K Number
K230362
Device Name
Planar LED mate
Date Cleared
2023-09-11

(213 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Planar LED mate is a prescription-use only device intended for removal of unwanted body and/or facial hair.

Device Description

The Planar LED mate is a prescription-use only device intended for removal of unwanted body and/or facial hair. It utilizes a diode LED as a light source (780-850 nm). lt works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its emission activation is by finger switch.

The device contains a capacitive sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light. The device is for single-person use only.

The proposed Planar LED mate is equipped with a light depilation function that is based on the theory of selective photo thermolysis. There is abundant melanin in the hair follicle and hair stem. The melanin is distributed amidst the cells between the hair ball substrate and it can be transferred to the structure of hair stems (such as medulla, cortex, and hair cuticle). The Near infrared light can precisely target the melanin and apply the depilation treatment selectively. After the melanin absorbs Near infrared light energy, the temperature rises, in this way, the surrounding hair follicle tissues are damaged, and the hairs are removed.

The device has 5 level Output Intensity.

AI/ML Overview

This submission for the Planar LED mate (K230362) does not include a study that proves the device meets specific acceptance criteria related to its performance for hair removal. Instead, it relies on non-clinical tests to demonstrate substantial equivalence to predicate devices and adherence to safety and performance standards.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria for hair removal efficacy and corresponding device performance. The focus is on demonstrating that the device's technical characteristics and safety comply with relevant standards and are similar to predicate devices.

Acceptance Criteria CategorySpecific Acceptance Criteria (Not explicitly stated for hair removal efficacy)Reported Device Performance
BiocompatibilityNo cytotoxicityNo Cytotoxicity
No evidence of sensitizationNo evidence of sensitization
No evidence of irritationNo evidence of irritation
Electrical SafetyComply with ANSI/AAMI ES60601-1, IEC 60601-1-11Comply with ANSI/AAMI ES60601-1, IEC 60601-1-11
Electromagnetic Compatibility (EMC)Comply with IEC 60601-1-2Comply with IEC 60601-1-2
Laser SafetyComply with IEC 60601-2-83, IEC 62471Comply with IEC 60601-2-83, IEC 62471
Software Verification & ValidationAdherence to "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate Level of Concern)Software Verification and Validation Testing was conducted per guidance.
Usability & Label ComprehensionUsability and label comprehension evaluatedUsability and label comprehension study conducted.
Energy OutputAccuracy of energy output (Specific numerical criteria not provided)Performance Testing for Energy Output Accuracy conducted.
Hair Removal EfficacyNot explicitly stated in terms of quantitative criteria or success rates.The submission does not include clinical study results to demonstrate specific hair removal efficacy. It states the device is intended for "removal of unwanted body and/or facial hair" based on the theory of selective photothermolysis, similar to predicate devices.

2. Sample size used for the test set and the data provenance

  • For hair removal efficacy: Not applicable. No clinical test set involving human subjects for hair removal efficacy was conducted or reported in this submission.
  • For non-clinical tests (biocompatibility, electrical safety, EMC, laser safety, software, usability, energy output): The document does not specify the sample sizes for these tests (e.g., number of units tested, number of participants in usability study). The tests were likely performed in a laboratory setting, but the provenance of the data itself (e.g., from which specific test labs) is not detailed beyond adherence to international standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable for hair removal efficacy. As no clinical study was performed, there was no ground truth for this purpose established by medical experts for a test set.
  • For non-clinical safety and performance: Ground truth for these aspects is generally established by adherence to recognized international standards (e.g., ISO, IEC, ANSI/AAMI) for testing methodologies and acceptable limits, rather than by individual expert consensus on a test set. The experts involved would be those who performed the standard-compliant testing and validated the results. Their specific qualifications are not detailed in this summary.

