K Number
K230194
Device Name
HF Surgery Generator 400KHZ
Date Cleared
2023-03-22

(57 days)

Product Code
Regulation Number
878.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurgically.
Device Description
The HF SURGERY GENERATOR 400KHZ is used for electrosurgical cutting and coagulation of living human tissue. For this purpose, electrical energy from the power supply network is transformed into high frequency current which allows this surgical property. The device offers a multitude of different profiled current forms which are optimized for the different surgical requirements. It is equipped for monopolar and bipolar cutting and coagulation in micro- and macrosurgical operations.
More Information

No reference devices were used in this submission.

No
The summary describes a standard electrosurgical generator with different current forms, but there is no mention of AI or ML in the device description, intended use, or performance studies. The "Dialog" function is mentioned, but without further context, it does not indicate AI/ML.

Yes.
The device is used for cutting and coagulating tissue electrosurgically, indicating it is intended for a therapeutic purpose.

No

This device is used for cutting and coagulating tissue electrosurgically, which is an interventional or therapeutic function, not a diagnostic one.

No

The device description explicitly states it generates high frequency electrical current for cutting and coagulating tissue, which requires hardware components to produce and deliver this energy. The performance studies also include validation of electromagnetic compatibility, electrical safety, and performance on tissue, all indicative of a hardware device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: The intended use is for "generating high frequency electrical current for cutting and coagulating tissue electrosurgically." This is a surgical procedure performed on the patient, not a test performed on a sample taken from the patient.
  • Device Description: The description reinforces the surgical nature of the device, stating it's used for "electrosurgical cutting and coagulation of living human tissue."
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue samples, etc.), detecting specific analytes, or providing diagnostic information based on laboratory testing.

IVD devices are designed to perform tests on samples taken from the human body to provide information for the diagnosis, monitoring, or treatment of diseases or conditions. This device is a surgical tool used directly on the patient's tissue.

N/A

Intended Use / Indications for Use

The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurgically.

Product codes (comma separated list FDA assigned to the subject device)

GEI

Device Description

The HF SURGERY GENERATOR 400KHZ is used for electrosurgical cutting and coagulation of living human tissue. For this purpose, electrical energy from the power supply network is transformed into high frequency current which allows this surgical property. The device offers a multitude of different profiled current forms which are optimized for the different surgical requirements. It is equipped for monopolar and bipolar cutting and coagulation in micro- and macrosurgical operations.
The HF SURGERY GENERATOR 400KHZ is delivered non-sterile. The devices are reusable and do require cleaning and reprocessing during their use-life, i.e., users are required to clean and disinfect the device before every application and before returning for repairs. The HF SURGERY GENERATOR 400KHZ contains software. The software is classified as a Major Level of Concern. The HF SURGERY GENERATOR 400KHZ is not intended to contact the patient directly or indirectly. The HF SURGERY GENERATOR 400KHZ consists of a metal housing which does not have any patient contact.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

The product is intended for adult patients.

Intended User / Care Setting

This product is exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance tests were performed to validate the electromagnetic compatibility and electrical safety, performance on tissue (animal study), performance based on waveform comparison and the tissue study (bench testing), performance testing due to different software functions, changes in power consumption, cooling method, and "Dialog" function (bench testing), and transportation validation. Non-clinical performance testing, including Electromagnetic Compatibility, Electrical Safety, and Software Verification and Validation, demonstrated that the safety and performance of the device is equivalent to the predicate device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K193591

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

No reference devices were used in this submission.

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

March 22, 2023

Richard Wolf Medical Instruments Corporation Mike Mcandrew US Head of Regulatory - OA/OC 353 Corporate Woods Parkway Vernon Hills, Illinois 60061

Re: K230194

Trade/Device Name: HF Surgery Generator 400KHZ Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: January 23, 2023 Received: January 24, 2023

Dear Mike Mcandrew:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.03.22
15:31:45 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) To be determined

Device Name HF SURGERY GENERATOR 400KHZ

Indications for Use (Describe)

The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurgically.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in a dark red color. To the right of the text is a logo featuring a stylized dog standing next to a vertical line with a small object at the top and a paw print at the bottom, all enclosed in a rounded square.

