(28 days)
The ARIA Radiation Therapy Management product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Therapy Management also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments.
The ARIA Radiation Therapy Management product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Therapy Management also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA Radiation Therapy Management supports the integration of all data and images in one central database including archiving and restoration. The different ARIA Radiation Therapy Management features support the visualization, processing, manipulation and management of all data and images stored in the system. Images can also be imported through using DICOM, the available image import filters or by means of film digitizers.
The provided text is a 510(k) summary for Varian Medical Systems' ARIA Radiation Therapy Management (v18.0) device. It describes the device, its intended use, comparison to a predicate device, and performance data from non-clinical testing.
However, the document does not contain the following information necessary to describe acceptance criteria and the study that proves the device meets those criteria:
- Specific acceptance criteria: The document mentions "applicable requirements were met" and "safeguards against hazards functioned properly" but does not detail what these specific quantitative or qualitative acceptance criteria were for the software verification and validation.
- Reported device performance: While it states "test results showed that applicable requirements were met," it does not provide any specific performance metrics or data (e.g., accuracy, precision, error rates) that were measured and compared against acceptance criteria.
- Sample size and data provenance for test set: No information is given about a test set, as all testing was non-clinical software verification and validation.
- Experts for ground truth and adjudication method: These are typically relevant for studies involving human interpretation or clinical endpoints, which this submission explicitly states were not conducted.
- MRMC comparative effectiveness study: The document clearly states that "No data from animal studies or clinical tests have been included" and "no animal or clinical studies were conducted for the subject device." Therefore, no MRMC study was performed.
- Standalone performance: Since no clinical studies were performed, there are no reported standalone performance metrics in the context of interpretation or diagnosis. The performance mentioned refers to software functionality.
- Type of ground truth: Ground truth is usually associated with clinical or pathological verification in studies, which are absent here.
- Training set information: As this is primarily a software update and management system, not a machine learning algorithm that requires a "training set" in the conventional sense, this information is not applicable and not provided.
Based on the provided text, I can only state what is mentioned regarding performance, rather than providing the requested table and details about a clinical study:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (General) | Reported Device Performance |
|---|---|
| Applicable software requirements were met | Test results showed that applicable requirements were met |
| Safeguards against hazards functioned properly | Test results assured that safeguards against hazards functioned properly |
| Performs as intended in specified use conditions | Non-clinical data support the safety and demonstrate that ARIA Radiation Therapy Management should perform as intended |
| Software safety and effectiveness | ARIA Radiation Therapy Management is as safe and effective as the predicate |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not applicable. The submission describes software verification and validation, not a test set of patient data.
- Data Provenance: Not applicable. Testing was based on software functionality and engineering requirements.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. No test set involving human interpretation or clinical ground truth was used.
4. Adjudication method for the test set:
- Not applicable. No test set requiring expert adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No data from animal studies or clinical tests have been included in this pre-market submission" and "no animal or clinical studies were conducted for the subject device."
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, in the context of software verification and validation, the performance of the software (algorithm) was assessed standalone against its functional and safety requirements. However, this is not a standalone diagnostic performance study as typically understood for AI/ML devices. The document states: "The non-clinical data support the safety of the software verification and validation demonstrate that ARIA Radiation Therapy Management should perform as intended in the specified use conditions."
7. The type of ground truth used:
- For the software verification and validation, the "ground truth" was the specified engineering requirements, functional specifications, and regulatory standards (e.g., IEC 62304, IEC 62366-1, ISO 13485, ISO 14971). There was no clinical or pathological ground truth used as no clinical studies were performed.
8. The sample size for the training set:
- Not applicable. This is a software management system, not a machine learning model that typically involves a "training set" in the AI/ML sense.
9. How the ground truth for the training set was established:
- Not applicable, as no training set (in the AI/ML context) was described.
{0}------------------------------------------------
April 10, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Varian Medical Systems Inc. % Peter Coronado Sr. Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
Re: K230699
Trade/Device Name: ARIA Radiation Therapy Management (v18.0) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: March 10, 2023 Received: March 13, 2023
Dear Peter Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Lora Lora D. D. Weidner -S Date: 2023.04.10 Weidner -S 13:47:48 -04'00'
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230699
Device Name ARIA Radiation Therapy Management
Indications for Use (Describe)
The ARIA Radiation Therapy Management product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Therapy Management also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains three blue circles arranged horizontally against a white background. The circles are evenly spaced and appear to be of the same size. The color of the circles is a light, bright blue, and they are solid in color.
