(90 days)
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
Active in vitro and in Clinical Infections Against: Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates)
Gram-negative Bacteria Haemophilus influenzae
The BD BBL Sensi-Disc Lefamulin 20 µg (LMU-20) device is a 6 mm disc prepared by impregnating high quality absorbent paper with accurately determined amounts of Lefamulin. Discs are clearly marked on both sides with the code LMU-20. The code designates the agent Lefamulin (LMU) and the drug content (20 ug).
BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs are furnished in cartridges containing 50 discs each. The last disc in each cartridge is marked "X" and contains the drug as coded. BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs can be dispensed using a BD BBL Sensi-Disc Dispenser.
The provided document is a 510(k) Summary for the BD BBL Sensi-Disc Lefamulin 20µg (LMU-20), an antimicrobial susceptibility test disc. It details the device's intended use, performance characteristics, and comparison to a predicate device. However, it does not describe a study involving an AI/machine learning device or human readers. The acceptance criteria and performance data provided relate to the reproducibility and comparative performance of an antimicrobial susceptibility disc, which is a physical diagnostic tool for laboratory use, not a digital AI-based diagnostic system.
Therefore, many of the requested elements for describing the acceptance criteria and the study for an AI device (e.g., number of experts for ground truth, adjudication methods, MRMC studies, training set details) are not applicable to this document.
I will extract the relevant information regarding acceptance criteria and performance as presented for this specific device.
Device: BD BBL Sensi-Disc Lefamulin 20µg (LMU-20)
Device Type: Antimicrobial Susceptibility Test Disc (not an AI/machine learning device)
Purpose of Study: To demonstrate substantial equivalence to a legally marketed predicate device (HardyDisk AST Lefamulin 20µg (LMU20)) for determining antimicrobial susceptibility to Lefamulin.
Below is the information extracted and adapted based on the provided document, acknowledging that it is not for an AI device.
1. Table of Acceptance Criteria and Reported Device Performance
The document describes two main performance studies: Reproducibility and Method Comparison (Clinical and Challenge). The acceptance criteria for these are typically implied as meeting certain percentages of agreement or being within accepted ranges.
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Reproducibility | >95% agreement | Between Disc Lots: 100.00% (270/270) Across Readers: 98.89% (267/270) |
| Method Comparison | High Category Agreement (CA), low Major (MAJ) and Very Major (VMJ) errors compared to predicate device. | Overall Combined Performance (Clinical & Challenge Isolates): - Total Isolates Tested: 336 - Category Agreement (CA): 98.5% (331/336) - Very Major Errors (VMJ): 4 (adjusted potential VMJ error rate is 0% due to measurement differences) - Major Errors (MAJ): 1 |
| Specific Organisms: | ||
| Haemophilus influenzae | Combined CA: 96.4% (108/112); VMJ: 3, MAJ: 1 | |
| Staphylococcus aureus (methicillin-susceptible isolates) | Combined CA: 100.0% (112/112); VMJ: 0, MAJ: 0 | |
| Streptococcus pneumoniae | Combined CA: 99.1% (111/112); VMJ: 1, MAJ: 0 |
2. Sample Size Used for the Test Set and Data Provenance
-
Reproducibility Study:
- Sample Size: 15 organisms (5 Haemophilus influenzae, 5 Staphylococcus aureus, 5 Streptococcus pneumoniae), tested in triplicate with two disc lots on three separate days.
- Data Points: 270 data points (15 organisms x 2 disc lots x 3 days x 3 independent readers).
- Provenance: Conducted at one external site (country of origin not specified, but the applicant and regulatory body are US-based). The study appears to be prospective in nature for generating the performance data.
-
Method Comparison Study (Clinical and Challenge):
- Clinical Isolates: 270 isolates (90 Haemophilus influenzae, 90 Staphylococcus aureus (methicillin-susceptible), and 90 Streptococcus pneumoniae).
- Challenge Isolates: 66 isolates (22 Haemophilus influenzae, 22 Staphylococcus aureus (methicillin-susceptible), and 22 Streptococcus pneumoniae).
- Total Isolates: 336 (Clinical + Challenge).