510(k) Summary 5

Submitter l

Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, IL 60061

Phone: 847-913-1113, ext. 380 Mobile: 847-636-0887 Fax: 847-913-1489 E-Mail: MMcAndrew@richardwolfusa.com

Contact Person, Mike McAndrew, US Head of Regulatory – QA/QC Date Prepared: March 21, 2023

Legal Manufacturer

Richard Wolf GmbH Pforzheimer Straße 32 75438 Knittlingen

ll Device

| Product
Specification | Regulation
Number and | Product
Code | Device
Class |

--------------------------------------------------------------------------------------

Table 5-1: Device classification

| Brand name | Trade name | Model Number | Product
classification
name | Regulation
Number and
Name | Product
Code | Device
Class | Review
Panel |
|------------|-----------------------------------|--------------|----------------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------|-----------------|-----------------------------------|
| 2260 | HF SURGERY
GENERATOR
400KHZ | 2260003 | Electrosurgical
, Cutting &
Coagulation &
Accessories | 878.4400
Electrosurgical
cutting and
coagulation
device and
accessories | GEI | Class II | General
and Plastic
Surgery |
| N/A | FOOTSWITCH 2
PEDALS L 5M | 2260021 | | | | | |

Document 001 510(k) Summary

4

Image /page/4/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in red. To the right of the words is a logo of a dog with a rod through its mouth, inside of a rounded square.

| N/A
HF CONNECTION
CABLE BIPO L 5M | N/A
HF CONNECTION
CABLE BIPO L 3M |
|-----------------------------------------|-----------------------------------------|
| 8108253 | 8108233 |

��� Predicate Device

Trade name of Predicate Device: ARC 400 510(k) Number: K193591 Requlatory Class: Class II Product Code: GEI Manufacturer: BOWA-Electronics GmbH & Co. KG

The predicate has not been subject to a design-related recall. No reference devices were used in this submission.

Subject Device Description 5.1

5.1.1 Device Identification

Table 5-2: Device identification

| Brand
name | Model
Number | Trade name | Package unit |
|---------------|-----------------|-------------------------------|--------------|
| 2260 | 2260003 | HF SURGERY GENERATOR 400KHZ | 1 |
| N/A | 2260021 | FOOTSWITCH 2 PEDALS L 5M | 1 |
| N/A | 8108233 | HF CONNECTION CABLE BIPO L 3M | 1 |
| N/A | 8108253 | HF CONNECTION CABLE BIPO L 5M | 1 |

5.1.2 Device characteristics

Delivered sterile / non-sterile

The HF SURGERY GENERATOR 400KHZ is delivered non-sterilization is not required.

Single use / reusable

The devices are reusable and do require cleaning and reprocessing during their use-life, i.e., users are required to clean and disinfect the device before every application and before returning for repairs. Methods of cleaning and reprocessing are detailed in the Instruction for Use.

5

Image /page/5/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in a dark red color. To the right of the text is a logo featuring a stylized image of a wolf's head with a nail through it, enclosed in a rounded square.

Software

The HF SURGERY GENERATOR 400KHZ contains software. The software is classified as a Major Level of Concern.

Materials with patient contact

The HF SURGERY GENERATOR 400KHZ is not intended to contact the patient directly or indirectly.

5.1.3 Brief written description of the Device

The HF SURGERY GENERATOR 400KHZ is used for electrosurgical cutting and coagulation of living human tissue. For this purpose, electrical energy from the power supply network is transformed into high frequency current which allows this surgical property. The device offers a multitude of different profiled current forms which are optimized for the different surgical requirements. It is equipped for monopolar and bipolar cutting and coagulation in micro- and macrosurgical operations.

