nurminen method

² One SLT participant was reported with a post-treatment AE of 'elevated IOP' before the 6-Month visit (Day 199), although the IOP was NOT elevated >10 mmHg more than baseline.

3 One Eagle participant was reported to have a cataract at the 6-month visit and subsequently underwent cataract surgery. However, this participant's cataract was pre-existing with severe opacity (3+) at the time of enrollment. A protocol deviation was reported for enrollment of this participant despite them having a pre-existing severe cataract.

{15}------------------------------------------------

Ocular EventsEagleSLT
N = 96N = 91
# of Eventsn (%)# of Eventsn (%)
Serious Adverse Events11 (1.0%)--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Acute optic neuropathy11 (1.0%)-----
Non-Serious Adverse Events76 (6.3%)77 (7.7%)
Cataract progression33 (3.1%)44 (4.4%)
Conjunctivitis1 (1.0%)-----
Diabetic retinopathy1 (1.0%)--
Elevated IOP (IOP increase from baseline11 (1.0%)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
≥ 10 mmHg)2
Foreign body sensation1 (1.0%)---
Cataract----11 (1.1%)
Cataract surgery--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11 (1.1%)
Guttata----11 (1.1%)
Any adverse events87 (7.3%)77 (7.7%)

Table 4: Study Eye Adverse Events from 6-12 Months (Safety Population)

The counts (n) are the number of participants reported with the corresponding events. % = n / N x 100%. Multiple events could be reported for the same participant.

¹ Miettinen-Nurminen method

2 Steroid-induced elevated IOP occurred in one Eagle participant and was due to use of nasal steroid. This participant had a similar event during the first 6 months.

{16}------------------------------------------------

PostProceduren (%)1Dn (%)7Dn (%)1Mn (%)3Mn (%)6Mn (%)12Mn (%)
Eagle (96 Eyes)
N96959494929188
Increase by +33 (3.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +26 (6.3%)5 (5.3%)1 (1.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +116 (16.7%)3 (3.2%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +0.55 (5.2%)10 (10.5%)2 (2.1%)0 (0.0%)0 (0.0%)1 (1.1%)1 (1.1%)
No Change66 (68.8%)77 (81.1%)91 (95.8%)94 (100.0%)92 (100.0%)90 (97.8%)81 (92.0%)
Decrease by +0.50 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)1 (1.1%)
Decrease by +10 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)5 (5.7%)
Decrease by +20 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase +0.5 ormore (95% CI)¹30 (31.3%)(22.2%,41.5%)18 (18.9%)(11.6%,28.3%)3 (3.2%)(0.7%,9.0%)0 (0.0%)(0.0%,3.8%)0 (0.0%)(0.0%,3.9%)1 (1.1%)(0.0%,5.9%)1 (1.1%)(0.0%,6.2%)
Not Reported0010010
Total96959594929288
SLT (91 Eyes)
N89909189878681
Increase by +30 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +21 (1.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +19 (9.9%)3 (3.3%)2 (2.2%)1 (1.1%)1 (1.1%)1 (1.2%)1 (1.2%)
Increase by +0.59 (9.9%)8 (8.8%)1 (1.1%)1 (1.1%)0 (0.0%)0 (0.0%)1 (1.2%)
No Change70 (76.9%)79 (86.8%)88 (96.7%)87 (97.8%)86 (98.9%)85 (98.8%)75 (91.5%)
Decrease by +0.50 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)2 (2.4%)
Decrease by +10 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)2 (2.4%)
Decrease by +20 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase +0.5 ormore (95% CI)¹19 (20.9%)(13.1%,30.7%)11 (12.1%)(6.2%,20.6%)3 (3.3%)(0.7%,9.3%)2 (2.2%)(0.3%,7.9%)1 (1.1%)(0.0%,6.2%)1 (1.2%)(0.0%,6.3%)2 (2.4%)(0.3%,8.5%)
Not Reported2100001
Total91919189878682

