(59 days)
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
The Hubly Electric Drill is a battery powered cranial drill. It is single-use disposable and provided ethylene oxide sterilized. The drill is designed to create optimal burr holes, in the emergency room, at the bedside, or in the operating room. The system is designed to streamline bedside intracranial access and facilitate the treatment of emergent conditions in adult patients.
The device is a battery-powered hand drill which can be used one-handed using either hand. The drill is trigger-activated after removal of the battery pull tab. The drill has a single speed and turns off when the trigger is released. The drill bit has depth indicators at 5 and 10mm depth, which the physician may use to visually gauge depth of penetration while drilling.
The drill also has an auto-stop feature which detects when the bit breaks through the inner table of the skull and immediately stops the drill. The device has an LED indicator which indicates to the user (green) when they are applying enough force and (red) when the drill stops.
The drill features mechanical plunge prevention with a tapered stainless steel drill may be reactivated any number of times using the trigger if the physician desires.
The provided text describes performance testing for the Hubly Electric Drill. However, it does not contain specific numerical acceptance criteria or numerical reported device performance in a table format, nor does it detail a study proving the device meets specific acceptance criteria in the way requested.
Instead, the document outlines various tests conducted to demonstrate the device's substantial equivalence to a predicate device and its safety and effectiveness.
Here's an attempt to extract and synthesize the information based on the request, noting where information is not explicitly provided in the text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a table with explicit numerical acceptance criteria and corresponding reported device performance values. It generally states that the device "meets all design specifications and requirements" for bench tests and "all in vivo requirements were met" for the animal study.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify a numerical sample size for any of the performance tests (bench testing, animal study).
- Data Provenance:
- Bench Testing: Conducted by Hubly Inc. (implied, as they are the applicant).
- Animal Study: Conducted in a GLP (Good Laboratory Practice) compliant setting using a sheep model. The location/country is not specified.
- Retrospective/Prospective: The tests described (bench, animal) are prospective studies conducted to support the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
- This information is not provided in the document. The studies described are performance tests and an animal study, not studies relying on expert review for ground truth establishment.
4. Adjudication Method for the Test Set
- This information is not applicable/provided. The described tests are instrumental/procedural performance tests and an animal study, not studies involving human reader interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not performed on human subjects." The studies focused on bench performance and an animal model.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- The Hubly Electric Drill is a physical medical device (a drill), not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable. The "software verification testing" was conducted on firmware responsible for the automatic stopping feature, but this is an embedded system, not a standalone AI application. The software implements "additional safety features" and has a "Moderate" level of concern.
7. Type of Ground Truth Used
- Bench Testing: Ground truth was implicitly based on engineering design specifications and requirements (e.g., battery life, depth of penetration, trigger performance).
- Animal Study: Ground truth was based on observational outcomes in the sheep model, specifically demonstrating that "Users (Test Device Evaluators) could safely use the Hubly Electric Drill for intracranial access without damaging the dura or brain tissue."
8. Sample Size for the Training Set
- This information is not applicable/provided. The device is a physical drill with embedded firmware, not a machine learning model that requires a training set in the conventional sense. The "software" mentioned is firmware, which is typically developed through traditional software engineering processes, not trained on data.
9. How the Ground Truth for the Training Set Was Established
- This information is not applicable/provided for the reasons stated in point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
May 4, 2023
Hubly Inc. Julie Byars Quality Lead 750 Warrenville Rd. Suite 303 Lisle, Illinois 60532
Re: K230619
Trade/Device Name: Hubly Electric Drill (H100) Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE Dated: March 1, 2023 Received: March 6, 2023
Dear Julie Byars:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Digitally signed by Adam D. Adam D. Pierce -S Date: 2023.05.04 Pierce -S 08:55:18 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230619
Device Name Hubly Electric Drill
Indications for Use (Describe)
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
| Type of Use (Select one or both, as applicable) | |
|---|---|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
A. Device Information:
| Category | Comments |
|---|---|
| Sponsor: | Hubly, Inc.Casey Grage750 Warrenville Rd. Suite 303Lisle, IL 60532(844) 482-5942 |
| Correspondent Contact Information: | Julie Byars750 Warrenville Rd. Suite 303Lisle, IL 60532901-848-4010 |
| Device Common Name: | Powered Cranial Drill |
| Device Regulation & Name: | 21 CFR 882.4310 - Powered simple cranial drills,burrs, trephines and their accessories |
| Classification & Product Code: | Class II, HBE |
| 510(k) Number: | K222391 |
| Device Proprietary Name: | Hubly Electric Drill |
Predicate Device Information:
| Predicate Device: | Phasor Drill |
|---|---|
| Predicate Device Manufacturer: | Biotex Inc. |
| Predicate Device Common Name: | Powered Cranial Drill |
| Predicate Device Premarket Notification#: | K161704 |
| Predicate Device Classification & Name: | 21 CFR 882.4310 - Powered simple cranial drills,burrs, trephines and their accessories |
| Predicate Device Classification &Product Code: | Class II, HBE |
B. Date Summary Prepared
May 4, 2023
C. Description of Device
The Hubly Electric Drill is a battery powered cranial drill. It is single-use disposable and provided ethylene oxide sterilized. The drill is designed to create optimal burr holes, in the emergency room, at the bedside, or in the operating room. The system is designed to streamline bedside intracranial access and facilitate the treatment of emergent conditions in adult patients.
{4}------------------------------------------------
Hubly, Inc. Hubly Electric Drill
The device is a battery-powered hand drill which can be used one-handed using either hand. The drill is trigger-activated after removal of the battery pull tab. The drill has a single speed and turns off when the trigger is released. The drill bit has depth indicators at 5 and 10mm depth, which the physician may use to visually gauge depth of penetration while drilling.