Image /page/5/Figure/8 description: The image shows a Richard Wolf HF-GENERATOR 2260. The device has a screen that displays the settings for different modes, including "M" and "B". The screen also shows that the device is currently "ON" and set to "Laparoscopy - Standard" mode, with options for "Mode" and "Menu".

Figure 05-1: Richard Wolf HF SURGERY GENERATOR 400KHZ

5.1.4 Materials of Use

The HF SURGERY GENERATOR 400KHZ consists of a metal housing which does not have any patient contact.

5.2 Indications for Use

Statement

The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurqically.

6

Image /page/6/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, black font above the word "WOLF", which is in a larger, red font. To the right of the text is a logo featuring a dog's head next to a vertical line with a small hand at the bottom, all enclosed in a rounded square.

User

This product is exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.

Patient population

The product is intended for adult patients.

The patient group intended for the application of the medical product herein described is not limited with regard to ethnicity, gender, body height, Before use, the attending physician must make sure that the product in view of its dimensions or settings can be used safely in the patient.

Explanation on differences to the predicate device

In comparison to the predicate device, the usage of the subject device is further restricted to adults. However, the full indication of use for the subject device is covered by the predicate device and reducing the scope does not lead to new questions regarding safety and effectiveness.

Comparison of Technological Characteristics with the Predicate Device 5.3

5.3.1 Overview Table

Table 5-3: Comparison of technical characteristics

ltemSubject devicePredicate devicesComparison
General information
Product nameHF SURGERY
GENERATOR 400KHZARC 400 Electrosurgical
UnitN/A
Brand name2260N/AN/A
Type No.2260003900400N/A
510(k) numberTBD / subject deviceK193591N/A
ManufacturerRichard Wolf GmbHBOWA-electronics GmbH
& Co. KGN/A
Product CodeGEIGEISame
Regulation
Number878.4400878.4400Same
Indications for
UseThe product is used for
generating high frequency
electrical current for
cutting and coagulating
tissue electrosurgically.The HF device is intended
exclusively for the
generation of electrical
power for monopolar and
bipolar cutting and
coagulation on tissue
structures in surgical
operations.Equivalent
ContraindicationsContraindications directly
related to the product are
presently unknown. If
medical contraindications
for the mentioned
indications are known, the
use of the products is not
permitted.
On the basis of the
patient's general condition
the doctor in charge must
decide whether or not the
planned use is possible.Do not use the HF device
if, in the opinion of an
experienced physician or
according to current
professional literature,
such use would endanger
the patient, due for
example to the general
condition of the patient, or
if other contraindications
are present.Equivalent
Prescription/
over-the-counter-
useRX onlyRX onlySame

7

Image /page/7/Picture/0 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, blocky, red font. To the right of the text is a logo featuring a stylized drawing of a wolf's head with a vertical line through it, enclosed in a rounded square.

