K Number
K230131
Date Cleared
2023-02-15

(29 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ReShape Calibration Tubes™ are indicated for use in gastric surgical procedures to decompress the stomach. drain and remove gastric fluid, irrigate, and act as a sizing guide.

Device Description

The ReShape Calibration Tubes™ is a flexible gastric tube designed to be used in gastric and bariatric surgical procedures and is available in three larger sizes. The trade name has been changed Gastrointestinal Boundary Identifier (GIBI HD ™) to ReShape Calibration Tubes ™ for these three (3) sizes.

The ReShape Calibration Tubes™ provide visible and tactile delineation of the gastroesophageal (GE) junction, and its location relative to the esophageal hiatus and antrum of the stomach. The device provides the ability to decompress the stomach, drain and remove gastric fluid and act as a sizing guide (Figure 1).

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called "ReShape Calibration Tubes™". The purpose of this submission is not to introduce a new device with novel performance criteria, but rather to establish substantial equivalence to a previously cleared predicate device due to a change in device name, minor edits to the Directions For Use (DFU), and the addition of a contraindication.

Therefore, the typical structure for describing acceptance criteria and a study proving a device meets them (as would be expected for a new device with performance claims) is not directly applicable here. This document primarily focuses on demonstrating that the changes made do not alter the established safety and effectiveness of the device as previously cleared.

Based on the provided text, here's how to address your request:

Key Takeaway: The device's performance was not re-evaluated in a new study for this 510(k) submission. Its performance has been established through the clearance of its predicate device (Gastrointestinal Boundary Identifier (GIBI HD ™) Calibration Tube, K221898). The current submission leverages the established performance of the predicate device to demonstrate substantial equivalence for the modified device.


1. A table of acceptance criteria and the reported device performance

Since this submission is based on substantial equivalence to a predicate device and no new performance data was generated for this specific submission, there isn't a direct "table of acceptance criteria and reported device performance" as you might find for a de novo submission or a device with new performance claims.

Instead, the acceptance criteria are implicitly that the ReShape Calibration Tubes™ are substantially equivalent to the predicate device (GIBI HD ™ Calibration Tube, K221898) across all relevant characteristics. The "performance" is considered to be the same as the predicate.

Here's how we can conceptualize it from the document:

Characteristic/Acceptance Criteria (for Substantial Equivalence to Predicate)Reported Device Performance (Same as Predicate)
Intended Use (same as predicate)Same
Principle of Operation (same as predicate)Same
Device Design (same as predicate)Same
Materials of Construction (same as predicate)Same
Usability (established via bench studies for predicate)Same (no new studies for this submission)
Safety (established via bench studies for predicate)Same (no new studies for this submission)
Effectiveness (established via bench studies for predicate)Same (no new studies for this submission)
Product Code (KNT)Same
Regulatory Class (Class II)Same
Regulation Number (21 C.F.R. 876.5980)Same
Regulation Name (Gastroenterology-Urology)Same
Outer Diameter / French Size (nominal, distal tip, deflated balloon)Same across all sizes
Tubing (Dual lumen)Same
Distal Side Holes (3 aspiration holes)Same
Distal Tip (Molded tip with 12 aspiration eyelets)Same
Connector for Suction (Adapter for room suction included)Same
Balloon + Inflation Valve (Inflation capacity ≥ 100 cc min)Same
Tubing Material (Silicone)Same
Markings (Indication marks at 30, 35, 40, 45, 50, 55, and 60 cm)Same
Sterility (Non-sterile, disposable, single patient use)Same
Shelf Life (2 years)Same

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "Bench testing was not conducted and remains same as the predicate device cleared under K221898." Therefore, there is no new test set, sample size, or data provenance information provided in this submission. The basis for substantial equivalence relies on the established performance of the predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. As no new testing or clinical studies were conducted for this submission, no new ground truth was established by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As no new testing or clinical studies were conducted for this submission, no adjudication method was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a calibration tube used in surgical procedures, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical device (calibration tube), not an algorithm or AI.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For this submission, the "ground truth" is effectively the established safety and effectiveness of the predicate device (GIBI HD ™ Calibration Tube, K221898) based on prior regulatory clearance. The current submission's "ground truth" for its acceptance is that its modifications do not alter its substantial equivalence to that predicate device. The document does not detail how the predicate's original performance data was established (e.g., specific ground truth for its bench studies).

8. The sample size for the training set

Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This device is a physical medical device, not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 15, 2023

ReShape Lifesciences Dov Gal VP of Regulatory, Quality, & Clinical 1001 Calle Amanecer San Clemente, CA 92673

Re: K230131

Trade/Device Name: ReShape Calibration Tubes Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: KNT Dated: January 16, 2023 Received: January 17, 2023

Dear Dov Gal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the

{1}------------------------------------------------

Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the text "Je An -S" in black font. The text is arranged horizontally, with "Je" on the left, "An" in the middle, and "-S" on the right. There is a faint, light blue watermark in the background that is difficult to discern.

Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K230131

Device Name ReShape Calibration TubesTM

Indications for Use (Describe)

ReShape Calibration Tubes™ are indicated for use in gastric surgical procedures to decompress the stomach. drain and remove gastric fluid, irrigate, and act as a sizing guide.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Confidential

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for ReShape Lifesciences. The logo features a stylized orange circle that is not fully complete. The text "ReShape" is written in a combination of orange and purple, with "Re" in orange and "Shape" in purple. Below "ReShape" is the word "Lifesciences" in purple.