4. Adjudication method for the test set

  • Not applicable for hair removal efficacy. No clinical test set was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a hair removal system, not an AI-assisted diagnostic or therapeutic imaging device that would typically involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a hardware-based light energy device for hair removal, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For hair removal efficacy: Not applicable. No clinical study was performed, so no "ground truth" for efficacy (e.g., based on clinical outcomes or hair counts) was established.
  • For device safety and technical performance: The "ground truth" is based on established engineering and biocompatibility standards (e.g., IEC 60601-1, ISO 10993, IEC 62471) which define acceptable limits and performance criteria for medical devices. Compliance with these standards serves as the benchmark.

8. The sample size for the training set

  • Not applicable. This device is a hair removal system and does not appear to involve machine learning algorithms that would require a "training set."

9. How the ground truth for the training set was established

  • Not applicable. As no training set is mentioned, this question is not relevant.

In summary: The provided 510(k) summary for the Planar LED mate (K230362) focuses on demonstrating substantial equivalence through a comparison of technological characteristics with predicate devices and comprehensive non-clinical testing against established safety and performance standards. Crucially, it explicitly states, "No clinical study is included in this submission," meaning there is no direct clinical data or study to demonstrate the device's hair removal efficacy against specific acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 11, 2023

Shenzhen Leaflife Technology Co., Ltd Cheng Qiang Regulatory Affairs Manager 4F, Bldg. C, JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang Dist., Shenzhen, Guangdong 518116 China

Re: K230362

Trade/Device Name: Planar LED mate Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 9, 2023 Received: February 10, 2023

Dear Cheng Qiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Tanisha L. Hithe -2023.09.11 Hithe -S 13:30:31 -04'00' Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230362

Device Name Planar LED mate

Indications for Use (Describe)

The Planar LED mate is a prescription-use only device intended for removal of unwanted body and/or facial hair.

Type of Use ( Select one or both, as applicable )
------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is written in a stylized font with the letters in blue and green. There is a green leaf above the letters. Below the company name is the company name written in Chinese.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

SUBMITTER l.

Shenzhen Leaflife Technology Co., Ltd 4F, Bldg. C, JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang Dist., Shenzhen, China

Contact Person: Cheng Qiang

Position: Regulatory Affairs Manager

Phone: 086-(0)17875910506

Fax: 086-0755-27215592

Primary Contact Person:Cheng Qiang
Regulatory Affairs Manager
Shenzhen Leaflife Technology Co., Ltd
Tel: 086-0755-27216609
Fax: 086-0755-27215592
Phone: 086-(0)17875910506
Email: cq@leaflife.cn
Secondary Contact Person:Vini Zeng
Regulatory Affairs Specialist
Shenzhen Leaflife Technology Co., Ltd
Tel: 086-0755-27216609
Fax: 086-0755-27215592

Email: zengvini@gmial.com

Date Prepared:

06/09/2023

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, with the word "LEAFLIFE" in blue, stylized font. Above the word is a green arc with a leaf at the end, and below the word is a blue arc. Underneath the word "LEAFLIFE" are two Chinese characters.

II. PROPOSED DEVICE

Trade Name:Planar LED mate
Model(s):LH-LNIRS
Common Name:Powered Laser Surgical Instrument
Classification Name:Laser surgical instrument for use in general and plasticsurgery and in dermatology (21 CFR 878.4810)
Regulation Class:II
Product Code:GEX
Review Panel:General & Plastic Surgery

III. PREDICATE DEVICE

Predicate device (Primary)
510(k) Number:K142845
Device Name:SILKPRO Laser Hair Removal System
Manufacturer:Wuhan Lotuxs Technology Company, Ltd.
Predicate device (1#)
510(k) Number:K222316
Device Name:IPL Hair Removal Device
Manufacturer:Shenzhen Goodwind Technology Development CO., LTD
Predicate device (2#)
510(k) Number:K220103
Device Name:LED Therapy Device
Manufacturer:Shenzhen Leaflife Technology Co., Ltd

The predicates have not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Planar LED mate is a prescription-use only device intended for removal of unwanted body and/or facial hair. It utilizes a diode LED as a light source (780-850 nm). lt works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its emission activation is by finger switch.