ItemSubject devicePredicate devicesComparison
Sterile/Non-sterile
deviceNon-sterile deviceNon-sterile deviceSame
Single use/
ReusableReusable deviceReusable deviceSame
Sterilization
methodNon-sterile deviceNon-sterile deviceSame
ReprocessingCleaning and disinfection
solution/wipes for low
level should be based on
alcohol. Examples for
suitable substances for
cleaning and disinfection
of surfaces:
• Propanol
• Ethanol
• Didecyldimethyla
mmoniumchloride
• Dimethylbenzyla
mmoniumchlorideApply the cleaning agent
and disinfectant.
Wipe down the HF device
only with cleaning agents
and disinfectants that are
approved in the country of
use for surface cleaning.Equivalent
Maintenance
intervals12 months12 monthsSame
Operating
ConditionAmbient temperature +10
°C to +40 °C
Relative humidity 30% to
75% (non-condensing)
Atmospheric pressure 700
hPa to 1060 hPaTemperature: +10 °C to
+40 °C
Relative humidity: 30 to
75%, non-condensing
Atmospheric pressure:
700 to 1600 hPa
Operating altitude (max.)
4000 m above sea levelEquivalent
Storage and
Transport
ConditionsAmbient temperature -20
°C to +60 °C
Relative humidity 10% to
90% (non-condensing)
Atmospheric pressure 700
hPa to 1060 hPaTemperature: -20 °C to
+50 °C
Relative humidity 0 to 75
%, non-condensing
Atmospheric pressure:
500 to 1600 hPaEquivalent
Operational modes / settings
Mode: Monopolar cutting
Modes1212Same
Max. Power400 W (at 500 Ω)400 W (at 200 Ω)Same
Output Frequency403 - 500 kHz350 kHzDifferent
Max. Voltage
Output3150 Vp1600 VpDifferent
Crest Factor1.4 - 4.41.5, 3.5Different
Wave FormsSinusoidal Constant
Sinusoidal Modulated
Sinusoidal AlternatingSinusoidal Constant
Sinusoidal Modulated
Sinusoidal Alternating
Cut/Coag/Pause PhasesEquivalent
Mode: Monopolar coagulation
Modes159Different
Max. Power200 W (at 350 Ω)250 W (at 500 Ω)Different
Output Frequency375 - 500 kHz350 kHzDifferent
Max. Voltage
Output5000 Vp5000 VpEquivalent
Crest Factor1.4 - 7.61.6 - 7.4Different
Wave FormsSinusoidal Constant
Sinusoidal Modulated
Pulse ModulatedSinusoidal Constant
Sinusoidal Modulated
Pulse ModulatedSame
ItemSubject devicePredicate devicesComparison
Modes84Different
Max. Power300 W (at 100 Ω)400W (at 75 Ω)Different
Output Frequency355 - 381 kHz350 kHzDifferent
Max. Voltage750 Vp500 VpDifferent
Output
Crest Factor1.4 - 2.31.5 - 1.6Different
Wave FormsSinusoidal ConstantSinusoidal ConstantSame
Mode: Bipolar coagulation
Modes99Equivalent
Max. Power250 W (at 75 Ω)350 W (at 25 Ω)Different
Output Frequency338 - 381 kHz350 kHzDifferent
Max. Voltage425 Vp550 VpDifferent
Output
Crest Factor1.4 - 1.71.5 - 3.8Different
Wave FormsSinusoidal ConstantSinusoidal ConstantSame
Pulse ModulatedPulse Modulated
Technological characteristics
StandardIEC 60601-1:2005 (19-4)IEC 60601-1: 2005 (19-4)Same
complianceIEC 60601-1-2:2014 (19-8)IEC 60601-1-2: 2014 (19-8)
IEC 60601-2-2:2017 (6-389)IEC 60601-2-2:2017 (6-389)
Voltage [V]100-240 V230 VEquivalent
Frequency [Hz]50/60 Hz50/60 HzSame
Power45 without HF output85 power consumption inEquivalent
consumption [VA]650 at max. output powerstandby mode
/ Current rating975 maximum power
[A]consumption with 300 W
HF output power
Protection classIISame
to EN/IEC 60601-
1;
(UL 2601-1/CSA
C22.2 No.601.1 -
for USA)
Protection againstCF applied part,
defibrillator-proofCFSame
electric shock
Degree ofIPX0IP 21Equivalent
protection against
liquids
Weight [kg]10.2 kg (22.4 lbs)approx. 12.5 kgDifferent
Dimensions300 x 160 x 402 mm430 x 180 x 475 mmDifferent
MaterialMetal & plasticsMetal & plasticsEquivalent
Single-UseNoNoSame
Light SourceNoNoSame
Battery OperatedNoNoSame
AC PoweredYesYesSame
Software version1.4.31.0.0.6.17.9Different
USB PortYesYesSame
PotentialYesYesSame
equalization
connector
LAN (Ethernet)
NetworkYesYesSame
ItemSubject devicePredicate devicesComparison
Foot switch
socket connectorYesYesSame
Control elements
/ user interfaceTouchscreenTouchscreenSame
Materials with patient contact
Materials with
direct patient
contactN/A - no materials with
direct patient
contactN/A - no materials with
direct patient
contactSame
Materials with
indirect patient
contactN/A - no materials with
indirect patient
contactN/A - no materials with
indirect patient
contactSame
Instruction for use
VersionGA-A339 / en / US / V0.6900-
400_IFU_V1.0.0.6.17.9_2
0140-S2-20130604-ENDifferent