Varian Medical Systems
3100 Hansen Way Palo Alto, CA 94304 650.493.4000
800.544.4636
varian.com
510(k) Summary
K230699
l. SUBMITTER
| Name and Address: | Varian Medical Systems, Inc.3100 Hansen WayPalo Alto, CA 94304 |
|---|---|
| Contact Person: | Peter J. CoronadoSr. Director, Regulatory Affairs+1 (650) 424-6320 (phone) +1 (650) 646-9200 (fax)submissions.support@varian.com |
| Date Prepared: | 7 April 2023 |
II. DEVICE
| Trade name: | ARIA Radiation Therapy Management, version 18.0 |
|---|---|
| Common name: | ARIA Radiation Therapy Management |
| Classification name: | Medical charged-particle radiation therapy system(21 CFR 892.5050) |
| Regulatory class: | Class II |
| Product code: | IYE |
III. PREDICATE DEVICE
Predicate Device: ARIA Radiation Therapy Management, version 16.1 MR3 (K221408)
No reference devices were used in this submission.
IV. DEVICE DESCRIPTION
The ARIA Radiation Therapy Management product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Therapy Management also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA Radiation Therapy Management supports the integration of all data and images in one central database including archiving and restoration. The different ARIA Radiation Therapy Management features support the visualization, processing, manipulation and management of all data and images stored in the system. Images can also be imported through using DICOM, the available image import filters or by means of film digitizers.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo is black and white.
V. INTENDED USE AND INDICATIONS FOR USE
The ARIA Radiation Therapy Management product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Therapy Management also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments.
(Same as predicate)
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The main changes in the subject device compared with the predicate device are as follows:
- . Support for Halcyon Gating plans and CBCTp mode
- MLC DB storage improvement
- CT Dose Index feature
- Display of Proton specific parameters
- . Updates to the RT Administration system
- O Support of the Proteus-One Treatment Delivery machine by IBA
- o Approval extension plug-in via Eclipse scripting API
- Updates to Offline Review which provides remote review of images
- o Removal of DLP display
- O Persistence of structure selection per patient
- Standardization of blending mode for gated images O
- DICOM updates
- O DICOM Transport Layer Security (TLS v.1.2)
- Cloud readiness: new DICOM attribute to allow referencing external prescriptions in DB o
- Infrastructure and architecture updates
- .NET 4.8 O
- O SQL Server 2019
- . FIPS Compliance
VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Software Verification and Validation Testing
Software verification and validation were conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management System standard and other FDA recognized consensus standards listed below. Test results showed that applicable requirements were met and assured that safeguards against hazards functioned properly.
The documentation in the submission was provided as recommended by the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). The level of concern is major because a failure or latent flaw in the software could directly result in serious injury or death to the patient or the operator.
No data from animal studies or clinical tests have been included in this pre-market submission.
Standards Conformance
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the logo is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo and text are both black.
ARIA Radiation Therapy Management conforms to the following regulatory standards including FDA recognized standards and references additional standards as applicable.
The devices conform with the following standards:
- IEC 62304:2006+A1:2015 Medical device software – Software lifecycle processes
- IEC 62366-1:2015 Medical devices – Part 1: Application of usability engineering to medical devices
- IEC 82304-1:2016 Health software – Part 1: General requirements for product safety
- IEC 62274:2005 Medical electrical equipment – Safety of radiotherapy record-and-verify systems
- IEC 61217:2011 Radiotherapy equipment - Coordinates, movements, and scales
ARIA RTM was designed and developed, including verification testing, within an established Quality System compliant to the following additional general (non-device-specific) standards:
- . EN ISO 13485:2016 Medical devices – Quality management systems – Requirements for regulatory purposes
- . EN ISO 14971:2019 Medical devices – Application of risk management to medical devices
- ISO 15223-1:2021 Medical devices Symbols to be used with medical device labels, labeling, and ● information to be supplied
- . EN ISO 20417:2021 – Information Supplied by the manufacturer of medical devices
- 21 CFR §820 - Quality System Regulation
CONCLUSIONS VIII.
Since the predicate device was cleared based only on the results of non-clinical testing, no animal or clinical studies were conducted for the subject device. The non-clinical data support the safety of the software verification and validation demonstrate that ARIA Radiation Therapy Management should perform as intended in the specified use conditions. ARIA Radiation Therapy Management is as safe and effective as the predicate that is currently marketed for the same intended use.
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.