510(k) Summary

I. Basic Information

510(k) OwnerReShape Lifesciences1001 Calle AmanecerSan Clemente CA 92673Phone: (844) 937-7374Establishment Registration No: 3013508647
Contact PersonDov Gal, VP of Regulatory, Quality, & Clinical
Date of summaryJanuary 16, 2023
Trade NameThe ReShape Calibration Tubes™
Common NameIntroduction/Drainage Catheter
ClassificationClass II, KNT
Classification NameGastrointestinal Tube and Accessories
Predicate DeviceReShape Lifesciences Gastrointestinal Boundary Identifier (GIBI HD ™)Calibration Tube, K221898

II. DEVICE DESCRIPTION

The subject device is the same as the Gastrointestinal Boundary Identifier (GIBI HD ™) Calibration Tube, which is also manufactured by our company, ReShape Lifesciences Inc., and was cleared under 510(k) K221898. The reason for the submission is the addition of a contraindication to the Directions for Use (DFU). Other changes described in this submission are the device name and minor edits to the DFU.

The ReShape Calibration Tubes™ is a flexible gastric tube designed to be used in gastric and bariatric surgical procedures and is available in three larger sizes. The trade name has been changed Gastrointestinal Boundary Identifier (GIBI HD ™) to ReShape Calibration Tubes ™ for these three (3) sizes.

The ReShape Calibration Tubes™ provide visible and tactile delineation of the gastroesophageal (GE) junction, and its location relative to the esophageal hiatus and antrum of the stomach. The device provides the ability to decompress the stomach, drain and remove gastric fluid and act as a sizing guide (Figure 1).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a clear, flexible tube, likely a medical device such as a feeding tube or catheter. The tube is coiled and has a blue clamp attached to it. One end of the tube has a bulbous tip with multiple holes, while the other end has a connector with a cap.

Figure 1: ReShape Calibration Tubes™ (Representative figure of all Models: B-2032, B-2040)

III. INDICATIONS FOR USE

ReShape Calibration Tubes™ are indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide.

IV. BASIS FOR SUBSTANTIAL EQUIVALENCE

Substantial equivalence of the ReShape Calibration Tubes™ to the predicate device (ReShape Lifesciences Gastrointestinal Boundary Identifier (GIBI HD ™) Calibration Tube, K221898) was already established through an evaluation of the indications for use, principle of operation, device design, materials of construction, and an assessment of usability, safety, and effectiveness via bench studies.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH V. THE PREDICATE DEVICE

A comparison of the ReShape Calibration Tubes and the predicate device is provided in Table 1.

FeaturePredicate DeviceReShape Lifesciences GastrointestinalBoundary Identifier (GIBI HD ™) CalibrationTube, K221898Subject DeviceReShape Calibration TubesEffect onSubstantialEquivalence
Product CodeKNTSameNone
Regulatory ClassClass IISameNone
Regulation Number21 C.F.R. 876.5980SameNone
Regulation NameGastroenterology-UrologySameNone
Table 1: Comparison of Subject Device and Predicate Device
--------------------------------------------------------------------

{5}------------------------------------------------

FeaturePredicate DeviceReShape Lifesciences GastrointestinalBoundary Identifier (GIBI HD ™) CalibrationTube, K221898Subject DeviceReShape Calibration TubesEffect onSubstantialEquivalence
Indications for UseThe GIBI HD™ is indicated for use ingastric and bariatric surgicalprocedures to decompress thestomach, drain and remove gastricfluid, irrigate, and act as a sizingguide.ReShape Calibration Tubes™are indicated for use ingastric and bariatric surgicalprocedures to decompressthe stomach, drain andremove gastric fluid, irrigate,and act as a sizing guide.None
Outer Diameter /French Size nominalB-2032 : 32FB-2036 : 36FB-2040: 40FSameNone
Outer Diameter /French Size distal tipB-2032 : 38FB-2036 : 42FB-2040 : 42FSameNone
Outer Diameter /French Size deflatedballoonB-2032 : 50FB-2036 : 52FB-2040: 56FSameNone
TubingDual lumenSameNone
Distal Side Holes3 aspiration holes proximal to theballoon to ensure a steady vacuumSameNone
Distal TipMolded tip with twelve (12)aspiration eyelets.SameNo impact tosubstantialequivalence.
Connector forSuctionThe catheter includes an adapterfor room suctionSameNone
Balloon + InflationValveThe balloon has an inflationcapacity ≥ 100 cc min.SameNone
Tubing MaterialSiliconeSameNone
MarkingsIndication marks at 30, 35, 40, 45,and 50, 55, and 60 centimeters.SameNone
SterilityNon-sterile, disposable, singlepatient useSameNone
Shelf Life2 yearsSameNone.

{6}------------------------------------------------

VI. PERFORMANCE DATA

Bench testing was not conducted and remains same as the predicate device cleared under K221898.

VII. CONCLUSION

Data presented in this submission demonstrate that the updated DFU is similar and equivalent to of the ReShape Gastrointestinal Boundary Identifier (GIBI HD) Calibration tube (K221898).

The ReShape Calibration Tubes™ device, and the predicate device (Gastrointestinal Boundary Identifier

(GIBI HD) Calibration tube):

  • have the same intended use,
  • · use the same principle of operation,
  • · incorporate the same design,
  • have the same construction and material, and
  • are both provided non-sterile.

In summary, the ReShape Calibration Tubes™ device described in this submission is substantially equivalent to the predicate device, the Gastrointestinal Boundary Identifier (GIBI HD) Calibration tube (K221898).

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.