K Number
K230660
Manufacturer
Date Cleared
2023-06-06

(88 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AviClear Laser System is indicated for the long-term treatment of mild to severe inflammatory acne vulgaris.

Device Description

The AviClear Laser System is an infrared diode laser device with a nominal wavelength of 1726 nm. Similar to other well-established infrared laser devices, the device uses a combination of the treatment spot size, beam characteristics, and tissue absorption and scattering coefficients at the output wavelength to deliver energy to tissues at depth. The wavelength of the laser energy, when combined with the pre, parallel (during energy delivery), and post cooling of epidermal and superficial dermal structures, causes selective heating of dermal tissue at different depths. The 1726 nm laser energy heats the chromophore, sebum within sebaceous glands, which results in controlled thermal injury of the sebaceous glands, the ultimate treatment target, thus reducing or eliminating sebum production and causing an improvement in acne vulgaris.

The diode laser and associated beam delivery optics, laser and electronics power supplies, control electronics, and cooling system are housed inside a console equipped with a touchscreen user interface. The laser treatment parameters are selected using the touchscreen.

The treatment handpiece has an integrated scanner for delivering treatment spot(s) in an operator-selected pattern; a temperature-controlled skincontact cooling window to provide thermal protection for the epidermis and superficial dermis and through which energy is delivered to the patient; and skin-contact pressure sensors that enable laser energy delivery when the system is in Ready mode and the footswitch is depressed.

AI/ML Overview

This document describes the regulatory clearance (K230660) for the AviClear Laser System. It is a 510(k) summary, which means the manufacturer is asserting their device is substantially equivalent to a legally marketed predicate device. In this case, the predicate device is a previous version of the AviClear Laser System (K213461). The current submission (K230660) appears to be primarily for minor design modifications to the handpiece and user interface, while leveraging clinical data from the prior clearance for efficacy and safety.

Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to the clinical study supporting the original clearance (K213461), which is then referenced to support the current substantial equivalence claim.

Here's the breakdown based on the provided text, focusing on the clinical study details:

1. A table of acceptance criteria and the reported device performance

The document primarily focuses on a single primary effectiveness objective and several secondary objectives rather than a predefined "acceptance criteria" table in the typical sense of a rigorous statistical hypothesis test with predefined alpha or beta. However, based on the primary effectiveness objective, we can construct the following:

Acceptance Criteria (Primary Effectiveness Objective)Reported Device Performance
Objective: Show that more than 50% of subjects enrolled were Responders who achieved Treatment Success.Primary Endpoint (12-weeks): Results for this primary endpoint are not explicitly reported in the provided text for K230660, but are stated to have resulted in FDA 510(k) clearance for K213461, implying the criterion was met. The document states: "The study's primary effectiveness objective was to show more that than 50% of subjects enrolled were Responders who achieved Treatment Success, where Treatment Success was defined as a subject with ≥50% fewer inflammatory acne lesions with respect to baseline."
Long-Term Effectiveness (52-weeks):Reported: At the 52-week post-final treatment follow-up visit, 91.5% of subjects achieved treatment success (p<0.001).

Secondary Effectiveness Analyses & Safety Performance:

AssessmentReported Performance
Improvement in Investigator's Global Assessment (IGA)By the 52-week follow-up visit, the number of subjects had increased to 66.2% for the "Clear or Almost Clear" skin categories.
Improvement from Baseline in Non-Inflammatory Lesion CountsReduced from 25% (p=0.001) at 12-week mark to 57.6% reduction from Baseline at the 52-week follow-up assessment (p<0.001).
Subject SatisfactionApproximately 70% of subjects reported they were either "Very likely" or "Likely" to have the laser treatment again at both the 12- and 52-week follow-up visits.
Treatment DiscomfortAll treatment sessions were well tolerated with an average discomfort score of 5.23 ± 1.15. No treatment sessions ended prematurely due to excessive treatment discomfort.
Serious Adverse Events (SAEs) & Unanticipated Adverse Device Effects (UADEs)No reports of SAEs and unanticipated adverse device effects of any severity (SUADEs/UADEs). No residual adverse events at the 52-week mark.
Common Adverse Events (at 12-weeks)- Mild, transient erythema (all subjects) - Edema (98.1% of subjects); self-resolved within hours to days - Mild acne purging/flare-up (45.2% of subjects); typically resolving with no medical intervention - Mild dryness (18.3% of subjects); typically resolving with topical skin moisturizers
Lack of Specific Adverse EventsNo reports of blistering, hyper- or hypopigmentation, or scarring adverse events in any subjects.