- Data Provenance: Performed at one U.S. site. The isolates likely represent a collection of retrospective or prospectively acquired clinical and challenge strains.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This section is not applicable as the document describes a traditional antimicrobial susceptibility test disc, not an AI/machine learning device requiring expert review for image-based ground truth. The "ground truth" for this device would be the established zone diameter readings from the predicate device and the reference standard MIC (Minimum Inhibitory Concentration) method, interpreted against recognized susceptibility interpretive criteria (e.g., CLSI, FDA STIC).
- Reproducibility: Results were visually read by three independent readers. Their specific qualifications are not detailed, but generally, operators in such settings are trained laboratory personnel.
- Method Comparison: Three independent operators participated in reading of test results. Their specific qualifications are not detailed.
4. Adjudication Method for the Test Set
This is not explicitly detailed for defining the "ground truth" as it is typically understood in AI studies (e.g., in cases of disagreement among experts).
- For reproducibility, readings from three independent readers were compared against a "modal or median" zone diameter value.
- For method comparison, the "ground truth" or reference for comparison was the FDA-cleared predicate disk's results and the MIC reference method. Discrepancies (errors) were simply reported without mentioning an explicit adjudication process beyond the direct comparison.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. This type of study (MRMC comparative effectiveness study comparing human readers with AI vs without AI assistance) is not applicable to this device. The device is a diagnostic disc, not an AI system.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable in the context of an AI algorithm. The device's performance is inherently linked to human interpretation of zone diameters. The performance stated in the table (e.g., Reproducibility, Category Agreement) reflects the device's performance when used by human operators.
7. The Type of Ground Truth Used
The "ground truth" for evaluating this device's performance is based on:
- Reference Method: Comparison against an FDA-cleared predicate device (HardyDisk AST Lefamulin 20µg (LMU20)) utilizing a standardized agar diffusion test procedure (Kirby-Bauer).
- Interpretive Criteria: Use of FDA-recognized susceptibility test interpretive criteria (e.g., from CLSI M100, FDA STIC website) which define susceptible, intermediate, and resistant categories based on zone diameters or MICs.
- MIC Results: Where errors occurred, the document mentions that potential VMJ errors were compared to "historical MIC results," implying an underlying MIC reference method for establishing the true susceptibility category.
8. The Sample Size for the Training Set
Not applicable. This is a physical diagnostic disc, not an AI model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, there is no AI training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION".
June 7, 2023
Becton, Dickinson and Company Katherine Cicala Regulatory Affairs Specialist 7 Loveton Circle Sparks, Maryland 21152-0999
Re: K230651
Trade/Device Name: BD BBL Sensi-Disc Lefamulin 20μg (LMU-20) Regulation Number: 21 CFR 866.1620 Regulation Name: Antimicrobial Susceptibility Test Disc Regulatory Class: Class II Product Code: JTN Dated: March 8, 2023 Received: March 9, 2023
Dear Katherine Cicala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ribhi Shawar -S
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230651
Device Name BD BBL Sensi-Disc Lefamulin 20ug (LMU-20)
Indications for Use (Describe)
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
Active in vitro and in Clinical Infections Against: Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates)
Gram-negative Bacteria Haemophilus influenzae
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Summary Preparation Date: June 06, 2023
I Background Information
A 510(k) Number
B Applicant
BD Diagnostic Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, Maryland 21152 Establishment Registration Number: 1119779
C Proprietary and Established Name
BD BBL Sensi-Disc Lefamulin 20 ug (LMU-20)
D Regulatory Information
| Product Code | Classification | Regulation Section | Panel |
|---|---|---|---|
| JTN | Class II | 21 CFR 866.1620 -Antimicrobial SusceptibilityTest Disc | MI - Microbiology |
II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for Lefamulin Antimicrobial Susceptibility Test Disc.
B Measurand:
Lefamulin Disc 20 µg (LMU-20)
C Type of Test:
Antimicrobial Susceptibility Test Disc
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
{4}------------------------------------------------
B Indication(s) for Use:
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semiquantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
Active in vitro and in Clinical Infections Against:
Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates)
Gram-negative Bacteria Haemophilus influenzae
C Special Conditions for Use Statement(s):
- . Rx - For Prescription Use Only
- Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and . efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
- Limitations
- The current absence of resistant isolates precludes defining any results other than ● "Susceptible." Isolates yielding results other than "Susceptible" should be submitted to a reference laboratory for further testing.