The device contains a capacitive sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light. The device is for single-person use only.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, and it features a stylized leaf. The word "LEAFLIFE" is written in a modern font, with the letters connected to each other. There is also some Chinese text below the word "LEAFLIFE".

The proposed Planar LED mate is equipped with a light depilation function that is based on the theory of selective photo thermolysis. There is abundant melanin in the hair follicle and hair stem. The melanin is distributed amidst the cells between the hair ball substrate and it can be transferred to the structure of hair stems (such as medulla, cortex, and hair cuticle). The Near infrared light can precisely target the melanin and apply the depilation treatment selectively. After the melanin absorbs Near infrared light energy, the temperature rises, in this way, the surrounding hair follicle tissues are damaged, and the hairs are removed.

The device has 5 level Output Intensity.

V. INTENDED USE

The Planar LED mate is a prescription-use only device intended for removal of unwanted body and/or facial hair.

vi. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Planar LED mate (LH-LNIRS) emits light energy to reduce hair growth, although compared to the predicate devices there may be differences in wavelengths, pulse durations, treatment spot sizes, and energy intensities. Please refer to the following table for details:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for LeafLife. The logo is in blue and green, and it features the word "LEAFLIFE" in a stylized font. There is a green leaf above the "F" in "LEAFLIFE", and a blue curved line below the word. Below the word "LEAFLIFE" are two Chinese characters.

ItemProposed DevicePredicate DeviceK142845 (Primary)Predicate Device (1#)K222316Predicate Device (2#)K220103Remark
Device namePlanar LED mateSILKPROIPL Hair Removal DeviceLED Therapy Device/
Product modelLH-LNIRS/ZHFIPL-II, ZHF-IPL-IIILM-LNIRA, LM-LNIRB/
K numberK230362K142845K222316K220103/
Product codeGEXOHTOHTOHTSame
Classificationregulation21 CFR 878.481021 CFR 878.481021 CFR 878.481021 CFR 878.4810Same
ltemProposed DevicePredicate DeviceK142845 (Primary)Predicate Device (1#)K222316Predicate Device (2#)K220103Remark
Intended UseThe Planar LED mate is aprescription-use only deviceintended for removal ofunwanted body and/or facialhair.SILKPRO is an over-thecounter device intended foradjunctive use with shavingfor hair removal sustainedwith periodic treatments.SILKPRO is also intended forpermanent reduction in hairregrowth defined as along-term, stable reductionin hair counts following atreatment regime.The IPL Hair RemovalDevice is anover-the-counter deviceintended for removal ofunwanted body hair.The LED Therapy Device isintended for hair removal,permanent hair reductionon all skin types(Fitzpatrick skin type I-VI),including tanned skin.Permanent hair reductionis defined as thelong-term, stablereduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after thecompletion of a treatmentregime.Same
Prescription useor notPrescription useOver-The-Counter UseOver-The-Counter UsePrescription useSame
Pulsing controlFinger switchFinger switchFinger switchFinger switchSame
Light TypeDiode LEDDiode laserIPLDiode LEDSimilar
LightWavelength780-850 nm810 nm510-1100nm780-850 nmSimilar
ItemProposed DevicePredicate DeviceK142845 (Primary)Predicate Device (1#)K222316Predicate Device (2#)K220103Remark
Spot size8mm×8mm9mm×9mmZHF-IPL-II: 3.6cm²ZHF-IPL-III: 4.1cm²(17mm*22mm) 3.74cm²(LM-LNIRA)Similar
OutputIntensity9.5-25J/cm²5J/cm², 10J/cm², 15J/cm²,20J/cm², 25J/cm²ZHF-IPL-II: 2.64.4J/cm²ZHF-IPL-III: 2.43.7J/cm²1-100J/cm²Similar
Pulse Duration108-320ms/7.2~10.8ms3-400msSimilar
Power SupplyAC 100-240V~ 50/60 HzAC 100~240V 50/60 HzAC 220240V 50/60 HzorAC 100120V 50/60 HzSimilar