2. Sample size used for the test set and the data provenance

  • Sample Size: 104 subjects
  • Data Provenance:
    • Country of Origin: Not explicitly stated in this document but typically for FDA submissions, clinical trials are often conducted in the US or under ICH GCP guidelines.
    • Retrospective or Prospective: Prospective, multi-center, clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not explicitly state the number of experts or their specific qualifications. For acne studies, the "Investigator's Global Assessment (IGA)" is typically determined by a qualified dermatologist or clinical investigator. The primary effectiveness measure ("≥50% fewer inflammatory acne lesions") would likely be a count performed by study personnel or automated systems, validated by clinical examination.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any specific adjudication method for establishing ground truth, such as multi-reader consensus.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a laser system for acne treatment, not an AI-assisted diagnostic device that would typically involve human readers interpreting images.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a direct treatment device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for effectiveness was based on:

  • Clinical Assessment: Reduction in inflammatory acne lesions (counted), and Investigator's Global Assessment (IGA) scores.
  • Patient Reported Outcomes: Subject Satisfaction data.
  • Safety Data: Adverse event reporting and discomfort scores.

8. The sample size for the training set

Not applicable. This is a direct treatment device, not a machine learning algorithm that requires a "training set" in the computational sense. The "training set" for the device's development would be the pre-clinical research and engineering associated with its design.

9. How the ground truth for the training set was established

Not applicable for the same reason as point 8. The "ground truth" for the device's design would stem from scientific understanding of laser-tissue interaction and acne pathophysiology.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

June 6, 2023

Cutera, Inc. Amogh Kothare VP, Clinical and Regulatory Affairs 3240 Bayshore Blvd. Brisbane, California 94005

Re: K230660

Trade/Device Name: AviClear Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 10, 2023 Received: March 10, 2023

Dear Amogh Kothare:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230660

Device Name

AviClear Laser System

Indications for Use (Describe)

The AviClear Laser System is indicated for the long-term treatment of mild to severe inflammatory acne vulgaris.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K230660 510(k) Summary

This 510(k) Summary of safety and effectiveness for the AviClear Laser System is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) Summary.

Applicant:Cutera, Inc.
Address:3240 Bayshore Blvd., Brisbane, CA 94005
Contact Person:Amogh Kothareakothare@cutera.comm. (415) 690-6796o. (415) 657-5730f. (415) 715-3730
Preparation Date:March 10, 2022
Device Trade Name:AviClear Laser System
Common Name:Dermatology Laser
Classification Name:Powered Laser Surgical Instrument, GEX, 21 CFR 878.4810
Legally MarketedPredicate Device:AviClear Laser System (K213461)
Indications for Use:The AviClear Laser System is indicated for the long-term treatment of mildto severe inflammatory acne vulgaris.
Device Description:The AviClear Laser System is an infrared diode laser device with a nominalwavelength of 1726 nm. Similar to other well-established infrared laserdevices, the device uses a combination of the treatment spot size, beamcharacteristics, and tissue absorption and scattering coefficients at the

e output wavelength to deliver energy to tissues at depth. The wavelength of the laser energy, when combined with the pre, parallel (during energy delivery), and post cooling of epidermal and superficial dermal structures, causes selective heating of dermal tissue at different depths. The 1726 nm laser enerqy heats the chromophore, sebum within sebaceous glands, which results in controlled thermal injury of the sebaceous glands, the ultimate treatment target, thus reducing or eliminating sebum production and causing an improvement in acne vulgaris.

The diode laser and associated beam delivery optics, laser and electronics power supplies, control electronics, and cooling system are housed inside a console equipped with a touchscreen user interface. The laser treatment parameters are selected using the touchscreen.

The treatment handpiece has an integrated scanner for delivering treatment spot(s) in an operator-selected pattern; a temperature-controlled skincontact cooling window to provide thermal protection for the epidermis and superficial dermis and through which energy is delivered to the patient; and skin-contact pressure sensors that enable laser energy delivery when the system is in Ready mode and the footswitch is depressed.

{4}------------------------------------------------

Section 7 510(k) Summary

  • Performance Data: IEC 60601-1/A1:2012 Medical Electrical Equipment - Part 1: ● General Requirements for Safety
    • IEC 60601-1-6:2010/A1:2013 Medical electrical equipment -O Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
    • IEC 60601-2-22:2007/A1:2012 Medical Electrical Equipment - O Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories
    • IEC 60825-1:2014 Safety of laser products Part 1: Equipment o classification and requirements
    • IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Disturbances
    • Software Verification and Validation Testing ●
    • Biocompatibility testing of patient-contact materials according to ISO-● 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • Data for the clinical study was previously submitted under K213461, Results of Clinical Study: resulting in FDA 510k clearance for the treatment of mild to severe inflammatory acne vulqaris. This is a prospective, multi-center, clinical study to assess the safety and effectiveness of the Cutera AviClear Laser System for the treatment of mild to severe acne vulgaris. The study's primary effectiveness objective was to show more that than 50% of subjects enrolled were Responders who achieved Treatment Success, where Treatment Success was defined as a subject with ≥50% fewer inflammatory acne lesions with respect to baseline. Safety was assessed by a review of the totality of all reported adverse events and their resolutions, including assessments of treatment discomfort.