Table 5: Slit Lamp Examination – Change in Anterior Chamber Cells from Screening by Visit (Safety Population)

1 Binomial distribution

{17}------------------------------------------------

PostProceduren (%)1Dn (%)7Dn (%)1Mn (%)3Mn (%)6Mn (%)12Mn (%)
Eagle (96 Eyes)
N96959494929188
Increase by +31 (1.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +22 (2.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +19 (9.4%)11 (11.6%)2 (2.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +0.58 (8.3%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)1 (1.1%)1 (1.1%)
No Change76 (79.2%)84 (88.4%)92 (96.8%)94 (100.0%)92 (100.0%)90 (97.8%)82 (93.2%)
Decrease by +0.50 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)2 (2.3%)
Decrease by +10 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)3 (3.4%)
Decrease by +20 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase +0.5 ormore (95% CI)¹20 (20.8%)(13.2%,30.3%)11 (11.6%)(5.9%,19.8%)2 (2.1%)(0.3%,7.4%)0 (0.0%)(0.0%,3.8%)0 (0.0%)(0.0%,3.9%)1 (1.1%)(0.0%,5.9%)1 (1.1%)(0.0%,6.2%)
Not Reported0010010
Total96959594929288
SLT (91 Eyes)
N89909189878681
Increase by +30 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +21 (1.1%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase by +18 (8.8%)2 (2.2%)1 (1.1%)1 (1.1%)1 (1.1%)1 (1.2%)1 (1.2%)
Increase by +0.56 (6.6%)5 (5.5%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
No Change74 (81.3%)82 (90.1%)89 (97.8%)87 (97.8%)85 (97.7%)84 (97.7%)75 (91.5%)
Decrease by +0.50 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)1 (1.2%)
Decrease by +10 (0.0%)1 (1.1%)1 (1.1%)1 (1.1%)1 (1.1%)1 (1.2%)4 (4.9%)
Decrease by +20 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Increase +0.5 ormore (95% CI)¹15 (16.5%)(9.5%,25.7%)7 (7.7%)(3.1%,15.2%)1 (1.1%)(0.0%,6.0%)1 (1.1%)(0.0%,6.1%)1 (1.1%)(0.0%,6.2%)1 (1.2%)(0.0%,6.3%)1 (1.2%)(0.0%,6.6%)
Not Reported2100001
Total91919189878682

Table 6: Slit Lamp Examination – Change in Anterior Chamber Flare from Screening by Visit (Safety Population)

1 Binomial distribution

{18}------------------------------------------------

Ocular Events1.1 mJ/shotN = 4n (%)1.2 mJ/shotN = 2n (%)1.3 mJ/shotN = 1n (%)1.4 mJ/shotN = 4n (%)1.5 mJ/shotN = 7n (%)1.6 mJ/shotN = 9n (%)1.7 mJ/shotN = 39n (%)1.8 mJ/shotN = 26n (%)1.9 mJ/shotN = 4n (%)TotalN = 96n (%)
Non-Serious Adverse Events1 (100.0%)1 (25.0%)2 (28.6%)4 (44.4%)10 (25.6%)15 (57.7%)1 (25.0%)34 (35.4%)
Blepharitis1 (3.8%)1 (1.0%)
Cataract progression1 (11.1%)2 (7.7%)3 (3.1%)
Conjunctivitis1 (11.1%)1 (2.6%)2 (2.1%)
Corneal erosion1 (2.6%)1 (1.0%)
Elevated IOP1 (14.3%)1 (3.8%)2 (2.1%)
Eye discharge1 (2.6%)1 (1.0%)
Eye discomfort1 (3.8%)1 (1.0%)
Eye itchiness1 (3.8%)1 (1.0%)
Eye pain1 (3.8%)1 (1.0%)
Foreign body sensation1 (25.0%)1 (2.6%)1 (3.8%)3 (3.1%)
Hordeolum externum1 (25.0%)1 (1.0%)
Mild or moderate anterior chamberinflammation1 (3.8%)1 (1.0%)
Periodical hyperlacrimation1 (2.6%)1 (1.0%)
Retinal tear1 (3.8%)1 (1.0%)
Subconjunctival hemorrhage,moderate1 (11.1%)1 (1.0%)
Subconjunctival hemorrhage,punctate1 (100.0%)2 (28.6%)1 (11.1%)7 (17.9%)9 (34.6%)20 (20.8%)
Superficial punctate keratitis2 (7.7%)2 (2.1%)
Transient blurred vision1 (25.0%)1 (1.0%)
Any adverse events0 Reportsfrom0 Reportsfrom1 Reportsfrom2 Reportsfrom3 Reportsfrom4 Reportsfrom12 Reportsfrom21 Reportsfrom1 Reportsfrom44 Reportsfrom
0 participants0.0%0 participants0.0%1 participant100.0%1 participant25.0%2 participants28.6%4 participants44.4%10 participants25.6%15 participants57.7%1 participant25.0%34 participants35.4%