8

Image /page/8/Picture/1 description: The image shows the logo for Richard Wolf. The words "RICHARD" and "WOLF" are stacked on top of each other, with "RICHARD" in black and "WOLF" in red. To the right of the text is a square with rounded corners containing a stylized image of a wolf's head and a medical instrument. The logo is simple and modern, with a clear focus on the company name.

9

Image /page/9/Picture/0 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, bolder, red font. To the right of the text is a logo featuring a stylized wolf head with a nail through it, enclosed in a rounded square.

5.3.2 Discussion

The subject device and the predicate device share the same fundamental technology. They both convert electrical energy from the grid into a high-frequency current which provides the surgical property. For that purpose, monopolar and bipolar high frequency current modes were provided to the corresponding socket module. The fundamental frequencies of about 400 kHz provided by the subject and the predicate device are comparable and substantial equivalent. Both generators feature a maximum power output of 400 W.

The general and technical characteristics that differ between the predicate and the subject device are differences in PCB layouts, software including the graphical user interface, number of output sockets and other technical characteristics such as power consumption and weight, dimensions. These differences do not question safety and effectiveness of the subject devices, several performance tests were performed, such as:

  • a. Validation of the electromagnetic compatibility and electrical safety (Section 17 -
    Electromagnetic Compatibility and Electrical Safety)

  • b. Validation of the performance on tissue (Section 19 Performing Testing Animal)

  • Validation of the performance based on waveform comparison and the tissue study (Section C. 18 - Performance Testing Bench)

  • ರ. Performance testing due to the different software functions, the change in power consumption, cooling method, and the "Dialog" function (Section 18 - Performance Testing Bench)

  • e. Transportation validation (Section 18 Performance Testing Bench)

Performance Testing 5.4

  • 5.4.1 Sterilization N/A

5.4.2 Biocompatibility

N/A

10

Image /page/10/Picture/1 description: The image contains the words "RICHARD WOLF" in a logo format. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, bolder, red font. To the right of the words is a logo of a wolf with a medical device. The logo is contained within a rounded square.

5.4.3 Electromaqnetic Compatibility and Electrical Safety

The predicate device and the device covered by this registration are classified in protection class I according to 60601-1. They both fulfil the electrical safety requirements of IEC 60601-2-2 and IEC 60601-1-2. Based on that, the devices are considered to be comparable in terms of electromagnetic compatibility and electrical safety.

5.4.4 Performance and Operational testing

The efficacy and safety of the HF SURGERY GENERATOR 400KHZ is documented by the verification and validation testing which confirms that the product meets all the requirements and specifications for overall design, basic safety, essential performance, and that the design inputs and specifications are met.

રે. રે Conclusions

The HF SURGERY GENERATOR 400KHZ has the same intended use and field of application as the legally marketed predicate device named in this submission.

The same performance tests were concluded for the subject device as for the predicate device. Therefore, no comparison testing was performed. Non-clinical performance testing, including Electromagnetic Compatibility. Electrical Safety, and Software Verification and Validation. demonstrated that the safety and performance of the device is equivalent to the predicate device. Differences in characteristics described in Section 12.2 do not raise any new questions regarding the safety and effectiveness of the HF SURGERY GENERATOR 400KHZ compared to the predicate device.

The HF SURGERY GENERATOR 400KHZ is substantially equivalent to the legally marketed predicate device ARC 400 (K193591).