104 subjects, 57% female and 43% male, 16 to 40 years (avg. 22.2 ± 5.5), diagnosed with mild, moderate, and severe acne vulgaris were enrolled. Subjects received up to 3 treatments and were followed up through 52weeks post final treatment, with 12-weeks being primary end-point.

Details surrounding the Primary, Secondary Effectiveness as well as Safety Analysis have been outlined in the previous 510(K) summary submitted under K213461.

  • . Long Term Effectiveness Analysis: At the 52-week post-finaltreatment follow-up visit, 91.5% of subjects achieved treatment success (p<0.001); thereby showing durable and efficacious results from the 12week post-final-treatment follow-up visit.

Secondary Effectiveness Analyses: ●

Improvement in Investigator's Global Assessment

  • By the 52-week follow-up visit, the number of subjects had O increased to 66.2% for the Clear or Almost Clear skin categories.

{5}------------------------------------------------

Section 7 510(k) Summary

Improvement from Baseline in Non-Inflammatory Lesion Counts

  • The non-inflammatory lesion counts reduced from 25% (p=0.001) o at 12-week mark to 57.6% reduction from Baseline at the 52week follow-up assessment (p<0.001).
    Subject Satisfaction

  • Approximately 70% of subjects reported they were either "Very O likely" or "Likely" to have the laser treatment aqain at both the 12and 52-week follow-up visits.

  • Safety Variable Analysis: All treatment sessions were well tolerated ● with an average discomfort score of 5.23 ± 1.15. No treatment sessions ended prematurely due to excessive treatment discomfort. No reports of serious adverse events (SAEs) and unanticipated adverse device effects of any severity (SUADEs/UADEs).

There were no residual adverse events at the 52-week mark. Adverse events through 12-week follow-up can be found outlined in previous 510(K) summary submitted under K213461. Most commonly reported adverse events from the 12-week mark are summarized below:

  • As is expected with laser procedures, all subjects experienced o mild, transient erythema, while 98.1% of subjects experienced edema. Erythema self-resolved within minutes to hours, while edema self-resolved within hours to days.
  • 45.2% of subjects experienced mild acne purging / flare-up; O typically resolving with no medical intervention.
  • 18.3% of subjects experienced mild dryness; typically resolving O with application of topical skin moisturizers.
  • There were no reports of blistering, hyper- or hypopigmentation, O or scarring adverse events in any subjects.

As demonstrated by the results of the clinical study, the AviClear Laser System continues to demonstrate safe and effective performance at the oneyear follow-up, thus supporting the claim for long-term treatment of mild to severe inflammatory acne vulgaris.

The technical specifications of the previously cleared AviClear Laser System and the subject device are identical, as tabulated below. However, the following minor design modifications were made to the subject device:

  • . Slimmer handpiece body (no change to patient contact end of handpiece)
  • Updated touchscreen user interface ●

Comparison of the Proposed Device and Predicate Device:

{6}------------------------------------------------

Section 7 510(k) Summary

Technical Specification Comparison

AviClear Laser System(current submission)AviClear Laser SystemK213461
Wavelength1726 nm1726 nm
Max FluenceSingle pulse mode: 30 J/cm²Double pulse mode: 20 J/cm²Single pulse mode: 30 J/cm²Double pulse mode: 20 J/cm²
Max Pulse Energy5 J5 J
Pulse DurationUp to 50 msUp to 50 ms
Spot Size3 mm or 10 mm 7-spothexagonal array3 mm or 10 mm 7-spothexagonal array
Laser TypeDiodeDiode
Output ModeQuasi-CWQuasi-CW
User InterfaceTouchscreenTouchscreen
Treatment Beam ActivationFootswitch and handpiece contactsensors verifying firm and evencontact between cooling windowand skin is made and maintainedFootswitch and handpiece contactsensors verifying firm and evencontact between cooling window andskin is made and maintained
Delivery SystemOptical fiber handpieceOptical fiber handpiece
Skin Cooling0°C to 5°C, Sapphire window0°C to 5°C, Sapphire window
Aiming Beam650 nm650 nm
HandpieceNon-sterile, reusable, cleanableNon-sterile, reusable, cleanable

Conclusion: As demonstrated by the results of the clinical study, the AviClear Laser System is safe and effective for the requested indication for use. The clinical study also confirmed that there are no new concerns for safety or efficacy for the updated indication for use. The minor cosmetic design changes to the handpiece and touchscreen do not impact the technical specifications nor the safety and effectiveness of the device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.