D Special Instrument Requirements:
Not applicable.
IV Device/System Characteristics:
A Device Description:
The BD BBL Sensi-Disc Lefamulin 20 µg (LMU-20) device is a 6 mm disc prepared by impregnating high quality absorbent paper with accurately determined amounts of Lefamulin. Discs are clearly marked on both sides with the code LMU-20. The code designates the agent Lefamulin (LMU) and the drug content (20 ug).
BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs are furnished in cartridges containing 50 discs each. The last disc in each cartridge is marked "X" and contains the drug as coded. BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs can be dispensed using a BD BBL Sensi-Disc Dispenser.
{5}------------------------------------------------
B Principle of Operation:
A suitable therapeutic agent can be determined using filter paper discs impregnated with specified concentrations of antimicrobial agents placed on the surface of a suitable test medium. The test is performed by inoculating pure cultures of clinical isolates onto the test medium and placing the AST disc on the surface of the medium. The antibiotic within the disc diffuses into the agar. After incubation, the zones of inhibition around the discs are measured and compared against recognized zone diameter ranges for the specific antimicrobial agent/organism combinations being tested.
V Substantial Equivalence Information:
A Predicate Device Name:
HardyDisk AST Lefamulin 20μg (LMU20)
B Predicate 510(k) Number:
C Comparison with Predicate:
| Device &Predicate Device: | Device:K230651 | Predicate:K192326 |
|---|---|---|
| Device Trade | BD BBL Sensi-Disc Lefamulin | HardyDisk AST Lefamulin 20 |
| Name | 20 µg (LMU-20) | ug (LMU20) |
| General Device Characteristic Similarities | ||
| Regulation | 866.1620 | Same |
| Product Code | JTN | Same |
| Intended Use | See Indications for Use below. | HardyDisk AST Disks are usedfor semi-quantitative in vitrosusceptibility testing by the agardiffusion test procedure (Kirby-Bauer) of rapidly growing andcertain fastidious bacterialpathogens. Standardized methodsfor agar diffusion testing havebeen described forEnterobacteriaceae,Staphylococcus spp.,Pseudomonas spp.,Acinetobacter spp., Listeriamonocytogenes, Enterococcusspp., and by modifiedprocedures, Haemophilus spp.,Neisseria gonorrhoeae, N.meningitidis and Streptococcusspp., including Streptococcuspneumoniae. |
| Device & | Device: | Predicate: |
| Predicate Device: | K230651 | K192326 |
| Indications for Use | BD BBL Sensi-DiscAntimicrobial SusceptibilityTest (AST) Discs are used inthe semi-quantitative agardiffusion test method for invitro susceptibility testing.BD BBL Sensi-Disc LefamulinDisc 20 µg (LMU-20) can beused to determine susceptibilityto Lefamulin against thefollowing bacteria, as describedin the FDA-approved packageinsert for this antimicrobialagent.Active in vitro and in ClinicalInfections Against:Gram-positive BacteriaStreptococcus pneumoniaeStaphylococcus aureus(methicillin-susceptibleisolates)Gram-negative BacteriaHaemophilus influenzae | HardyDisk AST Disks are usedfor semi-quantitative in vitrosusceptibility testing by the agardiffusion test procedure (Kirby-Bauer) of rapidly growing andcertain fastidious bacterialpathogens. Standardized methodsfor agar diffusion testing havebeen described forEnterobacteriaceae,Staphylococcus spp.,Pseudomonas spp.,Acinetobacter spp., Listeriamonocytogenes, Enterococcusspp., and by modifiedprocedures, Haemophilus spp.,Neisseria gonorrhoeae, N.meningitidis and Streptococcusspp., including Streptococcuspneumoniae.Use of HardyDisk ASTLefamulin 20µg (LMU20) for invitro agar diffusion susceptibilitytesting is indicated when there isthe need to determine thesusceptibility of bacteria toLefamulin.HardyDisk AST Lefamulin atconcentration 20µg can be usedto determine the zone diameter(mm) of Lefamulin against thefollowing bacteria for whichLefamulin has been shown to beactive both clinically and invitro:Streptococcus pneumoniaeStaphylococcus aureus(methicillin-susceptible isolates)Haemophilus influenzae |
| AntimicrobialAgent | Lefamulin | Same |
| Antimicrobial | ||
| Agent | 20 µg | Same |
| Concentration | ||
| Device &Predicate Device: | Device:K230651 | Predicate:K192326 |
| Interpretation | The user will interpret the zonediameter according to theestablished interpretive criteriafor the drug. | Same |
| Methodology | Kirby-Bauer Disk DiffusionSusceptibility Test Protocolrequires the user to determinecategorical interpretations(S/I/R) using the measured zonediameters. | Same |
| ResultInterpretationMethod | Measurement of zone size. | Same |
| General Device Characteristic Differences | ||
| ManufacturingSpecifications | BD's specifications | Hardy Diagnostics'specifications |
{6}------------------------------------------------
{7}------------------------------------------------
VI Standards/Guidance Documents Referenced:
- CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 33rd ed. CLSI ● supplement M100. Clinical and Laboratory Institute; 2023.
