K Number
K223823
Manufacturer
Date Cleared
2023-03-27

(96 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This single-use surgical mask EFXPLY3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile.

Device Description

The device is a flat pleated type mask the outer layer being blue and the inner layer being white, utilizing ear loops to secure the mask in place, and a nose bridge is also incorporated for a proper fit around the nose. The device number is EFX3PLYSMSK.

The device is manufactured with three layers, the inner and outer layers are made of spun bond polypropylene, and the middle layer is made of melt-blown polypropylene filter material. The model EFX3PLYSMSK device is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of polypropylene and spandex. The nose bridge contained within the device is placed between the layers of the surgical mask to allow the user to fit the surgical mask properly around their nose. The nose bridge is made up of galvanized with polyethylene.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Disposable Surgical Face Mask, based on the provided document:

Acceptance Criteria and Device Performance

Level 3 Performance

Test MethodologyAccepted Criteria (Level 3)Reported Device Performance (EFX3PLYSMSK)Result
Fluid Resistance ASTM F1862160mm Hg.95 of 96 samples passedPass
Particulate Filtration Efficiency ASTM F2299≥98%99.79% Filtration average efficiency across 96 samplesPass
Bacterial Filtration Efficiency ASTM F2101-19≥98%99.8% Filtration average efficiency across 96 samplesPass
Differential Pressure (Delta P) AMST F2100-19<6.0mm H2O /cm²3.56 mm H2O/cm² average pressure drop across 96 samplesPass
Flammability 16 CFR 1610Class 196 of 96 samples did not ignitePass

Level 2 Performance

Test MethodologyAccepted Criteria (Level 2)Reported Device Performance (EFX3PLYSMSK)Result
Fluid Resistance ASTM F1862120mm Hg.96 of 96 samples passedPass
Particulate Filtration Efficiency ASTM F2299≥98%99.79% Filtration average efficiency across 96 samplesPass
Bacterial Filtration Efficiency ASTM F2101-19≥98%99.8% Filtration average efficiency across 96 samplesPass
Differential Pressure (Delta P) AMST F2100-19<6.0mm H2O /cm²3.56 mm H2O/cm² average pressure drop across 96 samplesPass
Flammability 16 CFR 1610Class 196 of 96 samples did not ignitePass

Biocompatibility

Test ItemAccepted CriteriaReported Device PerformanceResult
CytotoxicityNo potential cytotoxicityUnder the conditions of the study, the device extract did not show potential toxicity to L 929 cellsPass
IrritationNon-irritatingUnder the condition of the study, the device is non-irritatingPass
SensitizationNon-sensitizingUnder the conditions of the study, the device is non-sensitizingPass

Study Information

Due to the nature of the device (Disposable Surgical Face Mask) and the provided document, much of the requested information regarding AI/software studies is not applicable. The document describes a non-clinical performance and biocompatibility study for a physical medical device, not an AI/software-based medical device.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: For performance tests (Fluid Resistance, Particulate Filtration, Bacterial Filtration, Differential Pressure, Flammability), 96 samples of the device (EFX3PLYSMSK) were used for each test. For biocompatibility tests, the sample size is not explicitly stated in numerical terms (e.g., number of test articles), but the results indicate testing was performed.
    • Data Provenance: The document does not specify the country of origin of the data directly, nor does it explicitly state whether the study was retrospective or prospective. Given these are physical performance tests, it's generally implied to be a prospective laboratory study on newly manufactured samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable to this type of study. "Ground truth" established by experts typically pertains to diagnostic accuracy studies involving image interpretation or clinical assessments, which are not relevant to the physical performance testing of a surgical mask. The acceptance criteria themselves serve as the 'ground truth' for performance.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This question is not applicable to this type of study. Adjudication methods are relevant for studies involving human interpretation or subjective assessments, not for objective physical performance testing against predefined criteria.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is specifically designed for evaluating diagnostic accuracy of AI systems, often in comparison to human reader performance, which is not relevant for a surgical mask.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. As this is a physical medical device (surgical mask), there is no AI algorithm involved.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this study is based on established regulatory standards and test methodologies (e.g., ASTM F1862, ASTM F2299, ASTM F2101-19, AMST F2100-19, 16 CFR 1610, ISO 10993-5, ISO 10993-10). The performance of the mask is measured against quantitative and qualitative acceptance criteria specified by these standards for surgical masks.
  7. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/machine learning device. The samples used for testing are production samples of the device.
  8. How the ground truth for the training set was established:

    • Not applicable. As there is no training set, there is no ground truth establishment for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

March 27, 2023

Efofex, Inc. % Timothy Kania Official Correspondent mdi Consultants. Inc. 55 Northern Blvd, Suite 200 Great Neck. New York 11021

Re: K223823

Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 20, 2022 Received: December 21, 2022

Dear Timothy Kania:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K223823

Device Name Disposable Surgical Mask

Indications for Use (Describe)

This single-use surgical mask EFXPLY3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

Contact Details

K223823

A. Applicant Details

Applicants NameEfofex, Inc.
Applicants Address1300 N Dixieland Road, Rogers, AR 72758
Contact PersonSiddhant Doshi
Phone Number917-213-2740
E-mailsid@efofexinc.com
Date Prepared12/14/22

Official Correspondent

Contact PersonTim Kania
Company Namemdi Consultants, Inc.
Address55 Northern Blvd. Suite 200Great Neck, NY 11021
Phone516-482-9001
EmailTim@mdiconsultants.com

B. Device Information

Trade NameDisposable Surgical Mask
Common NameDisposable Surgical Mask
ModelEFX3PLYSMSK
ClassificationII
Classification NameMask, Surgical
Product CodeFXX
Regulation No.21 CFR 878.4040

C. Legally Marketed Primary Predicate Device

Trade NameSurgical Face Mask, Model GFYY95
510 (K)K202761
Product CodeFXX
ManufacturerGuangdong Good Feeling Hygiene Material Tec Co., LTD.

{4}------------------------------------------------

D. Indications for use of the device

This single-use surgical mask EFXPL Y3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile.

E. Device description

The device is a flat pleated type mask the outer layer being blue and the inner layer being white, utilizing ear loops to secure the mask in place, and a nose bridge is also incorporated for a proper fit around the nose. The device number is EFX3PLYSMSK.

The device is manufactured with three layers, the inner and outer layers are made of spun bond polypropylene, and the middle layer is made of melt-blown polypropylene filter material. The model EFX3PLYSMSK device is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of polypropylene and spandex. The nose bridge contained within the device is placed between the layers of the surgical mask to allow the user to fit the surgical mask properly around their nose. The nose bridge is made up of galvanized with polyethylene.

{5}------------------------------------------------

DEVICEPROPOSED DEVICEPRIMARY PREDICATE DEVICECOMPARISON
ManufacturerEfofex, Inc1300 N Dixieland RoadRogers AR 72758Guangdong Good FeelingHygiene Material Tec Co. LTD.10 Phoenix Av. Eco-IndustrialPk. Guangdong Province, ChinaNA
510KK223823K202761NA
ClassificationClass II Device, FXX(21CFR878.4040)Class II Device, FXX(21CFR878.4040)Similar
Indications foruseThis single use surgical maskEFX3PLY3SMSK is intendedto be worn to protect boththe patient and health carepersonnel from transfer ofmicroorganisms, body fluidsand particulate material.These face masks areintended for use in infectioncontrol practices to reducethe potential exposure toblood and body fluids. This isa single use, disposal device,provided non-sterile.This single use surgical maskGFYY95 is intended to be wornto protect both the patient andhealth care personnel fromtransfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infectioncontrol practices to reduce thepotential exposure to bloodand body fluids. This is a singleuse, disposal device, providednon-sterile.Similar
DescriptionEar loops, flat pleated, threelayersEar loops, flat pleated, threelayersSimilar
Outer layerSpun-bond polypropyleneSpun-bond polypropyleneSimilar
Middle layerMelt-blown polypropylenefilterMelt-blown polypropylenefilterSimilar
Inner layerSpun- bound polypropyleneSpun- bound polypropyleneSimilar
Nose bridgeGalvanized wire withpolyethylene coatingGalvanized wire withpolyethylene coatingSimilar
Ear loopsPolypropylene & SpandexNylon & SpandexDifferent¹
ColorBlueBlueSimilar
Length175 mm180mmDifferent²
Width95 mm95mmSimilar
OTC useYesYesSimilar
SterilityNoNoSimilar
UseSingle use, DisposableSingle use, DisposableSimilar
ASTM F2100levelLevel 2 & 3Level 3Similar3
BiocompatibilityISO10993ISO10993Similar
  • F. Technological Characteristics Comparison with primary predicate device