Table 7: Study Eye Adverse Events at ≤6 Months, Stratified by Energy Level (Eagle Group Only) (Safety Population)

The counts (n) are the number of participants reported with the corresponding events. % = n / N x 100%. Multiple events could be reported for the same participant.

{19}------------------------------------------------

Table 8: Study Eye Adverse Events at >6 Months, Stratified by Energy Level (Eagle Group Only) (Safety Population)

Ocular Events1.1 mJ/shot1.2 mJ/shot1.3 mJ/shot1.4 mJ/shot1.5 mJ/shot1.6 mJ/shot1.7 mJ/shot1.8 mJ/shot1.9 mJ/shotTotal
N = 4n (%)N = 2n (%)N = 1n (%)N = 4n (%)N = 7n (%)N = 9n (%)N = 39n (%)N = 26n (%)N = 4n (%)N = 96n (%)
Serious Adverse Events1 (3.8%)1 (1.0%)
Acute optic neuropathy1 (3.8%)1 (1.0%)
Non-Serious Adverse Events1 (25.0%)1 (25.0%)1 (14.3%)1 (11.1%)1 (3.8%)1 (25.0%)6 (6.3%)
Cataract progression1 (25.0%)1 (11.1%)1 (3.8%)3 (3.1%)
Conjunctivitis1 (25.0%)1 (1.0%)
Diabetic retinopathy1 (25.0%)1 (1.0%)
Elevated IOP1 (14.3%)1 (1.0%)
Foreign body sensation1 (25.0%)1 (1.0%)
Any adverse events1 Reportsfrom1 participant25.0%0 Reportsfrom0 participants0.0%0 Reportsfrom0 participants0.0%2 Reportsfrom1 participant25.0%1 Reportsfrom1 participant14.3%1 Reportsfrom1 participant11.1%0 Reportsfrom0 participants0.0%2 Reportsfrom2 participants7.7%1 Reportsfrom1 participant25.0%8 Reportsfrom7 participants7.3%

The counts (n) are the number of participants reported with the corresponding events. % = n / N x 100%.

Multiple events could be reported for the same participant.

{20}------------------------------------------------

IX. Conclusions

The Eagle device has the same intended use as the legally marketed predicate device identified in this premarket notification. The IFU statement differs from that of the predicate, but the differences do not change the intended use of the device. The technological characteristics of the Eagle device differ from those of the predicate device, but the differences do not raise new or different types of questions of safety or effectiveness. The results of the non-clinical performance testing demonstrate that the Eagle device functions as intended. The results of the clinical performance testing support an acceptable safety and effectiveness profile that supports a determination of substantial equivalence. The non-clinical and clinical performance testing demonstrate that the Eagle device is substantially equivalent to the predicate device.

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.