- CLSI. Performance Standards for Antimicrobial Disk Susceptibility Tests. 13th ed. CLSI ● standard M02. Clinical and Laboratory Standards Institute; 2018.
- . CLSI. Methods for Antimicrobial Susceptibility Testing of Anaerobic Bacteria. 9th ed. CLSI standard M11. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
VII Performance Characteristics (if/when applicable):
Analytical Performance: A
1. Precision/Reproducibility:
Reproducibility was conducted at one external site using 15 organisms, tested in triplicate with two disc lots on three separate days using one lot of appropriate media (Mueller Hinton agar (MHA) for Staphylococcus aureus, Haemophilus Test Medium for Haemophilus influenzae, and MHA with 5% Sheep Blood for Streptococcus pneumoniae). Each test was visually read by three independent readers with results masked, resulting in 270 data points for evaluation (15 organisms x 2 disc lots x 1 media lot x 3 days x 3 independent readers = 270 data points).
The reproducibility study included 5 Haemophilus influenzae, 5 Staphylococcus aureus, and 5 Streptococcus pneumoniae. Reproducibility was calculated as the percent of results which were within ±3 mm difference in zone diameter comparing test results with the modal or median, if no mode is present, zone diameter value. Summary results between disk lots and across readers are shown in Table 1 below.
{8}------------------------------------------------
| Between Disc Lots | Across Readers | |||||
|---|---|---|---|---|---|---|
| Lot #1 | Lot #2 | All Lots | Reader #1 | Reader #2 | Reader #3 | All Readers |
| 100.00% | 100.00% | 100.00% | 96.67% | 100.00% | 100.00% | 98.89% |
| (135/135) | (135/135) | (270/270) | (87/90) | (90/90) | (90/90) | (267/270) |
Table 1: Reproducibility Summary
The reproducibility performance between disc lots and across readers is >95% and meets the acceptance criteria.
-
- Linearity:
Not applicable.
- Linearity:
-
- Analytical Specificity/Interference:
Not applicable.
- Analytical Specificity/Interference:
-
- Assay Reportable Range:
Not applicable.
- Assay Reportable Range:
-
- Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality Control (QC) Testing:
The CLSI-recommended quality control (QC) isolates, Haemophilus influenzae (ATCC 49247), Staphylococcus aureus (ATCC 25923), and Streptococcus pneumoniae (ATCC 49619), were tested a sufficient number of times (i.e., a minimum of 20 replicates per lot per reader). One Comparator disk lot and two BD disc lots were used. Each test was visually read by three independent readers, resulting in 129 data points (69 and 60 data points for Lot A and B, respectively) and 69 Comparator disk data points. The performance is shown in Table 2 below.