The drill also has an auto-stop feature which detects when the bit breaks through the inner table of the skull and immediately stops the drill. The device has an LED indicator which indicates to the user (green) when they are applying enough force and (red) when the drill stops.
The drill features mechanical plunge prevention with a tapered stainless steel drill may be reactivated any number of times using the trigger if the physician desires.
D. Indications for Use
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
E. Comparison of the Technological Characteristics
| Application DeviceHubly Electric Drill - K222391 | Predicate DevicePhasor Drill - K161704 | Impact on SubstantialEquivalence | |
|---|---|---|---|
| Company | Hubly, Inc. | Biotex, Inc. | N/A |
| RegulationNumber | 21 CFR 882.4310 | 21 CFR 882.4310 | Identical |
| Product Code | HBE - Class II | HBE - Class II | Identical |
| Intended Use &Indications foruse | The Hubly Electric Drill is asingle-use, sterile, disposabledevice intended for use onadult patients duringneurosurgical procedures fordrilling of cranial bone. | The Phasor Drill is a sterile,single-use, disposabledevice intended for use onadult patients duringneurosurgical proceduresfor drilling of cranial bone. | Identical |
| Technology | |||
| Power Source | Battery Power (removable) | Battery Power (removable) | Identical |
| Automatic stop | Auto-stop | None | No impact - Additionalsafety feature |
| Mechanicalplungeprevention | Tapered bit | Thumbscrew depth stop | No impact |
| Bit Material | Stainless Steel (316) | Stainless Steel (440A) | No impact |
| Drill activation | Trigger - On/Off | Button - On/Off | Identical |
| Drill rotation | Forward | Forward & Reverse | No impact |
| Flutecharacteristics | 2 Straight flutes | 2-3 Twist flutes | No impact – Straightflutes remove moreeasily without reverserotation. |
{5}------------------------------------------------
F. Summary of Supporting Data
The Hubly Electric Drill has the same intended use, patient population and classification as the predicate. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Sterile packaged for single-use, and disposable
- Operates on battery power (batteries removable for disposal)
- . Two state operation, single speed
- Employ mechanical plunge prevention mechanisms for safety
- . Plastic drill housing with stainless steel drill bit(s)
The following technical characteristics of the subject device differ from the predicate device:
- Use of tapered drill bit for mechanical plunge prevention (vs thumbscrew depth stop)
- Drill bit diameter & length, flute type
- . Addition of firmware based automatic stopping feature
- . Forward-only drill functionality
- EO sterilization (vs gamma)
The Hubly drill bit is larger in diameter and tapers to the tip for mechanical plunge prevention, whereas the predicate has a thumbscrew depth stop for plunge prevention. The Hubly drill bit has straight flutes and tapers and thus is easily removed from the burr hole without reverse functionality. The applicant device has an additional firmware-based automatic stopping safety feature, and for this reason is EO sterilized.
The proposed device does not differ from the predicate in any way which would negatively impact the safety or effectiveness of the device.
G. Discussion of Performance Testing
The following performance data were provided in support of the substantial equivalence determination.
BIOCOMPATIBILITY TESTING
The drill bit is the only patient-contacting material used in the Hubly device. 316 Stainless Steel certified to ASTM F899 was selected for its known biocompatibility with human tissues and fluids. A biological risk assessment was conducted for the device. It considers raw material, manufacturing processing, and full endpoint biocompatibility testing performed, and determines full compliance of the device with ISO 10993-1.
{6}------------------------------------------------
Hubly, Inc. Hubly Electric Drill
ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILTY (EMC)
Electrical safety and EMC testing were conducted on the Hubly device. The system complies with the IEC 60601-1 standard for safety and the IEC 60601-4-2 and 60601-4-2 standards for EMC.
SOFTWARE VERIFICATION TESTING
Software verification was conducted on the firmware responsible for the Hubly Electric Drill's automatic stopping feature. The device has an equivalent expected safety profile compared to the standard of care without utilizing software mitigations. The software implements additional safety features compared to the current standard of care.
The Hubly Electric Drill software is initially classified as a Class B software system according to IEC 62304 Section 4.3a based on the possible hazards to which the software component can contribute. For this reason, the resulting possible harm due to failure of software is considered non-serious. Based on the level of concern decision process (as described in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices), the level of concern for the Hubly Electric Drill has been determined to be "Moderate".
BENCH TESTING
The following benchtop performance tests were conducted. Results demonstrated that the drill meets all design specifications and requirements.
- Performance - battery/continuous use
- Depth of Penetration, bone thickness variation
- . Bit securement
- . Trigger performance
- Basic Safety and Essential Performance
- . EMC Testing (Basic Safety, Essential Performance, Electromagnetic disturbances)
ANIMAL STUDY
In the GLP animal study conducted, design validation and summative usability evaluation were conducted. A sheep model was selected for its close approximation of the characteristics of the human skull. This study demonstrated that Users (Test Device Evaluators) could safely use the Hubly Electric Drill for intracranial access without damaging the dura or brain tissue. Additionally, all in vivo requirements were met to support Usability Evaluation for the Hubly electric Drill.
CLINICAL STUDIES
Clinical testing was not performed on human subjects.
H. Conclusion
The predicate device was cleared based on bench testing alone. The performance and design validation testing conducted on the Hubly Electric Drill on the bench and in animals demonstrates that it performs equivalent to the predicate device that is currently marketed for the same intended use. The proposed device adds additional control and safety features, and hardware and
{7}------------------------------------------------
software verification and validation demonstrate that the Hubly drill will perform as safe and effective as the predicate device.
§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.
(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).