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Leaflife. The logo is a combination of text and a stylized leaf design. The text "LEAFLIFE" is written in a modern, sans-serif font, with the letters in blue and green. Above the text is a green leaf design, which is curved and stylized. Below the text are two Chinese characters in gray.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is a combination of text and a graphic. The text "LEAFLIFE" is written in a stylized font, with the top half of the letters in green and the bottom half in blue. Above the "F" is a green leaf, and to the right of the leaf are some green dots. Below the text is the Chinese translation of the name.

Discussion

The Planar LED mate (LH-LNRS) device, the K222316, devices are handheld devices, while the K220103 is a larger console device. The handheld designs allow them to be more easily transported and stored by users who can administer their own hair removal treatments. The wavelength range, pulse durations, and output intensities, of the Planar LED mate are similar to those the K142845 and K222316 devices, although difference may lead to some differences in the results seen with over-the-counter devices such as K142845 (diode laser) and K221316 (PL). The wavelength range characteristics of the Planar LED mate (LH-LNIRS) are considered to be essentially the same as the K220103 device, and while the maximum pulse duration are less, the Planar LED mate (LH-LNIRS) device's output characteristics are considered to be acceptable for this prescription-use only device for the intended use of hair removal.

{9}------------------------------------------------

Image /page/9/Figure/0 description: The image shows the logo for LEAFLIFE. The logo is blue and green, and it features a stylized leaf. The word "LEAFLIFE" is written in blue, with a green line running through the middle of the letters. Above the word is a green leaf with a green arc. Below the word are two Chinese characters.

ItemProposed DevicePredicate DeviceK142845 (Primary)Predicate Device (1#)K222316Predicate Device (2#)K220103Remark
CytotoxicityNo CytotoxicityNo CytotoxicityNo Cytotoxicity/Same
SensitizationNo evidence of sensitizationNo evidence ofsensitizationNo evidence ofsensitization/Same
IrritationNo evidence of irritationNo evidence of irritationNo evidence of irritation/Same
Electrical SafetyComply with ANSI/AAMIES60601-1, IEC 60601-1-11Comply with ANSI/AAMIES60601-1Comply with ANSI/AAMIES60601-1, IEC60601-1-11Comply with IEC60601-1.Same
EMCComply with IEC 60601-1-2Comply with IEC60601-1-2Comply with IEC60601-1-2Comply with IEC60601-1-2Same
Laser SafetyComply with IEC60601-2-83, IEC 62471Comply with IEC60601-2-22, IEC 60825Comply with IEC60601-2-83, IEC 62471Comply with IEC60601-2-57, IEC 62471Similar

{10}------------------------------------------------

VII. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:

ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ANSI/AAMI A2:2010/(R)2012, Medical electrical equipment — Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

IEC 60601-2-83, 2019 Medical electrical equipment - Part 2-83: Requirements for the basic safety and essential performance of home light therapy equipment.

IEC 62471:2006, Photobiological safety of lamps and lamp systems.

ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)

ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)

Performance Testing for Energy Output Accuracy.

Software Verification and Validation Testing was conducted per "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", and the level of concern was determined to be Moderate for the proposed device.

Usability and label comprehension study.

VIII. CLINICAL TEST CONCLUSION

No clinical study is included in this submission.

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for LEAFLIFE. The logo is composed of the word "LEAFLIFE" in a stylized font, with the top half of the letters in green and the bottom half in blue. Above the "L" is a green leaf. To the right of the word "LEAFLIFE" are two Chinese characters in gray.

IX. CONCLUSIONS

The Planar LED mate (LH-LNIRS) device and predicate devices emit pulsed light to reduce hair growth, and the information provided in this premarket notification supports that the device is substantially equivalent and that it can be used safely and effectively as a prescription-use only device intended for removal of unwanted body and/or facial hair.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.