| QC Organism | Zone Diameter(mm) | BD Lot 11(N) | BD Lot 21(N) | Comparatordisk2(N) |
|---|---|---|---|---|
| Haemophilus influenzaeATCC 49247Expected Range:22-28 mm | 19 | |||
| 20 | ||||
| 21 | ||||
| 22 | 6 | |||
| 23 | 14 | |||
| 24 | 1 | 21 | ||
| 25 | 9 | 4 | 24 | |
| 26 | 33 | 30 | 4 | |
| 27 | 25 | 25 | ||
| 28 | 1 | 1 | ||
| 29 | ||||
| 30 | ||||
| 31 |
{9}------------------------------------------------
| QC Organism | Zone Diameter(mm) | BD Lot 1¹(N) | BD Lot 2¹(N) | Comparatordisk²(N) |
|---|---|---|---|---|
| 23 | ||||
| 24 | ||||
| 25 | 1 | |||
| 26 | 3 | |||
| 27 | 11 | 6 | 15 | |
| Staphylococcus aureus ATCC | 28 | 17 | 16 | 31 |
| 25923 | 29 | 30 | 31 | 21 |
| Expected Range: | 30 | 13 | 7 | |
| 26-32 mm | 31 | |||
| 32 | ||||
| 33 | ||||
| 34 | ||||
| 35 | ||||
| 16 | ||||
| 17 | ||||
| 18 | ||||
| 19 | ||||
| 20 | ||||
| Streptococcus pneumoniae | 21 | 1 | ||
| ATCC 49619 | 22 | 7 | ||
| 23 | 3 | 26 | ||
| Expected Range: | 24 | 21 | 16 | 24 |
| 19-27 mm | 25 | 33 | 28 | 8 |
| 26 | 10 | 16 | 3 | |
| 27 | 2 | |||
| 28 | ||||
| 29 | ||||
| 30 |
ATCC = American Type Culture Collection
'Two BD disc lots were tested (Lot A and Lot B).
Two BD disc lots were tested (Lot A and Lot B).
2One Comparator disk lot was tested.
The BD disc QC performance is > 95% and is acceptable.
Inoculum Density Check:
Colony counts were conducted for all QC and reproducibility isolates, as well as 14.0% of clinical isolates. All were within the expected range.
-
- Detection Limit:
Not applicable.
- Detection Limit:
-
- Assay Cut-Off:
{10}------------------------------------------------
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
The BD BBL Sensi-Disc Lefamulin 20 ug (LMU-20) was compared with an FDA cleared disk of the same antimicrobial, mass/concentration, and content. The study was conducted at one testing site. Three independent operators participated in reading of test results with isolates evenly distributed to mimic testing at multiple sites. Testing was performed with one lot of discs from each test manufacturer (BD and comparator FDA-cleared disk), utilizing appropriate media following the method outlined in CLSI M02 and M100. An equivalent procedure to CLSI M11 was utilized for conducting colony counts.
Clinical:
Clinical testing was performed at one U.S. site with both the BD BBL Sensi-Disc Lefamulin disc and the comparator FDA cleared disk using a total of 270 clinical isolates including 90 Haemophilus influenzae, 90 Staphylococcus aureus (methicillin-susceptible isolates), and 90 Streptococcus pneumoniae.
Challenge:
Challenge testing was performed at one U.S site. A total of 66 challenge isolates were tested which included 22 Haemophilus influenzae, 22 Staphylococcus aureus (methicillinsusceptible isolates), and 22 Streptococcus pneumoniae.
Performance results for the total 336 clinical and challenge isolates are shown in Table 3. Performance of 39 MRSA isolates is not included in this table due to the lack of FDArecognized breakpoints for Lefamulin when testing MRSA.
