(31 days)
The G Surgical Marksman Spinal Deformity System is intended to provide immobilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
The G Surgical Marksman Spinal Deformity System consists of longitudinal members (rods), anchors (screws), connectors (crosslinks) and fasteners in a variety of sizes to accommodate differing anatomic requirements.
Unfortunately, I cannot fulfill your request to describe the acceptance criteria and study proving the device meets them, nor any of the other numbered items.
The provided document is a 510(k) clearance letter from the FDA for a medical device called the "G Surgical Marksman Spinal Deformity System." This type of document primarily confirms that the FDA has found the device to be "substantially equivalent" to predicate devices already on the market.
It explicitly states that the device is a "Thoracolumbosacral Pedicle Screw System" and is manufactured from specific titanium and cobalt chrome alloys.
The document details the device's indications for use (spinal immobilization and stabilization as an adjunct to fusion for various conditions), describes its components (rods, screws, connectors, fasteners), and mentions the materials used.
However, the provided text does not contain any information about acceptance criteria, clinical study results, performance data (beyond mechanical testing to ASTM F1717 and sterilization validations), sample sizes for testing, expert qualifications, ground truth establishment, or any details related to AI assistance or multi-reader multi-case studies.
Therefore, I cannot extract the information you've requested from this document. It appears to be a clearance for a physical orthopedic implant, not a diagnostic or AI-driven device that would typically involve the types of performance studies and metrics you've listed.
{0}------------------------------------------------
February 9, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
G Surgical LLC % Karen Warden, Phd President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026
Re: K230063
Trade/Device Name: G Surgical Marksman Spinal Deformity System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: January 6, 2023 Received: January 9, 2023
Dear Karen Warden, Phd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'neill -S 2023.02.09 '00'05- 16:36:42 -
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
G Surgical Marksman Spinal Deformity System
Indications for Use (Describe)
The G Surgical Marksman Spinal Deformity System is intended to provide immobilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
| Type of Use (Select one or both, as applicable) | ||
|---|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
| Date: | 6 January 2023 |
|---|---|
| Sponsor: | G Surgical LLC9433 Bee Cave Road |
| Building 3, Suite 101-AAustin, Texas 78733 USA | |
| Tel.: 512.494.4749 | |
| Sponsor Contact: | Don Grafton, Managing Director |
| 510(k) Contact: | Karen E. Warden, PhD |
| BackRoads Consulting Inc. | |
| PO Box 566 | |
| Chesterland, OH 44026 | |
| Office: 440.729.8457 | |
| Proposed Trade Name: | G Surgical Marksman Spinal Deformity System |
| Common Name: | Posterior pedicle screw system |
| Device Classification: | Class II |
| Requlation Name,Regulation Number,Product Code: | Thoracolumbosacral pedicle screw system, 888.3070, NKB |
| Submission purpose: | The subject 510(k) adds monoaxial screws and instruments to the clearedcomponents. |
| Device Description: | The G Surgical Marksman Spinal Deformity System consists of longitudinalmembers (rods), anchors (screws), connectors (crosslinks) and fasteners ina variety of sizes to accommodate differing anatomic requirements. |
| Indications for Use: | The G Surgical Marksman Spinal Deformity System is intended to provideimmobilization and stabilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion in thetreatment of the following acute and chronic instabilities or deformities:degenerative disc disease (defined as back pain of discogenic origin withdegeneration of the disc confirmed by history and radiographic studies),spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis,curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis,and/or failed previous fusion. |
| Materials: | G Surgical Marksman Spinal Deformity System is manufactured from Ti-6Al-4V ELI titanium alloy (ASTM F136) and Cobalt Chrome (per ASTMF1537). |
| Primary Predicate: | G Surgical Marksman System (G Surgical LLC - K193219) |
| Additional Predicates: | G Surgical Marksman MIS System (G Surgical LLC - K161516), G SurgicalPedicle System (G Surgical LLC - K081041) |
| Performance Data: | ASTM F1717 mechanical testing of worst case G Surgical MarksmanSystem constructs was relied upon in support of the G Surgical MarksmanSpinal Deformity System clearance. The testing included static and dynamiccompression bending and static torsion. An engineering rationale was usedto determine no new worse-case was introduced in addition to the predicatetesting used to support clearance.Sterilization validations were successfully performed to meet AAMI ST79parameters. |
| TechnologicalCharacteristics: | The G Surgical Marksman Spinal Deformity System Is identical to itself asthe primary predicate with respect to the following technologicalcharacteristics: intended use, basic design, material and dimensionalfeatures. |
| Conclusion: | The G Surgical Marksman Spinal Deformity System possesses the sameintended use and technological characteristics as the predicate devices.Therefore G Surgical Marksman Spinal Deformity System is substantiallyequivalent for its intended use. |