{6}------------------------------------------------

Discussion of Similarities and Differences

1: Differences in Ear Loop material have no effect on the safety or efficacy of the device. The subject device was tested for Biocompatibility and passed. Please see table 3 below for a summary of the Biocompatibility testing.

2: Differences in Dimensions have no effect on the safety or efficacy of the device. Please see table 2 below for a summary of the performance testing performed on this device.

3: The Level 2 and Level 3 masks are comprised of the same material which has passed performance testing for Level 3 protection. This has no effect on the safety or performance of the device.

G. Nonclinical test conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device meets the acceptance criteria for the following standards and the requirements stated in the guidance for industry and FDA staff; surgical mask-premarket notification ( 510 K ).

ISO 10993-5: 2009 Biological evaluation of medical devices - part 5: Test for In vitro cytotoxicity

ISO 10993-10: 2010 Biological Evaluation of Medical Devices – part 10: Test for irritation and skin sensitization

ASTM F2100, Standard Specification for performance of Materials Used in Face Masks.

ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood ( Horizontal Projection of Fixed Volume at A Known Velocity);

EN 14683, Medical Face Masks -- Requirements and Test Methods;

ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials Using a Biological Aerosol of Staphylococcus Aureus;

ASTM F2299, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

16 CFR 1610, Standard for the Flammability of clothing textiles;

{7}------------------------------------------------

Table 2 – Performance Testing Level 3
Test MethodologyProposed DeviceEFX3PLYSMSKAccepted CriteriaLevel 3Result
Fluid ResistancePerformance ASTMF186295 of 96 samplespassed160mm Hg.Pass
Particulate FiltrationEfficiencyASTM F229999.79% Filtrationaverage efficiencyacross 96 samples≥98%Pass
Bacterial FiltrationEfficiencyASTM F2101-1999.8% Filtrationaverage efficiencyacross 96 samples≥98%Pass
Differential Pressure(Delta P) AMSTF2100-193.56 mm H2O/cm²average pressure dropacross 96 samples<6.0mm H2O /cm²Pass
Flammability16 CFR 161096 of 96 samples didnot igniteClass 1Pass
Table 3 – Performance Testing Level 2
Test MethodologyProposed DeviceEFX3PLYSMSKAccepted CriteriaLevel 2Result
Fluid ResistancePerformance ASTMF186296 of 96 samplespassed120mm Hg.Pass
Particulate FiltrationEfficiencyASTM F229999.79% Filtrationaverage efficiencyacross 96 samples≥98%Pass
Bacterial FiltrationEfficiencyASTM F2101-1999.8% Filtrationaverage efficiencyacross 96 samples≥98%Pass
Differential Pressure(Delta P) AMSTF2100-193.56 mm H2O/cm²average pressure dropacross 96 samples<6.0mm H2O /cm²Pass
Flammability16 CFR 161096 of 96 samples didnot igniteClass 1Pass

{8}------------------------------------------------

Table 4 – Biocompatibility Testing
ItemProposed DeviceAccepted CriteriaResult
CytotoxicityUnder the conditionsof the study, thesubject device extractdid not showpotential toxicity to L929 cellsNo potentialcytotoxicityPass
IrritationUnder the conditionof the study, thedevice is non-irritatingNon-irritatingPass
SensitizationUnder the conditionsof the study, thedevice is nonsensitizingNon-sensitizingPass
  • H. Clinical Test Conclusions No clinical tests is included in this submission
  • l. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, is effective, and performs as well or better than the legally marketed primary predicate device number K202761, manufactured by Guangdong Good Feeling Hygiene Material Tec Co. LTD, single use surgical mask.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.