| Total | CA# | CA% | S (#) | NS (#) | VMJ* | MAJ | ||
|---|---|---|---|---|---|---|---|---|
| Haemophilus influenzae | ||||||||
| Clinical | 90 | 89 | 98.9 | 90 | 0 | 0 | 1 | |
| Challenge | 22 | 19 | 86.4 | 13 | 9 | 3 | 0 | |
| Combined | 112 | 108 | 96.4 | 103 | 9 | 3 | 1 | |
| Staphylococcus aureus (methicillin-susceptible isolates) | ||||||||
| Clinical | 90 | 90 | 100.0 | 90 | 0 | 0 | 0 | |
| Challenge | 22 | 22 | 100.0 | 10 | 12 | 0 | 0 | |
| Combined | 112 | 112 | 100.0 | 100 | 12 | 0 | 0 | |
| Streptococcus pneumoniae | ||||||||
| Clinical | 90 | 89 | 98.9 | 89 | 1 | 1 | 0 | |
| Challenge | 22 | 22 | 100.0 | 19 | 3 | 0 | 0 | |
| Combined | 112 | 111 | 99.1 | 108 | 4 | 1 | 0 | |
| Overall | ||||||||
| Clinical | 270 | 268 | 99.3 | 269 | 1 | 1 | 1 | |
| Challenge | 66 | 63 | 95.5 | 42 | 24 | 3 | 0 | |
| Combined | 336 | 331 | 98.5 | 311 | 25 | 4 | 1 |
Table 3: Overall Performance of the BD BBL Sensi-Disc Lefamulin Disc vs. Comparator disk
- Due to the lack of an interpretive category other than susceptible and considering that a 3 mm disk zone diameter is equivalent to one MIC doubling dilution, because all potential VMJ errors had disk zone diameter measurement differences
{11}------------------------------------------------
| Total | CA# | CA% | S (#) | NS (#) | VMJ* | MAJ | |
|---|---|---|---|---|---|---|---|
| ≤3 mm compared to the FDA-cleared comparator disk and were ≤ an MIC doubling dilution-equivalent compared to the |
historical MIC result, the adjusted potential VMJ error rate is 0%.
CA - Category Agreement S - Susceptible isolates
NS - Nonsusceptible isolates
MAJ - major errors VMJ - very major errors
Category Agreement (CA) is when the BD result interpretation agrees exactly with the FDA cleared comparator result interpretation. The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Haemophilus influenzae group (Table 3) was acceptable with 96.4% (108/112) CA. There were 3 very major errors and 1 major error.
The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Staphylococcus aureus (methicillin-susceptible isolates) group (Table 3) was acceptable with 100.0% (112/112) CA. There were no very major or major errors.
The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Streptococcus pneumoniae group (Table 3) was acceptable with 99.1% (111/112) CA. There was 1 very major error and no major errors.
As required under 511A(b)(2)(C)(ii)(1) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the 'Warnings and Precautions' section in the device labeling to address testing of non-indicated species:
"Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved."
Resistant Isolates:
A total of 336 clinical and challenge isolates were tested when the BD BBL Sensi-Disc Lefamulin was compared to the FDA cleared comparator disk. However, an insufficient number of resistant isolates were available for testing. The following limitation was added in the device labeling:
"The current absence of resistant isolates precludes defining any results other than "Susceptible." Isolates yielding results other than "Susceptible" should be submitted to a reference laboratory for further testing."
-
- Matrix Comparison
Not applicable.
- Matrix Comparison
C Clinical Studies:
-
- Clinical Sensitivity:
{12}------------------------------------------------
Not applicable
-
- Clinical Specificity:
Not applicable
- Clinical Specificity:
-
- Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
- Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
The FDA-identified interpretive criteria for Lefamulin are listed in Table 4.
| Table 4: Interpretive Categories and Breakpoints for Lefamulin | ||
|---|---|---|
| Minimum InhibitoryConcentration (µg/mL) | Disk Diffusion(zone diameter in mm) | |||||
|---|---|---|---|---|---|---|
| Pathogen | S | I | R | S | I | R |
| Staphylococcus aureus (methicillin-susceptible isolates) | $ \u22640.25 $ | - | - | $ \u226523 $ | - | - |
| Streptococcus pneumoniae | $ \u22640.5 $ | - | - | $ \u226519 $ | - | - |
| Haemophilus influenzae | $ \u22642 $ | - | - | $ \u226518 $ | - | - |
1 According to the FDA STIC Website, which recognizes CLSI M100-Ed33.
Note: The current absence of resistant isolates precludes defining any results other than "Susceptible". Isolates yielding MIC results other than "Susceptible" should be submitted to a reference laboratory for further testing.
S = Susceptible
I = Intermediate
R = Resistant
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/developmentresources/antibacterial-susceptibility-test-interpretive-criteria). The protocol outlined the specific procedures and acceptance criteria that Becton, Dickinson, and Company (BD Diagnostic Systems) intends to use to evaluate the BD BBL Sensi-Disc Lefamulin 20µg (LMU-20) when revised breakpoints for Lefamulin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, BD Diagnostic Systems will update the Lefamulin 20ug device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
§ 866.1620 Antimicrobial susceptibility test disc.
(a)
Identification. An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).